EVALUATION KIT AVAILABLE MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters General Description The MAX31850/MAX31851 cold-junction compensated, 1-WireM thermocouple-to-digital converters perform cold-junction compensation and digitize the signal from a K-, J-, N-, T-, S-, R-, or E-type thermocouple. The converters resolve temperatures to 0.25NC, allow readings as high as +1768NC and as low as -270NC, and exhibit thermocouple accuracy of 8 LSBs (2NC) for temperatures ranging from -200NC to +700NC. Communication with the master microcontroller is over a 1-Wire bus that by definition requires only one data line (and ground) for communication. Operating power can be obtained directly from the data line (“parasite power”), eliminating the need for an external power supply. Each device has a unique 64-bit serial code, which allows multiple units to function on the same 1-Wire bus. Therefore, it is simple to use one microcontroller (the master device) to monitor temperature from many thermocouples distributed over a large area. Features SCold-Junction Compensation S14-Bit, 0.25NC Resolution SVersions Available for K-, J-, N-, T-, S-, R-, and E-Type Thermocouples (see Table 1) S1-Wire Interface (Read-Only); Power Can Be Obtained from Interface (Parasite-Powered Mode) SDetects Thermocouple Shorts to GND or VDD SDetects Open Thermocouple Applications Industrial HVAC Appliances Medical Ordering Information appears at end of data sheet. Four location address inputs simplify mapping of individual units to specific locations. Block Diagram DQ COLD-JUCTION COMPENSATION S5 MAX31850 MAX31851 PARASITEPOWER CIRCUIT 64-BIT ROM AND 1-Wire PORT T+ CPP S4 ADC POWERSUPPLY SENSE T- VDD S1 S2 S3 FAULT DETECTION MEMORY CONTROL LOGIC REFERENCE VOLTAGE SCRATCHPAD ADDRESS PIN INPUT GND AD0 AD1 AD2 AD3 1-Wire is a registered trademark of Maxim Integrated Products, Inc. For related parts and recommended products to use with this part, refer to: www.maximintegrated.com/MAX31850.related For pricing, delivery, and ordering information, please contact Maxim Direct at 1-888-629-4642, or visit Maxim Integrated’s website at www.maximintegrated.com. 19-6725; Rev 0; 6/13 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters ABSOLUTE MAXIMUM RATINGS Supply Voltage Range (VDD to GND)...................-0.3V to +4.0V All Other Pins............................................. -0.3V to (VDD + 0.3V) Continuous Power Dissipation (TA = +70NC) TDFN (derate 16.70mW/NC above +70NC)............1333.30mW ESD Protection (All Pins, Human Body Model)....................±2kV Operating Temperature Range......................... -40NC to +125NC Junction Temperature......................................................+150NC Storage Temperature Range ........................... -65NC to +150NC Lead Temperature (soldering, 10s).................................+300NC Soldering Temperature (reflow) ......................................+260NC Stresses beyond those listed under “Absolute Maximum Ratings” may cause permanent damage to the device. These are stress ratings only, and functional operation of the device at these or any other conditions beyond those indicated in the operational sections of the specifications is not implied. Exposure to absolute maximum rating conditions for extended periods may affect device reliability. PACKAGE THERMAL CHARACTERISTICS (Note 1) TDFN Junction-to-Ambient Thermal Resistance (BJA)...........60NC/W Junction-to-Case Thermal Resistance (BJC)................30NC/W Note 1: Package thermal resistances were obtained using the method described in JEDEC specification JESD51-7, using a four-layer board. For detailed information on package thermal considerations, refer to www.maximintegrated.com/thermal-tutorial. DC ELECTRICAL CHARACTERISTICS (TA = -40NC to +125NC, unless otherwise noted.) (Note 2) PARAMETER SYMBOL Supply Voltage VDD Pullup Supply Voltage (Notes 3, 4) VPU Input Logic-Low VIL MIN TYP MAX UNITS Local power (Note 3, 4) CONDITIONS 3.0 3.3 3.7 V Parasite power (Note 3) 3.0 3.7 Local power (Note 3) 3.0 VDD (Notes 4, 6) -0.3 +0.8 2.2 lower of 3.7V or (VDD + 0.3V) Parasite power 3.0 lower of 3.7V or (VDD + 0.3V) VI/O = 0.4V (Note 4) 4.0 Local power Input Logic-High (Note 4) Output Sink Current VIH IL V V V mA Standby Supply Current IDDS (Notes 7, 8) 280 1000 nA Active Supply Current IDD VDD = 3.7V (Note 9) 900 1750 FA DQ Input Current IDQ (Note 10) 5 Power-Supply Rejection -0.3 Input Leakage Current (AD0–AD3 Pins) Thermocouple Input Bias Current Maxim Integrated FA (Note 3) IBTC TA = -40°C to +125°C, 100mV across the thermocouple inputs (Note 3) °C/V -1 +1 FA -100 +100 nA 2 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters THERMAL CHARACTERISTICS (3.0V P VDD P 3.6V, TA = -40NC to +125NC, unless otherwise noted.) (Note 2) PARAMETER MAX31850K Thermocouple Temperature Gain and Offset Error (41.276FV/NC nominal sensitivity) (Note 11) MAX31850J Thermocouple Temperature Gain and Offset Error (57.953FV/NC nominal sensitivity) (Note 11) MAX31850N Thermocouple Temperature Gain and Offset Error (36.256FV/NC nominal sensitivity) (Note 11) MAX31850T Thermocouple Temperature Gain and Offset Error (52.18FV/NC nominal sensitivity) (Note 11) MAX31850E Thermocouple Temperature Gain and Offset Error (76.373FV/NC nominal sensitivity) (Note 11) Maxim Integrated SYMBOL CONDITIONS MIN TYP MAX TTHERMOCOUPLE = -100NC to +100NC, TA = 0NC to +70NC (Note 12) -1 +1 TTHERMOCOUPLE = -200NC to +700NC, TA = -20NC to +85NC (Note 12) -2 +2 TTHERMOCOUPLE = +700NC to +1372NC, TA = -20NC to +85NC (Note 12) -4 +4 TTHERMOCOUPLE = -270NC to +1372NC, TA = -40NC to +125NC (Note 12) -6 +6 TTHERMOCOUPLE = -100NC to +100NC, TA = 0NC to +70NC (Note 12) -1 +1 TTHERMOCOUPLE = -210NC to +750NC, TA = -20NC to +85NC (Note 12) -2 +2 TTHERMOCOUPLE = -210NC to +1200NC, TA = -40NC to +125NC (Note 12) -4 +4 TTHERMOCOUPLE = -100NC to +100NC, TA = 0NC to +70NC (Note 12) -1 +1 TTHERMOCOUPLE = -200NC to +700NC, TA = -20NC to +85NC (Note 12) -2 +2 TTHERMOCOUPLE = +700NC to +1300NC, TA = -20NC to +85NC (Note 12) -4 +4 TTHERMOCOUPLE = -270NC to +1300NC, TA = -40NC to +125NC (Note 12) -6 +6 TTHERMOCOUPLE = -100NC to +100NC, TA = 0NC to +70NC (Note 12) -1 +1 TTHERMOCOUPLE = -270NC to +400NC, TA = -20NC to +85NC (Note 12) -2 +2 TTHERMOCOUPLE = -270NC to +400NC, TA = -40NC to +125NC (Note 12) -4 +4 TTHERMOCOUPLE = -120NC to +100NC, TA = -20NC to +85NC (Note 12) -1 +1 TTHERMOCOUPLE = -200NC to +700NC, TA = -20NC to +85NC (Note 12) -2 +2 TTHERMOCOUPLE = +700NC to +1000NC, TA = -20NC to +85NC (Note 12) -4 +4 TTHERMOCOUPLE = -270NC to +1000NC, TA = -40NC to +125NC (Note 12) -5 +5 UNITS NC NC NC NC NC 3 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters THERMAL CHARACTERISTICS (continued) (3.0V P VDD P 3.6V, TA = -40NC to +125NC, unless otherwise noted.) (Note 2) PARAMETER SYMBOL MAX31850R Thermocouple