RLAT 200K Report_RH1056AMW_Fab Lot 10217302.1.pdf

TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Radiation Lot Acceptance Testing (RLAT) of the RH1056AMW Precision,
High Speed, JFET Input Operational Amplifier for Linear Technology
Customer: Linear Technology, PO# 60038L
RAD Job Number: 11-307
Part Type Tested: RH1056AMW Precision, High Speed, JFET Input Operational Amplifier.
Traceability Information: Date Code: 1046A, Fab Lot Number: 10217302.1, Wafer Number: 12,
Assembly Lot Number: 599145.1. See photograph of unit under test in Appendix A.
Quantity of Units: 12 units received, 5 units for biased irradiation, 5 units for unbiased irradiation and 2
units for control. Serial numbers 1090, 1124, 1128, 1129 and 1131 were biased during irradiation, serial
numbers 1133, 1134, 1138, 1139 and 1140 were unbiased during irradiation and serial numbers 1141
and 1143 were used as control. See Appendix B for the radiation bias connection table.
Radiation and Electrical Test Increments: 50-300rad(Si)/s ionizing radiation with electrical test
increments: pre-irradiation, 25krad(Si), 50krad(Si), 100krad(Si) and 200krad(Si).
Pre-Irradiation Burn-In: Burn-In performed by Linear Technology prior to receipt by RAD.
Overtest and Post-Irradiation Anneal: No overtest. No anneal.
Radiation Test Standard: MIL-STD-750E TM1019 and/or MIL-STD-883H TM1019 Condition A.
Test Hardware and Software: LTS2020 Automated Tester, Entity ID TS04, Calibration Date: 04-2811, Calibration Due 04-28-12. LTS2101 Family Board, Entity ID FB02. LTS0607 Test Fixture, Entity
ID TF07. BGSS991003 RH1056 DUT Board. Test Program: RH1056LT.SRC
Facility and Radiation Source: Radiation Assured Devices' Longmire Laboratories, Colorado Springs,
CO. Gamma rays provided by JLSA 81-24 Co60 source. Dosimetry performed by Air Ionization
Chamber (AIC) traceable to NIST. RAD's dosimetry has been audited by DSCC and RAD has been
awarded Laboratory Suitability for MIL-STD-750 and MIL-STD-883 TM 1019.
Irradiation and Test Temperature: Room temperature controlled to 24°C±6°C per MIL-STD-883 and
MIL-STD-750.
RLAT Result: PASSED the total ionizing dose test to the maximum tested dose
level of 200krad(Si) with all parameters remaining within their datasheet
specifications.
An ISO 9001:2008 and DSCC Certified Company
1
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
1.0. Overview and Background
It is well known that total dose ionizing radiation can cause parametric degradation and ultimately
functional failure in electronic devices. The damage occurs via electron-hole pair production, transport
and trapping in the dielectric and interface regions. In discrete devices the bulk of the damage is
frequently manifested as a reduction in the gain and/or breakdown voltage of the device. The damage
will usually anneal with time following the end of the radiation exposure. Due to this annealing, and to
ensure a worst-case test condition MIL-STD-883 TM1019.8 calls out a dose rate of 50 to 300rad(Si)/s as
Condition A and further specifies that the time from the end of an incremental radiation exposure and
electrical testing shall be 1-hour or less and the total time from the end of one incremental irradiation to
the beginning of the next incremental radiation step should be 2-hours or less. The work described in
this report was performed to meet MIL-STD-883 TM1019.8 Condition A.
2.0. Radiation Test Apparatus
The total ionizing dose testing described in this final report was performed using the facilities at
Radiation Assured Devices' Longmire Laboratories in Colorado Springs, CO. The high dose rate total
ionizing dose (TID) source is a JLSA 81-24 irradiator modified to provide a panoramic exposure. The
Co-60 rods are held in the base of the irradiator heavily shielded by lead. During the radiation exposures
the rod is raised by an electronic timer/controller and the exposure is performed in air. The dose rate for
this irradiator in this configuration ranges from <1rad(Si)/s to a maximum of approximately
120rad(Si)/s, determined by the distance from the source. For high-dose rate experiments the bias boards
are placed in a radial fashion equidistant from the raised Co-60 rods with the distance adjusted to
provide the required dose rate. The irradiator calibration is maintained by Radiation Assured Devices
Longmire Laboratories using air ionization chamber (AIC) equipment calibrated with traceability to the
National Institute of Standards and Technology (NIST). Figure 2.1 shows a photograph of the JLSA
81-24 Co-60 irradiator at RAD's Longmire Laboratory facility.
RAD is currently certified by the Defense Supply Center Columbus (DSCC) for Laboratory Suitability
under MIL STD 750 and MIL-STD-883. Additional details regarding Radiation Assured Devices
dosimetry for TM1019 Condition A testing are available in RAD's report to DSCC entitled: "Dose Rate
Mapping of the J.L. Shepherd and Associates Model 81 Irradiator Installed by Radiation Assured
Devices".
An ISO 9001:2008 and DSCC Certified Company
2
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Figure 2.1. Radiation Assured Devices' high dose rate Co-60 irradiator. The dose rate is obtained by positioning
the device-under-test at a fixed distance from the gamma cell. The dose rate for this irradiator varies from
approximately 120rad(Si)/s close to the rods down to 1rad(Si)/s at a distance of approximately 2-feet.
An ISO 9001:2008 and DSCC Certified Company
3
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
3.0. Radiation Test Conditions
The RH1056AMW JFET Input Operational Amplifier described in this final report were irradiated using
a split 15V supply and with all pins tied to ground, that is biased and unbiased. See the TID Bias Table
in Appendix B for the full bias circuits. In our opinion, this bias circuit satisfies the requirements of
MIL-STD-883H TM1019.8 Section 3.9.3 Bias and Loading Conditions which states "The bias applied
to the test devices shall be selected to produce the greatest radiation induced damage or the worst-case
damage for the intended application, if known. While maximum voltage is often worst case some
bipolar linear device parameters (e.g. input bias current or maximum output load current) exhibit more
degradation with 0 V bias."
The devices were irradiated to a maximum total ionizing dose level of 200krad(Si) with incremental
readings at 25krad(Si), 50krad(Si) and 100krad(Si). Electrical testing occurred within one hour
following the end of each irradiation segment. For intermediate irradiations, the parts were tested and
returned to total dose exposure within two hours from the end of the previous radiation increment.
The TID bias board was positioned in the Co-60 cell to provide the required minimum of 50rad(Si)/s
and was located inside a lead-aluminum enclosure. The lead-aluminum enclosure is required under MILSTD-883H TM1019.8 Section 3.4 that reads as follows: "Lead/Aluminum (Pb/Al) container. Test
specimens shall be enclosed in a Pb/Al container to minimize dose enhancement effects caused by lowenergy, scattered radiation. A minimum of 1.5 mm Pb, surrounding an inner shield of at least 0.7 mm
Al, is required. This Pb/Al container produces an approximate charged particle equilibrium for Si and
for TLDs such as CaF2. The radiation field intensity shall be measured inside the Pb/Al container (1)
initially, (2) when the source is changed, or (3) when the orientation or configuration of the source,
container, or test-fixture is changed. This measurement shall be performed by placing a dosimeter (e.g.,
a TLD) in the device-irradiation container at the approximate test-device position. If it can be
demonstrated that low energy scattered radiation is small enough that it will not cause dosimetry errors
due to dose enhancement, the Pb/Al container may be omitted."
The final dose rate within the high dose rate lead-aluminum enclosure was determined using calibration
calculations based on air ionization chamber (AIC) dosimetry performed just prior to beginning the total
dose irradiations. The final dose rate for this work was 48.91rad(Si)/s with a precision of ±5%.
