Using a Novel Heartcut Device for Multidimensional GC Peak Sectioning to Reveal

Using a Novel Heartcut Device for Multidimensional GC Peak Sectioning to Reveal
Small Peaks Otherwise Obscured by Large Peaks in Mass Chromatograms
Daniel Pentek, Andrew Tipler, William D. Goodman, Adam J. Patkin, PerkinElmer Life and Analytical Sciences, Shelton, CT, United States .
1
Methylenechloride _NoCut_ 092
Deans’ Switch
Figure 2. Installing the wafer in the
holder
Scan EI+
35_150
5.00e7
100
5.47
3.30
4.56
6.16
7.19
1.98
2.87
100
0.00
0.50
1.00
1.50
2.00
2.5 0
3.14
3.33
3. 00
2.72
3 .50
4 .00
4.50
5.00
5.50
6.00
6.50
7.00
7.50
Time
8.00
3
0.00
0.50
1.00
1.50
2.00
2.50
3.00
The final analytical data is obtained by combining the
results from all the chromatograms shown in Figures 10 and 12.
Table 3 lists each impurity detected in each of the 5 DCM
samples. In each case there would have been co-elution of
some peaks if the sidecutting and heartcut sectioning
techniques were not deployed.
3.67
1.63
3.50
4.00
4.50
5.00
5.50
6.00
6.50
7.00
7.50
Time
8.00
Table 3. Tentative MS assignment of compound identities in DCM samples using
the solvent side cutting and heartcut sectioning technique.
DCM Sample
Figure 3. Wafer mounted in the
threaded holder
Figure 4. The holder containing the wafer
mounted inside the GC oven
Figure 7. FID chromatogram of DCM
Sample 3 showing small impurity peaks.
restrictor
restrictor
Scan EI+
35_150
5.00e7
100
Split/splitless injector
D-Swafer
D-Swafer
Retention
Time, min
Scan EI+
TIC
2.00e10
100
Scan EI+
TIC
2.00e10
100
3.30
All sample transferred
to Column 2
4.56
5.48
30m x 0.25mm x 1.0µm Elite-Wax
%
MS
MS
30m x 0.25mm x 1.0µm Elite-Wax
6.17
Figure 6.
6 Wafer heartcut system with
effluent from primary column cut to
secondary column.
Figure 5.
5 Wafer heartcut system with
effluent from primary column
directed to mid-point FID.
3
Experimental Conditions
Tables 1 and 2 give details of the analytical system and
method applied to examine 5 samples of dichloromethane
(DCM).
Table 1. Analytical System.
Setting
Clarus 600
Oven Temperature
p
Heartcut device
D-Swafer in D4 configuration
Carrier Gas
Split/Splitless
Injector
225°C
Carrier Gas Pressure (P1)
23psig (159kPa)
Split Flow
100mL/min
15m x 0.25mm x 1.0m Elite-1
Midpoint Pressure (P2)
Column 2
30m x 0.25mm x 1.0µm Elite Wax
Detector 1 (FID) Temperature
Restrictor
58cm x 0.10mm deactivated fused
silica
Air Flow rate
450mL/min
Hydrogen Flow Rate
45mL/min
Range
x20
Attenuation
Detector 2 (MS) Temperature
Mass Range
250°C
x64
200°C
15 to 150 Da
0.2s
Interscan delay
0.1s
Sample Injection
1µL by Autosampler in Fast
Mode
Swafer Switching Valve (V4) Timed Events
See Results Section
S l t Sid
tti
Solvent
Sidecutting
In Figures 7 and 8 we see the solvent peak dominates
the chromatography around it and probably obscures some
smaller peaks. The large amount of solvent entering the MS
system also raises some concerns.
A run was made with the heartcut switched to the
second column at the start of the run and switched to the FID
during the solvent peak elution then switched back again. This
sidecutting technique has the effect of removing a large
fraction of the solvent yet allowing the rest of the sample to
g
9 shows
enter the second column and the MS detector. Figure
a chromatogram run this way.
