SCXI-1120/D Calibration Procedure

CALIBRATION PROCEDURE
SCXI™-1120/D
Introduction
This document contains information and step-by-step instructions for
calibrating the National Instruments (NI) SCXI-1120 and SCXI-1120D
modules.
What Is Calibration?
Calibration consists of verifying the measurement accuracy of a module
and adjusting for any measurement error. Verification is measuring the
performance of the module and comparing these measurements to the
factory specifications. During calibration, you supply and read voltage
levels using external standards, then you adjust the module calibration
circuitry. This circuitry compensates for any inaccuracies in the module,
and returns the accuracy of the module to the factory specifications.
Why Should You Calibrate?
The accuracy of electronic components drifts with time and temperature,
which can affect measurement accuracy as a module ages. Calibration
restores these components to their specified accuracy and ensures that the
module still meets NI standards.
How Often Should You Calibrate?
The measurement requirements of your application determine how often
the SCXI-1120/D module needs to be calibrated to maintain accuracy.
NI recommends that you perform a complete calibration at least once every
year. You can shorten this interval to 90 days or six months based on the
demands of your application.
National Instruments™, NI™, ni.com™, and SCXI ™ are trademarks of National Instruments Corporation. Product and company names mentioned herein
are trademarks or trade names of their respective companies.
370268B-01
© 2000, 2001 National Instruments Corp. All rights reserved.
September 2001
Equipment and Other Test Requirements
This section describes the test equipment, software, documentation, and
test conditions required for calibrating the SCXI-1120/D modules.
Test Equipment
Calibration requires a high-precision voltage source with at least 50 ppm
accuracy and a multiranging 5 1/2 digit digital multimeter (DMM) with
15 ppm accuracy.
Instruments
NI recommends the following instruments for calibrating SCXI-1120/D
modules:
•
Calibrator—Fluke 5700A
•
DMM—NI 4060 or HP 34401A
If these instruments are not available, use the accuracy requirements listed
previously to select substitute calibration instruments.
Connectors
If you do not have custom connection hardware, you need the following
connectors:
•
Terminal block, such as the SCXI-1320
•
Shielded 68-pin connector cable
•
50-pin ribbon cable
•
50-pin breakout box
•
SCXI-1349 adapter
These components give easy access to the individual pins on the
SCXI-1120/D module front and rear connectors.
Software and Documentation
No special software or documentation is necessary to calibrate the
SCXI-1120/D module. This calibration document contains all the
information you need to complete the verification and adjustment
procedures. If you would like more information on the module, refer
to the SCXI-1120/D User Manual.
SCXI-1120/D Calibration Procedure
2
ni.com
Test Conditions
Follow these guidelines to optimize the connections and environment
during calibration:
•
Keep connections to the SCXI-1120/D module short. Long cables and
wires act as antennae, picking up extra noise and thermal offsets that
can affect measurements.
•
Use shielded copper wire for all cable connections to the device.
Use twisted-pair wire to eliminate noise and thermal offsets.
•
Maintain the temperature between 18–28 °C.
•
Keep relative humidity below 80%.
•
Allow a warm-up time of at least 15 minutes for the SCXI-1120/D
module to ensure the measurement circuitry is at a stable operating
temperature.
Calibration
The calibration procedure for the SCXI-1120/D module consists of the
following steps:
1.
Set up the module for testing.
2.
Verify the existing operation of the module to determine whether it is
operating within its specifications.
3.
Adjust the module with respect to a known voltage source.
4.
Verify that the module is operating within its specifications after
adjustments.
© National Instruments Corporation
3
SCXI-1120/D Calibration Procedure
Setting Up the Module
Refer to Figures 1 and 2 while performing the following steps to set up the
SCXI-1120/D module for verification:
1.
Remove the grounding screw from the module.
2.
Remove the cover on the module to access the potentiometers.
2
1
Grounding screw
2
1
Module cover
Figure 1. Grounding Screw and Module Cover Removal
3.
Remove the side plate of the SCXI chassis.
4.
Install the SCXI-1120/D into slot 4 of the SCXI chassis.
SCXI-1120/D Calibration Procedure
4
ni.com
®
SC
MA
IN
XI
FR
AM
E
L
NA TS
TIO EN
NA RUM
INST
3
2
1
1
Thumbscrews
2
Slot 4
3
Side plate
Figure 2. Side Plate Removal and Module Installation
The SCXI-1120/D module does not need to be connected to a data
acquisition (DAQ) device. Leave the configuration of the digital jumpers
W41–W43 and W46 unchanged because they do not affect this procedure.
