A4 LED eye safety(SAW84A0C)

Test Report issued under the responsibility of:
TEST REPORT
IEC 62471
Photobiological safety of lamps and lamp systems
Report Reference No......................: 10-1382-577-2
Date of issue.....................................: December 03, 2010
Total number of pages ....................: 19 pages
CB Testing Laboratory...................: KTL (Korea Testing Laboratory)
Address ............................................: 222-13, Guro-3dong, Guro-gu, Seoul, 152-718, Korea
Applicant’s name............................: SEOUL SEMICONDUCTOR
Address ............................................: 1B-36,727-5,WONSI-DONG,DANWON-GU,ANSAN-CITY,
KYUNGGI-DO,KOREA
Test specification:
Standard ...........................................: IEC 62471:2006 (First Edition) + European Group Difference and
National Differences
Test procedure .................................: N/A
Non-standard test method…………..: N/A
Test Report Form No......................: IEC62471A
TRF Originator..................................: VDE Testing and Certification Institute
Master TRF.......................................: Dated 2009-05
Copyright © 2009 IEC System for Conformity Testing and Certification of Electrical Equipment (IECEE),
Geneva, Switzerland. All rights reserved.
This publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged as
copyright owner and source of the material. IECEE takes no responsibility for and will not assume liability for damages resulting from
the reader's interpretation of the reproduced material due to its placement and context.
If this Test Report Form is used by non-IECEE members, the IECEE/IEC logo and the reference to the CB Scheme procedure shall be
removed.
This report is not valid as a CB Test Report unless signed by an approved CB Testing Laboratory
and appended to a CB Test Certificate issued by an NCB in accordance with IECEE 02.
Test item description .....................: LED Package
Trade Mark .......................................: Manufacturer ....................................: Same as the applicant
Model/Type reference ......................: AN4240
Ratings .............................................: AC 55V~, 20 mA (Exempt Group)
TRF No. IEC62471A
Page 2 of 19
Report No. : 10-1382-577-2
Testing procedure and testing location:
CB Testing Laboratory:
Testing location/ address.......................:
KTL (Korea Testing Laboratory)
222-13, Guro-3dong, Guro-gu, Seoul, 152-718, Korea
Associated CB Laboratory:
Testing location/ address.......................:
Tested by (name + signature) .....:
Jae-Hoon Jung
...............................
Approved by (+ signature) ............:
Jong-Koo Park
...............................
Testing procedure: TMP
Tested by (name + signature) .....:
Approved by (+ signature) ............:
Testing location/ address.......................:
Testing procedure: WMT
Tested by (name + signature) .....:
Witnessed by (+ signature)...........:
Approved by (+ signature) ...........:
Testing location/ address.......................:
Testing procedure: SMT
Tested by (name + signature) .....:
Approved by (+ signature) ............:
Supervised by (+ signature) ........:
Testing location/ address................ :
Testing procedure: RMT
Tested by (name + signature) .....:
Approved by (+ signature) ............:
Supervised by (+ signature) ........:
Testing location/ address.......................:
TRF No. IEC62471A
Page 3 of 19
Report No. : 10-1382-577-2
Summary of testing:
Tests performed (name of test and test clause):
Testing location:
All clauses
KTL (Korea Testing Laboratory)
222-13, Guro-3dong, Guro-gu, Seoul, 152-718,
Korea
Summary of compliance with National Differences:
The sample(s) tested complies with the requirements of 62471:2006 Compliance with European Group
Difference and National Differences, ATTACHMENT is recorded at the end of this report.
Copy of marking plate:
-
Item covered
We performed the test at Maximum voltage, AC63V
TRF No. IEC62471A
Page 4 of 19
Report No. : 10-1382-577-2
Test item particulars ..................................................:
Tested lamp ................................................................:
continuous wave lamps
pulsed lamps
Tested lamp system .....................................................: Lamp classification group.............................................:
exempt
risk 1
risk 2
risk 3
Lamp cap ....................................................................: Bulb.............................................................................: LED
Rated of the lamp .........................................................: Furthermore marking on the lamp................................: Seasoning of lamps according IEC standard ..............: Used measurement instrument....................................: See page 15
Temperature by measurement.....................................: 23.1 °C
Information for safety use .............................................: Not required
Possible test case verdicts:
–
test case does not apply to the test object ............ : N/A
–
test object does meet the requirement .................. : P (Pass)
–
test object does not meet the requirement ............ : F (Fail)
Testing:
Date of receipt of test item............................................ : 2010-11-04
Date (s) of performance of tests................................... : 2010-11-20until 2010-12-02
General remarks:
The test results presented in this report relate only to the object tested.
