Download

Programmable Digital Controller
E5AR-T
CSM_E5AR-T_DS_E_3_3
A new High-speed, High-precision Digital
Controller that is Programmable!
• Create up to 32 programs with up to 256 segments total.
• Coordinated operation for up to four channels with one Digital
Controller.
• 0.01°C High resolution for Pt input.
• High-speed sampling at 50 ms.
• Settings easily made from a computer using the CX-Thermo.
• RoHS compliance for world-wide application.
Refer to Safety Precautions for All E5@R
Models.
For the most recent information on models that have been certified for
safety standards, refer to your OMRON website.
Model Number Structure
■ Model Number Legend
Note: When your order, specify the power supply voltage.
E5AR-T@@@@@@@@-@@@
1 2 3 4 5 6 7 8
9
1. Control method
Blank:Standard or heating/cooling control
P:
Position proportional control
2. Output 1
R:
NO relay output + NO relay output
Q:
Pulse output/current output + pulse output
C:
Current output + current output
3. Output 2
R:
NO relay output + NO relay output
Q:
Pulse output/current output + pulse output
C:
Current output + current output
4. Auxiliary Outputs
Blank:None
4:
NO relay output + NO relay output
E:
5 transistor outputs + 5 transistor outputs
5. Communications
Blank:None
3:
RS-485 communications
6. Optional function
Blank:None
D:
4 event inputs
M:
4 event inputs + 4 event inputs
7. Input 1
B:
Universal-input and 2 event inputs
F:
Universal-input and FB
W:
Universal-input and universal-input
8. Input 2
Blank:None
W:
Universal-input and universal-input
9. Other
FLK: CompoWay/F communications
Note: The above model number legend is intended as a functional description of models. Not all possible combinations of functions are available.
Confirm model availability in Ordering Information when ordering.
Note: Be sure to read the precautions for correct use and other precautions in the following user's manual before using the Digital
Controller.
E5AR/ER Digital Controller User's Manual (Cat. No. Z182)
1
E5AR-T
Ordering Information
■ Digital Controllers
When your order, specify the power supply voltage.
Programmable Digital Controllers (100 to 240 VAC)
Size
Control type
Control mode
96 × 96 Basic control Standard control
mm
(1 loop)
Heating and cooling control
Outputs
(control/transfer)
Optional functions
Auxiliary
outputs
(SUB)
2 (pulse + pulse/cur- 4
rent)
Event
inputs
2
Model
Serial
communications
None
2 (current + current)
E5AR-TQ4B
E5AR-TC4B
2 (pulse + pulse/current)
RS-485
2 (current + current)
E5AR-TQ43B-FLK
E5AR-TC43B-FLK
2 (pulse + pulse/cur- 10 (See
rent)
note 1.)
10
E5AR-TQE3MB-FLK
2 (current + current)
E5AR-TCE3MB-FLK
4 (pulse + pulse/current + 2 current)
E5AR-TQCE3MB-FLK
2-loop control 2-loop standard control
Single-loop heating and cooling control
Single-loop cascade control
Single-loop control with remote SP
Single-loop proportional control
2 (pulse + pulse/cur- 4
rent)
4 (2 pulse + pulse/2
current)
10 (See
note 1.)
8
4-loop control 4-loop standard control
2-loop heating and cooling control
(See note 2.)
4 (4 current)
10 (See
note 1.)
8
4
RS-485
2 (current + current)
E5AR-TC43DW-FLK
E5AR-TQQE3MWFLK
RS-485
4 (2 pulse + pulse/2
current)
Control valve Single-loop position-proportional control Relay outputs
control
(1 open, 1 closed)
(1 loop)
Relay outputs
(1 open, 1 closed)
and 1 current
E5AR-TQ43DW-FLK
E5AR-TCCE3MWWFLK
E5AR-TQQE3MWWFLK
4
4
None
E5AR-TPR4DF
10 (See
note 1.)
8
RS-485
E5AR-TPRQE3MFFLK
Note 1: The outputs are transistor output.
2: Only for coordinated operation. (A separate program cannot be set for each channel.)
Programmable Digital Controllers (24 VAC/DC)
Size
Control type
Control mode
96 × 96 Basic control Standard control
mm
(1 loop)
Heating and cooling control
Outputs
(control/transfer)
Optional functions
Auxiliary
outputs
(SUB)
2 (pulse + pulse/cur- 4
rent)
Event
inputs
Model
Serial
communications
2
None
10
RS-485
E5AR-TQCE3MB-FLK
8
RS-485
E5AR-TQQE3MWFLK
2 (current + current)
E5AR-TQ4B
E5AR-TC4B
4 (pulse + pulse/current + 2 current)
4 (2 pulse + pulse/2
2-loop control 2-loop standard control
Single-loop heating and cooling control current)
Single-loop cascade control
Single-loop control with remote SP
Single-loop proportional control
10 (See
note 1.)
10 (See
note 1.)
4-loop control 4-loop standard control
2-loop heating and cooling control
(See note 2.)
10 (See
note 1.)
8
RS-485
E5AR-TCCE3MWWFLK
4
4
None
E5AR-TPR4DF
10 (See
note 1.)
