English User's Manual

User’s Guide
200A AC Current Clamp-on MultiMeter Adaptor
Model CA200
AC Current Measurements
Introduction
Congratulations on your purchase of the Extech CA200 200 Amp AC
Clamp-On MultiMeter Adaptor. With this clamp adaptor, an AC current
can be safely measured by simply clamping around a single conductor.
The adaptor outputs an AC mV signal that is proportional to the
measured AC current. The DMM mV reading is equivalent to the value of
the AC current. Careful use of this device will provide years of reliable
service.
Specifications
Sensor type
Induction coil
Measurement Range
0 to 200 Amps AC
Adaptor Output
1mV AC output to DMM per 1A AC
measurement
Frequency range
50/60Hz
Standards
CAT II 600V, CAT III 300V
Max output impedance
<1k
Accuracy
± (2.0% + 0.5A) + (accuracy of indicating device)
Temperature coefficient 0.1 times the specified accuracy/ 1oC (< 18°C or
> 28°C)
Jaw size
0.62" (16mm)
Operating Conditions
(0 to 30°C) 32 to 86°F, < 90% RH
(30 to 40°C) 86 to 104°F, <75% RH
(40 to 500°C) 104 to 122°F, <45% RH
Dimensions/Weight
4.3 x 1.9 x 1.3" (111 x 50 x 33mm);
4.5 oz. (129g)
NOTE: Accuracy is specified for a period of one year for ambient
o
o
o
o
conditions 18 C to 28 C (64 F to 84 F), < 80% RH.
1.
2.
3.
4.
Set the DMM to the AC mV range or to a low AC voltage range.
Connect the RED test lead to the DMM '+' Voltage terminal and
the BLACK lead to the '-' COM terminal.
Press the Jaw Trigger to open the jaw and clamp around a single
conductor.
Read the current on the DMM display (1mV AC per 1A of AC
current).
1
1.
2.
3.
2
Read mV
Set to ACV
Voltage Input
3
Note: Do not apply voltage to the clamp adaptor’s test leads
Note: For small currents, use the RELATIVE function (if available) on
the DMM with no wire clamped to zero the meter and improve
accuracy.
Note: Clamp a single conductor only, do not clamp hot and neutral
simultaneously.
Note: The clamp jaws must be fully closed for proper measurements.
On large conductors or line splitters, make sure there is no gap
between the upper and lower jaw.
Safety
Safety Symbols
This symbol, adjacent to another symbol or terminal,
indicates the user must refer to the manual for further
information.
This symbol indicates that double insulation is used.
WARNING:
This indicates that a potentially hazardous condition which,
if not avoided, could result in death or serious injury.
Safety Precautions
1. WARNING: Improper use of this meter can cause damage, shock,
injury or death. Read and understand this manual before operating
the meter.
2. Inspect the condition of the test leads and the meter itself for any
damage before operating the meter. Repair or replace any
damage before use.
3. Do not use on non-insulated conductors at voltages greater than
250VAC rms or DC.
Copyright © 2013‐2015 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form ISO‐9001 Certified www.extech.com CA200-en-GB_V2.3 1/15