a p p l i c at i o n N o t e Gas Chromatography/ Mass Spectrometry Authors Ruben Garnica Dawn May PerkinElmer, Inc. Shelton, CT USA Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8 Introduction U.S. EPA Method 8260C – Volatile Organic Compounds (VOCs) by Gas Chromatography Mass Spectrometry (GC/MS) is one of the most common environmental applications for GC/MS. This method outlines the analysis of volatile organic compounds in a variety of solid waste matrices including various air sampling trapping media, ground and surface water, soils, and sediments among others. The method requires not only demonstration of laboratory sample preparation and handling competence but instrument performance as well. The study presented here demonstrates the PerkinElmer® Clarus® SQ 8 GC/MS with purge and trap sample introduction both meets and exceeds the performance criteria set out in method 8260C and describes the analytical results and instrumental methodology. Experimental The PerkinElmer Clarus SQ 8C GC/MS operating in electron ionization mode with an Atomx purge and trap sample introduction system (Teledyne Tekmar, Mason, OH) was used to perform these experiments. The purge and trap conditions are presented in Table 1 and represent standard conditions for the analysis of method of VOCs by EPA Method 8260C. Table 1. Purge and Trap Instrument Conditions. Purge and Trap System: Tekmar – Atomx Trap Tekmar #9 trap Sample Size 5 mL Purge Parameters: Valve Oven Temp 140 °C Transfer Line Temp 140 °C Table 2. Clarus SQ 8C GC/MS conditions. Sample Mount Temp 90 °C Gas Chromatograph: PerkinElmer Clarus 680 Water Heater Temp 90 °C Sample Vial Temp 20 °C Analytical Column: Sample Equilibrate Time 0.00 min Injector Temperature: 220 °C Standby Flow 10 mL/min Carrier Gas: Helium @ 1.0 mL/min Purge Ready Temp 40 °C Split Flow: 80 mL/min Condensate Ready Temp 45 °C Oven Program: Temperature Hold Time Presweep Time 0.25 min Prime Sample Fill Volume 3.0 mL Sweep Sample Time Rate 2 min 10 °C/min 100 °C 0 min 30 °C/min 0.25 min 200 °C 4 min End Sweep Sample Flow 100 mL/min Mass Spectrometer: PerkinElmer Clarus SQ 8C Sparge Vessel Heater On Purge Time 11.00 min GC Transfer Line Temperature: Purge Flow 40 mL/min Purge Temp 40 °C Condensate Purge Temp 20 °C Dry Purge Time 2.00 min Dry Purge Flow 100 mL/min Dry Purge Temp 20 °C Water Needle Rinse Volume 7.0 mL Sweep Needle Time 0.50 min Desorb Preheat Temp 200 °C GC Start Signal Start of Desorb Desorb Time 2.00 min Drain Flow 300 mL/min Desorb Temp 200 °C Bake Parameters: 2 Elite 624 MS (30 m x 0.25 mm i.d. x 1.4 μm) 40 °C Desorb Parameters: The Clarus SQ 8C GC/MS conditions are presented in Table 2. The heated transfer line of the concentrator was connected directly to the injector port of the Clarus 680 GC. A 2 mm i.d. injector port liner was used inside the capillary injector. Split flow adjustments will increase or decrease the split ratio and the resulting sensitivity of the system and allow the analysts flexibility in optimizing their GC/MS system. Number of Water Bake Rinses 2 Water Bake Rinse Volume 7.0 mL Bake Rinse Sweep Time 0.25 min Bake Rinse Sweep Flow 100 mL/min Bake Rinse Drain Time 0.40 min Bake Time 6.00 min Bake Flow 250 mL/min Bake Temp 270 °C Condensate Bake Temp 200 °C 220 °C Ion Source Temperature:300 °C Function Type: Full Scan Solvent Delay: 0 – 0.50 min Scan Range: m/z 35 – 270 Scan Time: 0.20 sec Interscan Delay: 0.10 sec Calibration and performance standards were prepared from commercially available stock standards and diluted in Class-A volumetric flasks. Calibration standards were made from a combination of four multi-component standard mixes including a 76 component 8260 calibration mix, a 6 component 502.2 calibration gas mix, a 3 component VOA surrogate spike mix, and an 8260 internal standard mix. Stock standards were diluted to an intermediate concentration from which initial calibration