8
RS-485
E5AR-TPRQE3MFFLK
4 (4 current)
Control valve Single-loop position-proportional control Relay outputs
control
(1 open, 1 closed)
(1 loop)
Relay outputs
(1 open, 1 closed)
and 1 current
Note 1: The outputs are transistor output.
2: Only for coordinated operation. (A separate program cannot be set for each channel.)
2
E5AR-T
Inspection Results
■ Accessories (Order Separately)
If an inspection report is required, it can be ordered at the same time
as the Digital Controller using the following model number.
Terminal Cover
Inspection Report (Order Separately)
Model
E5AR-K
Descriptions
Terminal Cover for E5AR
Model
E53-COV14
Unit Label Sheet
Model
Y92S-L1
Rubber Packing
Model
Y92S-P4
Note: The Rubber Packing is provided with the Digital Controller.
3
E5AR-T
Specifications
■ Ratings
Supply voltage
(See note 2.)
CE marking
100 to 240 VAC, 50/60 Hz
UL certification
100 to 120 VAC, 50/60 Hz
Operating voltage range
24 VAC, 50/60 Hz; 24 VDC
85% to 110% of rated supply voltage
Power consumption
22 VA max. (with maximum load)
Sensor input (See note 3.)
Thermocouple: K, J, T, E, L, U, N, R, S, B, W
Platinum resistance thermometer: Pt100
Current input: 4 to 20 mA DC, 0 to 20 mA DC (including remote SP input)
Voltage input: 1 to 5 VDC, 0 to 5 VDC, 0 to 10 VDC (including remote SP input)
(Input impedance: 150 Ω for current input, approx. 1 MΩ for voltage input)
Control output
15 VA/10 W max. (with maximum load)
Voltage (pulse) output 12 VDC, 40 mA max. with short-circuit protection circuit (E5AR-TQQE3MW-FLK: 21 mA max.)
Current output
0 to 20 mA DC, 4 to 20 mA DC; load: 500 Ω max. (including transfer output)
(Resolution: Approx. 54,000 for 0 to 20 mA DC; Approx. 43,000 for 4 to 20 mA DC)
Relay output
Position-proportional control type (open, closed)
N.O., 250 VAC, 1 A (including inrush current)
Auxiliary output
Relay Output
N.O., 250 VAC, 1 A (resistive load)
Transistor Output
Maximum load voltage: 30 VDC; Maximum load current: 50 mA; Residual voltage: 1.5 V max.; Leakage current: 0.4 mA max.
Potentiometer input
100 Ω to 2.5 kΩ
Event input
Contact
Input ON: 1 kΩ max.; OFF: 100 kΩ min.
No-contact
Input ON: Residual voltage of 1.5 V max.; OFF: Leakage current of 0.1 mA max.
Short-circuit: Approx. 4 mA
Remote SP input
Refer to the information on sensor input.
Transfer output
Refer to the information on control output.
Control method
2-PID or ON/OFF control
Setting method
Digital setting using front panel keys or setting using serial communications
Indication method
7-segment digital display and single-lighting indicator
Character Height
PV display: 12.8 mm; SV display: 7.7 mm; MV display: 7.7 mm
Other functions
Depends on model.
Ambient operating temperature
−10 to 55°C (with no icing or condensation)
For 3 years of assured use: −10 to 50°C (with no icing or condensation)
Ambient operating humidity
25% to 85%
Storage temperature
−25 to 65°C (with no icing or condensation)
Note 1: Do not use an inverter output as the power supply. (Refer to Safety Precautions for All E5@R Models.)
2: The supply voltage (i.e., 100 to 240 VAC or 24 VAC/VDC) depends on the model. Be sure to specify the required type when ordering.
3: The Controller is equipped with multiple sensor input. Temperature input or analog input can be selected with the input type setting switch. There is basic
insulation between power supply and input terminals, power supply and output terminals, and input and output terminals.
■ Input Ranges
Platinum Resistance Thermometer, Thermocouple, Current, or Voltage Input
Input type
Pt100
Name
Temperature
range
(°C)
Thermocouple
Platinum
resistance
thermometer
2300
1800
1300
900
800
700
600
400
200
100
0
−100
−200
Setting
Minimum
setting unit
(SP and alarm)
Input type
setting switch
K
J
T
E
L
U
N
R
S
B
1700.0
1700.0
1800.0
Current
Voltage
mA
V
W
( W/Re
5-26 )
2300.0
1300.0
1300.0
850.0
850.0
850.0
500.0
20 to
4
600.0
400.0
400.0
400.0
150.00
20 to
10 to
5 to 1 5 to 0
0
0
100.0
0.0
−200.0 −150.00 −200.0
0
1
0.1°C
0.01
°C
2
−20.0
−100.0
−20.0
3
4
5
−100.0
−200.0
6
7
−200.0
−200.0
9
10
8
0.1°C
0.0
0.0
11
12
0.0
13
14
15
16
17
TC.PT
Set to TC.PT.
IN1
TYPE
ANALOG
18
19
(Depends on scaling and number
of decimal places.)
