EXTECH EC500

User's Guide
pH/Conductivity / TDS / Salinity /
Temperature Meter
ExStik® EC500
EC500
Introduction
®
Congratulations on your purchase of the ExStik EC500 pH/Conductivity/Total Dissolved
Solids (TDS) / Salinity meter. With the EC500’s dynamic cell-constant technology it is
possible to measure a wide range of Conductivity, TDS, and Salinity with the same
electrode. Careful use and maintenance will provide years of reliable service.
Powering the ExStik™
®
The ExStik uses four (4) CR2032 Lithium Ion Batteries (included). If the batteries are
®
weak, the ‘BAT’ indicator appears on the LCD. Press the ON/OFF key to turn the ExStik
®
on or off. The auto power off feature shuts the ExStik off automatically after 10 minutes of
inactivity to preserve battery life.
Getting Started
•
•
•
•
•
•
•
For new meters, remove the battery cap and then remove the battery insulating strip.
Remove the cap from the bottom of the ExStik to expose the pH electrode, reference
junction and conductivity electrodes.
Before the first use or after storage, soak the electrode in tap water or pH 4 buffer
solution for about 10 minutes.
White KCL crystals may be present in the cap or on the electrode. This is to be
expected depending on the length of time in storage. These crystals will dissolve
while soaking the electrode or they can be rinsed away with tap water.
For best results calibrate with pH 7 buffer solution first, then calibrate with the buffer
solution closest to the expected pH value of the solution or material to be tested.
To preserve the pH electrode life keep the sponge in the protective cap soaked with
tap water or pH 4 buffer solution.
For best results, calibrate for conductivity with a standard in the expected range of
the sample. For maximum accuracy calibrate from low conductivity value standards
to high value standards.
Warranty
EXTECH INSTRUMENTS CORPORATION (A FLIR COMPANY) warrants this instrument to be free
of defects in parts and workmanship for one year from date of shipment (a six month limited
warranty applies to sensors and cables). If it should become necessary to return the instrument for
service during or beyond the warranty period, contact the Customer Service Department at (781)
890-7440 ext. 210 for authorization or visit our website www.extech.com for contact information. A
Return Authorization (RA) number must be issued before any product is returned to Extech. The
sender is responsible for shipping charges, freight, insurance and proper packaging to prevent
damage in transit. This warranty does not apply to defects resulting from action of the user such as
misuse, improper wiring, operation outside of specification, improper maintenance or repair, or
unauthorized modification. Extech specifically disclaims any implied warranties or merchantability
or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or
consequential damages. Extech's total liability is limited to repair or replacement of the product.
The warranty set forth above is inclusive and no other warranty, whether written or oral, is
expressed or implied.
2
EC500 V2.3 3/08
Specifications
Display
2000 count LCD with Bargraph
pH Range
0.00 to 14.00
pH Accuracy
±0.01 pH typical
pH ATC Range
0°C to 90 C (32°F to 194 F)
pH Reference Junction
Permanent gel, non-refillable
Conductivity ranges
0 to 199.9µS/cm
200 to 1999µS/cm
2.00 to 19.99mS/cm
TDS ranges
0 to 99.9ppm or mg/L
(Variable ratio)
100 to 999ppm or mg/L
1.00 to 9.99ppt or g/L
o
o
Salinity range
0 to 99.9ppm
(Fixed ratio of 0.5)
100 to 999ppm
1.00 to 9.99ppt
TDS Ratio
0.4 to 1.0 adjustable
Salinity Ratio
0.5 Fixed
Conductivity ATC
2.0% per C
Temperature Range
-5.0°C to 90.0 C (23.0°F to 194 F)
Temperature Resolution
0.1 up to 99.9, 1 >100
Temperature Accuracy
±1°C; 1.8°F (from -5 to 50°C; 23 to 122°F)
Conductivity ATC Range
0.0°C to 60.0 C (32.0°F to 140 F)
Accuracy
Conductivity: ±2% full scale
TDS: ±2% full scale
Salinity: ±2% full scale
Measurement Storage
25 tagged (numbered) readings
Low battery indication
‘BAT’ appears on the LCD
Power
Four (4) CR2032 Lithium Ion Batteries
Auto power off
After 10 minutes (override available)
Operating conditions
-5°C to 50 C (23°F to 122 F)
Dimensions
40 x 200 x 40 mm (1.6 x 7.9 x 1.6”)
Weight
93 g (3.3 oz)
o
o
o
±3°C; 5.4°F (from 50 to 90°C; 122 to 194°F)
o
o
3
o
o
EC500 V2.3 3/08
Meter Description
Front Panel Description
1. Battery compartment cover
2. LCD Display
3. MODE/HOLD button
4. CAL/RECALL button
5. ON/OFF button
6. Electrode Collar
7. pH/Conductivity Electrode
(Note: The Electrode cap is not shown)
EC500
LCD Display
1. Bargraph display
2. Measurement units
3. Main display
4. Range calibration and
low battery indicators
5. Temperature display
pH
6. Renew indicator
7. Reading hold indicator
6
7
4
RENEW
EC500 V2.3 3/08
Measurement Procedure
Sample Preparation:
1.
