AGILENT 11974V

Agilent 11974 Series
Preselected Millimeter Mixers
RF Sections Available in Four Bands
from 26.5 to 75 GHz
Data Sheet
For the first time, preselected
signal analysis is available to
75 GHz. With preselection, you can
quickly and easily characterize all
signals of interest. No images or
multiples are displayed to confuse
the measurement. Time consuming
signal-identification routines are
eliminated.
Figure 1 shows how true signals can
be obscured by displayed multiples
and images in an unpreselected
system. Figure 2 shows how only the
true signals are displayed when an
Agilent Technologies 11974 Series
preselected millimeter mixer is used
in making the measurement.
Whether your millimeter-frequency
application is component or system
testing, signal surveillance or EMI
measurements, your software
Figure 1.
development and verification time will
be reduced and your measurement
times will be decreased. Simply place
a marker on a signal—all responses
are real—and perform a preselector
peak. The analyzer then displays
the amplitude and frequency of the
signal. No time is lost creating and
running signal identification routines,
which used to be required to separate
true responses from multiples and
images.
Each 11974 Series preselected
millimeter mixer is characterized for
conversion loss versus frequency.
Calibration accuracy ranges from
1 dB at 26.5 GHz to 2 dB at 75 GHz.
Just enter the conversion loss data
into your host spectrum analyzer
and then make amplitude-calibrated
measurements.
Choose your host spectrum analyzer
Agilent 11974 Series preselected
millimeter mixers are compatible with
Agilent’s 8566B, 8560 Series and
E4407B of rugged portables1, 70000
Series modular spectrum analyzers
and the E4440A/46A/48A PSA
Series1.
Figure 2.
1. Requires Option AYZ (external mixing).
Added Benefits of Preselected Millimeter Measurements
Reduced overload from many signals
Preselected millimeter mixers reduce
susceptibility to overloading by
numerous signals. Overloading can
cause your measurements to be
uncalibrated.
System amplitude accuracy
Millimeter preselection improves
system amplitude accuracy by
reducing the level of local oscillator
(LO) emissions back to the device
under test. Option 001 further
improves system amplitude accuracy
by adding an isolator in front of the
input filter. With Option 001, the RF
input VSWR is <1.6:1 in A, Q, and U
bands and <2.3:1 in V Band.
System configuration
All versions of 8563A spectrum
analyzers and 70907B external mixer
interface modules are fully compatible
with the 11974 Series.
To operate your 11974 Series
preselected millimeter mixer, simply
connect the tune+span output, local
oscillator output,1 and the IF input of
the spectrum analyzer to the mixer.
Select the appropriate analyzer using
the switch on the rear panel of the
mixer. Make two easy adjustments,
and you’re ready to make millimeter
measurements.
Greater dynamic range
Preselection improves dynamic range
when signals are sufficiently spaced.
The preselector allows only one of
these signals to reach the mixer at a
time. So internally produced distortion
products, such as third order
intermodulation, are reduced.
The 11974 Series connect to a
standalone power supply with a
convenient 6-foot cable. Use the
same power supply with any 11974
Series preselected mixer.
Ease of use
A full set of upgrade kits is available
to ensure compatibility with your
existing spectrum analyzers including
Agilent 8560A, 8561A/B, 8562A/B,
8566A/B, and E4440A/46A/48A PSA
Series. Refer to the compatibility table
to determine whether your spectrum
analyzer requires an upgrade.
Operators no longer need to learn
techniques of signal identification.
Anyone familiar with preselected
microwave spectrum analyzers can
quickly and easily make millimeter
measurements. Spurs from a device
under test are easily found. They
are not hidden by images or multiple
responses.
11974A
11474Q
11974U
11974V
26.5 GHz
33 GHz
40 GHz
50 GHz
60 GHz
75 GHz
1. Operation with 8566B spectrum analyzers requires the use of an 11975A LO amplifier to provide sufficient local oscillator power to the 11974 Series LO input.
