User's Guide Extech 700 Series 800 Amp Clamp Meters EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter Introduction Congratulations on your purchase of the Extech 700 Series Clamp meter. The series consists of the following models: EX710 AC Current Clamp meter EX720 True RMS AC Current Clamp with Temperature EX730 True RMS AC/DC Current Clamp with Temperature & DC zero function In addition to the parameters shown above in the individual model list, all models measure: AC/DC Voltage Resistance Capacitance Frequency Continuity Diode The Extech 700 Series features: Auto Power OFF Data Hold Peak Hold Backlit LCD display 2 EX700 Series Version 3.0 1/05 Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information. This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present Double insulation SAFETY NOTES • • • • Do not exceed the maximum allowable input range of any function. Do not apply voltage to meter when resistance function is selected. Set the function switch OFF when the meter is not in use. Remove the battery if meter is to be stored for longer than 60 days. WARNINGS • • • • Set function switch to the appropriate position before measuring. When measuring volts do not switch to current/resistance modes. Do not measure current on a circuit whose voltage exceeds 600V. When changing ranges always disconnect the test leads from the circuit under test. UL Note A UL mark does not indicate that this product has been evaluated for accuracy. CAUTIONS • • • • • • • Improper use of this meter can cause damage, shock, injury or death. Read and understand this user manual before operating the meter. Always remove the test leads before replacing the battery or fuses. Inspect the condition of the test leads and the meter itself for any damage before operating the meter. Repair or replace any damage before use. Use great care when making measurements if the voltages are greater than 25VAC rms or 35VDC. These voltages are considered a shock hazard. Always discharge capacitors and remove power from the device under test before performing Diode, Resistance or Continuity tests. Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty of connection to the recessed electrical contacts. Other means should be used to ensure that the terminals are not "live". If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Function Maximum Input A AC, A DC (A DC on Model EX730 only) 800A DC/AC V DC, V AC 600V DC/AC Resistance, Capacitance, Frequency, Diode Test 250V DC/AC Temperature (EX720 & EX730 only) 60V DC, 24V AC 3 EX700 Series Version 3.0 1/05 Description Meter Description (EX730 pictured) 1. Current clamp 2. Clamp opening trigger 3. Control buttons: 1 Data Hold Mode Peak Backlight Backlit LCD Display 5. COM negative input jack for black test lead 6. Rotary function switch 7. V·Ω·CAP·TEMP·Hz· lead 6 2 DCA Zero (EX730 only) 4. positive input jack for red 3 4 Display icons Description HOLD Data Hold Minus sign Negative reading display 0 to 3999 Measurement display digits DC ZERO DCA Zero (Model EX730 only) PEAK Peak capture AUTO Auto Range mode DC/AC Direct Current / Alternating Current BAT Low battery mV or V Milli-volts or Volts (Voltage) Ω Ohms (Resistance) A Amperes (Current) F Farad (Capacitance) Hz Hertz (Frequency) o o F and C n, m, µ, M, k •))) 7 5 Fahrenheit and Celsius units (Temperature) Unit of measure prefixes: nano, milli, micro, mega, and kilo Continuity test Diode test 4 EX700 Series Version 3.0 1/05 Specifications Function AC Current Range & Resolution Accuracy (% of reading + digits) EX720 EX710 EX730 40.00A AC ± (2.8% + 10d) ± (2.5% + 10d) ± (5% + 10d) 400.0A AC ± (2.8% + 8d) ± (2.5% + 8d) ± (5% + 10d) True RMS on EX720 & EX730 800A AC ± (3.0% + 8d) ± (2.8% + 5d) ± (2.8% + 10d) DC Current 40.00A DC EX730 only 400.0A DC NA NA ± (1.5% + 10d) ± (1.0% + 10d) ± (1.0% + 10d) ± (1.8% + 8d) ± (1.5% + 5d) ± (1.5% + 5d) 600V AC ± (2.5% + 8d) ± (2.0% + 5d) ± (2.0% + 5d) 400.0mV DC ± (0.8% + 2d) ± (0.8% + 2d) ± (0.8% + 2d) ± (1.5% + 2d) ± (1.5% + 2d) ± (1.5% + 2d) 600V DC ± (2.0% + 2d) ± (2.0% + 2d) ± (2.0% + 2d) 400.0Ω ± (1.0% + 4d) ± (1.0% + 4d) ± (1.0% + 4d) ± (1.5% + 2d) ± (1.5% + 2d) ± (1.5% + 2d) 50/60 Hz ± (5% + 10d) ± (5% + 10d) 800A DC 400.0mV AC AC Voltage 50/60Hz True RMS on EX720 & EX730 ± (2.8% + 10d) 4.000V AC 40.00V AC 400.0V AC 4.000V DC DC Voltage 40.00V DC 400.0V DC 4.000kΩ Resistance 40.00kΩ 400.0kΩ 4.000MΩ ± (2.5% + 3d) ± (2.5% + 3d) ± (2.5% + 3d) 40.00MΩ ± (3.5% + 5d) ± (3.5% + 5d) ± (3.5% + 5d) 4.000nF ± (5.0% + 30d) ± (5.0% + 30d) ± (5.0% + 30d) 40.00nF ± (5.0% + 20d) ± (5.0% + 20d) ± (5.0% + 20d) ± (3.0% + 5d) ± (3.0% + 5d) ± (3.0% + 5d) 400.0µF ± (4.0% + 10d) ± (4.0% + 10d) ± (4.0% + 10d) 400.0nF Capacitance Frequency 4.000µF 40.00µF 4.000mF ± (10% + 10d) ± (10% + 10d) ± (10% + 10d) 40.00mF unspecified unspecified unspecified 4.000kHz ± (1.5% + 2d) ± (1.5% + 2d) ± (1.5% + 2d) Sensitivity: 100V (<50Hz); 50V (50 to 400Hz); 5V (401 to 4000Hz) 5 EX700 Series Version 3.0 1/05 Function Range and Resolution Temperature Accuracy (% of reading + digits) EX710 EX720 o NA ± (3%rdg + 9 F) o NA ± (3%rdg + 5 C) -4 