Temperature Gain and Offset Error (10.506FV/NC nominal sensitivity) (Note 11) MAX31850S Thermocouple Temperature Gain and Offset Error (9.587FV/NC nominal sensitivity) (Note 11) CONDITIONS MIN +2 TTHERMOCOUPLE = -50NC to +700NC, TA = -20NC to +85NC (Note 12) -3 +3 TTHERMOCOUPLE = +700NC to +1768NC, TA = -20NC to +85NC (Note 12) -5 +5 TTHERMOCOUPLE = -50NC to +1768NC, TA = -40NC to +125NC (Note 12) -7 +7 TTHERMOCOUPLE = -50NC to +150NC, TA = 0NC to +70NC (Note 12) -2 +2 TTHERMOCOUPLE = -50NC to +700NC, TA = -20NC to +85NC (Note 12) -3 +3 TTHERMOCOUPLE = +700NC to +1768NC, TA = -20NC to +85NC (Note 12) -5 +5 TTHERMOCOUPLE = -50NC to +1768NC, TA = -40NC to +125NC (Note 12) -7 +7 Thermocouple Temperature Data Long-Term Drift Hot junction temperature = +400NC Internal Cold-Junction Temperature Error TA = -40NC to +100NC (Note 13) Cold-Junction Temperature Data Resolution TA = -40NC to +125NC Maxim Integrated tCONV MAX -2 (Note 14) UNITS NC NC Thermocouple Temperature Data Resolution Temperature Conversion Time (Thermocouple, Cold Junction, Fault Detection) TYP TTHERMOCOUPLE = -50NC to +100NC, TA = 0NC to +70NC (Note 12) 0.25 NC ±0.24 NC -2 +2 0.0625 72 NC NC 100 ms 4 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters 1-Wire TIMING CHARACTERISTICS (3.0V P VDD P 3.6V, TA = -40NC to +125NC, unless otherwise noted.) (Notes 2, 3) PARAMETER SYMBOL CONDITIONS MIN TYP MAX UNITS 8 µs 120 µs Time to Strong Pullup On tSPON Start Convert T command issued Time Slot tSLOT (Note 15) 60 Recovery Time tREC (Note 15) 1 Write-0 Low Time tLOW0 (Note 15) 60 120 µs Write-1 Low Time tLOW1 (Note 15) 1 15 µs Read Data Valid tRDV (Note 15) 15 µs Reset Time High tRSTH (Note 15) 480 Reset Time Low tRSTL (Notes 15, 16) 480 tPDHIGH (Note 15) 15 60 Presence Detect Low tPDLOW (Note 15) 60 240 µs Capacitance: DQ CIN/OUT (Note 17) 25 pF Capacitance: AD0–AD3 CIN_ADD (Note 17) 50 pF Presence Detect High µs µs µs µs Note 2: Limits are 100% production tested at TA = +25NC. Limits over the operating temperature range and relevant supply voltage range are guaranteed by design and characterization. Note 3: Limits are 100% production tested at TA = +25NC and +85NC. Limits over the operating temperature range and relevant supply voltage are guaranteed by design and characterization. Note 4: All voltages are referenced to GND. Currents entering the IC are specified positive and currents exiting the IC are negative. Note 5: The pullup supply voltage specification assumes that the pullup device is ideal, and therefore the high level of the pullup is equal to VPU. To meet the device’s VIH specification, the actual supply rail for the strong pullup transistor must include margin for the voltage drop across the transistor when it is turned on. Thus: VPU_ACTUAL = VPU_IDEAL + VTRANSISTOR. Note 6: To guarantee a presence pulse under low-voltage parasite power conditions, VILMAX, may have to be reduced to as low as 0.5V. Note 7: Standby current specified up to +70NC. Note 8: To minimize IDDS, DQ should be within the following ranges: VGND P VDQ P VGND + 0.3V or VDD - 0.3V P VDQ P VDD. Note 9: Active current refers to supply current during active temperature conversions. Note 10:DQ is high (high-impedance state with external pullup). Note 11:Not including cold-junction temperature error or thermocouple nonlinearity. Note 12:Guaranteed by design. These limits represent six sigma distribution for TA = +25NC to +85NC. Outside this temperature range, these limits are three sigma distribution. Note 13:Guaranteed by design. These limits represent a three sigma distribution. Note 14:After minimum VDD has been reached during power-up, wait 10ms before initiating temperature conversions. Note 15:See the 1-Wire Timing Diagrams. Note 16:Under parasite power, if tRSTL > 960Fs, a power-on reset (POR) may occur. Note 17:Represents the maximum capacitive load that may be applied to the pins and still maintain timing and logic state. Maxim Integrated 5 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters 1-Wire Timing Diagrams 1-Wire WRITE-ZERO TIME SLOT START OF NEXT CYCLE tSLOT tREC tLOW0 1-Wire READ-ZERO TIME SLOT tSLOT START OF NEXT CYCLE tREC tRDV 1-Wire RESET PULSE RESET PULSE FROM HOST tRSTL tRSTH 1-Wire PRESENCE DETECT PRESENCE DETECT tPDHIGH tPDLOW Maxim Integrated 6 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters Typical Operating Characteristics (VCC = +3.3V, TA = +25NC, unless otherwise noted.) 800 600 400 200 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0 20 40 60 80 -40 -20 100 120 0 20 40 60 80 -0.2 -40 -20 TA = +125°C TA = -40°C 0.5 TA = +85°C 0 0 20 40 60 80 100 120 TEMPERATURE (°C) 0.20 INTERNAL TEMPERATURE = +25°C 0.15 0.10 ADC ACCURACY (°C) ADC ACCURACY (°C) -0.1 ADC ACCURACY vs. ADC INPUT VOLTAGE ACROSS VCC MAX31850 toc04 VCC = 3.3V 1.5 0.05 VCC = 3.7V 0 -0.05 VCC = 3.3V -0.10 -0.15 VCC = 3.0V -0.20 -0.25 TA = +25°C -0.5 -0.30 0 20 40 ADC INPUT VOLTAGE (mV) Maxim Integrated 0 100 120 ADC ACCURACY vs. ADC INPUT VOLTAGE ACROSS TEMPERATURE 1.0 0.1 TEMPERATURE (°C) TEMPERATURE (°C) 2.0 0.2 -0.4 0 -40 -20 0.3 -0.3 0.05 0 VCC = 3.3V 0.4 MAX31850 toc05 1000 0.5 MEASUREMENT ERROR (°C) 1200 VCC = 3.7V 0.45 STANDBY SUPPLY CURRENT (nA) VCC = 3.7V INTERNAL TEMPERATURE SENSOR ACCURACY MAX31850 toc02 0.50 MAX31850 toc01 ACTIVE SUPPLY CURRENT (µA) 1400 STANDBY SUPPLY CURRENT vs. TEMPERATURE MAX31850 toc03 ACTIVE SUPPLY CURRENT vs. TEMPERATURE 60 0 20 40 60 ADC INPUT VOLTAGE (mV) 7 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters Pin Configuration TOP VIEW DNC AD3 AD2 AD1 AD0 10 9 8 7 6 MAX31850 MAX31851 EP + 1 2 3 4 5 GND T- T+ VDD DQ TDFN-EP (3mm x 4mm) Pin Description PIN NAME 1 GND 2 T- Thermocouple Input. See Table 1. Do not connect to GND. 3 T+ Thermocouple Input. See Table 1. 4 VDD Power-Supply Voltage 5 DQ Data Input/Output. Open-drain 1-Wire interface pin. Also provides power to the device when used in parasite-power mode (see the Parasite Power section.) 6 AD0 Location Address Input (Least Significant Bit) 7 AD1 Location Address Input 8 AD2 Location Address Input 9 AD3 Location Address Input (Most Significant Bit) 10 DNC Do Not Connect — EP Maxim Integrated FUNCTION Ground Exposed Pad. No internal connection. Connect to GND or leave unconnected. 