AGILENT E5818A

Agilent
E5818A LXI Class-B Trigger Box
Data Sheet
Features
Introduction
IEEE 1588 Benefits
• Remote access to
instruments via LAN
The Agilent E5818A LXI Class-B
Trigger Box is a device for handling
the IEEE 1588 Precision Time
Protocol (PTP) synchronization over
the Ethernet and for managing the
precision trigger control signals to
the legacy instrument that are not
compliant with LXI standards. The
IEEE 1588 PTP synchronization is
performed over the Ethernet by the LXI
trigger box.
The E5818 is adopting the IEEE
1588 precision time protocol (PTP)
synchronization over the Ethernet
network in nanosecond range
accuracy with minimal network and
local clock computing resources.
IEEE 1588 provides synchronization
accuracy as it meets current user
requirements and future user
requirements. It is using multicast
communications to synchronize time
between modules within a system.
• IEEE 1588 Time
Synchronization
• Output trigger at precise
timing
• Input trigger at precise
timing
• Peer-to-peer messaging
The E5818 LXI trigger box utilizes LXI
and IEEE 1588 to record and control
asynchronous events via TTL signal
input and output ports.
• Supports DHCP Protocol
• Built-in Web Access
E5818A Front and Rear Panel
Displays
• Battery supported RTC
A new IEEE 1588 device on the
network listens for the master clock
to send out its clock specification.
Every device on the network compares
its specification to the master. IEEE
1588 works best during asynchronous
events and asynchronous control. IEEE
1588 brings a common sense of time
to all LXI Class-B and devices.
IEEE 1588 is designed to fill a niche,
which is not well served by either
of the two dominant protocols, NTP
and GPS. IEEE 1588 is designed for
local systems that require a very high
degree of accuracy which goes beyond
what can be provided by NTP. It is also
designed for applications that cannot
bear the cost of a GPS receiver at
each node, or for which GPS signals
are inaccessible. 1
• Multiple Triggering Options
• High precision TCXO
• Simplifies event correlation
IEEE 1588 can accommodate a wide
range of systems and is designed
for supporting industrial network
protocols.
1
From John C. Eidson, Measurement, Control
and Communication Using IEEE 1588.
2
LXI Class-B Benefits
Application
The E5818A LXI Trigger Box is the
world’s first LXI Class-B triggering
device for precise synchronization
of non-LXI Class-B instruments over
LAN.
The E5818 is compatible with SCPI
and can be configured via a web
browser. For greater flexibility, it also
comes system-ready with a bundled
IVI driver so that users can easily
automate measurements and triggers
in the programming environment
of their choice, such as Agilent
VEE, Microsoft Visual Basic, C and
LabVIEW.
The E5818 provides a new type of
synchronization by using the IEEE
1588 standard for a PTP, which
enables precise time triggering and
stamping of events in the nanosecond
range. The new synchronization
is crucial, especially in automated
test systems that involve multiple
distributed devices with asynchronous
events.
The robust design of the E5818 trigger
box guarantees reliable operations for
demanding industries such as aviation,
automotive, test and measurement
and telecommunication industries.
The E5818 is able to support peerto-peer messaging and multicast
communication which can improve
test time, especially when heavy data
usage occur between devices.
The E5818 also provides ease of use
as it has the built-in web access to
communicate E5818 from a supported
web browser (Internet Explorer 6.0 or
higher or Firefox 2.0 or higher). You
can use this web access function to
configure and upgrade the E5818 in
LAN environment.
Compared to GPIB, LXI, and PXI, the
E5818 reduces the overall overhead
costs.
Furthermore, the E5818 provides
a more cost-effective solution for
demanding industries. The E5818
reduces the test system setup time
as it uses standard drivers such
as IVI drivers, reusing the existing
instruments and software, improving
the system throughput via LAN and
distributed intelligence, reducing
the size of the system and building
a versatile system to address future
needs.
3
Product Specifications
GENERAL REQUIREMENTS
Minimum System Requirements for
Agilent IO Libraries Suite
• Windows 2000/XP/Vista
• 450 MHz Pentium II (1 GHz 32-bit is recommended for Vista)
• 128 MB RAM (512 MB or greater is recommended for Vista)
(OS and Microsoft.NET Framework may require additional resources)
• 400 MB free disk space
Minimum System Requirements
(Client Computers)
• Available 10BASE-T/100BASE-TX LAN port
Supported Web Browser
• For E5818 web access: Internet Explorer 6.0 or higher, or Firefox 2.0 or
higher (Internet Explorer 7.0 is recommended for Vista)
Supported Standards
• VXI-11 Protocol
• IEEE-488.2 Compatible
• 10BASE-T/100BASE-TX Networks
• EIA-232
• LXI Standard Revision 1.1
• IEEE 1588 — 2002
Programming Language
• SCPI Command
• IVI-COM
• Web Controller
GENERAL CHARACTERISTICS
Time Stamp Input
Input has TTL compatible logic levels
Time Trigger Output
Output provides TTL compatible logic levels
Periodic Trigger of 1 Pulse Per Second
(1PPS)
For synchronization diagnostic use only. Issuing 1 Hz signal at 1 TTL output level
(rising edge).