standards were prepared. The calibration levels used in this study are presented in Table 3. Standards used for the Method Detection Limit (MDL) and precision experiments were prepared at appropriate concentration levels from intermediate standards. The internal standard mix was added to all samples such that a resulting concentration of 20 μg/L was achieved per 5 mL of sample. The tuning standard, 4-bromofluorobenzene, was prepared at 50 μg/L. Standards, stock standards, and associated QC/ QA samples were stored in a manner consistent with the guidelines set out in the method. Table 3. Calibration points employed in this study. Calibration Level Concentration (μg/L) 10.5 21.0 32.0 45.0 510 625 750 8100 9200 Results and Discussion The GC conditions were optimized to provide adequate separation while keeping the analytical runtime as short as possible. Figure 1 demonstrates a mid-level (25 μg/L) total ion chromatogram (TIC) from this analysis. All of the expected analyte separation is achieved while allowing for a maximum of throughput per unit laboratory time. The injection-to-injection time is further enhanced by the performance of the Clarus 680 GC oven. The oven incorporates a dual walled design with a unique air flow path to allow the fastest cooling times of any commercially available GC oven on the market. Air flow through the Clarus 680 GC is unidirectional eliminating the mixing of heated exhaust and cool ambient air allowing for the fastest transport of heat out of the GC oven. This high speed cooling allows the user to dramatically decrease the idle time of the instrument between analytical runs and in this study the injection-toinjection time, measured to be under 30 minutes, was limited by the purge and trap system. Table 4 presents the analytical results of these experiments including the Retention Time (RT), Average Relative Response Factors (Avg RRF) and Percent Relative Standard Deviation (% RSD) of the initial calibration, the Method Detection Limits (MDL), and the method precision as Percent Recovery (% Rec) and accuracy as % RSD. All of the analytes presented meet or exceed the minimum method requirements and boast excellent detection limits, precision and accuracy. Method Detection Limits were determined by analyzing seven replicate samples at a concentration of 0.5 μg/L and 1.0 µg/L. Precision and accuracy were measured by analyzing four replicates at a concentration of 25 μg/L. The majority of compounds in this report are calibrated over the entire concentration range as presented in Table 3. Compounds whose calibration range varied from the bulk of analytes are highlighted in Table 5. Alternate calibration ranges are expected for this method especially for compounds referred to as “bad actors”. In all cases, however, EPA method criteria are met. Overall improved results including MDL and precision and accuracy were obtained by operating the ion source at an elevated temperature than would regularly be used. Figure 2 shows a time selection of the seven TICs from the MDL study and illustrates the excellent repeatability of the system at a low concentration level. The inset presents the seven extracted ion chromatograms of naphthalene (EIC) at m/z = 128. The % RSD of the seven naphthalene EICs was measured to be 3.8%, which includes the sample handling performed by the purge and trap. 3 Figure 1. TIC of 8260C volatile organic standard mix at 25 μg/L. Numbering references compound as listed in Table 4. Figure 2. Zoom of the seven MDL samples. TICs overlaid to illustrate the excellent repeatability achieved at low levels. Inset shows the naphthalene EIC at m/z = 128; the % RSD was measured to be 3.8%. 