8 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters Detailed Description the introduction of noise errors from the thermocouple wires. The MAX31850/MAX31851 are sophisticated thermocouple-to-digital converters with a built-in 14-bit analogto-digital converter (ADC), cold-junction compensation sensing and correction, a digital controller, a 1-Wire data interface, and associated control logic. The devices are available in several versions, each optimized and trimmed for a specific thermocouple type (K, J, N, T, S, R, or E.). The thermocouple type is indicated in the suffix of the part number (e.g., MAX31850K). See the Ordering Information table for all options. Before converting the thermoelectric voltages into equivalent temperature values, it is necessary to compensate for the difference between the thermocouple cold-junction side (device ambient temperature) and a 0NC virtual reference. For a K-type thermocouple, the voltage changes by approximately 41FV/NC, which approximates the thermocouple characteristic with the following linear equation: The 1-Wire bus by definition requires only one data line (and ground) for communication with a central microcontroller. The data line requires a weak pullup resistor since all devices are linked to the bus through a three-state or open-drain port (i.e., the DQ pin). Four location address inputs simplify mapping of individual devices to specific locations. Each device has a unique 64-bit serial code, allowing multiple devices to function on the same 1-Wire bus. Therefore, it is simple to use one microcontroller to control many devices distributed over a large area. In this bus system, the microcontroller identifies and addresses devices on the bus using each device’s unique 64-bit code. Because each device has a unique code, the number of devices that can be addressed on one bus is virtually unlimited. The 1-Wire bus protocol, including detailed explanations of the commands and time slots, is described in the 1-Wire Bus System section. The scratchpad memory contains the 2-byte temperature register that stores the cold-junction-compensated thermocouple temperature data. A second 2-byte register stores the local cold-junction temperature. Both of these registers also store fault data for open thermocouple as well as shorts to supply and ground. Power may be obtained either from a power supply connected to VDD, or from the 1-Wire pullup resistor through the DQ pin when the bus is high. The high bus signal also charges an internal capacitor (CPP), which then supplies power to the device when the bus is low. This method of deriving power from the 1-Wire bus is referred to as “parasite power.” Temperature Conversion The devices include signal-conditioning hardware to convert the thermocouple’s signal into a voltage compatible with the input channels of the ADC. The T+ and T- inputs connect to internal circuitry that reduces Maxim Integrated VOUT = (41.276FV/NC) x (TR - TAMB) where VOUT is the thermocouple output voltage (FV), TR is the temperature of the remote thermocouple junction (NC), and TAMB is the temperature of the device (NC). Other thermocouple types use a similar straight-line approximation but with different gain terms. Note that the MAX31850/MAX31851 assume a linear relationship between temperature and voltage. Because all thermocouples exhibit some level of nonlinearity, apply appropriate correction to the device’s output data. Cold-Junction Compensation The function of the thermocouple is to sense a difference in temperature between two ends of the thermocouple wires. The thermocouple’s “hot” junction can be read across the operating temperature range (Table 1). The reference junction, or “cold” end (which should be at the same temperature as the board on which the device is mounted) can range from -55NC to +125NC. While the temperature at the cold end fluctuates, the device continues to accurately sense the temperature difference at the opposite end. The device senses and corrects for the changes in the reference junction temperature with cold-junction compensation. It does this by first measuring its internal die temperature, which should be held at the same temperature as the reference junction. It then measures the voltage from the thermocouple’s output at the reference junction and converts this to the noncompensated thermocouple temperature value. This value is then added to the device’s die temperature to calculate the thermocouple’s “hot junction” temperature. Note that the “hot junction” temperature can be lower than the cold junction (or reference junction) temperature. Optimal performance from the device is achieved when the thermocouple cold junction and the device are at the same temperature. Avoid placing heat-generating devices or components near the MAX31850/MAX31851 because this could produce cold-junction-related errors. 9 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters Table 1. Thermocouple Wire Connections and Nominal Sensitivities TYPE T- WIRE T+ WIRE TEMP RANGE (°C) SENSITIVITY (µV/°C) COLD-JUNCTION SENSITIVITY (µV/°C) (0NC TO +70NC) K Alumel Chromel -270 to +1372 41.276 (0NC to +1000NC) 40.73 J Constantan Iron -210 to +1200 57.953 (0NC to +750NC) 52.136 N Nisil Nicrosil -270 to +1300 36.256 (0NC to +1000NC) 27.171 R Platinum Platinum/Rhodium -50 to +1768 10.506 (0NC to +1000NC) 6.158 S Platinum Platinum/Rhodium -50 to +1768 9.587 (0NC to +1000NC) 6.181 T Constantan Copper -270 to +400 52.18 (0NC to +400NC) 41.56 E Constantan Chromel -270 to +1000 76.373 (0NC to +1000NC) 44.123 Conversion Functions During the conversion time, tCONV, three functions are performed: the temperature conversion of the internal cold-junction temperature, the temperature conversion of the external thermocouple, and the detection of thermocouple faults. When executing the temperature conversion for the internal cold-junction compensation circuit, the connection to signal from the external thermocouple is opened (switch S4) and the connection to the cold-junction compensation circuit is closed (switch S5). The internal T- reference to ground is still maintained (switch S3 is closed) and the connections to the fault-detection circuit are open (switches S1 and S2). When executing the temperature conversion of the external thermocouple, the connections to the internal fault-detection circuit are opened (switches S1 and S2 in the Block Diagram) and the switch connecting the coldjunction compensation circuit is opened (switch S5). The internal ground reference connection (switch S3) and the connection to the ADC (switch S4) are closed. This allows the ADC to process the voltage detected across Maxim Integrated the T+ and T- terminals. If T+ and T- are unconnected, the thermocouple temperature sign bit is 0 (MSbit of Scratchpad Byte 1), and the remainder of the thermocouple temperature value is 1. During fault detection, the connections from the external thermocouple and cold-junction compensation circuit to the ADC are opened (switches S4 and S5, see the Block Diagram). The internal ground reference on T- is also opened (switch S3). The connections to the internal fault-detection circuit are closed (switch S1 and S2). The fault-detection circuit tests for shorted connections to VDD or GND on the T+ and T- inputs, as well as looking for an open thermocouple condition. Bits 0, 1, and 2 of the internal (cold junction) temperature data are normally low. Bit 2 goes high to indicate a thermocouple short to VDD, bit 1 goes high to indicate a thermocouple short to GND, and bit 0 goes high to indicate a thermocouple open circuit. If any of these conditions exists, bit 0 of the cold-junction compensated thermacouple temperature data, which is normally low, also goes high to indicate that a fault has occurred. 