EXTECH 407113

User's Guide
Heavy Duty CFM Thermo-Anemometer
Model 407113
3
Introduction
Congratulations on your purchase of the Extech 407113 CFM meter. This handheld
meter can display Air Flow (CFM) with Area or Air Velocity with Temperature on a
10,000-count (0 to 9999) dual display LCD. Other functions include MIN/MAX
Record/Recall, RS-232 PC interface, and Data Hold. The metal vane sensor offers
rugged durability and higher temperature withstand than comparable devices. Careful
use of this meter will provide years of reliable service.
Specifications
General Specifications
Display
Measurements
Dual Display Multi-function 10,000-count (0 to 9999) LCD
Air Velocity: m/s, km/h, ft/min, knots, mph;
3
3
Air Flow: CMM (m /min) and CFM (ft /min);
Temperature: °C and °F
Data Hold
Freezes displayed reading
Sampling rate
One (1) reading per second
Sensors
Air velocity/flow sensor: Metal angled vane arms with lowfriction ball bearing. Temp. Sensor: Precision thermistor
MIN/MAX
Records/Recalls lowest and highest readings
Auto Power OFF
Automatic shut off after 15 minutes
PC Interface
RS-232 Serial Communications (16-bit data stream)
Over range indication
1___ appears on the LCD
Low battery indication
LBT appears on the LCD
Power supply
9V Battery (consumption 8.3mA approx.)
Operating conditions
Meter: 32 to 122 F (0 to 50 C); 80% RH max.
o
o
o
o
Dimensions / Weight
Sensor: 32 to 175 F (0 to 80 C)
Main instrument: 7.1 x 2.8 x1.3" (180 x 72 x 32mm)
Sensor head: 2.8” (72mm) diameter
Weight
0.84 lbs. (381g) for meter and sensor
2
Model 407113 V3.2 07/07
Range Specifications
Air Velocity Measurements
Range
Resolution
Accuracy (%rdg)
0.50 - 35.0 m/s
0.01 m/s<10
0.1 m/s >10
± (2% + 0.2m/s)
km/h (kilometers per hour)
1.8 – 126.0 km/h
0.1 km/h
± (2% + 0.8km/h)
ft/min (feet per minute)
100 – 6890 ft/min
1 ft/min
± (2% + 40ft/min)
mph (miles per hour)
1.1 – 78.3 mph
0.1 mph
± (2% + 0.4m/h)
knots (nautical miles per hour )
1.0 to 68 knots
0.1 knots
± (2% + 0.4knots)
Range
m/s (meters per second)
Air Flow Measurements
Resolution
Area
3
0.001 to 100
0 to 9,999m
3
CMM (cubic meters per minute)
0-999,900 m /min
CFM (cubic feet per minute)
0-999,900 ft /min
0.001 to 100
0 to 9,999ft
Range
Resolution
Accuracy
Air Temperature
o
o
32 to 175 F (0 to 80 C)
3
o
0.1 F/C
o
2
2
o
1.5 F (0.8 C)
Model 407113 V3.2 07/07
Meter Description
1.
Sensor Input
2.
RS-232 PC Interface jack
3.
LCD Display
4.
Vane
5.
Rubber Holster
6.
Keypad
7.
Battery compartment (rear)
3
Note: To access the rear battery compartment, first remove
the rubber holster that jackets the meter.
4
Model 407113 V3.2 07/07
Operation
Air Velocity Measurements
1. Connect the sensor to the sensor input jack on top of the meter.
2. Turn on the meter using the Power button.
3. Select the VELOCITY function using the FLOW / VELOCITY button. The LCD will
display VEL when the velocity mode is selected.
4. Select the desired temperature units using the C/F select button. The LCD will reflect
the current unit selection.
5. Select the desired air velocity units
using the UNIT button. The LCD will
reflect the current unit selection.
6. Place the sensor in the air current to
be measured with the
symbol on the inlet side of the vane
(see diagram).
Air Direction
7. View the air velocity and
temperature readings on the LCD
Display. The large main LCD
display shows the Air Velocity
reading. The lower LCD sub-display
shows the temperature reading.
Data Hold
1.
While taking measurements, press the HOLD button to freeze the LCD reading.
2.
The HOLD indicator will appear on the LCD when the display is in Data Hold mode.
3.
Press HOLD again to return to normal operation.
Maximum and Minimum Recording
The 407113 allows the user to record and view the highest (MAX) and lowest (MIN)
readings.
1. Press the RECORD/RECALL button once. The REC indicator will appear on the
display and the meter will begin recording the MAX and MIN values.
2. Press the RECORD/RECALL button to stop the MAX/MIN recording and display the
maximum reading. The MAX indicator along with the maximum reading will appear on
the LCD display
3. Press RECORD/RECALL again to view the minimum value. The MIN indicator along
with the minimum reading will appear on the LCD display.
4. To return to normal operation, press and hold the RECORD/RECALL button for
approx. 3 seconds. The display indicators REC, MAX, and MIN will disappear.
NOTE: Placing the meter in the RECORD/RECALL mode by pressing
the RECORD/RECALL button disables the AUTO POWER OFF feature.
NOTE: Handle the vane carefully. If the metal vane blades are inadvertently
bent or damaged erroneous readings may result.
