User Guide Indoor Air Quality Meter/Datalogger Model EA80 Introduction Congratulations on your purchase of the Extech EA80 Indoor Air Quality Meter. This meter measures Carbon Dioxide (CO2, ppm) levels, ambient Temperature and Relative Humidity (%RH). 16,200 readings can be logged by the meter and later transferred to a PC. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Description 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. Display Gas exhaust Gas inlet port PC Interface jack AC adapter socket (9V, 300mA) Power Button Data Hold Button Backlight Button MX/MN Button SET Button ALARM button MEM (memory) button (Enter) button READ button TIME button °C/°F button DEW button WET button Temperature / Humidity sensor probe AC Adaptor 2.5mm mono Plug Ø2.5 Three Tier LCD Display 2 EA80-en-GB_v3.3 09/15 Preparation for Use Power Supply The meter is powered by six (6) 1.5V ‘AAA’ alkaline batteries or by an AC adapter. Installing the Batteries Insert six (6) AAA batteries as indicated by the diagram located on the inside of the battery compartment. When the battery voltage drops below the operating voltage, the “ displayed indicating that the batteries need to be changed. + ” indicator will be AC Adapter The AC adapter allows the meter to be powered from a common AC wall outlet. When using the AC adapter, the batteries (if installed) will be bypassed. The AC adapter is not a battery charger. Gas Inlet Always ensure that the meter’s gas inlet port and gas exhaust are not blocked. Operation Note: Exhaled CO2 will affect the accuracy of the reading; do not hold meter near the face. Taking Measurements The sensor for Temperature, Humidity, Dew Point, and Wet Bulb measurements is located in the remote probe. Hold the probe in the air in the area to be tested. DO NOT immerse in liquid. The sensor for CO2 measurements is located at the top of the meter. Selecting temperature units of measure (C/F) Press the °C/°F button momentarily to toggle the temperature units. CO2 Measurement button to turn on the meter, 1. Press 2. The sensor requires a 30 second warm-up before displaying the CO2 measurement. 3. The meter sensor requires approximately10 minutes to stabilize in still air before the readings can be considered accurate. Moving the meter may decrease this stabilization time. 4. The primary display indicates the CO2 reading. The secondary display indicates temperature. The third display indicates %RH. 3 EA80-en-GB_v3.3 09/15 Humidity Measurement 1. Press the “ 2. The display will indicate the humidity reading (% RH) directly on the third display. 3. Hold the probe in the air in the area to be tested. Do NOT immerse in liquid. ” button to power the meter ON. 4. Allow adequate time for readings to stabilize. 5. Read the measurements on the LCD. Temperature Measurement 1. Press the “ 2. Press the °C/°F button momentarily to toggle the temperature units. ” button to power the meter ON. 3. The display will show the Temperature reading directly on the second display. 4. Allow adequate time for readings to stabilize. 5. Read the measurements on the LCD. Dew Point Temperature Measurement 1. Press the “ 2. Press the “DEW” button 3. The display will show the Dew Point Reading on the second display. ” button to power the meter ON. 4. Allow adequate time for the readings to stabilize 5. Read the measurements on the LCD. 6. Press the “DEW” button again to exit the dew point temperature reading. Wet Bulb Temperature Measurement 1. Press the “ 2. Press the “WET” button ” button to power the meter ON. 3. The display will show the Wet Bulb Reading on the second display. 4. Allow adequate time for the readings to stabilize 5. Read the measurements on the LCD. 6. Press the “WET” button again to exit the display. 4 EA80-en-GB_v3.3 09/15 CO2 Maximum and Minimum Recording Measurement 1. 2. 3. Press “MX/MN” button to enter the maximum / minimum recording mode, the “MAX” icon appears on the display. The maximum CO2 reading will be displayed and will be updated only when a higher reading occurs. MAX Press “TIME” button to enter time display mode. Displays 2 and 3 will indicate the time the maximum reading occurred. Press “MX/MN” button again. The “△” icon will appear with the current value and current time. 5. Press “MX/MN” button again cycle through the recorded MAX, MIN and current readings. 