10 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters VPU MAX31850 GND VPU µP DQ VDD 4.7kΩ 1-Wire BUS TO OTHER 1-Wire DEVICES Figure 1. Supplying the Parasite-Powered MAX31850/MAX31851 During Temperature Conversions MAX31850 GND VPU µP DQ VDD (EXTERNAL SUPPLY) VDD 4.7kΩ 1-Wire BUS TO OTHER 1-Wire DEVICES Figure 2. Powering the MAX31850/MAX31851 with an External Supply Powering the MAX31850/MAX31851 The MAX31850/MAX31851 can be powered by an external supply on the VDD pin, or they can operate in “parasite power” mode, which allows the device to function without a local external supply. Parasite power is useful for applications that require remote temperature sensing or those that are very space-constrained. Figure 1 shows the device’s parasite-power control circuitry, which “steals” power from the 1-Wire bus through DQ when the bus is high. The stolen charge powers the device while the bus is high, and some of the charge is stored on the internal parasite-power capacitor (CPP) to provide power when the bus is low. When the device is used in parasite-power mode, VDD must be connected to ground. In parasite-power mode, the 1-Wire bus and CPP can provide sufficient current to the device for most operations as long as the specified timing and voltage requirements are met (see the DC Electrical Characteristics and the 1-Wire Timing Characteristics tables). However, when the device is performing temperature conversions, the operating current can be as high as 1.5mA. This current can cause an unacceptable voltage drop across the weak 1-Wire pullup resistor and is more current than Maxim Integrated can be supplied by CPP. To ensure that the device has sufficient supply current, it is necessary to provide a strong pullup on the 1-Wire bus whenever temperature conversions are taking place. This can be accomplished by using a MOSFET to pull the bus directly to the rail as shown in Figure 1. The 1-Wire bus must be switched to the strong pullup within 10Fs (max) after a Convert T [44h] command is issued, and the bus must be held high by the pullup for the duration of the conversion (tCONV). No other activity can take place on the 1-Wire bus while the pullup is enabled. The device can also be powered by the conventional method of connecting an external power supply to VDD, as shown in Figure 2. The advantage of this method is that the MOSFET pullup is not required, and the 1-Wire bus is free to carry other traffic during the temperature conversion period. The use of parasite power is not recommended for temperatures above 100NC because the device may not be able to sustain communications due to the higher leakage currents that can exist at these temperatures. For applications in which such temperatures are likely, it is strongly recommended that the device be powered by an external power supply. 11 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters MSb LSb 8-BIT CRC CODE MSb 8-BIT FAMILY CODE (3Bh) 48-BIT SERIAL NUMBER LSb MSb LSb MSb LSb Figure 3. 64-Bit ROM Code SCRATCHPAD (POWER-UP STATE SHOWN IN PARENTHESES) BYTE 0 COLD-JUNCTION-COMPENSATED THERMOCOUPLE TEMPERATURE LSB AND FAULT STATUS (00h) BYTE 1 COLD-JUNCTION-COMPENSATED THERMOCOUPLE TEMPERATURE MSB (00h) BYTE 2 INTERNAL (COLD JUNCTION) TEMPERATURE AND FAULT STATUS LSB (00h) BYTE 3 INTERNAL (COLD JUNCTION) TEMPERATURE MSB (00h) BYTE 4 CONFIGURATION REGISTER* BYTE 5 RESERVED (FFh) BYTE 6 RESERVED (FFh) BYTE 7 RESERVED (FFh) BYTE 8 CRC *THE LOWER 4 BITS (AD[3:0]) OF THE CONFIGURATION REGISTER ARE HARDWIRED THROUGH AD0–AD3. Figure 4. Memory Map In some situations the bus master might not know whether the devices on the bus are parasite powered or powered by external supplies. The master needs this information to determine if the strong bus pullup should be used during temperature conversions. To get this information, the master can issue a Skip ROM [CCh] command, followed by a Read Power Supply [B4h] command, followed by a read time slot. During the read time slot, parasite-powered devices pull the bus low, and externally powered devices let the bus remain high. If the bus is pulled low, the master knows that it must supply the strong pullup on the 1-Wire bus during temperature conversions. 64-Bit ROM Code Each device contains a unique 64-bit code stored in ROM (Figure 3). The least significant 8 bits of the ROM code contain the device’s 1-Wire family code, 3Bh. The next 48 bits contain a unique serial number. The most significant 8 bits contain a cyclic redundancy check (CRC) byte that is calculated from the first 56 bits of the ROM code. See the CRC Generation section for a detailed explanation of the CRC bits. The 64-bit ROM code and associated Maxim Integrated ROM function control logic allow the device to operate as a 1-Wire device using the protocol detailed in the 1-Wire Bus System section. Scratchpad The device’s scratchpad is organized as shown in Figure 4. All memory commands are described in detail in the MAX31850/MAX31851 Function Commands section. Byte 0 and byte 1 of Scratchpad contain the least significant byte and the most significant byte of the thermocouple temperature register, respectively. Bytes 2 and 3 contain the LSB and MSB of the internal (cold-junction) temperature value, as well as fault status. Byte 4 contains the configuration information. Bytes 5, 6, and 7 are reserved for internal use by the device and cannot be overwritten; these bytes return all ones when read. Byte 8 of Scratchpad is read-only and contains the CRC code for bytes 0–7 of the scratchpad. The device generates this CRC using the method described in the CRC Generation section. 