DT10S DT10D

DT10S/D Series
DC/DC CONVERTER 10W, Reinforced Insulation, Medical Safety
FEATURES

Efficiency up to 82%

I/O Isolation 4200VAC with Reinforced Insulation,
rated for 300VAC Working Voltage
Medical Safety to UL/CSA/EN/IEC 60601-1 3rd

Edition

2 MOOP rated

Wide 2:1 Input Voltage Range

Fully regulated Output Voltage

Low Leakage Current

Operating Temp. Range –40°C to +75 °C

Input Filter meets EN 55022,class A and FCC, level A

Overload Protection

2”x 1” Plastic Package

3 Years Product Warranty
The DT10S/D series is a new range of high performance DC/DC converter modules with a reinforced insulation system .The I/Oisolation voltage is specified for 4200VACrms.The product comes in a compact 2”x1” industry standard package. All 15 models
features wide 2:1 input voltage range and fully regulated output voltage.The DT10S/D DC/DC converters offer an economical solution
for demanding applications in industrial and medical instrumentation requesting a certified supplementary or reinforced insulation
system to comply with industrial or latest medical safety standards.
Model List
Model
Input
Output
Number
Voltage
Voltage
(Range)
VDC
DT10S1205A
DT10S12051A
DT10S1212A
12
Output Current
Input Current
Ripple
Max.
@Max. Load @No Load
VDC
mA
mA(typ.)
5
1600
877
5.1
1600
907
12
835
1044
±12
±417
1042
±15
±333
1028
DT10S2405A
5
2000
541
DT10S24051A
5.1
2000
559
12
835
516
DT10D1212A
(9 ~ 18)
DT10D1215A
DT10S2412A
DT10D2412A
24
(18 ~ 36)
±12
±417
516
DT10D2415A
±15
±333
508
DT10S4805A
5
2000
271
5.1
2000
280
12
835
258
±12
±417
258
±15
±333
254
DT10S48051A
DT10S4812A
DT10D4812A
DT10D4815A
48
(36 ~ 75)
Reflected Max. capacitive
mA (typ.)
Load
Current
mA(typ.)
100
μF
470
220#
1000
20
50
470
220#
1000
10
25
(typ.)
@Max. Load
1000
30
Efficiency
470
220#
%
76
75
80
80
81
77
76
81
81
82
77
76
81
81
82
# For each output
DS_DT10S/D_12092015
E-mail: [email protected]
http://www.deltaww.com/dcdc
P1
Input Specifications
Parameter
Input Surge Voltage (1 sec. max.)
Start-Up Threshold Voltage
Under Voltage Shutdown
Short Circuit Input Power
Internal Power Dissipation
Conducted EMI
Model
12V Input Models
24V Input Models
48V Input Models
Min.
-0.7
-0.7
-0.7
Typ.
Max.
Unit
25
--50
--100
--7
8
9
13
15
18
VDC
30
33
36
----8.5
----16
----34
----3000
mW
----4000
mW
Compliance to EN 55022,class A and FCC part 15,class A
12V Input Models
24V Input Models
48V Input Models
12V Input Models
24V Input Models
48V Input Models
All Models
Output Specifications
Conditions
Min.
Typ.
Max.
Unit
Output Voltage Setting Accuracy
Parameter
At 50% Load and Nominal Vin
---
---
±1.0
%Vnom.
Output Voltage Balance
Dual Output, Balanced Loads
---
±0.5
±2.0
%
Vin=Min. to Max.
---
±0.3
±0.5
%
Io=15% to 100%
---
±0.5
±1.0
%
Io=5% to 100%
---
±0.6
±1.2
%
5V & 5.1V Output Models
---
75
100
mV P-P
---
100
150
mV P-P
Line Regulation
Load Regulation
Ripple & Noise (20MHz)
Other Output Models
Min.Load
No minimum Load Requirement
Over Load Protection
Transient Recovery Time
Transient Response Deviation
25% Load Step Change
Temperature Coefficient
120
150
---
%
---
300
600
μsec
---
±3
±5
%
---
±0.02
±0.05
%/℃
Short Circuit Protection
Continuous
Isolation, Safety Standards
Parameter
Conditions
Min.
Typ.
Max.
Unit
60 Seconds
4200
---
---
VACrms
Flash tested for 1 Second
6000
---
---
VPK
240VAC, 60Hz
---
---
10
μA
I/O Isolation Resistance
500 VDC
10
---
---
GΩ
I/O Isolation Capacitance
100KHz, 1V
--60
80
cUL/UL60950-1, CSA C22.2 No. 60950-1-03
