EXTECH 407123

User's Guide
Hot Wire Thermo-Anemometer
Model 407123
Introduction
Congratulations on your purchase of the Extech Hot Wire Anemometer. This instrument measures air
flow and temperature by placing the sensor into an airway such as a duct or a vent. The sensor is
situated at the end of the telescoping wand for convenience and has a protective sliding cover. The
meter includes an RS-232 PC Interface jack for use with the Model 407001 Data Acquisition software
and interface cable kit. This meter is shipped fully tested and calibrated and, with proper use, will provide
years of reliable service.
Specifications
General Specifications
Circuit configuration
Custom one-chip LSI microprocessor circuit
Display
3-1/2 digit (2000 count) dual LCD display
Measurements
m/s (meters per second), km/h (kilometers per hour), ft/min
(feet/per minute), knots (nautical miles per hour), MPH (miles per
hour), Temperature: °C, °F
Data hold
Freezes reading on display
Sensor Structure
Air velocity sensor: Glass bead thermistor
Temperature sensor: Precision thermistor
Min/Max Recording
Record and Recall Maximum (MAX) and Minimum (MIN) readings
PC Interface
RS 232 PC serial interface jack for use with Model 407001
software and interface cable kit
Operating
Temperature
32 to 122°F (0 to 50 °C)
Operating Humidity
Max. 80% RH
Power Supply
Six (6) 1.5V ‘AAA’ batteries
Power Consumption
Approx. 30 mADC
Weight
0.78 lb (355 g) including batteries & probe
Dimensions
Main instrument: 7.1 x 2.8 x 1.3” (180 x 72 x 32 mm)
Sensor: 0.5” (12mm) diameter
Telescope: 7’ (2.1m) maximum length with cable
Range Specifications
Units
Range
Resolution
Accuracy
m/s
0.2 to 20.0 m/s
0.1 m/s
±(3.0%rdg+0.3m/s)*
km/h
0.7 to 72.0 km/h
0.1 km/h
±(3.0%rdg+1.1km/h)*
ft/min
40 to 3940 ft/min
1 ft/min
±(3.0%rdg+59ft/min)*
MPH
0.5 to 45.0 MPH
0.1 MPH
±(3.0%rdg+0.67MPH)*
knots
0.4 to 31.0 knots
0.1 knots
±(3.0%rdg+0.58knots)*
*or, ±(1.0%FS+3d) whichever is greater
Temperature
o
o
o
o
o
32 to 122 F (0 to
0.1 F and C
1.5 F (0.8 C)
o
50 C)
Note: m/s: meters per second; km/h: kilometers per hour; ft/min: feet per minute; Knots:
nautical miles per hour; MPH: miles per hour
2
407123 V6.2 3/08
Description
Meter Description
1. LCD display – Indicates air velocity and
temperature readings, units of measure,
low battery icon, and other user alerts
9
2. Push-buttons:
• POWER: Turn meter ON and OFF
• HOLD: Freezes the displayed reading
• C/F: Select the temperature units
• RECORD: Press to track the highest
(MAX) and lowest (MIN) reading
• RECALL: Displays MAX/MIN readings
• UNIT: Air velocity units of measure
• ZERO: Press to zero the display
(sensor cover must be closed)
3. Protective holster – Rubber jacket that
surrounds the meter (must be removed
to access battery compartment)
4. Battery compartment - Located on the
lower back of the meter
5. PC Interface jack – Accepts a 3.5mm
plug from a PC interface cable (cable
and data acquisition software available
using part number 407001)
6. Sensor input jack – Insert the sensor
plug
7. Sensor opening – Air must flow through this opening for proper measurement
8. Telescoping sensor handle – Extends to 39” (1m)
9. Sensor protective cover – Slide DOWN to open when in use and slide UP to protect
sensor when not in use (close the cover when zeroing the meter)
1
Sensor Tip Description
2
4
1. Air Direction indentation
5
2. Air Velocity Sensor
3. Sensor cover in lower (measurement) open position
3
4. Sensor cover in upper (zero) closed position
5. Temperature sensor
3
407123 V6.2 3/08
Operation
Initialization and Zero
The meter should always be zeroed at temperature before use using the ZERO button.
Do not rely on the initial zero display.
Notes:
The meter does not display negative numbers.
Open antenna to desired length.
1. Connect the sensor to the input jack on top of the meter and open antenna to desired
length.
2. Turn on the meter using the Power button. The meter will perform a self-test during
which the display will count down from 9 to 0.
3. Select the desired temperature units using the C/F select button. The LCD will reflect
the selection.