Configuring the Gain Jumpers
Each input channel has two user-configurable gain stages. The first-stage
gain provides gains of 1, 10, 50, and 100. The second-stage gain provides
gains of 1, 2, 5, 10, and 20. Table 1 shows the jumper reference designators
for gain selection associated with each channel. Table 2 shows how to
position each jumper to select the desired gain for each channel.
Table 1. Gain Jumper Reference Designators
© National Instruments Corporation
Input Channel
Number
First-Stage
Gain Jumper
Second-Stage
Gain Jumper
0
W1
W9
1
W2
W10
2
W3
W11
5
SCXI-1120/D Calibration Procedure
Table 1. Gain Jumper Reference Designators (Continued)
Input Channel
Number
First-Stage
Gain Jumper
Second-Stage
Gain Jumper
3
W4
W12
4
W5
W13
5
W6
W14
6
W7
W15
7
W8
W16
Table 2. Gain Jumper Positions
Gain
First Stage
Second Stage
Setting
Jumper Position
1
D
10
C
50
B
100
A (factory setting)
1
A
2
B
5
C
10
D (factory setting)
20
E
To change the gain setting of a specified channel on the module, move
the appropriate jumper on the module to the position indicated in Table 2.
Refer to Table 1 for jumper reference designators, and Figure 3 for the
location of the jumpers.
SCXI-1120/D Calibration Procedure
6
ni.com
4
3
5
1
V25 V26 W27 W28 W29W30 W31V32V33 W34 W35 W36 V37 V38 V39 V40
6
2
W42
W43
7
W45
1
12
U14
R18
W6
8
GAINDIP
22 W14
AD705
Q12
AD
ABCDABCDEAB
9
10
11
13
12
Front
1
2
3
4
5
6
Thumbscrews
Front Connector
Product Name, Assembly Number,
and Serial Number
Output Null Adjust Potentiometers
Second-Stage Filter Jumpers
Rear Signal Connector
7
8
9
10
11
12
13
Rear
SCXIbus Connector
Input Null Adjust Potentiometers
First-Stage Gain Jumpers
Second-Stage Gain Jumpers
First-Stage Filter Jumpers
Terminal Block Mounting Hole
Grounding Screw
Figure 3. SCXI-1120/D Parts Locator Diagram
Note
The SCXI-1120D module has an additional fixed pre-stage gain of 0.5.
© National Instruments Corporation
7
SCXI-1120/D Calibration Procedure
The order of the settings for the first- and second-stage gain does not
matter as long as the first-stage gain multiplied by the second-stage
gain—multiplied by 0.5 when using the SCXI-1120D—equals the desired
final gain value.
•
SCXI-1120—To determine the overall gain of a given channel on the
SCXI-1120 module:
First-Stage Gain × Second-Stage Gain = Overall Gain
•
SCXI-1120D—To determine the overall gain of a given channel on the
SCXI-1120D module:
( First-Stage Gain × Second-Stage Gain ) × 0.5 = Overall Gain
Configuring the Filter Jumpers
Each input channel also has two user-configurable filter stages.
The SCXI-1120 module ships in the 4 Hz position and the SCXI-1120/D
module ships in the 4.5 kHz position. Refer to Table 3 or 4 to find the
correct jumper setting for the desired cutoff frequency. Figure 3 shows the
locations of the jumper blocks on the SCXI-1120/D modules. Verify that
both filter stages are set to the same filter setting to ensure that you achieve
the required bandwidth.
