ES51965(6600counts) Dual Display/PEAK Features • Low pass filter auto switching mode for ACA/ACV measurement • 6,600 counts dual LCD display • Band-gap reference voltage output • 128L QFP package • CE selection (DC/AC 1010V, DC/AC 610V) • 3.3V DC power supply • Serial data output (RS232 format) • Slow ADC Conversion rate : 2.8 times/s • LCD segment check when power on • Bar-graph ADC conversion rate: 28 times/s • Auto power off (30min/15min) • Full automatic measurement • Sleep state indicative signal output *Voltage measurement: 660.0mV – 1000V • Re-power on *Current measurement: μA/mA/A • On-chip buzzer driver *Frequency sub-display with Voltage or Current • Low battery detection mode (VAHz): 66.00Hz ~ 66.00kHz *Resistance measurement: 660.0Ω – 66.00MΩ *Resistance measurement for Clampmeter mode: 660.0Ω – 66.00KΩ *Capacitance measurement: 6.600nF – 66.00mF *Frequency counter with duty cycle display: 66.00Hz – 66.00MHz 1% – 99% (<10kHz) • Current modes for clamp meter with ZERO function • AC modes with frequency dual display • Diode measurement • Continuity check • 4 ADP modes with external reference voltage and independent “ADP” segment on LCD panel • Temperature mode with internal scale translation circuit from 0C to 0F • Push functions: *KEY function *PEAK function *Range change function *Relative function Description ES51965 is an integrated analog-to-digital converter (ADC) with dual 6,600-count LCD display (4-digit main-display and 4-digit sub-display), automatic range selection, and 3.3V DC power supply. Automatic range selection is provided for voltage (AC/DC) measurement, resistance measurement, current measurement, capacitance measurement, and frequency counter. Expensive and bulky mechanical range switches are not required. Other features include peak-hold, relative value display, offset removing feature for DCA clampmeter, data holding, maximum and minimum value holding, duty cycle measurement, diode measurement, temperature measurement, continuity checking, low battery detection, auto power off, re-power on, backlight driver, buzzer driver and RS232 data output. *Zero function: DCA clampmeter only *Data hold function with data record feature *MAX/MIN function *Back light function v2.5 1 09/04/14 ES51965(6600counts) Dual Display/PEAK Application Digital multimeter Clamp meter Pin Assignment 2 3 4 102 101 100 99 98 97 96 95 94 93 92 91 90 89 88 87 86 85 84 83 82 81 80 79 78 77 76 75 74 73 72 71 70 69 68 67 66 65 1 C MAX/MIN PEAK BKLIT RS232 BUZOUT BKOUT OSC2 OSC1 SHIFT APOSEL VST LPFOFF CESEL FC5 FC4 FC3 FC2 FC1 SLADC SDO C+ CLBAT9 VVV+ ES51965_QFP_128 V+ DGND AGND AGND CH+ CHCIH BUFH CAZH CL+ CLCIL CAZL BUFL RAZ IVSH IVSL OVX OVH OVSG OR1 VR5 VR4 VR3 VR2 TEST5 ACVL ACVH ADI ADO OHM C3 OHM C2 OHM C1 VRH VR VA+ VASGND B 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120 121 122 123 124 125 126 127 128 KEY REL RANGE HOLD LCDCAP ANNUC BP1 BP2 BP3 BP4 SEG01 SEG02 SEG03 SEG04 SEG05 SEG06 SEG07 SEG08 SEG09 SEG10 SEG11 SEG12 SEG13 SEG14 SEG15 SEG16 SEG17 SEG18 SEG19 SEG20 SEG21 SEG22 SEG23 SEG24 SEG25 SEG26 SEG27 SEG28 D D SEG29 SEG30 SEG31 SEG32 SEG33 SEG34 NC SEG35 FREQ BUZIN VBAR SLEEP LPFOUT LPC3 LPC2 LPC1 CSHCSH+ PMAX PMIN R1K R9K CA+ CAVR1 ADP C B A 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 A 64 63 62 61 60 59 58 57 56 55 54 53 52 51 50 49 48 47 46 45 44 43 42 41 40 39 1 v2.5 2 3 2 4 09/04/14 ES51965(6600counts) Dual Display/PEAK Pin Description Pin No 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 Symbol V+ DGND AGND AGND CH+ CHCIH BUFFH CAZH CL+ CLCIL CAZL BUFFL RAZ IVSH IVSL OVX OVH OVSG OR1 VR5 VR4 VR3 VR2 TEST5 ACVL Type O P/G P/G P/G IO IO O O O IO IO O O O O I I I O O O O O O O O O 28 ACVH O 29 30 31 32 33 34 35 36 ADI ADO OHMC3 OHMC2 OHMC1 VRH VR VA+ I O O O O O I I 37 VA- I 38 39 40 41 42 43 44 45 46 47 48 49 SGND ADP VR1 CACA+ R9K R1K PMIN PMAX CSH+ CSHLPFC1 G I I IO IO O O O O IO IO O v2.5 Description Output of on-chip DC-DC converter. Digital ground. Analog ground. Analog ground. Positive connection for reference capacitor of high-speed A/D. Negative connection for reference capacitor of high-speed A/D. High-speed integrator output. Connect to integral capacitor. High-speed buffer output pin. Connect to integral resistor. High-speed auto-zero capacitor connection. Positive connection for reference capacitor of high-resolution A/D. Negative connection for reference capacitor of high- resolution A/D. High-resolution integrator output. Connect to integral capacitor. High-resolution auto-zero capacitor connection. High-resolution Buffer output pin. Connect to integral resistor Buffer output pin in AZ and ZI phase. Current measurement input for 6600μA, 660mA and 66A modes. Current measurement input for 660μA, 66mA. Sense input for resistance/capacitance measurement Output connection for resistance measurement Sense low voltage for resistance/voltage measurement Reference resistor connection for 660.00Ω range Voltage measurement ÷10000 attenuator(1000V) Voltage measurement ÷1000 attenuator(660.0V) Voltage measurement ÷100 attenuator(66.00V) Voltage measurement ÷10 attenuator(6.600V) Buffer output of OVSG DC signal low input in ACV/ACA mode. Connect to negative output of external AC to DC converter. DC signal high input in ACV/ACA mode. Connect to positive output of external AC to DC converter. Negative input of internal AC to DC OP Amp. Output of internal AC to DC OP Amp. Filter capacitor connection for resistance mode. Filter capacitor connection for resistance mode. Filter capacitor connection for resistance mode. Output of band-gap voltage reference. Typically –1.23V Reference input voltage connection. Typically –400mV For ADP mode. De-integrating voltage positive input. The input should be higher than VA-. For ADP mode. De-integrating voltage negative input. The input should be lower than VA+. Signal Ground. Measurement input in ADP mode. Measurement Input. Connect to a precise 10MΩ resistor. Negative auto-zero capacitor connection for capacitor measurement Positive auto-zero capacitor connection for capacitor measurement Connect to a precise 9KΩ resister for capacitor measurement. Connect to a precise 1KΩ resister for capacitor measurement. Minimum peak hold output Maximum peak hold output. Cap for duty-cycle mode Cap for duty-cycle mode Capacitor C1 connection for internal low-pass filter 3 09/04/14 ES51965(6600counts) Dual Display/PEAK 50 51 52 53 54 LPFC2 LPFC3 LPFOUT SLEEP VBAR 55 BUZIN 56 FREQ 57,59-92 SEG35 – SEG01 93 BP4 94 BP3 95 BP2 96 BP1 97 ANNUNC O O O O I I I O O O O O O 98 99 100 101 102 LCDCAP HOLD RANGE REL_ZERO KEY I I I I I 103 104 105 106 