to 1400 F -20 to 760 C EX730 o ± (3%rdg + 9 F) o o ± (3%rdg + 5 C) o General Specifications Clamp jaw opening 1.2" (30mm) approx. Display 3-3/4 digits (4000 counts) backlit LCD Continuity check Threshold 40Ω; Test current < 0.5mA Diode test Test current of 0.3mA typical; Open circuit voltage < 3VDC typical Low Battery indication ‘BAT’ is displayed Over-range indication ‘OL’ display Measurement rate 2 readings per second, nominal PEAK Captures peaks >1ms Temperature sensor Type K thermocouple Input Impedance 10MΩ (VDC and VAC) AC bandwidth 50 to 400Hz (AAC and VAC) AC response True rms (AAC and VAC) on EX720 & EX730 Crest Factor 3.0 in 40A and 400A ranges, 1.4 in 800A range (50/60Hz and 5% to 100% of range) Operating Temperature 41 F to 104 F (5 C to 40 C) Storage Temperature o o o o o o o o -4 F to 140 F (-20 C to 60 C) o o Operating Humidity Max 80% up to 87 F (31 C) decreasing linearly to 50% at o o 104 F(40 C) Storage Humidity <80% Operating Altitude 7000ft. (2000meters) maximum. Battery One (1) 9V Battery Auto power OFF After approx. 25 minutes Dimensions & Weight Safety 9.0 x 3.1 x 2.0" (229 x 80 x 49mm); 10.7 oz. (303g) For indoor use and in accordance with the requirements for double insulation to IEC1010-1 (2001): EN61010-1 (2001) Overvoltage Category III 600V and Category II 1000V, Pollution Degree 2. 6 EX700 Series Version 3.0 1/05 Operation NOTES: Read and understand all Warning and Caution statements in this operation manual prior to using this meter. Set the function select switch to the OFF position when the meter is not in use. AC/DC Current Measurements (DC Current on Model EX730 only) WARNING: Ensure that the test leads are disconnected from the meter before making current clamp measurements. 1. Set the Function switch to the 800A, 400A, or 40A range. If the approx. range of the measurement is not known, select the highest range then move to the lower ranges if necessary. 2. Use the MODE button to select AC or DC (EX730 only). 3. Press the DC ZERO button to zero the meter display (EX730, DC only). 4. Press the trigger to open jaw. Fully enclose only one conductor. For optimum results, center the conductor in the jaw. 5. The clamp meter LCD will display the reading. Correct Incorrect AC/DC Voltage Measurements 1. Insert the black test lead into the negative COM terminal and the red test lead into the positive V·Ω·CAP·TEMP·Hz· terminal. 2. Set the function switch to the V Hz position. 3. Use the MODE button to select AC or DC Voltage. 4. Connect the test leads in parallel to the circuit under test. 5. Read the voltage measurement on the LCD display. V 7 EX700 Series Version 3.0 1/05 Resistance Measurements 1. Insert the black test lead into the negative COM terminal and the red test lead into the V·Ω·CAP·TEMP·Hz· positive terminal. 2. Set the function switch to the Ω position. 3. Touch the test probe tips across the circuit or component under test. 4. Read the resistance on the LCD display. Capacitance Measurements WARNING: To avoid electric shock, discharge the capacitor under test before measuring. 1. Set the function switch to the CAP position. 2. Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the V·Ω·CAP·TEMP·Hz· positive jack. 3. Touch the test probe tips across the part under test. If “dISC” appears in the display, remove and discharge the component. 4. Read the capacitance value in the display. 5. The display will indicate the proper decimal point and value. Note: For very large values of capacitance measurement it can take several minutes before the final reading stabilizes. CAP Frequency Measurements 1. Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the V·Ω·CAP·TEMP·Hz· positive jack. 2. Set the function switch to the V Hz Position. 3. Press and hold the MODE button to select the Frequency (Hz) function. “k Hz” will appear in the display. 4. Touch the test probe tips across the part under test. 5. Read the Frequency value on the display. 6. The display will indicate the proper decimal point and value. 7. Press and hold the MODE button again to return to the voltage mode. 8 Hz EX700 Series Version 3.0 1/05 Temperature Measurements (Models EX720 and EX730 only) 1. Set the function switch to the TEMP position. 2. Insert the Temperature Probe into the negative COM and the V·Ω·CAP·TEMP·Hz· positive jacks, observing polarity. 3. Touch the Temperature Probe head to the device under test. Continue to touch the part under test with the probe until the reading stabilizes. 4. Read the temperature on the display. The digital reading will indicate the proper decimal point and value. 5. Use the MODE button to select F or C. o o WARNING: To avoid electric shock, be sure the thermocouple probe has been removed before changing to another measurement function. Continuity Measurements 1. Insert the black test lead into the negative COM terminal and the red test lead into the V·Ω·CAP·TEMP·Hz· positive terminal. 2. Set the function switch to the •))) position. 3. Use the MODE button to select continuity “•)))”. The display icons will change when the MODE button is pressed. 