EXTECH FL700

User Guide
ExStik® FL700
Fluoride Meter
Introduction
The model FL700 is a system specifically designed for the quick and accurate measurement
of fluoride ions in drinking water and other aqueous samples. Unlike other electrode based
systems the FL700 consists of the sensing electrode, measuring electronics, and the display
in one convenient package. This meter is shipped fully tested; with proper use, this
instrument will provide years of reliable service.
Features
o
Automatic temperature compensation (±10 C of calibration temperature)
Automatic calibration
Stability sensing to optimize accuracy
Internal Datalogger for storing up to 25 readings
Direct reading of ppm units
Direct reading of relative mV units
Automatic shut down after 12 minutes to preserve battery life
Internal error detection
2
FL700
V1.1
11/07
Description
Meter Description
1. Battery compartment cover
2. LCD Display
3. MODE/HOLD button
4. CAL/RECALL button
5. ON/OFF button
6. Electrode Retaining Collar (ring)
7. Electrode Sensor
(Note: The Electrode storage cap is not shown this diagram)
Electrode Sensor Description
The sensing electrode is a europium doped lanthanum fluoride single
crystal that has been incorporated into a removable sensing module that
houses a reference electrode and temperature measurement system. The
high resistance electrode signals are impedance converted to a low resistance output in the
sensing module to ensure stable and noise free performance.
Reagent Tablets
The FL700 allows the users to follow the ASTM and EPA standard methodology using any
of the TISAB reagents and standards already in use. Also, Extech has developed tablet-form
TISAB which contains all of the essential and approved chemicals that are found in the
usual TISAB reagents.
The “dry” reagent does not contribute to sample dilution. The benefits of this method are:
1. No volumetric errors
2. Independent of sample size (± 20% of nominal sample volume)
3. Easy to use in the field or laboratory
4. Can be shipped more easily than liquid reagents
5. Lower cost per test
LCD Display
1. Bargraph display
1
2. Measurement units
2
3. Main display
4. Low (L) and High (H) Calibration icons
3
6
7
4
5
5. Data HOLD indicator
6. Low Battery indicator
7. Temperature display
3
FL700
V1.1
11/07
Quick Start
The following procedure describes a quick-start for the FL700 using the standard TISAB
chemistry. Detailed operating instructions can be found on the pages that follow.
Preparation
1.
Remove the FL700, electrode module and sample cup from its box. Remove caps from
module.
2.
Fit the electrode module onto the end of the meter body, making sure that the slots line
up correctly, and tightly turn the module retaining ring to secure the assembly.
3.
Wipe the fluoride crystal and reference junction with a damp tissue
Calibration
1.
Prepare a 1 ppm fluoride standard and TISAB reagent, or use pre made mixed TISAB
and 1 ppm standard
2.
Pour 15 – 20 mls of this standard solution into the sample cup
3.
Rinse the end of the FL700 module in TISAB solution and wipe thoroughly with paper
tissue
4.
Place the FL700 into the 1 ppm standard
5.
Switch the instrument on using the ON/OFF key. The instrument will now go through its
internal calibration
6.
1.0 ppm reading will stabilizes in ~35 seconds and the instrument will enter HOLD
mode
7.
When in HOLD mode press the CAL key and hold until 1.0 ppm and CAL appear in the
display. Release the CAL key
8.
Wait until the display stops blinking; the instrument will enter the HOLD mode
9.
The instrument is now calibrated and ready for use
Measurement
1.
Prepare unknown solution by adding TISAB reagent to the sample. Thoroughly wipe
the end of the FL700
2.
Place the FL700 into the prepared unknown sample
3.
If the display is indicating HOLD press the HOLD key to enter the Measure mode (the
HOLD display will switch off)
4.
After ~ 35 seconds the instrument will display the value of the unknown concentration
5.
The readings can be stored in memory by pressing the MODE/HOLD key for ~ 3
seconds
Storage and maintenance
1.
After use, store the electrode in an analyzed sample
2.
Thoroughly wipe the sensor with paper tissue. The flat ended sensors can be wiped
vigorously
3.
The fluoride module can be replaced once the automatic calibration no longer
sufficiently calibrates the instrument
4.
Other maintenance information is provided in a later section of this guide
4
FL700
V1.1
11/07
Operation
Preparation for use
1.
Remove the FL700, electrode module and sample cup from its box. Remove caps from
module
2.
Fit the electrode module onto the end of the meter body, making sure that the slots line
up correctly, and tightly turn the module retaining ring to secure the assembly
3.