107 MAX/MIN PEAK BKLIT RS232 BUZOUT I I I I O 108 BKOUT O 109 110 111 112 113 114 115 116 117 118 119 120 121 122 123 124 125 OSC2 OSC1 SHIFT APOSEL VST LPFOFF CESEL FC5 FC4 FC3 FC2 FC1 SLACDC SDO C+ CLBAT9 O I I I I I I I I I I I I O O O I 126 127 128 VVV+ P P O v2.5 Capacitor C2 connection for internal low-pass filter Capacitor C3 connection for internal low-pass filter Capacitor C1 connection for internal low-pass filter Sleep mode indicator, asserts low in SLEEP mode. In temperature mode, it is used to control decimal point. In µA or mA modes, it is used to control the ‘µ’ or ‘m’ sign. Pull to V- to enable the BUZOUT. Frequency counter input, offset V-/2 internally by the chip. LCD segment line 01 – 35 LCD backplane 4 LCD backplane 3 LCD backplane 2 LCD backplane 1 Square wave output at the backplane frequency, synchronized to BP1. ANNUNC can be used to control display annunciator. Connect a LCD segment to ANNUNC to turn it on; connect an LCD segment to its backplane to turn it off. LCD bias voltage bypass capacitor Pulse to V- to enable HOLD function. Pulse to V- to enable manual mode and manual range selection. Pulse to V- to enable Relative function or Zero function. Pulse to V- to change mode. In ADP mode, if this pin is connected to V-, the buzzer output will be off when the ADP input overflows. Pulse to V- to enable MAX/MIN function. Pulse to V- to enable Peak MAX/Peak MIN hold function Back light function. Pulse low to set BKOUT pin output. Assert low to enable serial data output. Outputs a 2KHz audio frequency signal for driving piezoelectric buzzer when BUZIN is low. If BKLIT function is enabled, this pin will change from V- to V+ For 60 sec, once press BKLIT pin again within 60 sec, this pin will Change back to V-. Crystal oscillator output connection Crystal oscillator input connection Pull to V- to enable display-shift feature on duty cycle mode Idle time selection for auto power off feature. Pull to V- to enable the auto range voltage mode to start from 660.0V Pull to V- to disable the internal Low pass filter in measurement CE selection feature control pin. Switch 5 for function selection. Switch 4 for function selection. Switch 3 for function selection. Switch 2 for function selection. Switch 1 for function selection. Select initial state. Serial data output Positive capacitor connection for on-chip DC-DC converter. Negative capacitor connection for on-chip DC-DC converter. Low battery configuration input. If battery is used, the low battery segment is displayed when the voltage of this pin is less than VRH (-1.23V) Negative supply voltage. Negative supply voltage. Output of on-chip DC-DC converter. 4 09/04/14 ES51965(6600counts) Dual Display/PEAK Absolute Maximum Ratings Characteristic Supply Voltage (V- to AGND) Analog Input Voltage V+ AGND/DGND Digital Input Power Dissipation. Flat Package Operating Temperature Storage Temperature Rating -4V V- -0.6 to V+ +0.6 V+ ≥ (AGND/DGND+0.5V) AGND/DGND ≥ (V- -0.5V) V- -0.6 to DGND +0.6 500mW 0℃ to 70℃ -25℃ to 125℃ Electrical Characteristics TA=25℃, V- = -3.3V Parameter Power supply Operating supply current In DCV mode Voltage roll-over error Voltage nonlinearity Symbol Test Condition VIDD Normal operation In sleep mode ISS REV 10MΩ input resistor Best case straight NLV line Input Leakage Zero input reading Band-gap reference voltage VREF Open circuit voltage for 660Ω measurement Open circuit voltage for other Ω measurement Peak to peak backplane drive voltage -3.5V≤ V ≤-2.2V Internal pull-high to 0V current AC frequency response at 6.600V range 3dB frequency for LPF mode Reference voltage temperature coefficient Capacitance measurement Accuracy2 10MΩ input resistor 100KΩ resistor between VRH and AGND TCRF Between V- pin and HOLD, RANGE, KEY, FC1-FC5 BKLIT,MAXMIN Between V- pin and RS232 ±1% ±5% LPF auto ON when Hz < 500Hz(+3%) 100KΩ resister Between VRH 0℃<TA<70℃ 6.6nF – 66mF Min. — — — — Typ. -3.3 1.8 5 — Max — 2.5 10 ±0.1 Units V mA µA %F.S1 — — ±0.1 %F.S1 -10 -000 1 000 10 +000 PA counts -1.3 -1.2 -1.1 V — -3.3 — V -1.19 -1.08 -0.97 V 3.0 3.1 3.2 V — 1.2 — µA — 11 — — — 40-400 400-2000 — — 7.5 HZ kHz — 100 — ppm/℃ -2.02 — 2.02 %F.S -3 — 3 counts Note: 1. Full Scale 2. When capacitance measurement (Clamp mode) is selected, the additional error +1% should be increased. v2.5 5 09/04/14 ES51965(6600counts) Dual Display/PEAK Function Description 1. Dual Display 1.1 Introduction ES51965 configures a dual display LCD driver for multi-meter applications. The dual display includes main-display and sub-display. Each display has 4 digits to show up to 6600 counts. The simplified representation of dual display is shown below. Main-display always shows current measurement value, however sub-display usually shows sub-measurement mode or some special reading, for example, maximum reading, minimum reading, and reference value of relative mode. By the dual display feature, ES51965 can also shows two different measurement mode at the same time, for example, AC voltage with frequency, frequency mode with duty cycle. It’s very convenient for user to observe the ac voltage and frequency at the same time without changing the rotary switch or pushing any function button. Other features achieved by dual display will be introduced in later sections. The full LCD configuration is shown at section 6.1 of page 27. Sub-Display Main-Display v2.5 6 09/04/14 ES51965(6600counts) Dual Display/PEAK 1.2 Dual display V.S. measurement mode The main-display always shows reading in main measurement mode, and sub-display shows reading in sub-measurement mode or special function. The configuration is shown at the table below. Main measurement mode Main-display Sub-display default DC voltage DC voltage N/A AC voltage AC voltage frequency DC current DC current N/A AC current for multimeter AC current frequency AC current for clampmeter AC current frequency Resistance mode Resistance N/A Capacitance mode Capacitance N/A Frequency mode Frequency Duty cycle Continuity Check Resistance N/A Diode mode Voltage N/A Temperature mode Temperature N/A ADP mode A/D result N/A Note: N/A means that sub-display shows nothing if no push function is enabled. 