4. Touch the test probe tips across the circuit or component under test. 5. If the resistance is < 40Ω, a tone will sound. Diode Test 1. Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the V·Ω·CAP·TEMP·Hz· positive jack 2. Turn the function switch to position. Use the MODE button to select the diode function if necessary (diode symbol will appear on the LCD when in Diode test mode) 3. Touch the test probe tips to the diode or semiconductor junction under test. Note the meter reading 4. Reverse the test lead polarity by reversing the red and black leads. Note this reading 5. The diode or junction can be evaluated as follows: If one reading displays a value (typically 0.400V to 0.900V) and the other reading displays OL, the diode is good. If both readings display OL the device is open. If both readings are very small or ‘0’, the device is shorted. 9 EX700 Series Version 3.0 1/05 Data Hold To freeze the LCD reading, press the HOLD button. While data hold is active, the HOLD icon appears on the LCD. Press the HOLD button again to return to normal operation. DC ZERO (Relative) (EX730 only) The DC ZERO is a relative feature and can be used in any function. 1. Press the DC ZERO button to zero the display. “ZERO” will appear in the display. The displayed reading is now the actual value less the stored “zero” value. 2. Press the DC ZERO button to view the stored value. “ZERO” will flash in the display. 3. To exit this mode, press and Hold the ZERO button until “ZERO” is no longer in the display. Peak Hold The Peak Hold function captures the peak AC or DC voltage or current. The meter can capture negative or positive peaks as fast as 1 millisecond in duration. 1. Turn the function switch to the A or V position. 2. Use the MODE button to select AC or DC (ADC EX730 only). 3. Allow time for the display to stabilize. 4. Press and Hold the PEAK button until “CAL” appears in the display. This procedure will zero the range selected. 5. Press the PEAK button, Pmax will display. 6. The display will update each time a higher positive peak occurs. 7. Press the PEAK button again, Pmin will display. The display will now update and indicate the lowest negative peak. 8. To return to normal operation, press and hold the PEAK button until the Pmin or Pmax indicator switches off. Note: If the Function switch position is changed after a calibration the Peak Hold calibration must be repeated for the new function selected. LCD Backlight Button The LCD is equipped with backlighting for easier viewing, especially in dimly lit areas. Press the backlight button to turn the backlight on. Press again to turn the backlight off. Note that the meter does have an auto power off feature as described below. Automatic Power OFF In order to conserve battery life, the meter will automatically turn off after approximately 25 minutes. To turn the meter on again, turn the function switch to the OFF position and then to the desired function position. 10 EX700 Series Version 3.0 1/05 Maintenance WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads from the input terminals, and turn OFF the meter before opening the case. Do not operate the meter with an open case. Cleaning and Storage Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. If the meter is not to be used for 60 days or more, remove the battery and store it separately. Battery Replacement 1. Remove the Phillips head screw that secures the rear battery door 2. Open the battery compartment 3. Replace the 9V battery 4. Secure the battery compartment Temperature Probe Replacement The replacement bead wire probe (with banana plug connectors) is Part Number TP873. Note: To use a Type K thermocouple probe that is terminated by a subminiature (flat blade) connector, a subminiature-to-banana plug adaptor (Part Number TP879) is required. 11 EX700 Series Version 3.0 1/05 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a six month limited warranty applies to sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department at (781) 890-7440 ext. 210 for authorization or visit our website www.extech.com for contact information. A Return Authorization (RA) number must be issued before any product is returned to Extech. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied. Calibration and Repair Services Extech offers repair and calibration services for the products we sell. Extech also provides NIST certification for most products. Call the Customer Service Department for information on calibration services available for this product. Extech recommends that annual calibrations be performed to verify meter performance and accuracy. Support line (781) 890-7440 Technical support: Extension 200; E-mail: [email protected] Repair & Returns: Extension 210; E-mail: [email protected] Product specifications subject to change without notice For the latest version of this User’s Guide, Software updates, and other up-to-the-minute product information, visit our website: www.extech.com Copyright © 2004 Extech Instruments Corporation All rights reserved including the right of reproduction in whole or in part in any form. 12 EX700 Series Version 3.0 1/05