Table 3. SCXI-1120 Filter Jumper Settings
First Filter Jumper
SCXI-1120/D Calibration Procedure
Input
Channel
Number
4 Hz
(Factory
Setting)
0
Second Filter Jumper
10 kHz
4 Hz
(Factory
Setting)
10 kHz
W17-A
W17-B
W25
W26
1
W18-A
W18-B
W27
W28
2
W19-A
W19-B
W29
W30
3
W20-A
W20-B
W31
W32
4
W21-A
W21-B
W33
W34
5
W22-A
W22-B
W35
W36
6
W23-A
W23-B
W37
W38
7
W24-A
W24-B
W39
W40
8
ni.com
Table 4. SCXI-1120D Filter Jumper Allocation
First Filter Jumper
Input
Channel
Number
4.5 kHz
(Factory
Setting)
0
Second Filter Jumper
22.5 kHz
4.5 kHz
(Factory
Setting)
22.5 kHz
W17-A
W17-B
W26
W25
1
W18-A
W18-B
W28
W27
2
W19-A
W19-B
W30
W29
3
W20-A
W20-B
W32
W31
4
W21-A
W21-B
W34
W33
5
W22-A
W22-B
W36
W35
6
W23-A
W23-B
W38
W37
7
W24-A
W24-B
W40
W39
Verifying the Operation of the Module
The verification procedure determines how well the SCXI-1120/D module
is meeting its specifications. You can use this information to select the
appropriate calibration interval for your application. Refer to the Setting Up
the Module section for information on how to configure the channel filter
and channel gain.
Complete the following steps to verify the operation of the SCXI-1120/D
module:
1.
Read the Test Conditions section in this document.
2.
Refer to Table 7 for the SCXI-1120 module or Table 8 for the
SCXI-1120D module for all acceptable settings for the module.
Although NI recommends verifying all ranges and gains, you can
save time by checking only those ranges used in your application.
3.
Set the channel filter for all channels on the module to 4 Hz for
the SCXI-1120 module or 4.5 kHz for the SCXI-1120D module.
4.
Set the channel gain on all channels to the gain you want to test,
starting with the smallest gain available for the module. Available
gains are shown in Tables 7 and 8.
5.
Connect the calibrator to the analog input channel you are testing,
starting with channel 0.
If you do not have an SCXI terminal block such as the SCXI-1320,
refer to Table 5 to determine the pins on the 96-pin front connector that
correspond to the positive and negative inputs of the specified channel.
© National Instruments Corporation
9
SCXI-1120/D Calibration Procedure
For example, the positive input for channel 0 is pin A32, which is
labeled CH0+. The negative input for channel 0 is pin C32, which is
labeled CH0–.
Table 5. SCXI-1120/D Front Connector Pin Assignments
SCXI-1120/D Calibration Procedure
Pin Number
Column A
Column B
Column C
32
CH0+
NP
CH0–
31
NP
NP
NP
30
CH1+
NP
CH1–
29
NP
NP
NP
28
NC
NP
NC
27
NP
NP
NP
26
CH2+
NP
CH2–
25
NP
NP
NP
24
CH3+
NP
CH3–
23
NP
NP
NP
22
NC
NP
NC
21
NP
NP
NP
20
CH4+
NP
CH4–
19
NP
NP
NP
18
CH5+
NP
CH5–
17
NP
NP
NP
16
NC
NP
NC
15
NP
NP
NP
14
CH6+
NP
CH6–
13
NP
NP
NP
12
CH7+
NP
CH7–
11
NP
NP
NP
10
NC
NP
NC
9
NP
NP
NP
8
NC
NP
RSVD
10
ni.com
Table 5. SCXI-1120/D Front Connector Pin Assignments (Continued)
Pin Number
Column A
Column B
Column C
7
NP
NP
NP
6
RSVD
NP
RSVD
5
NP
NP
NP
4
+5V
NP
MTEMP
3
NP
NP
NP
2
CHSGND
NP
DTEMP
1
NP
NP
NP
NP—No pin; NC—No connect
6.
© National Instruments Corporation
Connect the DMM to the output of the same channel to which the
calibrator was connected in step 5. Refer to Figure 4 to determine the
pins on the 50-pin rear connector that correspond to the positive and
negative outputs for the specified channel. For example, the positive
output for channel 0 is pin 3, which is labeled MCH0+. The negative
output for channel 0 is pin 4, which is labeled MCH0–.
11
SCXI-1120/D Calibration Procedure
AOGND
MCH0+
MCH1+
MCH2+
MCH3+
MCH4+
MCH5+
MCH6+
MCH7+
OUTREF
NC
NC
SERDATIN
DAQD*/A
SLOT0SEL*
NC
DIG GND
NC
SERCLK
NC
NC
RSVD
NC
NC
NC
1
3
5
7
9
11
13
15
17
19
21
23
25
27
29
31
33
35
37
39
41
43
45
47
49
2
4
6
8
10
12
14
16
18
20
22
24
26
28
30
32
34
36
38
40
42
44
46
48
50
AOGND
MCH0–
MCH1–
MCH2–
MCH3–
MCH4–
MCH5–
MCH6–
MCH7–
NC
NC
DIG GND
SERDATOUT
NC
NC
NC
NC
SCANCLK
NC
NC
NC
NC
NC
NC
NC
Figure 4. SCXI-1120/D Rear Connector Pin Assignments
7.