An ISO 9001:2008 and DSCC Certified Company
4
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
4.0. Tested Parameters
During the total ionizing dose characterization testing the following electrical parameters were measured
pre- and post-irradiation:
1. Positive Supply Current (A)
2. Negative Supply Current (A)
3. Input Offset Voltage (V)
4. Input Offset Current (A)
5. Positive Bias Current (A)
6. Negative Bias Current (A)
7. Large Signal Voltage Gain 2K (V/mV)
8. Large Signal Voltage Gain 1K (V/mV)
9. Common Mode Rejection Ratio (dB)
10. Power Supply Rejection Ratio (dB)
11. Positive Slew Rate (V/µs)
12. Negative Slew Rate (V/µs)
13. Positive Output Voltage Swing (V)
14. Negative Output Voltage Swing (V)
Appendix C details the measured parameters, test conditions, pre-irradiation specification and
measurement resolution for each of the measurements.
The parametric data was obtained as "read and record" and all the raw data plus an attributes summary
are contained in this report as well as in a separate Excel file. The attributes data contains the average,
standard deviation and the average with the KTL values applied. The KTL value used in this work is
2.742 per MIL-HDBK-814 using one sided tolerance limits of 90/90 and a 5-piece sample size. The
90/90 KTL values were selected to match the statistical levels specified in the MIL-PRF-38535
sampling plan for the qualification of a radiation hardness assured (RHA) component. Note that the
following criteria must be met for a device to pass the total ionizing dose test: following the radiation
exposure each of the 5 pieces irradiated under electrical bias shall pass the specification value. The units
irradiated without electrical bias and the KTL statistics are included in this report for reference only. If
any of the 5 pieces irradiated under electrical bias exceed the datasheet specifications, then the lot could
be logged as a failure.
An ISO 9001:2008 and DSCC Certified Company
5
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
5.0. Total Ionizing Dose Test Results
Based on this criterion the RH1056AMW JFET Input Operational Amplifier (from the lot traceability
information provided on the first page of this test report) PASSED the total ionizing dose test to the
maximum tested dose level of 200krad(Si) with all parameters remaining within their datasheet
specifications.
Figures 5.1 through 5.14 show plots of all the measured parameters versus total ionizing dose while
Tables 5.1 - 5.14 show the corresponding raw data for each of these parameters. In the data plots the
solid diamonds are the average of the measured data points for the sample irradiated under electrical bias
while the shaded diamonds are the average of the measured data points for the units irradiated with all
pins tied to ground. The black lines (solid or dashed) are the average of the data points after application
of the KTL statistics on the sample irradiated in the biased condition while the shaded lines (solid or
dashed) are the average of the data points after application of the KTL statistics on the sample irradiated
in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or
maximum specification value as defined in the datasheet and/or test plan.
The control units, as expected, show no significant changes to any of the parameters. Therefore we can
conclude that the electrical testing remained in control throughout the duration of the tests and the
observed degradation was due to the radiation exposure. Appendix D lists the figures used in this section
to facilitate the location of a particular parameter.
An ISO 9001:2008 and DSCC Certified Company
6
TID Report
11-307 110624 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Ps90%/90% (+KTL) Biased
8.00E-03
Positive Supply Current (A)
7.00E-03
6.00E-03
5.00E-03
4.00E-03
3.00E-03
2.00E-03
1.00E-03
0.00E+00
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.1. Plot of Positive Supply Current (A) versus total dose. The solid diamonds are the average of the
measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of
the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or
dashed) are the average of the data points after application of the KTL statistics on the samples irradiated under
electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application of the
KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or postirradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
7
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Table 5.1. Raw data for Positive Supply Current (A) versus total dose, including the statistical analysis,
specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Positive Supply Current (A)
Device
0
25
50
100
200
1090 5.12E-03 4.99E-03 4.93E-03 4.83E-03 4.68E-03
1124 4.53E-03 4.45E-03 4.36E-03 4.29E-03 4.15E-03
1128 4.61E-03 4.51E-03 4.44E-03 4.08E-03 3.95E-03
1129 5.16E-03 5.07E-03 4.99E-03 4.92E-03 4.79E-03
1131 4.87E-03 4.76E-03 4.71E-03 4.62E-03 4.51E-03
1133 5.27E-03 5.22E-03 5.17E-03 5.12E-03 4.95E-03
1134 4.88E-03 4.83E-03 4.79E-03 4.72E-03 4.59E-03
1136 4.69E-03 4.66E-03 4.61E-03 4.53E-03 4.36E-03
1139 4.75E-03 4.70E-03 4.65E-03 4.58E-03 4.41E-03
1140 4.92E-03 4.88E-03 4.82E-03 4.78E-03 4.62E-03
1141 4.76E-03 4.77E-03 4.76E-03 4.76E-03 4.76E-03
1143 5.20E-03 5.20E-03 5.20E-03 5.20E-03 5.20E-03
Biased Statistics
Average Biased
4.86E-03 4.76E-03 4.68E-03 4.55E-03 4.42E-03
Std Dev Biased
2.84E-04 2.76E-04 2.83E-04 3.55E-04 3.54E-04
Ps90%/90% (+KTL) Biased
5.64E-03 5.51E-03 5.46E-03 5.52E-03 5.39E-03
Ps90%/90% (-KTL) Biased
4.08E-03 4.00E-03 3.91E-03 3.58E-03 3.45E-03
Un-Biased Statistics
Average Un-Biased
4.90E-03 4.86E-03 4.81E-03 4.74E-03 4.59E-03
Std Dev Un-Biased
2.28E-04 2.24E-04 2.22E-04 2.31E-04 2.30E-04
Ps90%/90% (+KTL) Un-Biased
5.53E-03 5.47E-03 5.42E-03 5.38E-03 5.22E-03
Ps90%/90% (-KTL) Un-Biased
4.28E-03 4.24E-03 4.20E-03 4.11E-03 3.96E-03
Specification MAX
6.50E-03 7.00E-03 7.00E-03 7.00E-03 7.00E-03
Status
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2008 and DSCC Certified Company
8
TID Report
11-307 110624 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Ps90%/90% (-KTL) Biased
0.00E+00
Negative Supply Current (A)
-1.00E-03
-2.00E-03
-3.00E-03
-4.00E-03
-5.00E-03
-6.00E-03
-7.00E-03
-8.00E-03
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.2. Plot of Negative Supply Current (A) versus total dose. The solid diamonds are the average of the
measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of
the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or
dashed) are the average of the data points after application of the KTL statistics on the samples irradiated under
electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application of the
KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or postirradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
9
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table 5.2. Raw data for Negative Supply Current (A) versus total dose, including the statistical analysis,
specification and the status of the testing (pass/fail).
