SLPOWER MENB1130A1251F01

MENB1130
Universal Input 130 Watt Series
Medical Switch-Mode Power Supply
3 Year Warranty
•100-240VAC Universal Input
•Meets EISA2007, CEC Efficiency Level V,
EU (EC) No 278/2009 Phase II
•Desktop Style
•12V to 24V Single Output Models, up to 130W
•Modified and Custom Designs Available
•Regulated Output with Low Ripple
•Impact-Resistant Polycarbonate Enclosure
•No Load Power Consumption <0.5W
rd
•Meets UL/EN/IEC60601-1, 3
Edition
RoHS
Specifications
All Specifications are typical at nominal input, full load at 25°C unless otherwise stated.
AC Input
Input Current
100-240Vac, +/-10%, 47-63 Hz, 1∅
100 Vac: 2.2 A
Inrush Current
Input Fuse
60A peak @ 264Vac, cold start
3.15A, 250V Internal Primary Current Fuse provided
Efficiency
Meets International Efficiency Level V
MTBF
100,000 hours calculated
Hold-up Time
18 ms min. @ 115Vac, 60 ms min. @ 230 Vac
Overload Protection
Hiccup Mode
Short Circuit Protection
Topology
Output Voltage
See chart
Safety and EMC Approvals
Output Power
See chart
Dielectric Withstand
Ripple and Noise
1% pk-pk max., 20MHz BW
Line/Load Regulation
Transient Response
Excluding Cord: Line: +/- 1%, Load: +/-5%
500µs max., 50% load step, typical
Minimum Load
Not required
Case Material
Case Dimensions
drawing
Black 94V0 Polycarbonate
209.4mm x 64.4mm x 46mm , see mechanical
Hiccup Mode
Switching – Fixed Frequency Flyback
Operating Temperature
Storage Temperature
Relative Humidity
EN/IEC/CSA/UL60601-1
EMC: See chart
Input-Output: 5,656Vdc
Input-GND: 1,500 Vac, Output-GND: 500Vdc
0° to 40°C, no derating
-30 to +85°C
5% to 95%, non-condensing
Altitude
Output Connections
Weight
0 to 10,000 ft
Cable: 18AWG, 1,500mm, 4 conductor
Connector: Ault #3, 2.5mm or Ault #51, 6-pin Molex
700g
EMI/EMC Compliance
Conducted Emissions
FCC Part 15, Class B, EN55011 Class B, EN55024
Radiated Emissions
FCC Part 15, Class B, EN55011 Class B, EN55024
Voltage Fluctuations
EN61000-3-3, Line Flicker
Static Discharge Immunity
EN61000-4-2, 6kV Contact Discharge, 8kV air discharge
Radiated RF Immunity
EN61000-4-3, 3V/m.
EFT/Burst Immunity
EN61000-4-4, 2kV/5kHz..
Line Surge Immunity
EN61000-4-5, 1kV differential, 2kV common-mode
Conducted RF Immunity
EN61000-4-6, 3Vrms
Power Frequency Magnetic Field Immunity
EN61000-4-8, 3A/m
Voltage Dip Immunity
EN61000-4-11 crit. A, 100Vac 60Hz, 40%/5 cycles with 70% full load.
Line Frequency Harmonics
EN61000-3-2, Class A
______________________________________________________________________________________________________________________________ 1 of 2
SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com
MENB1130
Universal Input 130 Watt Series
Medical Switch-Mode Power Supply
Model Number
3 Year Warranty
Volts (V)
Output Current (max)
Max Watts
Ripple (Vp-p max)
MENB1130A1251F01
12 V
10.8 A
130 W
120mV
MENB1130A1551F01
15 V
8.67 A
130 W
150mV
MENB1130A1803F01
18 V
7.2 A
130 W
180mV
MENB1130A2403F01
24 V
5.4 A
130 W
240mV
Input Receptacle
Output Connector
Output Connector (12V & 15V models)
“#3” Connector is 2.5mm barrel type, center (+)
“#51” Connector is Molex 39-01-2060 or equivalent.
IEC320 C14 - Class I
Grounded (F)
“51" Connector Pinout: Pins 1, 4: (+), Pins 3, 6: (-). Mate = Molex 39-01-2061 or equiv.
Outline Drawing
Dimensions are in mm.
Data Sheet © 2011 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication.
Rev. 6-29-2012
However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
______________________________________________________________________________________________________________________________ 2 of 2
SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: [email protected] • www.slpower.com