RLAT 200K Report_RH117K_Fab Lot WP1058.1 W5.pdf

RLAT Report
09-413 090924 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Radiation Lot Acceptance Testing (RLAT) of the RH117K-Positive
Adjustable Regulator for Linear Technology
Customer: Linear Technology (PO# 53844L)
RAD Job Number: 09-413
Part Type Tested: Linear Technology RH117K Positive Adjustable Regulator
Commercial Part Number: RH117K
Traceability Information: Lot Date Code: 0926A, Assembly Lot# 533035.1, FAB Lot# WP1058.1, Wafer 5.
Information obtained from Linear Technology PO# 53844L. See photograph of unit under test in Appendix A.
Quantity of Units: 12 units total, 5 units for biased irradiation, 5 units for unbiased irradiation and 2 control
units. Serial numbers 108 to 112 were biased during irradiation, serial numbers 114 to 118 were unbiased during
irradiation and serial numbers 119 and 121 were used as controls. See Appendix B for the radiation bias
connection table.
Pre-Irradiation Burn-In: Burn-In performed by Linear Technology prior to receipt by RAD.
TID Dose Rate and Test Increments: 50-300rad(Si)/s with readings at pre-irradiation, 20, 50, 100, and
200krad(Si). Note that the LINEAR TECHNOLOGY datasheet guarantees post-irradiation performance to only
the 100krad(Si) dose level. Testing to 200krad(Si), as reported herein is an overtest of the datasheet guaranteed
radiation performance specifications.
TID Overtest and Post-Irradiation Anneal: No overtest or anneal.
TID Test Standard: MIL-STD-883G, Method 1019.7, Condition A
TID Electrical Test Conditions: Pre-irradiation, and within one hour following each radiation exposure.
Test Hardware: LTS2020 Tester, 2101 Family Board, 0606 Fixture and RH117 DUT Board
Facility and Radiation Source: Radiation Assured Devices Longmire Laboratories, Colorado Springs, CO using
the JLSA 81-24 high dose rate Co60 source. Dosimetry performed by CaF TLDs traceable to NIST. RAD’s
dosimetry has been audited by DSCC and RAD has been awarded Laboratory Suitability for MIL-STD-750 TM
1019.5
Irradiation and Test Temperature: Ambient room temperature for irradiation and test controlled to 24°C ±
6°C.
RLAT Result: PASSED. The units-under-test passed to the maximum
tested dose level of 200krad(Si) with no significant degradation observed
on any measured parameter.
An ISO 9001:2000 Certified Company
1
RLAT Report
09-413 090924 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
1.0. Overview and Background
It is well known that total dose ionizing radiation can cause parametric degradation and ultimately
functional failure in electronic devices. The damage occurs via electron-hole pair production, transport
and trapping in the dielectric and interface regions. In discrete devices the bulk of the damage is
frequently manifested as a reduction in the gain and/or breakdown voltage of the device. The damage
will usually anneal with time following the end of the radiation exposure. Due to this annealing, and to
ensure a worst-case test condition MIL-STD-883 TM1019.7 calls out a dose rate of 50 to 300rad(Si)/s as
Condition A and further specifies that the time from the end of an incremental radiation exposure and
electrical testing shall be 1-hour or less and the total time from the end of one incremental irradiation to
the beginning of the next incremental radiation step should be 2-hours or less. The work described in
this report was performed to meet MIL-STD-883 TM1019.7 Condition A.
2.0. Radiation Test Apparatus
The total ionizing dose testing described in this final report was performed using the facilities at
Radiation Assured Devices’ Longmire Laboratories in Colorado Springs, CO. The high dose rate total
ionizing dose (TID) source is a JLSA 84-21 irradiator modified to provide a panoramic exposure. The
Co-60 rods are held in the base of the irradiator heavily shielded by lead, during the radiation exposures
the rod is raised by an electronic timer/controller and the exposure is performed in air. The dose rate for
this irradiator in this configuration ranges from <1rad(Si)/s to a maximum of approximately
120rad(Si)/s, determined by the distance from the source. For high-dose rate experiments the bias
boards are placed in a radial fashion equidistant from the raised Co-60 rods with the distance adjusted to
provide the required dose rate. The irradiator calibration is maintained by Radiation Assured Devices
Longmire Laboratories using thermoluminescent dosimeters (TLDs)) traceable to the National Institute
of Standards and Technology (NIST). Figure 2.1 shows a photograph of the JLSA 81-24 Co-60
irradiator at RAD’s Longmire Laboratory facility.
RAD is currently certified by the Defense Supply Center Columbus (DSCC) for Laboratory Suitability
under MIL STD 750. Additional details regarding Radiation Assured Devices dosimetry for TM1019
Condition A testing are available in RAD’s report to DSCC entitled: “Dose Rate Mapping of the J.L.
Shepherd and Associates Model 81 Irradiator Installed by Radiation Assured Devices”
An ISO 9001:2000 Certified Company
2
RLAT Report
09-413 090924 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Figure 2.1. Radiation Assured Devices’ high dose rate Co-60 irradiator. The dose rate is obtained by
positioning the device-under-test at a fixed distance from the gamma cell. The dose rate for this
irradiator varies from approximately 120rad(Si)/s close to the rods down to 1rad(Si)/s at a distance of
approximately 2-feet.
An ISO 9001:2000 Certified Company
3
RLAT Report
09-413 090924 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
3.0. Radiation Test Conditions
The RH117K positive adjustable regulator described in this final report was irradiated using a split 15V
supply and with all pins tied to ground, that is biased and unbiased. See the TID Bias Table in Appendix
B for the full bias circuits. In our opinion, these bias circuits satisfy the requirements of MIL-STD883G TM1019.7 Section 3.9.3 Bias and Loading Conditions which states “The bias applied to the test
devices shall be selected to produce the greatest radiation induced damage or the worst-case damage for
the intended application, if known. While maximum voltage is often worst case some bipolar linear
device parameters (e.g. input bias current or maximum output load current) exhibit more degradation
with 0 V bias.”
The devices were irradiated to a maximum total ionizing dose level of 200krad(Si) with incremental
readings at 20krad(Si), 50krad(Si) and 100krad(Si). Electrical testing occurred within one hour
following the end of each irradiation segment. For the intermediate irradiations, the parts were tested
and returned to total dose exposure within two hours from the end of the previous radiation increment.
The TID bias board was positioned in the Co-60 cell to provide the required minimum of 50rad(Si)/s
and was located inside a lead-aluminum enclosure. The lead-aluminum enclosure is required under
MIL-STD-883G TM1019.7 Section 3.4 that reads as follows: “Lead/Aluminum (Pb/Al) container. Test
specimens shall be enclosed in a Pb/Al container to minimize dose enhancement effects caused by lowenergy, scattered radiation. A minimum of 1.5 mm Pb, surrounding an inner shield of at least 0.7 mm
Al, is required. This Pb/Al container produces an approximate charged particle equilibrium for Si and
for TLDs such as CaF2. The radiation field intensity shall be measured inside the Pb/Al container (1)
initially, (2) when the source is changed, or (3) when the orientation or configuration of the source,
container, or test-fixture is changed. This measurement shall be performed by placing a dosimeter (e.g.,
a TLD) in the device-irradiation container at the approximate test-device position. If it can be
demonstrated that low energy scattered radiation is small enough that it will not cause dosimetry errors
due to dose enhancement, the Pb/Al container may be omitted”.
