ETC 2700

02 Switch/Measure Systems
11/16/01
1:35 PM
Multimeter/Data Acquisition/
Switch Systems
Integra Series integrated switching, measurement, and datalogging solutions
2700
2750
Integra Series products combine precision measurement, switching, and control in a single, tightly integrated enclosure for either rack-mounted or benchtop
applications. These cost-effective, high performance test
platforms offer affordable alternatives to separate DMMs
and switch systems, dataloggers/recorders, plug-in card
data acquisition equipment, and VXI/PXI systems.
Integra Series plug-in switching and control modules
offer unmatched flexibility and testing efficiency for a
wide range of industries and applications. System
builders can create test solutions with a combination of
channel count, cost per channel, and system performance unmatched by any other single-box measurement
system. The input modules provide the flexibility to
vary the channel count from 20 to 200 (2-pole), apply a
stimulus to the device under test, route signals, control
system components, and make precision measurements
with up to 14 functions. Robust digital I/O capabilities
can be used for triggering, handshaking with other
automation equipment, and alarm limit outputs. Scan
rates of more than 200 channels/second (up to 2500
readings/second) increase test productivity.
• Combines functions of DMM,
switch system, and datalogger
• True 61⁄2-digit (22-bit) resolution
• Choice of 9 switch/control
plug-in modules
• Up to 200 differential input
channels (with 300V isolation)
for measurement and control
• Convenient front panel inputs
• TestPoint™ start-up software
• LabVIEW™, LabWindows™/CVI,
Visual Basic, C/C++, and
TestPoint drivers
• Optional ExceLINX-1A
datalogging software
SWITCH/MEASURE SYSTEMS
Page 50
Fast Setup and Operation
The Model 2700 and Model 2750 are fully integrated, off-the-shelf measurement and control systems.
Their DMM-like interfaces make it easy for users to collect data and/or perform troubleshooting within
minutes of installation and startup. Once sensor or DUT leads are hooked to the instrument’s input, use
the front panel controls to select the measurement function, range, filtering, scaling, trigger source, scanning sequence, alarms, and more. The TestPoint runtime start-up software included with both the Model
2700 and 2750 makes it easy to configure and use the system in a graphical “point-and-click” environment. This gives developers the basic tools needed to create a simple application without writing program code. If the application demands greater functionality, this free runtime can be modified with the
TestPoint software application package (sold separately).
The Advantage of Integrated Design
The Model 2700 and 2750 offer a variety of advantages over existing solutions for ATE and data acquisition
applications. For example, their flexible modular architecture and integrated measurement, switching, and
control capabilities save rack space by reducing the number of separate instruments needed. This design
also simplifies expanding the system as the number of channels grows or re-purposing it as new test
requirements evolve. Integrated signal conditioning, scaling, stimulus, filtering and I/O capabilities eliminate the need for external circuitry when designing and building data acquisition systems. This architecture also makes it unnecessary to open the computer to install plug-in boards. The Model 2700 and 2750
offer accuracy and repeatability superior to plug-in data acquisition boards and VXI/PXI systems, while
providing faster test times than typical DMM/switch systems. This makes it possible to combine higher
test yields with higher test throughput.
ACCESSORIES AVAILABLE
Screw terminals
use oversize
connectors for
easier, mistakefree wiring.
ExceLINX-1A
7788
7789
7790
7797
7705-MTC-2
7707-MTC-2
1.888.KEITHLEY
Excel Add-In Software for 2700 and 2750
Instruments
50-Pin D-Shell Connector Kit (2 each)(for
Models 7703, 7705 Modules w/D-sub
Connectors)
50-Pin/25-Pin D-Shell Kit (1 each)
50-Pin Male, 50-Pin Female and 25-Pin Male IDC
D-Shell Connector Kit (1 each) (Ribbon Cable
not Included)
Calibration Extender Board (for Model 2750)
50-Pin Male to Female D-Sub Cable, 2m
25-Pin Male to Female D-Sub Cable, 2m
(U.S. only)
www.keithley.com
A
G R E A T E R
M E A S U R E
O F
C O N F I D E N C E
Built-in measurement functions include:
• DCV
• ACV
• DCI
• ACI
• Resistance (2- or 4-wire, offset
compensation selectable)
• Dry circuit ohms (20mV clamp) 2750
only
• Temperature (with thermocouples,
RTDs, or thermistors)
• Frequency/Period
• Continuity
Ordering Information
These products are available with an
Extended Warranty.