For Conductivity, TDS or Salinity place the test sample in
a sample cup with enough depth (2.5cm minimum) to
cover the electrode. Stir the solution to remove any air
bubbles.
2.
For pH, place the tip of the electrode in the sample or
make contact with a wet surface.
Conductivity
TDS
Salinity
PH
Measurement:
8888
1.
Press the ON button. (
and then “SELF CAL” will
appear in the display during the turn-on diagnostics)
2.
Depress and hold the MODE/HOLD key to scroll to the desired
measurement mode.
3.
Insert the electrode into the sample making sure that the
electrodes are completely submersed.
4.
Slowly stir the solution with the electrode to remove air bubbles
if in the Conductivity, TDS or Salinity mode.
5.
If in the Conductivity, TDS or Salinity modes, the meter will
auto-range to the proper range and then display the reading.
(ppm)
EC500
(mg/l)
mgL
pH
L
M
H
5
EC500 V2.3 3/08
Changing Measurement Function
The meter can be set to measure pH, Conductivity, TDS or Salinity.
To change the mode:
1. Press and Hold the MODE/HOLD button for 2 seconds and the display will begin to
scroll through the units.
µS (Conductivity); pH; ppm S (Salinity); ppm (TDS); mg/l (TDS);
Note: The “HOLD” function cannot be on when changing the measurement function. If
“HOLD” is displayed in the lower left corner of the display, briefly press the MODE/HOLD
button to turn it off.
2. When the desired units are displayed, release the MODE/HOLD button.
TDS Compensation Ratio
The TDS value is determined by multiplying a conductivity reading by a known ratio
factor. The meter allows for selecting a conversion ratio in the range of 0.4 to 1.0. The
ratio varies with the application, but is typically set between 0.5 and 0.7.
Note: The stored ratio will briefly appear in the lower temperature display when the meter
is first turned on, or when changing measurement function to TDS.
Note: In the Salinity mode the ratio is fixed at 0.5.
To change the ratio, while in the TDS measurement mode (ppm or mg/l):
1. Press and release the CAL/RECALL button twice in succession.
The stored ratio will appear in the display.
2. Press the MODE/HOLD button to increase the ratio value in steps of
0.1.
3. When the desired ratio is displayed, press and release the
CAL/RECALL button to store the value and return to the normal
mode.
4. If no buttons are pressed for 5 seconds, the meter returns to measure mode.
Storing Readings
1. Press the MODE/HOLD button to store a reading. The storage
location number will be displayed on the lower display, while the
main display shows the stored reading. The meter will enter the
HOLD mode and the “HOLD” indicator will appear.
2. Press the MODE/HOLD button again to exit the HOLD mode and
return to normal operation.
3. If more than 25 readings are stored, previously stored readings
(starting with number 1) will be overwritten.
6
EC500 V2.3 3/08
Recalling Stored Readings
1. Press the CAL/RECALL button and then press the MODE/HOLD button. A location
number (1 through 25) will briefly appear and then the value stored in that location will
appear. The displayed units will flash, indicating that the storage recall mode is active.
19.5 °C
2. The last stored reading will be displayed first. Pressing and releasing the MODE/HOLD
button will scroll through the stored readings one at a time. The location number is
displayed first, followed by the reading stored in that location.
