Improved Sensitivity and Dynamic Range Using the Clarus SQ8 GC/MS System for EPA Method 8270D

a p p l i c at i o n N o t e
Gas Chromatography/
Mass Spectrometry
Authors
Yury Kaplan
Ruben Garnica
PerkinElmer, Inc.
Shelton, CT 06484 USA
Improved Sensitivity
and Dynamic Range
Using the Clarus SQ 8
GC/MS System for
EPA Method 8270D
Semi-Volatile Organic
Compound Analysis
Introduction
U.S. Environmental Protection
Agency (EPA) Method 8270D – SemiVolatile Organic Compounds by Gas
Chromatography/Mass Spectrometry
(GC/MS) – is a common and wide ranging
method employed in nearly all commercial
environmental laboratories. The analysis
focuses on the detection of trace level
semi-volatile organic compounds in
extracts from solid waste matrices, soils,
air sampling media and water samples.
The method lists over 200 compounds
however a majority of laboratories target
between 60 and 90 for most analyses.
The study presented here demonstrates
the PerkinElmer® Clarus® SQ 8 GC/MS, not only meets the method requirements
but provides users flexibility to satisfy their individual productivity demands.
An extended calibration range is presented as are the advantages of the Clarifi™
detector.
Experimental
The PerkinElmer Clarus SQ 8T GC/MS system operating in
electron ionization mode was used to perform these experiments. Experimental conditions are presented in Table 1
and represent standard conditions for EPA Method 8270D.
Minimum inlet reactivity was achieved using a deactivated
quartz liner (4-mm) without wool. As prescribed by the
method, a pressure pulse injection scheme was employed
and is described in Table 2. Data collection was achieved
using the TurboMass™ v6 GC/MS software with environmental reporting. The TurboMass software comes with a
built-in UltraTune™ (Standard – DFTPP/BFB) tuning procedure,
which was used to establish the base detector setting.
Table 1. GC and MS experimental conditions.
Analytical Column:
Elite-5MS (30 m x 0.25 mm id x 0.25 μm)
Calibration and performance standards were prepared
from commercially available stock standards and diluted in
Class-A volumetric flasks. Calibration standards were made
from a 76 component 8270 mega mix, a 15 component
acids fortification mix and a 6 component semi-volatile
internal standard mix. The acids were fortified with the
fortification mix to double the concentration relative to
the other components. Stock standards were diluted to
an intermediate concentration from which initial calibration standards were prepared. The calibration levels used
in this study are presented in Table 3. The internal standard
mix was added to all samples such that a resulting concentration of 40 μg/mL was achieved. The tuning standard,
Decafluorotriphenylphosphine (DFTPP), was prepared at
50 μg/mL. Standards, stock standards, and associated
QC/QA samples were stored in a manner consistent with
the guidelines set out in the method.
Injector Temperature: 300 °C
Carrier Gas:
Helium @ 1.0 mL/min
Oven Program:
Temperature Hold Time
Rate
37 °C
0.5 min
265 °C
0 min
6 °C/min
287 °C
0 min
25°C/min
320 °C
1.85 min
End
18 °C/min
“Acids”“Bases/Neutrals”
CalibrationConcentration Concentration
Level
(μg/mL)
(μg/mL)
12
1
24
2
38
4
412
6
GC Transfer Line
Temperature:
280 °C
MS Ion Source
Temperature:
280 °C
630 15
MS Function Type:
Full Scan
740 20
MS Solvent Delay:
0 – 2.30 min
880 40
MS Scan Range:
m/z 35 – 500
9160 80
MS Scan Time:
0.10 sec
10200
100
MS Interscan Delay:
0.10 sec
11300
150
520 10
Table 2. Pressure plus flow settings.
Time (min)
Split Flow (mL/min)
-0.504
-0.250
0.751
1.0050
5.0020
2
Table 3. Calibration points employed in this study.