2
Specifications
Unless noted, all specifications
describe warranted performance
under the following conditions:
5-minute warm-up from ambient
conditions; 1-year calibration cycle;
environmental requirements met;
IF and LO connections between
the Agilent 11974 and a compatible
spectrum analyzer are made with one
length of coax cable (part number
5061-5458). Typical performance
is non-warranted. Supplemental
characteristics are denoted by
“nominal”; these constitute nonwarranted functional performance
information derived during the design
process and are not tested on a
continuing basis.
Frequency
Frequency range
(LO harmonic number)
Amplitude
Amplitude
Displayed average noise level
Spectrum analyzers: 8566B, 71OOOC,
PSA Series (RBW = 10 Hz)
Image rejection (image positioned
2x fIF above tuned response)
Spectrum analyzers: 8566B, 71000C,
PSA Series
Maximum Typical
11974A
–111 dBm
–118 dBm
11974Q
–106 dBm
–116 dBm
11974U
–109 dBm
–117 dBm
11974V
–100 dBm
–109 dBm
Displayed average noise level
Spectrum analyzers: 8563A/62A/61B/
60A (RBW = 300 Hz)1
Maximum Typical
11974A
–96 dBm
–103 dBm
11974Q
–91 dBm
–101 dBm
11974U
–94 dBm
–104 dBm
11974V
–85 dBm
–94 dBm
O °C to
55 °C
20 °C to
30 °C
11974A
≤ –54 dBc
≤ –59 dBc
11974Q
≤ –50 dBc
≤ –55 dBc
11974U
≤ –50 dBc
≤ –55 dBc
11974V
≤ –50 dBc
(50 to 67 GHz)
≤ –55 dBc
11974V
≤ –40 dBc
(67 to 75 GHz)
≤ –45 dBc
Image rejection (image positioned
2x fIF above tuned response)
Spectrum analyzers: 8563A/62A/61B/
60A
O °C to
55 °C
20 °C to
30 °C
11974A
≤ –50 dBc
≤ –55 dBc
11974A (n = 8)
26.5 GHz to 40 GHz
Maximum conversion loss
RF to IF port (includes 1-m IF cable,
part number 5061-5458)
11974Q
≤ –45 dBc
≤ –50 dBc
11974Q (n = 10)
33 GHz to 50 GHz
11974A
44 dB
11974U
≤ –45 dBc
≤ –50 dBc
11974U (n = 10)
40 GHz to 60 GHz
11974Q
46 dB
50 GHz to 75 GHz
11974U
43 dB
11974V
≤ –45 dBc
(50 to 67 GHz)
≤ –50 dBc
11974V (n = 14)
11974V
57 dB
11974V
≤ –35 dBc
(67 to 75 GHz)
≤ –40 dBc
Preselector bandwidth
(3 dB points)
Minimum
Typical
11974A
80 MHz
130 MHz
11974Q
100 MHz
150 MHz
11974U
100 MHz
150 MHz
11974V
100 MHz
200 MHz
Center frequency accuracy,
residual FM, span accuracy
Using specifications of the spectrum
analyzer, evaluate characteristics with
reference to the harmonic numbers
listed below.
Conversion loss chart accuracy
20 °C to 30 °C: RSS of calibration
equipment errors (Add 0.7 dB if used
with 8563A/62A/61B/60A.)
0 °C to 55 °C: RSS of calibration
equipment errors plus worst-case
temperature effects of 11974 (Add
0.7 dB if used with 8563A/62A/61B/
60A.)
Multiple response
(Due to in-range applied signals mixing
with LO harmonics other than order n)
11974A (n = 8)
≤ –63 dBc
11974Q (n = 10)
≤ –60 dBc
11974U (n = 10)
≤ –60 dBc
11974V (n = 14)
(applied signal:
50 to 67 GHz)
≤ –60 dBc
(applied signal:
67 to 75 GHz)
≤ –55 dBc
0 °C to
55 °C
20 °C to
30 °C
11974A
≤ ±2.3 dB
≤ ±1.0 dB
11974Q
≤ ±2.3 dB
≤ +1.0 dB
Harmonic number
11974U
≤ ±2.6 dB
≤ +1.1 dB
1 dB gain compression (nominal)
11974A
n=8
11974V
≤ ±4.5 dB
≤ ±2.0 dB
11974A
≥ +6 dBm
11974Q
n = 10
11974Q
≥ 0 dBm
11974U
n = 10
11974U
≥ 0 dBm
11974V
n = 14
11974V
≥ +3 dBm
1. The minimum usable resolution bandwidth of the 8563A/61B/60A in external mixing mode is 300 Hz. The 8562A can be used with its minimum resolution band width of