3. To exit the storage mode, press the CAL/RECALL button and the meter will return to
normal operation, after displaying “End”.
Clear Stored Memory
With the unit on, press and hold ON/OFF for 4 seconds. “clr” will be briefly displayed
when the memory is cleared.
Changing Temperature Units
o
o
To change the displayed temperature units ( C or F):
1. With the unit OFF, press and hold down the CAL/RECALL button.
2. With the CAL/RECALL button depressed momentarily press the ON/OFF button.
When “SELF CAL” appears in the display release the CAL/RECALL button. The unit
will power on with temperature displayed in the new units.
Data Hold Mode
Press the MODE/HOLD button to hold (freeze) a reading in the display. The meter will
enter the HOLD mode and the “HOLD” indicator will appear.
Note: This also stores the reading.
Press the MODE/HOLD button again to return to normal operation.
Auto Power OFF
The auto power off feature automatically shuts the meter off 10 minutes after the most
recent button press.
Auto Power OFF Disengage Feature
To disable the Auto Power Off feature:
1.
Turn the unit on
2.
Press CAL/RECALL once (Quickly)
3.
Immediately and simultaneously press the MODE/HOLD and ON/OFF buttons for
approximately 2 seconds, until “oFF” is briefly displayed
To disengage this feature, turn the unit off with the ON/OFF button. The next time the unit
is powered up, Auto Power OFF mode will be engaged again.
Low Battery Indication
When the batteries become weak the “BAT” icon will appear in the display. Refer to the
Maintenance section for battery replacement information.
7
EC500 V2.3 3/08
Calibration - pH (1, 2, or 3 points)
1.
2.
3.
4.
5.
6.
Place the electrode into a buffer solution (4, 7, or 10). Press and hold the
CAL/RECALL key until “CAL” appears in the lower (temp.) display. When doing a 2
or 3 point calibration, calibrate with pH 7 buffer first, then follow with pH 4 then the pH
10 buffer.
®
The ExStik automatically recognizes the solution and calibrates itself to that value
(the circled number on the LCD will match the solution). Note that if the solution is
more than 1 pH unit off from the L (4), M (7), or H (10) pH buffer, or if the electrode
®
slope is low, the ExStik will assume an error and abort the calibration (End will be
displayed, and the unit will return to measure mode.)
During calibration, the pH reading flashes on the main display.
®
When calibration is complete, the ExStik automatically displays “SA”, then “End” and
returns to normal operation mode.
The appropriate circled indicator (L, M, or H) appears on the LCD when a particular
calibration or series of calibrations has been completed within one power on cycle.
®
When the ExStik is turned off, the circled indicator configuration and the calibration
data will be retained.
For a two or three point calibration, repeat steps 1-4.
CAL Reminder Display
When in pH measurement mode, a “CAL” icon will appear after 15 on/off cycles of the
meter without performing a calibration. The CAL display is simply a reminder to calibrate
pH, and will turn off when the pH electrode is recalibrated. The reminder does not affect
function in any way.
RENEW Display
A flashing ‘RENEW’ warning indicates that the probe is not performing to expected
specifications. If cleaning and recalibration does not cause the RENEW icon to disappear,
replace the probe (see optional accessories on the last page of this manual). The RENEW
display appears as a result of the pH electrode slope falling below 70% of a nominal slope.
Measurement and Display Considerations
•
•
If the unit appears to be locked (display frozen). It is possible that the Data Hold mode
has been inadvertently accessed by pressing the MODE/HOLD button. (“HOLD” will
be displayed in the bottom left of the LCD.) Simply press the MODE/HOLD button
again or turn the meter off and then on.
For maximum accuracy, allow sufficient time for the temperature of the probe to reach
the temperature of the sample before calibrating. This will be indicated by a stable
temperature reading on the display.
8
EC500 V2.3 3/08
Calibration - Conductivity
Meter accuracy verification should be performed on a periodic basis. Once per month is
the recommended cycle for normal use. If calibration is required, a conductivity
standardizing solution must be obtained. The meter can be calibrated in any or all of the
three ranges. Standardizing solutions of 84µS/cm, 1413µS/cm or 12.88mS/cm
(12,880µS/cm) are used for the automatic calibration recognition procedure. No other
calibration values are permitted.
