NI 9218 Calibration Procedure

CALIBRATION PROCEDURE
NI 9218
2-Channel, 24-Bit, Universal Analog Input Module
This document contains information for calibrating the National Instruments 9218. For more
information on calibration, visit ni.com/calibration.
Contents
Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 4
Initial Setup............................................................................................................................... 5
Verification ............................................................................................................................... 5
Current Accuracy Verification, 20 mA mode .................................................................. 6
Voltage Accuracy Verification, 60 V mode ..................................................................... 7
Voltage Accuracy Verification, 16 V mode ..................................................................... 9
Voltage Accuracy Verification, 65 mV mode .................................................................. 10
Voltage Accuracy Verification, IEPE 5 V mode.............................................................. 12
Bridge Accuracy Verification, 22 mV/V mode................................................................ 13
Adjustment................................................................................................................................ 16
Voltage Accuracy Adjustment, 16 V mode...................................................................... 16
Voltage Accuracy Adjustment, 60 V mode...................................................................... 18
Voltage Accuracy Adjustment, 65 mV mode................................................................... 19
Current Accuracy Adjustment, 20 mA mode ................................................................... 20
Voltage Accuracy Adjustment, IEPE 5 V mode .............................................................. 20
Bridge Accuracy Adjustment, 22 mV/V mode ................................................................ 21
EEPROM Update ..................................................................................................................... 23
Reverification ........................................................................................................................... 23
Accuracy Under Calibration Conditions .................................................................................. 23
Worldwide Support and Services ............................................................................................. 24
Pinout Signals ........................................................................................................................... 25
Software
Calibrating the NI 9218 requires the installation of NI-DAQmx 15.1or later on the calibration
system. You can download NI-DAQmx from ni.com/downloads. NI-DAQmx supports
LabVIEW, LabWindows™/CVI™, C/C++, C#, and Visual Basic .NET. When you install
NI-DAQmx you only need to install support for the application software that you intend to use.
Documentation
Consult the following documents for information about the NI 9218, NI-DAQmx, and your
application software. All documents are available on ni.com and help files install with the
software.
NI cDAQ-9174/9178 USB Chassis Quick Start
NI-DAQmx installation and hardware setup
NI 9218 Getting Started Guide
NI 9218 connection information
NI 9218 Datasheet
NI 9218 circuitry and specifications
NI-DAQmx Readme
Operating system and application software support in NI-DAQmx
LabVIEW Help
LabVIEW programming concepts and reference information about NI-DAQmx
VIs and functions
NI-DAQmx C Reference Help
Reference information for NI-DAQmx C functions and NI-DAQmx C properties
NI-DAQmx .NET Help Support for Visual Studio
Reference information for NI-DAQmx .NET methods and NI-DAQmx .NET
properties, key concepts, and a C enum to .NET enum mapping table
2
|
ni.com
|
NI 9218 Calibration Procedure
Test Equipment
Table 1 lists the equipment recommended for the performance verification and adjustment
procedures. If the recommended equipment is not available, select a substitute using the
requirements listed in Table 1.
Table 1. Recommended Equipment
Equipment
Recommended
Model
Calibrator
Fluke 5700A
Where Used
Verification
Adjustment
Minimum Requirements
Use a device that can source
voltages as high as ±50 V and
meets the following
specifications:
≤100 ppm gain error.
±60 V,
±16 V,
±65 mV
mode
±16 V,
±65 mV
mode
≤3 µV offset error.
<3 µV offset at voltages less
than 60 mV.
<200 mV offset at voltages
less than 15 V.
<1 mV offset at voltages less
than 60 V.
—
Use a device that can source
currents as high as ±15 mA,
<200 ppm gain error, and
offset error <1 µA.
IEPE ±5 V
mode
IEPE ±5 V
mode
Use a device that can source
1 kHz, 3 Vrms AC sinusoidal
signals with amplitude
uncertainty of 200 ppm
or less.
—
±60 V mode
± 20 mA
mode
DMM (x2)
NI 4071
Bridge,
±22 mV/V
mode
Bridge,
±22 mV/V
mode
NI 9218 Calibration Procedure
Use a DMM with <40 Ω of
error at resistances <400 kΩ.
For DMM1, use one DMM
with <50 ppm gain error, and
<50 µV of offset error at
voltages <3.5 V.
