NI VirtualBench VB-8012 Specifications

SPECIFICATIONS
NI VB-8012
NI VirtualBench™ All-In-One Instrument
Français
Deutsch
日本語
한국어
简体中文
ni.com/manuals
These specifications are for the National Instruments VirtualBench VB-8012 only. These
specifications are valid following 30 minutes of warmup and are typical at 25 °C unless
otherwise noted. For feature differences between the VirtualBench application for Windows
and iPad, go to ni.com/info and enter vbfeatures.
Mixed Signal Oscilloscope
Analog Channels
Vertical System
............................................................................
Number of channels
2 single-ended, non-isolated
Bandwidth (-3 dB)1
............................................................................
100 MHz
Resolution
............................................................................
8 bits
2
............................................................................
Accuracy
±2% of input, ±1% full scale (Vpk-pk)
............................................................................
Input coupling
DC, AC
1
2
Bandwidth using the accessory oscilloscope probe in 10X mode.
Indicates warranted specifications valid at Tcal ±5 °C. Temperature coefficients are
calculated using the temperature change from last external calibration.
Vertical sensitivity (range)
10 mV/div (100 mVpk-pk)
20 mV/div (200 mVpk-pk)
50 mV/div (400 mVpk-pk)
100 mV/div (1 Vpk-pk)
200 mV/div (2 Vpk-pk)
500 mV/div (4 Vpk-pk)
1 V/div (10 Vpk-pk)
2 V/div (20 Vpk-pk)
5 V/div (40 Vpk-pk)
Input
............................................................................
impedance
1 MΩ || 20 pF
Table 1. DC Offset Range
Range
Programmable Offset Range
10 mV/div, 20 mV/div, 50 mV/div
±5 V
100 mV/div, 200 mV/div, 500 mV/div,
±20 V
1 V/div, 2 V/div, 5 V/div
Acquisition modes
............................................................................
Sample, peak detect, averaging
Horizontal System
Maximum
............................................................................
sample rate
1 GS/s single channel,
500 MS/s/channel, dual channel
Maximum record length
............................................................................
1 MS/channel
2 |
ni.com
|
NI VirtualBench VB-8012 Specifications
Digital Channels/Logic Analyzer
Vertical System
Number
............................................................................
of channels
34
Maximum input frequency
............................................................................
100 MHz
Input voltage
............................................................................
0 to 5 V
Note Mixed signal oscilloscope digital channels are designed to withstand
accidental overvoltage from signals on the VB-8012 or similar devices. They are not
recommended for use with signals likely to exceed 0 to 5 V in normal operation.
............................................................................
Input threshold
Programmable, 0 to 2.0 V3
Threshold accuracy
............................................................................
350 mV
Additional/internal channels
............................................................................
Digital I/O lines, function generator start,
external trigger (TRIG), power line frequency
Horizontal System
............................................................................
Timing mode sample rate2
1 GS/s (down to ~15 kS/s)
............................................................................
Maximum external sample clock rate
100 MHz
Record length
Typical
....................................................................
1 MS
Minimum4
....................................................................
4 kS
Decimation
............................................................................External Sample Clock, 1:1, 2:1, and
n*4:1 where n is an integer
Maximum sample compression
............................................................................
215 to 1
3
4
Programmable input threshold is only available with the NI VirtualBench driver. The VirtualBench
application has an input threshold of 1.65 V.
Under most conditions, the logic analyzer can acquire 1 MS of data. Under some
conditions with very high sustained activity on multiple inputs, the logic analyzer may
only capture 4 kS of data.
NI VirtualBench VB-8012 Specifications
|
© National Instruments
|
3
Triggering
............................................................................
Trigger modes
Normal, Auto, Single, Force
Trigger sources
............................................................................
Oscilloscope analog channels, oscilloscope
digital channels, function generator start,
digital I/O lines, external trigger (TRIG),
power line frequency
Trigger types
Analog
....................................................................
Edge with hysteresis
Digital
....................................................................
Edge, glitch5, level, pattern
Trigger resolution
1....................................................................
GS/s mode
1 ns
500 MS/s mode
....................................................................
2 ns
Trigger export
....................................................................
Available through external trigger (TRIG)
Waveform Measurements
............................................................................
Oscilloscope time5
Period, frequency, positive duty cycle,
negative duty cycle, positive pulse width,
negative pulse width, rise time, fall time,
rise rate, fall rate
............................................................................