Calibration is always done in conductivity mode. Since salinity and TDS values are
calculated from conductivity values, this procedure also calibrates the salinity and TDS
ranges.
1.
2.
3.
4.
5.
Fill a sample cup with the standardizing solution.
Turn the meter ON and insert the electrode into the solution. Tap or move the
electrode in the sample to dislodge any air bubbles.
Press and hold the CAL/RECALL button (approximately 2 seconds) until “CAL”
appears in the lower (temp) display. The main display will start flashing.
The meter will automatically recognize and calibrate to the standardizing solution.
The display will briefly indicate “SA”, End and then return to the measurement mode
after a calibration.
Note: The “SA” will not appear if the calibration fails.
The “range calibrated” symbol will appear in the display for each range that is
calibrated during that power on cycle.
Low range, 84µS/cm
Medium range, 1413µS/cm
High range, 12.88mS/cm (12,880µS/cm)
Note: Each time the calibration mode is entered all calibration symbols on the display are
cleared, but only the calibration data for the currently calibrated range is replaced. The
other two ranges keep the existing calibration data, just the symbols are removed.
Calibration of all three ranges must be performed during one power on period for all three
range calibration symbols to appear.
Note: The meter allows for a 1, 2 or 3 point calibration. If calibration is done for more than
one point the lowest value standard should be done first to obtain the best accuracy.
Considerations and Techniques
•
•
•
•
•
•
Do not touch the inner surfaces of the conductivity electrodes. Touching the surface
of the platinized electodes may damage and reduce the life of the probe.
Store the electrode in the wetting cap with the sponge moistened with pH 4.01 buffer
solution.
Always rinse the electrode in de-ionized water between measurements to avoid cross
contamination of the sample. Double rinsing is recommended when high accuracy is
required.
Periodically, accumulated salt deposits from the reference electrode may build up in
the storage cap, and should be rinsed away. These deposits could affect measured
values of low conductivity samples.
When measuring low conductivity samples, extra care is recommended in rinsing the
probe to avoid contamination of the sample with electrolyte from the pH reference
electrode. This will only be a factor when measuring in the low range, and can be
further minimized by increasing the volume of the sample. (Example: Try a 200 to 500
mL sample.)
If the 20mL sample cup is to be used, then the electrode should not be allowed to sit
in the sample for any longer than necessary, to avoid electrolyte leakage into the
sample, raising the conductivity value.
9
EC500 V2.3 3/08
Operational Matrix
Function /
Resulting Action
On/Off
Power
Status
On or Off
Mode
Setting
Required Key Press Sequence
Any
Momentary press of the ON/OFF key
Calibration
On
pH or
Conductivity
Press & hold CAL/RECALL key for 2
seconds, until it enters CAL function
Store Reading
On
Any measure
mode
Momentary press of the
MODE/HOLD key
Hold Release
On
While In Hold
Mode
Momentary press of the
MODE/HOLD key
Enter Memory
Retrieval
On
Any measure
mode
Momentary press of the
CAL/RECALL key followed by a
momentary press of the
MODE/HOLD key (Within 4 seconds)
Scroll Stored
Readings
On
Memory
Recall
Momentary press of the
MODE/HOLD key
(Displays “last in first out”)
Exit Memory
Retrieval
On
Memory
Recall
Momentary press of the
CAL/RECALL key
Clear Stored
Memory
On
Any Measure
Mode
Press and hold the ON/OFF key for 4
seconds, until “clr” is displayed.
Change
Measurement Mode
On
Any
Press and hold the MODE/HOLD key
for at least 2 seconds (the modes will
scroll by until the key is released)
Enter Cond/TDS
Ratio
On
TDS (ppm or
mg/l)
Press and release the CAL/RECALL
key twice in quick succession
Change Cond/TDS
Ratio
On
TDS ratio
Momentary press of the
MODE/HOLD key (each key press
increases the ratio by 0.1, the value
cycles from 0.4 - 1.0)
Exit Cond/TDS
Ratio
On
TDS ratio
Momentary press of the
CAL/RECALL key
Change
Temperature Units
Off
n/a (off
mode)
Press and hold the CAL/RECALL key
then momentarily press the On/Off
key. Release the CAL/RECALL key
after the "SELF CAL" lights
Override Auto
Power Off
On
Any measure
mode
Momentarily press the CAL/RECALL
key then simultaneously press and
hold the ON/OFF & MODE/HOLD key
for approximately 2 seconds, until
“oFF” is displayed
Default Reset
OFF
n/a
Simultaneously press ON/OFF,
CAL/RECALL and MODE/HOLD
momentarily. “dFLt” will be displayed.
