SCXI-1300 Terminal Block Calibration Procedure

CALIBRATION PROCEDURE
SCXI™-1300 TERMINAL BLOCK
Introduction
This document contains information and step-by-step instructions for
verifying the temperature sensor performance of the National Instruments
SCXI-1300 terminal block.
What Is Calibration?
For SCXI-1300 terminal blocks, calibration is simply determining the
measurement accuracy of the components on the terminal block. Because
these components are not user-adjustable, calibration consists of a
verification procedure only.
Why Calibrate?
Electronic components drift with time, which can affect measurement
accuracy as the device ages. Calibration ensures that your SCXI-1300
terminal block is still meeting National Instruments standards. If the results
of the procedure indicate that your terminal block is out of specification,
return it to National Instruments for repair.
How Often Should You Calibrate?
The measurement accuracy requirements of your application determine
the calibration interval of your SCXI-1300 terminal block.
National Instruments recommends you perform a complete calibration at
least once every year. You can shorten this interval to six months or
90 days, based on the demands of your application.
National Instruments™, ni.com™, and SCXI ™ are trademarks of National Instruments Corporation. Product and company names mentioned herein are
trademarks or trade names of their respective companies.
370252A-01
© Copyright 2000 National Instruments Corp. All rights reserved.
July 2000
Equipment and Other Test Requirements
This section describes the equipment, software, documentation, and test
conditions required for verification.
Test Equipment
Verification requires a high-precision voltage source with at least 50 ppm
accuracy, and a multiranging 5 1/2 digit digital multimeter (DMM) with
15 ppm accuracy.
National Instruments recommends you use the following instruments for
verification of an SCXI-1300:
•
Calibrator—Fluke 5700A
•
DMM—NI 4060 or HP 34401A
If these instruments are not available, use the accuracy requirements listed
above to select a substitute calibration standard.
Software and Documentation
No software is required to verify the operation of the SCXI-1300. All
required documentation is found in this calibration procedure. However,
if you would like more information on the product, refer to the SCXI-1300
Terminal Block Installation Guide.
Test Conditions
Follow these guidelines to optimize the connections and the environment
during verification:
•
Keep connections to the SCXI-1300 terminal block short. Long cables
and wires act as antennae, picking up extra noise that can affect
measurements.
•
Use shielded copper wire for all cable connections to the device.
Use twisted-pair wire to eliminate noise and thermal offsets.
•
Keep relative humidity below 80%.
•
Maintain a temperature between 15 °C and 35 °C.
SCXI-1300 Terminal Block Calibration Procedure
2
ni.com
Verification Procedure
The following procedure verifies the performance of the temperature
sensor on your SCXI-1300:
1.
Connect a +5 VDC power source to the terminal block.
a.
Hold the terminal block vertically upright and view it from the
rear. The terminals on the 96-pin DIN connector are designated
as follows for the purposes of this procedure:
–
Column A is on the right, Column B is in the middle, and
Column C is on the left.
–
Row 1 is at the bottom and Row 32 is at the top.
–
Individual pins are identified by their column and row.
For example, “A3” denotes the terminal located in Column A
and Row 3.
This conforms to the labeling of the pins on the front connector of
a matching SCXI terminal block. It does not necessarily
correspond to the labeling of the pins on the rear of the terminal
block connector itself, which you can only view by opening the
terminal block enclosure.
2.
3.
© National Instruments Corporation
b.
Strip 0.5 in. of insulation from one end of a 22 AWG solid wire.
Insert the stripped end of the wire into terminal A1 on the 96-pin
female DIN connector on the rear of the terminal block. Attach the
other end of this wire to the positive terminal of the +5 VDC
power supply.
c.
Strip 0.5 in. of insulation from one end of a 22 AWG solid wire.
Insert the stripped end of the wire into terminal A2 on the 96-pin
female DIN connector on the rear of the terminal block. Attach the
other end of this wire to the negative terminal of the +5 VDC
power supply.
Connect a calibrated DMM to the temperature-sensor output of the
terminal block.
a.
Strip 0.5 in. of insulation from one end of a 22 AWG solid wire.
Insert the stripped end of the wire into terminal A3 on the 96-pin
female DIN connector on the rear of the terminal block. Attach the
other end of this wire to the positive input terminal of the
calibrated DMM.
b.
Connect the negative input terminal of the calibrated DMM to the
negative terminal of the +5 VDC power supply.
Place the terminal block in a temperature-controlled environment
where the temperature is between 15 °C and 35 °C.
3
SCXI-1300 Terminal Block Calibration Procedure
4.
When the terminal block temperature equilibrates with its
surroundings, measure the temperature sensor output Vmeas using a
calibrated DMM.
5.
Measure the actual temperature Tact in the temperature-controlled
environment using a calibrated instrument.
6.
Convert Vmeas (in volts) to measured temperature Tmeas (in degrees
Celsius) by multiplying Vmeas by 100.
7.
Compare Tact to Tmeas.
–
If (Tmeas − 0.9 °C) ≤ Tact ≤ (Tmeas + 0.9 °C), then the operation of
the terminal block temperature sensor has been verified.
–
If Tact < (Tmeas − 0.9 °C), the temperature sensor is nonfunctional.
Return the terminal block to National Instruments for repair or
replacement.
–
If Tact > (Tmeas + 0.9 °C), the temperature sensor is nonfunctional.
Return the terminal block to National Instruments for repair or
replacement.
You have completed the calibration verification procedure for your
SCXI-1300.
SCXI-1300 Terminal Block Calibration Procedure
4
ni.com