AIMTEC AME30

Series AME30-MAZ
30 Watt | AC-DC Converter
FEATURES:
•
•
•
•
•
•
•
•
4000VAC I/O isolation
Medical Switching power modules for PCB mounting
Operating temperature: -40 to +85°C
Universal Input: 90-260VAC, 47-440Hz, or 120-370VDC
Energy star compliant
RoHS compliant
Low Ripple and Noise
CE, cULus approvals
CB
Models
Single output
Model
AME30-3.3SMAZ
AME30-5SMAZ
AME30-12SMAZ
AME30-15SMAZ
AME30-24SMAZ
Input Voltage
(VAC/Hz)
Input Voltage
(VDC)
90-264/47-440
90-264/47-440
90-264/47-440
90-264/47-440
90-264/47-440
120-370
120-370
120-370
120-370
120-370
Input Voltage
(VAC/Hz)
Input Voltage
(VDC)
Output Voltage
(V)
Output Current
max
(A)
Efficiency
(%)
6
6
2.5
2
1.25
80
83
85
85
84
Output Current
max
(A)
Efficiency
(%)
3.3
5
12
15
24
Models
Dual output
Model
Output Voltage
(V)
AME30-5DMAZ
90-264/47-440
120-370
±5
±3
81
AME30-12DMAZ
90-264/47-440
120-370
±12
±1.25
84
AME30-15DMAZ
90-264/47-440
120-370
±15
±1
83
AME30-24DMAZ
90-260/47-440
120-370
±24
±0.625
82
NOTE: All specifications in this datasheet are measured at an ambient temperature of 25°C, humidity<75%, nominal input voltage and at
rated output load unless otherwise specified.
Input Specifications
Parameters
Conditions
Current
Inrush current <2ms
Leakage current
External fuse
Input dissipation
Typical
115 VAC
230 VAC
115 VAC
230 VAC
Recommended slow blow type
No load; 115VAC
No load; 264VAC
Start-up time
Maximum
Units
740
410
25
50
150
mA
mA
A
A
µA
A
3.15
<0.5
<0.8
117
W
ms
Output Specifications
Parameters
Voltage accuracy
Line regulation
Load regulation
Cross regulation
Maximum Capacitive load
Transient response deviation
Ripple & Noise*
Hold-up time (min)
Minimum Load Current
Conditions
Typical
(LL-HL)
0-100% load single
0-100% load dual
25% load - 1st out, 100% load – 2nd out
Depending of the model
25% load Step
20MHz bandwidth
115VAC
±2
±0.5
±1
±2
±5
470-23 000
±2
50
29
0
Maximum
Units
%
%
%
%
µF
% of Vout
mV p-p
ms
% of Max
*Ripple & Noise measured with 0.1μF M/C and 1μF E/C
www.aimtec.com
F 052e R8.D
Tel: +1 514 620 2722
1 of 3
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only
Series AME30-MAZ
30 Watt | AC-DC Converter
Isolation Specifications
Parameters
Conditions
Tested I/O voltage
Isolation Resistance
Typical
3 sec
Rated
Units
4000
VAC
MΩ
>1000
General Specifications
Parameters
Conditions
Switching frequency
Over current protection
Over voltage protection
Short circuit protection
Short Circuit restart
Operating temperature
Maximum case temperature
Storage temperature
Temperature coefficient
Cooling
Humidity
Case material
Weight
Dimensions (L X H X W)
MTBF
Typical
47
110
Zener diode clamp
Hiccup mode, indefinite
Auto recovery
-40 - +85
Auto recovery
With derating above 55ºC
Maximum
Units
140
110
KHz
%
%
o
C
C
o
C
% / oC
o
100
-40 to +95
0.02
Free air convection
Non condensing
95
% RH
Plastic resin + Fiberglass (flammability to UL 94V-0)
300
3.56 x 2.48 x 1.18 inches
90.5 x 63 x 30 mm
> 800 000 hrs (MIL-HDBK -217F, t=+25oC)
g
Environment Approval
Test
Shock
Vibration
Parameters
Wave form
Acceleration amplitude
Bump duration
Converter operation
Number of bumps
Test mode
Displacement
Acceleration
Converter operation
Conditions
Half sine wave
5gn
30 ms
Before and after test, body mounted (on chassis)
18 (3 in each direction for every axis)
Sweep sine, 10-100Hz, speed 0.05Hz/s
1 mm
3g, 3 loops 30min one cycle, 3h total, every axis tested
Before and after test, body mounted (on chassis)
Safety Specifications
Parameters
Agency approvals
Standards
www.aimtec.com
F 052e R8.D
cULus, CE, CB
Medical Electrical Equipment
Information technology Equipment
EMI - Conducted and radiated emission
Harmonic Current Emissions
Voltage fluctuations and flicker
Electrostatic Discharge Immunity
RF, Electromagnetic Field Immunity
Electrical Fast Transient/Burst Immunity
Surge Immunity
RF, Conducted Disturbance Immunity
Power frequency Magnetic Field Immunity
Voltage dips, Short Interruptions Immunity
IEC\EN\UL 60601-1, CSA-C22.2 No. 601.1-M90
EN 60950-1:2006+A11:2009
EN55011, class B
IEC/EN 61000-3-2, (EN60555-2)
IEC/EN 61000-3-3, (EN60555-3)
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-8
IEC 61000-4-11
Tel: +1 514 620 2722
2 of 3
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only
Series AME30-MAZ
30 Watt | AC-DC Converter
Pin Out Specifications
Pin
Single
Dual
1
2
3
4
5
AC Input (N)
AC Input (L)
+V Output
-V Output
N. C.
AC Input (N)
AC Input (L)
+V Output
Common
-V Output
Dimensions
Free Air Convection
Derating
55?C
Output Power (%)
100
NOTE: 1. Datasheets are updated as needed and as such, specifications
are subject to change without notice. Once printed or downloaded,
datasheets are no longer controlled by Aimtec; refer to www.aimtec.com for
the most current product specifications. 2. Product labels shown, including
safety agency certifications on labels, may vary based on the date
manufactured. 3. Mechanical drawings and specifications are for reference
only. 4. All specifications are measured at an ambient temperature of 25°C,
humidity<75%, nominal input voltage and at rated output load unless
otherwise specified. 5. Aimtec may not have conducted destructive testing
or chemical analysis on all internal components and chemicals at the time of
publishing this document. CAS numbers and other limited information are
considered proprietary and may not be available for release. 6. This
product is not designed for use in critical life support systems, equipment
used in hazardous environments, nuclear control systems or other such
applications which necessitate specific safety and regulatory standards
other the ones listed in this datasheet. 7. Warranty is in accordance with
Aimtec’s standard Terms of Sale available at www.aimtec.com.
80
Safe Operating Area
60
40
20
0
- 40
- 20
0
20
40
60
80
85
100
Ambient Temperature ?C
www.aimtec.com
F 052e R8.D
Tel: +1 514 620 2722
3 of 3
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only