This report shall not be reproduced, except in full, without the written approval of the Issuing testing laboratory.
"(See Enclosure #)" refers to additional information appended to the report.
"(See appended table)" refers to a table appended to the report.
Throughout this report a comma (point) is used as the decimal separator.
List of test equipment must be kept on file and available for review.
General product information:
<Electro-Optical characteristics at 55 V[rms]>
-
CCT = 3000 K
CRI = 85
PD = 0.76 W
<Absolute maximum ratings>
-
Vopt = 63 V[rms]
Topr = -30 ~ +85 ℃
-
Tstg = -40 ~ +120 ℃
TRF No. IEC62471A
Page 5 of 19
Report No. : 10-1382-577-2
IEC 62471
Clause
Requirement + Test
Result – Remark
4
EXPOSURE LIMITS
4.1
General
P
The exposure limits in this standard is not less than
0,01 ms and not more than any 8-hour period and
should be used as guides in the control of exposure
P
Detailed spectral data of a light source are generally see clause 4.3
required only if the luminance of the source exceeds
104 cd.m-2
P
4.3
Hazard exposure limits
P
4.3.1
Actinic UV hazard exposure limit for the skin and
eye
No UV hazard
Verdict
N/A
The exposure limit for effective radiant exposure is
30 J.m-2 within any 8-hour period
N/A
To protect against injury of the eye or skin from
ultraviolet radiation exposure produced by a
broadband source, the effective integrated spectral
irradiance , ES, of the light source shall not exceed
the levels defined by:
N/A
N/A
The permissible time for exposure to ultraviolet
radiation incident upon the unprotected eye or skin
shall be computed by:
N/A
N/A
4.3.2
Near-UV hazard exposure limit for eye
N/A
For the spectral region 315 nm to 400 nm (UV-A)
No Near-UV hazard
the total radiant exposure to the eye shall not
exceed 10000 J.m-2 for exposure times less than
1000 s. For exposure times greater than 1000 s
(approximately 16 minutes) the UV-A irradiance for
the unprotected eye, EUVA, shall not exceed 10 W.m2
.
N/A
The permissible time for exposure to ultraviolet
radiation incident upon the unprotected eye for time
less than 1000 s, shall be computed by:
N/A
N/A
4.3.3
Retinal blue light hazard exposure limit
N/A
To protect against retinal photochemical injury from
chronic blue-light exposure, the integrated spectral
radiance of the light source weighted against the
blue-light hazard function, B(λ), i.e., the blue-light
N/A
TRF No. IEC62471A
Page 6 of 19
Report No. : 10-1382-577-2
IEC 62471
Clause
Requirement + Test
Result – Remark
Verdict
weighted radiance , LB, shall not exceed the levels
defined by:
4.3.4
for t ≤ 104 s
N/A
for t > 104 s
N/A
Retinal blue light hazard exposure limit - small source
P
Thus the spectral irradiance at the eye Eλ, weighted see table 4.2
against the blue-light hazard function B(λ) shall not
exceed the levels defined by:
P
P
P
4.3.5
Retinal thermal hazard exposure limit
P
To protect against retinal thermal injury, the
see table 4.2
integrated spectral radiance of the light source, Lλ,
weighted by the burn hazard weighting function R(λ)
(from Figure 4.2 and Table 4.2), i.e., the burn hazard
weighted radiance, shall not exceed the levels
defined by:
P
(10 μs ≤ t ≤ 10 s)
4.3.6
P
Retinal thermal hazard exposure limit – weak visual stimulus
N/A
For an infrared heat lamp or any near-infrared
source where a weak visual stimulus is inadequate
to activate the aversion response, the near infrared
(780 nm to 1400 nm) radiance, LIR, as viewed by the
eye for exposure times greater than 10 s shall be
limited to:
N/A
N/A
4.3.7
Infrared radiation hazard exposure limits for the eye
N/A
The avoid thermal injury of the cornea and possible
delayed effects upon the lens of the eye
(cataractogenesis), ocular exposure to infrared
radiation, EIR, over the wavelength range 780 nm to
3000 nm, for times less than 1000 s, shall not
exceed:
N/A
TRF No. IEC62471A
Page 7 of 19
Report No. : 10-1382-577-2
IEC 62471
Clause
Requirement + Test
Result – Remark
Verdict
N/A
For times greater than 1000 s the limit becomes:
N/A
N/A
4.3.8
Thermal hazard exposure limit for the skin
N/A
Visible and infrared radiant exposure (380 nm to
3000 nm) of the skin shall be limited to:
N/A
N/A
5
MEASUREMENT OF LAMPS AND LAMP SYSTEMS
P
5.1
Measurement conditions
P
Measurement conditions shall be reported as part of
the evaluation against the exposure limits and the
assignment of risk classification.
