CALIBRATION PROCEDURE SCXI -1520 ™ This document contains information and step-by-step instructions for calibrating the National Instruments SCXI-1520 module. Equipment and Other Test Requirements This section describes the equipment, software, documentation, and test conditions required for calibrating the SCXI-1520. Test Equipment Calibration requires a high-precision voltage source with at least 50 ppm accuracy, a multiranging 5 1/2 digit digital multimeter (DMM) with 15 ppm accuracy, and a 16-bit data acquisition (DAQ) device. NI recommends the following instruments for calibrating the SCXI-1520. • Calibrator—Fluke 5700A • DMM—NI 4060 or Agilent 34401A • 16-bit DAQ device If these instruments are not available, use the accuracy requirements listed above to select a substitute calibration standard. If you do not have custom connection hardware, you need an SCXI-1314 terminal block and a shielded 68-pin connector cable. These components give easy access to the individual pins on the SCXI-1520 front connectors. You also need an SCXI-1349 cable adapter to connect the SCXI-1520 to the DAQ device. Note Software and Documentation Use the following software and documentation to calibrate the SCXI-1520. You can find them on the NI Web site at ni.com/support. • The latest version of the NI-DAQ driver software • NI-DAQ Function Reference Help • NI-DAQ User Manual for PC Compatibles • SCXI-1314 Installation Guide • SCXI-1349 Installation Guide Software The SCXI-1520 calibration procedure requires the latest version of the NI-DAQ driver on the calibration system. The driver supports a number of programming languages, including LabVIEW, Measurement Studio, Microsoft Visual C++, Microsoft Visual Basic, and Borland C++. When you install the driver, you need to install support only for the programming language you are using. Documentation The NI-DAQ Function Reference Help and the NI-DAQ User Manual for PC Compatibles contain detailed information on using the NI-DAQ driver. The online help includes detailed information on the driver functions. You can access the help by selecting Start»Programs»National Instruments DAQ»NI-DAQ Help. The user manual provides instructions on installing and configuring DAQ devices. The user manual also includes detailed information on creating applications that use the NI-DAQ driver. These are your primary references for writing your calibration utility. For more information on the SCXI-1520, refer to the SCXI-1520 User Manual. Test Conditions Follow these guidelines to optimize the connections and the environment during calibration. SCXI-1520 Calibration Procedure • Keep connections to the SCXI module short. Long cables and wires act as antennae, picking up extra noise and thermal offsets that can affect measurements. • Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to reduce noise. • Maintain the temperature between 18 and 28 °C. • Keep relative humidity below 80%. • Allow a warm-up time of at least 15 minutes for the SCXI module and 30 minutes for the DAQ device to ensure that the measurement circuitry is at a stable operating temperature. 2 ni.com Calibration The calibration process consists of the following steps: 1. Setting up the device for testing. 2. Verifying the existing operation of the device to determine whether it is operating within its specifications. 3. Adjusting the device with respect to a known voltage source. 4. Verifying that the device is operating within its specifications after adjustments. Setting Up the SCXI-1520 for Testing Complete the following steps to set up the SCXI-1520 for testing. 1. Install the SCXI-1520 in slot 1 of the SCXI chassis. 2. Using an SCXI-1349 cable adapter, connect a 68-to-68-pin cable between the SCXI-1520 module and the DAQ device installed in the host computer, as shown in Figure 1. 