Set the calibrator voltage to the value specified by the Test Point entry
listed in Table 7 for the SCXI-1120 module or Table 8 for the
SCXI-1120D module.
8.
Read the resulting output voltage on the DMM. If the output voltage
result falls between the Upper Limit and the Lower Limit values, the
module passed the test.
9.
Repeat steps 5 through 8 for the remaining test points.
10. Repeat steps 5 through 9 for the remaining analog input channels.
11. Repeat steps 4 through 10 for the remaining gain settings specified in
the appropriate table.
12. Repeat steps 3 through 11, but set the channel filter to 10 kHz for the
SCXI-1120 module or 22.5 kHz for the SCXI-1120D module.
You have completed verifying the operation of the module.
SCXI-1120/D Calibration Procedure
12
ni.com
Adjusting Offset Null Values of the Module
Complete the following steps to adjust the offset null value:
1.
Set the channel gain on all channels to a gain of 1. Set the filter value
to 4 Hz for the SCXI-1120 module or 4.5 kHz for the SCXI-1120D
module. Refer to the Setting Up the Module section in this document
for information on how to set the channel gain.
2.
Connect the calibrator to the analog input channel you want to adjust,
starting with channel 0. Refer to Table 5 to determine the pins on the
96-pin front connector that correspond to the positive and negative
inputs of the specified channel. For example, the positive input for
channel 0 is pin A32, which is labeled CH0+. The negative input for
channel 0 is pin C32, which is labeled CH0–.
3.
Connect the DMM to the output of the same channel to which the
calibrator was connected in step 2. Refer to Figure 4 to determine the
pins on the 50-pin rear connector that correspond to the positive and
negative outputs for the specified channel. For example, the positive
output for channel 0 is pin 3, which is labeled MCH0+. The negative
output for channel 0 is pin 4, which is labeled MCH0–.
4.
Set the calibrator to output 0.0 V.
5.
Adjust the output potentiometer of the channel until the DMM reading
is 0 ±3.0 mV. Refer to Figure 3 for the potentiometer location and
Table 6 for the potentiometer reference designator.
Table 6. Calibration Potentiometers Reference Designators
6.
© National Instruments Corporation
Input Channel Number
Input Null
Output Null
0
R08
R24
1
R10
R25
2
R12
R26
3
R14
R27
4
R16
R28
5
R18
R29
6
R20
R30
7
R21
R31
Set the channel gain on all channels to 1000.0.
13
SCXI-1120/D Calibration Procedure
7.
Adjust the input potentiometer of channel 0 until the DMM reading is
0 ±6.0 mV. Refer to Figure 3 for the potentiometer location and Table 6
for the potentiometer reference designator.
8.
Repeat steps 1 through 7 for the remaining analog inputs.
You have completed adjusting the module.
Verifying Adjusted Values
After you complete the adjustment procedure, it is important to verify the
accuracy of the adjusted values by repeating the procedure in the Verifying
the Operation of the Module section. Verifying the adjusted values ensures
the module is operating within its specifications after adjustments.
If the SCXI-1120/D module fails after calibration, return it to NI for repair or
replacement.
Note
SCXI-1120/D Calibration Procedure
14
ni.com
Specifications
Table 7 contains test specifications for SCXI-1120 modules. Table 8
contains test specifications for SCXI-1120D modules. If the module
was calibrated within the last year, the output from the module should
fall between the Upper Limit and Lower Limit values.