Accessories Supplied
TestPoint runtime start-up software;
LabVIEW, LabWindows™/CVI, Visual
Basic, C/C++, and TestPoint drivers;
manual; and Model 1751 Safety Test
Leads.
1.888.KEITHLEY
Rugged 50-pin D-sub connectors ensure dependability
and quick setup/teardown in production test racks.
Install up to five input modules in the 2750 mainframe. All switch/control modules are fully enclosed in impactresistant plastic for exceptional ruggedness. Three connector alternatives simplify connecting the modules to
DUTs. Rugged D-sub connectors allow quick, secure connections and are especially convenient when performing
routine maintenance or when the system is installed in a rack. IDC ribbon cable adapters are supplied with the
Model 7701, 7707, and 7709 modules for fast, uncomplicated hookups in production test and process monitoring
applications. Oversize screw-terminal connectors simplify setup in applications that require the greatest connection flexibility. Additional D-sub and IDC ribbon cable connector kits and pre-wired cable assemblies are sold
separately.
All modules are compatible with the two-slot Model 2700 Multimeter/Data Acquisition System and the five-slot
Model 2750 Multimeter/Switch System. When the application’s needs change, simply change modules. Integra
systems reconfigure themselves automatically.
Module
AC and DC Volts
AC and DC Current
Temperature:
T/C w/Automatic CJC
T/C w/External CJC
RTD
Thermistor
Resistance (2- or 4-wire)
Frequency
Event/Counter/Totalizer
Signal Routing/Control
Digital Input
Digital Output
Analog Output
Analog Input Channels
7700
•
•
7701
•
7702
•
•
7703
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
20
32
40
32
7705
7706
•
•
•
•
•
•
•
•
•
•
40
•
•
20
7707
•
7708
•
7709
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
10
40
6×8
(U.S. only)
www.keithley.com
A
G R E A T E R
M E A S U R E
O F
C O N F I D E N C E
SWITCH/MEASURE SYSTEMS
2700 DMM, Data Acquisition,
Datalogging System w/2 Slots
2750 DMM, Data Acquisition,
Switching, Datalogging
System w/5 Slots
7700 20-Channel Differential
Multiplexer Module
w/Automatic CJC and
Screw Terminals
7701 32-Channel Differential
Multiplexer Module with a
25- and 50-Pin Female D
Connector. Supplied with Male
IDC Ribbon Cable Connectors
7702 40-Channel Differential
Multiplexer Module w/Screw
Terminals
7703 32-Channel High Speed,
Differential Multiplexer Module
with 2 50-Pin Female D
Connectors. Includes 2 Mating
Connectors
7705 40-Channel Single-Pole
Control Module with 2 50-Pin
Female D Connectors.
Includes 2 Mating Connectors.
7706 All-in-One I/O Module: 20Channel Differential
Multiplexer w/Automatic CJC,
16 Digital Outputs, 2 Analog
Outputs, a Counter/Totalizer,
and Screw Terminals
7707 32-Channel Digital I/O
w/10-Channel Differential
Multiplexer Module with a 25Pin Female and 50-Pin Male D
Connectors. Supplied with
Mating IDC Ribbon Cable
Connectors
7708 40-Channel Differential
Multiplexer Module
w/Automatic CJC and Screw
Terminals
7709 6x9 Matrix Module with 25and 50-Pin Female D
Connectors. Supplied with
Male IDC Ribbon Cable
Connectors
Integra Series integrated switching, measurement, and datalogging solutions
Multimeter/Data Acquisition/
Switch Systems
2700
2750
02 Switch/Measure Systems
11/16/01
1:35 PM
Multimeter/Data Acquisition/
Switch Systems
Integra Series integrated switching, measurement, and datalogging solutions
2700
2750
SWITCH/MEASURE SYSTEMS
Page 52
Integra Series Module Selector Guide
This selector guide may prove helpful in identifying the best module for a specific application. Install up
to two modules at a time in the Model 2700 mainframe or five at a time in the Model 2750 mainframe.