8 1 7 2 ® 5 4 3 2 1 ADDRESS 3 6 SC XI 11 00 4 5 1 2 3 4 DAQ Device SH68-68 Cable SCXI-1349 Cable Adapter Thumbscrews 5 6 7 8 SCXI-1520 Module SCXI-1314 Terminal Block SCXI Chassis Power Switch SCXI-1000 Chassis Figure 1. Connecting an SCXI Chassis to a DAQ Device © National Instruments Corporation 3 SCXI-1520 Calibration Procedure 3. Configure the hardware with Measurement & Automation Explorer (MAX). For more information on configuring the system, refer to the documentation for the DAQ device and the SCXI-1520. Verifying Gain and Offset Errors The verification procedure determines how well the SCXI-1520 module is meeting its specifications. You can use this information to select the appropriate calibration interval for your application. Complete the following steps to verify gain and offset errors of the SCXI-1520. 1. Read the Test Conditions section. 2. Refer to the specifications in Table 2. 3. Call Calibrate_E_Series to minimize the uncertainty associated with the DAQ device measurements. Set the following parameters: 4. 5. SCXI-1520 Calibration Procedure • deviceNumber—The value assigned by MAX • calOP—ND_SELF_CALIBRATE • setOfCalConst—ND_USER_EEPROM_AREA • calRefVolts—0.0 Call AI_Configure to put the DAQ device in bipolar mode. Set the following parameters: • deviceNumber—The value assigned by MAX • chan—0 • inputMode—0 • inputRange—0 • polarity—0 • driveAIS—0 Call SCXI_Track_Hold_Setup to set up the track-and-hold circuitry of the SCXI-1520. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • inputMode—2 • source—0 • send—0 • holdCount—1 • DAQdeviceNumber—The value assigned by MAX 4 ni.com 6. Call SCXI_MuxCtr_Setup to configure the DAQ device counter for use as a multiplexer counter during SCXI channel scanning. Set the following parameters: • deviceNumber—The value assigned by MAX • enable—1 • scanDiv—1 • ctrValue—1 7. Connect the calibrator to the appropriate analog input, starting with channel 0. Refer to Table 1 to determine which pins on the 96-pin front connector correspond to the positive and negative inputs of the specified channel. For example, the positive input for channel 0 (S0+) is pin C32. The negative input for channel 0 (S0–) is pin B32. If you are using an SCXI-1314 terminal block, connect the calibrator to the S0+ and S0– inputs. 8. Call SCXI_SCAN_Setup to set up the SCXI chassis to perform a multiplexed scanning data acquisition. Set the following parameters: 9. • SCXIchassisID—The value assigned by MAX • numModules—1 • moduleList—{1}1 • numChans—{1} • startChans—{The channel number from step 7} • DAQdeviceNumber—The value assigned by MAX • modeFlag—0 Call SCXI_Configure_Filter to configure the filter on the module. Set the following parameters: • chassisID—The value assigned by MAX • moduleSlot—1 • channel—The channel number from step 7 • filterMode—0 • freq—0 • cutoffDivDown—0 • outClkDivDown—0 10. Call SCXI_Set_Potentiometer to set the fine potentiometer to the center position. Set the following parameters: 1 • SCXIchassisID—The value assigned by MAX • moduleSlot—1 Array notation: {X} denotes a one-element array initialized with the value X; [Y] denotes an uninitialized array of size Y. © National Instruments Corporation 5 SCXI-1520 Calibration Procedure • channel—The channel number from step 7 • value—2048 • pottype—1 11. Call SCXI_Set_Potentiometer to set the coarse potentiometer to the center position. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • channel—The channel number from step 7 • value—62 • pottype—0 12. Call SCXI_Set_Excitation to set the excitation voltage to zero. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • moduleChan—The channel number from step 7 • excitationType—2 • excitationValue—0 SCXI_Set_Excitation returns a 32-bit floating-point number, actualExcitationValuePtr. 13. Call SCXI_Configure_Connection to configure the channel for a full-bridge connection. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • moduleChan—The channel number from step 7 • connectionType—6 14. Call SCXI_Set_Gain to set the gain you want to test on the module. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • channel—The channel number from step 7 • gain—The gain value you want to test; you can find valid gain values in Table 2. 15. Set the calibrator to output the test point voltage from Table 2. 16. Call SCAN_Op to acquire the data. Set the following parameters: SCXI-1520 Calibration Procedure • deviceNumber—The value assigned by MAX • numChans—{1} 6 ni.com • chans—{0} • gains—1 • buffer—[5000] • count—5000 • sampleRate—10000 • scanRate—5000 17. Call SCXI_Scale to scale the data returned by SCAN_Op. Set the following parameters: • SCXIChassisID—The value assigned by MAX • moduleSlot—1 • channel—The channel number from step 7 • SCXIgain—The gain value from step 14 • TBgain—1 • DAQboard—The device number of the DAQ device • DAQchannel—0 • DAQgain—1 • numPoints—5000 • binArray—The array returned by SCAN_Op in step 16 The result is a set of scaled voltages called scaledArray. 18. Average the results returned by SCXI_Scale. Compare the averaged result to the upper and lower limits in Table 2. If the result falls within these limits, your module has passed this portion of the test. If the result is outside these limits, adjust the module using the steps in the Adjusting for Gain, Offset, and Excitation Errors section. 19. Repeat steps 14 through 18 for the remaining gains. 20. Repeat steps 7 through 19 for the remaining channels. If all channels fall within the upper and lower limits, the gain and offset do not need adjustment. Note Verifying Voltage Excitation This section describes how to verify the voltage excitation of the SCXI-1520 and how to calculate new voltage excitation constants. The calibration constants for the voltage excitation on the SCXI-1520 are calculated with respect to the onboard 2.5 V reference voltage. These calibration constants are stored on the module EEPROM and can be read by application software and used for software correction. You cannot physically adjust the excitation on the SCXI-1520. If the measured values of the excitation fall outside the limits given in Table 3, the module is not © National Instruments Corporation 7 SCXI-1520 Calibration Procedure operating within specifications. Return it to NI for repair. For more information about how to contact NI for repair, refer to the Signal Conditioning Technical Support Information document. 1. Read the Test Conditions section. 2. Connect the DAQ device to the SCXI-1520 using an SCXI-1349 cable adapter and a 68-to-68-pin cable. 3. Connect the DMM to the appropriate excitation output terminals, starting with channel 0. Refer to Table 1 to determine which pins on the 96-pin front connector correspond to the positive and negative excitation terminals of the specified channel. For example, the positive excitation terminal for channel 0 (P0+) is pin C30. The negative excitation terminal for channel 0 (P0–) is pin B30. If you are using an SCXI-1314 terminal block, connect the DMM to the P0+ and P0– output terminals of the SCXI-1314. 4. Call SCXI_Set_Excitation to set the excitation voltage to the desired value. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • moduleChan—The channel number from step 3 • excitationType—2 • excitationValue—The excitation value that you want to measure. Refer to Table 3 for a list of valid values. SCXI_Set_Excitation returns a 32-bit floating-point number called actualExcitationValuePtr. This number was stored in the SCXI-1520 EEPROM when the module was last calibrated, and represents the voltage excitation measured at that time. Record this value as Vexe. SCXI-1520 Calibration Procedure 5. Read the voltage from the DMM and record it as Vexm. You will use Vexm to verify the voltage excitation and, if necessary, to calculate new calibration constants in the Adjusting Calibration Constants section. 