DAQ for High-Voltage Measurements 16-Bit, 200 kS/s, 8 Analog Inputs New NI PXI-4204 • 8 analog inputs at 200 kS/s, 16-bit resolution • ±100 V input range • Programmable gain (0.5, 1, 10, 100) per channel • Programmable 2-pole Butterworth filter (6 Hz or 10 kHz) per channel • Differential simultaneous sampling inputs • Combicon screw terminal direct connectivity • NI-DAQ driver simplifies configuration and measurements Device NI PXI-4204 Bus PXI Operating Systems • Windows 2000/NT/XP Recommended Software • LabVIEW • LabWindows/CVI • Measurement Studio for Visual Studio.NET Other Compatible Software • Visual Basic, C/C++, and C# Driver Software (Included) • NI-DAQ 7 Analog Inputs 8 DI Resolution 16 bits Sampling Rate 200 kS/s Input Range ±0.5 to ±100 V Triggers Digital (2) Filter Settings 2-pole Butterworth (6 Hz or 10 kHz) Table 1. NI PXI-4204 Channel, Speed, and Resolution Specifications Driver Software Overview and Applications The National Instruments PXI-4204 module delivers accurate data acquisition for input ranges up to 100 V. With this module you can handle a broad variety of applications including: • 42 V automotive applications • High-voltage, multichannel data logging applications • Fuel cell and battery test applications Features The NI PXI-4204 is a full-featured data acquisition module with a ±100 V input range, 16-bit accuracy, and software-selectable filter and gain settings per channel. Programmable filter and gain settings ensure that the PXI-4204 achieves maximum accuracy over the entire ±100 V input range. In addition, the PXI-4204 is designed to work with LabVIEW 7 Express and NI-DAQmx. The DAQ Assistant in LabVIEW 7 can configure the PXI-4204 and acquire data through a menu-based window, eliminating the need to manually program the device. For high-channel-count, high-voltage applications, explore NI signal conditioning hardware. Visit ni.com/sigcon NI-DAQ 7 is the robust driver software included with all National Instruments data acquisition and signal conditioning products. This easy-to-use software tightly integrates the full functionality of your DAQ hardware to LabVIEW, LabWindows/CVI, and Measurement Studio for Visual Studio.NET. High-performance features include multidevice synchronization, networked measurements, and DMA data management. Bundled with NI-DAQ 7, the Measurement & Automation Explorer (MAX) utility simplifies the configuration of your measurement hardware with device test panels, interactive measurements, and scaled I/O channels. NI-DAQ also provides numerous example programs for LabVIEW and other application development environments to get you started with your application quickly. Ordering Information NI PXI-4204 ......................................................778745-4204 Includes NI-DAQ driver software. For more information on extended warranty and valueadded services, visit ni.com/services DAQ for High-Voltage Measurements 16-Bit, 200 kS/s, 8 Analog Inputs Specifications – PXI-4204 Complete Accuracy Tables Input Characteristics SSH Disabled Nominal Range (V) Positive FS Negative FS 100 -100 50 -50 5 -5 0.5 -0.5 % of Reading 24 Hrs 1 Year 0.063 0.070 0.031 0.038 0.061 0.068 0.061 0.068 Offset (mV) ± 16.8 ± 10.4 ± 4.7 ± 4.1 Absolute Accuracy Noise + Quantization (mV) Single Pt. Averaged ± 9.34 ± 0.92 ± 4.68 ± 0.57 ± 0.69 ± 0.40 ± 0.49 ± 0.40 Temp Drift (%/°C) 0.0027 0.0025 0.0027 0.0027 Absolute Accuracy at Full Scale (mV) 88 30 8.5 4.8 Percent Error at Full Scale (%) 0.088 0.060 0.17 0.97 SSH Enabled Nominal Range (V) Positive FS Negative FS 100 -100 50 -50 5 -5 0.5 -0.5 % of Reading 24 Hrs 1 Year 0.085 0.090 0.035 0.042 0.063 0.070 0.063 0.070 Offset (mV) ± 19.0 ± 12.6 ± 6.8 ± 6.3 Absolute Accuracy Noise + Quantization (mV) Single Pt. Averaged ± 9.35 ± 1.02 ± 4.70 ± 0.73 ± 0.82 ± 0.60 ± 0.66 ± 0.60 Temp Drift (%/°C) 0.0027 0.0025 0.0027 0.0027 Absolute Accuracy at Full Scale (mV) 110 34 11 7.2 Percent Error at Full Scale (%) 0.110 0.069 0.22 1.4 Note: Accuracies are valid for measurements following an internal calibration. Averaged numbers assume dithering and averaging of 100 single-channel readings. Measurement accuracies are listed for operational temperatures within ±1°C of internal calibration temperature and ±10°C of external or factory calibration temperature. Input Frequency Ranges DC to 60 Hz 100 Hz to 10 kHz CMRR >60 dB >40 dB Filter Type .................................................. 6 Hz or 10 kHz, 2-pole, Butterworth Stability 8 differential Module PXI-4204 Input signal ranges..................................... see accuracy tables Resolution .................................................. 