POWERBOX PBSE5125

MEDICAL & MILITARY
Medline 100 Series
1 0 0 W AT T S S I N G L E O U T P U T A C / D C M E D I C A L
Features
• High power density
• 12.5W per sq inch
• Reliable design
• Small footprint 51 x 102 x 32mm
• Safety standard IEC60601-1
• Overtemperature protection
• Overcurrent protection
• Class II, Double insulated
Specifications
ENVIRONMENTAL
INPUT
Nominal input range
100-240VAC
Voltage range
90-264VAC, 85V with 70% output power.
Frequency
47-63Hz.
Inrush current
Max 50A peak, cold start.
Input Current
Max 1.4A.
Input connector
Type AMP, MTA 156, 3-pole, part no 0-0640388-3.
Leakage current
<100µA
Inrush Current
20A @ 115V, 40A @ 230V
OUTPUT
Operating Temperature
0°C to +50°C, 70°C with derating
Storage temperature
–40°C to +75°C
Humidity
5% to 95% non-condensing
Derating
Derate from 100% at 50°C linearly to 50% at +70°C.
Derate to 80% output power at convection
cooling
Cooling
Convection or fan cooling. 6CFM (10m3/h)
required for 100W continuous output
GENERAL
Switching frequency
60-100kHz.
Power
Up to 100W, continuous at 100-240VAC input,
see table
MTBF
270,000 hours at 25°C ambient temperature
Lifetime prediction
Min 80,000 hours at 25°C, 230VAC at 70% load
Output Voltage
See table. Factory preset non adjustable
STANDARDS AND APPROVALS
Ripple & Noise
Typ. 1% p-p max (20MHz BW).
5% at <100VAC input.
Safety standards
IEC60601-1 including deviations for Europe, US &
Canada by Intertek Semko
Efficiency
87-90% @ 230VAC full load
C-Tick
AS/NZS CISPR11 Group 1, Class A
Hold up time
16ms min at 230VAC
Start up time
<2s max
RFI/EMI standards
EN60601-1-2
Line regulation
1% max full load. 5% at <100VAC input
Safety markings
S, ETL & CE
Load regulation
2% max @ 230VAC, 10-90% load change at
output terminal.
Isolation class
Class II. Double isolation
Overcurrent Protection
Approx 150%, auto recovery
Current limiting
characteristics
Constant current
MECHANICAL
Dimensions
51 x 102 x 32 mm
Weight
Max 360g
Overtemperature protection Yes, auto recovery
Overvoltage protection
Yes
Transient response
4ms max
Output connector
Type AMP, MTA 156, 4-pole, part no 0-0640388-4.
Technical Illustration
344
Selection Table
MODEL
NUMBER
OUTPUT
VOLTAGE
PBSE 5125
12Vdc
PBSE 5131
13.8Vdc
PBSE 5126
15Vdc
PBSE 5127
18Vdc
PBSE 5128
24Vdc
PBSE 5132
27.5Vdc
PBSE 5129
48Vdc
(*) At 6 CFM forced air.
CONTINUOUS
OUTPUT
POWER*
CONTINUOUS
OUTPUT
POWER*
8.3A
7.2A
6.7A
5.5A
4.2A
3.6A
2.1A
100W
100W
100W
100W
100W
100W
100W