6. Compare Vexm from step 5 to the upper and lower limits in Table 2. If Vexm is within these limits, the module has passed this portion of the test. If Vexm is outside these limits, the module is not operating within specifications. Return it to NI for repair. 7. Divide Vexe from step 4 by Vexm from step 5. If the resulting value is between 0.999 and 1.001, the calibration constants are within the required specifications. Do not use 0.0 V excitation for this step. 8. Repeat steps 4 through 7 for the remaining excitation levels you want to test on the module. 9. Repeat steps 3 through 8 for the remaining channels you want to test on the module. 8 ni.com Adjusting for Gain, Offset, and Excitation Errors The Determining Gain Errors and Adjusting Calibration Constants sections explain how to determine the gain, offset, and excitation errors of the SCXI-1520 and how to adjust the calibration constants to correct for these errors. Determining Gain Errors Complete the following steps to determine the gain errors of the SCXI-1520. 1. Read the Test Conditions section. 2. Refer to the specifications in Table 2. 3. Call Calibrate_E_Series to minimize the uncertainty associated with the DAQ device measurements. Set the following parameters: 4. 5. © National Instruments Corporation • deviceNumber—The value assigned by MAX • calOP—ND_SELF_CALIBRATE • setOfCalConst—ND_USER_EEPROM_AREA • calRefVolts—0.0 Call AI_Configure to put the DAQ device in bipolar mode. Set the following parameters: • deviceNumber—The value assigned by MAX • chan—0 • inputMode—0 • inputRange—0 • polarity—0 • driveAIS—0 Call SCXI_Track_Hold_Setup to configure the track-and-hold circuitry of the SCXI-1520. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • inputMode—2 • source—0 • send—0 • holdCount—1 • DAQdeviceNumber—The value assigned by MAX 9 SCXI-1520 Calibration Procedure 6. Call SCXI_MuxCtr_Setup to configure the DAQ device counter for use as a multiplexer counter during SCXI channel scanning. Set the following parameters: • deviceNumber—The DAQ device number assigned by MAX • enable—1 • scanDiv—1 • ctrValue—1 7. Connect the calibrator to the appropriate analog input, starting with channel 0. Refer to Table 1 to determine which pins on the 96-pin front connector correspond to the positive and negative inputs of the specified channel. For example, the positive input for channel 0 (S0+) is pin C32. The negative input for channel 0 (S0–) is pin B32. If you are using an SCXI-1314 terminal block, connect the calibrator to the S0+ and S0– inputs on the SCXI-1314. 8. Call SCXI_SCAN_Setup to set up the SCXI chassis to perform a multiplexed scanning data acquisition. Set the following parameters: 9. • SCXIchassisID—The value assigned by MAX • numModules—1 • moduleList—{1} • numChans—{1} • startChans—{The channel from step 7} • DAQdeviceNumber—The value assigned by MAX • modeFlag—0 Call SCXI_Configure_Filter to configure the filter on the module. Set the following parameters: • chassisID—The value assigned by MAX • moduleSlot—1 • channel—The channel from step 7 • filterMode—0 • freq—0 • cutoffDivDown—0 • outClkDivDown—0 SCXI_Configure_Filter returns the actual cutoff frequency that the module uses as a 64-bit floating-point number, actualFreq. SCXI-1520 Calibration Procedure 10 ni.com 10. Call SCXI_Set_Potentiometer to set the fine potentiometer to the center position. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • channel—The channel number from step 7 • value—2048 • pottype—1 11. Call SCXI_Set_Potentiometer to set the coarse potentiometer to the center position. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • channel—The channel number from step 7 • value—62 • pottype—0 12. Call SCXI_Set_Excitation to set the excitation voltage to zero. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • moduleChan—The channel number from step 7 • excitationType—2 • excitationValue—0 SCXI_Set_Excitation returns a 32-bit floating-point number called actualExcitationValuePtr. 