TC.PT
Set to ANALOG
IN1
TYPE
ANALOG
Note: The shaded area indicates the setting status at the time of purchase.
4
E5AR-T
■ Characteristics
Indication accuracy
Thermocouple input with cold junction compensation: (±0.1% of PV or ±1°C, whichever is greater) ±1 digit max. (See note 1.)
Thermocouple input without cold junction compensation: (±0.1% FS or ±1°C, whichever is smaller) ±1 digit (See note 2.)
Analog input: ±0.1% FS ±1 digit max.
Platinum resistance thermometer input: (±0.1% of PV or ±0.5°C, whichever is greater) ±1 digit max.
Position-proportional potentiometer input: ±5% FS ±1 digit max.
Control mode
Standard control (heating or cooling control), heating/cooling control, standard control with remote SP (2-input models only), heating/cooling control with remote SP (2-input models only), cascade standard control (2-input models only), cascade heating/cooling control (2-input
models only), proportional control (2-input models only), position-proportional control (control-valve control models only)
Influence of temperature Thermocouple input (R, S, B, W): (±1% of PV or ±10°C, whichever is greater) ±1 digit max.
Other thermocouple input: (±1% of PV or ±4°C, whichever is greater) ±1 digit max.
Influence of voltage
*K thermocouple at -100°C max.: ±10°C max.
Platinum resistance thermometer: (±1% of PV or ±2°C, whichever is greater) ±1 digit max.
Influence of EMS.
Analog input: (±1%FS) ±1 digit max.
(at EN61326-1)
Control period
0.2 to 99.0 s (in units of 0.1 s) for time-proportioning control output
Proportional band (P)
0.00% to 999.99% FS (in units of 0.01% FS)
Integral time (I)
0.0 to 3,999.9 s (in units of 0.1 s)
Derivative time (D)
0.0 to 3,999.9 s (in units of 0.1 s)
Hysteresis
0.01% to 99.99% FS (in units of 0.01% FS)
Manual reset value
0.0% to 100.0% (in units of 0.1% FS)
Alarm setting range
−19,999 to 99,999 EU (See note 3.)
(The decimal point position depends on the input type and the decimal point position setting.)
Input sampling period
50 ms
Insulation resistance
20 MΩ min. (at 500 VDC)
Dielectric strength
2,000 VAC, 50/60 Hz for 1 min (between charged terminals of different polarities)
Vibration resistance
(malfunction)
10 to 55 Hz, 20 m/s2 for 10 min each in X, Y, and Z directions
Shock resistance
(malfunction)
100 m/s2, 3 times each in X, Y, and Z directions
Inrush current
100 to 240-VAC models: 50 A max.
24 VAC/VDC models: 30 A max.
Weight
Controller only: Approx. 450 g; Mounting bracket: Approx. 60 g; Terminal cover: Approx. 30 g
Degree of protection
Front panel: NEMA4X for indoor use; Rear case: IP20; Terminals: IP00
Memory protection
Non-volatile memory (number of writes: 100,000)
Applicable standards
UL 61010C-1, CSA C22.2 No. 1010-1
(Power supply voltage: 100 to 120 VAC): Pollution degree 2/Overvoltage category 2
EN 61010-1 (IEC 61010-1) (Power supply voltage: 100 to 240 VAC): Pollution degree 2/Overvoltage category 2
EMC
EMI: EN61326-1 (See note 4.)
Radiated Interference Electromagnetic Field Strength: EN55011 Group 1 Class A
Noise Terminal Voltage: EN55011 Group 1 Class A
EMS: EN61326-1 (See note 4.)
ESD Immunity: EN61000-4-2:
4 kV contact discharge (level 2)
8 kV air discharge (level 3)
10 V/m (amplitude-modulated, 80 MHz to 1 GHz, 1.4 GHz to 2 GHz) (level 3)
Electromagnetic Immunity: EN61000-4-3:
Burst Noise Immunity:
EN61000-4-4:2 kV power line (level 3)
2 kV output line (relay output) (level 4)
1 kV measurement line, I/O signal line (level 4)
1 kV communications line (level 3)
Conducted Disturbance Immunity: EN61000-4-6:
3 V (0.15 to 80 MHz) (level 3)
Surge Immunity:
EN61000-4-5:1 kV line to line (power line, output line (relay output)) (level 2)
2 kV line to ground (power line, output line (relay output)) (level 3)
Power Frequency Magnetic Field Immunity: EN61000-4-8: 30 A/m (50 Hz) continuous field
Voltage Dip/Interrupting Immunity: EN61000-4-11:
0.5 cycle, 100% (rated voltage)
Note 1: K-, T-, or N-type thermocouple at −100°C max.: ±2°C ±1 digit max.
U- or L-type thermocouple: ±2°C ±1 digit max.
B-type thermocouple at 400°C max.: No accuracy specification.
R- or S-type thermocouple at 200°C max.: ±3°C ±1 digit max.
W-type thermocouple: (±0.3% of PV or ±3°C, whichever is greater) ±1 digit max.
