NI 778269-01

Low-Cost E Series Multifunction DAQ –
12 or 16-Bit, 200 kS/s, 16 Analog Inputs
NI E Series – Low-Cost
• 16 analog inputs at up to 200 kS/s,
12 or 16-bit resolution
• Up to 2 analog outputs at 10 kS/s,
12 or 16-bit resolution
• 8 digital I/O lines (TTL/CMOS);
two 24-bit counter/timers
• Digital triggering
• 4 analog input signal ranges
• NI-DAQ driver that simplifies
configuration and measurements
Operating Systems
Families
Other Compatible Software
•
•
•
•
•
• Visual Basic, C/C++, and C#
NI 6036E
NI 6034E
NI 6025E
NI 6024E
NI 6023E
Family
Bus
NI 6036E PCI, PCMCIA
NI 6034E
PCI
NI 6025E
PCI, PXI
NI 6024E PCI, PCMCIA
NI 6023E
PCI
110
• Windows 2000/NT/XP
• Real-time performance with LabVIEW
• Others such as Linux® and Mac OS X
Recommended Software
•
•
•
•
LabVIEW
LabWindows/CVI
Measurement Studio
VI Logger
Driver Software (included)
• NI-DAQ 7
Analog
Inputs
16 SE/8 DI
16 SE/8 DI
16 SE/8 DI
16 SE/8 DI
16 SE/8 DI
Input
Resolution
16 bits
16 bits
12 bits
12 bits
12 bits
Max
Sampling Rate
200 kS/s
200 kS/s
200 kS/s
200 kS/s
200 kS/s
Input Range
±0.05 to ±10 V
±0.05 to ±10 V
±0.05 to ±10 V
±0.05 to ±10 V
±0.05 to ±10 V
Analog
Outputs
2
0
2
2
0
Output
Resolution
16 bits
–
12 bits
12 bits
–
Output Rate Output Range Digital I/O Counter/Timers Triggers
±10 V
8
2, 24-bit
Digital
10 kS/s1
–
–
8
2, 24-bit
Digital
±10 V
8
2, 24-bit
Digital
10 kS/s1
±10 V
8
2, 24-bit
Digital
10 kS/s1
–
–
8
2, 24-bit
Digital
kS/s typical when using the single DMA channel for analog output. 1 kS/s maximum when using the single DMA channel for either analog input or counter/timer operations. 1 kS/s maximum for PCMCIA DAQCard devices in all cases.
Table 1. Low-Cost E Series Model Guide
Overview and Applications
National Instruments low-cost E Series multifunction data acquisition
devices provide full functionality at a price to meet the needs of the
budget-conscious user. They are ideal for applications ranging from
continuous high-speed data logging to control applications to high-voltage
signal or sensor measurements when used with NI signal conditioning.
Synchronize the operations of multiple devices using the RTSI bus or PXI
trigger bus to easily integrate other hardware such as motion control and
machine vision to create an entire measurement and control system.
Highly Accurate Hardware Design
NI low-cost E Series DAQ devices include the following features
and technologies:
Temperature Drift Protection Circuitry – Designed with components
that minimize the effect of temperature changes on measurements to
less than 0.0010% of reading/°C.
Resolution-Improvement Technologies – Carefully designed noise
floor maximizes the resolution.
Onboard Self-Calibration – Precise voltage reference included for
calibration and measurement accuracy. Self-calibration is completely
software controlled, with no potentiometers to adjust.
NI DAQ-STC – Timing and control ASIC designed to provide more
flexibility, lower power consumption, and a higher immunity to noise
and jitter than off-the-shelf counter/timer chips.
NI MITE – ASIC designed to optimize data transfer for multiple
simultaneous operations using bus mastering with one DMA channel,
interrupts, or programmed I/O.
NI PGIA – Measurement and instrument class amplifier that guarantees
settling times at all gains. Typical commercial off-the-shelf amplifier
components do not meet the settling time requirements for high-gain
measurement applications.
PFI Lines – Eight programmable function input (PFI) lines that you can
use for software-controlled routing of interboard and intraboard digital
and timing signals.
RTSI or PXI Trigger Bus – Bus used to share timing and control signals
between two or more PCI or PXI devices to synchronize operations.
RSE Mode – In addition to differential and nonreferenced single-ended
modes, NI low-cost E Series devices offer the referenced single-ended
(RSE) mode for use with floating-signal sources in applications with
channel counts higher than eight.
Onboard Temperature Sensor – Included for monitoring the operating
temperature of the device to ensure that it is operating within the
specified range.
Low-Cost E Series Multifunction DAQ – 12 or 16-Bit, 200 kS/s, 16 Analog Inputs
Models
Measurement Sensitivity1 (mV)
Nominal Range (V)
Positive FS
Negative FS
10
-10
5
-5
2.5
-2.5
2
-2
1
-1
0.5
-0.5
0.25
-0.25
0.2
-0.2
0.1
-0.1
0.05
-0.05
10
0
5
0
2
0
1
0
0.5
0
0.2
0
0.1
0
NI 6030E, NI 6031E,
NI 6032E, NI 6033E
0.0023
Full-Featured E Series
NI 6052E
NI 6070E, NI 6071E
0.0025
0.009
4.747
0.876
1.190
–
0.479
0.243
0.137
–
0.064
0.035
1.232
2.119
0.850
0.428
0.242
0.111
0.059
14.369
5.193
3.605
–
1.452
0.735
0.379
–
0.163
0.091
6.765
5.391
2.167
1.092
0.558
0.235
0.127
NI 6040E
0.008
Low-Cost E Series
NI 6034E, NI 6036E
NI 6023E, NI 6024E,
NI 6025E
0.0036
0.008
Basic
PCI-6013, PCI-6014
0.004
Absolute Accuracy (mV)
1.147
2.077
–
0.836
0.422
0.215
–
0.102
0.061
–
0.976
1.992
0.802
0.405
0.207
0.098
0.059
15.373
5.697
3.859
–
1.556
0.789
0.405
–
0.176
0.100
7.269
5.645
2.271
1.146
0.583
0.247
0.135
7.560
1.790
–
–
–
0.399
–
–
–
0.0611
–
–
–
–
–
–
–
16.504
5.263
–
–
–
0.846
–
–
–
0.106
–
–
–
–
–
–
–
8.984
2.003
–
–
–
0.471
–
–
–
0.069
–
–
–
–
–
–
–
Note: Accuracies are valid for measurements following an internal calibration. Measurement accuracies are listed for operational temperatures within ±1 °C of internal calibration temperature and ±10 °C of external or factorycalibration temperature. One-year calibration interval recommended. The Absolute Accuracy at Full Scale calculations were performed for a maximum range input voltage (for example, 10 V for the ±10 V range) after one year,
assuming 100 pt averaging of data.
1Smallest detectable voltage change in the input signal at the smallest input range.
Table 2. E Series Analog Input Absolute Accuracy Specifications
Models
Nominal Range (V)
Positive FS
Negative FS
10
-10
10
0
NI 6030E, NI 6031E,
NI 6032E, NI 6033E
Full-Featured E Series
NI 6052E
NI 6070E, NI 6071E
NI 6040E
Low-Cost E Series
NI 6034E, NI 6036E
NI 6023E, NI 6024E,
NI 6025E
Basic
PCI-6013, PCI-6014
Absolute Accuracy (mV)
1.430
1.201
1.405
1.176
8.127
5.685
8.127
5.685
2.417
–
8.127
–
3.835
–
Table 3. E Series Analog Output Absolute Accuracy Specifications
High-Performance,
Easy-to-Use Driver Software
NI-DAQ is the robust driver software that makes it easy to access the
functionality of your data acquisition hardware, whether you are a
beginning or advanced user. Helpful features include:
Automatic Code Generation – DAQ Assistant is an interactive
guide that steps you through configuring, testing, and programming
measurement tasks and generates the necessary code automatically
for NI LabVIEW, LabWindows/CVI, or Measurement Studio.
Cleaner Code Development – Basic and advanced software functions
have been combined into one easy-to-use yet powerful set to help you
build cleaner code and move from basic to advanced applications
without replacing functions.
High-Performance Driver Engine – Software-timed single-point input
(typically used in control loops) with NI-DAQ achieves rates of up to 50 kHz.
NI-DAQ also delivers maximum I/O system throughput with a
multithreaded driver.
Test Panels – With NI-DAQ, you can test all of your device functionality
before you begin development.
Scaled Channels – Easily scale your voltage data into the proper
engineering units using the NI-DAQ Measurement Ready virtual
channels by choosing from a list of common sensors and signals or
creating your own custom scale.
LabVIEW Integration – All NI-DAQ functions create the waveform data
type, which carries acquired data and timing information directly into
more than 400 LabVIEW built-in analysis routines for display of results in
engineering units on a graph.
For information on applicable hardware for NI-DAQ 7,
visit ni.com/dataacquisition.
Visit ni.com/oem for quantity discount information.
BUY ONLINE at ni.com or CALL (800) 813 3693 (U.S.)
2
Low-Cost E Series Multifunction DAQ – 12 or 16-Bit, 200 kS/s, 16 Analog Inputs
Recommended Accessories
Signal conditioning is required for sensor measurements or voltage
inputs greater than 10 V. National Instruments SCXI is a versatile,
high-performance signal conditioning platform, intended for
high-channel-count applications. NI SCC products provide portable,
flexible signal conditioning options on a per-channel basis. Both signal
conditioning platforms are designed to increase the performance and
reliability of your DAQ system, and are up to 10 times more accurate
than terminal blocks (please visit ni.com/sigcon for more details).
Refer to the table below for more information:
Sensor/Signals (>10 V)
System Description
DAQ Device
High-performance
PCI-60xxE, PXI-60xxE, DAQCard-60xxE
Low-cost, portable
PCI-60xxE, PXI-60xxE, DAQCard-60xxE
Signal Conditioning
SCXI
SCC
Signals (<10 V)1
System Description
Shielded
Shielded
Shielded
Low-cost
Low-cost
Low-cost
DAQ Device
PCI-60xxE
PXI-60xxE
DAQCard-60xxE
PCI-6025E/PXI-6025E
PCI-60xxE/PXI-60xxE
DAQCard-60xxE
Terminal Block
SCB-68
TB-2705
SCB-68
Two TBX-68s
CB-68LP
CB-68LP
Cable
SH6868-EP
SH6868-EP
SHC6868-EP
SH1006868
R6868
RC6868
1Terminal blocks do not provide signal conditioning (i.e., filtering, amplification, isolation, and so on),
which may be necessary to increase the accuracy of your measurements.
Table 4. Recommended Accessories
Ordering Information
PCI
NI PCI-6036E..........................................................................778465-01
NI PCI-6034E..........................................................................778075-01
NI PCI-6025E..........................................................................777744-01
NI PCI-6024E..........................................................................777743-01
NI PCI-6023E..........................................................................777742-01
PCMCIA
NI DAQCard-6036E ................................................................778561-01
NI DAQCard-6024E ................................................................778269-01
PXI
NI PXI-6025E..........................................................................777798-01
Includes NI-DAQ driver software.
BUY NOW!
For complete product specifications, pricing, and accessory information,
call (800) 813 3693 (U.S.) or go to ni.com/dataacquisition.
BUY ONLINE at ni.com or CALL (800) 813 3693 (U.S.)
3
NI Services and Support
SERVICE
NEEDS
NI has the services and support to meet
your needs around the globe and through
the application life cycle – from planning
and development through deployment
and ongoing maintenance. We offer
services and service levels to meet
customer requirements in research,
design, validation, and manufacturing.
Visit ni.com/services.
Training and Certification
NI training is the fastest, most certain route to productivity with our
products. NI training can shorten your learning curve, save development
time, and reduce maintenance costs over the application life cycle. We
schedule instructor-led courses in cities worldwide, or we can hold a
course at your facility. We also offer a professional certification program
that identifies individuals who have high levels of skill and knowledge on
using NI products. Visit ni.com/training.
Professional Services
Our Professional Services Team is comprised of NI applications engineers,
NI Consulting Services, and a worldwide National Instruments Alliance
Partner program of more than 600 independent consultants and
integrators. Services range from
start-up assistance to turnkey
system integration.
Visit ni.com/alliance.
OEM Support
We offer design-in consulting and product integration assistance if you
want to use our products for OEM applications. For information about
special pricing and services for OEM customers, visit ni.com/oem.
ni.com • (800) 813 3693
National Instruments • [email protected]
Local Sales and Technical Support
In offices worldwide, our staff is local to the country, giving you access
to engineers who speak your language. NI delivers industry-leading
technical support through online knowledge bases, our applications
engineers, and access to 14,000 measurement and automation
professionals within NI Developer Exchange forums. Find immediate
answers to your questions at ni.com/support.
We also offer service programs that provide automatic upgrades to
your application development environment and higher levels of technical
support. Visit ni.com/ssp.
Hardware Services
NI Factory Installation Services
NI Factory Installation Services (FIS) is the fastest and easiest way to
use your PXI or PXI/SCXI combination systems right out of the box.
Trained NI technicians install the software and hardware and configure
the system to your specifications. NI extends the standard warranty by
one year on hardware components (controllers, chassis, modules)
purchased with FIS. To use FIS, simply configure your system online with
ni.com/pxiadvisor.
Calibration Services
NI recognizes the need to maintain properly calibrated devices for
high-accuracy measurements. We provide manual calibration
procedures, services to recalibrate your products, and automated
calibration software specifically designed for use by metrology
laboratories. Visit ni.com/calibration.
Repair and Extended Warranty
NI provides complete repair services for our products. Express repair
and advance replacement services are also available. We offer
extended warranties to help you meet project life-cycle requirements.
Visit ni.com/services.
*342570A-01*
342570A-01
© 2006 National Instruments Corporation. All rights reserved. CVI, DAQCard, LabVIEW, Measurement Ready, Measurement Studio, MITE, National Instruments,
National Instruments Alliance Partner, NI, ni.com, RTSI, and SCXI are trademarks of National Instruments. Linux ® is the registered trademark of Linus Torvalds
in the U.S. and other countries. Other product and company names listed are trademarks or trade names of their respective companies.
A National Instruments Alliance Partner is a business entity independent from NI and has no agency, partnership, or joint-venture relationship with NI.
2006-6457-301-101-D