For DMM2, use a second
DMM with<50 ppm gain
error, and <2 µV of offset
error at voltages <50 mV.
|
© National Instruments
|
3
Table 1. Recommended Equipment (Continued)
Equipment
Recommended
Model
Where Used
Verification
Adjustment
Minimum Requirements
Bridge,
±22 mV
mode
TCR:  ππµ/°C
—
Bridge,
±22 mV
mode
cDAQ
Chassis
NI cDAQ-9178
All
All
—
Measurement
Adapter
NI 9982 (x2)
±16 V,
±65 mV,
Bridge
±22 mV/V,
IEPE ±5 V
mode
±60 V, ±16 V,
±65 mV,
Bridge
±22 mV/V,
IEPE ±5 V
mode
—
NI 9983 (x2)
±20 mA
mode
—
—
NI 9987 (x2)
±60 V mode
—
—
Discrete
Resistors
Resistor values:
10 kΩ (2x), 249 Ω
Test Conditions
The following setup and environmental conditions are required to ensure the NI 9218 meets
published specifications.
•
Keep connections to the device as short as possible. Long cables and wires act as antennas,
picking up extra noise that can affect measurements.
•
Verify that all connections to the device are secure.
•
Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire to
eliminate noise and thermal offsets.
•
Maintain an ambient temperature of 23 ±5 °C. The device temperature will be greater than
the ambient temperature.
•
Keep relative humidity below 80%.
•
Allow a warm-up time of at least 10 minutes to ensure that the NI 9218 measurement
circuitry is at a stable operating temperature.
•
Only one of each NI 9982, NI 9983, and NI 9987 is sufficient for this calibration procedure.
4
|
ni.com
|
NI 9218 Calibration Procedure
Initial Setup
Complete the following steps to set up the NI 9218.
1.
Install NI-DAQmx.
2.
Make sure the NI cDAQ-9178 power source is not connected.
3.
Connect the NI cDAQ-9178 to the system safety ground.
a.
Attach a ring lug to a 14 AWG (1.6 mm) wire.
b.
Connect the ring lug to the ground terminal on the side of the NI cDAQ-9178 using
the ground screw.
c.
Attach the other end of the wire to the system safety ground.
4.
Install the module in slot 8 of the NI cDAQ-9178 chassis. Leave slots 1 through 7 of the
NI cDAQ-9178 chassis empty.
5.
Connect the NI cDAQ-9178 chassis to your host computer.
6.
Connect the power source to the NI cDAQ-9178 chassis.
7.
Launch Measurement & Automation Explorer (MAX).
8.
Right-click the device name and select Self-Test to ensure that the module is working
properly.
9.
Ensure that the ambient temperature of your test environment is 23 °C ±5 °C.
Verification
This section provides instructions for verifying the NI 9218 specifications. The following
performance verification procedures describe the sequence of operations and test points required
to verify the NI 9218 and assumes that adequate traceable uncertainties are available for the
calibration references.
Note Each verification procedure requires the software configuration of a
measurement adapter. The software configuration is done through Measurement &
Automation Explorer. See KB 6T8GKTUX for how to configure your measurement
adapter in MAX.
Each verification may have different wiring schemes. Consult the
diagram on your measurement adapter or Figure 10 and 11 for the most appropriate
connection diagram.
Caution
Caution The analog inputs on the NI 9218 are not grounded (floating). Ensure that
a single point in the test system is connected to ground to prevent the entire system
from floating.
NI 9218 Calibration Procedure
|
© National Instruments
|
5
Current Accuracy Verification, 20 mA mode
Complete the following procedure to determine the As-Found status of the NI 9218.
1.
Set the calibrator to Standby mode (STBY).
2.
Connect the NI 9218 to the calibrator as shown in Figure 1.
AI+
–
AI–
EX+
DSUB/LEMO
+
CH 0
DSUB/LEMO
Figure 1. Current Accuracy Verification Connections
CH 1
EX–
Calibrator
NI 9983
AI+
AI–
EX+
EX–
NI 9983
NI 9218
3.
Set the calibrator current to a Test Point value indicated in Table 4.
4.
Set the calibrator to Operate mode (OPR).
5.
Configure the NI 9218 for the NI 9983 measurement adapters using MAX.
6.
Acquire and average samples.
a.
Create an AI current task on the NI 9218 according to Table 2.
Table 2. NI 9218 Configuration for Accuracy Verification
Min (mA)
Max (mA)
Scaled Units
Channel Configuration
-20
20
Amps
DC Coupled
b.
Configure the AI current task timing according to Table 3.
Table 3. NI 9218 Timing Configuration
7.
6
Sample Mode
Samples Per Channel
Rate (kHz)
Finite Samples
512 k
51.2
c.
Start the task.
d.
Average the readings and record the average.
e.
Stop and clear the task.
Set the calibrator to Standby mode (STBY).
|
ni.com
|
NI 9218 Calibration Procedure
8.
Compare the average to the limits in Table 4.
9.
Repeat steps 3 through 8 for each test point in Table 4.
Table 4. NI 9218 Verification Test Limits for Accuracy
Test Point
2-Year Limits
Location
Value (mA)
Lower Limit (mA)
Upper Limit (mA)
Max
15
14.929
15.071
Mid
0
-0.008
0.008
Min
-15
-15.071
-14.929
Note
The test limits listed in Table 4 are derived using the values in Table 32.
