English User's Manual

User’s Guide
400A AC/DC Clamp Meter
Model MA220
Introduction
Thank you for selecting the Extech MA220 AC/DC Clamp Meter. This meter measures AC/DC Current,
AC/DC Voltage, Resistance, Capacitance, Frequency, Duty Cycle, Temperature, Diode Test, and
Continuity. This professional meter, with proper care, will provide years of safe reliable service.
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to the
manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous
voltages may be present
Double insulation
SAFETY NOTES

Do not exceed the maximum allowable input range of any function

Do not apply voltage to meter when resistance function is selected.

Set the function switch OFF when the meter is not in use.

Remove the battery if meter is to be stored for longer than 60 days.
WARNINGS

Set function switch to the appropriate position before measuring.

When measuring volts do not switch to current/resistance modes.

Do not measure current on a circuit whose voltage exceeds 240V.

When changing ranges using the selector switch always disconnect the test leads from the
circuit under test.

Do not exceed the maximum rated input limits.
OVERVOLTAGE CATEGORY III 600V
This meter meets the IEC 610-1-95 standard for OVERVOLTAGE CATEGORY III 600V. Cat III
600V meters are protected against overvoltage transients in fixed installation at the distribution
level. Examples include switches in the fixed installation and some equipment for industrial use with
permanent connection to the fixed installation.
2
MA220-en-EU_V1.9 8/14
Meter Description
1.
Conductor jaws
2.
Jaw opening trigger
3.
Function select switch
4.
LCD Display
5.
ZERO button
6.
Data Hold and Backlight Button
7.
Mode select button
8.
Range select button
9.
Hz/%/Duty Cycle button
10.
COM input jack
11.
V/Temp jack
12.
Battery cover (rear)
AC
AC (alternating current)
DC
DC (direct currrent)
Minus sign
AutoRange mode
ZERO mode
Audible Continuity
Data Hold mode
Low Battery icon
AUTO
ZERO
•)))
HOLD
m
V
A
K
M
Ω
°F
°C
Diode test mode
milli
Volts
Amps
kilo
Mega
Ohms
Degrees Fahrenheit
Degrees Centigrade
3
MA220-en-EU_V1.9 8/14
Operation
Notice: Read and understand all WARNING and CAUTION statements listed in the safety section
of this operation manual prior to using this meter. Set the function select switch to the OFF position
when the meter is not in use.
DC/AC Current Measurements
Warning: Disconnect the test leads from the meter before making current
clamp measurements.
1.
2.
3.
4.
Set the Function switch to the 400ADC, 40ADC, 400AAC or 40AAC
range. If the range of the measured is not known, select the higher range
first then move to the lower range if necessary.
For DC current measurement, press the ZERO key to null the meter
display.
Press the trigger to open jaw. Fully enclose one conductor to be measured.
The clamp meter LCD will display the reading.
DC/AC Voltage Measurements
1. Set the rotary function switch to the Volts/Hz/% position.
2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead banana
plug into the positive (V/Temp) jack
3. Select AC or DC with the MODE button
4. Connect the test leads to the circuit under test
5. Read the voltage on the display. The display will indicate the proper decimal point and value.
Resistance Measurements
•))) CAP position.
1. Set the function switch to the 
2. Insert the black test lead banana plug into the negative (COM) jack
Insert the red test lead banana plug into the positive (V Temp jack.
3. Touch the test probe tips across the circuit or part under test. It is best to disconnect one side of
the part under test so the rest of the circuit will not interfere with the resistance reading.
4. Read the resistance on the display. The display will indicate the proper decimal point and value.
Continuity Check
•))) CAP position.
1. Set the function switch to the 
2. Push the mode button to indicate •))) on the display.
3. Insert the black lead banana plug into the negative (COM) jack
Insert the red test lead banana plug into the positive (V) jack.
4. Touch the test probe tips to the circuit or wire you wish to check.
5. If the resistance is less than approximately 150, the audible signal will sound. If the circuit is
open, the display will indicate “OL.”.
4
MA220-en-EU_V1.9 8/14
Diode Test
1. Turn the rotary switch to the 
•))) CAP position.
2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead banana
plug into the positive (V jack.
3. Push the mode button to indicate
on the display.
4. Touch the test probes to the diode under test. Typically for a normal diode, forward voltage will
indicate 0.4V to 0.7V. Reverse voltage will indicate “OL”. Shorted devices will indicate near 0V and
an open device will indicate “OL” in both polarities.
Capacitance Measurements
Warning: To avoid electrical shock, disconnect power to the unit under test and discharge all
capacitors before taking any capacitance measurements. Remove the batteries and unplug the
line cords.
1.
2.
3.
4.
5.
6.
Set the function switch to the 
•))) CAP position.
Push the mode button to indicate nF on the display.
Insert the black lead banana plug into the negative (COM) jack and
insert the red test lead banana plug into the positive (VTemp) jack.
Press the ZERO key to null the meter display.