Results and Discussion
An initial consideration when performing any method is to
determine the operating range with respect to on-column
concentration. The Clarus SQ 8 GC/MS allows the user to
control the linear range by controlling the Clarifi detector
voltage setting. The built-in UltraTune (Standard – DFTPP/
BFB) tune procedure adjusts the mass spectrometer providing
excellent spectral fidelity and sets the detector voltage such
that low sensitivity work can be performed. This standard
setting can be used directly and is recommended for most
applications. Additionally, the detector voltage can be
increased for added sensitivity when needed, or, as is the
case with this application note, the detector voltage can be
tuned to provide both sensitivity and a wide dynamic range
Scan EI+
7.22e8
198
100
442
77
%
255
127
69
51
110
50
39
0
57
43
78 93
186
117
148
93
275
206
129
107
167
143
193
224
441
244 256
296
243
293
423
323 334 365
343
393
443
444
443
493
543
m/z
Figure 2. Mass spectrum of DFTPP as analyzed by the tune evaluation shown
in Figure 1.
suitable for the analysis at hand. In this study the Clarifi
detector voltage was optimized such that both the low level
samples were measurable while the high level samples were
not saturating the detector. Specifically, the detector was
lowered 260V to the operational 1400V. The added sensitivity
of the Clarifi detector enhances a wider range of analyses
so that even at low voltage settings the low concentration
standards and samples are easily detected.
Figure 1. EPA 8270D tune evaluation as performed by the built-in software.
Passing result shown.
Scan EI+
TIC
63
100
43
32 36
23
25
20
57 58
54
53
62
46
27 28
38
29
60
71
52
2
%
67
47
12
8 13
24
10 15
18
41
56
48 50
51
59
72
68 69
65
66
31
55
16
17
30
1
0
3.00
5.00
7.00
9.00
11.00
13.00
15.00
17.00
Time
19.00
Figure 3. TIC of 40 μg/mL calibration sample. Retention Times are listed in Table 4 (Page 4).
3
Scan EI+
252
67
100
68
66
%
0
15.00
15.20
15.40
15.60
15.80
16.00
16.20
16.40
16.60
16.80
Time
17.00
Figure 4. Zoom of EIC at m/z = 252 showing separation of benzo[b]
fluoranthene (66) and benzo[k]fluoranthene (67). Benzo[a]pyrene (68)
also shown.
In addition to providing a solid foundation from which to
perform analyses the UltraTune (Standard – DFTPP/BFB)
function also produces a satisfactory tune, which meets the
tune evaluation requirements set out in Method 8270D.
Figure 1 demonstrates the passing DFTPP tune evaluation
sample while Figure 2 shows the mass spectrum utilized in
the tune evaluation test.
The GC conditions were optimized to provide analytical
separation and Figure 3 demonstrates a mid-level (40 μg/mL)
total ion chromatogram (TIC) from this analysis. All of the
expected analyte separation is achieved and a listing of
retention times (RT) measured using the instrument parameters
as described in Table 1 is presented. The Elite-5MS column
provided excellent separation and Figure 4, for example,
demonstrates the separation of the benzo[b]fluoranthene
and benzo[k]fluoranthene.
Table 4. Calibration data table with retention times listed in Figure 3 (Page 3).
#
Name
RT (min)
Average RRF
8270D Criteria RRF
1
2
Pyridine
3.09
0.2
8.98
Aniline
4.44
1.2
13.16
3Phenol
4.45
1.7
4
N-Nitrosodimethylamine
4.45
0.1
10.3
5
Bis (2-chloroethyl) ether
4.49
2.1
0.7
12.66
9.52
6
1,3-Dichlorobenzene
4.66
1.1
15.56
7
1,4-Dichlorobenzene - D4 (ISTD)
4.73
8.83
8
1,4-Dichlorobenzene
4.75
1.0
9.73
9
1,2-Dichlorobenzene
4.91
1.3
11.27
10
Benzyl Alcohol
4.94
1.0
11.86
11
Bis (2-chloroisopropyl) ether
5.05
1.7
9.34
122-Methylphenol
5.07
0.6
13