The final dose rate within the high dose rate lead-aluminum enclosure was determined based on TLD
dosimetry measurements (see previous section). The final dose rate for this work was 84.9rad(Si)/s with
a precision of ±5%.
An ISO 9001:2000 Certified Company
4
RLAT Report
09-413 090924 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
4.0. Tested Parameters
During the radiation lot acceptance testing the following pre- and post-irradiation electrical parameters
were measured:
1.
2.
3.
4.
5.
6.
7.
8.
9.
Reference Voltage
Line Regulation
Load Regulation VOUT≤5V
Load Regulation VOUT≥5V
Adjust Pin Current
Adjust Pin Current Change
Minimum Load Current
Current Limit VDIFF≤15V
Current Limit VDIFF=40V
The parametric data was obtained as read and record and all the raw data plus an attributes summary are
contained in a separate Excel file. The attributes data contains the average, standard deviation and the
average with the KTL values applied. The KTL value used in this work is 2.742 per MIL-HDBK-814
using one sided tolerance limits of 90/90 and a 5-piece sample size. The 90/90 KTL values were
selected to match the statistical levels specified in the MIL-PRF-38535 sampling plan for the
qualification of a radiation hardness assured (RHA) component. Note that the following criteria must be
met for a device to pass the RLAT: following the radiation exposure each of the 5 pieces shall pass the
specification value and the average value for the ten-piece sample must pass the specification value
when the KTL limits are applied. If either of these conditions is not satisfied following the radiation
exposure, then the lot could be logged as a failure.
5.0. Total Ionizing Dose Test Results
The RH117K positive adjustable regulator PASSED the RLAT to the maximum tested dose level of
200krad(Si) with none of the measured parameters showing any significant degradation. Figures 5.1 –
5.15 show plots of all the measured parameters versus total ionizing dose while Tables 5.1 – 5.15 show
the corresponding raw data for each of these parameters. Appendix D lists all the figures used in this
section for convenience.
In the data plots the solid diamonds are the average of the measured data points for the sample irradiated
under electrical bias while the shaded diamonds are the average of the measured data points for the units
irradiated with all pins tied to ground. The black lines (solid or dashed) are the average of the data
points after application of the KTL statistics on the sample irradiated in the biased condition while the
shaded lines (solid or dashed) are the average of the data points after application of the KTL statistics on
the sample irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-
An ISO 9001:2000 Certified Company
5
RLAT Report
09-413 090924 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
The control units, as expected, show no significant changes to any of the parameters. Therefore we can
conclude that the electrical testing remained in control throughout the duration of the tests and any
observed degradation was due to the radiation exposure.
An ISO 9001:2000 Certified Company
6
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Reference Voltage VDIFF=3V IL=10mA (V)
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
1.320
1.300
1.280
1.260
1.240
1.220
1.200
1.180
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.1. Plot of Reference Voltage VDIFF=3V IL=10mA (V) versus total dose. The data show no
significant change with radiation. The solid diamonds are the average of the measured data points for the
samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points
for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of
the data points after application of the KTL statistics on the samples irradiated under electrical bias while the
gray lines (solid and/or dashed) are the average of the data points after application of the KTL statistics on the
samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation
minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
7
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.1. Raw data for Reference Voltage VDIFF=3V IL=10mA (V) versus total dose,
including the statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Reference Voltage VDIFF=3V IL=10mA (V)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
1.249
1.256
1.251
1.247
1.246
1.249
1.251
1.248
1.250
1.245
1.251
1.250
20
1.249
1.241
1.241
1.242
1.246
1.249
1.248
1.248
1.248
1.245
1.251
1.250
50
1.248
1.238
1.237
1.240
1.246
1.234
1.239
1.248
1.245
1.245
1.252
1.249
100
1.248
1.238
1.237
1.240
1.236
1.234
1.239
1.248
1.245
1.241
1.252
1.249
200
1.249
1.249
1.228
1.239
1.235
1.233
1.235
1.246
1.245
1.224
1.252
1.250
1.250
0.004
1.261
1.239
1.244
0.004
1.254
1.234
1.242
0.005
1.255
1.228
1.240
0.005
1.253
1.227
1.240
0.009
1.265
1.215
1.249
1.248
1.242
1.241
1.237
0.002
0.002
0.006
0.005
0.009
1.255
1.252
1.258
1.256
1.262
1.242
1.243
1.227
1.227
1.212
1.200
1.200
1.200
1.200
1.200
PASS
PASS
PASS
PASS
PASS
1.300
1.300
1.300
1.300
1.300
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
8
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Reference Voltage VDIFF=40V IL=10mA (V)
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
1.320
1.300
1.280
1.260
1.240
1.220
1.200
1.180
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.2. Plot of Reference Voltage VDIFF=40V IL=10mA (V) versus total dose. The data show no
significant change with radiation. The solid diamonds are the average of the measured data points for the
samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points
for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of
the data points after application of the KTL statistics on the samples irradiated under electrical bias while the
gray lines (solid and/or dashed) are the average of the data points after application of the KTL statistics on the
samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation
minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
9
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.2. Raw data for Reference Voltage VDIFF=40V IL=10mA (V) versus total dose,
