User Guide Heavy Duty CFM Thermo

User Guide
Heavy Duty CFM Thermo-Anemometer
Model 407113
Introduction
Congratulations on your purchase of the Extech 407113 CFM meter. This handheld meter can
display Air Flow (CFM) with Area or Air Velocity with Temperature on a 10,000-count (0 to 9999)
dual display LCD. Other functions include MIN/MAX Record/Recall, RS-232 PC interface, and
Data Hold. The metal vane sensor offers rugged durability and higher temperature withstand than
comparable devices. Careful use of this meter will provide years of reliable service.
Specifications
General Specifications
Display
Measurements
Dual Display Multi-function 10,000-count (0 to 9999) LCD
Air Velocity: m/s, km/h, ft/min, knots, mph;
3
3
Air Flow: CMM (m /min) and CFM (ft /min); Temperature: °C and °F
Data Hold
Freezes displayed reading
Sampling rate
One (1) reading per second
Sensors
Air velocity/flow sensor: Metal angled vane arms with low-friction ball
bearing. Temp. Sensor: Precision thermistor
MIN/MAX
Records/Recalls lowest and highest readings
Auto Power OFF
Automatic shut off after 15 minutes
PC Interface
RS-232 Serial Communications (16-bit data stream)
Over range indication
1___ appears on the LCD
Low battery indication
LBT appears on the LCD
Power supply
9V Battery (consumption 8.3mA approx.)
Operating conditions
Meter: 32 to 122 F (0 to 50 C); 80% RH max.
Dimensions / Weight
Sensor: 32 to 175 F (0 to 80 C)
Main instrument: 7.1 x 2.8 x1.3" (180 x 72 x 32mm)
Sensor head: 2.8” (72mm) diameter
Weight
0.84 lbs. (381g) for meter and sensor
o
o
o
o
2
407113-EN v3.6 4/14
Range Specifications
Air Velocity Measurements
Range
Resolution
Accuracy (%rdg)
m/s (meters per second)
0.50 - 35.00 m/s
0.01 m/s
± (2% + 0.2m/s)
km/h (kilometers per hour)
1.8 – 126.0 km/h
0.1 km/h
± (2% + 0.8km/h)
ft/min (feet per minute)
100 – 6890 ft/min
1 ft/min
± (2% + 40ft/min)
mph (miles per hour)
1.1 – 78.3 mph
0.1 mph
± (2% + 0.4m/h)
knots (nautical miles per hour )
1.0 to 68 knots
0.1 knots
± (2% + 0.4knots)
Range
Resolution
Area
3
0.001 to 100
0 to 9,999m
3
Air Flow Measurements
CMM (cubic meters per minute)
0-999,900 m /min
CFM (cubic feet per minute)
0-999,900 ft /min
0.001 to 100
0 to 9,999ft
Range
Resolution
Accuracy
Air Temperature
o
o
32 to 175 F (0 to 80 C)
3
o
0.1 F/C
o
2
2
o
1.5 F (0.8 C)
407113-EN v3.6 4/14
Meter Description
1.
Sensor Input
2.
RS-232 PC Interface jack
3.
LCD Display
4.
Vane
5.
Rubber Holster
6.
Keypad
7.
Battery compartment (rear)
5
1
2
4
3
6
7
Note: To access the rear battery compartment, first remove the rubber holster that jackets the meter.
4
407113-EN v3.6 4/14
Operation
Air Velocity Measurements
1. Connect the sensor to the sensor input jack on top of the meter.
2. Turn on the meter using the Power button.
3. Select the VELOCITY function using the FLOW / VELOCITY button. The LCD will display VEL
when the velocity mode is selected.
4. Select the desired temperature units using the C/F select button. The LCD will reflect the current
unit selection.
5. Select the desired air velocity units
using the UNIT button. The LCD will
reflect the current unit selection.
6. Place the sensor in the air current to
be measured with the
symbol on the inlet side of the vane
(see diagram).
Air Direction
7. View the air velocity and
temperature readings on the LCD
Display. The large main LCD
display shows the Air Velocity
reading. The lower LCD sub-display
shows the temperature reading.
Data Hold
1.
While taking measurements, press the HOLD button to freeze the LCD reading.
2.
The HOLD indicator will appear on the LCD when the display is in Data Hold mode.
3.
Press HOLD again to return to normal operation.
Maximum and Minimum Recording
The 407113 allows the user to record and view the highest (MAX) and lowest (MIN) readings.
1. Press the RECORD/RECALL button once. The REC indicator will appear on the display and
the meter will begin recording the MAX and MIN values.
2. Press the RECORD/RECALL button to stop the MAX/MIN recording and display the maximum
reading. The MAX indicator along with the maximum reading will appear on the LCD display
3. Press RECORD/RECALL again to view the minimum value. The MIN indicator along with the
minimum reading will appear on the LCD display.
4. To return to normal operation, press and hold the RECORD/RECALL button for approx. 3
seconds. The display indicators REC, MAX, and MIN will disappear.
NOTE: Placing the meter in the RECORD/RECALL mode by pressing
the RECORD/RECALL button disables the AUTO POWER OFF feature.