12 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters Configuration Register Byte 4 of Scratchpad contains the configuration register, which is organized as shown in Configuration Register Format. The configuration register allows the user to read the programmed value of the address pins. The AD[3:0] bits report the pin-programmed location information. Pins connected to DQ are reported with logic 1, and pins connected to GND are reported as logic 0. Pins connected to DQ or GND through a resistor are valid logic 1s or logic 0s if the resistor is less than 10kI. Unconnected or high-impedance ( > 10kI) connections are indeterminate. Bits [7:4] are reserved for internal use and cannot be overwritten; they return a 1 when read. Configuration Register Format BIT 7 BIT 6 BIT 5 BIT 4 BIT 3 BIT 2 BIT 1 BIT 0 — — — — AD3 AD2 AD1 AD0 Note: Bits [3:0] are programmed through the four location programming address pins, AD[3:0]. Reading the configuration register provides location information on up to 16 individual devices. Table 2. Temperature Data Format Cold-Junction-Compensated Thermocouple Temperature Data (Bytes 0 and 1) BIT 7 6 5 4 3 2 1 0 LSByte (Scratchpad Byte 0) 23 22 21 20 2-1 2-2 Reserved 1 = Fault BIT 15 14 13 12 11 10 9 8 MSByte (Scratchpad Byte 1) Sign 210 29 28 27 26 25 24 4 3 2 1 0 1 = Short to GND 1 = Open Circuit Internal (Cold-Junction) Temperature Data (Bytes 2 and 3) BIT 7 6 5 LSByte (Scratchpad Byte 2) 2-1 2-2 2-3 2-4 Reserved 1 = Short to VDD BIT 15 14 13 12 11 10 9 8 MSByte (Scratchpad Byte 3) Sign 26 25 24 23 22 21 20 Table 3. Thermocouple Temperature Data Format Table 4. Internal (Cold-Junction) Temperature Data Format TEMPERATURE (NC) DIGITAL OUTPUT (D[31:18]) TEMPERATURE (NC) DIGITAL OUTPUT (D[15:4]) +1600.00 0110 0100 0000 00 +127.0000 0111 1111 0000 +1000.00 0011 1110 1000 00 +100.5625 0110 0100 1001 +100.75 0000 0110 0100 11 +25.0000 0001 1001 0000 +25.00 0000 0001 1001 00 0.0000 0000 0000 0000 0.00 0000 0000 0000 00 -0.0625 1111 1111 1111 -0.25 1111 1111 1111 11 -1.0000 1111 1111 0000 -1.00 1111 1111 1111 00 -20.0000 1110 1100 0000 -250.00 1111 0000 0110 00 -55.0000 1100 1001 0000 Maxim Integrated 13 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters CRC Generation shift register. After shifting in the 56th bit from the ROM or the most significant bit of byte 7 from the scratchpad, the polynomial generator contains the recalculated CRC. Next, the 8-bit ROM code or scratchpad CRC from the device must be shifted into the circuit. At this point, if the recalculated CRC was correct, the shift register contains all zeros. Additional information about the Maxim 1-Wire CRC is available in Application Note 27: Understanding and Using Cyclic Redundancy Checks with Maxim iButton® Products. CRC bytes are provided as part of the device’s 64-bit ROM code, in the 9th byte of Scratchpad. The ROM code CRC is calculated from the first 56 bits of the ROM code and is contained in the most significant byte of the ROM. The scratchpad CRC is calculated from the data in the scratchpad, and therefore changes when the data in it associated scratchpad changes. The CRC provides the bus master with a method of data validation when data is read from the device. To verify that data has been read correctly, the bus master must recalculate the CRC from the received data and then compare this value to either the ROM code CRC (for ROM reads) or to the scratchpad CRC (for scratchpad reads). If the calculated CRC matches the read CRC, the data has been received error-free. The comparison of CRC values and the decision to continue with an operation are determined entirely by the bus master. There is no circuitry inside the device that prevents a command sequence from proceeding if the device CRC (ROM or scratchpad) does not match the value generated by the bus master. 1-Wire Bus System The 1-Wire bus system uses a single bus master to control one or more slave devices. The MAX31850/ MAX31851 are always a slave. When there is only one slave on the bus, the system is referred to as a singledrop system; the system is multidrop if there are multiple slaves on the bus. All data and commands are transmitted least significant bit first over the 1-Wire bus. The following discussion of the 1-Wire bus system is broken down into three topics: hardware configuration, transaction sequence, and 1-Wire signaling (signal types and timing). The equivalent polynomial function of the CRC (ROM or scratchpad) is: Hardware Configuration CRC = X8 + X5 + X4 + 1 The 1-Wire bus has by definition only a single data line. Each device (master or slave) interfaces to the data line by using an open-drain or three-state port. This allows each device to “release” the data line when the device is not transmitting data so the bus is available for use by another device. The device’s 1-Wire port (DQ) is open drain with an internal circuit equivalent to that shown in Figure 6. The bus master can recalculate the CRC and compare it to the CRC values from the device using the polynomial generator shown in Figure 5. This circuit consists of a shift register and XOR gates, and the shift register bits are initialized to 0. Starting with the least significant bit of the ROM code or the least significant bit of byte 0 in the scratchpad, one bit at a time should shifted into the POLYNOMIAL = X8 + X5 + X4 + 1 1ST STAGE X0 2ND STAGE X1 3RD STAGE X2 4TH STAGE X3 5TH STAGE X4 6TH STAGE X5 7TH STAGE X6 8TH STAGE X7 X8 INPUT DATA Figure 5. CRC Generator iButton is a registered trademark of Maxim Integrated Products, Inc. Maxim Integrated 14 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters VPU BUS MASTER MAX31850/MAX31851 1-Wire PORT 4.7kΩ DQ Rx Tx Rx = RECEIVE Tx = TRANSMIT OPEN-DRAIN PORT PIN Rx 5µA TYP Tx 100Ω MOSFET Figure 6. Hardware Configuration The 1-Wire bus requires an external pullup resistor of approximately 5kI; thus, the idle state for the 1-Wire bus is high. If for any reason a transaction needs to be suspended, the bus must be left in the idle state if the transaction is to resume. Infinite recovery time can occur between bits so long as the 1-Wire bus is in the inactive (high) state during the recovery period. If the bus is held low for more than 480Fs, all components on the bus are reset. Transaction Sequence The transaction sequence for accessing the device is as follows: 1) Step 1: Initialization 2)Step 2: ROM Command (followed by any required data exchange) 3)Step 3: MAX31850/MAX31851 Function Command (followed by any required data exchange) It is very important to follow this sequence every time the device is accessed, as the device does not respond if any steps in the sequence are missing or out of order. An exception to this rule is the Search ROM [F0h] command. After issuing this ROM command, the master must return to step 1 in the sequence. Initialization All transactions on the 1-Wire bus begin with an initialization sequence. The initialization sequence consists of a reset pulse transmitted by the bus master followed by presence pulse(s) transmitted by the slave(s). The presence pulse lets the bus master know that slave devices (such as the MAX31850/MAX31851) are on the bus and Maxim Integrated are ready to operate. Timing for the reset and presence pulses is detailed in the 1-Wire Signaling section. ROM Commands After the bus master has detected a presence pulse, it can issue a ROM command. These commands operate on the unique 64-bit ROM codes of each slave device and allow the master to single out a specific device if many are present on the 1-Wire bus. These commands also allow the master to determine how many and what types of devices are present on the bus. There are four ROM commands, and each command is 8 bits long. The master device must issue an appropriate ROM command before issuing a MAX31850/MAX31851 function command. Figure 7 shows a flowchart for operation of the ROM commands. Search ROM [F0h] When a system is initially powered up, the master must identify the ROM codes of all slave devices on the bus, which allows the master to determine the number of slaves and their device types. The master learns the ROM codes through a process of elimination that requires the master to perform a Search ROM cycle (i.e., Search ROM command followed by data exchange) as many times as necessary to identify all the slave devices. If there is only one slave on the bus, the simpler Read ROM command can be used in place of the Search ROM process. For a detailed explanation of the Search ROM command procedure, refer to Application Note 937: Book of iButton® Standards. After every Search ROM cycle, the bus master must return to step 1 (initialization) in the transaction sequence. 