Manual - Extech Instruments

User Guide
Flexible AC Clamp Meter
3000 Amp / 600 Volt
Model 382400
Introduction
Congratulations on your purchase of the Model 382400 Extech AC Clamp device. The Model
382400 features:
ƒ True RMS AC measurements
ƒ 3000A maximum current
ƒ 600V maximum voltage
ƒ Fast Peak Hold feature
ƒ MAX-MIN and Data Hold features
ƒ Auto Power OFF
This meter is shipped fully test and calibrated. Careful use of this meter will provide years of reliable
service.
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to the
manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages
may be present
Double insulation
SAFETY NOTES
•
•
•
Do not exceed the maximum allowable measurement ranges.
Set the function switch to the OFF position when the meter is not in use.
Remove the batteries if the meter is to be stored for longer than 60 days.
WARNINGS
•
•
•
•
•
Do not install the flexible current probe around bare conductors carrying a voltage ranging from
30V to 600V unless wearing protective clothing and gloves suitable for high-voltage work.
Always inspect and check any damage of the current probe assembly before usage. Do not
use the flexible current probe if any damage is found.
Do not use the meter on circuits rated higher then 600V.
Set function switch to the appropriate position before measuring.
When measuring voltage do not switch to the current mode.
2
382400-EN-V1.2-3/12
CAUTIONS
•
•
•
•
•
Improper use of this meter can cause damage, shock, injury or death. Read and understand
this user manual before operating the meter.
Always remove the test leads and clamp jaw from circuits or devices under test before
replacing the battery or fuses.
Use great care when making measurements if the voltages are greater than 25VAC rms.
These voltages are considered a shock hazard.
Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty
of connection to the recessed electrical contacts. Other means should be used to ensure that
the terminals are not "live".
If the equipment is used in a manner not specified by the manufacturer, the protection provided
by the equipment may be impaired.
MAXIMUM INPUTS
Function
Maximum Input
Current (Amperes)
3000A AC
Voltage (Volts)
600V AC
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which
measures are taken to limit the transient overvoltages to an appropriate low level.
Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied
from the fixed installation.
Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.
Note – Examples include switches in the fixed installation and some equipment for industrial use
with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.
Note – Examples include electricity meters and primary over-current protection equipment
3
382400-EN-V1.2-3/12
Description
Meter Description
1.
Clamp conductor
2.
COM test lead input jack
3.
Voltage (V) test lead input jack
4.
LCD Display
5.
HOLD, MIN, MAX, PEAK mode button
6.
Function select switch
7.
50/60Hz select push-button
8.
Flex Clamp probe
9.
Clamp coupling mechanism
Display Description
H
Data Hold
MAX
Max reading
MIN
Min reading
Pk
Peak reading
Low battery
V
Volts (Voltage)
A
Amperes (Current)
HZ
Frequency
AP
Auto Power OFF
Minus sign
Negative reading display
0 to 9999
Measurement display digits
4
382400-EN-V1.2-3/12
Operation
NOTES: Read and understand all Warning and Caution statements in this operation manual prior to
using this meter. Set the function select switch to the OFF position when the meter is not in use.
AC True RMS Current Measurements
WARNING: Ensure that the test leads are disconnected from the meter before making current
clamp measurements.
1.
Set the Function switch to the A position.
2.
Select the fundamental frequency (50Hz or 60Hz) with the 50Hz-60Hz push-button. The LCD
display will reflect the selected frequency.
3.
Open the clamp loop by pressing the latch tabs and
disconnecting the clamp coupling connector.
4.
Fully enclose only one conductor centered inside the clamp
loop. Allow at least 25mm between the conductor and the
clamp coupling connector.
5.
The meter’s LCD will display the current (A) and frequency
(Hz) readings.
Note: For measurements lower than 3A, the meter’s frequency
display will indicate ‘0’.
‘OL’ is displayed when the measurement exceeds the
specified maximum limit.
AC True RMS Voltage Measurements
WARNING: Maximum input is 600Vac rms. Do not attempt to take any voltage measurement that
exceeds this limit. Exceeding the limit could cause electrical shock and damage to the meter.
1.
Insert the black test lead into the negative COM terminal and the red test lead into the positive
V terminal.
2.
Set the function switch to the V position.
3.
Select the fundamental frequency (50Hz or 60Hz) with the
50Hz-60Hz push-button. The LCD display will reflect the
selected frequency.
4.
Connect the test leads in parallel to the circuit under test.
5.
Read the voltage measurement (V) and frequency (Hz) on
the LCD display. ‘OL’ is displayed when the measurement
exceeds the specified maximum limit.
Note: For measurements lower than 30V, the meter’s
frequency display will indicate ‘0’.
5
382400-EN-V1.2-3/12
DATA HOLD, MAX-MIN, and PEAK Modes
1.
2.
Measure current or voltage as previously described.
Press the HOLD button to access the Data Hold mode; an H icon will be displayed. Use the
Data Hold mode to freeze a displayed reading.
3. Press again to access the MAX mode. In the MAX mode, only the highest reading will be
displayed. The MAX icon will be indicated on the LCD display.