Keithley’s Integra Systems provide precision measurement and control in a single, tightly integrated
enclosure for either rack-mounted or benchtop use. These cost-effective, high-performance test platforms
are affordable alternatives to separate DMMs and switch systems, dataloggers/recorders, and plug-in card
data acquisition equipment. Integra Series plug-in modules offer unmatched flexibility and testing efficiency for a wide range of industries and applications. Users can create test systems with a combination of
channel count, cost per channel, and system performance unmatched by any other single-box measurement system. The modules provide the flexibility to vary channel count from 20 to 200 (2-pole), apply a
stimulus to devices under test, route signals, control system components, and make precision measurements with up to 14 functions. Robust digital I/O capabilities can be used for triggering, handshaking with
other automation equipment, and alarm limit outputs. Scan rates of more than 200 channels/second (up
to 2500 readings/second) increase test productivity.
# Analog
Module Inputs
20
7700
Configuration
Multiplexer
w/CJC
Differential*
1 × 20 or
two 1 × 10
4-pole
1 × 10
Type of
Connector
Screw
terminals
Max.
Voltage
300V
Max.
Switched
Current
1A
Current
Measurement
Channels
2 channels
@ 3A
Digital
I/O
N/A
Switch
Speed
< 3 ms
Other
Automatic
thermocouple
linearization
7701
32
Multiplexer
1 × 32 or
two 1 × 16
1 × 16
D-sub (IDC)
150V
1A
N/A
N/A
< 3 ms
32 channels
of common-side
4-wire ohms
7702
40
Multiplexer
1 × 40 or
two 1 × 20
1 × 20
Screw
terminals
300V
1A
2 channels
@ 3A
N/A
< 3 ms
Maximum power
= 125VA
7703
32
Multiplexer
1 × 32 or
two 1 × 16
1 × 16
D-sub
(solder or crimp)
300V
500mA
N/A
N/A
< 1 ms
Reed relays
7705
40
Independent
SPST
N/A
N/A
D-sub
(solder or crimp)
300V
2A
N/A
N/A
< 3 ms
Software
programmable for
SPDT (Form C)
7706
20
Multiplexer
w/CJC
1 × 20 or
two 1 × 10
1 × 10
Screw
terminals
300V
1A
N/A
16 Digital
Out Only
< 3 ms
Two ±12V analog
output channels
& 100kHz event
counter/totalizer
7707
10
Digital I/O
1 × 10 or
two 1 × 5
1×5
D-sub
(IDC)
300V
1A
N/A
32 Digital
I/O
< 3 ms
Four 8-bit I/O ports
(33V, 100mA)
7708
40
Multiplexer
w/CJC
1 × 40 or
two 1 × 20
1 × 20
Screw
terminals
300V
1A
N/A
N/A
< 3 ms
Automatic
thermocouple
linearization
7709
48
6 × 8 Matrix
Yes
Yes
D-sub
(IDC)
300V
1A
N/A
N/A
< 3 ms
Connects to internal
DMM. Daisy chain
multiple cards to
create larger matrix
* Can be disconnected from internal DMM for routing external signals • Refer to the Model 2750 or Model 2700 brochure for additional information.
1.888.KEITHLEY
(U.S. only)
www.keithley.com
A
G R E A T E R
M E A S U R E
O F
C O N F I D E N C E
11/16/01
1:35 PM
Page 53
Important Features and Benefits
• Full per-channel configurability—Each channel
can be independently configured for making measurements. The parameters that can be chosen for
each channel include speed, range, resolution,
number of power line cycles (NPLC), filtering type,
offset compensation, math functions to be displayed, CJC type, RTD type, frequency gate time,
“m” and “b” values in mX + b format, HI/LO limits,
low Ω (Model 2750 only), ratio calculation, and
thermistor type.
• Channel monitor feature—Monitor any specific
input channel on the front panel display during a
scan. This feature can also serve as an analog trigger to initiate a scan sequence based on some
external factor, such as a temperature rising above
a pre-set limit. Only the data of interest is acquired,
so there’s no need to spend hours searching
through reams of normal readings to find anomalous data.