1.3 Dual display V.S. push function With dual display, user can observe current measurement result and special function at the same time. In other single-display DMM, it can not show the current value when some special function, for example, Peak-hold, Relative, Max/Min, is enabled. In order to see the current measurement value, the enabled function must be canceled. Push function Main-display Sub-display Enable Hold Stop updating Stop updating PEAK HOLD Current value Max. Peak or Min Peak Max/Min Current value Max. or Min. of current value 1 Relative Relative value Reference Value Zero Modified value2 N/A 1 Relative value = Current A/D result – Reference value 2 Modified value = Current A/D result – zero offset value. v2.5 7 09/04/14 ES51965(6600counts) Dual Display/PEAK 2. Operating Modes 2.1. Voltage Measurement A re-configurable voltage divider automatically provides a suitable range in voltage measurement mode. The following table summarizes the full-scale ranges in each configuration. Configuration 1 VR1 VR2 VR3 2 VR4 VR5 Full Scale Range 660.0mV 6.600V 66.00V 660.0V 1000V Divider Ratio 1 1/10 1/100 1/1000 1/10000 Resister Connection VR1 (10MΩ) VR2 (1.111MΩ) VR3 (101KΩ) VR4 (10.01KΩ) VR5 (1KΩ) Note: 1. 660.0mV range only exists in manual mode. 2. If VST (pin113) is set to V- level, the auto range mode will start from 660.0V range. 660.0mV – 66.00V ranges only exist in manual mode 2.1.2 CE selection ES51965 has a CE selection feature archived by configuring the pin CESEL. In automatic voltage mode, ES51965 will show OL when the voltage is over than the overflow level. If CESEL is connected to V- or DGND, ES51965 will have a 610.0V or 1010V overflow level in voltage mode relatively. If CESEL keeps floating, the overflow level will be set to 1510V in DCV and ACV mode. The configuration of CESEL is listed below. For fully automatic voltage modes: CESEL floating VDGND v2.5 DCV 1510V 610V 1010V 8 ACV 1010V 610V 1010V 09/04/14 ES51965(6600counts) Dual Display/PEAK 2.2 Resistance Measurement A re-configurable divider automatically provides a suitable full-scale range in resistance measurement mode. The following table summarizes the full-scale ranges and the reference resistors in each configuration. When the function code is set to FC1~5=(0,1,0,1,0), ES51965 enters the resistance mode for clampmeter. It only has 3 ranges from 660.0Ω to 66.00KΩ. Configuration OR1 OR2 OR3 OR4 OR5 OR6 Full Scale Range 660.0Ω 6.600KΩ 66.00KΩ 660.0KΩ 6.600MΩ 66.00MΩ Relative Resistor OR1 VR5 VR4 || VR1 VR3 || VR1 VR2 || VR1 VR1 Equivalent value 100Ω 1KΩ 10KΩ 100KΩ 1MΩ 10MΩ 2.3 Current measurement for multimeter ES51965 has 2 automatic and 1 manual current measurement modes for multimeter. The following table summarizes the full-scale range of each mode. When ES51965 operates in the current measurement modes for multimeter, it takes high input from pin IVSH or IVSL, low input from pin SGND with full scale 66mV and reference voltage from pin VR (-400mV). 1 VBAR Mode FC1~4 Full Scale Input Terminal Automatic1 1,1,0,1 1 660.0µA / 6600µA IVSL/IVSH V.S. SGND (66mV) Automatic2 1,1,1,1 1 66.00mA / 660.0mA IVSL/IVSH V.S. SGND (66mV) Manual 0,0,0,0 X 66.00A IVSH V.S. SGND (66mV) Note: 1. Connect VBAR to V- will disable the “µ2” / ”m2” symbol on LCD panel. v2.5 9 09/04/14 ES51965(6600counts) Dual Display/PEAK 2.4 Current measurement for clampmeter ES51965 has 2 automatic and 4 manual current measurement modes for Clampmeter. The following table summarizes the full-scale range of each mode. When ES51965 operate in the automatic mode1&2, it takes high input from IVSH/IVSL (higher range/lower range), low input from SGND with full scale 66mV and reference voltage from VR. When ES51965 operates in the manual mode1~4, it takes high input from ADP, low input from SGND with full scale 660mV and reference voltage from VA+ and VA-. 1 VBAR Full Scale Input Terminal Mode FC1~4 Automatic1 1,1,0,1 0 660.0A / 6600A IVSL/IVSH V.S. SGND (66mV) Automatic2 1,1,1,1 0 66.00A / 660.0A IVSL/IVSH V.S. SGND (66mV) Manual1 1,1,0,0 X 6.600A ADP V.S. SGND (660mV) Manual2 1,0,0,0 X 66.00A ADP V.S. SGND (660mV) Manual3 1,0,1,0 X 660.0A ADP V.S. SGND (660mV) Manual4 1,0,0,1 X 6600A ADP V.S. SGND (660mV) Note: 1. Connect VBAR to V- will disable the “µ2” / ”m2” symbol on LCD panel.. 2. In DC current modes for clampmeter, ES51965 provides Zero function for offset removing. 2.5 Low pass filter (LPF) mode for ACA/ACV mode A 3rd order low-pass filter with is built in ES51965. The LPF mode is active automatically when the input AC frequency of voltage or current is less than 500Hz. That means if the frequency is lower, the LPF mode (3dB BW = 7.5kHz typ.) will be helpful to filter-out the high frequency noise of the main signal. When LPFOFF (pin 114) is pulled to V-, the LPF mode will be disabled always. When PEAK mode is active, the auto LPF mode will be disabled temporarily until the PEAK mode is inactive. v2.5 10 09/04/14 ES51965(6600counts) Dual Display/PEAK 2.6 Capacitance Measurement The following table summarizes the eight ranges of capacitance measurement mode. Configuration C1 C2 C3 C4 C5 C6 C74 C84 Full Scale Range 6.600nF 66.00nF 660.0nF 6.600uF 66.00uF 660.0uF 6.600mF 66.00mF Relative Resistor Measurement Period 0.35 sec 0.35 sec 50k VR3 0.35 sec 101kΩ 0.35 sec 10.01kΩ 0.7 sec R9K / R1K 1.4 sec R9K / R1K 1.4 sec R9K / R1K 7.0 sec R9K / R1K Note: 1. In order to obtain an accurate reading, a capacitor must be discharged before measurement begins. The chip has a built-in discharge mode to automatically discharge the capacitor. In discharge mode, the main-display shows dIS.C 2. Discharging through the chip is quite slow. We recommend users to discharge the capacitor with some other apparatus. 3. The C2 range is calibrated by the VR near to the OVX pin. 4. The C1 range zero offset could be compensated by the small capacitors near to OVH pin. 5. When capacitance measurement (Clamp mode) is selected, C7-C8 range measurement period will be double. 2.7 Continuity Check Continuity check shares the same configuration with 660.0Ω manual resistance measurement mode and has buzzer output to indicate continuity. The buzzer generates 2KHz beep whenever the reading is less than 30Ω. 2.8 Diode Measurement Diode measurement mode shares the same configuration with 6.600V manual voltage measurement mode and has buzzer output to indicate continuity. The buzzer generates a 2KHz sound whenever the reading is less than 30mV. If the test circuit is open or the voltage drop between the two ports of the device (diode) under test is larger than 2V, the LCD panel will show “OL”. v2.5 11 09/04/14 ES51965(6600counts) Dual Display/PEAK 2.9 Frequency counter In frequency mode, main-display shows frequency and sub-display shows duty cycle at the same time. The time base of the frequency counter is derived from an external crystal oscillator by Tcounter = 4,000,000 Fosc where Fosc is the frequency of the crystal oscillator. Thus, the counter has a 1-second time base when a 4MHz oscillator is used. The frequency counter can select the proper