16 bits, 1 in 65,536 Input Range ± 100 V Gain Temperature Coefficient ± 25 ppm/Co Offset Temperature Coefficient ± 385 uV/Co Triggers Sampling Rate Maximum Sampling Rate .......................... 200 kS/s aggregate multi-channel Module PXI-4204 Module PXI-4204 PXI-4204 Filter Characteristics Number of Channels PXI-4204 CMRR Max per channel (scanning all channels) Scan Rate w/SSH* Scan Rate w/o SSH* 20 kS/s/ch 25 kS/s/ch Single Channel 333 kS/s *SSH = Simultaneous Sample and Hold Connector Screw Terminals SMB Connector Powered On ± 100 VDC ± 15 V Powered Off ± 100 VDC ± 15 V Overvoltage protection Inputs with overvoltage protection ............ CH<0..7> Input coupling ............................................ DC Data transfers ............................................ DMA, interrupts, programmed I/O Digital Triggers Number of triggers ............................... 2 Purpose ................................................ Start & stop trigger, gate, clock Source .................................................. PFI0/TRIG1 (front SMB connector), PXI_TRIG_0…PXI_TRIG_6 (PXI trigger bus) Slope .......................................................... Positive or negative; software selectable Compatibility......................................... 5V/TTL Response.............................................. Rising or falling edge Pulse width........................................... 10 ns minimum Impedance............................................ 10 kOhm Coupling................................................ DC PXI Trigger Bus Trigger lines .......................................... 6 Star trigger............................................ 1 Transfer Characteristics Nonlinearity Range ± 100 V Other Ranges ± 100 V, SSH Enabled Other Ranges, SSH Enabled Percent of Full Scale (Typical) 0.02% 0.01% 0.06% 0.02% Bus Interface PXI……….……………………….Master, slave Power Requirements Module PXI-4204 5 VDC (±5%) 1A DNL Module PXI-4204 Typical ± 0.5 LSB Maximum ± 1 LSB Gain Error ................................................... See Accuracy Table Offset Error ................................................ See Accuracy Table** Amplifier Characteristics Input Impedance Module PXI-4204 NMR (60 Hz) PXI-4204 Input Configuration Differential Input Single-Ended Normal Powered On 2 MΩ 1 MΩ 40 dB @ 6 Hz filter setting Powered Off 2MΩ 1MΩ Calibration Recommended warm-up time ............. External calibration interval................... Onboard calibration reference Level................................................ Temperature coefficient .................. Long-term stability .......................... 15 minutes 1 year 5.000V +/- 1mV + 0.6 ppm/oC 6 ppm/Sqrt(1000h) Physical Dimensions (not including connectors).......................................... 16.0 by 10.0 cm by 2.0 cm (6.3 x 3.9 x 0.79 in) Analog input signal connector.................. 16 x 1 minicombicon, 3.81 mm pitch Analog input signal mating connector ..... 16 x 1 minicombicon, 3.81 mm pitch, 28-16 AWG signal wire **Gain Error = Actual Input Voltage * % of Reading 2 National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • [email protected] • ni.com DAQ for High-Voltage Measurements 16-Bit, 200 kS/s, 8 Analog Inputs Specifications I/O Connectors Module PXI-4204 Environment: Operating temperature ......................... Storage temperature ............................ Relative humidity .................................. Maximum altitude ................................ Pollution degree (indoor use only)........ Front 16 screw terminals, 1 SMB 0 to 55 oC –20 to 70 oC 10 to 90%, non-condensing 2,000m 2 Certification and Compliance European Compliance Emissions................................................... EN 55011 Class A 10 meters Immunity.................................................... EN 61326:1997 + A2:2001, Table 1 EMC ........................................................... CE, C-Tick and FCC Part 15 (Class A) Compliant Safety......................................................... IEC 61010-1, EN 61010-1 North American Compliance Emissions................................................... FCC Part 15A above 1 GHz Immunity.................................................... EN 61326:1997 + A2:2001, Table 1 EMC ........................................................... CE, C-Tick and FCC Part 15 (Class A) Compliant Safety......................................................... IEC 61010-1, UL 3111-1, UL 61010B-1, CAN/SA C22.2 No. 1010.1 