Table 7. SCXI-1120 Specifications
4Hz filter setting
10kHz filter setting
Gain
Test Point
(V)
Upper Limit
(V)
Lower Limit
(V)
Upper Limit
(V)
Lower Limit
(V)
0.011
232.5
2.346996
2.303004
2.349248
2.300752
0.011
0
0.006888
–0.006888
0.009140
–0.009140
0.011
–232.5
–2.346996
–2.303004
–2.349248
–2.300752
0.021
186
3.751095
3.688905
3.753353
3.686647
0.021
0
0.006922
–0.006922
0.009180
–0.009180
0.021
–186
–3.751095
–3.688905
–3.753353
–3.686647
0.051
93
4.687000
4.613000
4.689236
4.610764
0.051
0
0.006784
–0.006784
0.009020
–0.009020
0.051
–93
–4.687000
–4.613000
–4.689236
–4.610764
0.11
46.5
4.686925
4.613075
4.689186
4.610814
0.11
0
0.006709
–0.006709
0.008970
–0.008970
0.11
–46.5
–4.686925
–4.613075
–4.689186
–.610814
0.21
23.25
4.686775
4.613225
4.689056
4.610944
0.21
0
0.006559
–0.006559
0.008840
–0.008840
0.21
–23.25
–4.686775
–4.613225
–4.689056
–4.610944
0.51
9.3
4.686353
4.613647
4.688626
4.611374
0.51
0
0.006138
–0.006138
0.008410
–0.008410
0.51
–9.3
–4.686353
–4.613647
–4.688626
–4.611374
1
4.65
4.691704
4.608296
4.693926
4.606074
1
0
0.011488
–0.011488
0.013710
–0.013710
1
–4.65
–4.691704
–4.608296
–4.693926
–4.606074
© National Instruments Corporation
15
SCXI-1120/D Calibration Procedure
Table 7. SCXI-1120 Specifications (Continued)
4Hz filter setting
10kHz filter setting
Gain
Test Point
(V)
Upper Limit
(V)
Lower Limit
(V)
Upper Limit
(V)
Lower Limit
(V)
2
2.325
4.690653
4.609347
4.692876
4.607124
2
0
0.010437
–0.010437
0.012660
–0.012660
2
–2.325
–4.690653
–4.609347
–4.692876
–4.607124
5
0.93
4.690498
4.609502
4.692726
4.607274
5
0
0.010282
–0.010282
0.012510
–0.012510
5
–0.93
–4.690498
–4.609502
–4.692726
–4.607274
10
0.465
4.690401
4.609599
4.692626
4.607374
10
0
0.010185
–0.010185
0.012410
–0.012410
10
–0.465
–4.690401
–4.609599
–4.692626
–4.607374
20
0.2325
4.690139
4.609861
4.692416
4.607584
20
0
0.009924
–0.009924
0.012200
–0.012200
20
–0.2325
–4.690139
–4.609861
–4.692416
–4.607584
50
0.093
4.690046
4.609954
4.692331
4.607669
50
0
0.009831
–0.009831
0.012115
–0.012115
50
–0.093
–4.690046
–4.609954
–4.692331
–4.607669
100
0.0465
4.689758
4.610242
4.692066
4.607934
100
0
0.009542
-0.009542
0.011850
–0.011850
100
–0.0465
–4.689758
–4.610242
–4.692066
–4.607934
200
0.02325
4.689464
4.610536
4.691936
4.608064
200
0
0.009248
–0.009248
0.011720
–0.011720
200
–0.02325
–4.689464
–4.610536
–4.691936
–4.608064
250
0.0186
4.689313
4.610687
4.692016
4.607984
250
0
0.009097
–0.009097
0.011800
–0.011800
250
–0.0186
–4.689313
–4.610687
–4.692016
–4.607984
500
0.0093
4.689443
4.610557
4.692731
4.607269
500
0
0.009227
–0.009227
0.012515
–0.012515
SCXI-1120/D Calibration Procedure
16
ni.com
Table 7. SCXI-1120 Specifications (Continued)
4Hz filter setting
10kHz filter setting
Gain
Test Point
(V)
Upper Limit
(V)
Lower Limit
(V)
Upper Limit
(V)
Lower Limit
(V)
500
–0.0093
–4.689443
–4.610557
–4.692731
–4.607269
1000
0.00465
4.693476
4.606524
4.698796
4.601204
1000
0
0.013260
–0.013260
0.018580
–0.018580
1000
–0.00465
–4.693476
–4.606524
–4.698796
–4.601204
2000
0.002325
4.703044
4.596956
4.712556
4.587444
2000
0
0.022828
–0.022828
0.032340
–0.032340
2000
–0.002325
–4.703044