range automatically or manually. Auto-range operation extends over seven decades, from 66.00Hz to 66.00MHz. The following table summarizes the full-scale range of the frequency counter. Range FR1* FR2 FR3 FR4 FR5 FR6 FR7 Full Scale 66.00Hz 660.0Hz 6.600KHz 66.00KHz 660.0KHz 6.600MHz 66.00MHz *If input frequency is less than 6.0Hz, ES51965 will show 0.00Hz 2.10 Temperature measurement mode Temperature measurement mode takes input signal from ADP pin and reference voltage from (VA+ - VA-). ES51965 has a built-in ºC-to-ºF scale translation circuit, and only needs an external ºC scale application circuit. The application circuit for ºF scale is not required. In temperature measurement mode, the default range is 6600 ºC or 9999 ºF. The VBAR pin is used to control the first decimal point (DP1) on the LCD panel. When VBAR is pulled to V-, DP1 will be turned on and the full scale becomes 660.0 ºC or 999.9 ºF. 2.11 Duty Cycle measurement The duty cycle measurement is available in frequency mode simultaneously. The duty cycle mode range is within 1.00 to 98.90% (< 10kHz@3Vpp). The minimum resolution is 0.01%. If the source frequency duty cycle is smaller than 0.9%, the 0.00% will be shown on the LCD display. If the duty cycle is larger than 98.9%, the OL will be shown on the LCD sub-display. If the SHIFT pin (pin 111) is pull to V-, a display shifted feature is available for duty-cycle mode. The minimum resolution is 0.1% when SHIFT is active. When the frequency input is zero, the duty cycle display will be 0.00% or OL shown. v2.5 12 09/04/14 ES51965(6600counts) Dual Display/PEAK 2.12 Auto power off and idle time selection ES51965 has a default auto power off function. If the meter is idle for more than the given Idle Time, the chip automatically turns the power off. The idle time to trigger the auto power off function is determined by pin ‘APOSEL’. If pin APOSEL is connected to V-, the Idle Time will be set to 30 minutes. If pin APOSEL is connected to DGND or floating, the Idle Time will be set to 15 minutes. When APO happens, the state of the meter is saved. The APO sign on the LCD panel indicates whether the auto power off is enabled or not. In some cases, user might want to disable Auto power off. There are two ways to disable this feature as following: 1. Power on the meter when any of the push functions, except for HOLD, is pressed down. 2. In addition, when RS232 output is active, the auto power off function is also disabled. Note: Powering on the meter while pressing HOLD and lasts 2 seconds turns on all LCD segments until HOLD is pressed again. 2.13 Sleep The meter enters sleep mode after auto power off. The SLEEP pin asserts low (V-) in the sleep mode, and asserts high (V+, not 0V) after re-power on. 2.14 Re-power on After auto power-off, pushing any of the push function or changing the rotary mode can turn on the meter again. If the meter is re-powered on by changing the rotary mode, the saved state is cleared. If the meter is re-powered on by push functions, the chip restores the saved state and enters HOLD mode. The LCD displays the saved value. v2.5 13 09/04/14 ES51965(6600counts) Dual Display/PEAK 3. Measurement Mode Switching Measurement mode depends on the logic level of SLACDC, FC1, FC2, FC3, FC4, FC5, and KEY. When FC5 is high, the measurement mode list is below. SLACDC 0 0 0 0 X 0 0 0 0 0 0 0 0 0 0 1 1 1 1 1 1 1 1 1 1 1 1 1 1 X FC1 1 1 1 0 1 1 1 1 1 0 0 0 0 0 0 1 1 1 0 1 1 1 1 0 0 0 0 0 0 0 FC2 0 1 1 0 1 1 0 0 0 0 0 1 0 1 1 0 1 1 0 1 0 0 0 0 0 1 0 1 1 1 FC3 1 0 1 0 1 0 0 1 0 1 0 1 1 1 0 1 0 1 0 0 0 1 0 1 0 1 1 1 0 0 FC4 1 1 1 0 0 0 0 0 1 1 1 1 0 0 0 1 1 1 0 0 0 0 1 1 1 1 0 0 0 1 Mode Voltage Measurement 12 Auto DC Current Measurement(µA) 12 Auto DC Current Measurement(mA) 1 66A DC Current Measurement(A) — 2 Manual DC 6.600A 2 Manual DC 66.00A 2 Manual DC 660.0A 2 Manual DC 6600A Resistance Measurement Continuity Check Resistance Measurement Frequency Measurement Capacitance Measurement Temperature Measurement Voltage Measurement 12 Auto AC Current Measurement(µA) 12 Auto AC Current Measurement(mA) 1 66A AC Current Measurement(A) 2 Manual AC 6.600A 2 Manual AC 66.00A 2 Manual AC 660.0A 2 Manual AC 6600A *ADP0 (6600) *ADP1 (660.0) *ADP2 (66.00) *ADP3 (6.600) Capacitance Measurement (Clamp)3 Temperature Measurement Resistance Measurement Function of KEY DCV ↔ ACV DCA ↔ ACA DCA ↔ ACA DCA ↔ ACA — DCA ↔ ACA DCA ↔ ACA DCA ↔ ACA DCA ↔ ACA Ω ↔ Continuity Continuity ↔ Diode Ω ↔ Continuity ↔ Diode — — ℃ ↔ ℉ ACV ↔ DCV ACµA ↔ DCµA ACmA ↔ DCmA ACA ↔ DCA ACA ↔ DCA ACA ↔ DCA ACA ↔ DCA ACA ↔ DCA — — — — — ℉ ↔ ℃ Ω ↔ Diode X means “don’t care”. *When FC5 is high, the ADP0, ADP1, ADP2 and ADP3 modes can display minus sign. 1 These modes could be designed for multimeter current modes, please refer to section 2.3. 2 These modes could be designed for clampmeter current modes, please refer to section 2.4. 3 The external protection resistance of OVH path is allowed to be 2kΩ PTC + max 2kΩ fixed resistor. The OVX capacitor 220pF Æ 1nF should be used. v2.5 14 09/04/14 ES51965(6600counts) Dual Display/PEAK When FC5 is low, the KEY function is disable. The measurement mode list is below. SLACDC 0 0 0 0 X 0 0 0 0 0 0 0 0 0 0 1 1 1 1 1 1 1 1 1 1 1 1 1 1 X FC1 1 1 1 0 1 1 1 1 1 0 0 0 0 0 0 1 1 1 0 1 1 1 1 0 0 0 0 0 0 0 FC2 0 1 1 0 1 1 0 0 0 0 0 1 0 1 1 0 1 1 0 1 0 0 0 0 0 1 0 1 1 1 FC3 1 0 1 0 1 0 0 1 0 1 0 1 1 1 0 1 0 1 0 0 0 1 0 1 0 1 1 1 0 0 FC4 1 1 1 0 0 0 0 0 1 1 1 1 0 0 0 1 1 1 0 0 0 0 1 1 1 1 0 0 0 1 Mode DC Voltage Measurement 12 Auto DC Current Measurement(µA) 12 Auto DC Current Measurement(mA) 1 66A DC Current Measurement(A) - 2 Manual DC 6.600A 2 Manual DC 66.00A 2 Manual DC 660.0A 2 Manual DC 6600A Resistance Measurement Continuity Check Diode Measurement Frequency Measurement Capacitance Measurement Temperature Measurement (℉) AC Voltage Measurement 12 Auto AC Current Measurement(µA) 12 Auto AC Current Measurement(mA) 1 66A AC Current Measurement(A) 2 Manual AC 6.600A 2 Manual AC 66.00A 2 Manual AC 660.0A 2 Manual AC 6600A *ADP0 (6600) *ADP1 (660.0) *ADP2 (66.00) *ADP3 (6.600) Capacitance Measurement (Clamp)3 Temperature Measurement (℃) 4 Resistance Measurement (Clamp) KEY - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - X means “don’t care”. *When FC5 is low, the ADP0, ADP1, ADP2 and ADP3 modes can NOT display minus sign. 1 These modes could be designed for multimeter current modes, please refer to section 2.3. 2 These modes could be designed for clampmeter current modes, please refer to section 2.4. 