100 Hz in external mixing mode.
3
Specifications
Amplitude
Third-order intercept (nominal)
(For two signals spaced less than
200 MHz apart)
11974A
≥ +13 dBm
11974Q
≥ +13 dBm
11974U
≥ +13 dBm
11974V
≥ +13 dBm
IF subharmonic response (nominal)
Intercept (2nd order; response at
fIF/2 above input signal)
fIF/2 = 160.7 MHz
(8566B/71000C)
fIF/2 = 155.35 MHz
(8563A/62A/61B/60A)
11974A
≥ +45 dBm
11974Q
≥ +45 dBm
11974U
≥ +45 dBm
11974V
≥ +30 dBm
Inputs and outputs
(all values are nominal)
Options
Option 001: Input isolator attached
and included in calibration. An input
isolator provides a specified broadband
RF input match. This is important
when the device under test requires
protection from out-of-band reflections
off the 11974’s input tracking filter.
With Option 001, the 11974 conversion
loss is calibrated with the input isolator
in place so that system amplitude
accuracy is maintained.
LO input
Connector type
SMA (f)
Frequency range
11974A
3.3 GHz to 5 GHz
11974Q
3.3 GHz to 5 GHz
11974U
4 GHz to 6 GHz
11974V
3.5 GHz to 5.3 GHz
Maximum LO
input level
+14.5 dBm to 16 dBm
VSWR
< 2.0:1 (50 Ω)
Specifications changes for Option 001
Amplitude
IF output
Connector type
SMA (f)
Match
< 1.6:1 (50 Ω)
Displayed average noise level and
maximum conversion loss: Add +2 dB
to the standard specifications for A, Q,
and U bands; add +3 dB for V band.
Tune + span input
Connector type
RF input VSWR
(across full waveguide band)
BNC (f)
Dynamic range
Refer to Figure 3
11974A, Q, U
< 1.6:1
11974V
< 2.3:1
Inputs and outputs
(all values are nominal)
–20
RF input
Waveguide size, flange type
UG 383/U-M
11974V
WR-15
UG 385/U
Maximum
RF input
level (CW,
peak, or
average)
+25 dBm
VSWR (at
peak of
preselector
filter)
< 3.0:1
Typical spur or noise level (dBc)
UG 383/U
WR-19
W
=
Second order
IF subharmonic
distortion
10
Hz
Hz
WR-22
11974U
RB
–50
0
11974Q
–40
30
UG 599/U
=
Flange
WR-28
W
Size
RB
11974A
Displayed
average
noise floor
–30
–60
Image and
multiple
response
–70
Third order
intermod
(≤ 200 MHz
signal separation)
–80
–90
–100
–110
–120
–130
–70
–60
–50
–40
–30
–20
–10
RF input (dBm)
Figure 3. Agilent 11974A system dynamic range
4
0
+10
+20
Specifications
General specifications
General specifications
Environmental specifications type
tested to MIL-T-28800C, Class 5
environmental conditions as listed
below:
Power requirements: 11974-60028
power supply (nominal)
Temperature
Non-operating
–40 °C to +75 °C
Operating
0 °C to +55 °C
Output
+50 V, 0.6 A dc
±15 V, 0.2 A dc
Input
100/120/220/240 V ac
50 to 60 Hz
Power
consumption
130 VA maximum
Relative humidity
Max. power
dissipation
95% ± 5% up to +40 °C
11974 Series
35 W
Accessories supplied
11974-60028
25 W
The following is a list of all accessories
and their part numbers supplied at
shipment:
Altitude
Non-operating
< 12,195 m
(< 40,000 feet)
Weight (nominal)
11974 Series
6.5 kg (14.3 lb)
Operating
< 3,048 m
(< 10,000 feet)
11974-60028
preselector power
supply
11.9 kg (26.21 lb)
Accessory
description
Part
number
Power supply
11974-60028
Maximum vibration level
Ball Driver, 3/32 inch