2: U- or L-type thermocouple: ±1°C ±1 digit
R- or S-type thermocouple at 200°C max.: ±1.5°C ±1 digit
3: “EU” (Engineering Unit) represents the unit after scaling. If a temperature sensor is used, it is either °C or °F.
4: Industrial electromagnetic environment (EN/IEC 61326-1 Table 2)
5
E5AR-T
■ Communications Specifications
Transmission path connection
Multiple points
Communications method
RS-485 (two-wire, half duplex)
Synchronization method
Start-stop synchronization
Baud rate
9,600, 19,200, or 38,400 bps
Transmission code
ASCII
Data bit length
7 or 8 bits
Stop bit length
1 or 2 bits
Error detection
Vertical parity (none, even, odd)
Block check character (BCC): CompoWay/F
CRC-16: Modbus
Flow control
None
Interface
RS-485
Retry function
None
Communications buffer
217 bytes
Communications response send wait time 0 to 99 ms, Default: 20 ms
■ Program Control Functions
Number of programs (patterns)
32 (with 8 segments/program)
Number of segments (steps)
32 (with 8 programs)
Maximum number of segments
256
Segment setting method
Time setting (Segment set with set point and time.)
Gradient setting (Segment set with set point, gradient, and time.)
Segment times
0 h 0 min to 99 h 59 min
0 min 0 s to 99 min 59 s
0 min 00.0 s to 99 min 59.9 s
Alarm group num- Number of groups
ber specifications Setting method
4
Reset operation
Select either stopping control or fixed SP operation.
Startup operation
PID sets
Set separately for each program.
Select continuing, resetting, manual operation, run mode, or ramp back operation.
Number of sets
8
Setting method
Set separately for each program (automatic PID group selection also supported).
Program status
control
Segment operation
Advance, hold, and back
Program operation
Program repetitions and program links
Wait operation
Wait method
Select from waiting at segment ends and always waiting.
Wait width setting
Wait width upper limit and lower limit set separately for each program.
Setting method
ON/OFF setting for each segment
Number of outputs
6
Number of ON/OFF
operations
3 each per output
Setting method
Set separately for each program.
Alarm SP function
Time signals
Select from ramp SP and target SP.
Segment outputs Number of outputs
Setting method
10
ON/OFF set for each segment.
Program status output
Program end output (pulse width can be set)
Segment number output
Program startup
operation
PV start
Select from segment 1 set point, slope-priority PV start, and time-priority PV start.
Standby
Standby
Operation end operation
Select from resetting, continuing control at final set point, and fixed SP control.
Number of event inputs
10 max.
6
E5AR-T
Wiring Terminals
■ E5AR-T (Programmable Type)
E5AR-TQ4B
+
+
−
A
100 to 240 VAC
!
2
3
4
Input power supply
depends on model.
100 to 240 VAC or 24
VAC/DC (no polarity)
24 VAC/DC
Auxiliary output
(Relay output)
B C D E
1
5
6
1
B
1
COM
2
SUB1
3
SUB2
4
5
6
+
+
−
−
24 VAC/DC
E5AR-TC4B
−
100 to 240 VAC
A
!
Input power supply
depends on model.
100 to 240 VAC or 24
VAC/DC (no polarity)
COM
SUB3
SUB4
2
B
1
COM
3
4
2
SUB1
3
SUB2
5
6
1
4
5
6
COM
SUB3
SUB4
3
4
3
4
5
6
5
6
F G H I
Event input
−
4
OUT1
+
5
Voltage output 12 V
−
40 mA or
6
Current output
4 to 20 mA DC, 500 Ω max. F
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
3
COM
+
−
I
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output
type setting.)
2
EV2
3
4
−
−
+
V
+
TC
PT
OUT1
Current output
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output
type setting.)
5
6
K
J K
Event input
1
OUT2
Current output
1
EV1
2
+
OUT2
Voltage output 12 V
40 mA
F G H I
J K
1
+
−
+
−
EV1
1
EV2
2
COM
3
2
3
+
4
5
6
F
−
−
−
+
V
I
+
TC
PT
E5AR-TC43B-FLK
Auxiliary output
(Relay output)
A
100 to 240 VAC
!
B C D E
Auxiliary output
(Relay output)
1
1
B
1
COM
2
B
1
COM
2
2
SUB1
2
SUB1
3
4
3
SUB2
3
4
3
SUB2
4
COM
4
COM
5
6
SUB3
5
6
SUB3
2
5
6
1
5
6
1
3
2
4
3
4
A
100 to 240 VAC
!
1
OUT1
+
5
Voltage output 12 V
40 mA or
−
6
Current output
4 to 20 mA DC, 500 Ω max.
F
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
4
5
6
K
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
E5AR-TQ43B-FLK
OUT2
Voltage output 12 V 40 mA −
Auxiliary output
(Relay output)
2
2
+
B(+)
RS-485
−
A(−)
+
B C D E
1
B C D E
SUB4
SUB4
2
3
4
5
6
5
6
F G H I
F G H I
J K
Event input
+
EV1
1
EV2
2
COM
3
−
−
I
+
V
−
PT
+
TC
4
5
6
K
OUT2
Current output
B(+)
RS-485
A(−)
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output
type setting.)