including the statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Reference Voltage VDIFF=40V IL=10mA (V)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
1.250
1.257
1.252
1.248
1.247
1.250
1.251
1.249
1.250
1.246
1.252
1.251
20
1.250
1.257
1.252
1.248
1.245
1.249
1.257
1.251
1.250
1.256
1.251
1.251
50
1.249
1.256
1.252
1.246
1.245
1.249
1.251
1.251
1.247
1.246
1.252
1.251
100
1.249
1.255
1.250
1.245
1.242
1.248
1.251
1.250
1.248
1.244
1.252
1.252
200
1.249
1.254
1.250
1.243
1.240
1.248
1.250
1.249
1.245
1.242
1.252
1.251
1.251
0.004
1.262
1.240
1.250
0.005
1.263
1.238
1.250
0.005
1.262
1.237
1.248
0.005
1.262
1.235
1.247
0.006
1.263
1.232
1.249
1.252
1.249
1.248
1.247
0.002
0.003
0.002
0.003
0.003
1.254
1.262
1.255
1.256
1.256
1.244
1.243
1.243
1.241
1.238
1.200
1.200
1.200
1.200
1.200
PASS
PASS
PASS
PASS
PASS
1.300
1.300
1.300
1.300
1.300
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
10
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Reference Voltage VDIFF=3V IL=1.5A (V)
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
1.320
1.300
1.280
1.260
1.240
1.220
1.200
1.180
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.3. Plot of Reference Voltage VDIFF=3V IL=1.5A (V) versus total dose. The data show no
significant change with radiation. The solid diamonds are the average of the measured data points for the
samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points
for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of
the data points after application of the KTL statistics on the samples irradiated under electrical bias while the
gray lines (solid and/or dashed) are the average of the data points after application of the KTL statistics on the
samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation
minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
11
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.3. Raw data for Reference Voltage VDIFF=3V IL=1.5A (V) versus total dose,
including the statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Reference Voltage VDIFF=3V IL=1.5A (V)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
1.249
1.256
1.252
1.248
1.246
1.248
1.252
1.248
1.250
1.245
1.252
1.250
20
1.248
1.261
1.252
1.248
1.251
1.248
1.252
1.250
1.250
1.242
1.252
1.249
50
1.247
1.256
1.250
1.245
1.250
1.243
1.242
1.250
1.250
1.242
1.252
1.250
100
1.247
1.251
1.250
1.246
1.246
1.245
1.243
1.248
1.248
1.240
1.251
1.250
200
1.244
1.250
1.249
1.241
1.241
1.240
1.240
1.247
1.241
1.240
1.252
1.250
1.250
0.004
1.261
1.240
1.252
0.005
1.267
1.237
1.250
0.004
1.261
1.238
1.248
0.002
1.254
1.242
1.245
0.004
1.257
1.233
1.249
1.248
1.245
1.245
1.242
0.003
0.004
0.004
0.003
0.003
1.256
1.259
1.257
1.254
1.250
1.241
1.238
1.234
1.235
1.233
1.200
1.200
1.200
1.200
1.200
PASS
PASS
PASS
PASS
PASS
1.300
1.300
1.300
1.300
1.300
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
12
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Reference Voltage VDIFF=40V IL=0.3A (V)
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
1.320
1.300
1.280
1.260
1.240
1.220
1.200
1.180
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.4. Plot of Reference Voltage VDIFF=40V IL=0.3A (V) versus total dose. The data show no
significant change with radiation. The solid diamonds are the average of the measured data points for the
samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points
for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of
the data points after application of the KTL statistics on the samples irradiated under electrical bias while the
gray lines (solid and/or dashed) are the average of the data points after application of the KTL statistics on the
samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation
minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
13
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.4. Raw data for Reference Voltage VDIFF=40V IL=0.3A (V) versus total dose,
including the statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Reference Voltage VDIFF=40V IL=0.3A (V)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
1.250
1.257
1.251
1.248
1.246
1.250
1.252
1.248
1.251
1.246
1.252
1.251
20
1.251
1.261
1.251
1.251
1.252
1.251
1.251
1.251
1.251
1.246
1.251
1.251
50
1.250
1.255
1.248
1.248
1.246
1.251
1.250
1.248
1.250
1.244
1.252
1.251
100
1.250
1.254
1.246
1.245
1.246
1.249
1.250
1.248
1.249
1.245
1.252
1.250
200
1.247
1.252
1.241
1.241
1.241
1.249
1.250
1.247
1.249
1.244
1.252
1.250
1.250
0.004
1.262
1.239
1.253
0.004
1.265
1.241
1.249
0.003
1.259
1.240
1.248
0.004
1.259
1.238
1.244
0.005
1.258
1.231
1.249
1.250
1.249
1.248
1.248
0.002
0.002
0.003
0.002
0.002
1.256
1.256
1.256
1.253
1.254
1.243
1.244
1.241
1.243
1.241
1.200
1.200
1.200
1.200
1.200
PASS
PASS
PASS
PASS
PASS
1.300
1.300
1.300
1.300
1.300
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
14
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Ps90%/90% (+KTL) Biased
3.50E-02
Line Regulation (%/V)
3.00E-02
2.50E-02
2.00E-02
1.50E-02
1.00E-02
5.00E-03
0.00E+00
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.5. Plot of Line Regulation (%/V) versus total dose. The data show no significant change with
radiation. The solid diamonds are the average of the measured data points for the samples irradiated under
electrical bias while the shaded diamonds are the average of the measured data points for the samples
irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of the data points
after application of the KTL statistics on the samples irradiated under electrical bias while the gray lines (solid
and/or dashed) are the average of the data points after application of the KTL statistics on the samples
irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum
and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
15
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.5. Raw data for Line Regulation (%/V) versus total dose, including the statistical
analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Line Regulation (%/V)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
1.80E-03
1.40E-03
1.10E-03
1.60E-03
1.60E-03
1.50E-03
1.50E-03
1.10E-03
