H HiTRON

H HiTRON
2010/6/9
AC/DC BOTH MEDICAL AND ITE APPLICATION EXTERNAL
ADAPTER UNIVERSAL INPUT SINGLE OUTPUT 45-50 WATTS
GREEN POWER SWITCHING POWER SUPPLIES HEMG49 SERIES
FEATURES:
 ACCOMMODATE UNIVERSAL AC SOURCES
 DESKTOP IEC320-2P/3P AC RECEPTACLE
 MEET MEDICAL STANDARDS IEC60601-1
AND ITE STANDARDS IEC60950-1
 EMI MEET EN 55011/ FCC CLASS B
 ENERGY STAR V2.0 LEVEL V/CEC LEVEL IV
 CE MARKING COMPLIANCE
SPECIFICATION
INPUT SPECIFICATION
OUTPUT SPECIFICATION
Input Voltage: Typical 90-264Vac.
Input Connector: IEC320-3P (C14) DT7 or 2P (C8) DT8
AC input receptacle.
Input Frequency: 47-63Hz.
Inrush Current: 12Arms @230Vac.
Input Current: 0.91A@115Vac & 0.57A @230Vac.
Dielectric Withstand: Meet IEC60601-1 & IEC60950-1.
EMI: Meet EN55011 / FCC Class B.
Hold-up Time: Typical 12mS @115Vac.
Typical 70mS @230Vac.
Over Temp. Protection: Optional (NTC ckt.).
Earth Leakage: Less than 0.3 mA.
No Load Power: Less than 0.3W @230Vac
Output Voltage: See Ratings Chart.
Output Current: See Ratings Chart.
Output Wattage: Typical 45-50Watts.
Output Connector & Cord: Optional.
Line Regulation: Typical 0.1%.
Load Regulation: Typical ±1.5-3.0%.
Noise & Ripple: 1.0% peak to peak.
OVP: Built-in by latch circuit.
Adjustability: Factory set.
Over Current Protection (OCP):
Fully protected against output overload and short circuit.
The PSU will shut down after OCP is activated.
Consult the factory for OCP setting.
GENERAL SPECIFICATION
Efficiency: Typical 86%-88% (various with the output
voltage)
Switching Frequency: Typical 65KHz.
Circuit Topology: Fixed Frequency Flyback circuit.
Transient Response: Output voltage returns in less than
5.5mS following a 50% load change.
Safety Standard: Meet Medical IEC60601-1 & ITE
IEC60950-1, Class I for DT7(C14) or Class II for DT8(C8)
Power Density: 3.3Watts. / Cubic inch.
Operating Temperature: 0°C to +40°C.
Storage Temperature: -20 to +85°C.
Temperature Coefficient: 0.04% /°C.
Cooling: Free air convection.
Construction: Impact resistant thermo-plastic
enclosure case.
Desktop Format.
NOTE: (1) All measurements are at nominal input, full load, and +25°C unless otherwise specified.
(2) Load regulation is measured at 115Vac or 230Vac in percentage to indicate the change in output voltage as the load varied from half load to full load (%).
(3) The exact obtainable load regulation depends upon the output cord selected and load current.
(4) Due to requests in market and advances in technology, specifications subject to change without notice.
OUTPUT VOLTAGE / CURRENT RATINGS CHART
2010/6/9
SINGLE OUTPUT
MODEL NO.
AC PLUG
O/P VOLTAGE
O/P CURRENT
HEMG49-S120400-7
DT7/IEC320-C14
12.0Vdc
4.0A
HEMG49-S120400-8
DT8/IEC320-C8
12.0Vdc
4.0A
HEMG49-S150330-7
DT7/IEC320-C14
15.0Vdc
3.3A
HEMG49-S150330-8
DT8/IEC320-C8
15.0Vdc
3.3A
HEMG49-S240210-7
DT7/IEC320-C14
24.0Vdc
2.1A
HEMG49-S240210-8
DT8/IEC320-C8
24.0Vdc
2.1A
MECHANICAL DIMENSIONS: MM [INCHES]
WEIGHT: 373.0g (13.2 Oz.)