6. Press “” button to exit this mode. 7. Press “TIME” button exit the time display mode. CO2 ppm d-h m:s Press “MX/MN” button again. The “MIN” icon will appear with the minimum value and its stamp time. 4. 8 MIN CO2 ppm d-h m:s CO2 ppm d-h m:s Data Hold 1. 2. Press the H button momentarily to freeze the displayed reading. The 'H' icon will appear on the upper left-hand side of the display. Press the H button again to return to normal operation (the 'H' hold icon will disappear). SET MAX MIN ALM R M 8888 88 Err INTV 8888 CO2 m:s DEW WET °C°F ppm CO2 CAL ppm 8888 CO2 d-h %RH ppm Backlight 1. Press the 2. The backlight will turn off automatically after 30 seconds. button to turn the backlight on or off. 5 EA80-en-GB_v3.3 09/15 CO2 Alarm Operation Setting the Alarm Limit Values 1. Press “ALARM” button to turn on the alarm function, the “ALM” icon, and current value are displayed. 2. Press “SET” button to enter High/Low limit value setting mode, the “SET” icon is displayed and hundred and thousand digits of the high limit value will flash. SET ALM CO2 ppm 3. Press “” or “” button to set desired value. 4. Press “” button to move the cursor to set the tens and ones digits. 5. Press “” or “” button to enter the desired value. 6. Press “” button to move the cursor to the hundred and thousand digits of the low limit value. 7. Press “” or “” button to set desired value. 8. Press “” button to move the cursor to move the cursor to set the tens and ones of the low limit value. 9. Press “” or “” button to set desired value. ppm CO2 CO2 ppm 10. To change any setting, press “” or “” button to move the cursor to desired high or low limit value position. 11. Press “” button to store these setting and exit this mode. Turning Alarm On / Off 1. 2. 3. 4. Press “ALARM” button to turn on the alarm function, “ALM” is displayed. When the CO2 value is below the low limit value, the meter displays “” mark and beeps. When the CO2 value exceeds the high limit value, the meter displays “” and beeps. To exit the ALARM function, press “ALARM” button again. Setting the Real Time Clock 1. CAL SET MAX MIN Press “SET” button to enter the real-time clock setting mode, “SET” is displayed and the minutes ALM are flashing. R 2. Press “” or “” button to set the minutes. M 88 Err INTV CO2 ppm 3. Press “” button to move the cursor to seconds. DEW WET 4. Press “” or “” button to set the seconds. °C°F CO2 m:s ppm CO2 d-h %RH ppm 5. Press “” button to move the cursor to days. 6. Press “” or “” button to set the day of the real time date. (Please note that this is not a calendar. Days are elapsed days up to 99 total) 7. Press “” button to move the cursor to the hours. (Please note that this is a 24 hour clock) 8. Press “” or “” button to set the hours. 9. To change any setting, press “” or “” button to move the cursor to desired position. 10. Press “” button to store the settings and exit this mode. 8888 8888 6 8888 EA80-en-GB_v3.3 09/15 Manual Datalogging Storing readings Press the "MEM" button. The LCD will display "M" and memory address number. The total memory size is 99 readings. Viewing readings Press the "READ" button to enter READ mode. The LCD display "R" and the memory address number. SET MAX MIN 8888 ALM R M 88 Err INTV 8888 CO2 m:s DEW WET °C°F ppm CO2 CAL the ppm 8888 CO2 d-h %RH ppm will Press " " or " " to scroll through the stored readings. Press " " enter to exit this mode. Deleting stored data Press the “SET” button three times. The LCD will display “CLr” and the meter will enter the Clear Memory mode. Press “” to clear the all manually stored readings. To abort, press the “SET” button twice and then press the “” button to exit the Clear Memory mode. Auto Datalogging Setting interval time Press “SET” two times. The “INTV” mark will appear and the meter will enter the Interval Time setting mode. Use the “” or “” button to select the desired interval time from 1 to 255 seconds. Press the “” button to store the value and to exit the mode. Auto Datalogging mode Press and hold “MEM” for two seconds (3 beeps), the LCD will display the “ ” icon and the memory location. The “M” will flash each time a recording is made. The maximum memory capacity is 16,200 recordings that can be divided into 99 sets. Press the “” button to exit this mode and to stop recording. Deleting Logged Data Press “SET” four times, the “CLr” and “ ” mark will be displayed. Press the “” button to clear the automatically stored data and to exit this mode. To abort, press the “SET” button again then press the “” button to exit. 7 EA80-en-GB_v3.3 09/15 Calibration CO2 Calibration 1. Press “ 2. Place the probe in a known CO2 reference for 10 minutes. 