1.30E-03
1.60E-03
1.30E-03
9.00E-04
20
1.50E-03
1.80E-03
1.20E-03
1.70E-03
1.70E-03
1.30E-03
1.10E-03
1.70E-03
1.10E-03
1.60E-03
1.20E-03
1.60E-03
50
1.30E-03
1.10E-03
1.10E-03
1.50E-03
1.20E-03
1.50E-03
1.20E-03
1.40E-03
1.70E-03
1.60E-03
8.00E-04
1.10E-03
100
1.60E-03
1.80E-03
1.90E-03
1.90E-03
2.10E-03
2.10E-03
1.30E-03
1.10E-03
1.90E-03
1.60E-03
1.30E-03
1.50E-03
200
2.00E-03
1.90E-03
5.70E-03
2.70E-03
7.20E-03
2.50E-03
2.50E-03
2.70E-03
4.10E-03
2.20E-03
1.80E-03
1.50E-03
1.50E-03
2.65E-04
2.23E-03
7.75E-04
1.58E-03
2.39E-04
2.23E-03
9.25E-04
1.24E-03
1.67E-04
1.70E-03
7.81E-04
1.86E-03
1.82E-04
2.36E-03
1.36E-03
3.90E-03
2.41E-03
1.05E-02
-2.70E-03
1.40E-03
2.00E-04
1.95E-03
8.52E-04
2.00E-02
PASS
1.36E-03
2.79E-04
2.13E-03
5.94E-04
2.00E-02
PASS
1.48E-03
1.92E-04
2.01E-03
9.53E-04
2.00E-02
PASS
1.60E-03
4.12E-04
2.73E-03
4.69E-04
3.00E-02
PASS
2.80E-03
7.48E-04
4.85E-03
7.48E-04
3.00E-02
PASS
An ISO 9001:2000 Certified Company
16
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Ps90%/90% (+KTL) Biased
Load Regulation VOUT<=5V (mV)
6.80E+01
5.80E+01
4.80E+01
3.80E+01
2.80E+01
1.80E+01
8.00E+00
-2.00E+00
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.6. Plot of Load Regulation VOUT<=5V (mV) versus total dose. The data show no significant
change with radiation. The solid diamonds are the average of the measured data points for the samples
irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the
samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of the
data points after application of the KTL statistics on the samples irradiated under electrical bias while the gray
lines (solid and/or dashed) are the average of the data points after application of the KTL statistics on the
samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation
minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
17
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.6. Raw data for Load Regulation VOUT<=5V (mV) versus total dose, including the
statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Load Regulation VOUT<=5V (mV)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
-5.39E-04
1.79E-03
2.06E-03
1.67E-03
9.10E-04
-2.19E-04
1.06E-03
-1.80E-03
7.75E-04
-3.54E-04
-5.10E-05
1.70E-03
20
-2.66E-03
2.02E-04
-3.37E-04
-1.70E-05
-2.36E-04
5.22E-04
1.85E-04
2.53E-04
-3.20E-04
4.21E-04
-2.86E-04
8.26E-04
50
-1.70E-05
6.07E-04
-5.10E-05
4.72E-04
5.22E-04
5.05E-04
8.09E-04
4.38E-04
1.11E-03
6.57E-04
-1.70E-05
8.59E-04
100
3.20E-04
3.54E-04
8.09E-04
8.76E-04
4.55E-04
1.13E-03
5.39E-04
1.01E-03
2.53E-04
3.88E-04
2.86E-04
2.02E-04
200
7.08E-04
1.89E-03
1.03E-03
1.23E-03
1.35E-03
5.05E-04
4.04E-04
1.16E-03
1.42E-03
4.72E-04
-4.72E-04
-2.02E-04
1.18E-03
1.05E-03
4.05E-03
-1.70E-03
-6.10E-04
1.17E-03
2.59E-03
-3.81E-03
3.07E-04
3.15E-04
1.17E-03
-5.57E-04
5.63E-04
2.61E-04
1.28E-03
-1.53E-04
1.24E-03
4.35E-04
2.43E-03
4.63E-05
-1.08E-04
1.13E-03
2.99E-03
-3.20E-03
1.50E+01
PASS
2.12E-04
3.26E-04
1.11E-03
-6.82E-04
4.20E+01
PASS
7.04E-04
2.69E-04
1.44E-03
-3.42E-05
4.80E+01
PASS
6.64E-04
3.86E-04
1.72E-03
-3.96E-04
6.00E+01
PASS
7.92E-04
4.64E-04
2.06E-03
-4.80E-04
6.00E+01
PASS
An ISO 9001:2000 Certified Company
18
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Ps90%/90% (+KTL) Biased
Load Regulation VOUT>=5V (%)
1.40E+00
1.20E+00
1.00E+00
8.00E-01
6.00E-01
4.00E-01
2.00E-01
0.00E+00
-2.00E-01
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.7. Plot of Load Regulation VOUT>=5V (%) versus total dose. The data show no significant change
with radiation. The solid diamonds are the average of the measured data points for the samples irradiated
under electrical bias while the shaded diamonds are the average of the measured data points for the samples
irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of the data points
after application of the KTL statistics on the samples irradiated under electrical bias while the gray lines (solid
and/or dashed) are the average of the data points after application of the KTL statistics on the samples
irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum
and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
19
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.7. Raw data for Load Regulation VOUT>=5V (%) versus total dose, including the
statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Load Regulation VOUT>=5V (%)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
-1.26E-01
-8.10E-02
-9.00E-03
-1.72E-01
-7.40E-02
-9.40E-02
-9.10E-02
-7.80E-02
-1.21E-01
-4.50E-02
-8.80E-02
-1.76E-01
20
-8.00E-02
-5.80E-02
-4.10E-02
-8.10E-02
-7.10E-02
-1.13E-01
-6.70E-02
-9.90E-02
-9.30E-02
-9.40E-02
-3.70E-02
-8.80E-02
50
-1.00E-01
-1.00E-01
-7.70E-02
-1.00E-01
-8.50E-02
-6.20E-02
-6.50E-02
-5.60E-02
-4.80E-02
-7.80E-02
-4.60E-02
-6.90E-02
100
-1.08E-01
-5.10E-02
-8.10E-02
-1.60E-02
-4.70E-02
-5.00E-02
-6.60E-02
-1.00E-01
-6.10E-02
-6.70E-02
-4.80E-02
-5.60E-02
200
-1.15E-01
-7.30E-02
-1.34E-01
-8.10E-02
-1.18E-01
-7.10E-02
-7.70E-02
-8.10E-02
-1.14E-01
-9.80E-02
-5.80E-02
-4.30E-02
-9.24E-02
6.10E-02
7.49E-02
-2.60E-01
-6.62E-02
1.68E-02
-2.00E-02
-1.12E-01
-9.24E-02
1.08E-02
-6.28E-02
-1.22E-01
-6.06E-02
3.51E-02
3.57E-02
-1.57E-01
-1.04E-01
2.60E-02
-3.29E-02
-1.76E-01
-8.58E-02
2.77E-02
-9.97E-03
-1.62E-01
3.00E-01
PASS
-9.32E-02
1.67E-02
-4.75E-02
-1.39E-01
8.40E-01
PASS
-6.18E-02
1.11E-02
-3.12E-02
-9.24E-02
9.60E-01
PASS
-6.88E-02
1.87E-02
-1.75E-02
-1.20E-01
1.20E+00
PASS
-8.82E-02
1.76E-02
-4.00E-02
-1.36E-01
1.20E+00
PASS
An ISO 9001:2000 Certified Company
20
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Adjust Pin Current VDIFF=2.5V IL=10mA (A)
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Ps90%/90% (+KTL) Biased
1.20E-04
1.00E-04
8.00E-05
6.00E-05
4.00E-05
2.00E-05
0.00E+00
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.8. Plot of Adjust Pin Current VDIFF=2.5V IL=10mA (A) versus total dose. The data show no
significant change with radiation. The solid diamonds are the average of the measured data points for the
samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points
for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of