OUT1
Current output
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output
type setting.)
+
−
+
−
+
−
J K
Event input
1
EV1
1
2
EV2
2
3
COM
3
4
5
6
F
+
−
I
−
−
+
+
V
PT
TC
4
5
6
K
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
Note
The power supply voltage must be 100 to 240 VAC or 24 VAC/DC for the E5AR-T to comply with CE
marking requirements.
The power supply voltage must be 100 to 120 VAC or 24 VAC/DC for the E5AR-T to comply with UL
requirements.
7
E5AR-T
E5AR-TQE3MB-FLK
E5AR-TCE3MB-FLK
Auxiliary output
(Transistor output)
−
100 to 240 VAC
!
A
B C D E
COM
+
1
SUB1(B), SUB6(C)
2
SUB2(B), SUB7(C)
3
4
SUB3(B), SUB8(C) +
+
SUB4(B), SUB9(C)
SUB5(B), SUB10(C)+
5
6
1
3
4
5
6
F G H I
J K
+
2
+
OUT2
Voltage output 12 V 40 mA
DE
1
EV3(E), EV7(D)
2
EV4(E), EV8(D)
3
EV5(E), EV9(D)
4
5
6
4
5
6
F
OUT1
+
Voltage output 12 V
−
40 mA or
Current output
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
+
EV1
1
EV2
2
COM
3
−
−
−
+
V
I
PT
+
TC
A
SUB1(B), SUB6(C)
B C D E
+
2
3
SUB3(B), SUB8(C) +
3
4
4
SUB4(B), SUB9(C) +
SUB5(B), SUB10(C) +
5
6
Event input
3
4
5
6
F G H I
J K
2
EV4(E),EV8(D)
3
EV5(E),EV9(D)
4
5
6
EV6(E),EV10(D)
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
1
Event input
2
EV1
1
4 to 20 mA DC, 500 Ω max.
3
0 to 20 mA DC, 500 Ω max. −
4
(Switched by output
+
type setting.)
5
OUT1
−
Current output
6
4 to 20 mA DC, 500 Ω max.
F
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
EV2
2
COM
3
+
DE
1
EV3(E),EV7(D)
COM
+
B(+)
RS-485
−
A(−)
4
5
6
K
2
SUB2(B), SUB7(C) +
2
OUT2
Current output
BC
1
1
5
6
1
Event input
3
−
!
4
5
6
COM
1
−
100 to 240 VAC
3
EV6(E), EV10(D)
B(+)
RS-485
A(−)
−
COM
2
+
Event input
2
Auxiliary output
(Transistor output)
BC
1
+
−
I
−
−
+
+
TC
V
PT
4
5
6
K
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
E5AR-TQCE3MB-FLK
Auxiliary output
(Transistor output)
100 to 240 VAC
24 VAC/DC
+
!
+
Input power supply
depends on model.
100 to 240 VAC or 24
VAC/DC (no polarity)
+
B(+)
RS-485
−
OUT2
A(−)
+
Voltage output 12 V
40 mA
−
OUT1
+
Voltage output 12 V
40 mA or
−
Current output
A
BC
1
SUB1(B), SUB6(C)
2
2
SUB2(B), SUB7(C) +
3
3
4
SUB3(B), SUB8(C) +
+
SUB4(B), SUB9(C)
+
SUB5(B), SUB10(C)
4
5
6
2
3
2
4
5
6
F
3
4
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
COM
+
5
6
1
1
-
B C D E
1
Event input
5
6
2
EV4(E), EV8(D)
3
EV5(E), EV9(D)
4
5
6
EV6(E), EV10(D)
F G H I
OUT4
Current output
4 to 20 mA DC, 500 Ω max.
+
0 to 20 mA DC, 500 Ω max.
(Switched by output
−
type setting.)
+
OUT3
Current output
−
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
J K
2
3
COM
+
−
−
I
+
V
−
PT
1
2
EV2
3
4
5
6
G
COM
Event input
EV1
1
DE
1
EV3(E), EV7(D)
+
TC
4
5
6
K
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
Note
The power supply voltage must be 100 to 240 VAC or 24 VAC/DC for the E5AR-T to comply with CE
marking requirements.
The power supply voltage must be 100 to 120 VAC or 24 VAC/DC for the E5AR-T to comply with UL
requirements.
8
E5AR-T
E5AR-TQ43DW-FLK (2-loop Control)
E5AR-TC43DW-FLK (2-loop Control)
Auxiliary output
(Relay output)
Auxiliary output
(Relay output)
SUB2
4
COM
5
6
SUB3
4
5
6
SUB3
SUB4
E
1
5
6
F G H I
EV3
2
EV4
3
EV5
J K
EV6
+
+
−
I
−
+V
+
TC
PT
I
+
V
PT
5
6
1
2
3
4
5
6
F G H I
J K
SUB4
Event input
E
1
EV3
2
EV4
3
EV5
4
5
6
EV6
COM
Input 2
2
3
4
−
5
6
+
TC K
−
−
Input 1
1
−
Input 2
3
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max. −
4
(Switched by output
+
type setting.)