5.26
0.7
5.26
0.9
3-Methylphenol
144-Methylphenol
0.7
10.66
11.71
0.6
8.59
15Hexachloroethane
5.28
0.5
0.3
10.62
16Nitrobenzene
5.41
0.4
0.2
9.91
17Isophorone
5.70
0.7
0.4
11.31
18
5.78
0.2
2-Nitrophenol
11.86
19Bis(2-chloroethoxy)methane
5.96
0.5
0.3
15.45
202,4-Dichlorophenol
6.08
0.3
0.2
16.69
21
1,2,4-Trichlorobenzene
6.14
0.4
14.19
22
Naphthalene - D8 (ISTD)
6.21
7.41
23Naphthalene
4
0.8
% RSD
6.24
1.1
0.7
9.63
24Hexachlorobutadiene
6.39
0.2
0.01
10.38
254-chloro-3-methyl-phenol
7.01
0.3
0.2
16.86
262-Methylnaphthalene
7.09
0.6
0.4
14.73
27
7.22
0.6
1-Methylnaphthalene
13.96
28Hexachlorocyclopentadiene
7.30
0.2
0.05
9.14
292-Chloronaphthalene
7.72
1.4
0.8
12.09
Table 4 continued
#
Name
RT (min)
Average RRF
8270D Criteria RRF
% RSD
302-Nitroaniline
7.93
0.4
0.01
9.54
31
8.19
1.4
0.01
10.67
8.25
0.3
0.2
13.95
33Acenaphthylene
8.25
2.0
0.9
8.27
34
8.45
Dimethyl phthalate
322,6-Dinitrotoluene
Acenaphthene - D10 (ISTD)
7.56
35Acenaphthene
8.49
1.5
0.9
9.85
363-Nitroaniline
8.49
0.2
0.01
14.4
372,4-Dinitrophenol
8.60
0.1
0.01
8.17
38Dibenzofuran
8.71
1.7
0.8
11.83
394-Nitrophenol
8.79
0.1
0.01
10.14
40
2,3,4,6-Tetrachlorophenol
8.85
0.2
12.71
41
2,3,5,6-Tetrachlorophenol
8.91
0.2
10.17
42Diethylphthalate
9.09
1.1
0.01
9.93
43Fluorene
9.15
1.3
0.9
11.93
444-Chlorodiphenylether
9.17
0.5
0.4
13.53
454,6-Dinitro-2-methylphenol
9.29
0.1
0.01
16.04
464-Nitroaniline
9.31
0.1
0.01
12.33
47N-Nitrosodiphenylamine
9.36
0.5
0.01
11.4
48
4-Bromophenyl phenyl ether
9.80
0.2
0.1
9.57
49
Azobenzene
9.81
0.1
10.13
50Hexachlorobenzene
9.88
0.2
0.1
10.17
51Pentachlorophenol
10.18
0.1
0.05
8.01
52
10.38
Phenathrene - D10 (ISTD)
3.48
53Phenanthrene
10.42
1.1
0.7
13.5
54Anthracene
10.48
1.1
0.7
15.16
55Carbazole
10.72
0.4
0.01
8.39
56
11.20
0.8
0.01
11.42
57Fluoroanthene
Di-n-butyl phthalate
11.97
1.0
0.6
8.52
58Pyrene
12.26
1.2
0.6
9.62
59
Butyl benzyl phthalate
13.16
0.4
0.01
12.35
60
Bis (2-ethylhexyl) adipate
13.26
0.5
14.38
61Benzo[a]anthracene
13.89
1.0
62
13.92
Chrysene - D12 (ISTD)
0.8
11.8
6.34
63Chrysene
13.96
1.0
0.7
16.57
64
Bis (2-ethylhexyl) phthalate
13.98
0.6
0.01
17.47
65
Di-n-Octyl phthalate
15.04
1.0
66Benzo[b]fluoranthene
15.68
1.2
0.7
12.55
67Benzo[k]fluoranthene
15.74
0.9
0.7
13.52
68Benzo[a]pyrene
16.29
1.0
0.7
9.46
69
16.42
Perylene - D12 (ISTD)
11.13
4.24
70Indeno(1,2,3-c,d)pyrene
18.00
1.2
0.5
8.72
71Dibenz[a,h]anthracene
18.12
1.0
0.4
9.25
72
18.42
1.0
0.5
7.98
Benzo[g,h,i] perylene
5
The results of the initial calibration curve for EPA Method
8270D are presented in Table 4 and include retention time,
average relative response factor (Avg. RRF), method criteria,
and percent relative standard deviation (% RSD). Satisfactory
results are obtained for all for the compounds listed.
Conclusion
In this application note the Clarus SQ 8 GC/MS system is
shown to provide wide concentration range applicability to
U.S. EPA Method 8270D, allowing the analysts flexibility
in performing experimentation to meet their needs. Satisfactory
analytical results were demonstrated over a concentration
range from 1.0 – 150 μg/mL for a majority of analyte
compounds using the full scan approach providing library
searchable spectra at all concentration levels. A number
of technological advances make the Clarus SQ 8 GC/MS
the ideal systems for laboratories wishing to perform high
throughput and sensitivity analyses with an ease of operation
that is currently unmatched.
PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com
For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2011, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.
010016_01