Negative Supply Current (A)
Device
1090
1124
1128
1129
1131
1133
1134
1136
1139
1140
1141
1143
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
-5.12E-03
-4.54E-03
-4.60E-03
-5.18E-03
-4.89E-03
-5.28E-03
-4.86E-03
-4.69E-03
-4.75E-03
-4.93E-03
-4.76E-03
-5.20E-03
Total
25
-5.02E-03
-4.45E-03
-4.50E-03
-5.07E-03
-4.77E-03
-5.23E-03
-4.84E-03
-4.66E-03
-4.69E-03
-4.88E-03
-4.77E-03
-5.20E-03
Dose (krad(Si))
50
100
-4.93E-03 -4.84E-03
-4.37E-03 -4.29E-03
-4.43E-03 -4.07E-03
-5.01E-03 -4.93E-03
-4.70E-03 -4.64E-03
-5.18E-03 -5.13E-03
-4.78E-03 -4.73E-03
-4.60E-03 -4.54E-03
-4.66E-03 -4.59E-03
-4.83E-03 -4.76E-03
-4.77E-03 -4.75E-03
-5.21E-03 -5.21E-03
200
-4.70E-03
-4.15E-03
-3.95E-03
-4.80E-03
-4.51E-03
-4.98E-03
-4.61E-03
-4.38E-03
-4.42E-03
-4.62E-03
-4.76E-03
-5.19E-03
-4.87E-03 -4.76E-03 -4.69E-03 -4.55E-03 -4.42E-03
2.91E-04 2.87E-04 2.85E-04 3.64E-04 3.60E-04
-4.07E-03 -3.98E-03 -3.91E-03 -3.56E-03 -3.43E-03
-5.67E-03 -5.55E-03 -5.47E-03 -5.55E-03 -5.41E-03
-4.90E-03
2.30E-04
-4.27E-03
-5.53E-03
-6.50E-03
PASS
-4.86E-03
2.25E-04
-4.24E-03
-5.48E-03
-7.00E-03
PASS
-4.81E-03
2.27E-04
-4.19E-03
-5.43E-03
-7.00E-03
PASS
-4.75E-03
2.33E-04
-4.11E-03
-5.39E-03
-7.00E-03
PASS
An ISO 9001:2008 and DSCC Certified Company
10
-4.60E-03
2.37E-04
-3.95E-03
-5.25E-03
-7.00E-03
PASS
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
1.00E-03
8.00E-04
Input Offset Voltage (V)
6.00E-04
4.00E-04
2.00E-04
0.00E+00
-2.00E-04
-4.00E-04
-6.00E-04
-8.00E-04
-1.00E-03
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.3. Plot of Input Offset Voltage (V) versus total dose. The solid diamonds are the average of the
measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of
the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or
dashed) are the average of the data points after application of the KTL statistics on the samples irradiated under
electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application of the
KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or postirradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
11
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table 5.3. Raw data for Input Offset Voltage (V) versus total dose, including the statistical analysis, specification
and the status of the testing (pass/fail).
Input Offset Voltage (V)
Device
1090
1124
1128
1129
1131
1133
1134
1136
1139
1140
1141
1143
0
-1.73E-04
-1.32E-04
-8.35E-05
1.56E-04
-1.78E-04
1.73E-04
-1.05E-04
-2.30E-04
-2.16E-04
-6.34E-05
1.07E-04
6.20E-05
Biased Statistics
Average Biased
-8.20E-05
Std Dev Biased
1.38E-04
Ps90%/90% (+KTL) Biased
2.98E-04
Ps90%/90% (-KTL) Biased
-4.62E-04
Un-Biased Statistics
Average Un-Biased
-8.82E-05
Std Dev Un-Biased
1.62E-04
Ps90%/90% (+KTL) Un-Biased
3.57E-04
Ps90%/90% (-KTL) Un-Biased
-5.33E-04
Specification MIN
-3.00E-04
Status
PASS
Specification MAX
3.00E-04
Status
PASS
Total
25
-2.33E-04
-1.96E-04
-1.12E-04
4.43E-05
-2.08E-04
9.92E-05
-1.79E-04
-2.93E-04
-2.81E-04
-1.49E-04
1.11E-04
5.97E-05
Dose (krad(Si))
50
100
-2.55E-04 -3.07E-04
-2.27E-04 -2.51E-04
-1.17E-04 -1.56E-04
2.72E-05 -2.32E-05
-1.95E-04 -2.24E-04
6.35E-05 -2.70E-06
-2.07E-04 -2.84E-04
-3.19E-04 -3.99E-04
-2.91E-04 -3.50E-04
-1.62E-04 -2.35E-04
1.30E-04 1.10E-04
8.42E-05 5.79E-05
200
-3.41E-04
-2.97E-04
-1.79E-04
-7.22E-05
-2.36E-04
-6.22E-05
-3.43E-04
-4.65E-04
-4.01E-04
-2.89E-04
1.07E-04
6.32E-05
-1.41E-04 -1.53E-04 -1.92E-04 -2.25E-04
1.13E-04 1.13E-04 1.09E-04 1.05E-04
1.69E-04 1.58E-04 1.07E-04 6.35E-05
-4.51E-04 -4.65E-04 -4.91E-04 -5.13E-04
-1.61E-04
1.58E-04
2.73E-04
-5.94E-04
-3.00E-04
PASS
3.00E-04
PASS
-1.83E-04
1.52E-04
2.33E-04
-5.98E-04
-3.70E-04
PASS
3.70E-04
PASS
-2.54E-04
1.54E-04
1.68E-04
-6.76E-04
-5.70E-04
PASS
5.70E-04
PASS
An ISO 9001:2008 and DSCC Certified Company
12
-3.12E-04
1.54E-04
1.11E-04
-7.35E-04
-8.70E-04
PASS
8.70E-04
PASS
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
4.00E-10
Input Offset Current (A)
3.00E-10
2.00E-10
1.00E-10
0.00E+00
-1.00E-10
-2.00E-10
-3.00E-10
-4.00E-10
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.4. Plot of Input Offset Current (A) versus total dose. The solid diamonds are the average of the
measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of
the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or
dashed) are the average of the data points after application of the KTL statistics on the samples irradiated under
electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application of the
KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or postirradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
13
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table 5.4. Raw data for Input Offset Current (A) versus total dose, including the statistical analysis, specification
and the status of the testing (pass/fail).
Input Offset Current (A)
Device
1090
1124
1128
1129
1131
1133
1134
1136
1139
1140
1141
1143
0
-5.00E-12
-3.00E-12
-3.00E-12
-5.00E-12
-3.00E-12
-2.00E-12
-7.00E-12
-5.00E-12
-3.00E-12
-5.00E-12
-6.00E-12
-4.00E-12
Biased Statistics
Average Biased
-3.80E-12
Std Dev Biased
1.10E-12
Ps90%/90% (+KTL) Biased
-7.96E-13
Ps90%/90% (-KTL) Biased
-6.80E-12
Un-Biased Statistics
Average Un-Biased
-4.40E-12
Std Dev Un-Biased
1.95E-12
Ps90%/90% (+KTL) Un-Biased
9.45E-13
Ps90%/90% (-KTL) Un-Biased
-9.75E-12
Specification MIN
-1.00E-11
Status
PASS
Specification MAX
1.00E-11
Status
PASS
Total
25
-1.10E-11
-9.00E-12
-6.00E-12
-1.20E-11
-8.00E-12
-4.00E-12
-7.00E-12
-7.00E-12
-2.00E-12
-7.00E-12
-5.00E-12
-4.00E-12
Dose (krad(Si))
50
100
-2.80E-11 -4.80E-11
-2.70E-11 -4.90E-11
-2.00E-11 -3.50E-11
-2.80E-11 -4.50E-11
-2.40E-11 -4.20E-11
-3.00E-12 -7.00E-12
-8.00E-12 -1.00E-11
-8.00E-12 -8.00E-12
-4.00E-12 -5.00E-12
-5.00E-12 -7.00E-12
-7.00E-12 -6.00E-12
-2.00E-12 -5.00E-12
200
-6.70E-11
-7.10E-11
-7.00E-11
-7.70E-11
-7.30E-11
-1.00E-11
-1.20E-11
-1.10E-11
-9.00E-12
-1.00E-11
-7.00E-12
-6.00E-12
-9.20E-12 -2.54E-11 -4.38E-11 -7.16E-11
2.39E-12 3.44E-12 5.63E-12 3.71E-12
-2.65E-12 -1.60E-11 -2.84E-11 -6.14E-11
-1.57E-11 -3.48E-11 -5.92E-11 -8.18E-11
-5.40E-12
2.30E-12
9.13E-13
-1.17E-11
-5.00E-11
PASS
5.00E-11
PASS
-5.60E-12
2.30E-12
7.13E-13
-1.19E-11
-1.50E-10
PASS
1.50E-10
PASS
-7.40E-12
1.82E-12
-2.42E-12
-1.24E-11
-2.50E-10
PASS
2.50E-10
PASS
An ISO 9001:2008 and DSCC Certified Company
14
-1.04E-11
1.14E-12
-7.27E-12
-1.35E-11
-3.50E-10
PASS
3.50E-10
PASS
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
2.50E-09
Positive Bias Current (A)
2.00E-09
1.50E-09
1.00E-09
5.00E-10
0.00E+00
-5.00E-10
-1.00E-09
-1.50E-09
-2.00E-09
-2.50E-09
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.5. Plot of Positive Bias Current (A) versus total dose. The solid diamonds are the average of the
measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of
the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or
dashed) are the average of the data points after application of the KTL statistics on the samples irradiated under
electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application of the
KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or postirradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
15
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table 5.5. Raw data for Positive Bias Current (A) versus total dose, including the statistical analysis,
specification and the status of the testing (pass/fail).