National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • [email protected] • ni.com 3 Global Services and Support NI has the services and support to meet your needs around the globe and through the application life cycle – from planning and development through deployment and ongoing maintenance – and they are tailored for customer requirements in research, design, validation, and manufacturing. We have direct operations in more than 37 countries and distributors in another 12 locations. Our local sales and support representatives are degreed engineers, ready to partner with you to find solutions that best fit your needs. Local Sales and Technical Support P LO DE P Training and Certification DEV E Y LO AN MA IN In offices around the globe, our staff is local to the country so that you have access to field engineers who speak your language and are available to consult on your unique needs. We also have a worldwide support organization staffed with Applications Engineers trained to quickly provide superior technical assistance. Use our online Request Support interface (ni.com/support) to define your question, then speak to or e-mail an Applications Engineer, or access more than 14,000 worldwide measurement and automation professionals within NI Developer Exchange Discussion Forums. ni.com/support also provides immediate answers to your questions through self-help troubleshooting, product reference, and application development resources. For advanced technical support and software maintenance services, sign up for IN PL Premier Support, a program that TA provides expanded hours of support availability and expedited phone and e-mail response time (typically four business hours). NI recognizes that both initial instruction and ongoing education contribute to your success. NI provides a variety of training alternatives, from self-paced tutorials and interactive CDs, to worldwide hands-on courses taught by experienced instructors – all designed so that you can choose how to learn about our products. Further, NI offers certifications acknowledging individual expertise in working with NI products and technologies. Visit ni.com/training for more information. ni.com Professional Services Our Professional Services team consists of National Instruments Applications Engineers, NI Consulting Services, and a worldwide Alliance Program (a network of 600 independent consultants and integrators). Our Professional Services team can offer services ranging from basic start-up assistance and collaborative development with your engineers, to turnkey system integration and maintenance of your system. Alliance Member In addition to our NI Alliance Partners, we have developed global relationships with many industry partners that range from computer software and hardware companies, such as Microsoft, Dell, Siemens, and Tektronix. By collaborating with these companies, you receive a complete spectrum of solutions – from components to turnkey systems. Find the Alliance Program member directory at ni.com/alliance Product Services NI GPIB products are warranted against defects in workmanship and material for one year from the date of shipment. To help you meet project lifecycle requirements, NI offers extended warranties for an additional charge. NI provides complete repair services for our products. Express repair and advanced replacement services are also available. Or, order your software and hardware installed in PXI and PXI/SCXI™ systems with NI Factory Installation Services. Ordering Made Easy Visit ni.com/products to browse product specifications, make comparisons, or access technical representatives via online chat or telephone. Worldwide customers can use a purchase order or credit card to buy in local currency and receive direct shipments from local NI offices. Our North American Customer Service Representatives are available Monday through Friday between 7 a.m. and 7 p.m. Central Time. Outside North America, please contact the NI office in your country. Order Status and Service Requests National Instruments brings you real-time status on current orders at ni.com/status Similarly, find out the status of open technical support incidents or hardware repair requests at ni.com/support/servicereq (800) 433-3488 National Instruments • Tel: (512) 683-0100 • Fax: (512) 683-9300 • [email protected] C This document represents a commitment from National Instruments to the environment. © 2003 National Instruments Corporation. All rights reserved. Product and company names listed are trademarks or trade names of their respective companies.