SENSORTECHNICS RPUM100G2A

RPU Series
Basic pressure sensors
FEATURES
· 0...35 mbar to 0...16 bar and
0...0.5 to 0...250 psi gage or differential
· High impedance bridge
· Miniature package
· Different pinning configurations
· Usable for wet/wet applications8
· Sensortechnics PRO services
MEDIA COMPATIBILITY
All media compatible with
port 1:
- polyetherimide
- silver-filled silicone
- silicon nitride
port 29:
- polyetherimide
- fluor-silicone
- silicon
ELECTRICAL CONNECTION
SPECIFICATIONS
Maximum ratings
Supply voltage
Temperature limits
Storage
Operating
12 V
+Vs
1
-55 to +100°C
-40 to +85°C
4
Lead temperature (10 sec. soldering)
Humidity limits
Vibration (0 to 2000 Hz)
(qualification tested)
Mechanical shock
(qualification tested)
E / 11702 / A
-
300°C
Output
0...100 %RH
2
20 g sine
+
3
150 g
1/5
www.sensortechnics.com
RPU Series
Basic pressure sensors
PRESSURE SENSOR CHARACTERISTICS
(Vs = 10.0 ± 0.01 V, tamb = 25 °C)
Part number
Operating pressure
Proof pressure1
RPUM035...
RPUM050...
RPUM100...
RPUM250...
RPUB001...
RPUB002...
RPUB005...
RPUB010...
RPUB016...
RPUP0x5...
RPUP001...
RPUP005...
RPUP015...
RPUP030...
RPUP100...
RPUP250...
0...35 mbar
0...50 mbar
0...100 mbar
0...250 mbar
0...1 bar
0...2 bar
0...5 bar
0...10 bar
0...16 bar
0...0.5 psi
0...1 psi
0...5 psi
0...15 psi
0...30 psi
0...100 psi
0...250 psi
1.4 bar
1.4 bar
1.4 bar
1.4 bar
3 bar
4 bar
12 bar
35 bar
35 bar
2 0 p si
2 0 p si
2 0 p si
4 5 p si
6 0 p si
2 0 0 p si
5 0 0 p si
Full-scale span2
Min.
Typ.
Max.
24 mV
21 mV
44 mV
61 mV
159 mV
232 mV
113 mV
84 mV
134 mV
24 mV
30 mV
85 mV
165 mV
240 mV
156 mV
145 mV
35 mV
33 mV
66 mV
83 mV
218 mV
320 mV
163 mV
123 mV
197 mV
35 mV
45 mV
115 mV
225 mV
330 mV
225 mV
212 mV
46 mV
44 mV
88 mV
105 mV
276 mV
410 mV
215 mV
163 mV
260 mV
46 mV
60 mV
145 mV
285 mV
420 mV
294 mV
280 mV
Typ.
Max.
Unit
PERFORMANCE CHARACTERISTICS
(Vs = 10.0 ± 0.01 V, tamb = 25 °C)
Characteristics
Zero pressure offset
Temperature effects (0 to 50 °C)4
Temperature effects on bridge impedance4
Non-linearity (BSL)3
Repeatability and hysteresis5
Long term stability7
Input impedance
Output impedance
Response time6
E / 11702 / A
Min.
-20
Offset
Span
4.0
4.0
+20
±2.0
-2000
+2200
±0.25
±0.15
±0.5
5.0
5.0
mV
ppm/°C
±1.0
% sp a n
6.0
6.0
1.0
kΩ
ms
2/5
www.sensortechnics.com
RPU Series
Basic pressure sensors
OUTLINE DRAWINGS10
RPU...G6A (single inline pinning, 1 x 4), gage pressure devices
12,7
Ø 6,4
7,9
Ø 5,1
Ø 1,93
2,29
4 x 1,02
3 x 2,54
P2
1,78
4 x 0,41
16,2
9,1
21,8
NOM 15,2
PIN 1
Ø 1,27
third angle projection
4 3 2 1
mass: 2 g
dimensions in mm
12,7
RPU...G2A (dual inline pinning, 2 x 2), gage pressure devices
12,7
PIN 3
PIN 4
Ø 6,4
7,9
Ø 5,1
Ø 1,93
PIN 1
PIN 2
P2
16,2
9,1
21,8
27,9
Ø 1,27
4 x 0,5
4 x 1,02
5,1
mass: 2 g
E / 11702 / A
12,7
4 x 0,41
8,4
third angle projection
dimensions in mm
3/5
www.sensortechnics.com
RPU Series
Basic pressure sensors
OUTLINE DRAWINGS10
RPU...D6A (single inline pinning, 1 x 4), differential pressure devices
12,7
Ø 6,4
Ø 1,93
PIN 1
Ø 1,27
3 x 2,54
NOM 15,2
7,9
Ø 5,1
2,29
P2
1,78
4 x 0,41
16,2
9,1
34,4
21,8
4 x 1,02
11,7
4 3 2 1
P1
mass: 2 g
12,7
third angle projection
dimensions in mm
Specification notes:
1. The maximum specified pressure which may be applied to the sensor without causing a permanent change in the output
characteristics.