Note
The uncertainty of the calibration standard is not included in these limits.
10. Disconnect the calibrator from the NI 9218.
Voltage Accuracy Verification, 60 V mode
Complete the following procedure to determine the As-Found status of the NI 9218.
1.
Set the calibrator to Standby mode (STBY).
2.
Connect the NI 9218 to the calibrator as shown in Figure 2.
AI+
–
AI–
Calibrator
DSUB/LEMO
+
CH 0
DSUB/LEMO
Figure 2. Voltage Accuracy Verification Connections
CH 1
NI 9987
AI+
AI–
NI 9987
3.
NI 9218
Set the calibrator voltage to a Test Point value indicated in Table 7.
4.
Set the calibrator to Operate mode (OPR).
5.
Configure the NI 9218 for the 9987 measurement adapters using MAX.
NI 9218 Calibration Procedure
|
© National Instruments
|
7
6.
Acquire and average samples.
a.
Create an AI voltage task on the NI 9218 according to Table 5.
Table 5. NI 9218 Configuration for Accuracy Verification
Min (V)
Max (V)
Scaled Units
Channel Configuration
-60
60
Volts
DC Coupled
b.
Configure the AI voltage task timing according to Table 6.
Table 6. NI 9218 Timing Configuration
7.
Sample Mode
Samples Per Channel
Rate (kHz)
Finite Samples
512 k
51.2
c.
Start the task.
d.
Average the readings and record the average.
e.
Stop and clear the task.
Set the calibrator to Standby mode (STBY).
8.
Compare the average to the limits in Table 7.
9.
Repeat steps 3 through 8 for each test point in Table 7.
Table 7. NI 9218 Verification Test Limits for Accuracy
Test Point
2-Year Limits
Location
Value (V)
Lower Limit (V)
Upper Limit (V)
Max
50
49.827
50.173
Mid
0
-0.008
0.008
Min
-50
-50.173
-49.827
Note
The test limits listed in Table 7 are derived using the values in Table 32.
Note
The uncertainty of the calibration standard is not included in these limits.
10. Disconnect the calibrator from the device.
8
|
ni.com
|
NI 9218 Calibration Procedure
Voltage Accuracy Verification, 16 V mode
Complete the following procedure to determine the As-Found status of the NI 9218.
1.
Set the calibrator to Standby mode (STBY).
2.
Connect the NI 9218 to the calibrator as shown in Figure 3.
AI+
–
AI–/EX–
Calibrator
DSUB/LEMO
+
CH 0
DSUB/LEMO
Figure 3. Voltage Accuracy Verification Connections
CH 1
NI 9982
AI+
AI–/EX–
NI 9218
NI 9982
3.
Set the calibrator voltage to a Test Point value indicated in Table 10.
4.
Set the calibrator to Operate mode (OPR).
5.
Configure the NI 9218 for the 9982 measurement adapters using MAX.
6.
Acquire and average samples.
a.
Create an AI voltage task on the NI 9218 according to Table 8.
Table 8. NI 9218 Configuration for Accuracy Verification
Min (V)
Max (V)
Scaled Units
Channel Configuration
-16
16
Volts
DC Coupled
b.
Configure the AI voltage task timing according to Table 9.
Table 9. NI 9218 Timing Configuration
c.
7.
Sample Mode
Samples Per Channel
Rate (kHz)
Finite Samples
512 k
51.2
Start the task.
d.
Average the readings and record the averages.
e.
Stop and clear the task.
Set the calibrator to Standby mode (STBY).
NI 9218 Calibration Procedure
|
© National Instruments
|
9
8.
Compare the averages to the limits in Table 10.
9.
Repeat steps 3 through 8 for each test point in Table 10.
Table 10. NI 9218 Verification Test Limits for Accuracy
Test Point
2-Year Limits
Location
Value (V)
Lower Limit (V)
Upper Limit (V)
Max
12
11.987
12.013
Mid
0
-0.002
0.002
Min
-12
-12.013
-11.987
Note
The test limits listed in Table 10 are derived using the values in Table 32.
Note
The uncertainty of the calibration standard is not included in these limits.
10. Disconnect the calibrator from the device.
Voltage Accuracy Verification, 65 mV mode
Complete the following procedure to determine the As-Found status of the NI 9218.
1.
Set the calibrator to Standby mode (STBY).
2.
Connect the NI 9218 to the calibrator as shown in Figure 4.
AI+
–
AI–
EX+
Calibrator
EX–
CH 0
CH 1
NI 9982
AI+
AI–
EX+
EX–
DSUB/LEMO
+
DSUB/LEMO
Figure 4. Voltage Accuracy Verification Connections
NI 9982
NI 9218
3.