Positive Bias Current (A)
Device
1090
1124
1128
1129
1131
1133
1134
1136
1139
1140
1141
1143
0
1.40E-11
1.20E-11
1.40E-11
1.30E-11
1.90E-11
1.60E-11
1.30E-11
1.30E-11
1.40E-11
1.40E-11
2.00E-11
1.40E-11
Total
25
6.30E-11
5.40E-11
5.80E-11
5.00E-11
6.00E-11
3.50E-11
2.70E-11
2.50E-11
2.30E-11
3.10E-11
2.50E-11
1.40E-11
Dose (krad(Si))
50
100
1.34E-10 2.78E-10
1.20E-10 2.81E-10
1.12E-10 2.35E-10
1.33E-10 2.63E-10
1.31E-10 2.49E-10
4.20E-11 5.90E-11
3.90E-11 6.20E-11
3.60E-11 5.40E-11
3.90E-11 5.60E-11
3.80E-11 6.20E-11
2.40E-11 2.40E-11
1.40E-11 1.40E-11
200
4.71E-10
5.07E-10
4.79E-10
5.07E-10
4.89E-10
1.04E-10
9.60E-11
9.40E-11
9.20E-11
9.50E-11
1.70E-11
1.60E-11
Biased Statistics
Average Biased
1.44E-11 5.70E-11 1.26E-10 2.61E-10 4.91E-10
Std Dev Biased
2.70E-12 5.10E-12 9.62E-12 1.94E-11 1.63E-11
Ps90%/90% (+KTL) Biased
2.18E-11 7.10E-11 1.52E-10 3.15E-10 5.35E-10
Ps90%/90% (-KTL) Biased
6.99E-12 4.30E-11 9.96E-11 2.08E-10 4.46E-10
Un-Biased Statistics
Average Un-Biased
1.40E-11 2.82E-11 3.88E-11 5.86E-11 9.62E-11
Std Dev Un-Biased
1.22E-12 4.82E-12 2.17E-12 3.58E-12 4.60E-12
Ps90%/90% (+KTL) Un-Biased
1.74E-11 4.14E-11 4.47E-11 6.84E-11 1.09E-10
Ps90%/90% (-KTL) Un-Biased
1.06E-11 1.50E-11 3.29E-11 4.88E-11 8.36E-11
Specification MIN
-5.00E-11 -2.50E-10 -5.00E-10 -1.00E-09 -2.00E-09
Status
PASS
PASS
PASS
PASS
PASS
Specification MAX
5.00E-11 2.50E-10 5.00E-10 1.00E-09 2.00E-09
Status
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2008 and DSCC Certified Company
16
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
2.50E-09
Negative Bias Current (A)
2.00E-09
1.50E-09
1.00E-09
5.00E-10
0.00E+00
-5.00E-10
-1.00E-09
-1.50E-09
-2.00E-09
-2.50E-09
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.6. Plot of Negative Bias Current (A) versus total dose. The solid diamonds are the average of the
measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of
the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or
dashed) are the average of the data points after application of the KTL statistics on the samples irradiated under
electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application of the
KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or postirradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
17
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Table 5.6. Raw data for Negative Bias Current (A) versus total dose, including the statistical analysis,
specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Negative Bias Current (A)
Device
0
25
50
100
200
1090 1.80E-11 7.60E-11 1.65E-10 3.39E-10 5.67E-10
1124 1.50E-11 6.40E-11 1.48E-10 3.38E-10 5.83E-10
1128 1.60E-11 6.80E-11 1.35E-10 2.85E-10 5.51E-10
1129 1.70E-11 6.20E-11 1.63E-10 3.27E-10 6.35E-10
1131 2.10E-11 6.80E-11 1.58E-10 3.04E-10 5.85E-10
1133 1.80E-11 3.80E-11 4.70E-11 6.60E-11 1.12E-10
1134 1.80E-11 3.40E-11 4.50E-11 7.30E-11 1.17E-10
1136 1.70E-11 3.10E-11 4.10E-11 6.70E-11 1.15E-10
1139 1.60E-11 2.50E-11 4.30E-11 6.10E-11 1.02E-10
1140 1.60E-11 3.50E-11 4.20E-11 6.90E-11 1.10E-10
1141 2.20E-11 2.70E-11 2.70E-11 2.70E-11 1.80E-11
1143 1.80E-11 1.60E-11 1.60E-11 1.70E-11 1.90E-11
Biased Statistics
Average Biased
1.74E-11 6.76E-11 1.54E-10 3.19E-10 5.84E-10
Std Dev Biased
2.30E-12 5.37E-12 1.24E-11 2.35E-11 3.15E-11
Ps90%/90% (+KTL) Biased
2.37E-11 8.23E-11 1.88E-10 3.83E-10 6.71E-10
Ps90%/90% (-KTL) Biased
1.11E-11 5.29E-11 1.20E-10 2.54E-10 4.98E-10
Un-Biased Statistics
Average Un-Biased
1.70E-11 3.26E-11 4.36E-11 6.72E-11 1.11E-10
Std Dev Un-Biased
1.00E-12 4.93E-12 2.41E-12 4.38E-12 5.81E-12
Ps90%/90% (+KTL) Un-Biased
1.97E-11 4.61E-11 5.02E-11 7.92E-11 1.27E-10
Ps90%/90% (-KTL) Un-Biased
1.43E-11 1.91E-11 3.70E-11 5.52E-11 9.53E-11
Specification MIN
-5.00E-11 -2.50E-10 -5.00E-10 -1.00E-09 -2.00E-09
Status
PASS
PASS
PASS
PASS
PASS
Specification MAX
5.00E-11 2.50E-10 5.00E-10 1.00E-09 2.00E-09
Status
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2008 and DSCC Certified Company
18
TID Report
11-307 110624 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Ps90%/90% (-KTL) Biased
Large Signal Voltage Gain 2K (V/mV)
3.00E+02
2.50E+02
2.00E+02
1.50E+02
1.00E+02
5.00E+01
0.00E+00
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.7. Plot of Large Signal Voltage Gain 2K (V/mV) versus total dose. The solid diamonds are the average
of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the
average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid
and/or dashed) are the average of the data points after application of the KTL statistics on the samples irradiated
under electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application
of the KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or
post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
19
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table 5.7. Raw data for Large Signal Voltage Gain 2K (V/mV) versus total dose, including the statistical
analysis, specification and the status of the testing (pass/fail).
