ETC DCXL20DN

SURSENSE Ultra Low Pressure Sensors
Millivolt Output
High Accuracy Compensation
Full Scale Pressure Ranges
FEATURES
• Position sensitivity to
± 5 mV/g, typical
• Offset Warm-up drift
< ± 50 mV, typical
• Temperature
Compensated over
0°C to +50 °C
• Available in Gage and
Differential Pressure
Ranges
• Long Term Stability
< ± 100 µV, typical
• Combined Linearity and
Hysteresis Error
< ± 0.25 % Span
TYPICAL APPLICATIONS
• Medical Instrumentation
• HVAC
• Industrial Instrumentation
• Environmental Controls
WARNING
PERSONAL INJURY
DO NOT USE these products
as safety or emergency stop
devices or in any other
application where failure of the
product could result in
personal injury.
Failure to comply with these
instructions could result in
death or serious injury.
Sensing and Control
DCXL Series
The DCXL line of ultra low pressure sensors is based upon

SURSENSE , a proprietary technology, which reduces all output offset or
common mode errors.
These sensors use a silicon micromachined sensing element which
features a unique stress concentration enhanced structure to provide a
highly stable linear output that is proportional to applied pressure. Output
offset errors due to changes in temperature, warm-up, long term stability
and position sensitivity have all been significantly reduced when compared
to conventional sensors.
The DCXL Series sensors provide a precision calibrated ratiometric

