Variable-temperature Stage Options for FT-IR Microscopes and Imaging Systems

P r o d u c t
N o t e
FT-IR Spectroscopy
Figure 1. Linkam variable
temperature sample cell.
Figure 2. Hot Stage accessory with heating block.
Variable-temperature
Stage Options for
FT-IR Microscopes
and Imaging Systems
There is an increased interest in the microscopic
analysis of samples using FT-IR variable-temperature
microscopy or imaging. Application areas include
polymer studies, pharmaceutical and food analysis
and geological measurements.
PerkinElmer offers two options for variable temperature analyses – Linkam variable temperature
sample cell and the hot stage accessory.
Linkam Variable Temperature Sample Cell
This kit is designed for either the Spotlight™ 200 or Spotlight 400 systems. It
includes the Linkam stage, the temperature controller and a modified Spotlight
stage casting.
The silver cell has a tapered hole to allow maximum transmittance of IR without
interference. The cell will operate from ambient to 600 ˚C with the option of
extending this range down to -196 ˚C.
Spectra can be collected in either transmission or reflectance mode. The cell is
supplied complete with a BaF2 sample window. KBr or ZnSe windows are also
available as options.
Hot Stage Accessory
Specifications
Temperature range -196 to 600 °C
Fast heating rates up to 130 °C/min
Fast cooling rates up to 130 °C/min
Temperature stability <0.1 °C
16 mm X,Y sample manipulation
Sample area 22 mm diameter
Gas tight chamber for atmospheric
control
100 ohm platinum resistor sensor
1/10th Din
Light aperture -1.3 mm Ø for high
accuracy control
The hot stage accessory consists of a heating block, shown
in Figure 2, which accepts IR windows, and a temperature
controller. The cell is supplied with a set of KBr windows. The
heating block contains an integral thermocouple where the
temperature is digitally displayed in degrees Centigrade on the
controller. The hot stage can heat samples up to 230 ˚C in 1 ˚
increments. A target temperature may be selected and maintained. The hot stage is held in the slide clip on the sample
stage of a PerkinElmer® FT-IR MultiScope™, AutoImage™ or any
Spectrum™ Spotlight system. The microscope does not require
any modification in order to add the hot stage accessory.
Ordering information
N1870184
Hot Stage Accessory 110-120V 60Hz
Silver heating block for high
thermal conductivity
N1870188
Hot Stage Accessory 220-240V 50Hz
Direct injection of the coolant
into the silver block
Windows
Range of IR transparent windows
available including BaF2, ZnSe and KBr
Water cooled stage body for high
temperature work (>300 °C)
Stage body size 137 x 92 x 22 mm
N9302611
Barium Fluoride Disk (13 x 1 mm)
N9302612
Barium Fluoride Disk (13 x 2 mm)
N9302613
Zinc Selenide Disk (13 x 1 mm)
N9302614
Sodium Chloride Disk (13 x 2 mm)
N9302615
Potassium Bromide Disk (13 x 2 mm)
Ordering information
L1860634
Linkam Hot Stage Kit 220-240V, 50Hz
L1860635
Linkam Hot Stage Kit 110-130V, 60Hz
L1860636
Linkam Hot Stage Kit 220V, 60Hz (Korea)
L1860632
Cooler for Hot Stage 110V
L1860633
Cooler for Hot Stage 240V
PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com
For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2009, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.
008663_01