P
5.1.1
5.1.2
5.1.3
5.1.4
5.1.5
Lamp ageing (seasoning)
Not lamps
N/A
Seasoning of lamps shall be done as stated in the
appropriate IEC lamp standard.
N/A
Test environment
N/A
For specific test conditions, see the appropriate IEC
lamp standard or in absence of such standards, the
appropriate national standards or manufacturer’s
recommendations.
N/A
Extraneous radiation
No extraneous radiation
N/A
Careful checks should be made to ensure that
extraneous sources of radiation and reflections do
not add significantly to the measurement results.
N/A
Lamp operation
N/A
Operation of the test lamp shall be provided in
accordance with:
N/A
–
the appropriate IEC lamp standard, or
N/A
–
the manufacturer’ s recommendation
N/A
Lamp system operation
Not lamp system
N/A
The power source for operation of the test lamp
shall be provided in accordance with:
N/A
–
the appropriate IEC standard, or
N/A
–
the manufacturer’ s recommendation
N/A
TRF No. IEC62471A
Page 8 of 19
Report No. : 10-1382-577-2
IEC 62471
Clause
Requirement + Test
Result – Remark
Verdict
5.2
Measurement procedure
P
5.2.1
Irradiance measurements
N/A
Minimum aperture diameter 7mm.
N/A
Maximum aperture diameter 50 mm.
N/A
The measurement shall be made in that position of
the beam giving the maximum reading.
N/A
The measurement instrument is adequate
calibrated.
N/A
5.2.2
Radiance measurements
5.2.2.1
Standard method
N/A
The measurements made with an optical system.
N/A
The instrument shall be calibrated to read in
absolute radiant power per unit receiving area and
per unit solid angle to acceptance averaged over the
field of view of the instrument.
N/A
5.2.2.2
5.2.3
P
Alternative method
P
Alternatively to an imaging radiance set-up, an
irradiance measurement set-up with a circular field
stop placed at the source can be used to perform
radiance measurements.
P
Measurement of source size
27.76 mrad
The determination of α, the angle subtended by a
source, requires the determination of the 50%
emission points of the source.
5.2.4
Pulse width measurement for pulsed sources
P
P
Continuous sources
The determination of Δt, the nominal pulse duration
of a source, requires the determination of the time
during which the emission is > 50% of its peak
value.
N/A
N/A
5.3
Analysis methods
P
5.3.1
Weighting curve interpolations
P
To standardize interpolated values, use linear
interpolation on the log of given values to obtain
intermediate points at the wavelength intervals
desired.
5.3.2
5.3.3
see table 4.1
P
Calculations
P
The calculation of source hazard values shall be
performed by weighting the spectral scan by the
appropriate function and calculating the total
weighted energy.
P
Measurement uncertainty
TRF No. IEC62471A
N/A
Page 9 of 19
Report No. : 10-1382-577-2
IEC 62471
Clause
Requirement + Test
Result – Remark
The quality of all measurement results must be
quantified by an analysis of the uncertainty.