5
−
OUT1
6
Current output
4 to 20 mA DC, 500 Ω max.
F
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
4
5
6
COM
+
B(+)
1
RS-485
−
OUT2
A(−)
2
Current output
+
B C D E
Input 1
1
−
I
+
−
V
PT
−
−
Event input
3
4
4
OUT1
+
Voltage output 12 V
5
40 mA or
−
Current output
6
4 to 20 mA DC, 500 Ω max. F
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
3
3
4
3
3
4
2
3
2
2
!
A
+
2
SUB2
COM
SUB1
100 to 240 VAC
+
−
+
OUT2
Voltage output 12 V 40 mA −
SUB1
2
B C D E
5
6
1
1
2
1
A
−
+
B(+)
RS-485
A(−)
COM
1
COM
100 to 240 VAC
!
B
1
B
1
TC
−
+
+
I
+
V
PT
TC
2
3
4
5
6
K
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
E5AR-TQQE3MW-FLK (2-loop Control)
Auxiliary output
(Transistor output)
24 VAC/DC
+
−
100 to 240 VAC
−
!
A
+
40 mA or
Current output
−
2
2
3
3
4
SUB3(B), SUB8(C) +
4
5
6
1
SUB5(B), SUB10(C) +
SUB4(B), SUB9(C) +
Event input
2
3
3
4
4
5
6
5
6
F
F G H I
J K
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
−
3
4
5
6
G
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
Note
+
+
2
3
EV5(E), EV9(D)
4
EV6(E), EV10(D)
5
6
1
−
−
2
+
TC
3
Input 2
I
+
V
PT
−
−
−
2
DE
1
EV4(E), EV8(D)
+
Voltage output 12 V
40 mA or
Current output
+
−
Voltage output 12 V
40 mA
OUT3
5
6
EV3(E), EV7(D)
COM
1
OUT4
BC
1
SUB2(B), SUB7(C) +
2
1
COM
SUB1(B), SUB6(C) +
1
−
Input power supply
depends on model.
100 to 240 VAC or 24
VAC/DC (no polarity)
+
B(+)
RS-485
−
OUT2
A(−)
+
Voltage output 12 V
40 mA
−
OUT1
+
Voltage output 12 V
B C D E
−
Input 1
I
+
V
PT
+
TC
4
5
6
K
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
The power supply voltage must be 100 to 240 VAC or 24 VAC/DC for the E5AR-T to comply with CE
marking requirements.
The power supply voltage must be 100 to 120 VAC or 24 VAC/DC for the E5AR-T to comply with UL
requirements.
9
E5AR-T
E5AR-TCCE3MWW-FLK (4-loop Control)
E5AR-TQQE3MWW-FLK (4-loop Control)
Auxiliary output
(Transistor output)
B C D E
−
+
2
3
SUB3(B), SUB8(C) +
+
SUB4(B), SUB9(C)
3
4
SUB5(B), SUB10(C) +
5
6
1
+
B(+)
1
RS-485
−
OUT2
A(−)
2
Current output
+
4 to 20 mA DC, 500 Ω max.
3
0 to 20 mA DC, 500 Ω max. −
(Switched by output
4
type setting.)
+
5
OUT1
−
Current output
6
4 to 20 mA DC, 500 Ω max.
F
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
2
SUB2(B), SUB7(C) +
2
DE
1
3
4
EV3(E), EV7(D)
2
5
6
EV4(E), EV8(D)
3
EV5(E), EV9(D)
4
EV6(E), EV10(D)
5
6
F G H I
J K
COM
+
2
3
+
Input 1(K)
Input 3(J)
1
−
I
+
−
−
Voltage output 12 V 21 mA +
−
OUT1
Voltage output 12 V
21 mA or
Current output
+
−
1
2
3
4
SUB3(B), SUB8(C)
+
5
6
1
SUB5(B), SUB10(C) +
SUB4(B), SUB9(C)
Event input
3
4
4
5
6
5
6
F
F G H I
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
J K
V
+
TC
PT
2
Voltage output 12 V 21 mA +
3
−
OUT3
Voltage output 12 V
21 mA or
Current output
−
−
+
+
TC
V
I
PT
5
6
+
−
3
4
5
6
DE
1
EV3(E), EV7(D)
2
EV4(E), EV8(D)
3
EV5(E), EV9(D)
4
EV6(E), EV10(D)
5
6
2
Input 2(K)
Input 4(J)
3
4
5
6
G
Input 1(K)
Input 3(J)
1
−
+
I
−
V
PT
2
+
TC
3
4
−
−
+
I
V
PT
5
6
+
TC
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
J, K
J, K
E5AR-TPRQE3MF-FLK
Auxiliary output
(Relay output)
−
!
COM
2
2
SUB1
−
3
4
3
SUB2
+
Input power supply
depends on model.
100 to 240 VAC or 24
VAC/DC (no polarity)
4
COM
5
6
1
5
6
SUB3
B C D E
SUB4
Event input
2
3
4
5
6
J K
OUT2
OUT1
3
4
5
6
F
2
EV4
3
EV5
4
5
6
COM
Potentiometer
1
2
EV3
EV6
F G H I
O
W
Relay output
250 VAC 1 A
Closed
Open
C
+
−
I
−
−
+
+
TC
V
E
1
PT
24 VAC/DC
+
100 to 240 VAC
!
