Designed for front-illuminated CCD area image sensor
C7027 is a multichannel detector head for use with front-illuminated FFT-CCD area image sensors (S7017/S7847 series) that offer high quantum
efficiency. C7027 incorporates a low-noise driver/amplifier circuit that provides reliable operation from simple external signals. C7027 includes a
highly stable temperature controller that cools the sensor to a preset temperature level (Ts= -50 ˚C) as soon as the power is turned on. If the
cooler fails and causes internal circuitry to overheat, the built-in protection circuit automatically turns off the power to the thermoelectric cooler.
Despite its compact size, the housing configuration is designed for good heat dissipation, and threaded mounting holes on the front panel allow
connections to other devices such as monochrometers.
Features
Applications
l Designed for front-illuminated CCD area image sensor
l Line binning operation/area scanning operation *1
l Driver/amplifier circuit for low noise CCD operation
l Highly stable temperature controller
Cooling temperature: -50 ± 0.05 ˚C
l Simple signal input operation
l Compact configuration
l Fluorescence spectroscopy
l Raman spectroscopy
l Other low-light-level detection
The table below shows CCD area image sensors applicable for C7027
Since C7027 does not include a CCD area image sensor, so select the desired sensor and order it separately.
CCD area image sensor
Type No.
Number of pixels
Number of active pixels
S7017-1007
S7017-1008
S7847-1007 (High IR sensitivity type)
S7847-1008 (High IR sensitivity type)
1044 × 128
1044 × 256
1044 × 128
1044 × 256
1024 × 124
1024 × 252
1024 × 124
1024 × 252
Active area
[mm (H) × mm (V)]
24.576 × 2.976
24.576 × 6.048
24.576 × 2.976
24.576 × 6.048
*1: The FFT-CCD was originally designed as a two-dimensional image sensor. However, it can be operated like a linear image sensor
having a large active area by transferring all the pixel signals in the vertical direction to the horizontal register (this is referred to as
line binning).
Note: C7027 can be operated from C7557 multichannel detector head controller, but cannot be cooled below -50 °C due to cooling.
capacity limitation.
1
CCD multichannel detector head
C7027
■ Absolute maximum ratings
Parameter
Supply voltage (for digital circuitry)
Symbol
Min.
Typ.
Max.
VD1
-0.5
+7
VA1+
+18
Supply voltage (for analog circuitry)
VA1-18
VA2
+30
VD2
+7.25
Supply voltage
Vp
+7.25
+14
VF
Digital input voltage
VD (1, 2)
+30 *2
Operating temperature
Topr
+10
Storage temperature
Tstg
0
+50
*2: The maximum value may be lower, depending on the drive frequency and the number of sensor pixels.
Unit
V
°C
■ Electrical characteristics
(Ta=25 °C, VD1=+5 V, VA1+=+15 V, VA1-= -15 V, VA2=+24 V, VD2=+7 V, Vp=+7 V, VF=+12 V, unless otherwise noted)
Parameter
High level
Low level
Digital input
CLK frequency
Data video readout frequency
Start pulse width
High level (Io= -6 mA)
Low level (Io=+6 mA)
Digital output
Symbol
VIH
VIL
fCLK
fv
tst
VIH
VIL
Min.
+2.0
-0.5
1/fCLK
+2.0
-
Typ.
-
Max.
+5.0
+0.8
1
fCLK/4
+0.8
Unit
V
V
MHz
Hz
s
V
V
VD1
VA1+
VA1VA2
VD2
Vp
VF
-
+4.75
+14.5
-14.5
+23.5
+6.75
+6.75
+11.75
-
+5.0
+15.0
-15.0
+24.0
+7.0
+7.0
+12.0
+2.0
+100
+5.25
+15.5
-15.5
+24.5
+7.25
+7.25
+12.75
+200
+100
-100
+30
+30
+4.0
-
V
V
V
V
V
V
V
mA
mA
mA
mA
mA
A
mA
Power supply operating conditions
Digital circuitry
Analog circuitry
Voltage
Other
VD1 (+5 VDC)
VA1+ (+15 VDC)
VA1- (-15 VDC)
VA2 (+24 VDC)
VD2 (+7 VDC)
Vp (+7 VDC)
Ta=25 °C
VF (+12 VDC)
Current
■ Electrical and optical characteristics
(Ta=25 °C, Ts= -50 °C, VD1=+5 V, VA1+=+15 V, VA1-= -15 V, VA2=+24 V, VD2=+7 V, Vp=+7 V, VF=+12 V)
Parameter
Spectral response range
Full well capacity
Vertical
Horizontal
Conversion gain *3
Dark current
Min.
