Automation & Drives Technical Support Information No. 134 Revision 4.0 August 2006 Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe Page 1 of 12 Ergänzende Service-Informationen zum MicroSAM - Teil 3 Entsprechend der wachsenden Anzahl von MicroSAM Applikationen im Feld und den daraus resultierenden Erfahrungen ergeben sich folgende aktuelle Service–Informationen: - Umbau von Analysenmodulen ab Erzeugnis-Stand 03 (ES03). Der TCD_2A wird am „Split/Proben-Ausgang“ versetzt. Diese Änderung gilt für alle Module (A01-A10). Carrier gas Backflush / Sample vent EPC 1 Sample in EPC 3 Sample out TCD 1B TCD 2A Live switching A01 ... A03 TCD 3B Membrane valve Live Injection TCD 4B Column 1a TCD 3A Column 1b TCD 1A TCD 4A Column 3 Column 2 TCD 2B EPC 2 Carrier gas Backflush / Sample vent EPC 1 Sample in EPC 3 Sample out TCD 1B TCD 2A Live switching A04 ... A10 TCD 4A Membrane valve Live Injection TCD 4B TCD 1A Column 1a Column 1b Column 2 TCD 2B EPC 2 © SIEMENS AG Karlsruhe - A&D PI 2 AS 2 - Niko Benas +49 721 595 7216 - [email protected] Als Betriebsgeheimnis anvertraut. Alle Rechte vorbehalten. Der Inhalt wurde auf Übereinstimmung mit der beschriebenen Hard- und Software geprüft, dennoch können Abweichungen nicht ausgeschlossen werden. Technische Änderungen vorbehalten. Proprietary data. All rights reserved. While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full agreement. Technical data are subject to change. Automation & Drives Technical Support Information No. 134 Revision 4.0 Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe August 2006 Page 2 of 12 Nun kann visuell überprüft werden, ob der Ausspülzeitpunkt optimal ist bzw. die Fläche des Dosierpeaks ihr Maximum erreicht hat: Maximale Dosierpeakfläche hier Flushloop ON bei -4 sec Sample ON bei -4 sec Dosierpeakfläche zu klein hier Flushloop ON bei -8 sec Sample ON bei -8 sec © SIEMENS AG Karlsruhe - A&D PI 2 AS 2 - Niko Benas +49 721 595 7216 - [email protected] Als Betriebsgeheimnis anvertraut. Alle Rechte vorbehalten. Der Inhalt wurde auf Übereinstimmung mit der beschriebenen Hard- und Software geprüft, dennoch können Abweichungen nicht ausgeschlossen werden. Technische Änderungen vorbehalten. Proprietary data. All rights reserved. While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full agreement. Technical data are subject to change. Automation & Drives Technical Support Information No. 134 Revision 4.0 Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe August 2006 Page 3 of 12 - Da an dieser Stelle nicht gemessen wird, sollte der Detektor geschont und die Brückenspannung bzw. die Brückenleistung des TCD_2 auf ein Minimum eingestellt werden: max. 70 mW für Trägergas Helium/Wasserstoff und max. 20 mW für Trägergas Stickstoff. - Das Membranventil wird ab sofort bespült (ES03). Die Gas-Versorgung erfolgt ausgangsseitig über das LIVET Stück (EPC-3), die ja bei korrektem Abgleich keine Probe beinhaltet. Die Entsorgung erfolgt über den Port 6, der bisher zur Kontrolle von möglichen Undichtigkeiten diente (Entsorgung pneumatischer Stecker). - Der MicroSAM besitzt seit Artikelstand 4 (AS04) vier Digital-Ausgänge und vier Digital-Eingänge statt bisher zwei: © SIEMENS AG Karlsruhe - A&D PI 2 AS 2 - Niko Benas +49 721 595 7216 - [email protected] Als Betriebsgeheimnis anvertraut. Alle Rechte vorbehalten. Der Inhalt wurde auf Übereinstimmung mit der beschriebenen Hard- und Software geprüft, dennoch können Abweichungen nicht ausgeschlossen werden. Technische Änderungen vorbehalten. Proprietary data. All rights reserved. While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full agreement. Technical data are subject to change. Automation & Drives Technical Support Information No. 134 Revision 4.0 Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe August 2006 Page 4 of 12 © SIEMENS AG Karlsruhe - A&D PI 2 AS 2 - Niko Benas +49 721 595 7216 - [email protected] Als Betriebsgeheimnis anvertraut. Alle Rechte vorbehalten. Der Inhalt wurde auf Übereinstimmung mit der beschriebenen Hard- und Software geprüft, dennoch können Abweichungen nicht ausgeschlossen werden. Technische Änderungen vorbehalten. Proprietary data. All rights reserved. While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full agreement. Technical data are subject to change. Automation & Drives Technical Support Information No. 134 Revision 4.0 Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe August 2006 Page 5 of 12 - Ein neues LCD-Display (ab AS05) finden nun Verwendung, das sequentiell die