5
Model 407113 V3.2 07/07
Air Flow (Volume) Measurements (CMM / CFM)
1. Connect the sensor to the sensor input jack on top of the meter.
2. Turn on the meter using the POWER button.
3. Select the FLOW mode using the FLOW/VELOCITY button. The LCD will display
FLOW CFM or FLOW CMM when the flow function has been selected.
4. Select the desired air flow units: CMM (cubic meters per minute) or CFM (cubic feet
per minute) using the UNIT button. The LCD will reflect the selection.
5. Measure the dimensions of the duct or vent and calculate the area in square feet or
square meters
Note: If the dimensional measurements are made in inches (or cm), convert them feet (or
meters) before calculating the square area.
2
2
6. Press the AREA button to begin entering the area in m or ft . Use the button to
increment the flashing digit, use the button to decrement, use the button to select
the next digit, and use the button to set the decimal point.
7. Press the ENTER/RESET button after the area value has been entered. The bottom
2
2
display will indicate the area entered in ft or m . The main LCD displays the Air Flow in
CFM (cubic feet per minute) or CMM (cubic meters per minute). If the CFM or CMM
reading exceeds 9999, use the displayed X10 or X100 multiplier to calculate the
reading.
8. Note that for FLOW measurements, three modes apply: The normal, default mode,
where the actual flow is indicated and the two modes described below under 2/3 MAX
Flow and AVG Flow.
Note: The temperature function is not active in the FLOW mode.
2/3V MAX Flow Mode
In this mode the meter will display two-thirds the measured Flow (Volume). To access this
mode, ensure that the meter is in the FLOW mode (via the FLOW/VELOCITY button) and
then press the FLOW MODE button until the 2/3V MAX display icon appears. Now the
Flow display will be 2/3 the actual measurement. To return to the normal display mode,
press the FLOW MODE button until the 2/3VMAX and the AVG icons switch off.
AVG (Average) Flow Mode
In this mode the meter will display the AVERAGE Flow for up to 20 readings. To access
this mode, ensure that the meter is in the FLOW mode (via the FLOW/VELOCITY button)
and press the FLOW MODE button until the AVG display icon appears. Now each time the
AVG/START button is pressed a reading will be taken and averaged. The lower LCD
display becomes a 20-reading counter and with each reading taken the counter is
incremented. Allow 3 or 4 seconds between AVG/START button presses for the display
counter to update. The main LCD displays the averaged Air Flow value. To return to the
normal display mode, press the FLOW MODE button until the AVG icon switches off.
RS-232 PC Interface
The 407113 is equipped with a 3.5mm phone jack (meter top) for connection to a PC for
TM
data acquisition purposes. To obtain PC interface cabling and Windows data acquisition
software (Model 407001), contact Extech Instruments or an authorized distributor.
Instructions for use are provided with the data acquisition software/hardware kits.
6
Model 407113 V3.2 07/07
Battery Replacement
When the LBT icon appears on the LCD, the 9V battery must be replaced.
1. Remove the rubber holster that surrounds the entire meter
2. Slide off the rear battery compartment
3. Replace the 9V battery
4. Affix the battery compartment cover and the meter holster
Useful Equations and Conversions
Area equation for rectangular or square ducts
Area (A) = Width (W) x Height (H)
Height (H)
Width (W)
Area equation for circular ducts
Area (A) = 6 x r2
Where 6 = 3.14 and r2 = radius x
radius
Radius
Cubic equations
CFM (ft3/min) = Air Velocity (ft/min) x Area (ft2)
CMM (m3/min) = Air Velocity (m/sec) x Area (m2) x 60
NOTE: Measurements made in inches
must be converted to feet or meters before using the above formulae.
Unit of Measure Conversion Table
m/s
ft/min
knots
km/h
MPH
1
196.87
1.944
3.6
2.24
1 ft/min
0.00508
1
0.00987
0.01829
0.01138
1 knot
0.5144
101.27
1
1.8519
1.1523
1 km/h
0.2778
54.69
0.54
1
0.6222
1 MPH
0.4464
87.89
0.8679
1.6071
1
1 m/s
7
Model 407113 V3.2 07/07
Warranty
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects
in parts and workmanship for one year from date of shipment (a six month limited warranty
applies to sensors and cables). If it should become necessary to return the instrument for
service during or beyond the warranty period, contact the Customer Service Department at
(781) 890-7440 ext. 210 for authorization or visit our website www.extech.com for contact
information. A Return Authorization (RA) number must be issued before any product is
returned to Extech. The sender is responsible for shipping charges, freight, insurance and
proper packaging to prevent damage in transit. This warranty does not apply to defects
resulting from action of the user such as misuse, improper wiring, operation outside of
specification, improper maintenance or repair, or unauthorized modification. Extech
specifically disclaims any implied warranties or merchantability or fitness for a specific
purpose and will not be liable for any direct, indirect, incidental or consequential damages.
Extech's total liability is limited to repair or replacement of the product. The warranty set
forth above is inclusive and no other warranty, whether written or oral, is expressed or
implied.
Calibration and Repair Services
Extech offers complete repair and calibration services for all of the products we sell.
For periodic calibration, NIST certification on most products or repair of any Extech
product, call customer service for details on services available. Extech recommends that
calibration be performed on an annual basis to ensure calibration integrity.
Support line (781) 890-7440
Technical support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451
Copyright © 2005 Extech Instruments Corporation.
All rights reserved including the right of reproduction in whole or in part in any form.