Oscilloscope voltage6
High, low, amplitude, maximum, minimum,
peak-to-peak, overshoot, undershoot, RMS,
mean, cycle RMS, cycle mean
Logic
............................................................................
analyzer time6
Period, frequency, positive duty cycle,
negative duty cycle, positive pulse width,
negative pulse width
Waveform Math
7
Operations
............................................................................
A + B, A - B, A * B, FFT
5
6
7
4 |
Glitch triggers are only available with the NI VirtualBench driver.
Waveform measurements are only available in the VirtualBench application.
Waveform math is only available in the VirtualBench application.
ni.com |
NI VirtualBench VB-8012 Specifications
Function Generator (FGEN)
............................................................................
Waveforms
Sine, square, ramp/triangle, DC, arbitrary8
............................................................................
Update rate
125 MS/s
Resolution
............................................................................
14 bits
Number of channels
............................................................................
1
Output impedance
............................................................................
50 Ω
Switchable filter9
............................................................................
36 MHz lowpass, 7-pole, elliptical
Sine
....................................................................
Maximum frequency
20 MHz
Total harmonic distortion
1 MHz
............................................................
-55 dBc
10 MHz
............................................................
-50 dBc
Spurious free dynamic range
....................................................................
-70 dB at 1 MHz (non-harmonic)
Phase noise (1 MHz)
....................................................................
-115 dBc/Hz at 10 kHz offset
Square
Maximum frequency
....................................................................
5 MHz
Rise/fall time
....................................................................
<20 ns (10% to 90%)
Overshoot
....................................................................
<5%
Jitter
....................................................................
8 ns cycle-to-cycle
Ramp/triangle maximum frequency
............................................................................
150 kHz
Accuracy (with >10 kΩ load)
Amplitude (1 kHz sine)
....................................................................
±(1% of setting + 5 mV)
DC
....................................................................
±(1% of setting + 5 mV)
Output range
50 Ω
....................................................................
±6 V
Hi-Z (>10 kΩ)
....................................................................
±12 V
DC offset
50 Ω
....................................................................
±6 V
Hi-Z (>10 kΩ)
....................................................................
±12 V
Note The combination of signal amplitude and DC offset cannot exceed the output
range specifications.
8
9
Arbitrary waveforms are only available with the NI VirtualBench driver.
Switchable filters are only available with the NI VirtualBench driver. The VirtualBench
application automatically enables the lowpass filter in sine mode.
NI VirtualBench VB-8012 Specifications |
© National Instruments
|
5
Frequency
Accuracy
....................................................................
≤100 ppm
Resolution
....................................................................
1 μHz
Arbitrary waveform
Points
....................................................................
1 MS
Sample rate
....................................................................
125 MS/s
Flatness
............................................................................
±0.3 dB to 20 MHz
Protection
............................................................................
Short-circuit protected
Triggering
Trigger
............................................................................
types
Start of buffer10
Trigger
............................................................................
resolution
8 ns
Trigger export
............................................................................
Available through external trigger (TRIG)
Digital I/O
Number of channels
............................................................................
8
Direction control
............................................................................
Input or output, software-selectable
Logic level
............................................................................
5 V compatible LVTTL input,
3.3 V LVTTL output
Drive
............................................................................
strength
4 mA
Input voltage
............................................................................
0 to 5 V
Note Digital I/O lines are designed to withstand accidental overvoltage from
signals on the VB-8012 or similar devices. They are not recommended for use with
signals likely to exceed 0 to 5 V in normal operation.
............................................................................
DIO channel pull resistors
10 kΩ, pull-down
External Power
3.3 V output
....................................................................
Voltage
3.3 V ±10%
Current
....................................................................
20 mA
10
The function generator can only produce a trigger.
6 | ni.com |
NI VirtualBench VB-8012 Specifications
Digital Multimeter
............................................................................
Functions
DC voltage, AC voltage, DC current,
AC current, resistance, diode, continuity
Resolution
............................................................................
5½ digits
Sample rate
............................................................................
5 S/s
Caution Do not use this device for connection to signals or for measurements
within Measurement Categories III or IV. For more information about Measurement
Categories, refer to the Safety Voltages section.
Input protection
Resistance,
....................................................................
diode
Up to 300 V DC
DC and AC voltage
....................................................................
Up to 300 V DC or 265 V ACrms,
400 V AC peak
DC and AC current
............................................................
