H HiTRON AC/DC MEDICAL APPLICATION EXTERNAL ADAPTER UNIVERSAL INPUT SINGLE OUTPUT 22-24 WATTS SWITCHING POWER SUPPLIES HES22 SERIES FEATURES: UNIVERSAL AC SOURCES DESKTOP IEC320 AC RECEPTACLE MEDICAL STANDARDS EN60601-1/ UL2601-1 EMI MEET EN55011 / FCC CLASS B CE MARKING COMPLIANCE SPECIFICATION INPUT SPECIFICATION OUTPUT SPECIFICATION Input Voltage: 90-264Vac typical. Input Connector: IEC320-3P DT7 AC receptacle. Input Frequency: 47-63Hz. Inrush Current: 30A @ 230Vac &15A @ 110Vac typical. Input Current: 0.60A @ 115Vac typical. 0.30A @ 230Vac typical. Dielectric Withstand: Meet IEC60601-1. EMI: Meet EN 55011 & FCC CLASS B. Hold-up Time: 16mS @115Vac typical. 64mS @230Vac typical. Earth Leakage: Less than 0.1 mA @264Vac. Output Voltage: See Ratings Chart. Output Current: See Ratings Chart. Output Wattage: 22-24Watts typical. Output Connector & Cord: Optional. Line Regulation: 0.1% typical. Load Regulation: ±1.5-3.0 % typical. Noise & Ripple: 1.0% typical peak to peak. OVP: Built-in, by latched circuit. Adjustability: Factory set. Overload Protection (OLP): Fully protected against output overload and short circuit. OLP set at about 125-150% rating output wattage. Consult the factory for OLP setting. GENERAL SPECIFICATION Efficiency: 80% typical. Switching Frequency: 75K Hz typical. Circuit Topology: Fixed Frequency Flyback circuit. Transient Response: Output voltage returns in less than 3mS following a 50% load change. Safety Standard: Meet Medical standards EN 60601-1/ UL 2601-1 Class I. Power Density: 1.75Watts / Cubic inch. Operating Temperature: 0 to +25/35/40°C without derating. (various with output voltage) Storage Temperature: -20 to +85°C. Temperature Coefficient: 0.04% /°C typical. Cooling: Free air convection. Construction: Impact resistant thermo- plastic enclosure case SE100 94V.0 Nylon. Desktop construction format. NOTE: (1) All measurements are at nominal input, full load, and +25°C unless otherwise specified. (2) Load regulation is measured at 115Vac or 230Vac in percentage to indicate the change in output voltage as the load varied from half load to full load(%). (3) The exact obtainable load regulation depends upon the output cord selected and load current. Upper data are for 6 ft. (2 m) cord AWG#18 wires. (4) Due to requests in market and advances in technology, specifications subject to change without notice. For the details of safety approval, please consult the factory. Hitron Electronics Corporation. B4-11, P.O. Box 26-110 Kaohsiung Export Processing Zone Kaohsiung 806 Taiwan TEL: 886-7-8210410/FAX: 886-7-8419917, 886-7-8210511/E-mail: [email protected] /Web-site: www.hitron-e.com 1 MODEL NO. DESCRIPTION: HES22 - X X X Y Y Y OUTPUT CURRENT OUTPUT VOLTAGE SERIES NAME (HiTRON External Switcher) : About 22W OUTPUT VOLTAGE/CURRENT RATINGS CHART SINGLE OUTPUT MODEL NO. AC PLUG O/P VOLTAGE O/P CURRENT HES22-033400 DT7 3.3Vdc 4.00A HES22-050400 DT7 5.0Vdc 4.00A HES22-075300 DT7 7.5Vdc 3.00A HES22-090250 DT7 9.0Vdc 2.50A HES22-120200 DT7 12.0Vdc 2.00A HES22-150160 DT7 15.0Vdc 1.60A HES22-210115 DT7 21.0Vdc 1.15A HES22-240100 DT7 24.0Vdc 1.00A HES22-300080 DT7 30.0Vdc 0.80A MECHANICAL DIMENSIONS: MM [INCHES] WEIGHT: 206.0g (6.23 Oz.). Hitron Electronics Corporation. B4-11, P.O. Box 26-110 Kaohsiung Export Processing Zone Kaohsiung 806 Taiwan TEL: 886-7-8210410/FAX: 886-7-8419917, 886-7-8210511/E-mail: [email protected] /Web-site: www.hitron-e.com 2