4. Press again to access the MIN mode. In the MIN mode, only the lowest reading will be
displayed. The MIN icon will be indicated on the LCD display.
5. Press again to access the PEAK mode. The PEAK display icon will be indicated on the LCD
display.
6. Press and hold the HOLD button for 3 seconds to return to the normal operating mode.
Note: The sampling time of the PEAK Mode is 39µS (50Hz) or 33µS (60Hz).
HOLD, MAX and MIN display the TRMS value
Automatic Power OFF
In order to conserve battery life, the meter will automatically turn off after approximately 15 minutes.
To turn the meter on again, turn the function switch to the OFF position and then to the desired
function position.
To defeat the Auto Power OFF feature, press and hold the HOLD button while turning the meter on.
6
382400-EN-V1.2-3/12
Specifications
AC Current
50/60Hz, Auto Range, True RMS, Crest Factor <4
Conductor should be positioned at the center of the flexible loop
Range
Resolution
0 to 300.0 AAC
Accuracy *
0.1A
300.0 to 999.9 A
1000 to 3000 A
± 1.0% of range
1A
*Position sensitivity is 2% of range; External field effect of < 40A/m and 200mm from the coupling is 1%
of range; Temperature coefficient is 0.02% of reading/oC.
AC Voltage
50/60Hz, Auto Range, True RMS, Crest Factor <4, Input Impedance 10M ohm, Overload
protection 800V
Range
Resolution
Accuracy
4.0V to 600.0 VAC
0.1V
± (0.5% reading+ 5 digits)
Frequency
ACV value > 30VAC or ACA value > 30A
Range
Resolution
45 – 65
0.1
Probe Specifications
Probe length
24” (610mm)
Bending diameter
1.4” (35mm)
Connector diameter
0.9” (23mm)
Cable diameter
0.5 “(14mm)
Cable length (probe to meter)
6.7” (170mm)
Cable length (meter to output)
6.7” (170mm)
General Specifications
Display
Range selection
Low Battery indication
Over-range indication
Measurement rate
Peak Hold
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Operating Altitude
Battery
Power consumption
Auto power OFF
Weight
Accuracy
± 0.2Hz
Four (4) digit (9999 counts) LCD
Automatic
Battery symbol is displayed
‘OL’ is displayed
2 readings per second, nominal
Captures peaks 39µS (50Hz) or 33µS (60Hz)
14°F to 185 °F (-10°C to 85°C)
-4°F to 185°F (-20°C to 85°C)
15% to 85%
15% to 85%
7000ft. (2000m) max.
Two (2) 1.5V ‘AA’ batteries
10mA approx.
After approx. 15 minutes
15.2 oz. (430g)
Safety
For indoor use and in accordance with the requirements for double
insulation to EN61010-1-032 Category III 600V; Pollution Degree 2.
Approvals
CE
7
382400-EN-V1.2-3/12
Maintenance
WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads
from the input terminals, and turn OFF the meter before opening the case. Do not operate the meter
with an open case.
Inspection
Inspect the clamp for damage before each use; have the unit repaired before use if damage is
evident.
Cleaning and Storage
Periodically wipe the meter and probe with a damp cloth and mild detergent; do no use abrasives or
solvents. If the meter is not to be used for 60 days or more, remove the batteries and store
separately.
Battery Replacement
When the low battery symbol appears on the LCD, replace the two 1.5V ‘AA’
batteries.
1.
Remove the center Phillips head screw that secures the rear battery door
(see diagram)
2.
Open the battery compartment
3.
Replace the 1.5V ‘AA’ batteries
4.
Secure the battery compartment
You, as the end user, are legally bound (EU Battery ordinance) to return all used
batteries, disposal in the household garbage is prohibited! You can hand over your
used batteries / accumulators at collection points in your community or wherever batteries
/ accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the
end of its lifecycle
8
382400-EN-V1.2-3/12
Warranty
EXTECH INSTRUMENTS CORPORATION (a FLIR company) warrants this instrument to be free of defects in parts and
workmanship for one year from date of shipment (a six month limited warranty applies to sensors and cables). If it should
become necessary to return the instrument for service during or beyond the warranty period, contact the Customer
Service Department for authorization. Visit our website www.extech.com for contact information. A Return Authorization
(RA) number must be issued before any product is returned to Extech. The sender is responsible for shipping charges,
freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting
from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or
repair, or unauthorized modification. Extech specifically disclaims any implied warranties or merchantability or fitness for a
specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is
limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether
written or oral, is expressed or implied.
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also provides NIST certification for most
products. Call the Customer Service Department for information on calibration services available for this product. Extech
recommends that annual calibrations be performed to verify meter performance and accuracy.
Support lines: U.S. 877-439-8324, Intl. 603-324-7800
Technical support: Option 3; E-mail: [email protected]
Repair & Returns: Option 4; E-mail: [email protected]
Product specifications subject to change without notice
Visit our website: www.extech.com
Extech Instruments Corporation, 9 Townsend West, Nashua, NH 03063
ISO 9001 Certified since 1995
9
382400-EN-V1.2-3/12