the data points after application of the KTL statistics on the samples irradiated under electrical bias while the
gray lines (solid and/or dashed) are the average of the data points after application of the KTL statistics on the
samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation
minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
21
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.8. Raw data for Adjust Pin Current VDIFF=2.5V IL=10mA (A) versus total dose,
including the statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Adjust Pin Current VDIFF=2.5V IL=10mA (A)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
4.01E-05
4.04E-05
3.84E-05
4.02E-05
4.06E-05
4.07E-05
3.95E-05
3.93E-05
4.04E-05
3.91E-05
4.01E-05
4.01E-05
20
3.94E-05
3.98E-05
3.83E-05
3.97E-05
4.06E-05
3.98E-05
3.91E-05
3.89E-05
4.06E-05
3.86E-05
3.96E-05
3.96E-05
50
3.93E-05
4.03E-05
3.83E-05
3.94E-05
4.11E-05
3.98E-05
3.91E-05
3.86E-05
4.00E-05
3.89E-05
4.06E-05
4.00E-05
100
3.97E-05
4.07E-05
3.86E-05
3.97E-05
4.05E-05
4.02E-05
3.93E-05
3.90E-05
4.01E-05
3.85E-05
4.04E-05
4.02E-05
200
3.89E-05
3.97E-05
3.76E-05
3.89E-05
3.96E-05
3.92E-05
3.81E-05
3.80E-05
3.92E-05
3.80E-05
4.01E-05
3.98E-05
3.99E-05
8.78E-07
4.23E-05
3.75E-05
3.96E-05
8.16E-07
4.18E-05
3.73E-05
3.97E-05
1.07E-06
4.26E-05
3.67E-05
3.99E-05
8.42E-07
4.22E-05
3.76E-05
3.89E-05
8.36E-07
4.12E-05
3.66E-05
3.98E-05
6.99E-07
4.17E-05
3.79E-05
1.00E-04
PASS
3.94E-05
7.91E-07
4.16E-05
3.72E-05
1.00E-04
PASS
3.93E-05
5.92E-07
4.09E-05
3.76E-05
1.00E-04
PASS
3.94E-05
7.48E-07
4.15E-05
3.74E-05
1.00E-04
PASS
3.85E-05
6.13E-07
4.02E-05
3.68E-05
1.00E-04
PASS
An ISO 9001:2000 Certified Company
22
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Ps90%/90% (+KTL) Biased
Adjust Pin Current VDIFF=5V IL=10mA (A)
1.20E-04
1.00E-04
8.00E-05
6.00E-05
4.00E-05
2.00E-05
0.00E+00
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.9. Plot of Adjust Pin Current VDIFF=5V IL=10mA (A) versus total dose. The data show no
significant change with radiation. The solid diamonds are the average of the measured data points for the
samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points
for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of
the data points after application of the KTL statistics on the samples irradiated under electrical bias while the
gray lines (solid and/or dashed) are the average of the data points after application of the KTL statistics on the
samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation
minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
23
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.9. Raw data for Adjust Pin Current VDIFF=5V IL=10mA (A) versus total dose,
including the statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Adjust Pin Current VDIFF=5V IL=10mA (A)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
4.01E-05
4.05E-05
3.89E-05
3.97E-05
4.09E-05
4.05E-05
3.97E-05
3.92E-05
4.09E-05
3.92E-05
4.01E-05
4.00E-05
20
3.92E-05
4.03E-05
3.84E-05
3.97E-05
4.05E-05
4.00E-05
3.90E-05
3.89E-05
4.02E-05
3.88E-05
3.95E-05
3.97E-05
50
3.95E-05
4.05E-05
3.85E-05
3.94E-05
4.11E-05
3.99E-05
3.90E-05
3.89E-05
3.99E-05
3.89E-05
4.05E-05
4.00E-05
100
3.97E-05
4.04E-05
3.85E-05
3.98E-05
4.05E-05
4.00E-05
3.93E-05
3.89E-05
4.04E-05
3.88E-05
4.02E-05
4.00E-05
200
3.89E-05
3.95E-05
3.78E-05
3.89E-05
3.98E-05
3.92E-05
3.81E-05
3.81E-05
3.93E-05
3.82E-05
4.01E-05
3.99E-05
4.00E-05
7.55E-07
4.21E-05
3.79E-05
3.96E-05
8.41E-07
4.19E-05
3.73E-05
3.98E-05
1.02E-06
4.26E-05
3.70E-05
3.98E-05
7.96E-07
4.19E-05
3.76E-05
3.90E-05
7.52E-07
4.10E-05
3.69E-05
3.99E-05
7.77E-07
4.20E-05
3.77E-05
1.00E-04
PASS
3.94E-05
6.50E-07
4.12E-05
3.76E-05
1.00E-04
PASS
3.93E-05
5.49E-07
4.08E-05
3.78E-05
1.00E-04
PASS
3.94E-05
6.98E-07
4.14E-05
3.75E-05
1.00E-04
PASS
3.86E-05
6.20E-07
4.03E-05
3.69E-05
1.00E-04
PASS
An ISO 9001:2000 Certified Company
24
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Adjust Pin Current VDIFF=40V IL=10mA (A)
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Ps90%/90% (+KTL) Biased
1.20E-04
1.00E-04
8.00E-05
6.00E-05
4.00E-05
2.00E-05
0.00E+00
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.10. Plot of Adjust Pin Current VDIFF=40V IL=10mA (A) versus total dose. The data show no
significant change with radiation. The solid diamonds are the average of the measured data points for the
samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points
for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of
the data points after application of the KTL statistics on the samples irradiated under electrical bias while the
gray lines (solid and/or dashed) are the average of the data points after application of the KTL statistics on the
samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation
minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
25
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.10. Raw data for Adjust Pin Current VDIFF=40V IL=10mA (A) versus total dose,
including the statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Adjust Pin Current VDIFF=40V IL=10mA (A)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
4.02E-05
4.07E-05
3.88E-05
3.99E-05
4.10E-05
4.05E-05
3.97E-05
3.92E-05
4.09E-05
3.90E-05
4.03E-05
4.00E-05
20
3.94E-05
4.02E-05
3.85E-05
3.97E-05
4.07E-05
3.99E-05
3.90E-05
3.90E-05
4.05E-05
3.86E-05
3.97E-05
3.96E-05
50
3.95E-05
4.04E-05
3.83E-05
3.95E-05
4.10E-05
4.00E-05
3.90E-05
3.89E-05
4.01E-05
3.90E-05
4.05E-05
4.01E-05
100
3.97E-05
4.04E-05
3.86E-05
3.99E-05
4.06E-05
4.02E-05
3.93E-05
3.89E-05
4.03E-05
3.88E-05
4.02E-05
4.01E-05
200
3.89E-05
3.96E-05
3.80E-05
3.89E-05
3.98E-05
3.94E-05
3.82E-05
3.82E-05
3.94E-05
3.82E-05
4.05E-05
3.99E-05
4.01E-05
8.59E-07
4.25E-05
3.77E-05
3.97E-05
8.38E-07
4.20E-05
3.74E-05
3.97E-05
1.03E-06
4.26E-05
3.69E-05
3.99E-05
7.82E-07
4.20E-05
3.77E-05
3.90E-05
7.15E-07
4.10E-05
3.71E-05
3.99E-05
8.57E-07
4.22E-05
3.75E-05
1.00E-04
PASS
3.94E-05
7.99E-07
4.16E-05
3.72E-05
1.00E-04
PASS
3.94E-05
5.90E-07
4.10E-05
3.78E-05
1.00E-04
PASS
3.95E-05
7.08E-07
4.14E-05
3.76E-05
1.00E-04
PASS
3.87E-05
6.39E-07
4.04E-05
3.69E-05
1.00E-04
PASS
An ISO 9001:2000 Certified Company