Large Signal Voltage Gain 2K (V/mV)
Device
1090
1124
1128
1129
1131
1133
1134
1136
1139
1140
1141
1143
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
2.66E+02
2.91E+02
2.82E+02
2.58E+02
2.61E+02
2.59E+02
2.52E+02
2.96E+02
2.80E+02
2.57E+02
2.64E+02
2.52E+02
Total
25
2.63E+02
2.89E+02
2.82E+02
2.58E+02
2.66E+02
2.54E+02
2.51E+02
2.97E+02
2.87E+02
2.62E+02
2.71E+02
2.52E+02
Dose (krad(Si))
50
100
2.65E+02 2.74E+02
2.86E+02 2.86E+02
2.82E+02 2.93E+02
2.54E+02 2.59E+02
2.64E+02 2.72E+02
2.52E+02 2.56E+02
2.48E+02 2.50E+02
2.89E+02 2.89E+02
2.87E+02 2.84E+02
2.66E+02 2.58E+02
2.73E+02 2.68E+02
2.50E+02 2.51E+02
200
2.68E+02
2.84E+02
2.94E+02
2.56E+02
2.67E+02
2.68E+02
2.52E+02
2.91E+02
3.00E+02
2.60E+02
2.71E+02
2.52E+02
2.71E+02
1.44E+01
3.11E+02
2.32E+02
2.71E+02
1.36E+01
3.09E+02
2.34E+02
2.70E+02
1.33E+01
3.07E+02
2.34E+02
2.77E+02
1.31E+01
3.13E+02
2.41E+02
2.74E+02
1.52E+01
3.15E+02
2.32E+02
2.69E+02
1.85E+01
3.20E+02
2.18E+02
1.50E+02
PASS
2.70E+02
2.05E+01
3.26E+02
2.14E+02
1.50E+02
PASS
2.68E+02
1.91E+01
3.21E+02
2.16E+02
1.50E+02
PASS
2.67E+02
1.78E+01
3.16E+02
2.18E+02
1.00E+02
PASS
2.74E+02
2.05E+01
3.30E+02
2.18E+02
7.50E+01
PASS
An ISO 9001:2008 and DSCC Certified Company
20
TID Report
11-307 110624 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Ps90%/90% (-KTL) Biased
Large Signal Voltage Gain 1K (V/mV)
3.00E+02
2.50E+02
2.00E+02
1.50E+02
1.00E+02
5.00E+01
0.00E+00
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.8. Plot of Large Signal Voltage Gain 1K (V/mV) versus total dose. The solid diamonds are the average
of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the
average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid
and/or dashed) are the average of the data points after application of the KTL statistics on the samples irradiated
under electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application
of the KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or
post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
21
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table 5.8. Raw data for Large Signal Voltage Gain 1K (V/mV) versus total dose, including the statistical
analysis, specification and the status of the testing (pass/fail).
Large Signal Voltage Gain 1K (V/mV)
Device
1090
1124
1128
1129
1131
1133
1134
1136
1139
1140
1141
1143
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
2.47E+02
2.72E+02
2.63E+02
2.37E+02
2.43E+02
2.44E+02
2.36E+02
2.73E+02
2.66E+02
2.39E+02
2.48E+02
2.37E+02
Total
25
2.45E+02
2.71E+02
2.67E+02
2.41E+02
2.49E+02
2.42E+02
2.36E+02
2.73E+02
2.71E+02
2.38E+02
2.54E+02
2.37E+02
Dose (krad(Si))
50
100
2.46E+02 2.50E+02
2.70E+02 2.72E+02
2.72E+02 2.78E+02
2.37E+02 2.44E+02
2.51E+02 2.57E+02
2.39E+02 2.41E+02
2.29E+02 2.42E+02
2.69E+02 2.75E+02
2.72E+02 2.66E+02
2.43E+02 2.48E+02
2.52E+02 2.52E+02
2.36E+02 2.37E+02
200
2.54E+02
2.70E+02
2.75E+02
2.39E+02
2.56E+02
2.43E+02
2.43E+02
2.80E+02
2.68E+02
2.42E+02
2.53E+02
2.37E+02
2.52E+02
1.47E+01
2.92E+02
2.12E+02
2.55E+02
1.34E+01
2.92E+02
2.18E+02
2.55E+02
1.52E+01
2.97E+02
2.14E+02
2.60E+02
1.45E+01
3.00E+02
2.20E+02
2.59E+02
1.41E+01
2.97E+02
2.20E+02
2.52E+02
1.70E+01
2.98E+02
2.05E+02
1.30E+02
PASS
2.52E+02
1.84E+01
3.03E+02
2.01E+02
1.30E+02
PASS
2.50E+02
1.89E+01
3.02E+02
1.98E+02
1.30E+02
PASS
2.55E+02
1.55E+01
2.97E+02
2.12E+02
8.70E+01
PASS
2.55E+02
1.77E+01
3.04E+02
2.07E+02
6.50E+01
PASS
An ISO 9001:2008 and DSCC Certified Company
22
TID Report
11-307 110624 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Ps90%/90% (-KTL) Biased
Common Mode Rejection Ratio (dB)
1.20E+02
1.00E+02
8.00E+01
6.00E+01
4.00E+01
2.00E+01
0.00E+00
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.9. Plot of Common Mode Rejection Ratio (dB) versus total dose. The solid diamonds are the average of
the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the
average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid
and/or dashed) are the average of the data points after application of the KTL statistics on the samples irradiated
under electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application
of the KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or
post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
23
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table 5.9. Raw data for Common Mode Rejection Ratio (dB) versus total dose, including the statistical analysis,
specification and the status of the testing (pass/fail).
Common Mode Rejection Ratio (dB)
Device
1090
1124
1128
1129
1131
1133
1134
1136
1139
1140
1141
1143
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
9.76E+01
9.71E+01
9.54E+01
1.03E+02
1.16E+02
9.28E+01
8.83E+01
9.84E+01
1.05E+02
9.34E+01
9.56E+01
9.69E+01
Total
25
9.69E+01
9.65E+01
9.63E+01
1.02E+02
1.13E+02
9.24E+01
8.80E+01
9.92E+01
1.04E+02
9.28E+01
9.53E+01
9.69E+01
Dose (krad(Si))
50
100
9.68E+01 9.69E+01
9.67E+01 9.67E+01
9.61E+01 9.61E+01
1.02E+02 1.02E+02
1.13E+02 1.12E+02
9.23E+01 9.21E+01
8.79E+01 8.78E+01
9.94E+01 9.94E+01
1.03E+02 1.03E+02
9.29E+01 9.28E+01
9.54E+01 9.56E+01
9.69E+01 9.69E+01
200
9.71E+01
9.67E+01
9.59E+01
1.03E+02
1.14E+02
9.22E+01
8.77E+01
9.91E+01
1.03E+02
9.30E+01
9.53E+01
9.70E+01
1.02E+02
8.26E+00
1.24E+02
7.91E+01
1.01E+02
6.93E+00
1.20E+02
8.18E+01
1.01E+02
7.05E+00
1.20E+02
8.16E+01
1.01E+02
6.86E+00
1.20E+02
8.20E+01
1.01E+02
7.43E+00
1.22E+02
8.08E+01
9.55E+01
6.22E+00
1.13E+02
7.84E+01
8.60E+01
PASS
9.52E+01
6.16E+00
1.12E+02
7.83E+01
8.60E+01
PASS
9.52E+01
6.13E+00
1.12E+02
7.84E+01
8.60E+01
PASS
9.50E+01
6.03E+00
1.12E+02
7.85E+01
8.60E+01
PASS
9.50E+01
5.97E+00
1.11E+02
7.86E+01
8.60E+01
PASS
An ISO 9001:2008 and DSCC Certified Company
24
TID Report
11-307 110624 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Ps90%/90% (-KTL) Biased
Power Supply Rejection Ratio (dB)
1.20E+02
1.00E+02
8.00E+01
6.00E+01
4.00E+01
2.00E+01
0.00E+00
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.10. Plot of Power Supply Rejection Ratio (dB) versus total dose. The solid diamonds are the average of
the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the
average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid
and/or dashed) are the average of the data points after application of the KTL statistics on the samples irradiated
under electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application
of the KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or
post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
25
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table 5.10. Raw data for Power Supply Rejection Ratio (dB) versus total dose, including the statistical analysis,
specification and the status of the testing (pass/fail).