DMM A current
Internal ceramic fuse, 10 A 250 V, time-delay,
5 × 20 mm, T 10A H 250V
connector fuse
(Cooper Bussmann part number S505H-10-R
at www.cooperbussmann.com)
DMM mA current
............................................................
Internal ceramic fuse, 1.25 A 250 V,
time-delay, 5 × 20 mm, T 1.25A H 250V
connector fuse
(Cooper Bussmann part number S505H-1.25-R
at www.cooperbussmann.com)
Caution Fuses are located on bottom of device underneath door. Use Phillips #1
screwdriver for removal. Ensure all hazardous voltages are disconnected from the
device prior to removal of door.
Fuse When this fuse symbol is marked on a device, take proper precautions.
Maximum
............................................................................
common-mode voltage
300 V DC or ACrms
NI VirtualBench VB-8012 Specifications |
© National Instruments
|
7
DC
Table 2. DC Voltage Accuracy
Range
Input Impedance
1-Year Accuracy2
± (% of Reading +
% of Range)
Temperature Coefficient2
± (% of Reading +
% of Range)/°C
100 mV* >10 GΩ, 10 MΩ
0.015 + 0.005
0.001 + 0.0005
1V
>10 GΩ, 10 MΩ
0.015 + 0.005
0.001 + 0.0005
10 V
>10 GΩ, 10 MΩ
0.015 + 0.005
0.001 + 0.0005
100 V
10 MΩ
0.035 + 0.005
0.005 + 0.0005
300 V
10 MΩ
0.035 + 0.005
0.005 + 0.0005
* Add 15 µV if not immediately following offset null.
Table 3. DC Current Accuracy
Range
Burden Voltage
1-Year Accuracy2
± (% of Reading +
% of Range)
Temperature Coefficient2
± (% of Reading +
% of Range)/°C
10 mA
<0.03 V
0.070 + 0.020
0.0035 + 0.0010
100 mA
<0.3 V
0.070 + 0.003
0.0020 + 0.0010
1A
<0.03 V
0.130 + 0.025
0.0065 + 0.0010
10 A*
<0.3 V
0.130 + 0.004
0.0045 + 0.0010
* 30 seconds on, 30 seconds off. Add 300 ppm/A for currents >2.2 A. After measuring >5 A,
wait two minutes to get full accuracy in the 1 A range.
Table 4. DC Resistance Accuracy (2-Wire)*, 1 V Open Circuit Voltage
Range
Short-Circuit
Current
1-Year Accuracy2
± (% of Reading +
% of Range)
Temperature Coefficient2
± (% of Reading +
% of Range)/°C
100 Ω
170 μA
0.018 + 0.050
0.0010 + 0.0005
1 kΩ
170 μA
0.018 + 0.005
0.0010 + 0.0005
10 kΩ
70 μA
0.018 + 0.005
0.0010 + 0.0005
8 | ni.com
|
NI VirtualBench VB-8012 Specifications
Table 4. DC Resistance Accuracy (2-Wire)*, 1 V Open Circuit Voltage (Continued)
Range
Short-Circuit
Current
1-Year Accuracy2
± (% of Reading +
% of Range)
Temperature Coefficient2
± (% of Reading +
% of Range)/°C
100 kΩ
10 μA
0.018 + 0.005
0.0010 + 0.0005
1 MΩ
1.1 μA
0.035 + 0.005
0.0040 + 0.0005
10 MΩ
1.1 μA
0.085 + 0.005
0.0100 + 0.0005
1.3 + 0.005
0.1000 + 0.0005
100 MΩ 1.1 μA
* Perform offset nulling.
Caution The input terminals of the DMM are not protected for electromagnetic
interference. As a result, the DMM may experience reduced measurement accuracy
or other temporary performance degradation when connected to unshielded test leads
in an environment with radiated or conducted radio frequency electromagnetic
interference.
DC
............................................................................
continuity accuracy range11
100 Ω
............................................................................
DC diode test range
2V
............................................................................
Effective
Common-Mode Rejection
>100 dB
Ratio (CMRR), 1 kΩ
resistance in LO lead
Normal-Mode
............................................................................
Rejection Ratio
>100 dB
(NMRR), 50/60 Hz ±0.1%
Overrange
............................................................................
105% of range except 300 V
11
DC continuity is only available in the VirtualBench application.