3 The external protection resistance of OVH path is allowed to be 2kΩ PTC + max 2kΩ fixed resistor. The OVX capacitor should be modified from 220pF Æ 1nF. 4 This mode has 3 ranges only from 660.0Ω(with continuity check) to 66.00KΩ. v2.5 15 09/04/14 ES51965(6600counts) Dual Display/PEAK 4. Push function All the enabled push functions will be reset when the measurement mode is changed by rotary switch. Change measurement mode by KEY function will reset enabled Range, Hold, Max/Min, REL, PEAK-hold and Zero functions. The following table lists the available function versus every measurement mode. Function Range Hold Max/Min REL PEAK Zero O O O O AC X O O O O AC X O O O AC AC DC Resistance O O O O X X Frequency O O X X X X Capacitance O O O O X X Continuity X O O O X X Diode X O O O X X Temperature X O O O X X ADP X O O O O3 X Mode Voltage 1 Current Mode for Multimeter 2 Current Mode for Clampmeter Note: 1 Include automatic μA, automatic mA and manual 66A modes, please refer to section 2.3. 2 Include 2 automatic modes and 4 manual modes, please refer to section 2.4. 3 When FC5=0, Peak-hold function is not available for ADP mode. 4.1. HOLD and Delay Hold feature HOLD mode makes the meter stop updating the LCD panel. This mode can be nested in most of the special modes. Enabling HOLD function in automatic mode makes the meter switch to manual mode, but the full-scale range remains the same. ES51965 provides a Delay HOLD feature. To activate Delay HOLD feature, press down the HOLD bottom and last for 2 seconds. The meter will delay for 6 seconds, than enters HOLD mode. In the 6-secnod delayed time, the HOLD symbol on LCD panel will blink. HOLD function and the delayed state can be cancelled by changing the measurement mode, pressing RANGE, or push HOLD again. v2.5 16 09/04/14 ES51965(6600counts) Dual Display/PEAK Normal mode Push < 2s HOLD mode (Auto/Manual) Push < 2s (Manual) Push > 2s Push < 2s Push > 2s 6sec past Delay State (Auto/Manual) 4.2. Range RANGE pin switches to and from automatic and manual mode, and while in manual mode, changes the full-scale range. The following figure shows the state transition. 1 push < 1 sec Automatic Mode Manual Mode 1 push > 1 sec Range up 1 push < 1 sec Measurement Mode Auto Manual VRi → VRi + 1, V(DC/AC) VR2 – VR5 VR5 → VR1 R1 → R2, Auto µA(DC/AC) R1 – R2 R2 → R1 R1 → R2 Auto mA(AC/DC) R1-R2 R2 → R1 66 A(DC/AC) fixed fixed R1 → R2 Auto 66A/660A R1-R2 (DC/AC) R2 → R1 R1 → R2 Auto 660A/6600A R1-R2 (DC/AC) R2 → R1 Ci → Ci + 1, Capacitance C1 – C8 C8 → C1 ORi → ORi + 1, Ω OR1 – OR6 OR6 → OR1 ORi → ORi + 1, Ω (Clamp) OR1 – OR3 OR3 → OR1 Continuity fixed fixed Diode fixed fixed FRi → FRi + 1 Frequency FR1 – FR7 FR7 → FR1 Control Range Initial Range 660.0mV – 1000V 6.600V 660.0µA – 6600µA 660.0µA 66.00mA – 660.0mA 66.00mA 66.00A 66.00A 66.00A – 660.0A 66.00A 660.0A – 6600A 660.0A 6.600nF– 66.00mF 6.600nF 660.0Ω – 66.00MΩ 660.0Ω 660.0Ω – 66.00KΩ 660.0Ω 660.0Ω 6.600V 660.0Ω 6.600V 66.00Hz – 66.00MHz 66.00Hz Note: Pushing RANGE resets all existing special modes. 4.3. KEY See Section “Measurement Mode Switching” for the function of this pin. v2.5 17 09/04/14 ES51965(6600counts) Dual Display/PEAK 4.4. REL + HOLD In REL mode, the main-display shows DN+K - DN(relative value), where N = 1, 2, 3, … ,and the DN value is shown on sub-display. DN(reference value) is the last value before REL is pushed, and DN+K is the current value. The meter returns to normal operation if REL is pressed again. Pressing HOLD in REL mode makes the meter stop updating the LCD panel. push REL Original State (Auto/Manual) push HOLD 〝REL〞active. 〝REL〞and〝H〞active. Hold the relative value The relative value push REL push HOLD Note: It’s possible that relative value exceeds 6,600 or –6,600 counts, but never exceeds 9,999 or –9,999 counts. Such relative values are displayed. The LCD shows OL in REL mode only if DN or DN+K is more than 6,600 counts or the relative value is more than 9,999. 4.5. Max/Min + HOLD The meter displays maximum or minimum value of input in Max/Min mode. When Max/Min is pressed for the first time, sub-display shows maximum value. Sub-display shows minimum value, when it is pressed again. Main-display always shows current value in Max/Min mode. The meter returns to normal operation if Max/Min is pressed and held for longer than one second. Pressing HOLD in Max/Min mode makes the meter stop updating the maximum or the minimum value. 1 push<1s Original state (Automatic/manual) 1 push>1s “MAX” active max value 1push<1s 1 push>1 sec “MIN” active min value 1push<1s Push HOLD Push HOLD “MIN” “H” active “MAX” “H” active hold max value. 1push<1s hold min value. 1push<1s v2.5 18 09/04/14 ES51965(6600counts) Dual Display/PEAK 4.6. Relative + Max/Min + HOLD Max/Min mode can be nested in REL mode. Sub-display shows maximum or minimum value which is relative to the reference and shows the reference value when Max/Min is pressed in REL mode. Main-display always shows current relative value. Pressing HOLD under REL + Max/Min makes the meter stop updating the LCD panel. Original State (Auto/Manual) Push REL MD: Main-Display SD: Sub-Display Push REL 〝REL〞active. MD: relative value, SD: reference value Push Max/Min Push Max/Min > 1s 〝REL〞and〝Max〞active. 〝REL〞and〝Min〞active. 〝REL〞〝Max〞〝Min〞 SD: Maximum rel. value SD: Minimum rel. value blink. SD: Reference value Push Max/Min Push Max/Min Push HOLD 〝REL〞,〝Max〞, 〝H〞active. Hold max. rel. value Push HOLD 〝REL〞, 〝Min〞, Push Max/Min 〝REL〞 〝Max〞 , 〝Min〞 , , 〝H〞blink. Hold the 〝H〞active. Hold min. rel. value Push HOLD Push Max/Min reference value Push Max/Min 4.7. PEAK-HOLD function ES51965 provides Peakhold function to caputure the maximum or minimum peak value. To ecnter peak mode, press PEAK key less than 1 second. A self-calibration process will execute automatically before normal peak-hold operation. The precision of Peak mode measurement can be enhanced by calibration. In peak mode, main-display shows current value of signal, and sub-display shows PMAX or PMIN value which is selected by PEAK input. Peak mode takes from input PMIN or PMIN pin. The peak values are stored in external capacitors. Polyester capacitors are recommended. To exit PEAK-hold function, press PEAK key larger than 1 seconds. v2.5 19 09/04/14 ES51965(6600counts) Dual Display/PEAK 1 push > 1 s 1 push > 1 s Original State (Auto/Manual) 1 push < 1 s CALIBRATION 〝Pmax〞active. SD: Max. Peak Value 1 push < 1 s 〝Pmin〞active. SD: Min. Peak Value push HOLD push HOLD 〝Pmax〞 〝HOLD〞 , active. 〝Pmin〞 〝HOLD〞 , active. 