Power Supply Rejection Ratio (dB)
Device
1090
1124
1128
1129
1131
1133
1134
1136
1139
1140
1141
1143
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
1.03E+02
1.08E+02
9.89E+01
1.00E+02
1.39E+02
9.57E+01
9.24E+01
9.52E+01
1.24E+02
1.06E+02
1.08E+02
1.02E+02
Total
25
1.03E+02
1.08E+02
9.90E+01
1.01E+02
1.27E+02
9.59E+01
9.25E+01
9.52E+01
1.26E+02
1.06E+02
1.08E+02
1.02E+02
Dose (krad(Si))
50
100
1.04E+02 1.04E+02
1.09E+02 1.10E+02
9.89E+01 9.89E+01
1.02E+02 1.02E+02
1.25E+02 1.25E+02
9.59E+01 9.61E+01
9.25E+01 9.26E+01
9.51E+01 9.50E+01
1.31E+02 1.34E+02
1.06E+02 1.07E+02
1.08E+02 1.08E+02
1.02E+02 1.02E+02
200
1.05E+02
1.11E+02
9.85E+01
1.02E+02
1.23E+02
9.65E+01
9.28E+01
9.49E+01
1.43E+02
1.08E+02
1.08E+02
1.02E+02
1.10E+02
1.65E+01
1.55E+02
6.45E+01
1.08E+02
1.13E+01
1.39E+02
7.68E+01
1.08E+02
1.05E+01
1.36E+02
7.88E+01
1.08E+02
1.05E+01
1.37E+02
7.92E+01
1.08E+02
9.40E+00
1.34E+02
8.20E+01
1.03E+02
1.29E+01
1.38E+02
6.71E+01
9.00E+01
PASS
1.03E+02
1.37E+01
1.41E+02
6.55E+01
9.00E+01
PASS
1.04E+02
1.58E+01
1.47E+02
6.08E+01
9.00E+01
PASS
1.05E+02
1.72E+01
1.52E+02
5.79E+01
9.00E+01
PASS
1.07E+02
2.10E+01
1.65E+02
4.96E+01
9.00E+01
PASS
An ISO 9001:2008 and DSCC Certified Company
26
TID Report
11-307 110624 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Ps90%/90% (-KTL) Biased
2.00E+01
1.80E+01
Positive Slew Rate (V/µs)
1.60E+01
1.40E+01
1.20E+01
1.00E+01
8.00E+00
6.00E+00
4.00E+00
2.00E+00
0.00E+00
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.11. Plot of Positive Slew Rate (V/µs) versus total dose. The solid diamonds are the average of the
measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of
the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or
dashed) are the average of the data points after application of the KTL statistics on the samples irradiated under
electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application of the
KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or postirradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
27
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table 5.11. Raw data for Positive Slew Rate (V/µs) versus total dose, including the statistical analysis,
specification and the status of the testing (pass/fail).
Positive Slew Rate (V/µs)
Device
1090
1124
1128
1129
1131
1133
1134
1136
1139
1140
1141
1143
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
1.82E+01
1.75E+01
1.28E+01
1.84E+01
1.79E+01
1.85E+01
1.47E+01
1.77E+01
1.73E+01
1.82E+01
1.78E+01
1.42E+01
Total
25
1.79E+01
1.72E+01
1.28E+01
1.83E+01
1.77E+01
1.84E+01
1.47E+01
1.75E+01
1.72E+01
1.83E+01
1.80E+01
1.43E+01
Dose (krad(Si))
50
100
1.76E+01 1.78E+01
1.72E+01 1.74E+01
1.28E+01 1.29E+01
1.79E+01 1.80E+01
1.75E+01 1.75E+01
1.83E+01 1.86E+01
1.46E+01 1.47E+01
1.75E+01 1.76E+01
1.70E+01 1.73E+01
1.83E+01 1.85E+01
1.80E+01 1.79E+01
1.42E+01 1.43E+01
200
1.84E+01
1.74E+01
1.28E+01
1.85E+01
1.81E+01
1.85E+01
1.47E+01
1.79E+01
1.74E+01
1.81E+01
1.79E+01
1.42E+01
1.70E+01
2.33E+00
2.34E+01
1.06E+01
1.68E+01
2.25E+00
2.29E+01
1.06E+01
1.66E+01
2.14E+00
2.25E+01
1.07E+01
1.67E+01
2.17E+00
2.27E+01
1.08E+01
1.70E+01
2.40E+00
2.36E+01
1.05E+01
1.73E+01
1.53E+00
2.15E+01
1.31E+01
1.00E+01
PASS
1.72E+01
1.51E+00
2.13E+01
1.31E+01
1.00E+01
PASS
1.71E+01
1.53E+00
2.13E+01
1.30E+01
9.00E+00
PASS
1.73E+01
1.57E+00
2.16E+01
1.30E+01
9.00E+00
PASS
1.73E+01
1.53E+00
2.15E+01
1.31E+01
9.00E+00
PASS
An ISO 9001:2008 and DSCC Certified Company
28
TID Report
11-307 110624 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Ps90%/90% (+KTL) Biased
0.00E+00
Negative Slew Rate (V/µs)
-1.00E+01
-2.00E+01
-3.00E+01
-4.00E+01
-5.00E+01
-6.00E+01
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.12. Plot of Negative Slew Rate (V/µs) versus total dose. The solid diamonds are the average of the
measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of
the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or
dashed) are the average of the data points after application of the KTL statistics on the samples irradiated under
electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application of the
KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or postirradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
29
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table 5.12. Raw data for Negative Slew Rate (V/µs) versus total dose, including the statistical analysis,
specification and the status of the testing (pass/fail).
Negative Slew Rate (V/µs)
Device
1090
1124
1128
1129
1131
1133
1134
1136
1139
1140
1141
1143
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
-3.80E+01
-3.49E+01
-3.80E+01
-3.67E+01
-3.92E+01
-3.74E+01
-3.55E+01
-3.55E+01
-3.60E+01
-3.73E+01
-3.71E+01
-3.63E+01
Total Dose (krad(Si))
25
50
100
-3.68E+01 -3.66E+01 -5.62E+01
-3.42E+01 -3.45E+01 -3.54E+01
-3.63E+01 -3.61E+01 -3.70E+01
-3.79E+01 -3.84E+01 -5.47E+01
-3.77E+01 -3.74E+01 -5.81E+01
-3.92E+01 -4.03E+01 -5.79E+01
-3.53E+01 -3.60E+01 -3.75E+01
-3.52E+01 -3.54E+01 -3.60E+01
-3.57E+01 -3.58E+01 -3.63E+01
-3.67E+01 -3.80E+01 -5.78E+01
-3.70E+01 -3.70E+01 -3.70E+01
-3.61E+01 -3.62E+01 -3.62E+01
200
-5.78E+01
-3.86E+01
-3.77E+01
-5.55E+01
-5.89E+01
-5.81E+01
-3.86E+01
-4.05E+01
-3.72E+01
-5.75E+01
-3.70E+01
-3.63E+01
-3.74E+01
1.65E+00
-3.28E+01
-4.19E+01
-3.66E+01
1.49E+00
-3.25E+01
-4.07E+01
-3.66E+01
1.47E+00
-3.26E+01
-4.06E+01
-4.83E+01
1.11E+01
-1.78E+01
-7.88E+01
-4.97E+01
1.06E+01
-2.05E+01
-7.88E+01
-3.63E+01
9.41E-01
-3.38E+01
-3.89E+01
-1.00E+01
PASS
-3.64E+01
1.65E+00
-3.19E+01
-4.09E+01
-1.00E+01
PASS
-3.71E+01
2.03E+00
-3.15E+01
-4.27E+01
-9.00E+00
PASS
-4.51E+01
1.17E+01
-1.31E+01
-7.71E+01
-9.00E+00
PASS
-4.64E+01
1.05E+01
-1.77E+01
-7.51E+01
-9.00E+00
PASS
An ISO 9001:2008 and DSCC Certified Company
30
TID Report
11-307 110624 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Ps90%/90% (-KTL) Biased
1.34E+01
Positive Output Voltage Swing (V)
1.32E+01
1.30E+01
1.28E+01
1.26E+01
1.24E+01
1.22E+01
1.20E+01
1.18E+01
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.13. Plot of Positive Output Voltage Swing (V) versus total dose. The solid diamonds are the average of
the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the
average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid
and/or dashed) are the average of the data points after application of the KTL statistics on the samples irradiated
under electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application
of the KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or
post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
31
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table 5.13. Raw data for Positive Output Voltage Swing (V) versus total dose, including the statistical analysis,
specification and the status of the testing (pass/fail).