2. Span is the algebraic difference between the output voltage at full-scale pressure and the output at zero pressure. Span is
ratiometric to the supply voltage.
3. Non-linearity refers to the Best Straight Line fit, measured for offset pressure, full scale pressure and 1/2 full scale pressure.
4. Error band of the offset voltage, span or bridge impedance in the specified temperature range, relative to the 25°C reading.
5. Repeatability, the deviation in output readings for successive application of any given input pressure (all other conditions
remaining constant). Hysteresis, the error defined by the deviation in output signal obtained when a specific pressure point is
approached first with increasing pressure, then with decreasing pressure or vice versa (all other conditions remaining constant).
6. Response time for 0 to full-scale pressure step change, readings taken at 10 % and 90 % of full-scale pressure.
7. Long term stability of offset and span over a period over one year.
8. The sensors might be used on both ports for media compatible with the components specified under "Media Compatibility".
9. Other sealing materials are available on request. Minimum order quantities might be required.
10. Other pressure port styles, like barbed ones, luers, modular, M5, needle style or flow through connection, are
available on request. Minimum order quantities might be required.
E / 11702 / A
4/5
www.sensortechnics.com
RPU Series
Basic pressure sensors
ORDERING INFORMATION - AVAILABLE LISTINGS
Note:
Preferred listings are highlighted in grey
Pressure range
Gage pressure
Differential pressure
Single inline pinning
Dual inline pinning
Single inline pinning
35 mbar
RPUM035G6A
-
RPUM035D6A
50 mbar
RPUM050G6A
RPUM050G2A
RPUM050D6A
100 mbar
RPUM100G6A
RPUM100G2A
RPUM100D6A
250 mbar
RPUM250G6A
RPUM250G2A
RPUM250D6A
1 bar
RPUB001G6A
RPUB001G2A
RPUB001D6A
2 bar
RPUB002G6A
RPUB002G2A
RPUB002D6A
5 bar
RPUB005G6A
RPUB005G2A
RPUB005D6A
10 bar
RPUB010G6A
RPUB010G2A
RPUB010D6A
16 bar
RPUB016G6A
RPUB016G2A
RPUB016D6A
0.5 psi
RPUP0x5G6A
-
RPUP0x5D6A
1 p si
RPUP001G6A
RPUP001G2A
RPUP001D6A
5 p si
RPUP005G6A
RPUP005G2A
RPUP005D6A
1 5 p si
RPUP015G6A
RPUP015G2A
RPUP015D6A
3 0 p si
RPUP030G6A
RPUP030G2A
RPUP030D6A
6 0 p si
-
-
-
1 0 0 p si
RPUP100G6A
RPUP100G2A
RPUP100D6A
2 5 0 p si
RPUP250G6A
RPUP250G2A
RPUP250D6A
Sensortechnics PRO services:
· Extended guarantee period of 2 years
· Improved performance characteristics
· Custom product modifications and adaptations even for small quantities
· Advanced logistics models for supply inventory and short delivery times
· Technical support through application engineers on the phone or at your site
· Fastest possible technical response for design and QA engineers
... plus other services on request
Sensortechnics reserves the right to make changes to any products herein. Sensortechnics does not assume any liability arising out of the
application or use of any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others.
E / 11702 / A
5/5
www.sensortechnics.com