MEDICAL & MILITARY Medline 100 Series 1 0 0 W AT T S S I N G L E O U T P U T A C / D C M E D I C A L Features • High power density • 12.5W per sq inch • Reliable design • Small footprint 51 x 102 x 32mm • Safety standard IEC60601-1 • Overtemperature protection • Overcurrent protection • Class II, Double insulated Specifications ENVIRONMENTAL INPUT Nominal input range 100-240VAC Voltage range 90-264VAC, 85V with 70% output power. Frequency 47-63Hz. Inrush current Max 50A peak, cold start. Input Current Max 1.4A. Input connector Type AMP, MTA 156, 3-pole, part no 0-0640388-3. Leakage current <100µA Inrush Current 20A @ 115V, 40A @ 230V OUTPUT Operating Temperature 0°C to +50°C, 70°C with derating Storage temperature –40°C to +75°C Humidity 5% to 95% non-condensing Derating Derate from 100% at 50°C linearly to 50% at +70°C. Derate to 80% output power at convection cooling Cooling Convection or fan cooling. 6CFM (10m3/h) required for 100W continuous output GENERAL Switching frequency 60-100kHz. Power Up to 100W, continuous at 100-240VAC input, see table MTBF 270,000 hours at 25°C ambient temperature Lifetime prediction Min 80,000 hours at 25°C, 230VAC at 70% load Output Voltage See table. Factory preset non adjustable STANDARDS AND APPROVALS Ripple & Noise Typ. 1% p-p max (20MHz BW). 5% at <100VAC input. Safety standards IEC60601-1 including deviations for Europe, US & Canada by Intertek Semko Efficiency 87-90% @ 230VAC full load C-Tick AS/NZS CISPR11 Group 1, Class A Hold up time 16ms min at 230VAC Start up time <2s max RFI/EMI standards EN60601-1-2 Line regulation 1% max full load. 5% at <100VAC input Safety markings S, ETL & CE Load regulation 2% max @ 230VAC, 10-90% load change at output terminal. Isolation class Class II. Double isolation Overcurrent Protection Approx 150%, auto recovery Current limiting characteristics Constant current MECHANICAL Dimensions 51 x 102 x 32 mm Weight Max 360g Overtemperature protection Yes, auto recovery Overvoltage protection Yes Transient response 4ms max Output connector Type AMP, MTA 156, 4-pole, part no 0-0640388-4. Technical Illustration 344 Selection Table MODEL NUMBER OUTPUT VOLTAGE PBSE 5125 12Vdc PBSE 5131 13.8Vdc PBSE 5126 15Vdc PBSE 5127 18Vdc PBSE 5128 24Vdc PBSE 5132 27.5Vdc PBSE 5129 48Vdc (*) At 6 CFM forced air. CONTINUOUS OUTPUT POWER* CONTINUOUS OUTPUT POWER* 8.3A 7.2A 6.7A 5.5A 4.2A 3.6A 2.1A 100W 100W 100W 100W 100W 100W 100W