einzelnen Probenkomponenten und deren Konzentrationen anzeigt: - Neue RSP-Boards A5E00315395 mit Ausgabestand 2 und Erzeugnisstand 2 werden nun gefertigt und ab sofort eingesetzt. In der Vergangenheit gab es vereinzelt Fälle, in denen die Betriebsspannung des MicroSAM unter dem Nominalwert von 24V lag. Dabei kam es zu einem deutlich wahrnehmbaren ‚Klackernen’-Geräusch. Dieses war zurückzuführen auf wiederholend anziehende und abfallende Relais der WLD Stromversorgung. Dieses wiederum wurde verursacht durch ein Schaltungsproblem in der Unterspannungserkennung, die normalerweise eine geordnete Abschaltung beim Ausschalten des SAM durchführt. Es gab Fälle, in denen diese Abschaltung irrtümlich bereits bei Unterschreitung von 23V die TCD Versorgung unterbricht. Die neuen RSP-Platinen haben nun eine Abschaltschwelle von ca. 16V. © SIEMENS AG Karlsruhe - A&D PI 2 AS 2 - Niko Benas +49 721 595 7216 - [email protected] Als Betriebsgeheimnis anvertraut. Alle Rechte vorbehalten. Der Inhalt wurde auf Übereinstimmung mit der beschriebenen Hard- und Software geprüft, dennoch können Abweichungen nicht ausgeschlossen werden. Technische Änderungen vorbehalten. Proprietary data. All rights reserved. While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full agreement. Technical data are subject to change. Automation & Drives Technical Support Information No. 134 Revision 4.0 Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe August 2006 Page 6 of 12 - Ab sofort wird der Akku für die interne CMOS Uhr eingebaut (ab AS04). Diese kann jedoch auch nachträglich in jeden MicroSAM mit RSP Platine 1.4 montiert werden. Ersatzteil Batteriebaustein A5E00713507A: - Ab AS04 können mit dem MicroSAM bis zu vier Ströme analysiert werden, wobei davon je nach Applikation einer oder zwei als Kalibierströme genutzt werden können. Mit dieser Erweiterung kann eine Vielzahl von neuen Applikationen bedient werden, die einen Gas-Chromatograph mit mehreren Probenströmen benötigen © SIEMENS AG Karlsruhe - A&D PI 2 AS 2 - Niko Benas +49 721 595 7216 - [email protected] Als Betriebsgeheimnis anvertraut. Alle Rechte vorbehalten. Der Inhalt wurde auf Übereinstimmung mit der beschriebenen Hard- und Software geprüft, dennoch können Abweichungen nicht ausgeschlossen werden. Technische Änderungen vorbehalten. Proprietary data. All rights reserved. While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full agreement. Technical data are subject to change. Automation & Drives Technical Support Information No. 134 Revision 4.0 August 2006 Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe Page 7 of 12 Complementary Service-Informations for the MicroSAM – Part 3 According to the increasing number of MicroSAM applications in the field and the experiences resulting from it following new service informations arised: - New design of the analytical modules since Product-Class 03 (ES03). The TCD_2A is displaced at the outlet of the Backflush/Sample vent. This modification is valid for all modules (A01-A10). Carrier gas Backflush / Sample vent EPC 1 Sample in EPC 3 Sample out TCD 1B TCD 2A Live switching A01 ... A03 TCD 3B Membrane valve Live Injection TCD 4B Column 1a TCD 3A Column 1b TCD 1A TCD 4A Column 3 Column 2 TCD 2B EPC 2 Carrier gas Backflush / Sample vent EPC 1 Sample in EPC 3 Sample out TCD 1B TCD 2A Live switching A04 ... A10 TCD 4A Membrane valve Live Injection TCD 4B TCD 1A Column 1a Column 1b Column 2 TCD 2B EPC 2 © SIEMENS AG Karlsruhe - A&D PI 2 AS 2 - Niko Benas +49 721 595 7216 - [email protected] Als Betriebsgeheimnis anvertraut. Alle Rechte vorbehalten. Der Inhalt wurde auf Übereinstimmung mit der beschriebenen Hard- und Software geprüft, dennoch können Abweichungen nicht ausgeschlossen werden. Technische Änderungen vorbehalten. Proprietary data. All rights reserved. While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full agreement. Technical data are subject to change. Automation & Drives Technical Support Information No. 134 Revision 4.0 Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe August 2006 Page 8 of 12 Now it can be visual verified, if the rinse point is optimised in time respectively if the injection peak offers maximum area: Maximum injection peak area here Flushloop ON at -4 sec Sample ON at -4 sec Injection peak area too small here Flushloop ON at -8 sec Sample ON at -8 sec © SIEMENS AG Karlsruhe - A&D PI 2 AS 2 - Niko Benas +49 721 595 7216 - [email protected] Als Betriebsgeheimnis anvertraut. Alle Rechte vorbehalten. Der Inhalt wurde auf