13. Call SCXI_Configure_Connection to configure the channel for a full-bridge connection. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • moduleChan—The channel number from step 7 • connectionType—6 14. Call SCXI_Set_Gain to set the gain you want to test on the module. Set the following parameters: © National Instruments Corporation • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • channel—The channel number from step 7 • gain—The gain value you want to test. Refer to Table 2 for gain values and test points. 11 SCXI-1520 Calibration Procedure 15. Set the calibrator to output the test point voltage from Table 2. 16. Call SCAN_Op to acquire the data. Set the following parameters: • deviceNumber—The device number of the DAQ device • numChans—{1} • chans—{0} • gains—1 • buffer—[5000] • count—5000 • sampleRate—10000 • scanRate—5000 17. Call DAQ_VScale to scale the data returned by SCAN_Op. Set the following parameters: • deviceNumber—The value assigned to the DAQ device by MAX • chan—0 • gain—1 • gainAdjust—1 • offset—0 • count—5000 • binArray—The array returned by SCAN_Op The result is voltArray, a set of scaled voltages measured with respect to the output of the SCXI module. 18. Average the results returned by DAQ_VScale. 19. Record the averaged voltage from step 18 as Vmp and the calibrator voltage as Vinp. 20. Set the calibrator to the negative test-point value for the same gain. 21. Repeat steps 16 through 18. 22. Record the averaged voltage from step 18 as Vmn and the calibrator voltage as Vinn. You now have two pairs of data points, (Vmp, Vinp) and (Vmn, Vinn), to use in the Gain Calibration Constants section. 23. Repeat steps 14 through 22 for the remaining gains you want to adjust on this module. 24. Repeat steps 7 through 23 for the remaining channels you want to adjust on this module. SCXI-1520 Calibration Procedure 12 ni.com Adjusting Calibration Constants Complete the following steps to adjust the calibration constants on the SCXI-1520. Gain Calibration Constants 1. Calculate the gain adjustment for each channel using the following equation: V mp – V mn GainAdjust = ----------------------------------------------( V inp – V inn ) × gain where GainAdjust is a value used later in this process Vmp is the measured output of the module given a positive input voltage Vinp Vmn is the measured output of the module given a negative input voltage Vinn gain is the gain setting of the module 2. 3. © National Instruments Corporation Call SCXI_Adv_Module_Write to write the new gain calibration constants to the module. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • moduleClass—2 • address—2 × iquo + 9, where iquo is the integer part of the module channel divided by 2 • numBytes—4 • data—The value of GainAdjust represented as 4 bytes in a 4-element array stored little-endian • whichAccess—0 • regMethod— –1 • subAddress—(296 × irem + 722) + (4 × index), where irem is the integer remainder of the module channel divided by 2 and index is the index from Table 2 that corresponds to the gain you are adjusting Repeat steps 1 and 2 for the remaining gains and channels. 13 SCXI-1520 Calibration Procedure Offset Calibration Constants 1. 2. Call SCXI_Calibrate and set the following parameters: • SCXIChassisID—The value assigned by MAX • moduleSlot—1 • moduleChan—The number of the channel you want to calibrate • opCode—0 • calibrationArea—0 • SCXIgain—1 • refVoltage—1 • DAQdevice—The value assigned by MAX • DAQchan—0 Repeat step 1 for each of the remaining channels. Excitation Calibration Constants Note This adjustment is not needed for the 0.0 V excitation. 1. 