NI NI5105

60 MS/s, 60 MHz, 12-Bit, 8-Channel Digitizers
NI 5105
• 8 channels simultaneously
sampled at 12-bit resolution
• 60 MS/s real-time sampling
• 60 MHz bandwidth
• 50 mVpp to 30 Vpp input range
• 72 dBc SFDR
• 16, 128, or 512 MB of onboard memory
• Edge, window, hysteresis, and
digital triggering
Calibration
• Gain, offset, frequency response, and
timing self-calibration
• 2-year external calibration interval
Operating Systems
• Windows Vista/XP/2000
• LabVIEW Real-Time
Recommended Software
• LabVIEW
• LabWindows™/CVI
• Measurement Studio for
Visual Studio
• LabVIEW SignalExpress
Driver Software (included)
• NI-SCOPE driver
• LabVIEW Express VIs
• Scope Soft Front Panel
Overview
National Instruments 5105 high-resolution
digitizers feature eight 60 MS/s
simultaneously sampled input channels
with 12-bit resolution, 60 MHz bandwidth,
and up to 512 MB of memory in a compact,
3U PXI/PXI Express or PCI device. An
NI 5105 uses the National Instruments
Synchronization and Memory Core (SMC)
architecture, so you can combine multiple
devices to build up to 136 phase-coherent channels in a single
PXI chassis. You can also synchronize an NI 5105 with other analog and
digital instruments to create mixed-signal test systems. An NI 5105 is
ideal for a wide range of applications including ultrasonic nondestructive
test (NDT), medical imaging, scientific research, military/aerospace, and
consumer electronics.
Applications
Imaging
Ultrasonic nondestructive test
Optical coherence tomography
Medical imaging
Aerospace/Defense
RADAR, SONAR, and LIDAR
Satellite
Signal intelligence
Consumer Electronics
DVD, DVD-R, and PVR
Set-top box
Gaming console
Communications
xDSL
Eight 60 MS/s, 12-Bit Input Channels
for Time and Frequency Analysis
• 60 MHz input bandwidth with antialias and noise filters
• >72 dBc spurious-free dynamic range (SFDR)
• Independent channel-selectable 50 mVpp to 30 Vpp input ranges
• Independent channel-selectable 50 Ω or 1 MΩ input impedance
• 2-year calibration interval and 0 to 55 °C operating temperature
Deep Onboard Memory
• 16, 128, or 512 MB of onboard memory
• Capture more than 1 million triggered waveforms in multiple
record mode, with hardware trigger rearming
• Stream data continuously from onboard memory to
host memory or disk
Triggering, Clocking, and Synchronization
• Edge, window, hysteresis, and digital triggering
• Pretrigger and posttrigger acquisition in single- and
multiple-record mode
• Internal 60 MHz clock or external clock from 4 to 65 MHz
• Phase lock to PXI 10 MHz reference or external reference
from 1 to 20 MHz
• Timestamp-triggered events with 100 ps resolution
Software
• IVI-compliant NI-SCOPE driver for NI LabVIEW and LabWindows/CVI
as well as Microsoft C++ and Visual Basic with more than
50 built-in measurements
• Scope Soft Front Panel for interactive control
Ordering Information
NI PXI-5105 ........................................................................779685-0M1
NI PCI-5105 ........................................................................779686-0M1
1M
(onboard memory): 1 (16 MB), 2 (128 MB), 3 (512 MB)
Includes NI-SCOPE driver and Scope Soft Front Panel.
Recommended PXI Switch
NI PXI-2593 (500 MHz mux/matrix) ......................................778793-01
BUY NOW!
For complete product specifications, pricing, and accessory
information, call 800 813 3693 (U.S.) or go to ni.com/digitizers.
60 MS/s, 60 MHz, 12-Bit, 8-Channel Digitizers
Specifications
These specifications are valid for 0 to 55 °C‚ unless otherwise stated.
Acquisition System
Number of channels............................ 8 simultaneously sampled
Vertical resolution............................... 12 bits
Bandwidth (-3 dB)
50 Ω
60 MHz
55 MHz
Full-Scale Input Range
All except 50 mV
50 mV
Bandwidth limit filters
(software selectable) ..........................
Maximum sample rate........................
Onboard sample memory....................
Multirecord mode
Maximum number of records .........