NI VirtualBench VB-8012 Specifications |
© National Instruments
|
9
AC
Table 5. AC Voltage Accuracy
Range (rms)
100 mV, 1 V,
10 V, 100 V,
265 V
Peak Voltage
Frequency
1-Year Accuracy2
± (% of Reading +
% of Range)
Temperature
Coefficient2
± (% of Reading +
% of Range)/°C
±210 mV,
20 Hz to
±2.1 V, ±21 V, 45 Hz
±210 V, ±400 V
45 Hz to
65 Hz
0.91 + 0.10
0.01 + 0.005
0.30 + 0.05
0.01 + 0.005
65 Hz to
1 kHz
0.21 + 0.05
0.01 + 0.005
1 kHz to
5 kHz
0.12 + 0.05
0.01 + 0.005
5 kHz to
20 kHz
0.35 + 0.05
0.01 + 0.005
Table 6. AC Current Accuracy
Range
(rms)
5 mA
50 mA
10 |
ni.com
Peak
Current
Burden
Voltage
(rms)
±10.5 mA <0.02 V
±105 mA
|
<0.2 V
1-Year Accuracy2
± (% of Reading +
% of Range)
Temperature
Coefficient2
± (% of Reading +
% of Range)/°C
20 Hz to
1 kHz
0.20 + 0.01
0.01 + 0.005
1 kHz to
5 kHz
0.60 + 0.01
20 Hz to
1 kHz
0.20 + 0.01
1 kHz to
5 kHz
0.50 + 0.01
Frequency
NI VirtualBench VB-8012 Specifications
0.01 + 0.005
Table 6. AC Current Accuracy (Continued)
Range
(rms)
500 mA
5A
Peak
Current
±1.05 A
±10.5 A
Burden
Voltage
(rms)
<0.02 V
<0.2 V
1-Year Accuracy2
± (% of Reading +
% of Range)
Temperature
Coefficient2
± (% of Reading +
% of Range)/°C
20 Hz to
1 kHz
0.15 + 0.01
0.01 + 0.005
1 kHz to
5 kHz
0.50 + 0.01
20 Hz to
1 kHz
0.25 + 0.03
1 kHz to
5 kHz
0.60 + 0.03
Frequency
0.01 + 0.005
Caution The input terminals of the DMM are not protected for electromagnetic
interference. As a result, the DMM may experience reduced measurement accuracy
or other temporary performance degradation when connected to unshielded test leads
in an environment with radiated or conducted radio frequency electromagnetic
interference.
Input
............................................................................
impedance
10 MΩ || 200 pF
CMRR, 1 kΩ resistance in LO lead
............................................................................
>70 dB (DC to 60 Hz)
DC Power Supply
............................................................................
Outputs
0 to +6 V/0 to 1 A,
0 to +25 V/0 to 500 mA (isolated),
0 to -25 V/0 to 500 mA (isolated)
Note The +25 V and -25 V channels are bank isolated from ground but not from
each other.
NI VirtualBench VB-8012 Specifications |
© National Instruments
|
11
Table 7. DC Accuracy/Resolution
Output
DC output2
Type
+6 V
+25 V
-25 V
Voltage
0 to +6 V
0 to +25 V
0 to -25 V
Current12
1A
500 mA
500 mA
Programming accuracy2,12
± (% of reading + offset)
Voltage
0.1% + 5 mV
0.1% + 20 mV 0.1% + 20 mV
Current
0.2% + 10 mA
0.15% + 4 mA 0.15% + 4 mA
Readback accuracy2,13
± (% of reading + offset)
Voltage
0.1% + 5 mV
0.1% + 20 mV 0.1% + 20 mV
Current
0.2% + 10 mA
0.15% + 4 mA 0.15% + 4 mA
Programming resolution
Voltage
1.7 mV
6.5 mV
6.5 mV
Current
0.30 mA
0.15 mA
0.15 mA
Voltage
0.41 mV
1.7 mV
1.7 mV
Current
70 µA
35 µA
35 µA
Voltage
0.01% + 25 mV 0.03% + 5 mV 0.03% + 5 mV
Readback resolution
Load regulation
± (% of reading + offset)14
Overvoltage protection
............................................................................
30 V (25 V channels) and 10 V (6 V channel)
Reverse voltage protection
............................................................................
400 mV
External Trigger (TRIG)
............................................................................
Direction control
Input or output, software-selectable
Logic level
............................................................................