Wipe the fluoride crystal and reference junction with a damp tissue (cleaning
techniques are addressed later in this guide)
Powering the FL700
The FL700 uses four (4) CR2032 lithium ion batteries. Press the ON/OFF button to turn
the meter on or off. If the batteries are weak, the ‘BAT’ indicator appears on the LCD. The
auto power off feature shuts the meter off automatically after approximately 12 minutes of
inactivity. The auto power off feature may be temporarily disabled for convenience or for
extended polarization time.
Power-On Diagnostics
1.
When the meter is switched ON the LCD displays “SELF” and “CAL” while the meter
runs a diagnostic routine
2.
During this time the meter is recalling the user calibration data, performing self
diagnostics & initializing the circuitry
3.
When completed, the meter proceeds to the normal measurement mode
Calibration
The FL700 can be calibrated between 1.0 ppm and 10 ppm or between 0.5 and 5.0 ppm
Fluoride ion. The following calibration procedure assumes the normal 1.0 to 10 ppm range
has been chosen.
1. Prepare a 1 ppm fluoride standard and TISAB reagent, or use pre made mixed TISAB
and 1 ppm standard
2.
Pour 15 – 20 mls of this standard solution into the sample cup
3.
Rinse the end of the FL700 module in TISAB solution and wipe thoroughly with paper
tissue
4.
Place the FL700 in the 1 ppm standard and switch the instrument ON using the
ON/OFF key. The instrument will now run its self-calibration
5.
The instrument will enter the HOLD mode when stabilized in the 1 ppm solution
6.
Press the CAL key; CAL will appear in the display followed by 0.5ppm and 5.0 ppm.
Continue holding until 1.0 ppm is shown. Release the CAL key. After the display stops
blinking the instrument will enter the HOLD mode
7.
The instrument is now calibrated and ready for use. The circled L and H icons on the
display indicate that the low range (L) and high range (H) calibrations have been
completed
Calibration frequency
Calibrate the FL700 prior to each new measurement batch or if more than 12 hours has
elapsed since the last calibration.
5
FL700
V1.1
11/07
Slope Adjustment
1. Slope adjustment although not a frequent requirement can be carried out by following the
instructions in Calibration steps above except for the fact that a 10 ppm standard is used
after calibrating with the 1 ppm standard
2. Press the Cal button until 10 ppm appears. Slope adjustment is then complete
Other standards
As mentioned the FL700 can also be calibrated between 0.5 and 5.0 ppm F. Follow the
calibration instructions above but substitute 0.5 ppm for 1.0 ppm and 5.0 ppm for 10 ppm.
Measurements
1.
Prepare an unknown solution by adding TISAB reagent to the sample in the same dilution
ratio as for the calibration procedure. Mix thoroughly
2.
Rinse the end of the FL700
3.
Place the FL700 into the prepared unknown sample. If the instrument is in the HOLD
mode, press MODE/HOLD to unlock HOLD
4.
After 25 seconds, the instrument will display the value of the unknown concentration and
will then enter the HOLD mode
Note: The readings can be stored in the memory by pressing the MODE/HOLD key for ~ 3
seconds as explained in a subsequent section of this user guide.
Temperature Units (ºF / ºC)
1.
With the unit OFF, press and hold the CAL/RECALL button
2.
With the CAL/RECALL button depressed, momentarily press the ON/OFF button to turn
the unit ON
3.
The CAL/RECALL button can be released when ‘Self Cal’ is shown in the display
4.
To switch back to the previous unit of measure, repeat steps 1 through 3.
Auto-Power OFF Feature
The auto power off feature automatically shuts the meter off 12 minutes after the most recent
button press.
Disabling the Auto-Power OFF Feature
With the unit ON, momentarily press the CAL/RECALL button, then quickly press and hold
both the MODE/HOLD and ON/OFF buttons until ‘oFF’ is displayed. To restore the Auto Power
Off Feature (auto power OFF enable) simply turn the meter off and on again using the ON/OFF
button.
Low Battery Indication
When the battery voltage falls below the operating threshold, ‘BAT’ will appear on the display.
Refer to the Maintenance section for battery replacement information.
6
FL700
V1.1
11/07
Storing Readings
Up to 25 readings can be stored in memory for later recall.
1. With the meter in the HOLD mode, press and hold the MODE/HOLD button for three (3)
seconds to store a reading. Release the button when the memory location number
appears on the lower display.
2.
After approx. 30 seconds (measurement duration) the meter will return to the HOLD mode
and another reading can then be stored.
3.
If more than 25 readings are stored, previously stored readings (starting with reading
number 1) are overwritten.