–4.596956
–4.712556
–4.587444
1Value
available only when used with the SCXI-1327 high-voltage terminal block
Table 8. SCXI-1120D Specifications
4.5KHz filter setting
22.5KHz filter setting
Gain
Test Point
(V)
Upper Limit
(V)
Lower Limit
(V)
Upper Limit
(V)
Lower Limit
(V)
0.011
232.5
2.351764
2.298236
2.365234
2.284766
0.011
0
0.006230
–0.006230
0.019700
–0.019700
0.011
–232.5
–2.351764
–2.298236
–2.365234
–2.284766
0.0251
186
4.698751
4.601249
4.733819
4.566181
0.0251
0
0.007683
–0.007683
0.042750
–0.042750
0.0251
–186
–4.698751
–4.601249
–4.733819
–4.566181
0.051
93
4.697789
4.602211
4.768769
4.531231
0.051
0
0.006720
–0.006720
0.077700
–0.077700
0.051
–93
–4.697789
–4.602211
–4.768769
–4.531231
0.11
46.5
4.698899
4.601101
4.841289
4.458711
0.11
0
0.007830
–0.007830
0.150220
–0.150220
0.11
–46.5
–4.698899
–4.601101
–4.841289
–4.458711
0.251
18.6
4.701669
4.598331
5.028819
4.271181
© National Instruments Corporation
17
SCXI-1120/D Calibration Procedure
Table 8. SCXI-1120D Specifications (Continued)
4.5KHz filter setting
22.5KHz filter setting
Gain
Test Point
(V)
Upper Limit
(V)
Lower Limit
(V)
Upper Limit
(V)
Lower Limit
(V)
0.251
0
0.010600
–0.010600
0.337750
–0.337750
0.251
–18.6
–4.701669
–4.598331
–5.028819
–4.271181
0.5
9.3
4.697331
4.602669
4.703726
4.596274
0.5
0
0.006355
–0.006355
0.012750
–0.012750
0.5
–9.3
–4.697331
–4.602669
–4.703726
–4.596274
1
4.65
4.697416
4.602584
4.710876
4.589124
1
0
0.006440
-0.006440
0.019900
–0.019900
1
–4.65
–4.697416
–4.602584
–4.710876
–4.589124
2.5
1.86
4.697883
4.602117
4.732426
4.567574
2.5
0
0.006908
–0.006908
0.041450
–0.041450
2.5
–1.86
–4.697883
–4.602117
–4.732426
–4.567574
5
0.93
4.698726
4.601274
4.768726
4.531274
5
0
0.007750
–0.007750
0.077750
–0.077750
5
–0.93
–4.698726
–4.601274
–4.768726
–4.531274
10
0.465
4.700796
4.599204
4.841236
4.458764
10
0
0.009820
–0.009820
0.150260
–0.150260
10
–0.465
–4.700796
–4.599204
–4.841236
–4.458764
25
0.18
5.070004
3.929996
4.870004
4.129996
25
0
0.530350
–0.530350
0.330350
–0.330350
25
–0.18
–5.070004
–3.929996
–4.870004
–4.129996
50
0.086
4.360392
4.239608
4.825892
3.774108
50
0
0.022500
–0.022500
0.488000
–0.488000
50
–0.086
–4.360392
–4.239608
–4.825892
–3.774108
100
0.038
3.879624
3.720376
4.810624
2.789376
100
0
0.039800
–0.039800
0.970800
–0.970800
100
–0.038
–3.879624
–3.720376
–4.810624
–2.789376
SCXI-1120/D Calibration Procedure
18
ni.com
Table 8. SCXI-1120D Specifications (Continued)
4.5KHz filter setting
22.5KHz filter setting
Gain
Test Point
(V)
Upper Limit
(V)
Lower Limit
(V)
Upper Limit
(V)
Lower Limit
(V)
250
0.0125
3.277438
2.972563
4.830188
1.419813
250
0
0.091500
–0.091500
0.056751
–1.644250
250
–0.0125
–3.277438
–2.972563
–4.830188
–1.419813
500
0.006
3.273770
2.726230
4.810770
1.189230
500
0
0.176000
–0.176000
1.713000
–1.713000
500
–0.006
–3.273770
–2.726230
–4.810770
–1.189230
1000
0.0029
3.416058
2.383942
4.895058
0.904942
1000
0
0.342000
–0.342000
1.821000
–1.821000
1000
–0.0029
–3.416058
–2.383942
–4.895058
–0.904942
1Value
available only when used with the SCXI-1327 high-voltage terminal block
© National Instruments Corporation
19
SCXI-1120/D Calibration Procedure