4. Select the desired air velocity units using the UNIT button. The LCD will reflect the
selection.
5. Place the sensor cover in the UP (ZERO) position.
6. Place the sensor in the area to be measured and allow a short time for the sensor to
reach the temperature of the air under test.
7. Press the ZERO button to zero the meter.
Taking Measurements
1. Zero the meter as described above.
2. Slide the air velocity sensor cover down.
3. Place the sensor in the air current to be measured. Have the air flow meet the sensor
head in the direction of the indentation on the sensor head.
4. View the air velocity and temperature readings on the LCD Display. The large main
LCD display indicates the Air Velocity reading. The lower LCD sub-display indicates
the temperature reading.
4
407123 V6.2 3/08
MIN and MAX Function
1. To begin capturing the Minimum (MIN) and Maximum (MAX) readings, press the
RECORD button. The ‘REC’ icon will appear on the LCD.
2. Take readings as described previously.
3. Press the RECALL button to view the Maximum reading encountered since the
RECORD button was pressed. The ‘MAX’ display icon will appear along with the
maximum reading.
4. Press RECALL again to view the minimum reading (MIN).
5. Press the RECORD button to exit this mode and return to normal operation.
Data Hold
To freeze the LCD display, press the HOLD button. The 'DH' display icon will appear on
the LCD. Press the HOLD button again to return to normal operation (the 'DH' hold icon will
switch off).
5
407123 V6.2 3/08
Battery Replacement
When the low battery indicator (LBT) appears on the LCD or if the LCD does not switch on
when the POWER button is pressed, replace the batteries.
1. Remove the protective rubber holster that encapsulates the meter.
2. Open the rear battery compartment by prying the compartment off with a flat-head
screwdriver or a coin.
3. Replace the six ‘AAA’ 1.5V batteries observing polarity.
4. Replace the battery compartment cover and the protective holster.
PC Interface
The meter is equipped with an RS-232 PC Interface jack at the top of the meter (next to the
sensor input jack). This port is for use with the Extech Part Number 407001 Data
Acquisition software and interface cable kit. Directions for use are supplied with the
407001 kit. The interface kit allows the user to take measurements and view/save the
readings on a PC.
6
407123 V6.2 3/08
Useful Equations and Conversions
Area equations
The volume of air flowing through a duct or vent can be determined by taking the area of
the duct in square units (i.e. square feet) and multiplying this value by the measured
2
3
linear velocity (i.e., feet per minute). This gives: ft/min x ft = ft /min (CFM)
R
H
W
A = pi x R
A=WxH
2
Cubic equations
3
2
CFM (ft /min) = Air Velocity (ft/min) x Area (ft )
3
2
CMM (m /min) = Air Velocity (m/sec) x Area (m ) x 60
Units Conversion Table
m/s
ft/min
knots
km/hr
MPH
1
196.87
1.944
3.6
2.24
1 ft/min
0.00508
1
0.00987
0.01829
0.01138
1 knot
0.5144
101.27
1
1.8519
1.1523
1 km/hr
0.2778
54.69
0.54
1
0.6222
1 MPH
0.4464
87.89
0.8679
1.6071
1
1 m/s
7
407123 V6.2 3/08
Warranty
EXTECH INSTRUMENTS CORPORATION (A FLIR COMPANY) warrants this instrument
to be free of defects in parts and workmanship for one year from date of shipment (a six
month limited warranty applies to sensors and cables). If it should become necessary to
return the instrument for service during or beyond the warranty period, contact the
Customer Service Department at (781) 890-7440 ext. 210 for authorization or visit our
website www.extech.com for contact information. A Return Authorization (RA) number
must be issued before any product is returned to Extech. The sender is responsible for
shipping charges, freight, insurance and proper packaging to prevent damage in transit.
This warranty does not apply to defects resulting from action of the user such as misuse,
improper wiring, operation outside of specification, improper maintenance or repair, or
unauthorized modification. Extech specifically disclaims any implied warranties or
merchantability or fitness for a specific purpose and will not be liable for any direct, indirect,
incidental or consequential damages. Extech's total liability is limited to repair or
replacement of the product. The warranty set forth above is inclusive and no other
warranty, whether written or oral, is expressed or implied.
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also
provides NIST certification for most products. Call the Customer Care Department for
information on calibration services available for this product. Extech recommends that
annual calibrations be performed to verify meter performance and accuracy.
Support line (781) 890-7440
Technical support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451
Copyright © 2008 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in any form.
8
407123 V6.2 3/08