Recalling Stored Readings
1.
Momentarily press the CAL/RECALL button and then within 4 seconds momentarily
press the MODE/HOLD. The last stored data point location will be displayed (1 to 25).
Each time the MODE/HOLD button is momentarily pressed the next most recently stored
data point will be displayed.
2.
After the last data point is displayed, pressing the MODE/HOLD button again returns the
display to the beginning of the list.
3.
Pressing the CAL/RECALL button at anytime stops the data retrieval process and returns
the meter to the normal measurement mode.
Clearing Stored Readings
1.
With the unit ON press and hold the ON/OFF button for 4 seconds
2.
When “clr” is shown in the main display the memory is cleared.
7
FL700
V1.1
11/07
Maintenance
Battery Replacement
1.
Twist off the battery compartment cover
2.
Holding the battery housing in place with a finger, pull out the battery carrier using the two
small tabs
3.
Replace the four (4) CR2032 batteries observing proper polarity
4.
Replace the battery carrier, reattach the battery compartment cap and tighten securely
Electrode Replacement
1.
To remove the electrode, first turn the instrument OFF and then unscrew and remove the
electrode retaining collar. (turn the collar counter-clockwise to remove)
2.
Gently rock the electrode from side to side, pulling it away from the meter until it
disconnects
3.
To attach an electrode, align the positioning “keys” on the electrode and the main body
housing and then carefully push the electrode into the meter socket until it is fully seated
4.
Tighten the electrode retaining collar firmly enough to seal the electrode with the meter
Electrode Storage
1.
The module and can be stored wet or dry. If stored dry it will be necessary to allow
approximately 15 minutes of soaking in a fluoride solution before the specified
performance can be achieved. It is recommended that the electrode be stored wet in the
last test solution used by the instrument (fluoride ion plus TISAB reagent).
2.
The instrument will give an error code when the electrode can no longer be calibrated
3.
If the instrument will not calibrate, clean the fluoride electrode surface with the Extech
cleaning paste and recalibrate the instrument. If the meter still does not calibrate, replace
the electrode.
Extech cleaning paste
Extech offers a unique cleaning procedure for the lanthanum fluoride crystal that removes any
coating from the surface of the crystal and dramatically increases response time. This is
available from Extech and a sample of this paste is shipped with the product.
8
FL700
V1.1
11/07
Specifications
Range
0.10 to 9.99ppm (mg/l)
Accuracy
± 3% of reading or ± 0.1ppm (whichever is greater)
Resolution
0.1ppm
Display
2000 count, Dual function 3 ½ digit LCD with Bargraph,
Display size: 24 mm x 20 mm
Electrode
Europium doped lanthanum fluoride single crystal
Electrode life
6 months minimum
Response Time
90% of change in less than 30 seconds (typical)
Operating Temp. Range
32 to 140°F (0 to 60°C)
ATC Range
32 to 140°F (0 to 60°C)
Measurement Storage
25 tagged (numbered) data sets with recall
Battery Power
Four (4) CR2032 button batteries
Low Battery Indication
‘BAT’ appears on the LCD
Auto Power Off
After 12 minutes of inactivity
Dimensions/Weight
1.4 x 6.8 x 1.6” (36 x 173 x 41mm); 7.4 oz (210g)
9
FL700
V1.1
11/07
Warranty
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in
parts and workmanship for one year from date of shipment (a six month limited warranty
applies to sensors and cables). If it should become necessary to return the instrument for
service during or beyond the warranty period, contact the Customer Service Department at
(781) 890-7440 ext. 210 for authorization or visit our website www.extech.com for contact
information. A Return Authorization (RA) number must be issued before any product is
returned to Extech. The sender is responsible for shipping charges, freight, insurance and
proper packaging to prevent damage in transit. This warranty does not apply to defects
resulting from action of the user such as misuse, improper wiring, operation outside of
specification, improper maintenance or repair, or unauthorized modification. Extech specifically
disclaims any implied warranties or merchantability or fitness for a specific purpose and will not
be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is
limited to repair or replacement of the product. The warranty set forth above is inclusive and no
other warranty, whether written or oral, is expressed or implied.
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also provides
NIST certification for most products. Call the Customer Care Department for information on
calibration services available for this product. Extech recommends that annual calibrations be
performed to verify meter performance and accuracy.
Support line (781) 890-7440
Technical support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451
Copyright © 2007 Extech Instruments Corporation.
All rights reserved including the right of reproduction in whole or in part in any form.
10
FL700
V1.1
11/07