Positive Output Voltage Swing (V)
Device
1090
1124
1128
1129
1131
1133
1134
1136
1139
1140
1141
1143
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
Total
25
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
Dose (krad(Si))
50
100
1.32E+01 1.32E+01
1.32E+01 1.32E+01
1.32E+01 1.32E+01
1.32E+01 1.32E+01
1.32E+01 1.32E+01
1.32E+01 1.32E+01
1.32E+01 1.32E+01
1.32E+01 1.32E+01
1.32E+01 1.32E+01
1.32E+01 1.32E+01
1.32E+01 1.32E+01
1.32E+01 1.32E+01
200
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01
1.32E+01 1.32E+01 1.32E+01 1.32E+01 1.32E+01
6.11E-03 5.63E-03 6.10E-03 5.52E-03 5.77E-03
1.32E+01 1.32E+01 1.32E+01 1.32E+01 1.32E+01
1.32E+01 1.32E+01 1.32E+01 1.32E+01 1.32E+01
1.32E+01
6.36E-03
1.32E+01
1.32E+01
1.20E+01
PASS
1.32E+01
5.27E-03
1.32E+01
1.32E+01
1.20E+01
PASS
1.32E+01
5.37E-03
1.32E+01
1.32E+01
1.20E+01
PASS
1.32E+01
4.56E-03
1.32E+01
1.32E+01
1.20E+01
PASS
An ISO 9001:2008 and DSCC Certified Company
32
1.32E+01
4.82E-03
1.32E+01
1.32E+01
1.20E+01
PASS
TID Report
11-307 110624 R1.0
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Ps90%/90% (+KTL) Biased
Negative Output Voltage Swing (V)
-1.18E+01
-1.20E+01
-1.22E+01
-1.24E+01
-1.26E+01
-1.28E+01
-1.30E+01
-1.32E+01
0
50
100
150
200
Total Dose (krad(Si))
Figure 5.14. Plot of Negative Output Voltage Swing (V) versus total dose. The solid diamonds are the average of
the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the
average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid
and/or dashed) are the average of the data points after application of the KTL statistics on the samples irradiated
under electrical bias while the gray lines (solid and/or dashed) are the average of the data points after application
of the KTL statistics on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or
post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan.
An ISO 9001:2008 and DSCC Certified Company
33
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table 5.14. Raw data for Negative Output Voltage Swing (V) versus total dose, including the statistical analysis,
specification and the status of the testing (pass/fail).
Negative Output Voltage Swing (V)
Device
1090
1124
1128
1129
1131
1133
1134
1136
1139
1140
1141
1143
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
-1.31E+01
-1.31E+01
-1.31E+01
-1.31E+01
-1.31E+01
-1.31E+01
-1.32E+01
-1.31E+01
-1.31E+01
-1.31E+01
-1.31E+01
-1.31E+01
Total Dose (krad(Si))
25
50
100
-1.31E+01 -1.31E+01 -1.31E+01
-1.31E+01 -1.31E+01 -1.31E+01
-1.31E+01 -1.31E+01 -1.31E+01
-1.31E+01 -1.31E+01 -1.31E+01
-1.31E+01 -1.31E+01 -1.31E+01
-1.31E+01 -1.31E+01 -1.31E+01
-1.31E+01 -1.31E+01 -1.31E+01
-1.31E+01 -1.31E+01 -1.31E+01
-1.31E+01 -1.31E+01 -1.31E+01
-1.31E+01 -1.31E+01 -1.31E+01
-1.31E+01 -1.31E+01 -1.31E+01
-1.31E+01 -1.31E+01 -1.31E+01
200
-1.31E+01
-1.30E+01
-1.30E+01
-1.31E+01
-1.31E+01
-1.31E+01
-1.31E+01
-1.30E+01
-1.30E+01
-1.31E+01
-1.31E+01
-1.31E+01
-1.31E+01 -1.31E+01 -1.31E+01 -1.31E+01 -1.31E+01
1.43E-02 1.65E-02 1.70E-02 2.08E-02 2.81E-02
-1.31E+01 -1.31E+01 -1.31E+01 -1.30E+01 -1.30E+01
-1.32E+01 -1.32E+01 -1.31E+01 -1.31E+01 -1.31E+01
-1.31E+01
1.85E-02
-1.31E+01
-1.32E+01
-1.20E+01
PASS
-1.31E+01
1.96E-02
-1.31E+01
-1.32E+01
-1.20E+01
PASS
-1.31E+01
1.97E-02
-1.31E+01
-1.32E+01
-1.20E+01
PASS
An ISO 9001:2008 and DSCC Certified Company
34
-1.31E+01
2.25E-02
-1.30E+01
-1.32E+01
-1.20E+01
PASS
-1.31E+01
2.92E-02
-1.30E+01
-1.31E+01
-1.20E+01
PASS
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
6.0. Summary / Conclusions
The total ionizing dose testing described in this final report was performed using the facilities at
Radiation Assured Devices' Longmire Laboratories in Colorado Springs, CO. The high dose rate total
ionizing dose (TID) source is a JLSA 81-24 irradiator modified to provide a panoramic exposure. The
Co-60 rods are held in the base of the irradiator heavily shielded by lead, during the radiation exposures
the rod is raised by an electronic timer/controller and the exposure is performed in air. The dose rate for
this irradiator in this configuration ranges from <1rad(Si)/s to a maximum of approximately
120rad(Si)/s, determined by the distance from the source.
The parametric data was obtained as "read and record" and all the raw data plus an attributes summary
are contained in this report as well as in a separate Excel file. The attributes data contains the average,
standard deviation and the average with the KTL values applied. The KTL value used in this work is
2.742 per MIL-HDBK-814 using one sided tolerance limits of 90/90 and a 5-piece sample size. The
90/90 KTL values were selected to match the statistical levels specified in the MIL-PRF-38535
sampling plan for the qualification of a radiation hardness assured (RHA) component. Note that the
following criteria must be met for a device to pass the total ionizing dose test: following the radiation
exposure each of the 5 pieces irradiated under electrical bias shall pass the specification value. The units
irradiated without electrical bias and the KTL statistics are included in this report for reference only. If
any of the 5 pieces irradiated under electrical bias exceed the datasheet specifications, then the lot could
be logged as a failure.
Based on this criterion the RH1056AMW JFET Input Operational Amplifier (from the lot date code
identified on the first page of this test report) PASSED the total ionizing dose test to the maximum
tested dose level of 200krad(Si) with all parameters remaining within their datasheet specifications.