λ
-
Fw
Sv
S7017
S7847
Readout noise
Dynamic range
Photo response non-uniformity *4
*3: Including the circuit gain.
*4: Measured at 50 % of the full well capacity.
2
Symbol
DS
Nr
DR
PRNU
150000
300000
-
Typ.
200 to
1100
300000
600000
15
0.03
0.3
15
75000
-
Max.
Unit
-
nm
0.1
1
±10
e
-
µV/e
-
-
e /pixel/s
-
e rms
%
CCD multichannel detector head
C7027
■ Specifications for temperature controller
(Ta=25 °C, Ts= -50 °C, VD1=+5 V, VA1+=+15 V, VA1-= -15 V, VA2=+24 V, VD2=+7 V, Vp=+7 V, VF=+12 V)
Parameter *5
Symbol
Min.
Typ.
Max.
Unit
Cooling temperature
Ts
-51
-50
-49
°C
Temperature control range
-0.05
+0.05
°C
∆Ts
Power dissipation of TE-cooler element
Pp
15
W
Cool down time to reset temperature
to
5
min.
Setting temperature for overheat protection *6
To
+45
°C
*5: Other functions include error display, automatic power off, and detection of electrical opens and shorts by the thermosensor.
*6: Temperature at back side of housing (between housing and fun).
■ Spectral response (without window)
(Typ. Ta = 25 °C)
50
QUANTUM EFFICIENCY (%)
S7847 SERIES
40
30
20
S7017 SERIES
10
0
400
500
600
700
800
900
1000 1100 1200
WAVELENGTH (nm)
KMPDB0104EA
■ Device structure, line output format
IG1V IG2V ISV
7
5
6
SS
3
TG
25
P1V
24
P2V
23
......
V
8
RD
9
......
......
H
S1023
S1024
D11
D12
D13
D14
D15
D16
D17
D18
D19
D20
RG
D1
D2
D3
D4
D5
D6
D7
D8
D9
D10
S1
S2
1
V=124, 252
H=1024
OS 10
21 IG1H
11
OD
12
OG
13
SG
4 BLANK
18 19
20
IG2H
1024 2 ISOLATOIN P2H P1H
4 OPTICAL
SIGNAL OUT
4 BLANK
BLACK
2 ISOLATOIN
4 OPTICAL
BLACK
KMPDC0085EA
Pixel format
Blank
4
22 ISH
Optical Black
4
Left ← Horizontal Direction → Right
Isolation
Effective
Isolation
2
1024
2
Optical Black
4
Blank
4
Top ← Vertical Direction → Bottom
Isolation
Effective
Isolation
2
124 or 252
2
3
CCD multichannel detector head
C7027
■ Block diagram
P+
Vp (+7 V)
P. GND
VD2 (+7 V)
D. GND
PLED
TH1
TH2
VF (+12 V)
F. GND
Temp Monitor
Cooling Control
VD1 (+6 V)
D. GND
VA1+ (+15 V)
VA1- (-15 V)
A. GND
TEMPERATURE CONTROLLER
TE-COOLER
OD
RD, ISV, ISH
OG
IG1H, IG2H
IG1V, IG2V
P1V, P2V, TG
TH.H
REGULATOR
P1H, P2H
SG
RG
VA2 (+24 V)
Start
CLK
VEE
VDD
DRIVER
TIMING
GENERATOR
EOS
Trigger
Mode Select
SS
BUFFER
Data Video
OS
PRE-AMP
INVERTER
CCD AREA IMAGE SENSOR
S7017/S7847 SERIES
CLAMP
LOW-PASS
FILTER
CCCKACCC0159EB
■ Timing chart of line binning operation
Start
CLK
P1V
P2V
TG
P1H
P2H
SG
RG
Clamp
EOS
Trigger
Data Video
H-ch10
V-ch1
H-ch11
V-ch2
H-ch(N-10) H-ch(N-9)
V-ch(M)
H-ch1
H-ch2
H-ch(N)
M=128, 256 N=1044
V-ch1--V-ch4, V-ch (M-1)--V-ch (M): ISOLATION PIXELS
H-ch1--H-ch4, H-ch (N-3)--H-ch (N): BLANK PIXELS
KACCC0160EC
4
CCD multichannel detector head
C7027
■ Timing
chart of area
scanning
operation
g
g
(
)
Start
CLK
P1V
P2V
TG
P1H
P2H
SG
RG
Clamp
EOS
Trigger
• • • • • • • • •
BLOCK 1
BLOCK M
Data Video
Start
CLK
P1V
P2V
TG
P1H
P2H
SG
RG
Clamp
EOS
Trigger
Data Video
V-ch1
H-ch1
H-ch10 H-ch11
H-ch (N-10) H-ch (N-9)
H-ch (N)
NOTE) M=128, 256 N=1044
KACCC0161EA
■ Pin connection of "SIGNAL I/O" connector
15-pin D-sub connector
8
7
15
6
14
5
13
4
12
3
11
2
10
1
9
KACCC0069EA
Pin No.