5 V compatible LVTTL input,
3.3 V LVTTL output
Drive
............................................................................
strength
4 mA
Input voltage
............................................................................
0 to 5 V
Note The external trigger line is designed to withstand accidental overvoltage from
signals on the VB-8012 or similar devices. It is not recommended for use with
signals likely to exceed 0 to 5 V in normal operation.
12
13
14
12 |
Minimum programmable current limit is 1% of range.
Programming and readback accuracy specified at no load.
Change in output voltage for any load within range.
ni.com
|
NI VirtualBench VB-8012 Specifications
Connectivity
Wired Interface
USB specification
............................................................................
USB 2.0 Hi-Speed
Wireless Interface
Table 8. Network Protocols and Ports Used
Port
Protocol
Function
Port 80/TCP
HTTP
Device configuration (web, MAX)
Port 443/TCP
HTTP
Device configuration (web, MAX)
Port 3580/TCP
Service locator
Device configuration (web, MAX)
Port 9090/TCP
Configuration only
VirtualBench instrument protocol
Port 5353/UDP
Multicast DNS
Device discovery
Network IP configuration
............................................................................
IPv4, DHCP Client/Server
Radio mode
............................................................................
IEEE 802.11 b,g,n
Wireless modes
............................................................................
AP mode (default), client mode
Frequency band
............................................................................
2.4 GHz ISM
Channel width
............................................................................
20 MHz
Channels
............................................................................
USA 1-11, International 1-13
(12 and 13 client mode only)
TX power
............................................................................
+10 dBm max (10 mW)
Security
............................................................................
Open, WPA, WPA2, WPA2-Enterprise
Enterprise
............................................................................
security EAP types
EAP-TLS, EAP-TTLS/MS-CHAPv2,
PEAPv0/MS-CHAPv2
Antenna
............................................................................External RP-SMA omnidirectional dipole
NI VirtualBench VB-8012 Specifications
|
© National Instruments
|
13
Power Requirements
Caution The protection provided by the VirtualBench hardware can be impaired if
it is used in a manner not described in the NI VB-8012 Safety, Environmental, and
Regulatory Information document.
............................................................................
Voltage input range
100 to 240 VAC, 50/60 Hz
Power consumption
............................................................................
100 W maximum
Power input connector
............................................................................
IEC C13 power connector
Power disconnect
............................................................................
The AC power cable provides main power
disconnect. Do not position the equipment so
that it is difficult to disconnect the power
cable. Depressing the front panel power button
does not inhibit the internal power supply.
Calibration
............................................................................
Calibration cycle (digital multimeter,
1 year
mixed signal oscilloscope,
function generator,
DC power supply)
Specified
............................................................................
temperature
Tcal ± 5 °C
............................................................................
Warmup time
30 minutes
14 |
ni.com
|
NI VirtualBench VB-8012 Specifications
Physical Characteristics
Dimensions
Enclosure
....................................................................25.40 cm × 19.05 cm × 7.39 cm
(10.00 in. × 7.50 in. × 2.91 in.)
Enclosure with connectors and
....................................................................
25.40 cm × 23.37 cm × 13.97 cm
(10.00 in. × 9.20 in. × 5.50 in.)
antenna
Note Use the VirtualBench instrument in a horizontal orientation. Allow at least
10.16 cm (4.0 in.) of clearance in front and behind the VirtualBench instrument for
USB, power, and common connector cabling.
............................................................................
Weight
2.05 kg (4 lb 8.3 oz)
Connectivity
Mixed signal oscilloscope
....................................................................
BNC
Logic analyzer
....................................................................
2x20 shrouded IDC header
External trigger
....................................................................
BNC
Function generator
....................................................................
BNC
Digital I/O
Type
............................................................
Pluggable screw terminal, 3.5 mm
(14 position)
Screw terminal wiring
............................................................
0.1 to 2.0 mm2 (30 to 14 AWG)
Torque
............................................................0.25 N · m (2.2 lb · in.)
Digital multimeter
....................................................................
4 mm banana jacks
DC power supply
Type
............................................................
Pluggable screw terminal, 3.81 mm
(6 position)
Screw terminal wiring
............................................................
0.1 to 2.0 mm2 (30 to 14 AWG)
Torque
............................................................0.25 N · m (2.2 lb · in.)
Security cable slot
............................................................................
1, complies with Kensington security slot
dimensions
If you need to clean the device, wipe it with a dry towel.