Übereinstimmung mit der beschriebenen Hard- und Software geprüft, dennoch können Abweichungen nicht ausgeschlossen werden. Technische Änderungen vorbehalten. Proprietary data. All rights reserved. While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full agreement. Technical data are subject to change. Automation & Drives Technical Support Information No. 134 Revision 4.0 Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe August 2006 Page 9 of 12 - Because of no measurement at this place, the bridge voltage repectively the bridge current of the TCD_2 should be adjusted to a minimum: max. 70 mW for carrier gas helium/hydrogen and max. 20 mW for carrier gas nitrogen. - The diaphragm valve is now continously rinsed (ES03). The gas supply takes place from the outlet of the LIVET piece (EPC-3), which should not contain any sample if it’s correct adjusted. The disposal takes place on port 6, which was so far a verification of possible leaks (outlet from pneumatic plug). - The MicroSAM owns since Article-Class 4 (AS04) four digital-outputs and four digital-inputs in place of two: © SIEMENS AG Karlsruhe - A&D PI 2 AS 2 - Niko Benas +49 721 595 7216 - [email protected] Als Betriebsgeheimnis anvertraut. Alle Rechte vorbehalten. Der Inhalt wurde auf Übereinstimmung mit der beschriebenen Hard- und Software geprüft, dennoch können Abweichungen nicht ausgeschlossen werden. Technische Änderungen vorbehalten. Proprietary data. All rights reserved. While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full agreement. Technical data are subject to change. Automation & Drives Technical Support Information No. 134 Revision 4.0 Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe August 2006 Page 10 of 12 © SIEMENS AG Karlsruhe - A&D PI 2 AS 2 - Niko Benas +49 721 595 7216 - [email protected] Als Betriebsgeheimnis anvertraut. Alle Rechte vorbehalten. Der Inhalt wurde auf Übereinstimmung mit der beschriebenen Hard- und Software geprüft, dennoch können Abweichungen nicht ausgeschlossen werden. Technische Änderungen vorbehalten. Proprietary data. All rights reserved. While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full agreement. Technical data are subject to change. Automation & Drives Technical Support Information No. 134 Revision 4.0 Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe August 2006 Page 11 of 12 - A new LCD-display (since AS05) is now in usage, which shows sequential the particular sample components and its concentrations: - New RSP-Boards A5E00315395 Article-Class 2 and Product-Class 2 are now manufactured and in use. In some circumstances, the operating voltage of the MicroSAM was below the nominal value of 24V. This resulted sometimes to a noticeable ‘ticking’ noise. The cause was a repetitive activation and deactivation of the TCD power supply relais. In some cases the shutoff of the TCD power supply took place below 23 V. The new RSP boards own a shutoff threshold of approx. 16V. © SIEMENS AG Karlsruhe - A&D PI 2 AS 2 - Niko Benas +49 721 595 7216 - [email protected] Als Betriebsgeheimnis anvertraut. Alle Rechte vorbehalten. Der Inhalt wurde auf Übereinstimmung mit der beschriebenen Hard- und Software geprüft, dennoch können Abweichungen nicht ausgeschlossen werden. Technische Änderungen vorbehalten. Proprietary data. All rights reserved. While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full agreement. Technical data are subject to change. Automation & Drives Technical Support Information No. 134 Revision 4.0 Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe August 2006 Page 12 of 12 - With immediate effect an accumulator for the internal CMOS clock is installed (since AS04). This part can be installed additionally in every MicroSAM with RSP board 1.4. Spare part No. is A5E00713507A: - Since AS04 the MicroSAM is able to analyze up to four streams, whereby one or two could be used as calibration streams. With this new feature the MicroSAM serves considerably more applications, which require gas chromatographs with multiple streams. © SIEMENS AG Karlsruhe - A&D PI 2 AS 2 - Niko Benas +49 721 595 7216 - [email protected] Als Betriebsgeheimnis anvertraut. Alle Rechte vorbehalten. Der Inhalt wurde auf Übereinstimmung mit der beschriebenen Hard- und Software geprüft, dennoch können Abweichungen nicht ausgeschlossen werden. Technische Änderungen vorbehalten. Proprietary data. All rights reserved. While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full agreement. Technical data are subject to change.