Set the calibrator voltage to a Test Point value indicated in Table 13.
4.
Set the calibrator to Operate mode (OPR).
5.
Configure the NI 9218 for the NI 9982 measurement adapters using MAX.
10
| ni.com
|
NI 9218 Calibration Procedure
6.
Acquire and average samples.
a.
Create an AI voltage task on the NI 9218 according to Table 11.
Table 11. NI 9218 Configuration for Accuracy Verification
Min (mV)
Max (mV)
Scaled Units
Channel Configuration
-65
65
Volts
DC Coupled
b.
Configure the AI voltage task timing according to Table 12.
Table 12. NI 9218 Timing Configuration
7.
Sample Mode
Samples Per Channel
Rate (kHz)
Finite Samples
512 k
51.2
c.
Start the task.
d.
Average the readings and record the averages.
e.
Stop and clear the task.
Set the calibrator to Standby mode (STBY).
8.
Compare the averages to the limits in Table 13.
9.
Repeat steps 3 through 8 for each test point in Table 13.
Table 13. NI 9218 Verification Test Limits for Accuracy
Test Point
2-Year Limits
Location
Value (mV)
Lower Limit (mV)
Upper Limit (mV)
Max
50
49.906
50.094
Mid
0
-0.024
0.024
Min
-50
-50.094
-49.906
Note
The test limits listed in Table 13 are derived using the values in Table 32.
Note
The uncertainty of the calibration standard is not included in these limits.
10. Disconnect the calibrator from the device.
NI 9218 Calibration Procedure
|
© National Instruments
|
11
Voltage Accuracy Verification, IEPE 5 V mode
Complete the following procedure to determine the As-Found status of the NI 9218.
1.
Connect the NI 9218 to the calibrator as shown in Figure 5.
AI+
–
AI-
Calibrator
CH 0
CH 1
NI 9982
AI+
AINI 9982
2.
DSUB/LEMO
+
DSUB/LEMO
Figure 5. Voltage Accuracy Verification Connections
NI 9218
Configure calibrator according to Table 14.
Table 14. NI 9218 Verification Test Limits for Accuracy
Test Point
2-Year Limits
Signal Type
Frequency
(kHz)
Amplitude
(Vrms)
Lower Limit
(Vrms)
Upper Limit
(Vrms)
Sinewave
1
1
0.998
1.002
3
2.994
3.006
Note
The test limits listed in Table 14 are derived using the values in Table 32.
Note
The uncertainty of the calibration standard is not included in these limits.
3.
Configure the NI 9218 for the NI 9982 measurement adapters using MAX.
4.
Acquire and calculate the AC magnitude.
a.
Create an AI voltage task on the NI 9218 according to Table 15.
Table 15. NI 9218 Configuration for Phase Matching Verification
12
Min (V)
Max (V)
Scaled Units
Channel Configuration
-5
5
Volts
AC Coupled
| ni.com
|
NI 9218 Calibration Procedure
5.
6.
Use the DAQmx Channel Property node to configure the following properties:
a.
AI.Coupling: AC
b.
AI.Excit.Src: Internal
c.
AI.Excit.VoltageorCurrent: Current
d.
AI.Excit.Val: 0.002
Configure the AI voltage task timing according to Table 16.
Table 16. NI 9218 Timing Configuration
Sample Mode
Samples Per Channel
Rate (kHz)
Finite Samples
512 k
51.2
a.
Set the calibrator to Operate mode (OPR).
b.
Start the task.
Note Wait 6 seconds for the AC coupling filter on the NI 9218 to settle before
acquiring samples.
c.
Calculate and record the AC magnitude for each channel on the NI 9218.
NI recommends using the Extract Single Tone Information VI to calculate the
magnitude.
Note The output of the Extract Single Tone Information is in Vpk. To convert to
RMS, use the following equation:
Vrms = (Vpk) ÷ (√2)
d.
Disable the calibrator.
e.
Stop and clear the task.
7.
Compare the AC magnitude to the limits in Table 14.
8.
Repeat steps 2 through 7 for each test point in Table 14.
9.
Disable the output of the calibrator and disconnect it from the NI 9218.
Bridge Accuracy Verification, 22 mV/V mode
Complete the following steps to verify the full-bridge accuracy of an NI 9218.
1.
Connect a resistor network to the NI 9218 and two DMMs as shown in Figures 6, 7, or 8
for the appropriate test point value indicated in Table 17. Observe the following conditions
when connecting the equipment to the NI 9218.
•
Connect the leads of DMM2 as close as possible to the AI+ and AI- terminals on the
NI 9982.
•
Connect the leads of DMM1 as close as possible to the RS+ and RS- leads of the
NI 9982.