2. Call SCXI_Adv_Module_Read to get the value of the 2.5 V reference voltage on the SCXI-1520. Use the following parameters: • SCXIchassisID—The value assigned by MAX • slot—1 • moduleClass—2 • address—3 • numBytes—4 • data—The array that returns the value of the SCXI-1520 2.5 V reference voltage. The data is returned as 4 bytes in a 4-element array stored little-endian. • whichAccess—0 • regMethod— –1 • subAddress—104 Calculate the excitation calibration constant to be written to the module. Use the following equation: ExcitationAdjust = Vexm/Vref_2.5 where Vexm is the measured value of the excitation voltage from step 5 of the Verifying Voltage Excitation section Vref_2.5 is the value of the 2.5 V reference voltage from step 1 SCXI-1520 Calibration Procedure 14 ni.com 3. 4. Call SCXI_Adv_Module_Write to write the new excitation adjust calibration constants to the module. Set the following parameters: • SCXIchassisID—The value assigned by MAX • moduleSlot—1 • moduleClass—2 • address—2 × iquo + 9, where iquo is the integer part of the module channel divided by 2 • numBytes—4 • data—The value of ExcitationAdjust represented as 4 bytes in a 4-element array stored little-endian • whichAccess—0 • regMethod— –1 • subAddress—(296 × irem + 654) + (4 × index), where irem is the integer remainder of the module channel divided by 2, and index is the index from Table 3 that corresponds to the excitation voltage you are adjusting Repeat steps 2 and 3 for the remaining excitation voltages and channels. Reading New Calibration Constants to Memory You can call SCXI_Get_Chassis_Info and SCXI_Get_Module_Info to find variables (parameters) for SCXI_Set_Config. Refer to the Traditional NI-DAQ Help for detailed function help. Note Reset NI-DAQ to read new calibration constants by calling SCXI_Set_Config with the following parameters: • SCXIchassisID—The value assigned by MAX • chassisType—0 for SCXI-1000, 1 for SCXI-1001, 4 for PXI-1010, 5 for PXI-1011 • address—0–31 as found in MAX • commMode—1 • commPath—MIO device number • numSlots—physical number of slots in chassis, 4, 8, or 12 • moduleTypes—array of length numSlots, with each element corresponding to a slot in the chassis. Use –1 if the index corresponds to an empty slot. Use 51 if the index corresponds to a slot containing an SCXI-1520. © National Instruments Corporation 15 SCXI-1520 Calibration Procedure • opModes—array of length numSlots, use –1 for an empty slot, 0 for multiplex mode, 1 for parallel mode, and 2 for parallel mode using the secondary connector of the DAQ device • DAQboards—array of size numSlots, with 0 for all slots with no DAQ device cabled to the module, and the MIO device number for the slot containing the module cabled to the DAQ device Verifying Adjusted Values After completing the adjustments to the SCXI-1520, it is important that you verify the analog input operation by repeating the steps listed in the Verifying Gain and Offset Errors and Verifying Voltage Excitation sections. Reverifying after adjustments ensures that the SCXI-1520 is operating within its specifications. Front Signal Connector Pin Assignments The pin assignments for the SCXI-1520 front signal connector are shown in Table 1. The front signal connector is a special 96-pin DIN C male connector through which you make all signal connections. The terminal assignments are as follows: • SX+ and SX– are for analog input • RSX+ and RSX– are for remote sense • PX+ and PX– are for excitation output • SCAX are for shunt calibration circuit A • SCBX are for shunt calibration circuit B where X is the channel number. The negative terminals are listed in Column B and the positive terminals are listed in Column C. The pins labeled RSVD are reserved. Do not make any connections to these pins. SCXI-1520 Calibration Procedure 16 ni.com Table 1. Front Signal Pin Assignments Front Connector Diagram Pin Number Column A 32 SCB0 S0– S0+ 31 SCB0 RS0– RS0+ 30 SCB1 P0– P0+ 32 29 SCB1 SCA0 SCA0 31 28 RSVD S1– S1+ 30 27 RSVD RS1– RS1+ 26 RSVD P1– P1+ 25 RSVD SCA1 SCA1 24 SCB2 S2– S2+ 25 23 SCB2 RS2– RS2+ 24 22 SCB3 