English Datasheet

Oyster® Multifunction Process Calibrator
Precision Source and Measure
For thermocouples (Types J, K, and T), mA, mV, and V devices
Features:
• Source 0 to 24mA with 0.01mA resolution
• Measure 0 to 50mA (0.1mA resolution)
• Select Type J, K, or T thermocouple for precision
calibrations in °F/°C or mV
• Output 0 to 1999mV or 0 to 10.00V
• Measure 0 to 19.99V
• Single/continuous step function for Voltage and Current
• Self-contained 24V loop power supply drives
current loads up to 1000Ω
• Memory for 5 user settable output values
of each type for fast calibrations
Complete with six AA batteries, test leads
with detachable alligator clips, temperature
calibration cable with spade lug connector,
carrying case, neckstrap, and 100V-240V
AC Universal adaptor (50/60Hz)
Specifications
DC Source
Current
Voltage
Type J
Type K
Type T
Loop Power
Max. Load
DC Measure
Current
Voltage
Type J
Type K
Type T
Dimensions/Weight
Range
Accuracy (of reading)
0 to 24mA, -25% to +125%
0 to 10V
-58 to 1830°F (-50 to 1000°C)
-58 to 2498°F (-50 to 1370°C)
-184 to 752°F (-120 to 400°C)
24V
1000 @ 24mA
±(0.075% + 1 digit) or ± 3 digits, whichever is greater
±(0.075% + 1 digit) or ± 3 digits, whichever is greater
±(0.15% + 1.8°F or 1°C)
±(0.15% + 1.8°F or 1°C)
±(0.15% + 1.8°F or 1°C)
±(10µV ± 1 digit)
0 to 50mA, -25% to +230%
±(0.075% + 1 digit) or ± 3 digits, whichever is greater
0 to 20V
±(0.075% + 1 digit) or ± 3 digits, whichever is greater
-58 to 1830°F (-50 to 1000°C)
±(0.15% + 1.8°F or 1°C)
-58 to 2498°F (-50 to 1370°C)
±(0.15% + 1.8°F or 1°C)
-184 to 752°F (-120 to 400°C)
±(0.15% + 1.8°F or 1°C)
3.8 x 4.7 x 2.0" (96 x 118 x 45mm)/12oz (340g)
Ordering Information:
412400 ......................Multifunction Calibrator
412400-NIST ............412400 with NIST Certificate
www.extech.com
Specifications subject to change without notice.
Copyright © 2007-2014 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form.