26
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Adjust Pin Current Change IL=10mA-1.5A (A)
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
6.00E-06
4.00E-06
2.00E-06
0.00E+00
-2.00E-06
-4.00E-06
-6.00E-06
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.11. Plot of Adjust Pin Current Change IL=10mA-1.5A (A) versus total dose. The data show no
significant change with radiation. The solid diamonds are the average of the measured data points for the
samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points
for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of
the data points after application of the KTL statistics on the samples irradiated under electrical bias while the
gray lines (solid and/or dashed) are the average of the data points after application of the KTL statistics on the
samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation
minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
27
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.11. Raw data for Adjust Pin Current Change IL=10mA-1.5A (A) versus total dose,
including the statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Adjust Pin Current Change IL=10mA-1.5A (A)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
0.00E+00
-3.10E-07
-6.50E-07
1.80E-07
-1.30E-07
-6.20E-07
-3.60E-07
1.80E-07
-2.70E-07
2.70E-07
1.80E-07
0.00E+00
20
0.00E+00
-4.00E-08
1.80E-07
-1.80E-07
4.50E-07
4.00E-08
2.00E-08
0.00E+00
-1.80E-07
5.10E-07
2.00E-07
-4.00E-08
50
-2.90E-07
5.10E-07
3.10E-07
-2.70E-07
-2.70E-07
0.00E+00
2.70E-07
-4.00E-08
0.00E+00
5.60E-07
2.50E-07
0.00E+00
100
2.00E-08
1.60E-07
9.00E-08
-3.60E-07
2.90E-07
0.00E+00
5.30E-07
2.00E-07
-2.90E-07
-1.30E-07
1.30E-07
3.10E-07
200
-2.00E-07
2.50E-07
5.30E-07
7.00E-08
0.00E+00
4.50E-07
-4.00E-08
-2.00E-07
3.30E-07
-2.00E-08
-4.50E-07
-8.70E-07
-1.82E-07
3.17E-07
6.88E-07
-1.05E-06
8.20E-08
2.43E-07
7.47E-07
-5.83E-07
-2.00E-09
3.83E-07
1.05E-06
-1.05E-06
4.00E-08
2.45E-07
7.11E-07
-6.31E-07
1.30E-07
2.76E-07
8.86E-07
-6.26E-07
-1.60E-07
3.76E-07
8.70E-07
-1.19E-06
-5.00E-06
PASS
5.00E-06
PASS
7.80E-08
2.57E-07
7.83E-07
-6.27E-07
-5.00E-06
PASS
5.00E-06
PASS
1.58E-07
2.57E-07
8.61E-07
-5.45E-07
-5.00E-06
PASS
5.00E-06
PASS
6.20E-08
3.17E-07
9.32E-07
-8.08E-07
-5.00E-06
PASS
5.00E-06
PASS
1.04E-07
2.74E-07
8.54E-07
-6.46E-07
-5.00E-06
PASS
5.00E-06
PASS
An ISO 9001:2000 Certified Company
28
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Adjust Pin Current Change VDIFF=2.5V-40V (A)
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
6.00E-06
4.00E-06
2.00E-06
0.00E+00
-2.00E-06
-4.00E-06
-6.00E-06
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.12. Plot of Adjust Pin Current Change VDIFF=2.5V-40V (A) versus total dose. The data show no
significant change with radiation. The solid diamonds are the average of the measured data points for the
samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points
for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of
the data points after application of the KTL statistics on the samples irradiated under electrical bias while the
gray lines (solid and/or dashed) are the average of the data points after application of the KTL statistics on the
samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation
minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
29
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.12. Raw data for Adjust Pin Current Change VDIFF=2.5V-40V (A) versus total dose,
including the statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Adjust Pin Current Change VDIFF=2.5V-40V (A)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
-1.80E-07
-2.70E-07
-1.80E-07
0.00E+00
-2.70E-07
0.00E+00
-3.10E-07
-1.80E-07
7.00E-08
-2.50E-07
1.80E-07
-3.60E-07
20
-1.80E-07
-1.60E-07
4.00E-08
9.00E-08
9.00E-08
2.00E-08
-1.30E-07
4.00E-08
-1.10E-07
-2.70E-07
2.00E-08
2.00E-08
50
-1.60E-07
-2.20E-07
-2.00E-08
-9.00E-08
2.20E-07
4.00E-08
-1.30E-07
0.00E+00
-1.10E-07
-1.10E-07
0.00E+00
-1.80E-07
100
-2.50E-07
-1.80E-07
-4.00E-08
9.00E-08
-3.30E-07
-1.80E-07
9.00E-08
-2.00E-07
-1.60E-07
-2.00E-07
-1.80E-07
2.00E-08
200
7.00E-08
-7.00E-08
-9.00E-08
2.00E-08
-4.00E-08
-2.00E-07
-9.00E-08
-4.00E-08
-1.10E-07
0.00E+00
2.00E-08
-4.00E-08
-1.80E-07
1.10E-07
1.22E-07
-4.82E-07
-2.40E-08
1.35E-07
3.46E-07
-3.94E-07
-5.40E-08
1.71E-07
4.14E-07
-5.22E-07
-1.42E-07
1.68E-07
3.18E-07
-6.02E-07
-2.20E-08
6.61E-08
1.59E-07
-2.03E-07
-1.34E-07
1.63E-07
3.13E-07
-5.81E-07
-5.00E-06
PASS
5.00E-06
PASS
-9.00E-08
1.26E-07
2.55E-07
-4.35E-07
-5.00E-06
PASS
5.00E-06
PASS
-6.20E-08
7.66E-08
1.48E-07
-2.72E-07
-5.00E-06
PASS
5.00E-06
PASS
-1.30E-07
1.24E-07
2.10E-07
-4.70E-07
-5.00E-06
PASS
5.00E-06
PASS
-8.80E-08
7.60E-08
1.20E-07
-2.96E-07
-5.00E-06
PASS
5.00E-06
PASS
An ISO 9001:2000 Certified Company
30
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Ps90%/90% (+KTL) Biased
6.00E-03
Minimum Load Current (A)
5.00E-03
4.00E-03
3.00E-03
2.00E-03
1.00E-03
0.00E+00
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.13. Plot of Minimum Load Current (A) versus total dose. The data show no significant change with
radiation. The solid diamonds are the average of the measured data points for the samples irradiated under
electrical bias while the shaded diamonds are the average of the measured data points for the samples
irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of the data points
after application of the KTL statistics on the samples irradiated under electrical bias while the gray lines (solid
and/or dashed) are the average of the data points after application of the KTL statistics on the samples
irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum
and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
31
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.13. Raw data for Minimum Load Current (A) versus total dose, including the
statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Minimum Load Current (A)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
1.86E-03
1.92E-03
1.88E-03
1.92E-03
2.02E-03
1.90E-03
1.85E-03
1.92E-03
1.96E-03
1.96E-03
1.94E-03
1.87E-03
20
1.85E-03
1.92E-03
1.73E-03
1.81E-03