6
Verdict
N/A
LAMP CLASSIFICATION
For the purposes of this standard it was decided that see table 6.1
the values shall be reported as follows:
–
for lamps intended for general lighting service,
the hazard values shall be reported as either
irradiance or radiance values at a distance
which produces an illuminance of 500 lux, but
not at a distance less than 200 mm
–
for all other light sources, including pulsed lamp 200 mm used
sources, the hazard values shall be reported at
a distance of 200 mm
P
N/A
P
6.1
Continuous wave lamps
P
6.1.1
Except Group
P
In the except group are lamps, which does not pose See table 6.1
any photobiological hazard. The requirement is met
by any lamp that does not pose:
P
6.1.2
–
an actinic ultraviolet hazard (ES) within 8-hours
exposure (30000 s), nor
N/A
–
a near-UV hazard (EUVA) within 1000 s, (about
16 min), nor
N/A
–
a retinal blue-light hazard (LB) within 10000 s
(about 2,8 h), nor
P
–
a retinal thermal hazard (LR) within 10 s, nor
P
–
an infrared radiation hazard for the eye (EIR)
within 1000 s
N/A
Risk Group 1 (Low-Risk)
N/A
In this group are lamps, which exceeds the limits for
the except group but that does not pose:
N/A
–
an actinic ultraviolet hazard (ES) within 10000 s,
nor
N/A
–
a near ultraviolet hazard (EUVA) within 300 s, nor
N/A
–
a retinal blue-light hazard (LB) within 100 s, nor
N/A
–
a retinal thermal hazard (LR) within 10 s, nor
N/A
–
an infrared radiation hazard for the eye (EIR)
within 100 s
N/A
Lamps that emit infrared radiation without a strong
visual stimulus and do not pose a near-infrared
retinal hazard (LIR), within 100 s are in Risk Group 1.
N/A
TRF No. IEC62471A
Page 10 of 19
Report No. : 10-1382-577-2
IEC 62471
Clause
Requirement + Test
6.1.3
Risk Group 2 (Moderate-Risk)
This requirement is met by any lamp that exceeds
the limits for Risk Group 1, but that does not pose:
6.1.4
6.2
Result – Remark
Verdict
N/A
See table 6.1
N/A
–
an actinic ultraviolet hazard (ES) within 1000 s
exposure, nor
N/A
–
a near ultraviolet hazard (EUVA) within 100 s, nor
N/A
–
a retinal blue-light hazard (LB) within 0,25 s
(aversion response), nor
N/A
–
a retinal thermal hazard (LR) within 0,25 s
(aversion response), nor
N/A
–
an infrared radiation hazard for the eye (EIR)
within 10 s
N/A
Lamps that emit infrared radiation without a strong
visual stimulus and do not pose a near-infrared
retinal hazard (LIR), within 10 s are in Risk Group 2.
N/A
Risk Group 3 (High-Risk)
N/A
Lamps which exceed the limits for Risk Group 2 are
in Group 3.
N/A
Pulsed lamps
N/A
Pulse lamp criteria shall apply to a single pulse and
to any group of pulses within 0,25 s.
N/A
A pulsed lamp shall be evaluated at the highest
nominal energy loading as specified by the
manufacturer.
N/A
The risk group determination of the lamp being
tested shall be made as follows:
N/A
–
a lamp that exceeds the exposure limit shall be
classified as belonging to Risk Group 3 (HighRisk)
N/A
–
for single pulsed lamps, a lamp whose weighted
radiant exposure or weighted radiance does is
below the EL shall be classified as belonging to
the Exempt Group
N/A
–
for repetitively pulsed lamps, a lamp whose
weighted radiant exposure or weighted radiance
dose is below the EL, shall be evaluated using
the continuous wave risk criteria discussed in
clause 6.1, using time averaged values of the
pulsed emission
N/A
TRF No. IEC62471A
Page 11 of 19
Report No. : 10-1382-577-2
IEC 62471
Clause
Requirement + Test
Result – Remark
Verdict
Table 4.1
Spectral weighting function for assessing ultraviolet hazards for skin and eye
N/A
Wavelength1
λ, nm
UV hazard function
SUV(λ)
Wavelength
λ, nm
UV hazard function
SUV(λ)
200
0,030
313*
0,006
205
0,051
315
0,003
210
0,075
316
0,0024
215
0,095
317
0,0020
220
0,120
318
0,0016
225
0,150
319
0,0012
230
0,190
320
0,0010
235
0,240
322
0,00067
240
0,300
323
0,00054
245
0,360
325
0,00050
250
0,430
328
0,00044
254*
0,500
330
0,00041
255
0,520
333*
0,00037
260
0,650
335
0,00034
265
0,810
340
0,00028
270
1,000
345
0,00024
275
0,960
350
0,00020
280*
0,880
355
0,00016
285
0,770
360
0,00013
290
0,640
365*
0,00011
295
0,540
370
0,000093
297*
0,460
375
0,000077
300
0,300
380
0,000064
303*
0,120
385
0,000053
305
0,060
390
0,000044
308
0,026
395
0,000036
310
0,015
400
0,000030
Wavelengths chosen are representative: other values should be obtained by logarithmic interpolation at
intermediate wavelengths.