1 MΩ
60 MHz
35 MHz
24 MHz antialias
60 MS/s real-time
16, 128, or 512 MB
1 MΩ
50 mV range:
±(1% of input + 0.25% of FS + 600 µV)
200 mV, 1 V, and 6 V ranges:
±(0.65% of input + 0.25% of FS + 600 µV)
30 V range:
±(0.75% of input + 0.25% of FS + 600 µV)
Channel-to-channel crosstalk ............. ≤-80 dB at 1 MHz
Spectral Characteristics (typical)
Dynamic performance (50 Ω input impedance with
10 MHz, -1 dBFS input signal)
THD (dBc)
-75
-75
-75
-75
THD (dBc)
-72
-75
-65
-68
SINAD (dB)
50
59
61
59
SINAD = Signal-to-noise and distortion
DC accuracy (0 V offset setting)
SFDR (dBc)
–
72
72
72
SFDR (dBc)
–
70
65
65
THD = Total harmonic distortion
Accuracy
Full-Scale Input Range
50 mV
200 mV
1V
6V
Full-Scale Input Range
50 mV
200 mV
1V
6V
SFDR = Spurious-free dynamic range
Onboard memory/(512 B x
number of enabled channels)
Input impedance ................................. 50 Ω and 1 MΩ || 50 pF,
software selectable
Full-scale input range
50 Ω................................................ 50 mV, 200 mV, 1 V, 6 V
1 MΩ .............................................. 50 mV, 200 mV, 1 V, 6 V, 30 V
Maximum input overload
50 Ω ............................................... 7 Vrms with | peaks | ≤10 V
1 MΩ .............................................. | peaks | ≤42 V
Input coupling ..................................... AC, DC (AC coupling on 1 MΩ only)
AC coupling cutoff frequency (-3 dB).. 12 Hz
50 Ω
All ranges:
±(1% of input + 0.25% of FS + 600 µV)
Dynamic performance (1 MΩ input impedance with
10 MHz, -1 dBFS input signal)
SINAD (dB)
59
62
62
62
RMS Noise (24 MHz filter enabled)
Full-Scale Input Range
50 mV
200 mV
1V
6V
30 V (1 MΩ only)
50 Ω
19 µVrms
56 µVrms
290 µVrms
1.68 mVrms
–
1 MΩ
60 µVrms
72 µVrms
300 µVrms
2.16 mVrms
9 mVrms
Timebase System
Timebase options................................ Internal, PXI star, external (PFI 1)
Internal
Internal sample clock frequency ........ 60 MS/s sampling rate with
decimation by n where
1 ≤ n ≤ 65,535
Timebase accuracy.............................. ±25 ppm (±0.0025%)
External
External clock range............................ 8 to 65 MHz, variable with
decimation by n where
1 ≤ n ≤ 65,535
External reference sources ................. PFI 1 (SMB connector), PXI
backplane 10 MHz
External reference range .................... 1 to 20 MHz in 1 MHz increments
External clock/reference amplitude.... Sine wave: 0.65 to 2.8 Vpp
(0 to 13 dBm)
Square wave: 0.2 to 2.8 Vpp
External clock/reference impedance .. 50 Ω, AC coupled
Trigger System
Modes ................................................. Edge, hysteresis, window,
digital, immediate, software
Sources................................................ CH 0 to CH 7, PXI_Trig <0..6>,
PFI 1, PXI star, software
Slope ................................................... Rising or falling
Hysteresis............................................ Fully programmable
Sensitivity............................................ 2% FS
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)
2
60 MS/s, 60 MHz, 12-Bit, 8-Channel Digitizers
Intermodule SMC Synchronization Using
NI-TClk (typical)
Skew.................................................... 500 ps
<10 ps after manual adjustment
Power Requirements (typical)
PXI
PCI
+3.3 VDC
1.5 A
1.7 A
+5 VDC
1.7 A
2A
+12 VDC
200 mA
20 mA
-12 VDC
25 mA
OA
Total Power
16.15 W
15.85 W
Environment
Operating temperature1 ...................... 0 to 55 °C (meets
IEC-60068-2-1 and IEC-60068-2-2)
Storage temperature........................... -40 to 70 °C (meets
IEC-60068-2-1 and IEC-60068-2-2)
Relative humidity ................................ 10 to 90%, noncondensing
(meets IEC-60068-2-56)
10
to 45 °C in PXI-101x and 1000/B chassis.
For access to certifications, marks, and DoCs, visit
ni.com/certification.
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)
3
NI Services and Support
NI has the services and support to meet
your needs around the globe and through
the application life cycle – from planning
and development through deployment
and ongoing maintenance. We offer
services and service levels to meet
customer requirements in research,
design, validation, and manufacturing.
Visit ni.com/services.
Training and Certification
NI training is the fastest, most certain route to productivity with our
products. NI training can shorten your learning curve, save development
time, and reduce maintenance costs over the application life cycle. We
schedule instructor-led courses in cities worldwide, or we can hold a
course at your facility. We also offer a professional certification program
that identifies individuals who have high levels of skill and knowledge on
using NI products. Visit ni.com/training.
Professional Services
Our NI Professional Services team is composed of NI applications
and systems engineers and a worldwide National Instruments Alliance
Partner program of more than 600 independent consultants and
integrators. Services range
from start-up assistance to
turnkey system integration.
Visit ni.com/alliance.
OEM Support
We offer design-in consulting and product integration assistance if you
want to use our products for OEM applications. For information about
special pricing and services for OEM customers, visit ni.com/oem.
ni.com • 800 813 3693
National Instruments • [email protected]
Local Sales and Technical Support
In offices worldwide, our staff is local to the country, giving you access
to engineers who speak your language. NI delivers industry-leading
technical support through online knowledge bases, our applications
engineers, and access to 14,000 measurement and automation
professionals within NI Developer Exchange forums. Find immediate
answers to your questions at ni.com/support.
We also offer service programs that provide automatic upgrades to
your application development environment and higher levels of technical
support. Visit ni.com/ssp.
Hardware Services
NI Factory Installation Services
NI Factory Installation Services (FIS) is the fastest and easiest way to
use your PXI or PXI/SCXI combination systems right out of the box.
Trained NI technicians install the software and hardware and configure
the system to your specifications. NI extends the standard warranty by
one year on hardware components (controllers, chassis, modules)
purchased with FIS. To use FIS, simply configure your system online
with ni.com/pxiadvisor.
Calibration Services
NI recognizes the need to maintain properly calibrated devices for
high-accuracy measurements. We provide manual calibration
procedures, services to recalibrate your products, and automated
calibration software specifically designed for use by metrology
laboratories. Visit ni.com/calibration.
Repair and Extended Warranty
NI provides complete repair services for our products. Express repair
and advance replacement services are also available. We offer
extended warranties to help you meet project life-cycle requirements.
Visit ni.com/services.
*351444B-01*
351444B-01
2006-7317-101-D
© 2007 National Instruments Corporation. All rights reserved. CVI, LabVIEW, Measurement Studio, National Instruments, National Instruments Alliance Partner, NI, ni.com,
SCXI, and SignalExpress are trademarks of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Other product and company names listed
are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from National Instruments and has no agency,
partnership, or joint-venture relationship with National Instruments.