−
Input power supply
depends on model.
100 to 240 VAC or 24
VAC/DC (no polarity)
A
B C D E
1
+
1
SUB1(B), SUB6(C)
2
SUB2(B), SUB7(C) +
3
4
SUB3(B), SUB8(C)
+
SUB4(B), SUB9(C)
+
SUB5(B), SUB10(C) +
5
6
1
2
OUT4
+
Voltage output 12 V
3
40 mA
−
4
OUT3
Voltage output 12 V +
5
40 mA or
−
Current output
6
4 to 20 mA DC,
G
500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
−
COM
Event input
2
3
4
5
6
2
3
4
5
6
DE
1
EV3(E), EV7(D)
2
EV4(E), EV8(D)
3
EV5(E), EV9(D)
4
EV6(E), EV10(D)
5
6
COM
F G H I
BC
1
J K
1
2
3
B(+)
RS-485
A(−)
+
−
4
5
6
K
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
Note
Auxiliary output
(Transistor output)
+
1
B
1
A
100 to 240 VAC
−
24 VAC/DC
+
2
COM
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
E5AR-TPR4DF
BC
1
+
2
3
OUT4
4
−
4
5
6
G
A(−)
+
1
1
Input 2(K)
Input 4(J)
RS-485
OUT2
+
SUB2(B), SUB7(C)
2
4
5
6
−
COM
SUB1(B), SUB6(C) +
B C D E
A
1
!
B(+)
Event input
−
OUT4
Current output
4 to 20 mA DC, 500 Ω max.
+
0 to 20 mA DC, 500 Ω max.
(Switched by output
−
type setting.)
+
OUT3
Current output
−
4 to 20 mA DC, 500 Ω max.
0 to 20 mA DC, 500 Ω max.
(Switched by output type setting.)
100 to 240 VAC
+
A
1
SUB1(B), SUB6(C) +
−
!
−
Input power supply
depends on model.
100 to 240 VAC or 24
VAC/DC (no polarity)
−
COM
100 to 240 VAC
+
+
24 VAC/DC
BC
1
−
Auxiliary output
(Transistor output)
OUT2
OUT1
Potentiometer
O
1
2
3
4
5
6
F
W
Relay output
250 VAC 1 A
Closed
C
+
−
Open
I
1
2
3
4
−
−
+
+
V
PT
TC
5
6
K
(Thermocouple)
(Current) (Voltage)
(Resistance temperature input sensor)
The power supply voltage must be 100 to 240 VAC or 24 VAC/DC for the E5AR-T to comply with CE
marking requirements.
The power supply voltage must be 100 to 120 VAC or 24 VAC/DC for the E5AR-T to comply with UL
requirements.
10
E5AR-T
Nomenclature
E5AR-T
Rear
Operation
Indicators
SUB1 SUB2 SUB3 SUB4 WAIT
FSP
Channel
Indicator
PV
Bar Graph
SV
Program Status
Indicators
CMW
Display No. 2
PRG.
SEG
MANU
OUT1 OUT2 OUT3 OUT4
PF1
Display No. 1
RST
Display No. 3
RSP HOLD
PF2
RUN/RST
Function Key 1/
Run/Reset Key
Function Key 2
Level Key
Mode Key
Down Key
Up Key
Dimensions
Note: All units are in millimeters unless otherwise indicated.
E5AR-T
110
95
11.5
96
3
SUB1 SUB2 SUB3 SUB4 WAIT
91 × 91
2
FSP
PV
SV
96
CMW
OUT1 OUT2 OUT3 OUT4
PF1
PRG.
SEG
MANU
RST
111
RSP HOLD
PF2
RUN/RST
Crimp terminal size: M3
Mounting Bracket (Accessory)
Terminal cover (E53-COV14: order separately)
Panel Cutouts
+0.8
920
110 min.
+0.8
920
120 min.
• Recommended panel thickness is 1 to 8 mm.
• Group mounting is not possible. (Maintain the specified
mounting space between Controllers.)
• When two or more Controllers are mounted, make sure
that the surrounding temperature does not exceed the
allowable operating temperature specified in the specifications.
11
E5AR-T
■ Accessories (Order Separately)
Terminal Cover
E53-COV14 (for E5AR)
95
95
28.5
10.1
Unit Label Sheet
Y92S-L1
11.8
4.8
Rubber Packing
Y92S-P4 (for DIN96 × 96)
Order the Rubber Packing separately if it becomes lost or damaged. (Refer to page 3.)
The Rubber Packing can be used to achieve an IP66 degree of protection.
(Deterioration, shrinking, or hardening of the rubber packing may occur depending on the operating
environment. Therefore, periodic replacement is recommended to ensure the level of waterproofing
specified in NEMA4. The time for periodic replacement depends on the operating environment. Be
sure to confirm this point at your site. Consider one year a rough standard. OMRON shall not be liable for the level of water resistance if the customer does not perform periodic replacement.)