Inspection of Figure 9 shows that much of the solvent
has been removed by the sidecutting method. This removal is
better illustrated by Figure 10 which shows the two
chromatograms at a larger scale. This is a highly effective
technique for keeping solvent away from the MS detector.
1.50
2.00
2.50
3.00
3.50
4.00
4.50
5.00
5.50
6.00
6.50
7.00
7.50
Time
8.00
0
1 00
1.00
2 00
2.00
3 00
3.00
4 00
4.00
5 00
5.00
6 00
6.00
7 00
7.00
8 00
8.00
Time
9 00
9.00
Figure 10. Chromatograms shown in
Figures 8 and 9 plotted together at a larger
scale to show the efficacy of sidecutting
for solvent removal.
branched chain pentene*
√
3.68+
dichloroethylene*
√
4
5
√
3
√
√
√
√
√
√
3.75
branched chain hexane*
3.87
acetone
3.90
branched chain hexane*
√
√
√
√
√
branched chain hexene*
4.56
dichloroethylene*
4.65
ethanol
4.70
isopropanol
4.91#
trimethyl oxirane
5.31
1-chlorobutane
5.48
2-chloro-2-methylbutane
5.79
cyclohexene
y
√
√
√
√
√
√
√
√
√
√
√
√
√
√
6.02
acetonitrile
6.17
2-butanone
√
7.08
hexyl alcohol*
√
√
7.19
chloroform
√
√
√
√
Solvent Sectioning
6
Methylenechloride_1-78_099
100
100
%
0
1
2
3
4
5
Scan EI+
35_150
5.00e7
5.06
6
6
Methylenechloride_1-76_098
6
5
7
Methylenechloride_1-74_097
100
%
4
Scan EI+
35_150
5.00e7
3.46
7
Methylenechloride_1-72_096
100
%
%
Scan EI+
35_150
5.00e7
3
7
Methylenechloride_1-70_095
100
%
Scan EI+
35_150
5.00e7
2
3.68
7
Methylenechloride_1-68_094
100
%
Scan EI+
35_150
5.00e7
1
3.41 3.68
7
Methylenechloride_1-66_093
100
%
1.67 1.68
1.69
1.70 1.71
1.72
1.73 1.74
1.75
1.76 1.77
1.78
1.79
1.80
1.81 1.82
1.83
1.84 1.85
Figure 11. Sectioning the DCM solvent
peak into six 0.02-minute heartcuts.
Time
Scan EI+
35_150
5.00e7
0
7
1.00
This time would be needed to fully
y map
p the
obscured components. In the samples examined
here, only two additional peaks were found in
the sectioned chromatograms and so the
method could be optimized just to apply
heartcuts to the affected sections and so reduce
the number of runs necessary.
Scan EI+
35_150
5.00e7
100
%
Summary
Thi side-cut
This
id
t and
dh
heartcut
t t ttechnique
h i
provides a comprehensive and reliable method
of revealing the low-level impurities of solvents
using MS as the detection system. Although the
solvent peak sectioning process entails several
repeat chromatograms of the same sample,
these runs are fairly short and isothermal and so
the total analytical time is just 50 minutes.
23+16psig 100mLsplit 1.68 Cut, 04-Feb-2009 + 20:32:17
Methylenechloride_NoCut_092
80
2
+ Peak co-eluting with solvent in Column 1
# Peak co-eluting with solvent in Column 2
* Isomer not determined
Although this sidecutting technique allows the sample to
be processed on the MS without the potential damage and
interference from the solvent peak, it does not take into
account any peaks which will co-elute with the solvent on the
primary column – these peaks would not enter the secondary
column or be seen by the MS.
Close examination of Figure 9 reveals that two peaks are
missing from this chromatogram at approximately 3.42 and
3.67 minutes that were present in Figure 8. These clearly
must co-elute with the solvent on the primary column.