1 push < 1 s SD: Max. Peak Value SD: Min. Peak Value 4.8. ZERO function In manual DC 6.600A, 66.00A, 660.0A and 6600A, auto DC 66.00A/660.0A and auto DC 660.0A/6600A mesurement modes, ES51965 provides Zero function to remove the residual current value. Push ZERO bottom less than one second to enter Zero mode. In Zero mode, main-display shows DN+K - DN, where N = 1, 2, 3, … , DN is the last conversion value before Zero is pushed, and DN+K is the current conversion value. Sub-display shows nothing if no other function is activated. If Zero is pushed again in Zero mode, main-display will update the DN value and displays the DN+K - DN again. The meter returns to normal operation if Zero is pressed and held for longer than one second. Pressing HOLD in Zero mode makes the meter stop updating the LCD panel. In 2-range auto DCA modes for clampmeter (660.0A/6600A or 66.00A/660.0A), the system will stay in automatic mode, even if the Zero function is activated. In other words, it could achieve real automatic operation. In automatic mode, Zero function could not be entered from higher range, but it could be still activated if current range is lower one. This is because most residual current value is so small that the range should stay at lower one in automatic mode. When enter Zero mode from lower range, the system will store the nonzero counts (residual current value). If the range goes up to higher one automatically, the non-zero counts will be divided by ten. So this function will still work well in automatic modes. v2.5 20 09/04/14 ES51965(6600counts) Dual Display/PEAK State diagram for Zero mode: Original State Press Zero Press Zero > 1s ‘Zero’ Active Press Zero < 1s Main-display: DN+K - DN State diagram for Zero + HOLD mode: Original State Press Zero Press Zero > 1s ‘Zero’ Active Press Zero < 1s Main-display: DN+K - DN Press Hold ‘Zero’ ‘Hold’ Active LCD stops updating State diagram for Zero + Max/Min + HOLD mode: Original State Press Zero Press Max/Min Press Zero > 1s ‘Zero’ Active MD: DN+K - DN SD: No data Press Max/Min > 1s ‘Zero’ ‘Max’ Active MD: DN+K - DN SD: Max. DN+K - DN Value Press HOLD Press Zero < 1s ‘Zero’ ‘Min’ Active MD: DN+K - DN SD: Min. DN+K - DN Value Press Max/Min < 1s ‘Zero’ ‘Max’ ‘Hold’ Active SD: Max. DN+K - DN Value Press HOLD ‘Zero’ ‘Min’ ‘Hold’ Active SD: Min. DN+K - DN Value Press Max/Min < 1s v2.5 21 09/04/14 ES51965(6600counts) Dual Display/PEAK 5. Serial Data Output The RS232 function will be activated if the RS232 pin is pulled to and asserts at V-. The serial data sent to SDO pin once every A/D conversion cycle. The data format complies with JIS 7Bits transmission code with a baud rate of 19230. The host can use RS232 interface to read the data. A single data packet includes a start bit (always 0), 7 data bits, an odd parity check bit, and a stop bit (always 1). The high and low voltage levels correspond to DGND and V- respectively. SDO remains at 1 (high) when it is inactive. Hence the start bit (0) could be used as the triggering signal to begin the reading process. The following figure shows the data format of a single packet. The LSB is sent first and the MSB is sent last. single package 0V 0 -3V D0 ~D6 LSB p 1 P MSB One data block consists of 17 packets, or 170 bits. The following figure shows the format of a data block. The M_range or S_range packet indicates the full-scale range of main measurement mode or sub-measurement mode. M_digit3 through M_digit0 are the conversion result of main measurement mode, and S_digit3 through S_digit0 are the conversion result of sub-measurement mode(Hz or duty). The function packet indicates the measurement mode of the meter. Status, option1, option2 and option3 give the status of the meter. CR and LF are delimiters used to separate the blocks. all package 0 10 M_range 0 10 S_range 10 M_digit1 0 10 10 0 Option1 1 S_digit2 10 Function 10 Option2 M_digit2 S_digit3 S_digit0 1 10 10 10 0 M_digit3 M_digit0 S_digit1 10 1 Status 10 Option3 1 CR 1 LF v2.5 22 09/04/14 ES51965(6600counts) Dual Display/PEAK 5.1. FUNCTUON This packet indicates the measurement mode of the meter. The following table summarizes the transmitted code for each mode. Note that the encoding of this packet is different from the encoding of FC1-FC5 switch. Code 0111011 2 0111101 2 0111111 0110000 0111001 0110011 0110101 0110001 0110010 0110110 0110100 0111110 Measurement Mode Voltage Auto µA Current Auto mA Current 66 A current Manual A Current Ω Continuity Diode Frequency Capacitance 1 Temperature ADP VBAR=0 VBAR=1 Don’t care Auto µA Current Auto mA Current Auto 660.0A/6600A Auto 66.00A/660.0A Don’t care 1 When the function code = 0110100, the TF bit in the Status packet determines whether the unit is Celsius or Fahrenheit. 2 When the function code = 0111101 or 0111111, the measurement mode is determined by VBAR bit of STATUS packet. 5.2. RANGE This packet indicates the full-scale range of the meter. When the meter operates in continuity mode or diode mode, this packet is always 0110000 since the full-scale ranges in these modes are fixed. The following table lists the code for each range in each measurement mode. Ω Code V *2-range auto A 66 A Manual A ADP 0110000 6.600V Lower Range(IVSL) 66.00 A 6.600A ADP3 660.0Ω 0110001 66.00V Higher Range(IVSH) 66.00A ADP2 6.600KΩ 0110010 660.0V 660.0A ADP1 66.00KΩ 0110011 1000V 6600A ADP0 660.0KΩ 0110100 660.0mV 6.600MΩ 0110101 66.00MΩ 0110110 0110111 *It includes auto 660.0/6000µA, 66.00/660.0mA, 66.00A/660.0A, 660.0A/6000A. v2.5 23 Frequency Capacitor 66.00Hz 6.600nF 660.0Hz 66.00nF 6.600KHz 660.0nF 66.00KHz 6.600µF 660.0KHz 66.00µF 6.600MHz 660.0µF 66.00MHz 6.600mF 66.00mF 09/04/14 ES51965(6600counts) Dual Display/PEAK 5.3. Mdigit_3 – Mdigit_0 and Sdigit_3 – Sdigit_0 Mdigit_3 – Mdigit_0 is the measurement result of main measurement mode. This result is shown on main-display of LCD panel. Sdigit_3 – Sdigit_0 is the measurement result of sub-measurement mode, for example, frequency in AC voltage/current mode or duty cycle in frequency mode. This result will be shown on sub-display of LCD panel. Mdigit3 and Sdigit3 is the most significant digit, and Mdigit0 and Sdigit0 is the least significant digit. Digit 0 1 2 3 4 5 6 7 8 9 Code 0110000 0110001 0110010 0110011 0110100 0110101 0110110 0110111 0111000 0111001 5.4. STATUS The format of this package is shown below. The TF field is meaningful only when the Function packet indicates Temperature mode. In Temperature mode, the TF is 1 if the unit is °C and is 0 of the unit is °F. BATT field is one when battery low condition is true. The Judge bit is available in duty cycle mode. If the Judge bit is 1, the final result should be calculated by (100-X) % which X is the output data from Sdigit_3~Sdigit0. If the Judge bit is 0, then the final data is X %. The VBAR will be 1 only when the VBAR pin is connected to V-. 