* Emission lines of a mercury discharge spectrum.
1
TRF No. IEC62471A
Page 12 of 19
Report No. : 10-1382-577-2
IEC 62471
Clause
Requirement + Test
Table 4.2
Spectral weighting functions for assessing retinal hazards from broadband optical
sources
Wavelength
nm
300
305
310
315
320
325
330
335
340
345
350
355
360
365
370
375
380
385
390
395
400
405
410
415
420
425
430
435
440
445
450
455
460
465
470
475
480
485
490
495
500-600
600-700
700-1050
1050-1150
1150-1200
1200-1400
TRF No. IEC62471A
Result – Remark
Blue-light hazard function
B (λ)
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,013
0,025
0,05
0,10
0,20
0,40
0,80
0,90
0,95
0,98
1,00
1,00
0,97
0,94
0,90
0,80
0,70
0,62
0,55
0,45
0,40
0,22
0,16
10[(450-λ)/50]
0,001
Verdict
Burn hazard function
R (λ)
0,1
0,13
0,25
0,5
1,0
2,0
4,0
8,0
9,0
9,5
9,8
10,0
10,0
9,7
9,4
9,0
8,0
7,0
6,2
5,5
4,5
4,0
2,2
1,6
1,0
1,0
10[(700-λ)/500]
0,2
0,2.100,02(1150-λ)
0,02
P
Page 13 of 19
Report No. : 10-1382-577-2
IEC 62471
Clause
Requirement + Test
Table 5.4
Summary of the ELs for the surface of the skin or cornea (irradiance based values)
Hazard
Name
Relevant equation
Result – Remark
Verdict
N/A
Wavelength
range
nm
Exposure
duration
sec
Limiting
aperture
rad (deg)
EL in terms of
constant irradiance
W•m-2
Actinic UV
skin & eye
ES = ∑Eλ • S(λ) • ∆λ
200 – 400
< 30000
1,4 (80)
30/t
Eye UV-A
EUVA = ∑Eλ • ∆λ
315 – 400
≤1000
>1000
1,4 (80)
10000/t
10
Blue-light
small source
EB = ∑Eλ • B(λ) • ∆λ
300 – 700
≤100
>100
< 0,011
100/t
1,0
Eye IR
EIR = ∑Eλ • ∆λ
780 –3000
≤1000
>1000
1,4 (80)
18000/t 0,75
100
Skin thermal
EH = ∑Eλ • ∆λ
380 – 3000
< 10
2π sr
20000/t 0,75
Table 5.5
Summary of the ELs for the retina (radiance based values)
Hazard Name
Relevant equation
P
Wavelength
range
nm
Exposure
duration
sec
Field of view
radians
EL in terms of
constant radiance
W•m-2•sr-1)
0,011•√(t/10)
0,011
0,0011•√t
0,11
106/t
106/t
106/t
100
Blue light
LB = ∑Lλ • B(λ) • ∆λ
300 – 700
0,25 – 10
10-100
100-10000
≥ 10000
Retinal
thermal
LR = ∑Lλ • R(λ) • ∆λ
380 – 1400
< 0,25
0,25 – 10
0,0017
0,011•√(t/10)
50000/(α•t 0,25)
50000/(α•t 0,25)
Retinal
thermal
(weak visual
stimulus)
LIR = ∑Lλ • R(λ) • ∆λ
780 – 1400
> 10
0,011
6000/α
TRF No. IEC62471A
Page 14 of 19
Report No. : 10-1382-577-2
IEC 62471
Clause
Requirement + Test
Table 6.1
Emission limits for risk groups of continuous wave lamps
Risk
Actinic UV
Action
spectrum
SUV(λ)
Near UV
Result – Remark
Verdict
P
Emission Measurement
Symbol
Es
EUVA
Units
Exempt
-2
W•m
-2
W•m
-2
-1
Low risk
Mod risk
Limit
Result
Limit
Result
Limit
Result
0,001
-
0,003
-
0,03
-
10
-
33
-
100
-
100
29.2
10000
-
4000000
-
Blue light
B(λ)
LB
W•m •sr
Blue light,
small source
B(λ)
EB
W•m-2
1,0*
-
1,0
-
400
-
Retinal
thermal
R(λ)
LR
W•m-2•sr-1
28000/α
= 1008646
11006.2
28000/α
-
71000/α
-
Retinal
thermal,
weak visual
stimulus**
R(λ)
LIR
W•m-2•sr-1
6000/α
-
6000/α
-
6000/α
-
EIR
W•m-2
100
-
570
-
3200
-
IR radiation,
eye
*
**
TRF No. IEC62471A
Small source defined as one with α < 0,011 radian. Averaging field of view at 10000 s is 0,1 radian.