millivolt output using the enhanced stability of SURSENSE technology.
Each sensor features calibrated offset, full scale span and thermal error
calibration to ensure the highest possible accuracy for flow pressure
measurement. These highly stable sensors feature an industry standard
ported package with improved stress isolation for printed circuit board
mount applications.
Product is patented by US patent 6023978.
WARNING
MISUSE OF DOCUMENTATION
• The information presented in this product sheet is for reference only. Do
not use this document as product installation information.
• Complete installation, operation, and maintenance information is
provided in the instructions supplied with each product.
Failure to comply with these instructions could result in death or serious
injury.
SURSENSE Ultra Low Pressure Sensors
Millivolt Output
DCXL Series
DCXL Electrical Specifications at 12 Vdc Excitation, 25 °C
Excitation Voltage
Common Mode Voltage, % Excitation
Span*except DCXL01DN
Span*DCXL01DN
Null*
Offset Temperature Shift (0 °C to +50 °C [+32 °F to +122 °F])**
except DCXL01DN
Offset Temperature Shift (0 °C to +50 °C [+32 °F to +122 °F])**
DCXL01DN
Span Temperature Shift (0 °C to +50 °C [+32 °F to +122 °F])**
Linearity, Hysteresis Error***
Operating Temperature
Storage Temperature
Offset Warm-up Shift****except DCXL01DN
Offset Warm-up Shift****DCXL01DN
Offset Position Sensitivity (±1g) except DCXL01DN,
DCXL05DN and DCXL10DN
Offset Position Sensitivity (±1g) DCXL01DN
Offset Position Sensitivity (±1g) DCXL05DN, DCXL10DN
Offset Long Term Stability (1 year) except DCXL01DN
Offset Long Term Stability (1 year) DCXL01DN
*Note 1:
**Note 2:
***Note 3:
****Note 4:
Min.
Typ.
Max.
Units
3.0
-19
9.0
- 500
12.0
50
20.0
10.0
0
16.0
-21.0
11.0
+500
Vdc
%
mV
mV
µV
-150
--
+150
µV
- 250
-200
--25
-40
---
0
0
0.05
± 50
± 100
+250
+200
0.25
+85
+125
---
µV
µV
%Span
°C
°C
µV
µV
------
±5
± 50
± 10
± 100
± 200
------
µV
µV
µV
µV
µV
The voltage added to the offset voltage at full scale pressure. Nominally the output voltage range is 1.0 V to 6.0 V.
Shift is relative to 25 °C
Measured at ½ full scale rated pressure using BFSL
Shift is within the first hour of excitation applied to the device.
DCXL Series Pressure Ratings in H2O (by Catalog Listing)
DCXLOPERATING PRESSURE RANGE
PROOF
BURST
COMMON MODE
01D
1.0
100
200
50
05D
5.0
150
300
50
10D
10.0
150
300
50
20D
20.0
300
450
50
DCXL Series Resistance (by Catalog Listing)
DCXL
INPUT RESISTANCE
OUTPUT RESISTANCE
01D
4.5
1.5
05D
10
1.5
2 Honeywell • Sensing and Control
10D
13
1.5
20D
10
2
30D
12
1.5
UNIT
KO
KO
30D
30.0
450
750
50
SURSENSE Ultra Low Pressure Sensors
Millivolt Output
DCXL Series
MOUNTING DIMENSIONS (for reference only) mm/in
DCXL SERIES, DCXL CERAMIC PACKAGE
APPLICATION INFORMATION
MEDIA COMPATIBILITY, WETTED MATERIALS
Port 2
P2 Port: Media must be compatible
with nylon housing, epoxy adhesive
and silicon.
Port 1
P1 Port: Dry Gases Only. Media
must be compatible with epoxy
based adhesive.
Testing to Non
Excitation Voltage
Standard Conditions,
Common mode pressure. Standard
changes to:
testing has positive pressure to port
2 with ambient pressure to port 1.
Ratiometric Output: The output voltage of the sensor is
ratiometric, proportional, to the excitation voltage. For this
model sensor, all specifications will change proportionally
to any changes in the excitation voltage. The excitation
may vary between 3vots to 16 volts. All specifications will
nominally be changed by a ratio of V EXCITATION/12.0
volts. For example: if the excitation voltage is 5.0 volts,
then both the full scale output voltage and the offset
voltage nominals would be 5/12th the specified value.
Equivalent Circuit
Pinout
1.
N/C
2.
+V Supply
3.
+V Output
Pressure Compatibility:
4.
-V Supply
Measures differential or gage pressure and vacuum.
Pressure may be applied to either port. For pressure to the
low pressure port, the output polarity is reversed.
5.
-V Output
6.
N/C
Honeywell • Sensing and Control 3
SURSENSE Ultra Low Pressure Sensors
Millivolt Output
DCXL Series
DCXL SERIES ORDER GUIDE
(Not all product combinations are released.)
DCXL
05
D
N
Series
Package Type
DCXL Compensated and calibrated, mV
Pressure Range
01, 05, 10, 20, 30 inches H2 O
WARRANTY/REMEDY
Honeywell warrants goods of
its manufacture as being free of
defective materials and faulty
workmanship. Contact your local
sales office for warranty
information. If warranted goods
are returned to Honeywell during
the period of coverage,
Honeywell will repair or replace
without charge those items it
finds defective. The foregoing is
Buyer’s sole remedy and is in
lieu of all other warranties,
expressed or implied,
including those of
merchantability and fitness for
a particular purpose.
N- Plastic
Pressure Reference
G - Gage
D - Differential
Specifications may change
without notice. The information
we supply is believed to be
accurate and reliable as of this
printing. However, we assume
no responsibility for its use.
While we provide application
assistance personally, through
our literature and the Honeywell
web site, it is up to the customer
to determine the suitability of the
product in the application.
For application assistance,
current specifications, or name of
the nearest Authorized
Distributor, contact a nearby
sales office. Or call:
1-800-537-6945 USA
1-800-737-3360 Canada
1-815-235-6847 International
FAX
1-815-235-6545 USA
INTERNET
www.honeywell.com/sensing
[email protected]
Sensing and Control
Honeywell
11 West Spring Street
Freeport, Illinois 61032
Printed with Soy Ink
on 50% Recycled Paper
008055-2-EN IL50 GLO 0801 Printed in USA
Copyright  1997- 2001 by Honeywell Data Instruments Inc. All rights reserved.