0 BIT6 v2.5 1 BIT 5 1 BIT 4 TF BIT 3 BATT BIT 2 24 Judge BIT1 VBAR BIT 0 09/04/14 ES51965(6600counts) Dual Display/PEAK 5.5. OPTION 1 DC field indicates that the meter operates in DC measurement mode, either voltage or current. AC field indicates that the meter operates in AC measurement mode, either voltage or current. AUTO field is set to 1 if the meter operates in automatic mode, and is set to 0 when the meter operates in manual mode. SIGN field indicates whether the minus sign of main measurement on the LCD panel is on or off. 0 BIT 6 1 BIT 5 1 BIT 4 DC BIT 3 AC BIT 2 AUTO BIT 1 SIGN BIT 0 1 BIT 5 1 BIT 4 OLM BIT 3 OLS BIT 2 BIT1 BIT 0 5.6. OPTION 2 0 BIT 6 OLM bit is 1 when the main measurement mode is over range. In PEAK mode, the OLM bit is active only when ADC counts larger than 9999. OLS bit is 1 when the sub-measurement mode (Hz or duty) is over range in voltage/current with frequency or frequency with duty cycle modes. 5.7. OPTION 3 In peak mode, the meter outputs and repeats current result, maximum peak value and minimum peak value sequentially no mater what the state of peak shown on LCD panel is PMAX or PMIN. PEAKC, PMAX or PMIN bit indicates the output data is current result of signal, or the stored voltage at the maximum peak or minimum peak capacitor. VAHz bit is set to high when the sub-display is available for HZ mode of ACV/ACA. If MAX/MIN or REL key is pushed actively, the Hz mode will be inactive and the VAHz bit will be set to low. 0 BIT 6 1 BIT 5 1 BIT 4 PEAKC BIT 3 PMAX BIT 2 PMIN BIT 1 VAHz BIT 0 5.8. CR Carriage return. The transmitted code is 0001101. 5.9. LF Line feed. The transmitted code is 0001010. v2.5 25 09/04/14 ES51965(6600counts) Dual Display/PEAK 6. Miscellaneous The conditions, which the meter turns on the buzzer, include: (1) Changing measurement mode generates one beep. (2) Pressing any of the push functions generates one beep, if the function is valid. (3) Power on and re-power on generate one beep. (4) Input overflow in voltage and current mode generates one beep every 0.3 seconds (or 3.33 beeps per second.) (5) Continuity(diode) check generates a continuous 2KHz beep whenever the measurement is less then 30Ω(30mV) (6) Auto power off generates a 2KHz beep which lasts for 1.5 seconds. The following figures show the output waveform from the BUZOUT pin. 0.5 mS (a) Continuous 2KHz beep 0.15 sec 0.3 sec (b) 3.33 beep/sec v2.5 26 09/04/14 ES51965(6600counts) Dual Display/PEAK 6.1. LCD Panel S01 S02 BP1 ADP0 ADP1 S03 S04 S05 S06 S07 S08 S09 S10 S11 S12 APO AC MINUS 4F 4A 4B 3F 3A 3B 2F DC 4E 4G 4C 3E 3G 3C 2E BP2 ADP2 ADP3 RS232 LBAT bar7 bar8 bar11 bar12 bar13 bar14 bar17 bar18 bar15 bar16 bar19 bar20 4D DP3 bar3 bar4 bar5 bar6 bar9 bar10 bar21 bar22 S13 S14 S15 S16 S17 S18 BP1 2A 2B 1F 1A 1B µ2 BP2 2G 2C 1E 1G 1C BP3 2D BP3 bar- bar0 BP4 bar1 bar2 bar23 bar24 bar25 bar26 bar27 bar28 bar29 bar30 bar31 bar32 bar33 bar34 bar35 bar36 S19 S20 S21 S22 S23 S24 n m2 µ1 V A m1 F ℃ ℉ 3D DP2 % M K Ω Hz bar45 bar46 bar47 bar48 bar49 bar50 bar51 bar52 bar53 bar54 bar55 bar56 bar57 bar58 bar59 bar60 bar63 bar64 S27 S28 S29 S30 S31 S32 S33 S34 S35 S1A S2B S2A S3B S3A S4B S4A AUTO PMAX S1G S1F S2G S2F S3G S3F S4G S4F MAX MANU PMIN BP3 S1C S1E S2C S2E S3C S3E S4C SMINUS MIN ZERO BP4 S1D SDP1 S2D SDP2 S3D SDP3 S4D S4E HOLD REL DP2 1D BP4 bar38 bar39 bar40 bar41 bar42 bar43 bar44 S25 S26 BP1 S1B BP2 bar37 S31 S32 S35 S34 S33 S29 S30 S27 S28 Hz2 % K2 Hz2 BP1 ADP0 APO AC ADP1 DC BP2 ADP2 LBAT ADP3 RS232 ℃℉ µ2 m2 V A n µ1 m1 F M K Ω Hz 66 BP3 0 S01 S02 v2.5 K2 S25 S26 S24 MAX MIN HOLD AUTO MANU ZERO REL PMAX PMIN BP4 bar61 bar62 bar65 bar66 10 S03 S04 20 30 S05 S07 S09 S06 S08 S10 27 40 S11 S13 S12 50 S14 S16 S17 S19 S21 S15 S18 S20 60 S22 S23 09/04/14 ES51965(6600counts) Dual Display/PEAK LCD Backplane Waveform V- +3.1 V- +3.1 V- V- V- +3.1 V- +3.1 V- V- V- +3.1 V- +3.1 VV- +3.1 VV- +3.1 V- 6.2. LCD display on condition LCD Annunciator V A Ω F % Hz ADP1 ADP2 ADP3 ADP4 DC AC AUTO MANU HOLD REL Max / Min PMAX/PMIN ZERO m1 µ1 n m2 µ2 M v2.5 Condition In voltage measurement mode, and diode measurement mode. In current measurement mode. In resistance measurement mode, and continuity mode. In capacitance measurement mode. In continuity check mode. In diode mode. In duty cycle measurement. In frequency mode. When ADP1 mode is active When ADP2 mode is active When ADP3 mode is active When ADP4 mode is active In DC voltage or DC current mode. In AC voltage or AC current mode. When automatic full scale range selection is enabled. In manual mode. When HOLD function is enabled. When Relative function is enabled. When Max/Min function is enabled. When PEAK-hold function is enabled. When ZERO function is enabled. In capacitor measurement mode and the full scale range is in the order of mF. In capacitor measurement mode and the full scale range is in the order of uF. In capacitor measurement mode and the full scale range is in the order of nF. In voltage or current measurement mode and the full scale range is in the order of 10-3. In current measurement mode and the full scale range id in the order of uA. In resistance measurement mode and the full scale range is in the order of MΩ 28 09/04/14 ES51965(6600counts) Dual Display/PEAK K ℃ ℉ APO RS232 v2.5 In resistance measurement mode and the full scale range is in the order of KΩ In temperature measurement mode and when the unit is ℃ In temperature measurement mode and when the unit is ℉ When auto power off function is enabled. When RS232 output is enabled. 29 09/04/14 ES51965(6600counts) Dual Display/PEAK 6.3. Operating Timing ES51965 incorporates a dual slope ADC with four phases: ZI, AZ, INT and DINT. The timing of each phase is listed below. The Bar-graph ADC is 10X faster than slow ADC. *Voltage / Ohm /ADP / Manual Current (for clampmeter) measurement: Phase ZI AZ INT DINT Lo-speed 60ms 24ms 100ms 166ms Hi-speed 6ms 2.4ms 10ms 16.6ms PEAK 256ms 120ms 100ms 246ms *Current mode for multimeter / Current mode for clampmeter Phase ZI AZ INT DINT Lo-speed 60ms 24ms 1000ms 166ms Hi-speed 6ms 2.4ms 100ms 16.6ms *Capacitance measurement: Range 6.600nF 66.00nF 660.0nF 6.600μF 66.00μF 660.0μF 6.600mF 66.00mF Total Measurement Time 0.35sec 0.35sec 0.35sec 0.35sec 0.70sec 1.40sec 1.40sec(max) 7.0sec(max) *Frequency + Duty cycle mode measurement In range 66.00Hz~66MHz, the measurement cycle = 1.05 sec P.S. In the frequency measurement with auto mode, if the range is changed, the internal clock rate will increase ten times and the new measurement cycle becomes 1/10 times of the original cycle until the range is stable. v2.5 30 09/04/14 