LAN
Allows communication with an external controller
AC Input
Input Voltage Range: 100-240 V ±10% @ 47 to 63 Hz
Power: 25 VA peak (7.5 Watt typical)
Protection Class
Protection Class 1
Connectors
9-pin RS-232 (for debugging purposes), RJ-45 LAN, BNC Connectors
Power Cord
US/Canada: P/N: 8120-1378 (UL Listed, CSA Certified) 10A/250V. All other cord
sets: Appropriate foreign country approval associated with cord for per country
of use.
Dimensions
With bumper: 227.3 mm (W) X 237.4 mm (L) X 59.3 mm (H)
Without bumper: 212.3 mm (W) X 230.4 mm (L) X 43.25 mm (H)
Weight
3.085 kg
Rack Mount Kit Dimensions
Full rack: 242.8 mm (W) X 238.05 mm (L) X 43.25 mm (H)
Half rack: 27.5 mm (W) X 226.2 mm (L) X 43.25 mm (H)
Non-volatile Memory Battery
Lithium Manganese Dioxide, approximate lifetime 5 years at 25 °C
User Memory Clear
Statement:
• All user-configurable data may be erased from the system without trace for
security purposes
• All LAN configurations can be reset through the LAN Reset button
• All Trigger Box configurations are stored in a flash file system and can be
deleted. However, once deleted from the system, the user has to restart all the
applications manually.
BNC Protection circuit
All BNCs will have its own set of overcurrent protection circuits
Indicators
PWR, LAN, 1588
Safety Compliance
Certified by: IEC 61010-1:2001/EN 61010-1:2001, CAN/CSA-C22.2 No. 61010-104, IEC 61326-2002/EN 61326:1997+A1:1998+A3:2003
Warranty
1 year
4
GENERAL SYSTEM SPECIFICATIONS
Input Characteristics
Input Impedance
15M Ohm
Input Threshold:
Voltage Threshold High
+2.3 V
Voltage Threshold Low
+0.8 V
Minimum Pulse Width
20 ns
Output Characteristics
Output Impedance
50 Ohm nominal
Output Threshold:
Output High
1.7 V typical for 50 Ohm load; 4.3 V typical for 1M Ohm load
Output Low
0 V typical for 50 Ohm load; 0 V typical for 1M Ohm load
Output Current
+/−24 µA, max
Square Wave Rise/Fall Time
(10 to 90%) for 50 Ohm load: 0.95 ns min, 1.15 ns max
Internal Timer Resolution
20 ns
TCXO Characteristics
Frequency
50 MHz
Initial Accuracy
+/−2.0 ppm
Aging Per Year
+/−1.0 ppm
Temperature Stability (0 to 55 °C)
+/−2.5 ppm (−30 to 75%)
1588 Synchronization Accuracy 1
1.5 m Ethernet Direct Connection
+/−2 ns mean, 10 ns standard deviation
2
+/−3 ns mean, 17 ns standard deviation
Via Hirschmann Boundary Clock
Via Switch 3
+/−12 ns mean, 11 ns standard deviation
1
The master unit was configured to connect to an external clock source for greater accuracy. The slave unit clock is based on the TCXO. A sync
interval for E5818 and boundary clock was set to 1 s. Ethernet connection speed is 100 Mb/s. For E5818 connection using switch and boundary
clock, no 1588 traffic was present as the network is isolated. Measurements were taken in room temperature for a duration of 8 hours for
each configuration.
