WSA5-M 3 0 Wa t t M e d i c a l Grade Switchmode Power Supplies series Wide Input Voltage 90 to 264 VAC, 47 to 63 Hz Output Voltage Tolerance: ±2.5% Typical 0 to 40°C Operating Temperature Short Circuit and Overload Protection LPS Efficiency Level IV Available Features • More Output Voltage Options - 7V to 48V • Fixed-In-Place Input Cord • Wall Plug-in Configuration • Output Cordsets and Connector Configurations • Optional ON/OFF Swtich Please contact Astrodyne/Jerome Industries at 908-353-5700 Model Number WSA512M WSA515M WSA524M Rated Power 30 W 30 W 30 W WWW.ASTRODYNE.COM Output Voltage Output Amps 12 Volts(DC) 15 Volts(DC) 24 Volts(DC) 2.5 Amps 2.0 Amps 1.25 Amps PHONE: 1-908-353-5700 FAX: 1-908-353-1021 3 0 Wa t t M e d i c a l Grade Switchmode Power Supplies INPUT SPECIFICATIONS Input Voltage Range Frequency Range series GENERAL SPECIFICATIONS 90-264 VAC 47-63 Hz Leakage Current @ 275VAC, 60Hz Input and Worst Case Measurement Condition including Humidity Preconditioning per IEC60601-1 Clause 19 Class 1 Input Configuration Earth Leakage Current Normal Condition (NC): 157uA Single Fault Condition (SFC): 231.4uA Enclosure Leakage Current NC: 1.36uA SFC: 2.08uA Patient Leakage Current to Output Negative Cord Connection NC: 58uA SFC: 89uA Patient Leakage Current, Mains Voltage Applied to Output Negative Cord Connection NC: N/A SFC: 153uA Class 2 Input Configuration Enclosure Leakage Current NC: 1.28uA SFC: 2.06uA Patient Leakage Current to Output Negative Cord Connection NC: 64uA SFC: 112uA Patient Leakage Current, Mains Voltage Applied to Output Negative Cord Connection NC: N/A SFC: 204uA OUTPUT SPECIFICATIONS Voltage and Current Output Voltage Variance Load Regulation, FL-NL Ripple/Noise @ FL Short Circuit Protection Hold-Up Time @ 120 Vin WSA5-M See Selection Charts ± 5% @ FL ± 5%, max. 100mV p-p max. Continuous 20mS min. Isolation 4000 Volts(AC) Efficiency Level IV EMC (Note 1) Meets requirements of European EMC Directive and the FCC Emissions FCC 47 CFR part 15B, CISPR11B (EN55011B), CISPR22B (EN55022B) EN61000-6-3 (light industrial) and EN61000-3-2 Immunity CISPR24 (EN55024), EN61326 and EN61000-6-1 Safety (c)UL: 60601-1 (File # E160908) TUV: EN60601-1 (File # B050624072019) MTBF @ 40°C 100,000 Hrs min. PHYSICAL SPECIFICATIONS Size Weight 3.25” x 2.72” x 1.89” 7.76oz (220g) ENVIRONMENTAL SPECIFICATIONS Operating: Operating. Temperature 0 to +40°C Humidity 10% to 95%, non-cond Atmospheric Pressure 70 to 106 kPa Storage: Storage Temperature -40 to +75°C Humidity 10% to 95%, non-cond Atmospheric Pressure 50 to 106 kPa NOTES 1. The EMC characteristics of the final system comprised of both the power supply and the main system must be evaluated to the applicable EMC environment regulations All specifications are typical at nominal input, full load, and 25°C unless otherwise noted WWW.ASTRODYNE.COM PHONE: 1-908-353-5700 FAX: 1-908-353-1021 WSA5-M 3 0 Wa t t M e d i c a l Grade Switchmode Power Supplies series MECHANICAL DIMENSIONS STYLE A 2.72 3.25 OUTPUT CORD Length: 6 FT Gauge: 18 # of Wires: 2 Color: Black 1.89 Shown as desktop with IEC 320 Inlet Fixed input cord or wall plug-in configuration A B C .374” 9.5mm .216” 5.5mm .08 3” 2.1mm Call for engineering assistance at (908) 353-5700 or fax us at (908) 353-1021 WWW.ASTRODYNE.COM PHONE: 1-908-353-5700 FAX: 1-908-353-1021