4 Table 4. Summarized initial calibration results of the analyzed VOCs from 0.5 – 200 μg/L for most compounds. A hyphen (-) indicates value either not provided by the method or not generated in this work. A plus sign (+) indicates calibration calculated using linear regression with R2 presented in table. Avg # Compound RT RRF 1 Dichlorodifluoromethane + 8260C Minimum RRF % RSD 8260C % RSD Criteria MDL (µg/L) Precision Accuracy (% Rec) (% RSD) 1.38 0.440.1 14 200.1490 2.9 2Chloromethane 1.59 0.72 0.1 0.99920.990 0.15 96 3 Vinyl chloride 1.66 0.790.1 8.2 200.2195 1.8 4Bromomethane 2.00 5 Chloroethane 2.08 0.400.1 4.9 200.2499 1.9 6Trichlorofluoromethane 2.29 7 Diethyl ether 2.58 0.37 - 6.7 200.1198 4.6 0.26 0.57 0.1 0.1 18 5.9 20 20 0.18 0.06 105 109 4.9 4.1 1.2 8 1,1,2-Trichloro-1,2,2-trifluor2.78 0.380.1 13 200.1599 1.1 91,1-Dichloroethene 2.82 10 Carbon disulfide 3.02 1.460.1 5.2 200.0593 1.4 11 Allyl chloride+ 3.21 0.17 - 0.99400.990 0.05 76 2.7 12Acetonitrile 3.29 0.20 - 0.99990.990 0.38 96 5.5 13 Methylene chloride 3.37 0.510.1 16 200.2994 2.5 14 trans-1,2 Dichloroethene 3.59 15 Acrylonitrile 3.67 0.44 - 8.8 200.1497 4.6 161,1,-Dichloroethane 4.06 17 Chloroprene 4.11 0.70 - 11 200.2589 1.7 182,2,-Dichloropropane+ 4.63 0.15 19cis-1,2,-Dichloroethene 4.66 0.53 0.1 9.6 20 20 Methyl acrylate 4.76 0.92 - 4.2 20 21Propionitrile 4.84 0.22 - 4.4 20 22 Tetrahydrofuran 4.92 0.53 - 13 200.1586 4.7 23Bromochloromethane 4.92 24 Methacrylonitrile 4.96 0.83 - 8.9 200.1290 4.7 25Chloroform 4.99 0.69 0.2 4.2 20 0.09 103 2.1 261,1,1-Trichloroethane 5.14 0.55 0.1 5.5 20 0.20 107 0.9 27 Pentafluorobenzene (ISTD) 5.22 28 Carbon tetrachloride 5.27 0.38 0.1 12 20 0.13 108 29 1,1-Dichloropropene 5.32 0.43 - 5.9 200.2589 1.6 30 Isobutyl alcohol 5.49 0.09 - 11.5 200.2289 5.6 31 Benzene 5.53 1.330.5 6.9 200.2290 1.6 32 1,2-dichloroethane-d4 (Surr) 5.55 33 1,2-Dichloroethane 5.64 0.420.1 9.8 200.0991 2.0 34 1,4-Difluorobenzene (ISTD) 6.00 35Trichloroethene 6.25 0.32 0.2 8.9 20 0.28 103 36 1,2-Dichloropropane 6.56 0.470.1 6 200.1596 1.8 37 Methyl methacrylate 6.67 0.39 - 6.2 200.1791 3.9 38Dibromomethane 6.70 0.19 - 4.9 20 0.10 106 2.7 39Bromodichloromethane 6.87 0.28 0.2 7.4 20 0.17 102 1.9 + + 0.44 0.46 0.98 0.26 0.1 0.1 0.2 15 9.3 9.2 20 20 20 0.30 0.34 0.16 - 0.99960.990 - - 4.5 20 102 105 107 1.8 1.3 2.3 56 7.4 0.24 106 1.9 0.14 100 5.3 0.12 100 5.7 0.09 107 3.6 1.2 2.5 402-Nitropropane 7.23 0.06 94 4.7 41cis-1,3-Dichloropropene 7.41 0.47 0.2 8.7 20 0.08 90 1.6 42 Toluene-d8 (Surr) 7.66 - 0.99960.990 - 5 Table 4 continued Avg # Compound RT RRF 8260C Minimum RRF % RSD 8260C % RSD Criteria MDL (µg/L) Precision Accuracy (% Rec) (% RSD) 43Toluene 7.74 0.67 0.4 6.5 20 0.22 100 1.5 44trans-1,3-Dichloropropene 8.11 0.39 0.1 17 20 0.11 93 2.5 45 Ethyl methacrylate 8.16 0.66 - 11 20 0.07 101 2.6 461,1,2-Trichloroethane 8.31 0.28 0.1 4.3 20 0.06 96 1.9 47Tetrachloroethene 8.35 0.43 0.2 20 20 0.17 123 4.6 481,3-Dichloropropane 8.50 0.65 - 4.4 20 0.06 94 3.1 49Chlorodibromomethane 8.69 0.28 0.1 12 20 0.10 108 3.7 50 Ethylene dibromide 8.81 0.40 0.1 3.6 20 0.07 111 2.8 51 Chlorobenzene-d5 (ISTD) 9.21 52Chlorobenzene 9.24 1.08 0.5 4.7 20 0.23 102 1.0 53Ethylbenzene 9.31 1.34 0.1 2.2 20 0.20 104 1.4 541,1,1,2-Tetrachloroethane 9.31 0.27 - 7.8 20 0.19 105 1.9 55 p- & m-Xylene 9.41 1.09 0.1 4.4 20 0.23 101 1.6 56o-Xylene 9.73 1.15 0.1 2.5 20 0.19 103 2.2 3.8 20 0.18 57Styrene 9.76 1.09 0.3 58Bromoform+ 9.91 0.22 0.1 10.01 1.61 0.1 59Isopropylbenzene 0.99960.990 0.10 2.8 20 0.18 106 1.8 104 2.8 106 1.8 + 60cis-1,4-Dichloro-2-butene 10.120.22 - 0.9986 0.9900.09 68 6 61 Bromofluorobenzene (Surr) 10.15 62Bromobenzene 10.25 631,1,2,2-Tetrachloroethane 10.28 64n-Propylbenzene 651,2,3-Trichloropropane 2.2 0.98 - 8.5 20 0.15 98 2.3 1.17 0.3 5.9 20 0.16 93 9.9 10.30 3.21 - 4.8 20 0.20 101 1.1 10.32 1.76 - 5.8 20 0.10 91 3.3 0.9900.05 71 66trans-1,4-Dichloro-2-butene+10.330.25 - 0.9991 2.2 672-Chlorotoluene 10.38 1.88 - 4.4 20 0.19 102 0.9 681,3,5-Trimethylbenzene 10.43 3.00 - 6.3 20 0.21 104 1.1 694-Chlorotoluene 10.47 1.99 - 7.6 20 0.15 100 3.2 70tert-Butylbenzene 10.64 2.57 - 9.2 20 0.18 94 1.3 711,2,4-Trimethylbenzene 