To enable these (and possibly other) peaks that co-elute with
the solvent to be transferred to the second column for
separation, a peak sectioning technique was used to deliver
time-incremented narrow heartcuts of the solvent peak from
sample Figure 11 shows
successive runs of the same DCM sample.
how the solvent peak was sectioned into six 0.02-minute
heartcuts that each produced chromatograms shown in Figure
12. This approach allows the area under the solvent peak on
the first column to be fully mapped by the second without
exposing the MS detector to large amounts of solvent.
16psig (110kPa)
Scan Time
1.00
5
Helium
Temperature
Flame Ionization
Clarus 600 MS
4
0.50
Value
Detector 2
Column1
5
0.00
Figure 9. MS TIC with DCM solvent peak
sidecut. The switching valve was turned off
between 1.68 and 1.80 minutes but was on
for the rest of the run.
60°C isothermal for 8 minutes
Detector 1
7.19
2.72
Table 2. Analytical Conditions.
Description
GC
3.41+
4.56
0
Methylenechloride_Total_119
%
P1
P1
1
2-methylbutane
%
5.08
15m x 0.25mm x 1.0µm Elite-1
15m x 0.25mm x 1.0µm Elite-1
MS Identification
3.30
23+16psig 100mLsplit Nosolvent, 05-Feb-2009 + 10:23:10
Methylenechloride_NoSolvent_118
Solvent eliminated by
sidecutting
Solvent peak
largely removed
P2
P2
Figure 8. MS total ion chromatogram
(TIC) of DCM Sample 3.
23+16psig 100mLsplit Nosolvent, 05-Feb-2009 + 10:35:34
Methylenechloride_NoSolvent_119
FID
FID
Split/splitless injector
Injector
For this work, a D-SwaferTM microchannel pneumatic switch was
used as shown in Figures 1 to 4. This device is about the same
size as a nickel coin. The internal microchannels are fabricated
using laser etching. They are fully chemically deactivated to
handle reactive analytes.
In this instance the wafer was configured in a classical
heartcutting configuration (Deans’ switch) to enable sample cuts
to be directed from the effluent of the first column into the inlet
of the second column as shown in Figures 5 and 6. The cutting
operation is controlled through a solenoid valve programmed by
GC timed events to apply a switching gas to direct the primiary
column effluent between the wafer two outlets.
Methylenechloride_Total_119
%
%
Component
Figure 1. The D-Swafer Deans’
switching microchannel wafer
A: F ID
5.38e5
100
Solvents are widely used in the
pharmaceutical and food industries for a variety of
purposes. It is important that such solvents are
carefully QC tested prior to use to ensure that no
unsafe levels of impurities are present.
GC is normally the preferred technique for the
determination of impurities in solvents. The
inclusion of a mass spectrometric detector enables
the identities and quantification of trace-level
impurities to be established.
Because many solvents are produced by
fractional distillation, their impurities will have
similar boiling points to that of the solvent. Thus in
GC, the retention times will be similar to that of
the solvent and the risk of co-elution can be high.
Furthermore, if the MS is kept active during
solvent elution, contamination of the ion source or
analyzer may result and the risk of filament
damage is greatly increased.
This poster describes a heartcutting technique
that allows the entire injected sample to reach the
detector and yet resolve the issues with solvent
peak resolution and potential detector damage.
2
23+16psig 100mLsplit Total, 05-Feb-2009 + 11:55:02
23+16psig 100mLsplit No Cut, 04-Feb-2009 + 20:07 :30
Introduction
2.00
3.00
4.00
5.00
6.00
7.00
Figure 12. Chromatograms from
successive 0.02-minute heartcuts
8.00
9.00
Time
Although we have shown the application of
this technique just to samples of
dichloromethane,
dichloromethane the same approach could be
extended to other solvents or any sample where
there is an interest in identifying and quantifying
compounds at low levels that co-elute with other
relatively large peaks.
PerkinElmer, Inc., 940 Winter Street, Waltham, MA USA (800) 762-4000 or (+1) 203 925-4602 www.perkinelmer.com