Safety Voltages
Connect only voltages that are within these limits.
NI VirtualBench VB-8012 Specifications |
© National Instruments
|
15
DMM Isolation Voltages
Hazardous Voltage This icon denotes a warning advising you to take precautions
to avoid electrical shock.
Channel-to-earth ground
Continuous
....................................................................
300 V, Measurement Category II
Withstand
....................................................................
3,000 Vrms, verified by a 5 s dielectric
withstand test
Measurement Category II is for measurements performed on circuits directly connected to the
electrical distribution system. This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.
Caution Do not connect the VirtualBench hardware to signals or use for
measurements within Measurement Categories III or IV.
DC Power Supply Isolation Voltages
............................................................................
+25 V and -25 V-to-earth ground
60 VDC, Measurement Category I
continuous
Note Measurement Categories CAT I and CAT O (Other) are equivalent. The input
circuits are not intended for direct connection to the MAINs building installations of
Categories CAT II, CAT III, or CAT IV.
Environmental
Operating
............................................................................
temperature
0 to 40 °C
Storage temperature
............................................................................
-20 to 70 °C
Operating humidity
............................................................................
10 to 90% RH, noncondensing
DMM full accuracy at 10 to 80%
Storage
............................................................................
humidity
5 to 95% RH, noncondensing
Cooling
............................................................................
Forced air circulation (positive pressurization)
through a fan. Fan speed automatically adjusts
according to operating conditions. Intake and
exhaust locations are on rear of device. Ensure
that the intake and exhaust locations are not
obstructed.
16 |
ni.com
|
NI VirtualBench VB-8012 Specifications
Pollution Degree
............................................................................
2
Maximum altitude
............................................................................
2,000 m
Indoor use only.
Shock and Vibration
............................................................................
Operational shock
30 g peak, half-sine, 11 ms pulse (Tested
in accordance with IEC 60068-2-27. Test
profile developed in accordance with
MIL-PRF-28800F.)
Random vibration
Operating
....................................................................
5 to 500 Hz, 0.3 grms
Nonoperating
....................................................................
5 to 500 Hz, 2.4 grms (Tested in accordance
with IEC-60068-2-64. Nonoperating test
profile exceeds the requirements of
MIL-PRF-28800F, Class 3.)
Safety
This product meets the requirements of the following standards of safety for electrical
equipment for measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the Online
Product Certification section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for radio equipment and
for telecommunication terminal equipment:
•
EN 61000-6-4 (IEC 61000-6-4): Emissions
•
EN 61000-6-2 (IEC 61000-6-2): Immunity
•
EN 55022 (CISPR 22): Class A emissions
•
EN 55024 (CISPR 24): Immunity
•
AS/NZS CISPR 22: Class A emissions
NI VirtualBench VB-8012 Specifications
|
© National Instruments
|
17
•
•
FCC 47 CFR Part 15B: Class A emissions
ICES-001: Class A emissions
Note For EMC declarations, certifications, and additional information, refer to the
Online Product Certification section.
CE Compliance
This product meets the essential requirements of applicable European Directives, as follows:
•
2006/95/EC; Low-Voltage Directive (safety)
•
2004/108/EC; Electromagnetic Compatibility Directive (EMC)
•
1995/5/EC; Radio and Telecommunications Terminal Equipment (R&TTE) Directive
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance
information. To obtain product certifications and the DoC for this product, visit ni.com/
certification, search by model number or product line, and click the appropriate link in the
Certification column.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible
manner. NI recognizes that eliminating certain hazardous substances from our products is
beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact
web page at ni.com/environment. This page contains the environmental regulations and
directives with which NI complies, as well as other environmental information not included in
this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of the product life cycle, all NI products must be
disposed of according to local laws and regulations. For more information about
how to recycle NI products in your region, visit ni.com/environment/weee.
电子信息产品污染控制管理办法(中国 RoHS)
中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物
质指令(RoHS)。关于 National Instruments 中国 RoHS 合规性信息,请登录
ni.com/environment/rohs_china。(For information about China RoHS
compliance, go to ni.com/environment/rohs_china.)
18 |
ni.com
|
NI VirtualBench VB-8012 Specifications
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on National Instruments trademarks.
Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents
covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the
patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about
end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export
Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and
how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES
AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS.
U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the
applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2014 National Instruments. All rights reserved.
371527C-01
Sep14