10.69 2.94 - 4.8 20 0.19 105 1.2 72sec-Butylbenzene 10.79 3.48 - 5.1 20 0.20 107 1.7 734-Isopropyltoluene 10.88 2.98 - 4.5 20 0.22 97 1.0 741,3-Dichlorobenzene 10.89 1.64 0.6 9 20 0.22 98 0.7 75 1,4-Dichlorobenzene-d4 (ISTD) 10.93 761,4-Dichlorobenzene 10.95 1.65 0.5 9.4 20 0.18 97 1.4 77n-Butylbenzene 11.14 2.10 - 6.5 20 0.17 98 1.2 781,2-Dichlorobenzene 11.19 1.60 0.4 4.5 20 0.12 102 1.5 791,2-Dibromo-3-chloropropane 11.69 0.33 0.05 9.8 20 0.08 80Nitrobenzene+ 11.84 0.06 811,2,4-Trichlorobenzene 12.24 1.19 0.2 9.3 20 82Hexachlorobutadiene 12.32 0.50 - 16 20 83Naphthalene 12.46 3.60 - 7.3 841,2,3-Trichlorobenzene 12.65 1.13 0.2 7.2 - 98 3.6 102 5.4 0.16 92 1.3 0.23 91 1.3 20 0.06 102 2.4 20 0.09 94 1.6 0.99990.990 - Table 5. Alternate calibration range compounds. Calibration Concentration Range Compound 0.5 – 100 μg/L Tetrachloroethene 1.0 – 200 μg/L Bromomethane Allyl chloride Acetonitrile Isobutyl alcohol 2.0 – 200 μg/L 2,2-Dichloropropane 2-Nitropropane Nitrobenzene 1.0 – 400 µg/L p- & m-Xylene Design advancements in the Clarus SQ 8 mass spectrometer allow such a wide operating range with the Clarifi™ detector providing much of the advantage. The detector’s enhanced sensitivity enables full scan mass spectrometry methods to be performed resulting in library searchable spectra at the lowest possible concentration levels. With the added sensitivity of SIFI™ (selected ion full ion) or SIM (selected ion monitoring) data acquisition at even lower limits are possible for specific compounds. The voltage setting on the Clarifi detector was important in developing the method for such a wide concentration range and deserves mention here. The UltraTune™ (Standard – DFTPP/BFB) function built into the TurboMass™ v6.0 software is designed to set the voltage such that the m/z = 69 peak of the tune gas FC43 is approximately 80% in the tune window. This setting is appropriate for general instrument analyses however users are encouraged to fine tune this setting to suit their specific analytical needs. In our case this was achieved by analyzing two samples, the high and low concentration level of our calibration, at voltage settings below that set during UltraTune. The optimum voltage setting produced measurable signal at the lowest level while not saturating the detector at the highest level. In our case the initial UltraTune setting was 1570V with test measurements performed at -50V and -100V of this value. The analysis was performed at 1520V. The stability of the Clarus SQ 8 system allows this setting to be used for extended periods of time and need only be reproduced periodically. The UltraTune (Standard – DFTPP/BFB) function also produces a satisfactory tune which meets the Tune Evaluation requirements set out in the method. Figure 3 demonstrates the passing BFB Tune Evaluation Sample. Figure 4 shows the mass spectrum utilized in the Tune Evaluation Test and all of the necessary peak intensity comparisons are achieved. Figure 3. TurboMass 6.0 EPA BFB tune evaluation results. Figure 4. Mass spectrum of EPA BFB tune evaluation used for test in Figure 3. Conclusion The analysis of VOCs by purge and trap GC/MS following EPA Method 8260C is easily achievable over a wide concentration range using the Clarus SQ 8 GC/MS system. Satisfactory analytical results were achieved for a concentration range from 0.5 – 200 μg/L for a majority of analyte compounds using the full scan approach providing library searchable spectra at all concentration levels. Excellent detection limits are reported as are accuracy and precision measurements. A number of technological advances make the Clarus SQ 8 GC/MS the ideal systems for laboratories wishing to perform high throughput and sensitivity analyses with an ease of operation currently unmatched. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2012, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 010017B_01