15 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters Read ROM [33h] This command can be used only when there is one slave on the bus. It allows the bus master to read the slave’s 64-bit ROM code without using the Search ROM command procedure. If this command is used when there is more than one slave present on the bus, a data collision occurs when all the slaves attempt to respond at the same time. Match ROM [55h] The Match ROM command followed by a 64-bit ROM code sequence allows the bus master to address a specific slave device on a multidrop or single-drop bus. Only the slave that exactly matches the 64-bit ROM code sequence responds to the function command issued by the master; all other slaves on the bus wait for a reset pulse. Skip ROM [CCh] The master can use this command to address all devices on the bus simultaneously without sending out any ROM code information. For example, the master can make all devices on the bus perform simultaneous temperature conversions by issuing a Skip ROM command followed by a Convert T [44h] function command. Note that the Read Scratchpad command can follow the Skip ROM command only if there is a single slave device on the bus. In this case, time is saved by allowing the master to read from the slave without sending the device’s 64-bit ROM code. A Skip ROM command followed by a Read Scratchpad command causes a data collision on the bus if there is more than one slave because multiple devices attempt to transmit data simultaneously. MAX31850/MAX31851 Function Commands After the bus master has used a ROM command to address the MAX31850/MAX31851 with which it wishes to communicate, the master can issue one of the MAX31850/MAX31851 function commands. These commands allow the master to read from the device’s scratchpad memories, initiate temperature conversions, and determine the power-supply mode. The MAX31850/ MAX31851 function commands are summarized in Table 5 and illustrated by the flowchart in Figure 8. Convert T [44h] This command initiates a single thermocouple temperature conversion, which consists of measuring the internal (cold junction) temperature, measuring the thermocouple temperature, and detecting any faults. Following the conversion, the resulting cold-junction-compensated thermocouple data, internal temperature data, and fault status are stored in the two 2-byte temperature registers in the scratchpad memory, and the MAX31850 returns to its lowpower idle state. If the device is being used in parasitepower mode, within 10Fs (max) after this command is issued the master must enable a strong pullup on the 1-Wire bus for the duration of the conversion (tCONV) as described in the Powering the MAX31850/MAX31851 section. If the device is powered by an external supply, the master can issue read time slots after the Convert T command and the device responds by transmitting 0 while the temperature conversion is in progress and 1 when the conversion is done. In parasite-power mode, this notification technique cannot be used because the bus is pulled high by the strong pullup during the conversion. Table 5. MAX31850/MAX31851 Function Command Set COMMAND Convert T (Note 1) Read Scratchpad (Note 2) Read Power Supply (Note 3) DESCRIPTION PROTOCOL 1-Wire BUS ACTIVITY AFTER COMMAND IS ISSUED Initiates temperature conversion. 44h The device transmits conversion status to master (not applicable for parasite-powered devices). Reads the 9-byte scratchpad including the CRC byte. BEh The device transmits up to 9 data bytes to master. The 9th byte is the CRC byte. Signals the device’s power-supply mode to the master. B4h The device transmits supply status to the master. Note 1: For parasite-powered devices, the master must enable a strong pullup on the 1-Wire bus during temperature conversions. No other bus activity can take place during this time. Note 2: The master can interrupt the transmission of data at any time by issuing a reset. Note 3: During the read time slot, parasite-powered devices pull the DQ bus low, and externally powered units let the bus remain high. Maxim Integrated 16 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters MASTER Tx RESET PULSE INITIALIZATION SEQUENCE DEVICE Tx PRESENCE PULSE MASTER Tx ROM COMMAND 33h READ ROM? Y N 55h MATCH ROM? F0h SEARCH ROM? N Y Y DEVICE Tx SERIAL NUMBER 6 BYTES DEVICE Tx CRC BYTE N Y DEVICE Tx BIT 0 MASTER Tx BIT 0 N N BIT 0 MATCH? Y DEVICE Tx FAMILY CODE 1 BYTE CCh SKIP ROM? DEVICE Tx BIT 0 MASTER Tx BIT 0 BIT 0 MATCH? N Y DEVICE Tx BIT 1 MASTER Tx BIT 1 BIT 1 MATCH? DEVICE Tx BIT 1 MASTER Tx BIT 1 N N BIT 1 MATCH? Y Y DEVICE Tx BIT 63 MASTER Tx BIT 63 BIT 63 MATCH? DEVICE Tx BIT 63 MASTER Tx BIT 63 N N BIT 63 MATCH? Y Y MASTER Tx FUNCTION COMMAND Figure 7. ROM Commands Flowchart Maxim Integrated 17 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters 44h CONVERT T? MASTER Tx FUNCTION COMMAND N Y N DEVICE BEGINS CONVERSION MASTER ENABLES STRONG PULLUP ON DQ DEVICE CONVERTING TEMPERATURE? DEVICE CONVERTS TEMPERATURE N Y MASTER Rx “0s” N Y PARASITE POWER? MASTER Rx “1s” B4h READ POWER SUPPLY? MASTER DISABLES STRONG PULLUP N Y N MASTER Rx “1s” PARASITE POWER? BEh READ SCRATCHPAD? Y Y MASTER Rx “0s” MASTER Rx DATA BYTE FROM SCRATCHPAD MASTER Tx RESET? Y N N HAVE 8 BYTES BEEN READ? Y MASTER Rx SCRATCHPAD CRC BYTE RETURN TO INITIALIZATION SEQUENCE FOR NEXT TRANSACTION Figure 8. MAX31850/MAX31851 Function Commands Flowchart Maxim Integrated 18 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters SEARCH ALL ROM IDs ON BUS AND STORE ROM IDs (F0h COMMAND) INCREMENT COUNTER N=N+1 N > NMAX? N BUILDING CROSS-REFERENCE TABLE USING ROM IDs AND 4-BIT ADDRESSES Y DONE NMAX IS THE NUMBER OF ROM IDs FOUND MASTER Tx NEXT ROM ID READ SCRATCHPAD (USE AD3−AD0 FROM CONFIG REGISTER) MATCH ROM ID TO ADDRESS AND ADD TO CROSS-REFERENCE TABLE CROSS-REFERENCE TABLE ROM ID ROM ID(0) ROM ID(1) ROM ID(2) ROM