An ISO 9001:2008 and DSCC Certified Company
35
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Appendix A: Photograph of Packing Label and a Sample Unit-Under-Test to Show Part
Traceability
An ISO 9001:2008 and DSCC Certified Company
36
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Appendix B: Radiation Bias Connections
TID Radiation Biased Conditions: Extracted from Linear Technology RH1056 Precision, High Speed,
JFET Input Operational Amplifier Datasheet.
Pin Function
Connection / Bias
1
NC
NC
2
BAL
NC
3
-IN
To Pin 7 via 10kΩ Resistor
4
+IN
5
V-
6
BAL
To 8V via 10kΩ Resistor
To -15V
To GND via 0.1µF Capacitor
NC
7
OUT
8
V+
9
NC
To Pin 3 via 10kΩ Resistor
To +15V
To GND via 0.1µF Capacitor
NC
10
NC
NC
TID Radiation Unbiased Conditions: All pins connected to Ground.
Pin Function Connection / Bias
1
NC
GND
2
BAL
GND
3
-IN
GND
4
+IN
GND
5
V-
GND
6
BAL
GND
7
OUT
GND
8
V+
GND
9
NC
GND
10
NC
GND
An ISO 9001:2008 and DSCC Certified Company
37
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Figure B.1. Irradiation bias circuit for the units to be irradiated under electrical bias. This figure was extracted
from Linear Technology RH1056 Precision, High Speed, JFET Input Operational Amplifier Datasheet.
Figure B.2. Package drawing (for reference only). This figure was extracted from Linear Technology RH1056
Precision, High Speed, JFET Input Operational Amplifier Datasheet.
Absolute Maximum Rating:
Parameter
Max Rating
Supply Voltage
±20V
Differential Input Voltage ±40V
Input Voltage
±20V
An ISO 9001:2008 and DSCC Certified Company
38
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Appendix C: Electrical Test Parameters and Conditions
The expected ranges of values as well as the measurement conditions are taken from Linear Technology
RH1056 Precision, High Speed, JFET Input Operational Amplifier Datasheet. All electrical tests for this
device are performed on one of Radiation Assured Device's LTS2020 Test Systems. The LTS2020 Test
System is a programmable parametric tester that provides parameter measurements for a variety of
digital, analog and mixed signal products including voltage regulators, voltage comparators, D to A and
A to D converters. The LTS2020 Test System achieves accuracy and sensitivity through the use of
software self-calibration and an internal relay matrix with separate family boards and custom personality
adapter boards. The tester uses this relay matrix to connect the required test circuits, select the
appropriate voltage / current sources and establish the needed measurement loops for all the tests
performed. The measured parameters and test conditions are shown in Table C.1.
A listing of the measurement precision/resolution for each parameter is shown in Table C.2. The
precision/resolution values were obtained from test data or from the DAC resolution of the LTS-2020
for the particular test shown, whichever is greater. To generate the precision/resolution shown in Table
C.2, one of the units-under-test was tested repetitively (a total of 10-times with re-insertion between
tests) to obtain the average test value and standard deviation. Using this test data MIL-HDBK-814 90/90
KTL statistics were applied to the measured standard deviation to generate the final measurement range.
This value encompasses the precision/resolution of all aspects of the test system, including the LTS2020
mainframe, family board, socket assembly and DUT board as well as insertion error. In some cases, the
measurement resolution is limited by the internal DACs, which results in a measured standard deviation
of zero. In these instances the precision/resolution will be reported back as the LSB of the DAC.
Note that the testing and statistics used in this document are based on an “analysis of variables”
technique, which relies on small sample sizes to qualify much larger lot sizes (see MIL-HDBK-814, p.
91 for a discussion of statistical treatments). Not all measured parameters are well suited to this
approach due to inherent large variations. If necessary, larger samples sizes could be used to qualify
these parameters using an “attributes” approach.
An ISO 9001:2008 and DSCC Certified Company
39
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table C.1. Measured parameters and test conditions for the RH1056AMW Precision, High Speed, JFET Input
Operational Amplifier.
Parameter
Symbol
Test Conditions
Positive Supply Current (A)
ICC
VS=+/-15V
Negative Supply Current (A)
IEE
VS=+/-15V
Input Offset Voltage (V)
VOS
VS=+/-15V
Input Offset Current (A)
IOS
VS=+/-15V
Positive Bias Current (A)
IB+
VS=+/-15V
Negative Bias Current (A)
IB-
VS=+/-15V
Large Signal Voltage Gain 2K (V/mV) AOL_RL2K VS=+/-15V VO= +/-10V RL=2K
Large Signal Voltage Gain 1K (V/mV) AOL_RL1K VS=+/-15V VO= +/-10V RL=2K
Common Mode Rejection Ratio (dB)
CMRR
VS=+/-15V VCM= +/-11V
Power Supply Rejection Ratio (dB)
PSRR
VS=+/-10V to +/-18V
Positive Slew Rate (V/µs)
Slew+
AV=1 RL=2K VO=+/-5V
Negative Slew Rate (V/µs)
Slew-
AV=1 RL=2K VO=+/-5V
Positive Output Voltage Swing (V)
VOUT+
VS=+/-15V RL=2K
Negative Output Voltage Swing (V)
VOUT-
VS=+/-15V RL=2K
An ISO 9001:2008 and DSCC Certified Company
40
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
TID Report
11-307 110624 R1.0
Table C.2. Measured parameters, pre-irradiation specifications and measurement precision for the RH1056AMW
Precision, High Speed, JFET Input Operational Amplifier.
Pre-Irradiation Specification
Parameter
MIN
MAX
Positive Supply Current (A)
Measurement Precision/Resolution
6.50E-03
±3.20E-05
Negative Supply Current (A)
-6.50E-03
±3.20E-05
Input Offset Voltage (V)
-3.00E-04
3.00E-04
±1.04E-05
Input Offset Current (A)
-1.00E-11
1.00E-11
±1.52E-12
Positive Bias Current (A)
-5.00E-11
5.00E-11
±4.00E-12
Negative Bias Current (A)
-5.00E-11
5.00E-11
±4.00E-12
Large Signal Voltage Gain 2K (V/mV)
1.50E+02
±1.60E+01
Large Signal Voltage Gain 1K (V/mV)
1.30E+02
±1.60E+01
Common Mode Rejection Ratio (dB)
8.60E+01
±6.39E-01
Power Supply Rejection Ratio (dB)
9.00E+01
±1.52E-01
Positive Slew Rate (V/µs)
1.00E+01
±4.10E-01
Negative Slew Rate (V/µs)
-1.00E+01
Positive Output Voltage Swing (V)
1.20E+01
±4.10E-01
±1.60E-02
Negative Output Voltage Swing (V)
-1.20E+01
An ISO 9001:2008 and DSCC Certified Company
41
±1.60E-02
TID Report
11-307 110624 R1.0
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Appendix D: List of Figures Used in the Results Section (Section 5)
5.1.
5.2.
5.3.
5.4.
5.5.
5.6.
5.7.
5.8.
5.9.
5.10.
5.11.
5.12.
5.13.
5.14.
Positive Supply Current (A)
Negative Supply Current (A)
Input Offset Voltage (V)
Input Offset Current (A)
Positive Bias Current (A)
Negative Bias Current (A)
Large Signal Voltage Gain 2K (V/mV)
Large Signal Voltage Gain 1K (V/mV)
Common Mode Rejection Ratio (dB)
Power Supply Rejection Ratio (dB)
Positive Slew Rate (V/µs)
Negative Slew Rate (V/µs)
Positive Output Voltage Swing (V)
Negative Output Voltage Swing (V)
An ISO 9001:2008 and DSCC Certified Company
42