NI 9218 Calibration Procedure
|
© National Instruments
|
13
Figure 6. Bridge Accuracy Verification Connections (Configuration 1)
10 k˖
HI
DMM 2
249 ˖
LO
Vb
RS+
EX+
AI+
AI–
EX–
RS–
DSUB/LEMO
Va
CH x
HI
DMM 1
LO
NI 9982
NI 9218
10 k˖
Figure 7. Bridge Accuracy Verification Connections (Configuration 2)
Va
HI
DMM 2
249 ˖
LO
Vb
RS+
EX+
AI+
AI–
EX–
RS–
NI 9982
DSUB/LEMO
10 k˖
CH x
HI
DMM 1
LO
NI 9218
10 k˖
Figure 8. Bridge Accuracy Verification Connections (Configuration 3)
10 k˖
HI
DMM 2
249 ˖
LO
Vb
RS+
EX+
AI+
AI–
EX–
RS–
DSUB/LEMO
Va
NI 9982
10 k˖
14
| ni.com
|
NI 9218 Calibration Procedure
CH x
HI
DMM 1
LO
NI 9218
Table 17. NI 9218 Verification Test Limits for Accuracy
Test Point
2-Year Limits
Excitation
Value
Location
Value
Lower Limit
(mV/V)
Upper Limit
(mV/V)
2V
Max
Configuration 1
-0.050
0.050
Mid
Configuration 2
-0.036
0.036
Min
Configuration 3
-0.050
0.050
Max
Configuration 1
-0.022
0.022
Mid
Configuration 2
-0.007
0.007
Min
Configuration 3
-0.022
0.022
3.3 V
Note
The test limits listed in Table 17 are derived using the values in Table 32.
Note
The uncertainty of the calibration standard is not included in these limits.
2.
Configure DMM1 for a voltage measurement in the 10 V range.
3.
Enable autozero on DMM1.
4.
Configure DMM2 for a voltage measurement in the 100 mV range.
5.
Enable autozero on DMM2.
6.
Configure the NI 9218 for the NI 9982 measurement adapters using MAX.
7.
Acquire and average samples.
a.
Create an AI bridge with excitation task on the NI 9218 according to Table 18.
Table 18. NI 9218 Configuration for Accuracy Verification
Min
(V/V)
Max
(V/V)
Channel
Configuration
Excitation
Value (V)
Excitation
Source
Bridge
Configuration
-0.022
0.022
DC Coupled
2
Internal
Full-Bridge
3.3
b.
Configure the AI Bridge with excitation task timing according to Table 19.
Table 19. NI 9218 Timing Configuration
c.
Sample Mode
Samples Per Channel
Rate (kHz)
Finite Samples
512 k
51.2
Start the task.
NI 9218 Calibration Procedure
|
© National Instruments
|
15
d.
Record the DMM1 measurements as VDMM1.
e.
Record the DMM2 measurement as VDMM2.
Perform the DMM measurements and the NI 9218 measurements as
simultaneously as possible in order to reduce error from drift in the resistors.
Note
f.
Average the NI 9218 readings and record the average as Ratiochannel.
g.
Perform the following calculation using the two DMM measurements.
V DMM2
Ratio ref = -----------------V DMM1
h.
Perform the following calculation using the recorded Ratioref and Ratiochannel values.
Accuracy = Ratiochannel - Ratioref
i.
Compare the result of step h to the limits in Table 17.
j.
Stop and clear the task.
8.
Repeat steps 2 through 7 for each test point in Table 17.
9.
Repeat steps 1 through 8 for each excitation value specified in Table 17.
10. Disconnect the DMMs and resistors from the NI 9218.
11. Repeat steps 1 through 10 for each channel on the NI 9218.
Adjustment
The following performance adjustment procedures describe the sequence of operation required
to adjust the NI 9218.
Voltage Accuracy Adjustment, 16 V mode
Complete the following procedure to adjust the accuracy performance of the NI 9218.
1.
Set the calibrator to Standby mode (STBY).
2.
Connect the NI 9218 to the calibrator as shown in Figure 3.
3.
Configure the NI 9218 for the 9982 measurement adapters using MAX.
4.
Adjust the NI 9218.
16
a.
Initialize a calibration session on the NI 9218. The default password is NI.
b.
Input the external temperature in degrees Celsius using the DAQmx Set Temperature
C Series Calibration function.
| ni.com
|
NI 9218 Calibration Procedure
c.
Call and configure the DAQmx Setup 9218 Calibration function according to
Table 20.
Table 20. Setup Configuration for 16 V Accuracy Adjustment
Min (V)
Max (V)
Physical Channels
Measurement Type
-16
16
cDAQMod8/AIx
Voltage
d.