1.90E-03
1.79E-03
1.73E-03
1.83E-03
1.87E-03
1.87E-03
1.85E-03
1.77E-03
50
1.87E-03
1.94E-03
1.75E-03
1.85E-03
1.94E-03
1.83E-03
1.75E-03
1.85E-03
1.90E-03
1.89E-03
1.83E-03
1.77E-03
100
1.90E-03
1.96E-03
1.79E-03
1.89E-03
2.00E-03
1.85E-03
1.81E-03
1.90E-03
2.00E-03
1.94E-03
1.85E-03
1.79E-03
200
2.02E-03
2.04E-03
1.90E-03
2.04E-03
2.14E-03
1.98E-03
1.96E-03
2.02E-03
2.14E-03
2.06E-03
1.85E-03
1.79E-03
1.92E-03
6.14E-05
2.09E-03
1.75E-03
1.84E-03
7.84E-05
2.06E-03
1.63E-03
1.87E-03
8.03E-05
2.09E-03
1.65E-03
1.91E-03
8.16E-05
2.13E-03
1.68E-03
2.03E-03
8.39E-05
2.26E-03
1.80E-03
1.92E-03
4.70E-05
2.05E-03
1.79E-03
5.00E-03
PASS
1.82E-03
5.94E-05
1.98E-03
1.65E-03
5.00E-03
PASS
1.84E-03
6.08E-05
2.01E-03
1.68E-03
5.00E-03
PASS
1.90E-03
7.66E-05
2.11E-03
1.69E-03
5.00E-03
PASS
2.03E-03
6.97E-05
2.22E-03
1.84E-03
5.00E-03
PASS
An ISO 9001:2000 Certified Company
32
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Ps90%/90% (-KTL) Biased
2.40E+00
Current Limit VOUT=15V (A)
2.20E+00
2.00E+00
1.80E+00
1.60E+00
1.40E+00
1.20E+00
1.00E+00
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.14. Plot of Current Limit VOUT=15V (A) versus total dose. The data show a small increase
(improvement) with radiation. The solid diamonds are the average of the measured data points for the samples
irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the
samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of the
data points after application of the KTL statistics on the samples irradiated under electrical bias while the gray
lines (solid and/or dashed) are the average of the data points after application of the KTL statistics on the
samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation
minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
33
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.14. Raw data for Current Limit VOUT=15V (A) versus total dose, including the
statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Current Limit VOUT=15V (A)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
2.03E+00
2.07E+00
2.08E+00
2.06E+00
2.12E+00
2.06E+00
1.97E+00
2.06E+00
2.01E+00
2.06E+00
2.07E+00
2.00E+00
20
2.09E+00
2.11E+00
2.12E+00
2.11E+00
2.15E+00
2.11E+00
2.02E+00
2.11E+00
2.06E+00
2.11E+00
2.07E+00
2.00E+00
50
2.14E+00
2.16E+00
2.16E+00
2.16E+00
2.21E+00
2.15E+00
2.06E+00
2.16E+00
2.11E+00
2.16E+00
2.07E+00
2.00E+00
100
2.17E+00
2.20E+00
2.21E+00
2.19E+00
2.25E+00
2.19E+00
2.10E+00
2.20E+00
2.15E+00
2.20E+00
2.07E+00
2.00E+00
200
2.22E+00
2.24E+00
2.25E+00
2.25E+00
2.29E+00
2.24E+00
2.15E+00
2.25E+00
2.20E+00
2.25E+00
2.06E+00
2.00E+00
2.07E+00
3.05E-02
2.15E+00
1.99E+00
2.11E+00
2.45E-02
2.18E+00
2.05E+00
2.16E+00
2.64E-02
2.24E+00
2.09E+00
2.20E+00
2.69E-02
2.28E+00
2.13E+00
2.25E+00
2.46E-02
2.31E+00
2.18E+00
2.03E+00
4.24E-02
2.15E+00
1.91E+00
1.50E+00
PASS
2.08E+00
4.08E-02
2.19E+00
1.97E+00
1.50E+00
PASS
2.13E+00
4.34E-02
2.25E+00
2.01E+00
1.50E+00
PASS
2.17E+00
4.09E-02
2.28E+00
2.06E+00
1.50E+00
PASS
2.21E+00
4.41E-02
2.34E+00
2.09E+00
1.50E+00
PASS
An ISO 9001:2000 Certified Company
34
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Ps90%/90% (-KTL) Biased
8.00E-01
Current Limit VOUT=40V (A)
7.00E-01
6.00E-01
5.00E-01
4.00E-01
3.00E-01
2.00E-01
1.00E-01
0.00E+00
0
20
40
60
80
100
120
140
160
180
Total Dose (krad(Si))
Figure 5.15. Plot of Current Limit VOUT=40V (A) versus total dose. The data show an increase
(improvement) with radiation. The solid diamonds are the average of the measured data points for the samples
irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the
samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the average of the
data points after application of the KTL statistics on the samples irradiated under electrical bias while the gray
lines (solid and/or dashed) are the average of the data points after application of the KTL statistics on the
samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation
minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
35
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table 5.15. Raw data for Current Limit VOUT=40V (A) versus total dose, including the
statistical analysis, specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Current Limit VOUT=40V (A)
Device
108
109
110
111
112
114
115
116
117
118
119
121
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
4.64E-01
4.48E-01
4.70E-01
4.64E-01
4.75E-01
4.70E-01
4.31E-01
4.81E-01
4.57E-01
4.63E-01
4.69E-01
4.52E-01
20
5.18E-01
5.01E-01
5.29E-01
5.18E-01
5.29E-01
5.24E-01
4.90E-01
5.46E-01
5.13E-01
5.24E-01
4.68E-01
4.51E-01
50
5.79E-01
5.51E-01
5.85E-01
5.79E-01
5.85E-01
5.79E-01
5.35E-01
5.96E-01
5.68E-01
5.79E-01
4.68E-01
4.51E-01
100
6.29E-01
5.96E-01
6.29E-01
6.24E-01
6.40E-01
6.24E-01
5.85E-01
6.46E-01
6.24E-01
6.24E-01
4.68E-01
4.46E-01
200
6.90E-01
6.51E-01
7.01E-01
6.96E-01
7.12E-01
6.79E-01
6.57E-01
7.12E-01
6.96E-01
6.96E-01
4.73E-01
4.46E-01
4.64E-01
1.02E-02
4.92E-01
4.36E-01
5.19E-01
1.15E-02
5.50E-01
4.88E-01
5.76E-01
1.42E-02
6.15E-01
5.37E-01
6.24E-01
1.65E-02
6.69E-01
5.78E-01
6.90E-01
2.32E-02
7.54E-01
6.26E-01
4.60E-01
1.87E-02
5.12E-01
4.09E-01
3.00E-01
PASS
5.19E-01
2.03E-02
5.75E-01
4.64E-01
3.00E-01
PASS
5.71E-01
2.27E-02
6.34E-01
5.09E-01
3.00E-01
PASS
6.21E-01
2.21E-02
6.81E-01
5.60E-01
3.00E-01
PASS
6.88E-01
2.09E-02
7.45E-01
6.31E-01
3.00E-01
PASS
An ISO 9001:2000 Certified Company
36
RLAT Report
09-413 090924 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
6.0. Summary / Conclusions
The total ionizing dose testing described in this final report was performed using the facilities at
Radiation Assured Devices’ Longmire Laboratories in Colorado Springs, CO. The high dose rate total
ionizing dose (TID) source is a JLSA 84-21 irradiator modified to provide a panoramic exposure. The
Co-60 rods are held in the base of the irradiator heavily shielded by lead, during the radiation exposures
the rod is raised by an electronic timer/controller and the exposure is performed in air. The dose rate for
this irradiator in this configuration ranges from <1rad(Si)/s to a maximum of approximately
120rad(Si)/s, determined by the distance from the source.