10
EC500 V2.3 3/08
Maintenance
Battery Replacement
1.
Twist off the battery compartment cap
2.
Holding the battery housing in place with a finger, pull out the battery carrier using the
two small tabs.
Replace the four (4) CR2032 batteries observing polarity.
Replace the battery compartment cap
3.
4.
Electrode Replacement
1.
To remove an electrode, unscrew and completely remove the electrode collar (turn
the collar counter-clockwise to remove).
2.
Gently rock the electrode from side to side, pulling it downwards, until it disconnects
from the meter.
3.
To attach an electrode, carefully plug the electrode into the meter socket (note that
the electrode connector is keyed, ensuring proper connection).
4.
Tighten the electrode collar firmly enough to make a good seal (a rubber gasket seals
the electrode with the meter).
11
EC500 V2.3 3/08
Cleaning Recommendations
When cleaning the probe, take care not to scratch or damage the sensing surface or the
platinized electrode surfaces.
Contaminant
Cleaning Solution
Instructions
Water soluble substances
Deionized water
Soak or scrub with a soft
brush. Recondition in 4 or 7
buffer for 1 hour.
Grease & Oil
Warm water and household
detergent
Soak or scrub with a soft
brush, maximum of 10
minutes. Rinse thoroughly
with DI water, recondition in
4 or 7 buffer for 1 hour.
Heavy grease & Oil
Alcohol
Maximum of 5 minute soak,
scrub with a soft brush.
Rinse thoroughly with DI
water, recondition in 4 or 7
buffer for 1 hour.
Lime and hydroxide
coatings
10% acetic acid
Soak until coating dissolved,
maximum of 5 minutes. .
Rinse thoroughly with DI
water, recondition in 4 or 7
buffer for 1 hour.
Please Note: Since the EC500 does not have a refillable reference electrolyte chamber, it is
important not to soak the electrode in the above solutions for more than the recommended
times. To do so may cause a reference potential shift, which will cause degradation in
performance or failure.
12
EC500 V2.3 3/08
Troubleshooting
Problem
Possible Cause
Action
Reading is frozen
Unit is in “HOLD” mode
Press MODE/HOLD key to
exit “HOLD” mode
“BAT” message
Batteries are low
Replace batteries
Unit will not calibrate in pH
Low pH slope
Replace electrode, see
reorder information
Unit will not calibrate in pH
Clogged or contaminated
reference junction
Clean junction (see
cleaning instructions)
Unit will not calibrate in pH
Damaged or worn out
sensing membrane
Replace electrode, see
reorder information
Unit will not calibrate in pH
Contaminated pH buffers
Use fresh buffers
Unit will not calibrate in
conductivity mode
Contaminated conductivity
standards
Use fresh standards
Unit will not calibrate in
conductivity mode
Dirty probe
Clean conductivity probe
(See cleaning instructions)
Unit will not calibrate in
conductivity mode
Damaged conductivity probe
Replace electrode, see
reorder information
Unit will not calibrate in
conductivity mode
Trapped air bubbles
Tap or stir to release air
bubbles
Unit will not turn on
Batteries are low or dead
Replace batteries
Unit will not turn on
Batteries installed with
incorrect polarity
Replace batteries,
observe polarity
“RENEW” message
pH sensor needs
recalibration
Recalibrate unit
“RENEW” message
pH sensor slope has fallen
below acceptable limits
Replace electrode, see
reorder information
Unit will not respond to any
key presses
Internal fault
Perform hard reboot:
Remove batteries, hold
ON/OFF switch down for 5
seconds, replace batteries
Use fresh buffers
Support line (781) 890-7440
Technical support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Copyright © 2008 Extech Instruments Corporation (a FLIR company).
All rights reserved including the right of reproduction in whole or in part in any form.
www.extech.com
13
EC500 V2.3 3/08
14
EC500 V2.3 3/08