NOTE: Handle the vane carefully. If the metal vane blades are inadvertently
bent or damaged erroneous readings may result.
5
407113-EN v3.6 4/14
Air Flow (Volume) Measurements (CMM / CFM)
1. Connect the sensor to the sensor input jack on top of the meter.
2. Turn on the meter using the POWER button.
3. Select the FLOW mode using the FLOW/VELOCITY button. The LCD will display FLOW CFM or
FLOW CMM when the flow function has been selected.
4. Select the desired air flow units: CMM (cubic meters per minute) or CFM (cubic feet per minute)
using the UNIT button. The LCD will reflect the selection.
5. Measure the dimensions of the duct or vent and calculate the area in square feet or square
meters
Note: If the dimensional measurements are made in inches (or cm), convert them feet (or meters)
before calculating the square area.
2
2
button to increment
6. Press the AREA button to begin entering the area in m or ft . Use the
the flashing digit, use the
button to decrement, use the button to select the next digit, and
use the button to set the decimal point.
7. Press the ENTER/RESET button after the area value has been entered. The bottom display will
2
2
indicate the area entered in ft or m . The main LCD displays the Air Flow in CFM (cubic feet per
minute) or CMM (cubic meters per minute). If the CFM or CMM reading exceeds 9999, use the
displayed X10 or X100 multiplier to calculate the reading.
8. Note that for FLOW measurements, three modes apply: The normal, default mode, where the
actual flow is indicated and the two modes described below under 2/3 MAX Flow and AVG
Flow.
Note: The temperature function is not active in the FLOW mode.
2/3V MAX Flow Mode
In this mode the meter will display two-thirds the measured Flow (Volume). To access this mode,
ensure that the meter is in the FLOW mode (via the FLOW/VELOCITY button) and then press the
FLOW MODE button until the 2/3V MAX display icon appears. Now the Flow display will be 2/3 the
actual measurement. To return to the normal display mode, press the FLOW MODE button until the
2/3VMAX and the AVG icons switch off.
AVG (Average) Flow Mode
In this mode the meter will display the AVERAGE Flow for up to 20 readings. To access this mode,
ensure that the meter is in the FLOW mode (via the FLOW/VELOCITY button) and press the FLOW
MODE button until the AVG display icon appears. Now each time the AVG/START button is pressed
a reading will be taken and averaged. The lower LCD display becomes a 20-reading counter and
with each reading taken the counter is incremented. Allow 3 or 4 seconds between AVG/START
button presses for the display counter to update. The main LCD displays the averaged Air Flow
value. To return to the normal display mode, press the FLOW MODE button until the AVG icon
switches off.
RS-232 PC Interface
The 407113 is equipped with a 3.5mm phone jack (meter top) for connection to a PC for data
acquisition purposes. For streaming of data to a PC via the RS232 Output jack, the optional
407001-USB kit (RS232 to USB cable and driver CD) along with the 407001 software (available free
at www.extech.com) are required. Instructions for use are provided with the data acquisition
software/hardware kits.
6
407113-EN v3.6 4/14
Battery Replacement
When the LBT icon appears on the LCD, the 9V battery must be replaced.
1. Remove the rubber holster that surrounds the entire meter
2. Slide off the rear battery compartment
3. Replace the 9V battery
4. Affix the battery compartment cover and the meter holster
Useful Equations and Conversions
Area equation for rectangular or square ducts
Height (H)
Area (A) = Width (W) x Height (H)
Width (W)
Area equation for circular ducts
Area (A) = pi x r2
Where pi = 3.14 and r2 = radius x radius
Radius
Cubic equations
CFM (ft3/min) = Air Velocity (ft/min) x Area (ft2)
CMM (m3/min) = Air Velocity (m/sec) x Area (m2) x 60
NOTE: Measurements made in inches
must be converted to feet or meters before using the above formulae.