Call the DAQmx Get 9218 Calibration Adjustment Points function to obtain an array
of recommended calibration voltages for the NI 9218.
e.
Set the calibrator to a reference value determined by the array of recommended
calibration voltages.
f.
Set the calibrator to Operate mode (OPR).
g.
Call and configure the DAQmx Adjust 9218 Calibration function according to
Table 21.
Table 21. Adjustment Configuration
Physical Channel
Reference Value
cDAQMod8/AIx
The recommended calibration voltage
Note
5.
The NI 9218 acquisition starts each time the adjustment function is called.
h.
Set the calibrator to Standby mode (STBY).
i.
Repeat steps e through h for each point in the array of recommended calibration
voltages.
j.
Use the DAQmx Close External Calibration to close and commit the calibration
session.
k.
Repeat steps c through j for each channel.
Disconnect the NI 9218 from the calibrator.
NI 9218 Calibration Procedure
|
© National Instruments
|
17
Voltage Accuracy Adjustment, 60 V mode
Complete the following procedure to adjust the accuracy performance of the NI 9218.
1.
You must complete the Voltage Accuracy Adjustment, 16 V mode section before starting
this section.
2.
Connect the NI 9218 to the DMM as shown in Figure 9.
3.
Configure the NI 9218 for the NI 9982 measurement adapters using MAX.
HI
AI+
LO
AI–
DMM
4.
DSUB/LEMO
Figure 9. Voltage Accuracy Adjustment Connections
CH x
NI 9982
NI 9218
Adjust the NI 9218.
a.
Initialize a calibration session on the NI 9218. The default password is NI.
b.
Input the external temperature in degrees Celsius using the DAQmx Set Temperature
C Series Calibration function.
c.
Call and configure the DAQmx Setup 9218 Calibration function according to
Table 22.
Table 22. Setup Configuration for 60 V Accuracy Adjustment
Min (V)
Max (V)
Physical Channels
Measurement Type
-60
60
cDAQMod8/AIx
Voltage
d.
Set the DMM to the most appropriate range for 390 kΩ resistance measurement.
e.
Measure the resistance.
f.
Call and configure the DAQmx Adjust 9218 Calibration function according to
Table 23.
Table 23. Adjustment Configuration
5.
18
Physical Channel
Reference Value
cDAQMod8/AIx
The reading from the DMM
g.
Use the DAQmx Close External Calibration to close and commit the calibration
session.
h.
Repeat steps c through g for each channel.
Disconnect the DMM from the NI 9218.
| ni.com
|
NI 9218 Calibration Procedure
Voltage Accuracy Adjustment, 65 mV mode
Complete the following procedure to adjust the accuracy performance of the NI 9218.
1.
Set the calibrator to Standby mode (STBY).
2.
Connect the NI 9218 to the calibrator as shown in Figure 4.
3.
Configure the NI 9218 for the NI 9982 measurement adapters using MAX.
4.
Adjust the NI 9218.
a.
Initialize a calibration session on the NI 9218. The default password is NI.
b.
Input the external temperature in degrees Celsius using the DAQmx Set Temperature
C Series Calibration function.
c.
Call and configure the DAQmx Setup 9218 Calibration function according to
Table 24.
Table 24. Setup Configuration for 65 mV Accuracy Adjustment
Min (V)
Max (V)
Physical Channels
Measurement Type
-0.065
0.065
cDAQMod8/AIx
Voltage
d.
Call the DAQmx Get 9218 Calibration Adjustment Points function to obtain an array
of recommended calibration voltages for the NI 9218.
e.
Set the calibrator to a reference value determined by the array of recommended
calibration voltages.
f.
Set the calibrator to Operate mode (OPR).
g.
Call and configure the DAQmx Adjust 9218 Calibration function according to
Table 25.
Table 25. Adjustment Configuration
Physical Channel
Reference Value
cDAQMod8/AIx
The recommended calibration voltage
Note
5.
The NI 9218 acquisition starts each time the adjustment function is called.
h.
Set the calibrator to Standby mode (STBY).
i.
Repeat steps e through h for each point in the array of recommended calibration
voltages.
j.
Use the DAQmx Close External Calibration to close and commit the calibration
session.
k.
Repeat steps c through j for each channel.
Disconnect the NI 9218 from the calibrator.
NI 9218 Calibration Procedure
|
© National Instruments
|
19
Current Accuracy Adjustment, 20 mA mode
Complete the following procedure to adjust the accuracy performance of the NI 9218.
1.
You must complete the Voltage Accuracy Adjustment, 65 mV mode section before starting
this section.
2.
Adjust the NI 9218.
a.
Initialize a calibration session on the NI 9218. The default password is NI.
b.
Input the external temperature in degrees Celsius using the DAQmx Set Temperature
C Series Calibration function.
c.
