Introduction
The ADLINK USB-1900 series provides a range of USB 2.0-based multi-functional DAQ modules.
The USB-1901 and USB-1902 16-bit 250 kS/s USB 2.0-based high-performance DAQ modules allow
four different voltage input ranges, while the USB-1903 features additional built-in precision current-tovoltage resistors capable of direct measurement of current signal from 0 to 20 mA.
The USB-1900 series is USB bus powered and equipped with removable screw-down terminals for easy
device connectivity. The attached multi-functional stand can be used for desktop, rail, or wall mounting.
Suitable for mixed-signal tests, laboratory research, and factory automation, the USB-1900 series provide
a simple measurement solution at an affordable price.
Features
■
■
■
■
■
■
■
■
■
USB 2.0 high-speed function
USB bus power
Removable screw-down terminal module
Lockable USB cable for secure connectivity
16-CH 250 kS/s voltage input (USB-1901/1902)
8-CH 250 kS/s current input (USB-1903)
2-CH 1 MS/s voltage output (USB-1902/1903)
Analog and Digital triggering
Free testing application included (U-Test)
■Operating Systems
• Windows 7/Vista/XP
■Driver Support
• UD-DASK for VB.NET/VC.NET/VB/VC++/BCB/Delphi
• DAQPilot for LabVIEW™
• DAQ-MTLB for MATLAB®
Standard Shipped Accessories
• One pair of 20-pin removable screw
terminals
• Module stand
• 2 M USB Type A to USB Mini-B
cable with lockable connector
• Rail-mount kit
Pin Assignment
USB-1901/1902
ECLK
20
40
AOTG*
NC
19
39
NC
18
GPO3
USB-1903
ECLK
20
40
AOTG
AITG
NC
19
39
AITG
38
GPI7
NC
18
38
GPI7
17
37
GPI6
GPO3
17
37
GPI6
GPO2
16
36
GPI5
GPO2
16
36
GPI5
GPO1
15
35
GPI4
GPO1
15
35
GPI4
GPO0
14
34
GPI3
GPO0
14
34
GPI3
DGND
13
33
GPI2
DGND
13
33
GPI2
AGND
12
32
GPI1
AGND
12
32
GPI1
*AO1
11
31
GPI0
AO1
11
31
GPI0
*AO0
10
30
DGND
AO0
10
30
DGND
AGND
9
29
AISE
AGND
9
29
AISE
AI7(AIL3)
8
28
AI15(AIL7)
CI3-
8
28
CI7-
AI6(AIH3)
7
27
AI14(AIH7)
CI3+
7
27
CI7+
AI5(AIL2)
6
26
AI13(AIL6)
CI2-
6
26
CI6-
AI4(AIH2)
5
25
AI12(AIH6)
CI2+
5
25
CI6+
AI3(AIL1)
4
24
AI11(AIL5)
CI1-
4
24
CI5-
AI2 (AIH1)
3
23
AI10(AIH5)
CI1+
3
23
CI5+
AI1 (AIL0)
2
22
AI9(AIL4)
CI0-
2
22
CI4-
AI0(AIH0)
1
21
AI8(AIH4)
CI0+
1
21
CI4+
* Not available for USB-1901
Ordering Information
■USB-1901
16-CH 16-Bit 250kS/s Analog Input USB DAQ
■USB-1902
16-CH 16-Bit 250kS/s Multi-Function USB DAQ
■USB-1903
8-CH 16-Bit Current Input Multi-Function USB DAQ
2-7
www.adlinktech.com
Optional Accessories
■RST-20P
One pair of 20-pin removable
screw terminals
■USB-2M-L
2 M USB Type A to USB Mini-B cable
with lockable connector
USB DAQ
Specifications
Model Name
USB-1901
USB-1902
USB-1903
Analog Input
Resolution
Number of channels
16-bit
16 SE / 8 Pseudo-diff, voltage input
Maximum sampling rate (single channel)
8 Current inputs
250 kS/s(Multiplexing, channel-gain-queue)
Programmable gain
1, 5, 10, 50 1
Input range (Voltage)
± 10 V, ± 2 V, ± 1 V, ± 200 mV
N/A
Input range (Current)
Offset error
± 0.1 mV (gain=1)
Gain error
± 0.05% of FSR (gain=1)
N/A
0-20 mA
±0.01 mA (typical)
-3dB small signal bandwidth (gain=1)
±0.05% of FSR (typical)
600 kHz
CMRR (gain=1)
90 dB
-
SFDR (gain=1)
108 dB
-
SINAD (gain=1)
89 dB
-
THD (gain=1)
102 dB
-
SNR (gain=1)
89 dB
-
ENOB (gain=1)
14.5-bit
-
FIFO buffer size
4 k samples
Trigger sources
Software, external digital, analog trigger ( from one of analog input channels)
Trigger mode
Post trigger, delay trigger, retrigger, gate trigger
External conversion source
Yes (up to 250 kS/s)
Input coupling
DC
Overvoltage protection
Input impedance
Continuous ± 24 V
High impedance > 1 GΩ
Data transfer
249.5 Ω (input resistor)
Programmed I/O, continuous (USB bulk transfer mode)
Analog Output
Number of channels
N/A
Resolution
-
2 voltage outputs
16-bit
Maximum update rate
-
1 MS/s (simultaneous update)
Output range
-
± 10 V
Offset error
-
± 0.15 mV
Gain error
-
± 0.05% of FSR
INL
-
≦1 LSB
DNL
-
< 1 LSB
Output driving capacity
± 5 mA
Slew rate
-
2.2 V/μs
Settling time (0.1% of Full scale)
-
26 μs
Rising time
-
6 μs
Falling time
-
6 μs
FIFO
-
10 k samples (2-CH sharing)
Output mode
-
Programmed I/O, continuous (USB bulk transfer mode)
Function I/O
Mode*
Digital I/O
General timer/counter
Pulse generation
Digital I/O, general timer/counter, pulse generation
8 DI / 4 DO (TTL level)
Two 32-bit, base clock: 80 MHz, external to 10 MHz
Two PWM outputs (Modulation frequency: 0.01 Hz to 5 MHz; Duty cycle: 1%-99%)
General Specifications
Interface
I/O connector
Operating temperature
Storage temperature
Relative humidity
Power requirements
Dimensions
USB 2.0 high speed, mini-USB connector
Two 20-pin screw terminals
0 to 55°C
-20 to 70°C
5 to 95% non-condensing
5V@400 mA (USB bus powered)
114 mm (H) x 156.5 mm (L) x 41.3 mm (W) (4.5 x 6.16 x 1.63 in.)
(without connector and stand)
Note: The function I/O share the same I/O pins. Only one of these modes can be selected.
Updated 03-28, 2012. ©2012 ADLINK Technology, Inc. All Rights Reserved. All specifications are subject to change without further notice.
www.adlinktech.com 2-8