Involves evaluation of non-GLS source
Page 15 of 19
Furthermore remarks:
- Spectral irradiance for Blue light hazard
- Spectral irradiance for Retinal thermal hazard
- Measuring Instrument
Name
Maker
Model No.
Spectrometer
OceanOptics
HR4000
Illuminance meter
Gigahertz-optik
X1
TRF No. IEC62471A
Report No. : 10-1382-577-2
Page 16 of 19
Report No. : 10-1382-577-2
.:
IEC62471A - ATTACHMENT
Clause
Requirement + Test
Result - Remark
Verdict
ATTACHMENT TO TEST REPORT IEC 62471
EUROPEAN GROUP DIFFERENCES AND NATIONAL DIFFERENCES
Photobiological safety of lamps and lamps systems
Differences according to................: EN 62471:2008
Attachment Form No......................: EU_GD_IEC62471A
Attachment Originator ...................: IMQ S.p.A.
Master Attachment .........................: 2009-07
Copyright © 2009 IEC System for Conformity Testing and Certification of Electrical Equipment
(IECEE), Geneva, Switzerland. All rights reserved.
4
4.1
CENELEC COMMON MODIFICATIONS (EN)
EXPOSURE LIMITS
Contents of the whole Clause 4 of IEC 62471:2006
moved into a new informative Annex ZB
Clause 4 replaced by the following:
Limits of the Artificial Optical Radiation Directive
(2006/25/EC) have been applied instead of those
fixed in IEC 62471:2006
General
First paragraph deleted
TRF No. IEC62471A
⎯
See appended Table 6.1
P
Noted
⎯
Page 17 of 19
Report No. : 10-1382-577-2
IEC62471A - ATTACHMENT
Clause
Table 6.1
Requirement + Test
Result - Remark
Verdict
Emission limits for risk groups of continuous wave lamps (based on EU Directive 2006/25/EC)
P
Emission Measurement
Risk
Actinic UV
Action
spectrum
SUV(λ)
Near UV
Symbol
Units
Exempt
Low risk
Mod risk
Limit
Result
Limit
Result
Limit
Result
Es
W•m-2
0,001
-
-
-
-
-
EUVA
W•m-2
0,33
-
-
-
-
-
Blue light
B(λ)
LB
W•m-2•sr-1
100
29.2
10000
-
4000000
-
Blue light,
small source
B(λ)
EB
W•m-2
0,01*
-
1,0
-
400
-
Retinal
thermal
R(λ)
LR
W•m-2•sr-1
28000/α =
1008646
11006.2
28000/α
-
71000/α
-
3200
-
545000
Retinal
thermal,
weak visual
stimulus**
R(λ)
LIR
W•m-2•sr-1
-
0,0017≤ α ≤ 0,011
6000/α
-
0,011≤ α ≤ 0,1
IR radiation,
eye
TRF No. IEC62471A
EIR
W•m-2
100
-
570
-
Page 18 of 19
Report No. : 10-1382-577-2
IEC62471A - ATTACHMENT
Clause
*
**
NOTE
Requirement + Test
Result - Remark
Small source defined as one with α < 0,011 radian. Averaging field of view at 10000 s is 0,1 radian.
Involves evaluation of non-GLS source
The action functions: see Table 4.1 and Table 4.2
The applicable aperture diameters: see 4.2.1
The limitations for the angular subtenses: see 4.2.2
The related measurement condition 5.2.3 and the range of acceptance angles: see Table 5.5.
TRF No. IEC62471A
Verdict
Page 19 of 19
Photographs
<Fig. 1>
TRF No. IEC62471A
Report No. : 10-1382-577-2