• Front/rear switch—Switching between the front
and rear panel measurement inputs is as easy as
pressing a button. Users can select the front panel
inputs for tasks such as system set-up and verification, manual probing, troubleshooting, and calibration, while the rear panel inputs through the
modules allow fast, automated multiplexing
and control.
• Battery-backed set-up memory—Up to four different set-up configurations can be stored in onboard
memory. If the line power fails during a scan, the
system will resume scanning where it stopped
once power is restored.
• 110K reading data buffer—The mainframe’s nonvolatile wrap-around reading memory allows continuous, unattended datalogging over long periods—data in the buffer can be transferred to a PC
controller automatically as new data is acquired.
The real-time clock can be used to time- and datestamp readings for later review and interpretation.
• 2 TTL-level digital inputs—For implementing external triggers to initiate a scan sequence.
• 5 “per-channel” HI/LO alarm limit TTL outputs —
Trigger external alarms or perform other control
functions without a PC controller.
• Dry circuit ohms (20mV clamp)—Protects sensitive
devices from damage and prevents self-heating
errors during testing (Model 2750 only).
• Virtual channel—Stores the results of channel-to
channel ratio and average math operations.
• Onboard statistical analysis—Mathematical functions available at the push of a button are channel
average, mX+b scaling, minimum, maximum, average, and standard deviation.
• GPIB and RS-232 interfaces
• 3-year warranty
1.888.KEITHLEY
Temperature Capabilities
Integra Series mainframes support three major types of temperature sensors with built-in
signal conditioning and 300V isolation: thermocouples, RTDs, and thermistors. To begin
using a sensor, simply hook it up and the instrument does the rest. If a thermocouple is
broken or disconnected, the instrument will alert the operator. The mainframes also support three methods for cold-junction compensation (CJC): automatic (built-in), external
(built-in), and simulated.
Software Solutions
Whether the task calls for a simple start-up package to acquire
several channels of data or the tools to create a fully custom
acquisition and analysis solution, Keithley has the software needrunning
ed to get the most performance from a Model 2700 or 2750
Multimeter/Switch System. Our broad range of software solutions s o f t w a r e
makes it easy to get applications “Up & Running” quickly and economically.
&
Our free start-up
software package
provides basic datalogging capabilities,
so a system can be
up and running
almost immediately. It
also can be used to
configure instrument
functions. Data from
multiple channels
from a single instrument can be saved to
disk; up to eight
channels of data can
be graphed automatically and multiple
configurations can be
saved to disk.
The new version of
ExceLINX is an economical, easy-to-use,
add-in utility for
Microsoft Excel.
Within minutes of
installing ExceLINX1A on a PC, users
can acquire data
directly from the
Model 2700 or 2750,
then employ Excel’s
graphics, charting,
and analysis capabilities to turn that data
into useful information. No programming
is required to use it—
a few mouse clicks
are all it takes to configure channels, set
parameters, and get
your measurements
into Excel.
(U.S. only)
www.keithley.com
A
G R E A T E R
M E A S U R E
O F
C O N F I D E N C E
Integra Series integrated switching, measurement, and datalogging solutions
Multimeter/Data Acquisition/
Switch Systems
2700
2750
SWITCH/MEASURE SYSTEMS
02 Switch/Measure Systems
02 Switch/Measure Systems
11/16/01
1:35 PM
Multimeter/Data Acquisition/
Switch Systems
Integra Series integrated switching, measurement, and datalogging solutions
2700
2750
SWITCH/MEASURE SYSTEMS
Page 54
Instrument Drivers
For experienced programmers who prefer to build
fully custom systems from scratch, Keithley provides a Series 2700/2750 instrument driver for use
with Application Development Environments such
as LabVIEW™, LabWindows™/CVI, Visual Basic,
C/C++, and TestPoint. This IVI-style driver (VISA
based) supports all of the functionality of the
Model 2700 and 2750. Numerous examples and an
on-line help utility are provided to help programmers get their applications up and running.
