FUJI EDS3-104C

INFRARED CO2 CONTROLLER
DATA SHEET
ZFP9
This is a non-dispersion type infrared CO2 gas analyzer. It
accurately measures the concentration of CO2 gas in the air,
indicates the result and simultaneously outputs an analog
signal. Since upper and lower limit alarm functions are standard it can also be used as a controller.
This instrument is suitable for use as a CO2 meter in green
houses, building ventilation systems, CA (Controlled Atmosphere) storage facilities, and so on.
FEATURES
1.
2.
3.
4.
Use of high-performance infrared-ray system
Excellent stability and easy maintenance
Control with upper/lower limit alarm functions possible
Compact and lightweight (Approx. 3kg)
SPECIFICATIONS
Functional specifications
Measuring system:
NDIR (non-dispersion infrared ray system)
Measured gas: CO2 in the air
Measuring range: 0 to 0.2%, 0 to 0.3%, 0 to 1%, 0 to 5%,
0 to 10%, 0 to 20% CO2 (as specified)
Output signal:
4 to 20mA DC nonlinear (max. load resistance 350Ω)
or 0 to 100mV DC nonlinear (output resistance 100Ω)
Note) Linear ouput is available only for 0 to 0.2%
range, 4 to 20mA DC or 0 to 100 mV DC.
Alarm function: Alarm setting range; 0 to 100% FS
Setting method; set value display on
indicator with internal upper/lower
limit setting VR's
Output; relay contact (2 x 1c), 250V AC,
3A (resistive load)
Alarm hysteresis width; ± 4% of scale
length (at 50% FS)
Indication; upper limit indicator lamp (red)
— lights up when CO2 concentration
exceeds upper limit.
lower limit indicator lamp (red)
— lights up when CO2 concentration
is below the lower limit.
Indicator:
CO2 concentration actual scale, moving
coil type, JIS 2.5 class
Power supply:
100, 115, 200, 220V AC ± 10%, 50/60Hz
(as specified)
Power consumption:
Approx. 18VA
Ambient temperature:
0 to 40°C
Ambient humidity:
90% RH or less
Storage temperature:
–20 to 50°C
Sample gas temperature:
0 to 50°C
Gas sampling: Aspirating pump (with power ON-OFF
switch), membrane filter built in, sample
gas flow rate approx. 0.6r/min
Performance specifications
Repeatability:
Zero point drift:
Response time:
Warmup time:
Linearity:
± 1% FS
Within ± 10%/6 months
Within 10 seconds (for 90% response)
Approx. 30 minutes
± 3% FS (only for 0 to 0.2% range)
Structure and material
Enclosure:
Case material:
Indoor dust-proof type
Case cover (ABS resin), base plate (steel
plate)
Surface color:
Case (Munsell 3.1Y7.0/1.1)
Lid (Munsell 9.2YR4/1)
Material of parts contacting gas:
SUS304, corrosion-resistant aluminum
Mounting:
Wall mounting
Outer dimensions (H x W x D):
257 x 220 x 85mm
Mass {Weight}: Approx. 3kg
Piping:
4.8mm dia. hose end (sample gas inlet)
Wiring:
M3 screw terminal
EDS3-104c
Date
Apr. 1, 2011
ZFP9
SCOPE OF DELIVERY
CODE SYMBOLS
Controller and standard accessories (filter paper – 5 sheets,
fuse – 2)
1 2 3 4 5 6 7 8
Z F P 9
Standard gas for zero/span calibration (type ZBM)
Pressure regulator for standard gas (type ZBD610)
Flowmeter for zero/span calibration (type ZBD472)
OUTLINE DIAGRAM
Measuring range
0 to 0.3%
0 to 1%
0 to 5%
0 to 10%
0 to 20%
0 to 0.2%
A
B
C
D
E
G
ITEMS TO BE ORDERED SEPARATELY
Output signal
DC 0 to 100mV nonlinear
DC 4 to 20mA nonlinear
DC 0 to 100mV linear*
DC 4 to 20mA linear*
A
B
C
D
1
2
4
5
(Unit:mm)
Description
1
Power supply
100V AC 50/60Hz
115V AC 50/60Hz
220V AC 50/60Hz
200V AC 50/60Hz
* Linear output is available only for 0 to 0.2% range.
(5th code : G)
220
257
85
Rear panel diagram
7
137
23
8
16
R3.5
12mm dia.
35
Mounting hole
36
44
Sample gas inlet
(for 4.8mm dia. hose end)
21.5
22.5
137
Cable inlets (22mm dia.)
(with rubber bushing)
28
120
7mm dia.
mounting hole
2
35
Cable inlets (22mm dia.)
(with rubber bushing)
Sample gas outlet
(for 4.8mm dia. hose end)
(attached for more than 0 to 1% range)
CONNECTION DIAGRAM
1
2
3
100V, 115V
Ground 200V, 220V
terminal AC 50/60Hz
power supply
(as specified)
4
5
6
7
8
9
NO
COM
NC
NO
COM
NC
10
11
OUTPUT
Lower limit
Upper limit
Output signal
0 to 100mV DC
or 4 to 20mA DC
Alarm output
Lower limit alarm:4 – 5 become conductive when CO2
sample gas drops below set value, while 5 – 6
tive when concentration exceeds set value.
Upper limit alarm:8 – 9 become conductive when CO2
sample gas drops below set value, while 7 – 8
tive when concentration exceeds set value.
concentration in
become conducconcentration in
become conduc-
CAUTIONS ON WIRING
Use vinyl insulated cable with a cross-sectional area of
1.25mm2 or more for the power supply and alarm contact
output.
Use two-core shielded cable for the output signal cable in
order to avoid the influence of inductive noise.
Be sure to connect shielded wires to the ground terminal.
Keep the output signal cable away from the power line and
alarm contact output line.
The product conforms to the requirements of the
Electromagnetic compatibility Directive 89/336/EEC as
detailed within the technical construction file number
TZ734574. The applicable standards used to demonstrate
compliance are :
EN 50081-1 : 1991 CLASS A Conducted and Radiated emissions
EN 50082-1 : 1992
Radiated immunity, ESD and FBT
ZFP
3
ZFP9
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan