NI NIUSB-6009

Low-Cost, Bus-Powered Multifunction DAQ for USB –
12- or 14-Bit, up to 48 kS/s, 8 Analog Inputs
NI USB-6008, NI USB-6009
• 8 analog inputs at 12 or 14 bits,
up to 48 kS/s
• 2 analog outputs at 12 bits,
software-timed
• 12 TTL/CMOS digital I/O lines
• 32-bit, 5 MHz counter
• Digital triggering
• Bus-powered
• 1-year warranty
Recommended Software
•
•
•
•
Other Compatible Software
• C#, Visual Basic .NET
• ANSI C/C++
Operating Systems
•
•
•
•
•
LabVIEW
LabVIEW SignalExpress
LabWindows™/CVI
Measurement Studio
Windows Vista (32- and 64-bit)/XP/2000
Mac OS X1
Linux®1
Windows Mobile1
Windows CE1
Measurement Services
Software (included)
• NI-DAQmx driver software
• Measurement & Automation
Explorer configuration utility
• LabVIEW SignalExpress LE
1You
need to download NI-DAQmx
Base for these operating systems.
Product
USB-6009
USB-6008
1 SE
Bus
USB
USB
Analog
Inputs1
8 SE/4 DI
8 SE/4 DI
= single ended, DI = differential
Input
Resolution
(bits)
14
12
Max
Sampling Rate
(kS/s)
48
10
Input
Range
(V)
±1 to ±20
±1 to ±20
Analog
Outputs
2
2
Output
Resolution
(bits)
12
12
Output
Rate
(Hz)
150
150
Output
Range
(V)
0 to 5
0 to 5
Digital
I/O
Lines
12
12
32-Bit
Counter
1
1
Trigger
Digital
Digital
2Software-timed
Overview and Applications
With recent bandwidth improvements and new innovations from
National Instruments, USB has evolved into a core bus of choice for
measurement applications. The NI USB-6008 and USB-6009 are lowcost entry points to NI flagship data acquisition (DAQ) devices. With
plug-and-play USB connectivity, these modules are simple enough for
quick measurements but versatile enough for more complex
measurement applications.
The USB-6008 and USB-6009 are ideal for a number of applications
where low cost, small form factor, and simplicity are essential.
Examples include:
• Data logging – quick and easy environmental or voltage data logging
• Academic lab use – student ownership of DAQ hardware for
completely interactive lab-based courses (Academic pricing available.
Visit ni.com/academic for details.)
• OEM applications as I/O for embedded systems
Recommended Software
National Instruments measurement services software, built around
NI-DAQmx driver software, includes intuitive application programming
interfaces, configuration tools, I/O assistants, and other tools
designed to reduce system setup, configuration, and development time.
National Instruments recommends using the latest version of NI-DAQmx
driver software for application development in NI LabVIEW, LabVIEW
SignalExpress, LabWindows/CVI, and Measurement Studio software.
To obtain the latest version of NI-DAQmx, visit
ni.com/support/daq/versions.
NI measurement services software speeds up your development with
features including:
• A guide to create fast and accurate measurements with no
programming using the DAQ Assistant.
• Automatic code generation to create your application in LabVIEW.
• LabWindows/CVI; LabVIEW SignalExpress; and C#, Visual Studio .NET,
ANSI C/C++, or Visual Basic using Measurement Studio.
• Multithreaded streaming technology for 1,000 times
performance improvements.
• Automatic timing, triggering, and synchronization routing
to make advanced applications easy.
• More than 3,000 free software downloads available at
ni.com/zone to jump-start your project.
• Software configuration of all digital I/O features without
hardware switches/jumpers.
• Single programming interface for analog input, analog output,
digital I/O, and counters on hundreds of multifunction DAQ hardware
devices. M Series devices are compatible with the following versions
(or later) of NI application software – LabVIEW, LabWindows/CVI, or
Measurement Studio versions 7.x; and LabVIEW SignalExpress 2.x.
Low-Cost, Bus-Powered Multifunction DAQ for USB – 12- or 14-Bit, up to 48 kS/s, 8 Analog Inputs
Every M Series data acquisition device also includes a copy of
LabVIEW SignalExpress LE data-logging software, so you can quickly
acquire, analyze, and present data without programming. The NI-DAQmx
Base driver software is provided for use with Linux, Mac OS X,
Windows Mobile, and Windows CE operating systems.
Recommended Accessories
The USB-6008 and USB-6009 have removable screw terminals for easy
signal connectivity. For extra flexibility when handling multiple wiring
configurations, NI offers the USB-600x Connectivity Kit, which includes
two extra sets of screw terminals, extra labels, and a screwdriver.
In addition, the USB-600x Prototyping Kit provides space for adding
more circuitry to the inputs of the USB-6008 or USB-6009.
Information for Student Ownership
To supplement simulation, measurement, and automation theory courses
with practical experiments, NI has developed the USB-6008 and USB-6009
student kits, which include the LabVIEW Student Edition and a ready-to-run
data logger application. These kits are exclusively for students, giving them
a powerful, low-cost, hands-on learning tool. Visit ni.com/academic for
more details.
Information for OEM Customers
For information on special configurations and pricing, call (800) 813 3693
(U.S. only) or visit ni.com/oem. Go to the Ordering Information section
for part numbers.
NI USB DAQ for OEMs
Shorten your time to market by integrating world-class National
Instruments OEM measurement products into your embedded system
design. Board-only versions of NI USB DAQ devices are available for
OEM applications, with competitive quantity pricing and available
software customization. The NI OEM Elite Program offers free 30-day
trial kits for qualified customers. Visit ni.com/oem for more information.
Ordering Information
NI USB-60081 ........................................................................779051-01
NI USB-60091 ........................................................................779026-01
NI USB-6008 OEM ................................................................193132-02
NI USB-6009 OEM ................................................................193132-01
NI USB-6008 Student Kit1,2 ..................................................779320-22
NI USB-6009 Student Kit1,2 ..................................................779321-22
NI USB-600x Connectivity Kit ..............................................779371-01
NI USB-600x Prototyping Kit ................................................779511-01
1 Includes
NI-DAQmx software, LabVIEW SignalExpress LE, and a USB cable.
2 Includes
LabVIEW Student Edition.
BUY NOW!
For complete product specifications, pricing, and accessory
information, call 800 813 3693 (U.S. only) or go to ni.com/usb.
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)
2
Low-Cost, Bus-Powered Multifunction DAQ for USB – 12- or 14-Bit, up to 48 kS/s, 8 Analog Inputs
Specifications
Typical at 25 °C unless otherwise noted.
Analog Input
Absolute accuracy, single-ended
Range
±10
Typical at 25 ˚C (mV)
14.7
Maximum (0 to 55 ˚C) (mV)
138
Absolute accuracy at full scale, differential1
Range
±20
±10
±5
±4
±2.5
±2
±1.25
±1
Typical at 25 ˚C (mV)
14.7
7.73
4.28
3.59
2.56
2.21
1.70
1.53
Maximum (0 to 55 ˚C) (mV)
138
84.8
58.4
53.1
45.1
42.5
38.9
37.5
Number of channels............................ 8 single-ended/4 differential
Type of ADC ........................................ Successive approximation
ADC resolution (bits)
Module
USB-6008
USB-6009
Differential
12
14
Single-Ended
11
13
Maximum sampling rate (system dependent)
Module
USB-6008
USB-6009
Maximum Sampling Rate (kS/s)
10
48
Input range, single-ended................... ±10 V
Input range, differential...................... ±20, ±10, ±5, ±4, ±2.5, ±2,
±1.25, ±1 V
Maximum working voltage ................. ±10 V
Overvoltage protection ....................... ±35 V
FIFO buffer size ................................... 512 B
Timing resolution ................................ 41.67 ns (24 MHz timebase)
Timing accuracy .................................. 100 ppm of actual sample rate
Input impedance ................................. 144 kΩ
Trigger source...................................... Software or external digital trigger
System noise....................................... 5 m Vrms (±10 V range)
Analog Output
Absolute accuracy (no load) ............... 7 mV typical, 36.4 mV maximum
at full scale
Number of channels............................ 2
Type of DAC ........................................ Successive approximation
DAC resolution.................................... 12 bits
Maximum update rate ........................ 150 Hz, software-timed
1 Input
Output range .......................................
Output impedance...............................
Output current drive............................
Power-on state....................................
Slew rate.............................................
Short-circuit current............................
0 to +5 V
50 Ω
5 mA
0V
1 V/µs
50 mA
Digital I/O
Number of channels............................ 12 total
8 (P0.<0..7>)
4 (P1.<0..3>)
Direction control ................................. Each channel individually
programmable as input or output
Output driver type
USB-6008........................................ Open-drain
USB-6009........................................ Each channel individually
programmable as push-pull or
open-drain
Compatibility ....................................... CMOS, TTL, LVTTL
Internal pull-up resistor ...................... 4.7 kΩ to +5 V
Power-on state.................................... Input (high impedance)
Absolute maximum voltage range...... -0.5 to +5.8 V
Digital logic levels
Level
Input low voltage
Input high voltage
Input leakage current
Output low voltage (I = 8.5 mA)
Output high voltage (push-pull, I = -8.5 mA)
Output high voltage (open-drain, I = -0.6 mA, nominal)
Output high voltage (open-drain, I = -8.5 mA,
with external pull-up resistor)
Min
-0.3
2.0
–
–
2.0
2.0
Max
0.8
5.8
50
0.8
3.5
5.0
Units
V
V
µA
V
V
V
2.0
–
V
Counter
Number of counters ............................
Resolution ...........................................
Counter measurements.......................
Pull-up resistor....................................
Maximum input frequency..................
Minimum high pulse width.................
Minimum low pulse width..................
Input high voltage ...............................
Input low voltage ................................
1
32 bits
Edge counting (falling edge)
4.7 kΩ to 5 V
5 MHz
100 ns
100 ns
2.0 V
0.8 V
Power available at I/O connector
+5 V output (200 mA maximum) ......... +5 V typical
+4.85 V minimum
+2.5 V output (1 mA maximum) .......... +2.5 V typical
+2.5 V output accuracy ....................... 0.25% max
Voltage reference temperature drift... 50 ppm/°C max
voltages may not exceed the working voltage range.
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)
3
Low-Cost, Bus-Powered Multifunction DAQ for USB – 12- or 14-Bit, up to 48 kS/s, 8 Analog Inputs
Physical Characteristics
Safety and Compliance
If you need to clean the module, wipe it with a dry towel.
Dimensions (without connectors) ....... 6.35 by 8.51 by 2.31 cm
(2.50 by 3.35 by 0.91 in.)
Dimensions (with connectors) ............ 8.18 by 8.51 by 2.31 cm
(3.22 by 3.35 by 0.91 in.)
Weight (without connectors) .............. 59 g (2.1 oz)
Weight (with connectors) ................... 84 g (3 oz)
I/O connectors..................................... USB series B receptacle
(2) 16-position (screw-terminal)
plug headers
Screw-terminal wiring ........................ 16 to 28 AWG
Screw-terminal torque........................ 0.22 to 0.25 N•m
(2.0 to 2.2 lb•in.)
Safety
Power Requirement
USB (4.10 to 5.25 VDC)....................... 80 mA typical
500 mA maximum
USB suspend....................................... 300 µA typical
500 µA maximum
Environmental
The USB-6008 and USB-6009 are intended for indoor use only.
Operating environment
Ambient temperature range ........... 0 to 55 °C (tested in accordance
with IEC-60068-2-1
and IEC-60068-2-2)
Relative humidity range ................. 10 to 90%, noncondensing
(tested in accordance
with IEC-60068-2-56)
Storage environment
Ambient temperature range ........... -40 to 85 °C (tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2)
Relative humidity range ................. 5 to 90%, noncondensing
(tested in accordance
with IEC-60068-2-56)
Maximum altitude............................... 2,000 m
(at 25 °C ambient temperature)
Pollution degree.................................. 2
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA 61010-1
Note: For UL and other safety certifications, refer to the product label
or visit ni.com/certification, search by model number or product line,
and click the appropriate link in the Certification column.
Electromagnetic Compatibility
This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement, control,
and laboratory use:
• EN 61326 EMC requirements; Minimum Immunity
• EN 55011 Emissions; Group 1, Class A
• CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A
Note: For EMC compliance, operate this device according to
product documentation.
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE marking, as follows:
• 2006/95/EC; Low-Voltage Directive (safety)
• 2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Note: Refer to the Declaration of Conformity (DoC) for this product for
any additional regulatory compliance information. To obtain the DoC for
this product, visit ni.com/certification, search by model number or
product line, and click the appropriate link in the Certification column.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers: At the end of their life cycle, all products must be
sent to a WEEE recycling center. For more information about WEEE
recycling centers and National Instruments WEEE initiatives, visit
ni.com/environment/weee.htm.
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)
4
IN
TA
PL
AN
MA
IN
NI Services and Support
P
DE
LO
Y
EL
OP
SERVICE
NEEDS
V
DE
NI has the services and support to meet
your needs around the globe and through
the application life cycle – from planning
and development through deployment
and ongoing maintenance. We offer
services and service levels to meet
customer requirements in research,
design, validation, and manufacturing.
Visit ni.com/services.
Training and Certification
NI training is the fastest, most certain route to productivity with our
products. NI training can shorten your learning curve, save development
time, and reduce maintenance costs over the application life cycle. We
schedule instructor-led courses in cities worldwide, or we can hold a
course at your facility. We also offer a professional certification program
that identifies individuals who have high levels of skill and knowledge on
using NI products. Visit ni.com/training.
Professional Services
Our Professional Services Team is comprised of NI applications engineers,
NI Consulting Services, and a worldwide National Instruments Alliance
Partner program of more than 600 independent consultants and
integrators. Services range from
start-up assistance to turnkey
system integration.
Visit ni.com/alliance.
OEM Support
We offer design-in consulting and product integration assistance if you
want to use our products for OEM applications. For information about
special pricing and services for OEM customers, visit ni.com/oem.
ni.com • (800) 813 3693
National Instruments • [email protected]
Local Sales and Technical Support
In offices worldwide, our staff is local to the country, giving you access
to engineers who speak your language. NI delivers industry-leading
technical support through online knowledge bases, our applications
engineers, and access to 14,000 measurement and automation
professionals within NI Developer Exchange forums. Find immediate
answers to your questions at ni.com/support.
We also offer service programs that provide automatic upgrades to
your application development environment and higher levels of technical
support. Visit ni.com/ssp.
Hardware Services
NI Factory Installation Services
NI Factory Installation Services (FIS) is the fastest and easiest way to
use your PXI or PXI/SCXI combination systems right out of the box.
Trained NI technicians install the software and hardware and configure
the system to your specifications. NI extends the standard warranty by
one year on hardware components (controllers, chassis, modules)
purchased with FIS. To use FIS, simply configure your system online with
ni.com/pxiadvisor.
Calibration Services
NI recognizes the need to maintain properly calibrated devices for
high-accuracy measurements. We provide manual calibration
procedures, services to recalibrate your products, and automated
calibration software specifically designed for use by metrology
laboratories. Visit ni.com/calibration.
Repair and Extended Warranty
NI provides complete repair services for our products. Express repair
and advance replacement services are also available. We offer
extended warranties to help you meet project life-cycle requirements.
Visit ni.com/services.
*351378C-01*
351378C-01
2008-9621-301-101-D
©2008 National Instruments. All rights reserved. CVI, LabVIEW, Measurement Studio, National Instruments, National Instruments Alliance Partner, NI, ni.com, SCXI, and SignalExpress
are trademarks of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Windows is a registered trademark of Microsoft Corporation in the
United States and other countries. Linux® is the registered trademark of Linus Torvalds in the U.S. and other countries. Other product and company names listed are trademarks or trade
names of their respective companies. A National Instruments Alliance Partner is a business entity independent from NI and has no agency, partnership, or joint-venture relationship with NI.