Touch the test probe tips to the capacitor you wish to check.
Read the capacitance value on the display.
Frequency or % Duty Cycle Measurements
1.
2.
3.
4.
5.
Turn the rotary switch to the Volts Hz % position.
Insert the black test lead banana plug into the negative (COM) jack and the red test lead banana
plug into the positive (V jack.
Select Hz or % with the HZ/% button.
Touch the test probe tips to the circuit under test.
Read the frequency on the display.
Temperature Measurements
1.
2.
Turn the rotary switch to the Temp position.
Insert the Temperature Probe into the negative (COM) and the (VTemp jacks, making sure to
observe correct polarity.
3. Select °C or °F with the MODE button.
4. Touch the temperature probe head to the part whose temperature you wish to measure. Keep the
probe touching the part under test until the reading stabilizes.
5. Read the temperature on the display.
Warning: To avoid electrical shock, be sure the thermocouple has been removed before changing
to another measurement function.
Auto/Manual Ranging
The meter turns on in Autoranging mode. Press the RANGE button to enter manual ranging. Each
press of the range button will step to the next range as indicated by the units and decimal point
location. Press and hold the RANGE button for two seconds to return to Autoranging mode.
Note: Manual ranging does not function in AC Current or Diode and Continuity check functions. In
Temperature function, it will change the resolution from 0.1° to 1°.
5
MA220-en-EU_V1.9 8/14
Data Hold
To freeze the LCD meter reading, press the HOLD button. While data hold is active, the HOLD
display icon appears on the LCD. Press the HOLD button again to return to normal operation.
Backlight
Press and hold the HOLD button for >2 seconds to turn the backlight on/off.
Note: The HOLD feature will activate when the backlight is turned on. Press the HOLD button again
to exit the Hold feature.
Zero Button
Zeros Capacitance and DC Current measurements. Also allows the user to offset the meter by
using the displayed value as the zero reference value. Press the ZERO key momentarily to activate
and to exit Zero mode.
Maintenance
WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads
from the input terminals and turn OFF the meter before opening the case. Do not operate with open
case.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. If
the meter is not to be used for periods of longer than 60 days, remove the batteries and store them
separately
Battery Replacement
1. Remove the two rear Phillips head screws
2.
Open the battery compartment
3.
Replace the two 1.5V AAA batteries.
4.
Re-assemble the meter
You, as the end user, are legally bound (Battery ordinance) to return all used batteries
and accumulators; disposal in the household garbage is prohibited!
You can hand over your used batteries / accumulators at collection points in your
community or wherever batteries / accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the
end of its lifecycle
6
MA220-en-EU_V1.9 8/14
Specifications
Function
Range
DC Current
40.00 ADC
0-20.00 ADC
20.00-40.00 ADC
± (3% + 6 digits)
400.0 ADC
0-300.0 ADC
± (2.5% + 6 digits)
AC Current
Accuracy (of reading)
40.00 AAC
400.0 AAC
DC Voltage
AC Voltage
Resistance
Capacitance
Frequency
Duty Cycle
Temperature
± (2.5% + 6 digits)
300.0-400.0 ADC
± (3.5% + 6 digits)
0-20.00 AAC
20.00-40.00 AAC
0-300.0 AAC
300.0-400.0 AAC
± (3% + 10 digits)
± (5% + 10 digits)
± (3% + 10 digits)
± (5% + 10 digits)
± (0.8% + 3 digits)
400.0mV
4.000V
± (1.5% + 3 digits)
40.00V
400.0V
600V
± (2.0% + 3 digits)
400.0mV
± (1% + 10 digits)
4.000V
± (2% + 5 digits)
40.00V
400.0V
600V
± (2% + 5 digits)
± (1.0% + 4 digits)
400.0
4.000k
± (1.5% + 2 digits)
40.00k
400.0k
± (2.5% + 3 digits)
4.000 M
± (3.5% + 5 digits)
40.00M
40.00nF
± (5% + 3 digits)
400.0nF
± (3% + 5 digits)
4.000µF
± (3.5% + 5 digits)
40.00µF
100.0µF
± (5% + 5 digits)
5.000Hz
± (1.5% + 5 digits)
50.00Hz
± (1.2% + 2 digits)
500.0Hz
Sensitivity: 5~5KHz:
5.000KHz
10Vrms min.
5KHz~150KHz: 40Vrms
50.00KHz
min.
150.0KHz
0.5% to 99.0%
± (1.2% + 2 digits)
Pulse Width: 100µs-100ms, Frequency: 5Hz to 150KHz
-50.0 to 400.0°C
-50.0 to -20.0°C
± 7°C
-20.0 to 400.0°C
± (3% + 5 °C)
400 to 1000°C
400 to 1000°C
-58.0 to 400.0°F
-58.0 to 0°F
± 14°F
0 to 400.0°F
± (2.5% + 6 digits)
400 to 1832°F
± (3% + 7°F)
400 to 1832°F
7
MA220-en-EU_V1.9 8/14
Jaw size
23mm (0.9") approx.
Display
Continuity
4000 count LCD
Audible tone < 150 approx.
Diode Test
Open circuit voltage < 1.5VDC;
AC V bandwidth
50Hz to 400Hz
Test current 0.3mA (typical)
AC A bandwidth
50/60Hz
Low battery indication
“ ” is displayed
Overrange indication
“OL” is displayed
Auto Power OFF
After 30 minutes
Measurement rate
2 per second, nominal
Input Impedance
7.8MΩ (V DC and V AC)
Operating Temperature
5ºC to 40ºC (41ºF to 104ºF)
Storage Temperature
-20 C to 60 C (-4 F to 140 F)
o
o
o
o
Operating Humidity
Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at 45ºC (113ºF)
Storage Humidity
<80%
Operating Altitude
2000 meters (6560ft) operating
Batteries
(2) 1.5V AAA batteries
Weight
200g (0.44lb)
Size
200x50x35mm (7.87” x 1.97” x 1.38”)
Safety
For indoor use and in accordance with the requirements for double
insulation to IEC1010-1 (1995): EN61010-1 (1995) Overvoltage
Category III 600V, Pollution Degree 2.