ID(3) AD3−AD0 0000 0001 0010 0011 ROM ID(12) ROM ID(13) ROM ID(14) ROM ID(15) 1100 1101 1110 1111 NOTE: TEMPERATURE SENSORS ARE ADDRESSED BY ROM ID, NOT BY BINARY ADDRESS. Figure 9. Building a Cross-Reference Table Read Scratchpad [BEh] This command allows the master to read the contents of Scratchpad. The data transfer starts with the least significant bit of byte 0 and continues through the scratchpad until the 9th byte (byte 8, CRC) is read. The master can issue a reset to terminate reading at any time if only part of the scratchpad data is needed. The CRC is computed while data is read from bytes 0–7, and is shifted out as byte 8. Read Power Supply [B4h] The master device issues this command followed by a read time slot to determine if any devices on the bus are using parasite power. During the read time slot, parasite-powered devices pull the bus low, and externally powered devices do not pull the bus low. See the Powering the MAX31850/ MAX31851 section for more information. Building a Cross-Reference Table The procedure in Figure 9 uses the Search ROM command to find all MAX31850/MAX31851s on the 1-Wire bus (16 maximum) and then reads each configuration register to match the ROM IDs to the hardwired addresses. Maxim Integrated 1-Wire Signaling The device uses a strict 1-Wire communication protocol to ensure data integrity. Several signal types are defined by this protocol: reset pulse, presence pulse, write-zero, write-one, read-zero, and read-one. The bus master initiates all these signals except the presence pulse. All MAX31850/MAX31851s on the bus are slaves. Initialization Procedure: Reset and Presence Pulses All communication with the device begins with an initialization sequence that consists of a reset pulse from the master followed by a presence pulse from the device. This is illustrated in Figure 10. When the device sends the presence pulse in response to the reset, it is indicating to the master that it is on the bus and ready to operate. During the initialization sequence, the bus master transmits (Tx) the reset pulse by pulling the 1-Wire bus low for 480Fs (min). The bus master then releases the bus and goes into receive mode (Rx). When the bus is released, the 5kI pullup resistor pulls the 1-Wire bus high. When the device detects this rising edge, it waits 15Fs to 60Fs 19 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters MASTER Tx RESET PULSE 480µs MINIMUM VPU MAX31850/MAX31851 WAITS 15µs TO 60µs MASTER Rx 480µs MINIMUM MAX31850/MAX31851 Tx PRESENCE PULSE 60µs TO 240µs 1-Wire BUS GND BUS MASTER PULLING LOW MAX31850/MAX31851 PULLING LOW RESISTOR PULLUP Figure 10. Initialization Timing and then transmits a presence pulse by pulling the 1-Wire bus low for 60Fs to 240Fs. Read/Write Time Slots The bus master writes data to the device during write time slots and reads data from the device during read time slots. One bit of data is transmitted over the 1-Wire bus per time slot. Write Time Slots There are two types of write time slots: write-one time slots and write-zero time slots. The bus master uses a write-one time slot to write a logic 1 to the slave and a write-zero time slot to write a logic 0 to the slave. All write time slots must have a 60Fs (min) duration with a 1Fs (min) recovery time between individual write slots. Both types of write time slots are initiated by the master pulling the 1-Wire bus low (Figure 11). To generate a write-one time slot, after pulling the 1-Wire bus low, the bus master must release the 1-Wire bus within 15Fs. When the bus is released, the 5kI pullup resistor pulls the bus high. To generate a write-zero time slot, after pulling the 1-Wire bus low, the bus master must continue to hold the bus low for the duration of the time slot (at least 60Fs). The slave samples the 1-Wire bus during a window that lasts from 15Fs to 60Fs after the master initiates the write time slot. If the bus is high during the sampling window, a 1 is written to the slave. If the line is low, a 0 is written to the device. Maxim Integrated Read Time Slots The slave can only transmit data to the master when the master issues read time slots. Therefore, the master must generate read time slots immediately after issuing a Read Scratchpad [BEh] command or Read Power Supply [B4h] command, so that the device can provide the requested data. In addition, the master can generate read time slots after issuing a Convert T [44h] command to verify the operation status as explained in the MAX31850/MAX31851 Function Commands section. All read time slots must be 60Fs (min) in duration with a 1Fs (min) recovery time between slots. A read time slot is initiated by the master device pulling the 1-Wire bus low for a minimum of 1Fs (tINIT) and then releasing the bus (Figure 11). After the master initiates the read time slot, the slave begins transmitting a 1 or 0 on bus. The slave transmits a 1 by leaving the bus high and transmits a 0 by pulling the bus low. When transmitting a 0, the slave releases the bus by the end of the time slot, and the pullup resistor pulls the bus back to its high idle state. Output data from the slave is valid for 15Fs after the falling edge that initiated the read time slot. Therefore, the master must release the bus and then sample the bus state within 15Fs from the start of the slot. Figure 12 illustrates that the sum of tINIT, tRC, and the master sample window must be less than 15Fs for a read time slot. tRC is the rise time due to the resistive and capacitive characteristics of the bus. Figure 13 shows that system timing margin is maximized by keeping tINIT and tRC as short as possible and by locating the master sample time during read time slots towards the end of the 15Fs period. 20 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters START OF SLOT START OF SLOT MASTER WRITE-ONE SLOT MASTER WRITE-ZERO SLOT 1µs < tREC < ∞ 60µs < Tx “0” < 120µs > 1µs VPU 1-Wire BUS GND MAX31850/MAX31851 SAMPLES MAX31850/MAX31851 SAMPLES MIN 15µs TYP 15µs MIN MAX 30µs 15µs MASTER READ-ZERO SLOT TYP 15µs MAX 30µs MASTER READ-ONE SLOT 1µs < tREC < ∞ VPU 1-Wire BUS GND MASTER SAMPLES > 1µs MASTER SAMPLES > 1µs 15µs 45µs BUS MASTER PULLING LOW 15µs MAX31850/MAX31851PULLING LOW RESISTOR PULLUP Figure 11. Read/Write Time Slot Timing Diagram VPU VIH OF MASTER 1-Wire BUS GND tINIT > 1µs tRC MASTER SAMPLES 15µs BUS MASTER PULLING LOW RESISTOR PULLUP Figure 12. Detailed Master Read-One Timing Maxim Integrated 21 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters VPU VIH OF MASTER 1-Wire BUS GND tINIT = SMALL MASTER SAMPLES tRC = SMALL 15µs BUS MASTER PULLING LOW RESISTOR PULLUP Figure 13. Recommended Master Read-One Timing Table 6. Operation Example MASTER MODE DATA (LSB FIRST) Tx Reset Rx Presence Devices respond with presence pulse. Tx F0h Master issues Search ROM command Tx Reset Rx Presence Tx 55h Tx 64-bit