Call and configure the DAQmx Setup 9218 Calibration function according to
Table 26.
Table 26. Setup Configuration for 20 mA Accuracy Adjustment
Min (A)
Max (A)
Physical Channels
Measurement Type
-0.02
0.02
cDAQMod8/AIx
Current
d.
Call and configure the DAQmx 9218 Adjustment Calibration function according to
Table 27.
Table 27. Adjustment Configuration
Physical Channel
Reference Value
cDAQMod8/AIx
0
e.
Use the DAQmx Close External Calibration to close and commit the calibration
session.
f.
Repeat steps c through e for each channel.
Voltage Accuracy Adjustment, IEPE 5 V mode
Complete the following procedure to adjust the accuracy performance of the NI 9218.
1.
Connect the NI 9218 to the calibrator as shown in Figure 5.
2.
Configure the NI 9218 for the NI 9982 measurement adapters using MAX.
3.
Adjust the NI 9218.
20
a.
Initialize a calibration session on the NI 9218. The default password is NI.
b.
Input the external temperature in degrees Celsius using the DAQmx Set Temperature
C Series Calibration function.
| ni.com
|
NI 9218 Calibration Procedure
c.
Call and configure the DAQmx Set up 9218 Calibration function according to
Table 28.
Table 28. Setup Configuration for IEPE 5 V Accuracy Adjustment
Min (V)
Max (V)
Physical Channels
Measurement Type
-5
5
cDAQMod8/aix
Voltage
d.
Call the DAQmx Get 9218 Calibration Adjustment Points function to obtain an array
of recommended calibration amplitudes for the NI 9218.
Note This function provides values in terms of Vrms. A conversion is needed if
your Calibrator API requires a different signal description.
e.
Set the calibrator to a reference value determined by the array of recommended
calibration amplitudes. The signal should be a 1 kHz sinewave.
f.
Call and configure the DAQmx Adjust 9218 Calibration function according to
Table 29.
Table 29. Adjustment Configuration
Physical Channel
Reference Value
cDAQMod8/AIx
The recommended calibration amplitude
Note
4.
The NI 9218 acquisition starts each time the adjustment function is called.
g.
Repeat steps e through f for each point in the array of recommended calibration value.
h.
Use the DAQmx Close External Calibration to close and commit the calibration
session.
i.
Repeat steps c through h for each channel.
Disconnect the NI 9218 from the calibrator.
Bridge Accuracy Adjustment, 22 mV/V mode
Complete the following steps to adjust the full-bridge accuracy of the NI 9218.
1.
Connect the NI 9218 and DMMs to the resistor network as shown in Figures 6, 7, or 8
depending on the test point.
2.
Configure the NI 9218 for the NI 9982 measurement adapters using MAX.
3.
Adjust the NI 9218.
a.
Initialize a calibration session on the NI 9218. The default password is NI.
b.
Input the external temperature in degrees Celsius using the DAQmx Set Temperature
C Series Calibration function.
NI 9218 Calibration Procedure
|
© National Instruments
|
21
c.
Call and configure the DAQmx Setup 9218 Calibration function according to
Table 30.
Table 30. Setup Configuration for Bridge Accuracy Adjustment
Min (V/V)
Max (V/V)
Physical Channels
Measurement Type
-0.022
0.022
cDAQMod8/aix
Bridge
d.
Configure DMM1 for a voltage measurement in the 10 V range.
e.
Enable autozero on DMM1.
f.
Configure DMM2 for a voltage measurement in the 100 mV range.
g.
Enable autozero on DMM2.
h.
Call the DAQmx Get 9218 Calibration Adjustment Points function to obtain an array
of recommended calibration points for the NI 9218. The adjustment points can be
achieved by observing the following conditions.
•
i.
The max test point is achieved by following the diagram in Figure 6.
•
The mid test point is achieved by following the diagram in Figure 7.
•
The min test point is achieved by following the diagram in Figure 8.
Acquire voltage measurements with both DMMs. Perform the following calculation
using the two DMM measurements.
V DMM2
Ratio ref = -----------------V DMM1
Perform the DMM measurements simultaneously in order to reduce error
from drift in the resistors.
Note
j.
Call and configure the DAQmx Adjust 9218 Calibration function according to
Table 31.
Table 31. Adjustment Configuration
Physical Channel
Reference Value
cDAQMod8/AIx
Ratioref
Note
The NI 9218 acquisition starts each time the adjustment function is called.
k.
Repeat steps d through k for every value in the array.
l.
Close and commit the calibration session.
4.
Repeat steps 3c through 3l for each channel on the NI 9218.
5.
Disconnect the DMMs and the resistors from the NI 9218.
22
| ni.com
|
NI 9218 Calibration Procedure
EEPROM Update
When an adjustment procedure is completed, the NI 9218 internal calibration memory
(EEPROM) is immediately updated.