The parametric data was obtained as “read and record” and all the raw data plus an attributes summary
were presented in this report. The attributes data contains the average, standard deviation and the
average with the KTL values applied. The KTL value used was 2.742 per MIL HDBK 814 using onesided tolerance limits of 90/90 and a 5-piece sample size. Note that the following criteria was used to
determine the outcome of the testing: following the radiation exposure each parameter had to pass the
specification value and the average value for the five-piece sample must pass the specification value
when the KTL limits are applied. If these conditions were not both satisfied following the radiation
exposure, then the lot would be logged as an RLAT failure.
Based on these criteria, the RH117K positive adjustable regulator (from the lot of material described on
the first page of this report) PASSED the RLAT to the maximum tested dose level of 200krad(Si) with
none of the measured parameters showing any significant degradation. Note that the LINEAR
TECHNOLOGY datasheet guarantees post-irradiation performance only to the 100krad(Si) dose level.
Testing to 200krad(Si), as reported herein is an overtest of the datasheet guaranteed radiation
performance specifications.
An ISO 9001:2000 Certified Company
37
RLAT Report
09-413 090924 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Appendix A: Photograph of device-under-test to show part markings
An ISO 9001:2000 Certified Company
38
RLAT Report
09-413 090924 R1.0
Appendix B: TID Bias Connections
Biased Samples:
Pin
Function Connection / Bias
1
ADJ
2kΩ to -15V
2
VIN
To 15V,
0.1μF decoupling to -15V
3
CASE
VOUT
61.9Ω to -15V
Unbiased Samples:
Pin
Function Connection / Bias
1
ADJ
GND
2
VIN
GND
3
CASE
VOUT
GND
An ISO 9001:2000 Certified Company
39
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Figure B.1. Irradiation bias drawing for the units to be irradiated under electrical bias. This figure was
extracted from LINEAR TECHNOLOGY CORPORATION, RH117 Datasheet.
Figure B.2. K package drawing (for reference only). This figure was extracted from the LINEAR
TECHNOLOGY CORPORATION RH117 Datasheet.
An ISO 9001:2000 Certified Company
40
RLAT Report
09-413 090924 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Appendix C: Electrical Test Parameters and Conditions
All electrical tests for this device are performed on one of Radiation Assured Device’s LTS2020 Test
Systems. The LTS2020 Test System is a programmable parametric tester that provides parameter
measurements for a variety of digital, analog and mixed signal products including voltage regulators,
voltage comparators, D to A and A to D converters. The LTS2020 Test System achieves accuracy and
sensitivity through the use of software self-calibration and an internal relay matrix with separate family
boards and custom personality adapter boards. The tester uses this relay matrix to connect the required
test circuits, select the appropriate voltage / current sources and establish the needed measurement loops
for all the tests performed. The tests will be conducted using the LTS-2101 Linear Family Board, LTS0606 Socket Assembly and the RH117 DUT board. The measured parameters and test conditions are
shown in Tables C.1.
A listing of the measurement precision/resolution for each parameter is shown in Tables C.2. The
precision/resolution values were obtained either from test data or from the DAC resolution of the LTS2020. To generate the precision/resolution shown in Table C.2, one of the units-under-test was tested
repetitively (a total of 10-times with re-insertion between tests) to obtain the average test value and
standard deviation. Using this test data MIL-HDBK-814 90/90 KTL statistics were applied to the
measured standard deviation to generate the final measurement range. This value encompasses the
precision/resolution of all aspects of the test system, including the LTS2020 mainframe, family board,
socket assembly and DUT board as well as insertion error. In some cases, the measurement resolution is
limited by the internal DACs, which results in a measured standard deviation of zero. In these instances
the precision/resolution will be reported back as the LSB of the DAC.
Note that the testing and statistics used in this document are based on an “analysis of variables”
technique, which relies on small sample sizes to qualify much larger lot sizes (see MIL-HDBK-814, p.
91 for a discussion of statistical treatments). Not all measured parameters are well suited to this
approach due to inherent large variations. If necessary, larger samples sizes could be used to qualify
these parameters using an “attributes” approach.
An ISO 9001:2000 Certified Company
41
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
09-413 090924 R1.0
Table C.1. Measured parameters and test conditions RH117K.
TEST DESCRIPTION
TEST CONDITIONS
VDIFF=VIN-VOUT=3V, IL=10mA
VDIFF=40V, IL=10mA
Reference Voltage
VDIFF=3V, IL=1.5A
VDIFF=40V, IL=0.3A
Line Regulation
VDIFF=3V to 40V, IL=10mA
Load Regulation VOUT≤5V VDIFF=5V, VIN=6.25V, IL=10mA to 1.5A
Load Regulation VOUT≥5V VDIFF=5V, VIN=11.25V, IL=10mA to 1.5A
VDIFF=2.5V, IL=10mA
Adjust Pin Current
VDIFF=5V, IL=10mA
VDIFF=40V, IL=10mA
VDIFF=5V, IL=10mA to 1.5A
Adjust Pin Current Change
VDIFF=2.5V to 40V, IL=10mA
Minimum Load Current
VDIFF=40V
Current Limit VDIFF≤15V
VDIFF=15V
Current Limit VDIFF=40V
VDIFF=40V
Table C.2. Measured parameters, pre-irradiation specifications and measurement resolutions
for the RH117K.
Measured Parameter
Pre-Irradiation
Specification
Measurement
Resolution/Precision
Reference Voltage
1.25V±50mV
± 1.09E-03V
± 3.40E-04%/V
2.40E-04V
4.48E-03%
2.26E-06A
4.20E-07A
2.84E-05A
5.22E-03A
6.53E-03A
Line Regulation
0.02%/V MAX
Load Regulation VOUT≤5V
15mV MAX
Load Regulation VOUT≥5V
0.3% MAX
Adjust Pin Current
100µA MAX
Adjust Pin Current Change
± 5µA MAX
Minimum Load Current
5mA MAX
Current Limit VDIFF≤15V
1.5A MAX
Current Limit VDIFF=40V
0.3A MAX
An ISO 9001:2000 Certified Company
42
RLAT Report
09-413 090924 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Appendix D: List of Figures used in Section 5 (Total Ionizing Dose Test Results):
5.1
5.2
5.3
5.4
5.5
5.6
5.7
5.8
5.9
5.10
5.11
5.12
5.13
5.14
5.15
Reference Voltage VDIFF=3V IL=10mA (V)
Reference Voltage VDIFF=40V IL=10mA (V)
Reference Voltage VDIFF=3V IL=1.5A (V)
Reference Voltage VDIFF=40V IL=0.3A (V)
Line Regulation (%/V)
Load Regulation VOUT<=5V (mV)
Load Regulation VOUT>=5V (%)
Adjust Pin Current VDIFF=2.5V IL=10mA (A)
Adjust Pin Current VDIFF=5V IL=10mA (A)
Adjust Pin Current VDIFF=40V IL=10mA (A)
Adjust Pin Current Change IL=10mA-1.5A (A)
Adjust Pin Current Change VDIFF=2.5V-40V (A)
Minimum Load Current (A)
Current Limit VOUT=15V (A)
Current Limit VOUT=40V (A)
An ISO 9001:2000 Certified Company
43