UL60601-1,CSA C22.2 No.601-1,
IEC/EN 60950-1, IEC/EN 60601-1 3rd Edition, 2 MOOP
IEC60950-1 CB report, cUL/UL 60950-1 certificate
UL60601-1 UL certificate
pF
I/O Isolation Voltage (reinforced)
I/O Isolation Test Voltage
Leakage Current
Safety Standards
Approvals(Pending)
General Specifications
Parameter
Conditions
Switching Frequency
MTBF(calculated)
MIL-HDBK-217F@25℃, Ground
Benign
Min.
Typ.
Max.
Unit
120
150
180
KHz
1,000,000
---
---
Hours
Input Fuse
12V Input Models
24V Input Models
48V Input Models
3000mA Slow-Blow Type
1500mA Slow-Blow Type
750mA Slow-Blow Type
DS_DT10S/D_12092015
E-mail: [email protected]
http://www.deltaww.com/dcdc
P2
Environmental Specifications
Parameter
Operating Ambient Temperature Range (See Power
Derating Curve)
Case Temperature
Storage Temperature Range
Humidity (non condensing)
Altitude
Cooling
Lead Temperature (1.5mm from case for 10Sec.)
Conditions
Natural Convection
Min.
-40
Max.
+75
--+95
-50
+125
--95
--4000
Free-Air convection
--260
Unit
℃
℃
℃
% rel. H
m
℃
Power Derating Curve
100
Output Power (%)
80
Natural
Convection
20LFM
60
40
20
0
~
-40
0
20
40
60
80
100 110
Ambient Temperature C
Notes
1
2
3
4
5
6
7
Specifications typical at Ta=+25℃, resistive load, nominal input voltage and rated output current unless otherwise noted.
Transient recovery time is measured to within 1% error band for a step change in output load of 75% to 100%.
Ripple & Noise measurement bandwidth is 0-20MHz.
All DC/DC converters should be externally fused at the front end for protection.
Other input and output voltage may be available, please contact factory.
That “natural convection” is about 20LFM but is not equal to still air (0 LFM).
Specifications are subject to change without notice.
DS_DT10S/D_12092015
E-mail: [email protected]
http://www.deltaww.com/dcdc
P3
Mechancial Drawing
Mechanical Dimensions
Pin Connections
5.08
[0.20]
5.08
[0.20]
Bottom View
5
4
Pin
Single Output
Dual Output
1
+Vin
+Vin
2
-Vin
-Vin
3
+Vout
+Vout
4
No Pin
Common
5
-Vout
-Vout
50.8 [2.00]
1.00 [ 0.04]
1
45.72 [1.80]
2
12.0
[0.47]
►All dimensions in mm (inches)
3
►Tolerance: X.X±0.25 (X.XX±0.01)
10.16
[0.40]
10.16
[0.40]
2.54
[0.10]
X.XX±0.13 ( X.XXX±0.005)
►Pin diameter  1.0 ±0.05 (0.04±0.002)
25.4 [1.00]
Physical Outline
Case Size
:
50.8x25.4x12.0mm (2.0x1.0x0.47 inches)
Case Material
:
Non-Conductive Black Plastic (flammability to UL 94V-0 rated)
Pin Material
:
Copper Alloy with Gold Plate Over Nickel Subplate
Weight
:
24.5g
DS_DT10S/D_12092015
E-mail: [email protected]
http://www.deltaww.com/dcdc
P4
Part Numbering System
D
T
10
S
12
05
A
Form factor
Family series
Watt
Number of
Outputs
Input Voltage
Output Voltage
Option Code
D - DIP
P - SIP
S - SMD
A~Z
10 - 10W
12 - 12V
24 - 24V
48 - 48V
05 - 5V
051 - 5.1V
12 - 12V
15 - 15V
A - Std. Functions
S - Single
D - Dual
WARRANTY
Delta offers a three(3) years limited warranty. Complete warranty information is listed on our web site or is available upon request from Delta.
Information furnished by Delta is believed to be accurate and reliable. However, no responsibility is assumed by Delta for its use, nor for any
infringements of patents or other rights of third parties, which may result from its use. No license is granted by implication or otherwise under
any patent or patent rights of Delta. Delta reserves the right to revise these specifications at any time, without notice.
DS_DT10S/D_12092015
E-mail: [email protected]
http://www.deltaww.com/dcdc
P5