If you do not want to perform an adjustment, you can update the calibration date and onboard
calibration temperature without making any adjustments by initializing an external calibration,
setting the C Series calibration temperature, and closing the external calibration.
Reverification
Repeat the Verification section to determine the As-Left status of the device.
Note If any test fails Reverification after performing an adjustment, verify that you
have met the Test Conditions and have run verification with another measurement
adapter before returning your device to NI. Refer to Worldwide Support and Services
for assistance in returning the device to NI.
Accuracy Under Calibration Conditions
The values in the following table are based on calibrated scaling coefficients, which are stored
in the onboard EEPROM.
The following accuracy table is valid for calibration under the following conditions:
•
Ambient temperature 23 °C ± 5 °C
•
NI 9218 installed in slot 8 of an NI cDAQ-9178 chassis
•
Slots 1 through 7 of the NI cDAQ-9178 chassis are empty
Table 32. NI 9218 Accuracy Under Calibration Conditions*
*
Mode
Percent of Reading (Gain Error)
Offset Error
20 mA
0.42%
8 µ
60 V
0.33%
8 mV
16 V
0.09%
2 mV
65 mV
0.14%
24 V
IEPE 5 V
0.20%
—
Bridge 22 mV/V, 2 V
0.12%
36 µV/V
Bridge 22 mV/V, 3.3 V
0.12%
7.2 µV/V
The accuracy is inclusive of the NI 9982, NI 9983, and NI 9987.
NI 9218 Calibration Procedure
|
© National Instruments
|
23
Note For operational specifications, refer to the most recent NI 9218 Datasheet
online at ni.com/manuals.
Worldwide Support and Services
The National Instruments website is your complete resource for technical support. At ni.com/
support you have access to everything from troubleshooting and application development
self-help resources to email and phone assistance from NI Application Engineers.
Visit ni.com/services for NI Factory Installation Services, repairs, extended warranty, and
other services.
Visit ni.com/register to register your National Instruments product. Product registration
facilitates technical support and ensures that you receive important information updates from NI.
A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European
Communities using the manufacturer’s declaration of conformity. This system affords the user
protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC
for your product by visiting ni.com/certification. If your product supports calibration,
you can obtain the calibration certificate for your product at ni.com/calibration.
National Instruments corporate headquarters is located at 11500 North Mopac Expressway,
Austin, Texas, 78759-3504. National Instruments also has offices located around the world. For
telephone support in the United States, create your service request at ni.com/support or dial
1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the
Worldwide Offices section of ni.com/niglobal to access the branch office websites, which
provide up-to-date contact information, support phone numbers, email addresses, and current
events.
24
| ni.com
|
NI 9218 Calibration Procedure
Pinout Signals
Figure 10. NI 9218 with DSUB Pinout and Pin Assignments
DSUB PINOUT
PIN ASSIGNMENT BY MODE
Pin Full-Bridge ±65 mV
5
9
4
8
3
7
2
6
1
9
8
7
6
5
4
3
2
1
±16 V
IEPE TEDS
1
EX+
EX+*
EX+*
—
—
2
—
—
—
AI+
T+
3
EX–
EX–*
AI–, EX–*
AI–
T–
4
RS+
—
—
—
—
5
RS–
—
—
—
—
6
AI+
AI+
AI+
—
—
7
AI–
AI– †
—
—
—
8
SC
—
—
—
—
9
SC
—
—
—
—
* Optional sensor excitation.
†In ±65 mV mode, you must tie pin 7 (AI–) to
V+ V–
pin 3 (EX–).
NI 9218 Calibration Procedure
|
© National Instruments
|
25
Figure 11. NI 9218 with LEMO Pinout and Pin Assignments
LEMO PINOUT
4
3 2
PIN ASSIGNMENT BY MODE
1
10 9
5
4
6 7
3 2
8
6 7
Full-Bridge
±65 mV
±16 V
IEPE
TEDS
1
EX+
EX+*
EX+*
—
—
2
—
—
—
AI+
T1+
3
EX–
EX–*
AI–, EX–*
AI–
T1–, T2–
4
RS+
—
—
—
—
5
RS–
—
—
—
—
6
AI+
AI+
AI+
—
—
7
AI–
AI– †
—
—
—
8
SC
—
—
—
—
9
SC
—
—
—
—
10
—
—
—
—
T2+
1
10 9
5
Pin
8
V+
V+
V-
V-
* Optional sensor excitation.
† In ±65 mV mode, you must tie pin 7 (AI–) to
pin 3 (EX–).
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on NI trademarks. Other product and company
names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the
appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patents Notice at
ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your
NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy
and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE
ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data
contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in
FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2015 National Instruments. All rights reserved.
375372A-01
Sep15