The Rubber Packing does not need to be attached if a waterproof structure is not required.
ALL DIMENSIONS SHOWN ARE IN MILLIMETERS.
To convert millimeters into inches, multiply by 0.03937. To convert grams into ounces, multiply by 0.03527.
In the interest of product improvement, specifications are subject to change without notice.
12
Terms and Conditions Agreement
Read and understand this catalog.
Please read and understand this catalog before purchasing the products. Please consult your OMRON representative if you
have any questions or comments.
Warranties.
(a) Exclusive Warranty. Omron’s exclusive warranty is that the Products will be free from defects in materials and workmanship
for a period of twelve months from the date of sale by Omron (or such other period expressed in writing by Omron). Omron
disclaims all other warranties, express or implied.
(b) Limitations. OMRON MAKES NO WARRANTY OR REPRESENTATION, EXPRESS OR IMPLIED, ABOUT
NON-INFRINGEMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OF THE PRODUCTS. BUYER
ACKNOWLEDGES THAT IT ALONE HAS DETERMINED THAT THE
PRODUCTS WILL SUITABLY MEET THE REQUIREMENTS OF THEIR INTENDED USE.
Omron further disclaims all warranties and responsibility of any type for claims or expenses based on infringement by the
Products or otherwise of any intellectual property right. (c) Buyer Remedy. Omron’s sole obligation hereunder shall be, at
Omron’s election, to (i) replace (in the form originally shipped with Buyer responsible for labor charges for removal or
replacement thereof) the non-complying Product, (ii) repair the non-complying Product, or (iii) repay or credit Buyer an amount
equal to the purchase price of the non-complying Product; provided that in no event shall Omron be responsible for warranty,
repair, indemnity or any other claims or expenses regarding the Products unless Omron’s analysis confirms that the Products
were properly handled, stored, installed and maintained and not subject to contamination, abuse, misuse or inappropriate
modification. Return of any Products by Buyer must be approved in writing by Omron before shipment. Omron Companies shall
not be liable for the suitability or unsuitability or the results from the use of Products in combination with any electrical or
electronic components, circuits, system assemblies or any other materials or substances or environments. Any advice,
recommendations or information given orally or in writing, are not to be construed as an amendment or addition to the above
warranty.
See http://www.omron.com/global/ or contact your Omron representative for published information.
Limitation on Liability; Etc.
OMRON COMPANIES SHALL NOT BE LIABLE FOR SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES,
LOSS OF PROFITS OR PRODUCTION OR COMMERCIAL LOSS IN ANY WAY CONNECTED WITH THE PRODUCTS,
WHETHER SUCH CLAIM IS BASED IN CONTRACT, WARRANTY, NEGLIGENCE OR STRICT LIABILITY.
Further, in no event shall liability of Omron Companies exceed the individual price of the Product on which liability is asserted.
Suitability of Use.
Omron Companies shall not be responsible for conformity with any standards, codes or regulations which apply to the
combination of the Product in the Buyer’s application or use of the Product. At Buyer’s request, Omron will provide applicable
third party certification documents identifying ratings and limitations of use which apply to the Product. This information by itself
is not sufficient for a complete determination of the suitability of the Product in combination with the end product, machine,
system, or other application or use. Buyer shall be solely responsible for determining appropriateness of the particular Product
with respect to Buyer’s application, product or system. Buyer shall take application responsibility in all cases.
NEVER USE THE PRODUCT FOR AN APPLICATION INVOLVING SERIOUS RISK TO LIFE OR PROPERTY OR IN LARGE
QUANTITIES WITHOUT ENSURING THAT THE SYSTEM AS A WHOLE HAS BEEN DESIGNED TO ADDRESS THE RISKS,
AND THAT THE OMRON PRODUCT(S) IS PROPERLY RATED AND INSTALLED FOR THE INTENDED USE WITHIN THE
OVERALL EQUIPMENT OR SYSTEM.
Programmable Products.
Omron Companies shall not be responsible for the user’s programming of a programmable Product, or any consequence
thereof.
Performance Data.
Data presented in Omron Company websites, catalogs and other materials is provided as a guide for the user in determining
suitability and does not constitute a warranty. It may represent the result of Omron’s test conditions, and the user must correlate
it to actual application requirements. Actual performance is subject to the Omron’s Warranty and Limitations of Liability.
Change in Specifications.
Product specifications and accessories may be changed at any time based on improvements and other reasons. It is our
practice to change part numbers when published ratings or features are changed, or when significant construction changes are
made. However, some specifications of the Product may be changed without any notice. When in doubt, special part numbers
may be assigned to fix or establish key specifications for your application. Please consult with your Omron’s representative at
any time to confirm actual specifications of purchased Product.
Errors and Omissions.
Information presented by Omron Companies has been checked and is believed to be accurate; however, no responsibility is
assumed for clerical, typographical or proofreading errors or omissions.
2015.8
In the interest of product improvement, specifications are subject to change without notice.
OMRON Corporation
Industrial Automation Company
http://www.ia.omron.com/
(c)Copyright OMRON Corporation 2015 All Right Reserved.