HCX Series - First Sensor

HCX Series
Fully signal conditioned pressure transducer
FEATURES
· Pressure ranges from
±5 mbar to 5 bar differential,
1 and 2 bar absolute
· TTL power supply
· 0.5...4.5 V output
· Inline pinning for easy PCB-mounting
· Externally adjustable offset and span
SERVICE
Non-corrosive, non-ionic working fluids, such as dry
air and dry gases.
SPECIFICATIONS
ELECTRICAL CONNECTION
Maximum ratings
4.8 to 15 V
Excitation voltage
Output current
Source
Sink
Temperature limits
Operating
Storage
Compensated
-20 to 70°C
-40 to 85°C
0 to 50°C
Lead temperature (2...4 sec. soldering)
Humidity
Proof pressure1
HCXPM005..., HCXM010...
and HCXM020
HCXM050 to HCXM350
all other devices
Tightening torque
(mounting screws M2.5)
August 2007 / 026
+Vs
5 mA
5 mA
250°C
0 - 95 %RH
5
Vout
1
HCX(M)... 4
span adj.6
(optional)
3 2 6
V
offset adj.6
(optional)
-Vs
GND
700 mbar
1.4 bar
2 x rated pressure
10 Ncm
1/4
www.first-sensor.com
www.sensortechnics.com
HCX Series
Fully signal conditioned pressure transducer
PERFORMANCE CHARACTERISTICS
(unless otherwise noted, VS = 5 V, RL > 100 kΩ, tamb = 25°C)
Characteristics
Operating pressure
HCXPM005D6...
HCXM010D6...
HCXPM010D6...
HCXM020D6...
HCXM050D6...
HCXM100D6...
HCXM350D6...
HCX001...6...
HCX002...6...
HCX005...6...
all HCXPM...
HCXM010D6... / HCXM020D6...
all other devices
HCXPM...
all other devices
Zero pressure offset
Span5
Full scale output
Output at lowest specified pressure
4
Thermal effects (0 to 50°C)
Combined offset and span
Non-linearity and hysteresis (BSL)2
Long term stability3
Output impedance
Power supply rejection
Power consumption (no load)
HCXPM... only
HCXP...M005D6...
HCXM010D6... to HCXM050D6...
HCXM100D6..., HCXM350D6...
all other devices
HCXM020D6...
all other devices
Min.
-5
0
-10
0
0
0
0
0
0
0
2.40
0.40
0.45
1.95
3.95
Typ.
2.50
0.50
0.50
2.0
4.0
4.5
Max.
5
10
10
20
50
100
350
1000
2000
5000
2.60
0.60
0.55
2.05
4.05
Unit
mbar
V
0.5
0.5
0.1
±0.2
0.20
0.12
0.10
0.05
1.0
0.5
%FSO/°C
%FSO
Ω
50
Offset
Span
0.05
0.03
50
%FSO/V
mW
Specification notes:
1.
2.
3.
4.
5.
6.
Proof pressure is the maximum pressure which may be applied without causing damage to the sensing element.
Non-linearity - the maximum deviation of measured output at constant temperature, from "Best Straight Line" through
three points (offset pressure, full scale pressure and 1/2 full scale pressure).
Change after one year or 1 million pressure cycles.
Thermal effects tested and guaranteed from 0°C to 50°C relative to 25°C. All specifications shown are relative to 25°C.
Span is the algebraic difference between the output at full scale pressure and offset.
Under normal conditions external offset and span calibrations are not needed. In case fine trimming is required, offset
adjustments are possible to lower values only. Do not trim for nominal value minus 150 mV. Span adjustments are
possible to lower pressure range (higher gain). Do not trim for more than 15 % of full scale pressure.
August 2007 / 026
2/4
www.first-sensor.com
www.sensortechnics.com
HCX Series
Fully signal conditioned pressure transducer
OUTLINE DRAWING
NOM 35
Top view
HCX(M)...6H, HCXPM...6H
NOM Ø 2.6
NOM 29
P1:
High pressure port for
5 mbar and 10 mbar devices
P2:
High pressure port for
all other devices
(forward gage)
654321
SENSORTECHNICS
NOM 19 5.25
3
3
P2
P1
NOM 25
15 ±0.2
25
Ø 3.0
Ø 4.4
Ø 5.4
Ø 6.3
Ø 4.8
NOM 26
max. 13
pin size 0.5 x 0.35
12 ±0.5
NOM 2.54
mass: 14 g
HCX(M)...6V
min. 3.2
NOM
5 x 2.54
dimensions in mm
NOM 35
Top view
NOM Ø 2.6
NOM 29
P1:
High pressure port for
5 mbar and 10 mbar devices
P2:
High pressure port for
all other devices
654321
SENSORTECHNICS
NOM 19
3
P2
P1
3
Ø 6.3
25
15 ±0.2
39.6 +0.5/-0.3
NOM 6.8
August 2007 / 026
pin size 0.5 x 0.35
NOM 2.54
NOM 25
Ø 4.8
Ø 3.0
Ø 4.4
Ø 5.4
mass: 14 g
third angle projection
pin 1
max. 13
12 ±0.5
min. 3.2
third angle projection
NOM
5 x 2.54
pin 1
dimensions in mm
3/4
www.first-sensor.com
www.sensortechnics.com
HCX Series
Fully signal conditioned pressure transducer
ORDERING INFORMATION
Pressure range
Part number / Package version
Top facing ports
Side facing ports
differential / gage devices
0 to ±5 mbar
0 to 10 mbar
0 to ±10 mbar
0 to 20 mbar
0 to 50 mbar
0 to 100 mbar
0 to 350 mbar
0 to 1 bar
0 to 2 bar
0 to 5 bar
HCXPM005D6V
HCXM010D6V
HCXPM010D6V
HCXM020D6V
HCXM050D6V
HCXM100D6V
HCXM350D6V
HCX001D6V
HCX002D6V
HCX005D6V
HCXPM005D6H
HCXM010D6H
HCXPM010D6H
HCXM020D6H
HCXM050D6H
HCXM100D6H
HCXM350D6H
HCX001D6H
HCX002D6H
HCX005D6H
absolute devices
0 to 1 bar
0 to 2 bar
HCX001A6V
HCX002A6V
HCX001A6H
HCX002A6H
First Sensor reserves the right to make changes to any products herein. First Sensor does not assume any liability arising out of the application
or use of any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others.
August 2007 / 026
4/4
www.first-sensor.com