Unit of Measure Conversion Table
m/s
ft/min
knots
km/h
MPH
1
196.87
1.944
3.6
2.24
1 ft/min
0.00508
1
0.00987
0.01829
0.01138
1 knot
0.5144
101.27
1
1.8519
1.1523
1 km/h
0.2778
54.69
0.54
1
0.6222
1 MPH
0.4464
87.89
0.8679
1.6071
1
1 m/s
7
407113-EN v3.6 4/14
Warranty FLIR Systems, Inc. warrants this Extech Instruments brand device to be free of defects in parts and workmanship for one year from date of shipment (a six month limited warranty applies to sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department for authorization. Visit the website www.extech.com for contact information. A Return Authorization (RA) number must be issued before any product is returned. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. FLIR Systems, Inc. specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. FLIR’s total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied. Calibration, Repair, and Customer Care Services FLIR Systems, Inc. offers repair and calibration services for the Extech Instruments products we sell. NIST certification for most products is also provided. Call the Customer Service Department for information on calibration services available for this product. Annual calibrations should be performed to verify meter performance and accuracy. Technical support and general customer service is also provided, refer to the contact information provided below. Support Lines: U.S. (877) 439‐8324; International: +1 (603) 324‐7800 Technical Support: Option 3; E‐mail: [email protected] Repair & Returns: Option 4; E‐mail: [email protected] Product specifications are subject to change without notice Please visit our website for the most up‐to‐date information www.extech.com FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA Copyright © 2014 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form www.extech.com
8
407113-EN v3.6 4/14
Garantie FLIR Systems, Inc. garantit que cet appareil Extech Instruments est exempt de défauts matériaux et de fabrication pendant un an à partir de la date d’envoi (une garantie limitée de six mois s’applique aux capteurs et aux câbles). Si le renvoi de l’appareil pour réparation devient nécessaire durant ou après la période de garantie, contactez le service client pour autorisation. Pour obtenir les coordonnées, visitez le site Web suivant : www.extech.com. Un numéro d’autorisation de retour (AR) doit être délivré avant tout retour de produit. L’expéditeur prend à sa charge les frais d’expédition, le fret, l’assurance et l’emballage correct de l’appareil afin de prévenir toute détérioration durant le transport. Cette garantie ne s’applique pas aux dommages imputables à l’utilisateur, tels que l’usage impropre ou abusif, un mauvais câblage, une utilisation non conforme aux spécifications, un entretien ou une réparation incorrecte, ou toute modification non autorisée. FLIR Systems, Inc. déclinera spécifiquement toute garantie ou qualité marchande ou aptitude à l’emploi prévu, et ne sera en aucun cas tenu responsable pour tout dommage conséquent, direct, indirect ou accidentel. La responsabilité totale de FLIR est limitée à la réparation ou au remplacement du produit. La garantie définie ci‐dessus est inclusive et aucune autre garantie, écrite ou orale, n’est exprimée ou implicite.
Calibrage, réparation et services après‐vente FLIR Systems, Inc. offre des services de calibrage et de réparation pour les produits Extech Instruments que nous commercialisons. Nous fournissons également une certification NIST pour la plupart des produits. Contactez notre service client pour toute information sur les services de calibrage disponibles pour ce produit. Un calibrage doit être effectué chaque année pour vérifier les performances et la précision du mètre. Nous offrons également une assistance technique et un service à la clientèle. Veuillez vous reporter aux coordonnées fournies ci‐dessous. Lignes d’assistance: États‐Unis (877) 439‐8324; international: +1 (603) 324‐7800 Service d’assistance technique : Option 3 ; E‐mail : [email protected] Réparations et retours : Option 4 ; E‐mail : [email protected] Les spécifications produit sont sujettes à modifications sans préavis. Pour les toutes dernières informations, veuillez visiter notre site Web. www.extech.com FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA Copyright © 2014 FLIR Systems, Inc. Tous droits réservés, y compris la reproduction partielle ou totale sous quelque forme que ce soit. www.extech.com 9
407113-EN v3.6 4/14
Garantía FLIR Systems, Inc., garantiza este dispositivo marca Extech Instruments para estar libre de defectos en partes o mano de obra durante un año a partir de la fecha de embarque (se aplica una garantía limitada de seis meses para cables y sensores). Si fuera necesario regresar el instrumento para servicio durante o después del periodo de garantía, llame al Departamento de Servicio a Clientes para obtener autorización. Visite www.extech.com para Información de contacto. Se debe expedir un número de Autorización de Devolución (AD) antes de regresar cualquier producto. El remitente es responsable de los gastos de embarque, flete, seguro y empaque apropiado para prevenir daños en tránsito. Esta garantía no se aplica a defectos resultantes de las acciones del usuario como el mal uso, alambrado equivocado, operación fuera de las especificaciones, mantenimiento o reparación inadecuada o modificación no autorizada. FLIR Systems, Inc., rechaza específicamente cualesquier garantías implícitas o factibilidad de comercialización o idoneidad para cualquier propósito determinado y no será responsable por cualesquier daños directos, indirectos, incidentales o consecuentes. La responsabilidad total de FLIR está limitada a la reparación o reemplazo del producto. La garantía precedente es inclusiva y no hay otra garantía ya sea escrita u oral, expresa o implícita. Servicios de calibración, reparación y atención a clientes FLIR Systems, Inc., ofrece servicios de reparación y calibración para los productos que vendemos de Extech Instruments. Además ofrecemos certificación NIST para la mayoría de los productos. Llame al Departamento de Servicio al Cliente para solicitar información de calibración para este producto. Para verificar el funcionamiento y precisión se debe realizar la calibración anual. Además se provee Soporte Técnico y servicios generales al cliente, consulte la información de contacto en seguida. Líneas de soporte: EE.UU. (877) 439‐8324; Internacional: +1 (603) 324‐7800 Soporte Técnico Opción 3; correo electrónico: [email protected] Reparación / Devoluciones: Opción 4; correo electrónico: [email protected] Las especificaciones del producto están sujetas a cambios sin aviso Por favor visite nuestra página en Internet para la información más actualizada www.extech.com FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA Certificado ISO 9001 Copyright © 2014 FLIR Systems, Inc. Reservados todos los derechos, incluyendo el derecho de reproducción total o parcial en cualquier medio
www.extech.com
10
407113-EN v3.6 4/14