3. Press the “SET” button 5 times until “C-01” is displayed in the second display. 4. Press “►” or “◄” to select digits to adjust (flashing). 5. Press “▼” or “▲”to adjust the display to the reference value. 6. Press “” to store the value and exit the calibration mode. ” to turn the meter on. Humidity Calibration 1. Press “ 2. Place the probe in a known humidity reference for 60 minutes. 3. Press the “SET” button 6 times until “SET” and “CAL” are displayed on the LCD. 4. Press “▼” or “▲” to adjust the display to the reference value. 5. Press “” to store the value and exit the calibration mode. ” to turn the meter on. Temperature Calibration 1. Press “ 2. Place the probe in a known temperature reference for 60 minutes. 3. Press the “SET” button 7 times until “SET” and “CAL” are displayed on the LCD. 4. Press “▼” or “▲” to adjust the display to the reference value. 5. Press “” to store the value and exit the calibration mode. ” to turn the meter on. 8 EA80-en-GB_v3.3 09/15 Specifications Display Three Tier LCD Display Rate One reading per second Low Battery Indication The “ + ” icon is displayed when the battery voltage drops below the operating voltage Power Supply Six (6) AAA-size alkaline batteries or AC adapter (120V) Battery Life Approx. 8 hours using alkaline batteries (with backlight and Alarm functions OFF) Manual Data Memory Capacity 99 records Auto Datalogging Capacity 16,200 records in up to 99 data sets Operating Temperature Range 5°C to 50°C (41°F to 122°F) Storage Temperature Range -10°C to 60°C (-14°F to 140°F) Operating Humidity Range 10%RH to 90%RH, non-condensing Storage Humidity Range 10%RH to 90%RH, non-condensing Dimensions 158 (L) x 72(W) x 35(H) mm (6.22” x 2.83” x 1.38”) Weight 255g (0.56 lbs) approx. (including batteries) Accessories Instruction Manual, Batteries, AC Adaptor, Software CD ROM, and PC Interface Cable CO2 Specifications Sensing Range Sensing Resolution Accuracy Sensing Method Gas Sampling Mode Warm up time Response time Temperature Coefficient 0 to 6000ppm 1ppm 3% of reading or 50ppm, whichever is greater @ 101.4 kPa (29.92 inHg) and @ 25°C (77°F) Dual wavelength detector with non- dispersive infrared (NDIR) sensor Diffusion type 10 seconds < 10 minutes in still air Add 0.36% of reading per °C (0.2% of reading per °F) when deviating from calibration temperature Temperature & Humidity Specifications Relative Humidity Temperature 10% ~ 95% RH -20°C ~ +60°C (-4°C ~ +140°F) Resolution 0.1% RH 0.1°C (0.1°F) Accuracy 3%RH @ 25°C(77°F), 30~95%RH) 5%RH @ at 25°C(77°F), 10~30%RH) 0.5°C (0.9°F) Sensor type Precision capacitance sensor Thermistor Response time 45%RH95%RH≦1min 95%RH45%RH≦3min 10°C / 2sec Range 9 EA80-en-GB_v3.3 09/15 Maintenance Cleaning Periodically wipe the case with a dry cloth or a damp cloth with mild detergent. Do not use abrasives or solvents to clean this instrument. Battery Replacement When the “ ” symbol appears on the LCD, the six 1.5V ‘AAA’ must be replaced. 1. Turn the meter off 2. Remove the meter’s rubber protective jacket 3. Remove the flat-head screw at the rear of the meter 4. Remove the meter’s battery cover 5. Replace the batteries observing polarity 6. Affix the battery cover, secure the rear screw, and re-attach the meter’s rubber protective jacket You, as the end user, are legally bound (Battery ordinance) to return all used batteries and accumulators; disposal in the household garbage is prohibited! Bring used batteries / accumulators to the collection points in your community or wherever batteries / accumulators are sold! PC Interface, Software Installation and Operation Software operational instructions are located on the software disc. Install the software by inserting the disc in the PC CD-ROM drive and following the on-screen prompts. Open the Software manual (PDF) and read the Installation and Operation instructions. Copyright © 2013‐2015 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form ISO‐9001 Certified www.extech.com 10 EA80-en-GB_v3.3 09/15