Downloads for MD-Series Medical Grade Isolation Power Transformers

Medical Isolation Transformers
MD-Series
Power Range: Available in 250VA, 500VA, 1KVA, 2KVA and 2.4KVA
Primary: 120 or 240VAC, 50/60Hz (depending on model)
Insulation Class: Class F, 155°C
Dielectric Isolation (Input to Output): 3KVAC.
Low leakage current and low stray fields
Electrical Specifications (@20C):
Triad
Input
Output
Regulation
Efficiency
Width
Height
Depth
Weight
P/N
VA
Voltage
Voltage
Typ.
Typ.
(mm)
(mm)
(mm)
(kg)
MD-250-E
250
240VAC
120VAC
8%
92%
134
130
176
4.0
MD-250-U
250
120VAC
120VAC
8%
92%
134
130
176
4.0
MD-500-E
500
240VAC
120VAC
5%
94%
162
146
209
6.8
MD-500-U
500
120VAC
120VAC
5%
94%
162
146
209
4.5
MD-1000-E
1000
240VAC
120VAC
5%
94%
182
146
230
10.0
MD-1000-U
1000
120VAC
120VAC
5%
94%
182
146
230
10.0
MD-2000-U
2000
120VAC
120VAC
5%
94%
222
173
270
19.1
MD-2400-E
2400
240VAC
120VAC
5%
94%
222
173
270
19.1
Safety:
nd
Medical isolation transformer box is certified to UL 60601-1, 2
edition medical standard and CSA-C22.2.No. 601.1M90, 2005
nd
UL: (E352717), UL60601-1, 2 edition and CAN/CSA-C22.2 No. 601.1M90, 2005
Note: Not for use in life critical application
Connections:
Input: 11ft medical grade power cord with Nema plug.
Output: Duplex receptacle. Nema 5-15R. 2000VA and 2400VA models are equipped with two duplex receptacles.
Schematic:
L
155C
Thermal
Protector
L
O
A
D
G
N
Custom designs available.
RoHS Compliance: Meets the requirements of 2011/65/EU, known as the RoHS initiative.
* Upon printing, this document is considered “uncontrolled”. Please contact Triad Magnetics’ website for the most current version.
Web: www.TriadMagnetics.com
Phone 951-277-0757
Fax 951-277-2757
460 Harley Knox Blvd.
Perris, California 92571
Publish Date: June 6, 2016