Terminal name
1
Mode select
2
3
4
5
Data video
VA1+ (+15 V)
VA1- (-15 V)
VD1 (+5 V)
6
Start
7
CLK
8
EOS
9
10
11
12
13
14
15
A. GND
A. GND
VA2 (+24 V)
D. GND
D. GND
D. GND
Trigger
Description
Digital input signal used to select between the line binning operation and area scanning
operation.
HCMOS compatible.
High level or left open: line binning operation
Low level: area scanning operation
Analog video output. Positive polarity
Analog power supply
Analog power supply
Digital power supply
Digital input signal for initializing the circuit.
HCMOS compatible. Positive logic.
The interval of the Start pulses determines the integration time of the CCD image sensor.
Digital input signal for operating the circuit.
HCMOS compatible. Rising edge operation.
Digital input signal for indicating end-of-scan of the image sensor.
HCMOS compatible. Negative logic.
Analog ground
Analog ground
Analog power supply
Digital ground
Digital ground
Digital ground
Digital output signal for A/D conversion. HCMOS compatible. Positive logic.
5
CCD multichannel detector head
C7027
■ Pin connection of "TE CONTROL I/O" connector
9-pin D-sub connector
5
4
9
3
8
2
7
1
6
KACCC0075EA
Pin No.
1
2
Terminal name
VD2 (+7 V)
Temp monitor
Description
Digital power supply
Analog output signal of the temperature of the CCD image sensor
Digital input signal for starting to cool down. HCMOS compatible.
3
Cooling control High level or left open: cooling
Low level: stand-by
Power supply for the thermoelectric cooler in the CCD image sensor.
4
Vp (+7 V)
(Please use AWG 18 wire)
5
VF (+12 V)
Power supply for cooling fan
6
D. GND
Ground
7
D. GND
Ground
Power supply return of the thermoelectric cooler mounted in the CCD image sensor.
8
P. GND
(Please use AWG 18 wire)
9
F. GND
Power supply return for cooling fan
Available for using same power supply (+7 V) for "+VD2" and "+Vp".
42.1
79.5
5.0
■ Dimensional outline (unit: mm)
5.9
110.0
0.5
FOCAL PLANE
98.4
60.0
50.0
80.0
100.0
TE-control I/O
Signal I/O
40.0
(2 ×) M3 DEPTH 4
50.0
23.0
(4 ×) M3 DEPTH 4
KACCA0109EA
6
CCD multichannel detector head
C7027
■ Connection example
TE CONTROLLER
POWER SUPPLY VF (+12 V)
F. GND
VP (+7 V)
P. GND
VD2 (+7 V)
D. GND
CONTROLLER Cooling
D. GND
9
5
4
8
7
6
3
2
1
I/O SIGNAL
PULSE GENERATOR CLK
D. GND
Start
D. GND
POWER SUPPLY
VD1 (+5 V)
D. GND
VA1+ (+15 V)
VA1- (-15 V)
VA2 (+24 V)
A. GND
CONTROLLER
Mode Select
D. GND
S/H or A/D
Analog Input
A. GND
S/H Input
D. GND
OSCILLOSCOPE
CH1
GND
Ext.Trig
GND
15
8
7
14
13
12
11
10
6
5
4
3
2
9
1
KACCC0162EB
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions.
Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2007 Hamamatsu Photonics K.K.
HAMAMATSU PHOTONICS K.K., Solid State Division
1126-1 Ichino-cho, Higashi-ku, Hamamatsu City, 435-8558 Japan, Telephone: (81) 53-434-3311, Fax: (81) 53-434-5184, www.hamamatsu.com
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P.O.Box 6910, Bridgewater, N.J. 08807-0910, U.S.A., Telephone: (1) 908-231-0960, Fax: (1) 908-231-1218
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49) 08152-3750, Fax: (49) 08152-2658
France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: 33-(1) 69 53 71 00, Fax: 33-(1) 69 53 71 10
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road, Welwyn Garden City, Hertfordshire AL7 1BW, United Kingdom, Telephone: (44) 1707-294888, Fax: (44) 1707-325777
North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171 41 Solna, Sweden, Telephone: (46) 8-509-031-00, Fax: (46) 8-509-031-01
Italy: Hamamatsu Photonics Italia S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39) 02-935-81-733, Fax: (39) 02-935-81-741
Cat. No. KACC1069E05
May 2007 DN
7