WSA5-M series

WSA5-M
3 0 Wa t t M e d i c a l Grade Switchmode Power Supplies
series
Wide Input Voltage 90 to 264 VAC, 47 to 63 Hz
Output Voltage Tolerance: ±2.5% Typical
0 to 40°C Operating Temperature
Short Circuit and Overload Protection
LPS Efficiency Level IV
Available Features
• More Output Voltage Options - 7V to 48V
• Fixed-In-Place Input Cord
• Wall Plug-in Configuration
• Output Cordsets and Connector Configurations
• Optional ON/OFF Swtich
Please contact Astrodyne/Jerome Industries at 908-353-5700
Model Number
WSA512M
WSA515M
WSA524M
Rated Power
30 W
30 W
30 W
WWW.ASTRODYNE.COM
Output Voltage
Output Amps
12 Volts(DC)
15 Volts(DC)
24 Volts(DC)
2.5 Amps
2.0 Amps
1.25 Amps
PHONE: 1-908-353-5700
FAX: 1-908-353-1021
3 0 Wa t t M e d i c a l Grade Switchmode Power Supplies
INPUT SPECIFICATIONS
Input Voltage Range
Frequency Range
series
GENERAL SPECIFICATIONS
90-264 VAC
47-63 Hz
Leakage Current @ 275VAC, 60Hz Input and Worst Case Measurement Condition including Humidity Preconditioning per
IEC60601-1 Clause 19
Class 1 Input Configuration
Earth Leakage Current
Normal Condition (NC): 157uA Single Fault Condition (SFC): 231.4uA
Enclosure Leakage Current
NC:
1.36uA
SFC:
2.08uA
Patient Leakage Current to Output Negative Cord Connection
NC:
58uA
SFC:
89uA
Patient Leakage Current, Mains Voltage Applied to Output Negative Cord Connection
NC:
N/A
SFC:
153uA
Class 2 Input Configuration
Enclosure Leakage Current
NC:
1.28uA
SFC:
2.06uA
Patient Leakage Current to Output Negative Cord Connection
NC:
64uA
SFC:
112uA
Patient Leakage Current, Mains Voltage Applied to Output Negative Cord Connection
NC:
N/A
SFC:
204uA
OUTPUT SPECIFICATIONS
Voltage and Current
Output Voltage Variance
Load Regulation, FL-NL
Ripple/Noise @ FL
Short Circuit Protection
Hold-Up Time @ 120 Vin
WSA5-M
See Selection Charts
± 5% @ FL
± 5%, max.
100mV p-p max.
Continuous
20mS min.
Isolation
4000 Volts(AC)
Efficiency Level IV
EMC (Note 1)
Meets requirements of European
EMC Directive and the FCC
Emissions
FCC 47 CFR part 15B, CISPR11B
(EN55011B), CISPR22B
(EN55022B)
EN61000-6-3 (light industrial) and
EN61000-3-2
Immunity
CISPR24 (EN55024), EN61326 and
EN61000-6-1
Safety
(c)UL:
60601-1 (File # E160908)
TUV:
EN60601-1 (File # B050624072019)
MTBF @ 40°C
100,000 Hrs min.
PHYSICAL SPECIFICATIONS
Size
Weight
3.25” x 2.72” x 1.89”
7.76oz (220g)
ENVIRONMENTAL SPECIFICATIONS
Operating:
Operating. Temperature
0 to +40°C
Humidity
10% to 95%, non-cond
Atmospheric Pressure
70 to 106 kPa
Storage:
Storage Temperature
-40 to +75°C
Humidity
10% to 95%, non-cond
Atmospheric Pressure
50 to 106 kPa
NOTES
1. The EMC characteristics of the final system comprised of both the
power supply and the main system must be evaluated to the
applicable EMC environment regulations
All specifications are typical at nominal input, full load, and 25°C unless otherwise noted
WWW.ASTRODYNE.COM
PHONE: 1-908-353-5700
FAX: 1-908-353-1021
WSA5-M
3 0 Wa t t M e d i c a l Grade Switchmode Power Supplies
series
MECHANICAL DIMENSIONS
STYLE A
2.72
3.25
OUTPUT CORD
Length: 6 FT
Gauge: 18
# of Wires: 2
Color: Black
1.89
Shown as desktop with IEC 320 Inlet
Fixed input cord or wall plug-in configuration
A
B
C
.374”
9.5mm
.216”
5.5mm
.08 3”
2.1mm
Call for engineering assistance at
(908) 353-5700 or fax us at (908) 353-1021
WWW.ASTRODYNE.COM
PHONE: 1-908-353-5700
FAX: 1-908-353-1021