Manual - Extech Instruments

User's Guide
Datalogging Light Meter
with PC Interface
Model 401036
Introduction
Congratulations on your purchase of the Extech 401036 Datalogging Light Meter. This device
measures light intensity to 20,000 Foot-candles (Fc) or Lux. 16,000 measurement readings can be
logged by the meter and later transferred to PC using the RS-232 interface. Real-time meter/PC
logging is also supported. Careful use of this meter will provide years of reliable service.
Specifications
Display
3 ½ digit (2000 count) multifunction LCD display
Sensor
Silicon Photo-diode meets C.I.E. photopic curve V (λ)
Ranges / Resolution
19.99, 199.9, 1999, and 1999 x 10 Fc and Lux
Accuracy
± (3% reading + 5 digits)
Repeatability
± 2%
Datalogger memory
16,000 readings max.
PC Interface
RS-232C Serial Communications at 9600 bps
Range status indication
'OL' is displayed for measurement exceeding published limits
Sampling time
2.5 readings per second
Power Supply
9V battery (display includes low battery indicator);
50 hour typical battery life
Operating temperature
32 to 104 F (0 to 40 C)
Operating Humidity
< 80% RH
o
o
Weight
9.6 oz. (300g)
Dimensions
Meter: 5.7 x 2.8 x 1.5" (146 x 70 x 39mm);
Accessories
Carrying case, Windows
9V battery
Probe: 3.4 x 2.4 x 1.1" (87.5 x 60 x 29mm)
2
TM
software, PC Interface cable, &
401036-EN v3.2 07/13
Meter Description
1. 2000 Count LCD includes LUX, Fc, low battery, MAX, H (hold), & x10 indicators
2. ON/OFF key Turns the meter on or off
3. Max. Hold key When pressed, the LCD
displays only the maximum reading
1
4. Lux/Fc key Press to toggle the displayed
unit of measure. The LCD shows the
unit
6
2
3
4
5. Light Sensor (Lens cover not shown)
Collects light. Note that a tripod mount
is located on the rear of the sensor
6. Range key Press repeatedly to select
desired range: 20, 200, 2,000, 20,000
7
8
9
10
7. Data Hold key Press to freeze the
displayed measurement
8. Record/Erase key Press momentarily to
store one reading; Press and hold for 3
seconds to activate the continuous
recording mode. Press again to end
recording
5
9. RS-232 jack Stereo jack for PC to Meter
connection cable
10. Zero adjust Cover the sensor with
supplied lens cap and adjust the zero
pot for a 0.0 Fc/LUX display
Note that the battery compartment, tilt stand, and hanging mount are located on the rear of the
meter
Spectral Sensitivity Curve
= Light Meter
= V (λ)
100
Relative
Sensitivity (%)
80
60
40
20
400
450
500
550
600
650
700
Wavelength (nm)
3
401036-EN v3.2 07/13
Operation
1. Turn the meter ON or OFF with the ON/OFF key.
2. Press the Lux/Fc key to select the unit of measure for light intensity.
3. With the lens cap completely covering the light sensor, zero the meter by adjusting the screw on
the right side of the meter for an LCD reading of 0.0.
4. Remove the lens cap to allow the sensor to collect light.
5. Read the light intensity measurement on the LCD.
6. For over-range conditions, the ‘OL’ icon will display. Select a higher range by pressing the
Range key until a valid reading replaces the ‘OL’ display.
7. For the 20,000 range, multiply the displayed reading by a factor of 10.
Data Hold
Press the HOLD key to freeze the displayed reading. Press HOLD again to return to the normal
operating mode.
Maximum (MAX) reading
Press the MAX key to display only the highest reading. As higher measurements are made the
display updates accordingly. The ‘MAX’ icon appears on the LCD in this mode of operation. To
return to normal operation, press the MAX key again and the MAX indicator will extinguish.
Range selection
Press the RANGE key to select the appropriate measurement range. Start with the lowest range
(20) and work up to higher ranges as needed. If ‘OL’ (overload) is displayed, press the range key
until a valid reading is displayed.
Datalogger
REC/ERASE Key
The REC/ERASE key is used to record and erase data. Press the REC/ERASE key once to log one
reading (REC icon flashes once on the LCD). To automatically datalog readings at programmable
intervals (continuous recording mode), press and hold the REC/ERASE key for 3 seconds (REC
icon begins flashing repeatedly).
Press the REC/ERASE key any time while recording to exit the record mode. When the memory is
full (16,000 readings or 255 sets), the LCD will display ‘FULL’ and recording ends. To download and
view the stored data, connect the meter to a PC and perform the steps provided in the ‘PC Interface’
section of this manual.
To erase stored data, perform these steps:
1. With the meter turned off, press and hold the REC/ERASE key.
2. Turn the meter on by pressing the ON/OFF key and then immediately release the ON/OFF key.
3. Continue to hold the REC/ERASE key until ‘del’ flashes 3 times on the LCD. The meter
automatically returns to normal operation after the data are erased.
4
401036-EN v3.2 07/13
Sampling Rate
The sampling rate is the interval of time between logged readings. Note that there are two sampling
rate selections for two separate datalogging scenarios described below:
1. PC Sampling Rate
The PC sampling rate is the rate at which measurements are recorded by the PC while the
meter is connected to the PC. This is known as real-time recording since the readings are
being taken by the meter and recorded by the PC at the same time. Set the rate in the
supplied software program as described in the PC SAMPLING paragraph on page 6.
2. Datalogger Sampling Rate
This is the meter’s internal sampling rate used when the meter is logging readings remotely
in continuous recording mode (disconnected from the PC). The default rate is 1 second.
Change the rate in software by following the steps on page 6 for SAMPLING TIME with the
meter connected to the PC.
PC Interface
Connect the supplied interface cable to the phono jack on the right side of the light meter and to the
PC 9-pin COM port.
TM
Windows
Software
Refer to separate instructions supplied on the Software Program disk for the operation of the
TM
Windows Software.
5
401036-EN v3.2 07/13
Battery Replacement
The meter uses a 9V battery for its operational power and it has a 50 hour typical life span. Replace
the battery when the low battery icon appears on the meter’s LCD display.
1. Remove the three (3) Phillips screws on the rear of the meter.
2. Open the meter housing and replace the battery
3. Close the meter housing and secure the rear screws
6
401036-EN v3.2 07/13
Warranty FLIR Systems, Inc. warrants this Extech Instruments brand device to be free of defects in parts and workmanship for one year from date of shipment (a six month limited warranty applies to sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department for authorization. Visit the website www.extech.com for contact information. A Return Authorization (RA) number must be issued before any product is returned. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. FLIR Systems, Inc. specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. FLIR’s total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied. Calibration, Repair, and Customer Care Services FLIR Systems, Inc. offers repair and calibration services for the Extech Instruments products we sell. NIST certification for most products is also provided. Call the Customer Service Department for information on calibration services available for this product. Annual calibrations should be performed to verify meter performance and accuracy. Technical support and general customer service is also provided, refer to the contact information provided below. Support Lines: U.S. (877) 439‐8324; International: +1 (603) 324‐7800 Technical Support: Option 3; E‐mail: [email protected] Repair & Returns: Option 4; E‐mail: [email protected] Product specifications are subject to change without notice Please visit our website for the most up‐to‐date information www.extech.com FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA ISO 9001 Certified Copyright © 2013 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form www.extech.com 7