P2– P2+ 23 21 SCB3 SCA2 SCA2 22 20 RSVD S3– S3+ 19 RSVD RS3– RS3+ 18 RSVD P3– P3+ 17 RSVD SCA3 SCA3 17 16 SCB4 S4– S4+ 16 15 SCB4 RS4– RS4+ 15 14 SCB5 P4– P4+ 14 13 SCB5 SCA4 SCA4 12 RSVD S5– S5+ 11 RSVD RS5– RS5+ 10 10 RSVD P5– P5+ 9 9 RSVD SCA5 SCA5 8 8 SCB6 S6– S6+ 7 7 SCB6 RS6– RS6+ 6 6 SCB7 P6– P6+ 5 SCB7 SCA6 SCA6 4 RSVD S7– S7+ 2 3 RSVD RS7– RS7+ 1 2 RSVD P7– P7+ 1 RSVD SCA7 SCA7 A Column B C 29 28 27 26 21 20 19 18 13 12 11 5 Column B Column C 4 3 © National Instruments Corporation 17 SCXI-1520 Calibration Procedure Specification Tables Table 2 contains the specifications you should use when verifying and calibrating the gain and offset of the SCXI-1520. If the module has been calibrated within the last year, the test point value should fall between the Lower Limit (V) and Upper Limit (V) values. Table 2. SCXI-1520 Gain and Offset Calibration Table Test Point (V) Gain Lower Limit (V) Upper Limit (V) Index 4.900000 1 4.891100 4.908900 0 0.000000 1 –0.004000 0.004000 0 –4.900000 1 –4.908900 –4.891100 0 4.260870 1.15 4.253130 4.268609 1 0.000000 1.15 –0.003478 0.003478 1 –4.260870 1.15 –4.268609 –4.253130 1 3.769230 1.3 3.762385 3.776077 2 0.000000 1.3 –0.003077 0.003077 2 –3.769230 1.3 –3.776077 –3.762385 2 3.266670 1.5 3.260733 3.272600 3 0.000000 1.5 –0.002667 0.002667 3 –3.266670 1.5 –3.272600 –3.260733 3 2.722220 1.8 2.717278 2.727167 4 0.000000 1.8 –0.002222 0.002222 4 –2.722220 1.8 –2.727167 –2.717278 4 2.450000 2 2.445550 2.454450 5 0.000000 2 –0.002000 0.002000 5 –2.450000 2 –2.454450 –2.445550 5 2.227270 2.2 2.223227 2.231318 6 0.000000 2.2 –0.001818 0.001818 6 –2.227270 2.2 –2.231318 –2.223227 6 2.041670 2.4 2.037958 2.045375 7 0.000000 2.4 –0.001667 0.001667 7 SCXI-1520 Calibration Procedure 18 ni.com Table 2. SCXI-1520 Gain and Offset Calibration Table (Continued) Test Point (V) Gain Lower Limit (V) Upper Limit (V) Index –2.041670 2.4 –2.045375 –2.037958 7 1.814810 2.7 1.811519 1.818111 8 0.000000 2.7 –0.001481 0.001481 8 –1.814810 2.7 –1.818111 –1.811519 8 1.580650 3.1 1.577774 1.583516 9 0.000000 3.1 –0.001290 0.001290 9 –1.580650 3.1 –1.583516 –1.577774 9 1.361110 3.6 1.358639 1.363583 10 0.000000 3.6 –0.001111 0.001111 10 –1.361110 3.6 –1.363583 –1.358639 10 1.166670 4.2 1.164548 1.168786 11 0.000000 4.2 –0.000952 0.000952 11 –1.166670 4.2 –1.168786 –1.164548 11 0.875000 5.6 0.873411 0.876589 12 0.000000 5.6 –0.000714 0.000714 12 –0.875000 5.6 –0.876589 –0.873411 12 0.753850 6.5 0.752477 0.755215 13 0.000000 6.5 –0.000615 0.000615 13 –0.753850 6.5 –0.755215 –0.752477 13 0.653330 7.5 0.652147 0.654520 14 0.000000 7.5 –0.000533 0.000533 14 –0.653330 7.5 –0.654520 –0.652147 14 0.563220 8.7 0.562195 0.564241 15 0.000000 8.7 –0.000460 0.000460 15 –0.563220 8.7 –0.564241 –0.562195 15 0.490000 10 0.489110 0.490890 16 0.000000 10 –0.000400 0.000400 16 © National Instruments Corporation 19 SCXI-1520 Calibration Procedure Table 2. SCXI-1520 Gain and Offset Calibration Table (Continued) Test Point (V) Gain Lower Limit (V) Upper Limit (V) Index –0.490060 10 –0.490890 –0.489110 16 0.426090 11.5 0.425313 0.426861 17 0.000000 11.5 –0.000348 0.000348 17 –0.426090 11.5 –0.426861 –0.425313 17 0.376920 13 0.376238 0.377608 18 0.000000 13 –0.000308 0.000308 18 –0.376920 13 –0.377608 –0.376238 18 0.326670 15 0.326073 0.327260 19 0.000000 15 –0.000267 0.000267 19 –0.326670 15 –0.327260 –0.326073 19 0.272220 18 0.271728 0.272717 20 0.000000 18 –0.000222 0.000222 20 –0.272220 18 –0.272717 –0.271728 20 0.245000 20 0.244555 0.245445 21 0.000000 20 –0.000200 0.000200 21 –0.245000 20 –0.245445 –0.244555 21 0.222730 22 0.222323 0.223132 22 0.000000 22 –0.000182 0.000182 22 –0.222730 22 –0.223132 –0.222323 22 0.204170 24 0.203796 0.204538 23 0.000000 24 –0.000167 0.000167 23 –0.204170 24 –0.204538 –0.203796 23 0.181480 27 0.181152 0.181811 24 0.000000 27 –0.000148 0.000148 24 –0.181480 27 –0.181811 –0.181152 24 0.158060 31 0.157777 0.158352 25 0.000000 