SW SPDT 1 1 R1 90 R2 9 R3 0.9 R4 R5 SW2 SW3 SW10 SW11 SW12 uA uA mA mA A R6 100 C12 2 C13 C5 C15 C14 220nF 22nF 22nF SW4 R21 R7 R14 R15 R16 R20 10nF V- C21 100nF 470K 150K 100K 100K Detailed V+ 56K VR VA+ VA- 1 R13 OVH OVSG 100 1K 10.01K 101K 1.111M TEST5 ACVL ACVH ADI ADO C22 7.5V V + ZR6 0.1uF 10uF Power Switch 5.6V 0.1uF C17 220nF 47nF 470K C18 220nF R9 R10 R11 R12 C6 C2 10uF C1 ZR5 Regulator DC3.3V R8 0-100pF 0-100pF IVSH IVSL Q1 Q2 + 0.001 0.099 Metallized Polypropylene Film Capacitor : C17 Metallized Polyester Capacitor : C18 Polyester Capacitor : C14 , C16 , C32 , C33 Current measurement input terminals should be protected by voltage clamping devices. uA SW1 mA J1 1 C11 220pF + 470nF + D J6 20A Gnd J5 C23 C19 C9 VR2 10K J3 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 3 V+ DGND AGND AGND CH+ CHCIH BUFH CAZH CL+ CLCIL CAZL BUFL RAZ IVSH IVSL OVX OVH OVSG OR1 VR5 VR4 VR3 VR2 TEST5 ACVL ACVH ADI ADO OHMC3 OHMC2 OHMC1 VRH VR VA+ VASGND 100K 10M 1uF V+ V- C8 470nF 4 Y1 4MHz BKOUT BUZOUT RS232 BKLIT PEAK M AX/MIN U1 102 KEY 101 REL_ZERO 100 RANGE 99 HOLD 98 97 ANNUNC 96 BP1 95 BP2 94 BP3 93 BP4 92 SEG01 91 SEG02 90 SEG03 89 SEG04 88 SEG05 87 SEG06 86 SEG07 SEG08 85 84 SEG09 83 SEG10 82 SEG11 81 SEG12 80 SEG13 79 SEG14 SEG15 78 SEG16 77 SEG17 76 75 SEG18 74 SEG19 73 SEG20 SEG21 72 SEG22 71 SEG23 70 69 SEG24 SEG25 68 67 SEG26 66 SEG27 65 SEG28 + Q4 R28 2.2K PTC 1 C7 0.1uF V- J4 Q7 ES51965_QFP_128 2.2uF KEY REL RANGE HOLD LCDCAP ANNUC BP1 BP2 BP3 BP4 SEG01 SEG02 SEG03 SEG04 SEG05 SEG06 SEG07 SEG08 SEG09 SEG10 SEG11 SEG12 SEG13 SEG14 SEG15 SEG16 SEG17 SEG18 SEG19 SEG20 SEG21 SEG22 SEG23 SEG24 SEG25 SEG26 SEG27 SEG28 C29 + Q3 5 SW22 SW18 SW17 SW16 SW15 SW14 SW20 SW19 SW9 SW8 SW7 SW6 SW5 RANGE HOLD FC5 SLACDC FC4 FC3 FC2 FC1 R31 1N4148 D2 V- SW-SPDT V- REL_ZERO KEY MAX/MIN PEAK BKLIT Function Switch BKLIT PEAK MAX/MIN KEY REL_ZERO RANGE HOLD SLACDC FC5 FC4 FC3 FC2 FC1 10K 10K ( For IVSH/IVSL mode ) VR3 VR4 1uF + C4 Number 14-Apr-2009 F:\Protel file\KA026.ddb 10K C25 1uF C3 ADO R30 15K 6 4.7uF 100K R32 100K R33 Sheet of Drawn By: 6 Ver:4 ACVL ACVH TEST5 ADI C31 0.1uF Revision ES51965_AVG Circuit(Peak) 1N4148 D1 ( For ACV/ADP mode ) R29 15K Title B Size Date: File: S1 Average Circuit FIN 5 + 9K 4 Q6 R27 200 SEG34 SEG33 SEG32 SEG31 SEG30 SEG29 C24 4.7uF ADP R25 C26 R26 LPF 3dBBW = 7.5KHz VBAR BUZIN FREQ SEG35 VR1 50K R19 27K V- 1K 100pF C28 470pF SLEEP VIN R23 10K R24 Detailed 22nF 1 uA / mA J2 SW21 SW_RC Q5 BZ1 Buzzer Hi-impedance Buzzer 3 C16 C20 2.2nF C 1 BUZOUT Buzzer Circuit R/C/D/Cont Close to IC R22 10K 2 SDO SLACDC FC1 FC2 FC3 FC4 FC5 CESEL LPFOFF VST APOSEL En_Shift 0.1uF 0.1uF D C B A 09/04/14 31 v2.5 B A 1 R18 1.5K PTC LBAT9 C32 C33 180K R17 128 127 126 125 124 123 122 121 120 119 118 117 116 115 114 113 112 111 110 109 108 107 106 105 104 103 V+ VVLBAT9 CC+ SDO SLADC FC1 FC2 FC3 FC4 FC5 CESEL LPFOFF VST APOSEL SHIFT OSC1 OSC2 BKOUT BUZOUT RS232 BKLIT PEAK M AX/MIN ADP VR1 CACA+ R9K R1K PMIN PMAX CSH+ CSHLPC1 LPC2 LPC3 LPFOUT SLEEP VBAR BUZIN FREQ SEG35 NC SEG34 SEG33 SEG32 SEG31 SEG30 SEG29 470nF 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 C10 C27 + ES51965(6600counts) Dual Display/PEAK 7. Application Circuit 7.1 AVG circuit SW SPDT 1 1 R1 90 R2 9 R3 0.9 R4 R5 SW2 SW3 SW10 SW11 SW12 uA uA mA mA A R6 IVSH IVSL Q1 100 C12 2 C13 ZR5 C1 C5 C2 10uF SW4 R7 R14 R15 R16 R20 10nF V- C21 100nF 470K 150K 100K 100K Detailed V+ ADO 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 3 U1 V+ DGND AGND AGND CH+ CHCIH BUFH CAZH CL+ CLCIL CAZL BUFL RAZ IVSH IVSL OVX OVH OVSG OR1 VR5 VR4 VR3 VR2 TEST5 ACVL ACVH ADI ADO OHMC3 OHMC2 OHMC1 VRH VR VA+ VASGND 100K V+ V- C8 470nF ES51965_QFP_128 4 Y1 4MHz + 2.2uF KEY REL RANGE HOLD LCDCAP ANNUC BP1 BP2 BP3 BP4 SEG01 SEG02 SEG03 SEG04 SEG05 SEG06 SEG07 SEG08 SEG09 SEG10 SEG11 SEG12 SEG13 SEG14 SEG15 SEG16 SEG17 SEG18 SEG19 SEG20 SEG21 SEG22 SEG23 SEG24 SEG25 SEG26 SEG27 SEG28 C29 C7 0.1uF V- J4 Q7 S1 5 FC1 FC2 FC3 FC4 FC5 SLACDC HOLD RANGE REL_ZERO KEY MAX/MIN PEAK BKLIT SW22 SW18 SW17 SW16 SW15 SW14 SW20 SW19 SW9 SW8 SW7 SW6 SW5 Title 200 200 RANGE HOLD FC5 SLACDC FC4 FC3 FC2 C3 FC1 - VS 1 2 3 4 5 6 7 C25 V- 10uF U2 C4 ES636 500K - VS 14 13 12 11 10 9 8 + VS 10uF R29 10K Vin + VS En NC - VS NC CAV NC dB COMMON BUF out RL BUF in Iout VR5 + VS 4.7uF V- REL_ZERO KEY MAX/MIN PEAK BKLIT Function Switch FIN Number 14-Apr-2009 F:\Protel file\KA026.ddb 6 R30 200 Sheet of Drawn By: 6 + ACVH Revision Ver:4 ES51965_RMS(ES636) Circuit(Peak) ( For ACV/ADP mode ) VR4 SW-SPDT VR3 ( For IVSH/IVSL mode ) RMS Circuit 1 C30 22uF 102 KEY 101 REL_ZERO 100 RANGE HOLD 99 98 ANNUNC 97 BP1 96 BP2 95 BP3 94 BP4 93 SEG01 92 SEG02 91 SEG03 90 SEG04 89 SEG05 88 SEG06 87 SEG07 86 SEG08 85 SEG09 84 SEG10 83 SEG11 82 SEG12 81 SEG13 80 SEG14 79 SEG15 78 SEG16 77 SEG17 76 SEG18 75 SEG19 74 SEG20 73 SEG21 72 SEG22 71 SEG23 70 69 SEG24 68 SEG25 67 SEG26 66 SEG27 SEG28 65 ADO Q4 R28 2.2K PTC B Size Date: File: + Q3 + Q6 R27 200 5 + 10M LPF 3dBBW = 7.5KHz 4 + R25 R13 OVH OVSG 100 1K 10.01K 101K 1.111M TEST5 ACVL ACVH C22 7.5V V + ZR6 0.1uF 10uF Power Switch 5.6V 0.1uF 22nF C6 22nF C15 220nF 470K C17 C14 Regulator DC3.3V R8 C18 220nF R9 R10 R11 R12 220nF 47nF SW13 ACV RMS C23 + 0.001 0.099 Metallized Polypropylene Film Capacitor : C17 Metallized Polyester Capacitor : C18 Polyester Capacitor : C14 , C16 , C32 , C33 Current measurement input terminals should be protected by voltage clamping devices. uA SW1 J1 1 C11 220pF 0-100pF 0-100pF + 470nF + D mA 20A Gnd J5 Q2 C19 R21 56K VR VA+ VA- 1 BKOUT BUZOUT RS232 BKLIT PEAK M AX/MIN SEG34 SEG33 SEG32 SEG31 SEG30 SEG29 C9 VR2 10K J3 1uF 9K VBAR BUZIN FREQ SEG35 C28 470pF SLEEP C24 4.7uF ADP C26 R26 100pF VR1 50K R19 27K V- 1K C20 2.2nF VIN R23 10K R24 Detailed 22nF J6 J2 SW21 SW_RC Q5 BZ1 Buzzer Hi-impedance Buzzer 3 C16 1 uA / mA 1 BUZOUT Buzzer Circuit R/C/D/Cont Close to IC R22 10K 2 SDO SLACDC FC1 FC2 FC3 FC4 FC5 CESEL LPFOFF VST APOSEL En_Shift 0.1uF 0.1uF C27 2.2uF C31 LBAT9 128 127 126 125 124 123 122 121 120 119 118 117 116 115 114 113 112 111 110 109 108 107 106 105 104 103 C32 C33 C B A 1 R18 1.5K PTC V+ VVLBAT9 CC+ SDO SLADC FC1 FC2 FC3 FC4 FC5 CESEL LPFOFF VST APOSEL SHIFT OSC1 OSC2 BKOUT BUZOUT RS232 BKLIT PEAK M AX/MIN ADP VR1 CACA+ R9K R1K PMIN PMAX CSH+ CSHLPC1 LPC2 LPC3 LPFOUT SLEEP VBAR BUZIN FREQ SEG35 NC SEG34 SEG33 SEG32 SEG31 SEG30 SEG29 470nF 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 C10 180K R17 D C B A 09/04/14 32 v2.5 + ES51965(6600counts) Dual Display/PEAK 7.2 RMS circuit (ES636) ES51965(6600counts) Dual Display/PEAK 8. Package Information 8.1 128L QFP Outline drawing 8.2 Dimension parameters v2.5 33 09/04/14