ROM code Tx 44h Tx DQ line held high by strong pullup COMMENTS Master issues reset pulse. Master issues reset pulse. Devices respond with presence pulse. Master issues Match ROM command for desired address Master sends device ROM code. Master issues Convert T command. Master applies strong pullup to DQ for the duration of the conversion (tCONV). Tx Reset Rx Presence Master issues reset pulse. Devices respond with presence pulse. Tx 55h Master issues Match ROM command. Tx 64-bit ROM code Tx BEh Rx 9 data bytes Master sends device ROM code. Master issues Read Scratchpad command. Master reads entire scratchpad including CRC. The master then recalculates the CRC of the first 8 data bytes from the scratchpad and compares the calculated CRC with the read CRC (byte 9). If they match, the master continues; if not, the read operation is repeated. Operation Example Table 6 shows an operation example in which there are multiple devices on the bus using parasite power. The Maxim Integrated bus master initiates a temperature conversion in a specific MAX31850/MAX31851 and then reads Scratchpad and recalculates the CRC to verify the data. 22 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters 1-Wire BUS AD0 DQ VDD VDD MAX31850 MAX31851 AD1 AD2 GND AD3 DQ AD0 LOCATION 0 AD0 = GND AD1 = GND AD2 = GND AD3 = GND 1-Wire BUS AD0 DQ VDD MAX31850 MAX31851 AD1 AD2 GND AD3 DQ AD0 LOCATION 0 AD0 = GND AD1 = GND AD2 = GND AD3 = GND VDD VDD MAX31850 MAX31851 AD1 AD2 GND AD3 DQ AD0 LOCATION 1 AD0 = VDD AD1 = GND AD2 = GND AD3 = GND VDD MAX31850 MAX31851 AD1 AD2 GND AD3 DQ AD0 LOCATION 1 AD0 = DQ AD1 = GND AD2 = GND AD3 = GND VDD VDD MAX31850 MAX31851 AD1 AD2 GND AD3 DQ AD0 LOCATION 2 AD0 = GND AD1 = VDD AD2 = GND AD3 = GND VDD MAX31850 MAX31851 AD1 AD2 GND AD3 DQ AD0 LOCATION 2 AD0 = GND AD1 = DQ AD2 = GND AD3 = GND VDD VDD GND MAX31850 MAX31851 AD1 AD2 AD3 LOCATION 15 AD0 = VDD AD1 = VDD AD2 = VDD AD3 = VDD NOTE: AD3–AD0 CANNOT BE LEFT UNCONNECTED; EACH PIN MUST BE CONNECTED TO EITHER VDD OR GND. Figure 14. Address Programming Diagram—VDD Powered Applications Information Open and Shorted Thermocouple Detection The LSB of Byte 0 in Scratchpad is normally low and goes high if the thermocouple input is open or shorted to ground or VDD. Bits 0, 1, and 2 of Byte 2 are normally low. When bit 2 (SCV) is high, it indicates a thermocouple short to VDD. When bit 1 (SCG) is high, it indicates a thermocouple short to ground. When bit 0 (OC) is high, it indicates a thermocouple open circuit. VDD GND MAX31850 MAX31851 AD1 AD2 AD3 LOCATION 15 AD0 = DQ AD1 = DQ AD2 = DQ AD3 = DQ NOTE: AD3–AD0 CANNOT BE LEFT UNCONNECTED; EACH PIN MUST BE CONNECTED TO EITHER DQ OR GND. Figure 15. Address Programming Diagram—Parasite Powered Noise Considerations Because of the small signal levels involved, thermocouple temperature measurement is susceptible to power supply coupled noise. The effects of power-supply noise can be minimized by placing a 0.1FF ceramic bypass capacitor close to the VDD pin of the device and to GND. The input amplifier is a low-noise amplifier designed to enable high-precision input sensing. Keep the thermocouple and connecting wires away from electrical noise sources. It is strongly recommended to add a 10nF ceramic surfacemount differential capacitor, placed across the T+ and T- pins, in order to filter noise on the thermocouple lines. Maxim Integrated 23 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters Thermal Considerations Self-heating degrades the temperature measurement accuracy of the MAX31850/MAX31851 in some applications. The magnitude of the temperature errors depends on the thermal conductivity of the MAX31850 package, the mounting technique, and the effects of airflow. Use a large ground plane to improve the temperature measurement accuracy of the MAX31850. • Avoid mechanical stress and vibration, which could strain the wires. • When using long thermocouple wires, use a twisted pair extension wire. • Avoid steep temperature gradients. • Try to use the thermocouple wire well within its temperature rating. The accuracy of a thermocouple system can also be improved by following these precautions: • Use the proper sheathing material in hostile environments to protect the thermocouple wire. • Use the largest wire possible that does not shunt heat away from the measurement area. • Use extension wire only at low temperatures and only in regions of small gradients. • If a small wire is required, use it only in the region of the measurement, and use extension wire for the region with no temperature gradient. • Keep an event log and a continuous record of thermocouple resistance. Ordering Information PART THERMOCOUPLE TYPE MEASURED TEMP RANGE PIN-PACKAGE MAX31850KATB+ K -270NC to +1372NC 10 TDFN-EP* MAX31850KATB+T K -270NC to +1372NC 10 TDFN-EP* MAX31850JATB+ J -210NC to +1200NC 10 TDFN-EP* MAX31850JATB+T J -210NC to +1200NC 10 TDFN-EP* MAX31850NATB+ N -270NC to +1300NC 10 TDFN-EP* MAX31850NATB+T N -270NC to +1300NC 10 TDFN-EP* MAX31850TATB+ T -270NC to +400NC 10 TDFN-EP* MAX31850TATB+T T -270NC to +400NC 10 TDFN-EP* MAX31850EATB+ E -270NC to +1000NC 10 TDFN-EP* MAX31850EATB+T E -270NC to +1000NC 10 TDFN-EP* MAX31851SATB+ S -270NC to +1768NC 10 TDFN-EP* MAX31851SATB+T S -270NC to +1768NC 10 TDFN-EP* MAX31851RATB+ R -270NC to +1768NC 10 TDFN-EP* MAX31851RATB+T R -270NC to +1768NC 10 TDFN-EP* Note: All devices are specified over the -40°C to +125°C operating temperature range. +Denotes a lead(Pb)-free/RoHS-compliant package. T = Tape and reel. *EP = Exposed pad. Package Information For the latest package outline information and land patterns (footprints), go to www.maximintegrated.com/packages. Note that a “+”, “#”, or “-” in the package code indicates RoHS status only. Package drawings may show a different suffix character, but the drawing pertains to the package regardless of RoHS status. PACKAGE TYPE PACKAGE CODE OUTLINE NO. LAND PATTERN NO. 10 TDFN-EP T1034N+1 21-0268 90-0247 Maxim Integrated 24 MAX31850/MAX31851 Cold-Junction Compensated, 1-Wire Thermocouple-to-Digital Converters Revision History REVISION NUMBER REVISION DATE 0 6/13 DESCRIPTION Initial release PAGES CHANGED — Maxim Integrated cannot assume responsibility for use of any circuitry other than circuitry entirely embodied in a Maxim Integrated product. No circuit patent licenses are implied. Maxim Integrated reserves the right to change the circuitry and specifications without notice at any time. The parametric values (min and max limits) shown in the Electrical Characteristics table are guaranteed. Other parametric values quoted in this data sheet are provided for guidance. Maxim Integrated 160 Rio Robles, San Jose, CA 95134 USA 1-408-601-1000 © 2013 Maxim Integrated Products, Inc. 25 Maxim Integrated and the Maxim Integrated logo are trademarks of Maxim Integrated Products, Inc.