2
Hirschmann MM3-4TX1-RT with PTP precision between 2 modules < 80 ns
3
DLink - DES1005D switch used
TIME STAMP SYSTEM SPECIFICATIONS
Time Stamp Latency
62 ns
Time Stamp Input
2 ports
Event Log File
5000 data
Time Stamp Resolution
Min: 100 µs
Max: not applicable
Time Stamp Polarity
Rising or Falling Edge
Minimum Pulse Width
20 ns
5
TIME TRIGGER SYSTEM SPECIFICATIONS
Time Trigger Latency
67 ns
Trigger Pulse Width
1 µs
Time Trigger Output
2 ports
Event Log File
5000 data
Periodic Trigger
Supported
Programmability
Yes (SW)
Trigger Polarity
Rising or Falling Edge
Trigger Offset
12 hours (max)
Trigger Period
Min: 100 µs
Max: 12 hours
MINIMUM TRIGGER PERIOD AND TIMESTAMP RESOLUTION SETTINGS 4
TTL1/TTL2 5
EXT1/EXT2
100 µs
5
400 µs
TTL1/TTL2 + peer-to-peer LAN message 6
EXT1/EXT2 + peer-to-peer LAN message
4.5 ms
6
TTL(n) + EXT(n), with n ranging from 1 to 2
1 ms
7
200 µs
TTL(n) + EXT(n), with n ranging from 1 to 2 +
peer-to-peer LAN message 8
7 ms
TTL1 + TTL2 + EXT1 + EXT2 9
300 µs
TTL1 + TTL2 + EXT1 + EXT2 + peer-to-peer LAN
message 10
14 ms
4
Test is carried out on a network without P2P LAN messages from other LXI devices besides the device under test. Measurements were taken in
room temperature for duration of 12 hours for each configuration. These are the typical values for the setting
5
A single port, either TTL or EXT of each the E5818, is being exercised
6
A single port, either TTL or EXT of each the E5818, is being exercised, and peer-to-peer LAN message (WOR) generation upon any event
happening
7
A combination of either two ports of TTL or EXT of the E5818 is being exercised
8
A combination of either two ports of TTL or EXT of the E5818 is being exercised, and peer-to-peer LAN message (WOR) generation upon any
event happening
9
All four ports of the E5818 are being exercised
10
All four ports of the E5818 are being exercised, and peer-to-peer LAN message (WOR) generation upon any event happening
EXTERNAL TRIGGER SYSTEM SPECIFICATIONS
LXI LAN packets
Based on LAN0 to LAN7 or custom event ID with same domain
LAN MESSAGING SYSTEM SPECIFICATIONS
External Trigger
Based on Time Stamp 0 or Time Stamp 1
TRIGGER OUTPUT RESPONSE TIMES
Delay from EXT source to TTL Output
300 µs
Delay for Trigger relating to LAN Event source or
LAN Event output
Unknown
6
ENVIRONMENTAL SPECIFICATIONS
Operating Environment
0 °C to 55 °C — Indoor use, Pollution Degree 2
Installation Category
II (mains input)
Altitude (operating)
2000 m
Operating Humidity
Up to 90% at 40 °C non-condensing
Storage Humidity
Up to 90% at 65 °C non-condensing
Storage Temperature
−40 °C to +70 °C
Operating State
Continuous
Connection to Supply Mains
Detachable cord
Degree of Mobility
Portable (Bench and Cabinet)
SUPPORTED LANGUAGES AND APPLICATIONS
Applications (with Agilent Intuilink)
Microsoft Excel 97 and 2000
Microsoft Word 97 and 2000
Check the web for latest supported applications
Software Development
Visual Basic 5.0/6.0
Visual C++ 5.0/6.0
Visual Studio
Agilent VEE 5.0 or greater
BASIC for Windows
Visual Studio.NET
SUPPLEMENTARY INFORMATION
Maximum Number of Simultaneous
LXI Peer-to-peer Connections
8
SAFETY
7
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
Agilent Direct
www.agilent.com/find/agilentdirect
Quickly choose and use your test
equipment solutions with confidence.
Agilent
Open
www.agilent.com/find/open
Agilent Open simplifies the process
of connecting and programming
test systems to help engineers
design, validate and manufacture
electronic products. Agilent offers
open connectivity for a broad range
of system-ready instruments, open
industry software, PC-standard I/O
and global support, which are
combined to more easily integrate test
system development.
Remove all doubt
Our repair and calibration services
will get your equipment back to you,
performing like new, when promised. You will get full value out of
your Agilent equipment throughout its lifetime. Your equipment
will be serviced by Agilent-trained
technicians using the latest factory
calibration procedures, automated
repair diagnostics and genuine parts.
You will always have the utmost
confidence in your measurements.
Agilent offers a wide range of additional expert test and measurement services for your equipment,
including initial start-up assistance,
onsite education and training, as
well as design, system integration,
and project management.
For more information on repair and
calibration services, go to:
www.agilent.com/find/removealldoubt
www.agilent.com
For more information on Agilent Technologies’
products, applications or services, please contact
your local Agilent office. The complete list is
available at:
www.agilent.com/find/contactus
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008
Europe & Middle East
Austria
Belgium
Denmark
Finland
France
0820 87 44 11
32 (0) 2 404 93 40
45 70 13 15 15
358 (0) 10 855 2100
0825 010 700*
*0.125 € fixed network rates
Germany
01805 24 6333**
**0.14€/minute
www.lxistandard.org
LXI is the LAN-based successor to
GPIB, providing faster, more efficient
connectivity. Agilent is a founding
member of the LXI consortium.
Ireland
1890 924 204
Israel
972-3-9288-504/544
Italy
39 02 92 60 8484
Netherlands
31 (0) 20 547 2111
Spain
34 (91) 631 3300
Sweden
0200-88 22 55
Switzerland (French) 41 (21) 8113811(Opt 2)
Switzerland (German) 0800 80 53 53 (Opt 1)
United Kingdom
44 (0) 118 9276201
Other European Countries:
www.agilent.com/find/contactus
Revised: October 24, 2007
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2007, 2008
Printed in USA, March 15, 2008
5989-7290EN