31 –0.000129 0.000129 25 SCXI-1520 Calibration Procedure 20 ni.com Table 2. SCXI-1520 Gain and Offset Calibration Table (Continued) Test Point (V) Gain Lower Limit (V) Upper Limit (V) Index –0.158060 31 –0.158352 –0.157777 25 0.136110 36 0.135864 0.136358 26 0.000000 36 –0.000111 0.000111 26 –0.136110 36 –0.136358 –0.135864 26 0.116670 42 0.116455 0.116879 27 0.000000 42 –0.000095 0.000095 27 –0.116670 42 –0.116879 –0.116455 27 0.087500 56 0.087341 0.087659 28 0.000000 56 –0.000071 0.000071 28 –0.087500 56 –0.087659 –0.087341 28 0.075380 65 0.075248 0.075521 29 0.000000 65 –0.000062 0.000062 29 –0.075380 65 –0.075522 –0.075248 29 0.065330 75 0.065215 0.065452 30 0.000000 75 –0.000053 0.000053 30 –0.065330 75 –0.065452 –0.065215 30 0.056320 87 0.056220 0.056424 31 0.000000 87 –0.000046 0.000046 31 –0.056320 87 –0.056424 –0.056220 31 0.049000 100 0.048891 0.049109 32 0.000000 100 –0.000060 0.000060 32 –0.049000 100 –0.049109 –0.048891 32 0.042610 115 0.042507 0.042710 33 0.000000 115 –0.000059 0.000059 33 –0.042610 115 –0.042710 –0.042507 33 0.037690 130 0.037597 0.037788 34 0.000000 130 –0.000058 0.000058 34 © National Instruments Corporation 21 SCXI-1520 Calibration Procedure Table 2. SCXI-1520 Gain and Offset Calibration Table (Continued) Test Point (V) Gain Lower Limit (V) Upper Limit (V) Index –0.037690 130 –0.037788 –0.037597 34 0.032670 150 0.032577 0.032756 35 0.000000 150 –0.000057 0.000057 35 –0.032670 150 –0.032756 –0.032577 35 0.027220 180 0.027139 0.027305 36 0.000000 180 –0.000056 0.000056 36 –0.027220 180 –0.027305 –0.027139 36 0.024500 200 0.024421 0.024580 37 0.000000 200 –0.000055 0.000055 37 –0.024500 200 –0.024580 –0.024421 37 0.022270 220 0.022196 0.022350 38 0.000000 220 –0.000055 0.000055 38 –0.022270 220 –0.022350 –0.022196 38 0.020420 240 0.020342 0.020491 39 0.000000 240 –0.000054 0.000054 39 –0.020420 240 –0.020491 –0.020342 39 0.018150 270 0.018076 0.018220 40 0.000000 270 –0.000054 0.000054 40 –0.018150 270 –0.018220 –0.018076 40 0.015810 310 0.015737 0.015875 41 0.000000 310 –0.000053 0.000053 41 –0.015810 310 –0.015875 –0.015737 41 0.013610 360 0.013545 0.013678 42 0.000000 360 –0.000053 0.000053 42 –0.013610 360 –0.013678 –0.013545 42 0.011670 420 0.011603 0.011731 43 0.000000 420 –0.000052 0.000052 43 SCXI-1520 Calibration Procedure 22 ni.com Table 2. SCXI-1520 Gain and Offset Calibration Table (Continued) Test Point (V) Gain Lower Limit (V) Upper Limit (V) Index –0.011670 420 –0.011731 –0.011603 43 0.008750 560 0.008689 0.008811 44 0.000000 560 –0.000052 0.000052 44 –0.008750 560 –0.008811 –0.008689 44 0.007540 650 0.007479 0.007598 45 0.000000 650 –0.000052 0.000052 45 –0.007540 650 –0.007598 –0.007479 45 0.006530 750 0.006475 0.006592 46 0.000000 750 –0.000052 0.000052 46 –0.006530 750 –0.006592 –0.006475 46 0.005630 870 0.005575 0.005690 47 0.000000 870 –0.000052 0.000052 47 –0.005630 870 –0.005690 –0.005575 47 0.004900 1000 0.004843 0.004957 48 0.000000 1000 –0.000052 0.000052 48 –0.004900 1000 –0.004957 –0.004843 48 © National Instruments Corporation 23 SCXI-1520 Calibration Procedure Table 3. Excitation Limits Excitation (V) Lower Limit (V) Upper Limit (V) Index 0 –0.020000 0.020000 0 0.625 0.603125 0.646875 1 1.25 1.226250 1.273750 2 1.875 1.849375 1.900625 3 2.5 2.472500 2.527500 4 3.125 3.095625 3.154375 5 3.75 3.718750 3.781250 6 4.375 4.341875 4.408125 7 5 4.965000 5.035000 8 5.625 5.588125 5.661875 9 6.25 6.211250 6.288750 10 6.875 6.834375 6.915625 11 7.5 7.457500 7.542500 12 8.125 8.080625 8.169375 13 8.75 8.703750 8.796250 14 9.375 9.326875 9.423125 15 10 9.950000 10.050000 16 LabVIEW™, Measurement Studio™, National Instruments™, NI™, ni.com™, NI-DAQ™, and SCXI™ are trademarks of National Instruments Corporation. Product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your CD, or ni.com/patents. © 2002–2003 National Instruments Corp. All rights reserved. *370396B-01* 370396B-01 Sep03