8710-1539
5 Hz to 55 Hz
Dimensions (nominal)
Wrench, 5/16 inch
8710-0510
11974 Series 210.2 mm (8.3”) x
84.7 mm (3.3”) x
101.5 mm (4.0”)
Alignment tool
8710-0630
Four standard #4-40
screws for A band
3030-0221
Captive screws for Q, U,
and V bands, 4 each
1390-0671
3 cables with SMA (m)
connectors (1 m)
5061-5458
1 cable with BNC (m)
connectors (48 inches)
8120-1839
1 cable with BNC (m)
connectors (24 inches)
8120-2582
1 adapter with BNC (f)
connectors
1250-0080
1 stand for the 11974
83556-60010
2g
Maximum shock
30 g
Electromagnetic compatibility
Meets radiated and conducted
emissions of MIL-STD-461B, Part 7,
methods CE03 (Air Force) and RE02,
FTZ 526/527/79
11974-60028
preselector
power
supply
346.1 mm (13.6”) x
212.7 mm (8.4”) x
133.4 mm (5.25”)
5
Compatibility Information
The following table lists compatibility requirements of host spectrum analyzers. Installation by a service center or
personnel qualified by Agilent is required for the E4440A, E4446A, E4448A, and E4407B Option AYZ upgrade kits.
70000 modular measurement system with 70907 external mixer interface module
Model
S/N prefix
Compatibility kit
70907A
all
70907A-K74 upgrade kit required1
70907B
all
Fully compatible provided 70900A/B LO meets the following requirements:
Firmware date code YYMMDD
LO upgrade kit required
850730
70907B-098
860203
(Firmware + CPU)
861015
70907B-099
870501
(Firmware only)
880314
880901
890606
Fully compatible
Rugged portable spectrum analyzers
Model
S/N prefix
Compatibility kit
8560A
< 3003A
8560A-K74 upgrade kit required2
8560A
≥ 3003A
Fully compatible2
8561A
all
8561A-K74 upgrade kit required3
8561B
< 3003A
8561 B-K74 upgrade kit required
8561B
≥ 3003A
Fully compatible
8562A/B
all
8562A-K74 upgrade kit required3
8560E/EC
all
Fully compatible
8561E/EC
all
Fully compatible
8562E/EC
all
Fully compatible
8563A/E/EC
all
Fully compatible
all
Fully compatible
5
E4407B
PSA Series with Option AYZ
Model
S/N prefix
Compatibility kit
E4440A/46A/48A
all
Fully compatible
8566A/B with 11975A LO amplifier
Model
S/N prefix
4
Compatibility kit
8566A
< 2007A
No upgrade kit available
8566A
≥ 2007A4
8566AB conversion kit previously installed:
Firmware datecode < 89.10.10: 8566B-K74 upgrade kit required
Firmware datecode ≥ 89.10.10: 8566B-K75 upgrade kit required
8566AB conversion kit NOT previously installed:
(two kits required) 8566AB conversion kit, 8566B-K75 upgrade kit
8566B
< 29484
8566B-K74 upgrade kit required
8566B
≥ 29484
Fully compatible
1. Mixer bias capability is converted to preselector peak, resulting in activation of a preselector peaking with the BIAS PEAK function. Sweep times must be manually
controlled in wide spans to avoid oversweeping the preselector.
2. External mixing is not available with the 8560EC if a built-in tracking generator, Option 8560EC-002, has been installed.
3. Operation with 11974 Series will require entry of a frequency offset. Preselector peaking is performed manually with a screwdriver adjustment. Sweep times must be
manually controlled in wide spans to avoid oversweeping the preselector.
4. The serial number prefix on the 8566A/B refers to the RF section.
5. Requires Option AYZ (external mixing).
6
Ordering Information
11974A 26.5 to 40 GHz
Compatible spectrum analyzers
11974Q 33 to 50 GHz
E4440A 3 Hz to 26.5 GHz
high-performance
11974U 40 to 60 GHz
11974V 50 to 75 GHz
E4446A 3 Hz to 44 GHz
high-performance
Option 11974X-001: Input isolator
attached and included in calibration
(Note: X = A, Q, U or V)
E4448A 3 Hz to 50 GHz
high-performance
Option 11974X-003: Exclusion of
power supply (Note: X = A, Q, U or V)
8560EC 2.9 GHz rugged portable
5062-3989: Front handle kit for power
supply
5062-3957: Rack mount kit with
extended flange for power supply
(half-width)
8561EC 6.5 GHz rugged portable
8562EC 13.2 GHz rugged portable
8563EC 26.5 GHz rugged portable
8564EC 40 GHz rugged portable
Other related accessories available
from Agilent Technologies:
281A/B waveguide with 2.4-mm coax
adapters
382A attenuators for R, Q, and U
bands only
373D/G 20/50 dB fixed attenuators
370A/B/C 3 dB, 6 dB, and 10 dB
fixed attenuators for Q and U bands
only
800 Series waveguide bends, twists,
and straight sections
365A Series waveguide isolators for
R, Q, U, and V bands
8565EC 50 GHz rugged portable
Warranty and service
Standard warranty is 12 months.
Option R-51B: Return-to-Agilent
warranty and service
E4407B 26.5 GHZ rugged portable
71100C 2.9 GHz modular
71200C 22/26.5 GHz modular
71210C 22 GHz modular
Calibration
R-50C-001: Standard calibration
70907B external mixer interface
module
R-50C-002: Standard compliant
calibration
8566B 22 GHz benchtop
11975A LO amplifier
Note: An 8562A can be made
hardware compatible with the 11974
Series by installing the 8562A-K74
upgrade kit. Limitations of using an
8562A with the 11974 Series are
indicated in footnote 1 below.
1. Operation with 11974 Series will require entry of a frequency offset. Preselector peaking is performed manually with a screwdriver adjustment. Sweep times must be
manually controlled in wide spans to avoid oversweeping the preselector.
7
www.agilent.com
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
Remove all doubt
Our repair and calibration services
will get your equipment back to you,
performing like new, when promised. You will get full value out of
your Agilent equipment throughout its lifetime. Your equipment
will be serviced by Agilent-trained
technicians using the latest factory
calibration procedures, automated
repair diagnostics and genuine parts.
You will always have the utmost
confidence in your measurements.
For information regarding self
maintenance of this product, please
contact your Agilent office.
Agilent offers a wide range of additional expert test and measurement services for your equipment,
including initial start-up assistance,
onsite education and training, as
well as design, system integration,
and project management.
For more information on repair and
calibration services, go to:
www.agilent.com/find/removealldoubt
Product specifications and descriptions
in this document subject to change
without notice.
For more information on Agilent
Technologies’ products, applications or
services, please contact your local Agilent
office. The complete list is available at:
www.agilent.com/find/contactus
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008
Europe & Middle East
Austria
01 36027 71571
Belgium
32 (0) 2 404 93 40
Denmark
45 70 13 15 15
Finland
358 (0) 10 855 2100
France
0825 010 700*
*0.125 €/minute
Germany
07031 464 6333
Ireland
1890 924 204
Israel
972-3-9288-504/544
Italy
39 02 92 60 8484
Netherlands
31 (0) 20 547 2111
Spain
34 (91) 631 3300
Sweden
0200-88 22 55
Switzerland
0800 80 53 53
United Kingdom 44 (0) 118 9276201
Other European Countries:
www.agilent.com/find/contactus
Revised: October 6, 2008
© Agilent Technologies, Inc.
2005, 2008
Printed in USA, October 17, 2008
5952-2748