TYPICAL APPLICATIONS
• Production test of electronic
products and devices
• Accelerated stress testing (AST)
• Process monitor and control
• Device characterization/R&D
• Low ohms, multichannel
measurements
Which Integra Mainframe is the
Best Choice for the Application?
Use this selector guide to decide which Integra Series mainframe offers the combination of features and
capacity that’s right for a specific application. If testing requirements change in the future, switch/control
modules and test code can be easily re-used.
2700
80
96
100mΩ
No
2
55,000 rdgs
Half-rack width
No. of different input channels
Matrix crosspoints
Ohms resolution
Dry circuit ohms (20mV clamp)
No. of slots
Memory buffer
Size (2U height)
1.888.KEITHLEY
2750
200
240
1mΩ
Yes
5
110,000 rdgs
Full-rack width (19˝)
(U.S. only)
www.keithley.com
A
G R E A T E R
M E A S U R E
O F
C O N F I D E N C E
Multimeter/Data Acquisition/
Switch Systems
2700
2750
DC CHARACTERISTICS
VOLTAGE
RESISTANCE
2750 Only
2700 and 2750
DRY CIRCUIT
RESISTANCE
2750 Only
CONTINUITY (2W)
CURRENT
RANGE
RESOLUTION
100.0000
1.000000
10.00000
100.0000
1000.000
1.000000
10.00000
100.0000
1.000000
10.00000
100.0000
1.000000
10.00000
100.0000
mV
V
V
V
V
Ω
Ω
Ω
kΩ
kΩ
kΩ
MΩ
MΩ
MΩ
0.1
1.0
10
100
1
1
10
100
1
10
100
1.0
10
100
µV
µV
µV
µV
mV
µΩ
µΩ
µΩ
mΩ
mΩ
mΩ
Ω
Ω
Ω
10
10
1
1
100
10
10
0.7
0.7
1.000000
10.00000
100.0000
1.000000
1.000
20.00000
100.0000
1.000000
3.000000
Ω
Ω
Ω
kΩ
kΩ
mA
mA
A
A
1
10
100
1
100
10
100
1.0
10
µΩ
µΩ
µΩ
mΩ
mΩ
nA
nA
µA
µA
10
1
100
10
1
< 0.2
< 0.1
< 0.5
< 1.5
CHANNEL (RATIO) 10
INPUT
RESISTANCE
OR OPEN CKT.
VOLTAGE
>10
>10
>10
10
10
5.9
5.9
12.2
12.2
6.8
12.8
12.8
7.0
7.0
mA
mA
mA
mA
µA
µA
µA
µA // 10M Ω
µA // 10M Ω
mA
mA
µA
µA
mA
V
V
V
V
GΩ
GΩ
GΩ
MΩ ± 1%
MΩ ± 1%
V
V
V
V
V
V
V
V
V
20 mV
20 mV
20 mV
20 mV
12.2 V
ACCURACY: ±(ppm of reading + ppm of range)
(ppm = parts per million) (e.g., 10ppm = 0.001%)
TEMPERATURE
COEFFICIENT
0°–18°C & 28°–50°C
24 Hour
23°C±1°
90 Day
23°C±5°
1 Year
23°C±5°
15 + 30
15 + 6
10 + 4
15 + 6
20 + 6
80 + 40
20 + 20
20 + 20
20 + 6
20 + 6
20 + 6
20 + 6
150 + 6
800 + 30
25 + 35
25 + 7
20 + 5
35 + 9
35 + 9
80 + 40
80 + 20
80 + 20
80 + 6
80 + 6
80 + 10
80 + 10
200 + 10
2000 + 30
30 + 35
30 + 7
30 + 5
45 + 9
50 + 9
100 + 40
100 + 20
100 + 20
100 + 6
100 + 6
100 + 10
100 + 10
400 + 10
2000 + 30
(1 + 5)/°C
(1 + 1)/°C
(1 + 1)/°C
(5 + 1)/°C
(5 + 1)/°C
(8 + 1)/°C
(8 + 1)/°C
(8 + 1)/°C
(8 + 1)/°C
(8 + 1)/°C
(8 + 1)/°C
(8 + 1)/°C
(70 + 1)/°C
(385 + 1)/°C
80 + 40
25 + 40
25 + 40
25 + 90
40 + 100
60 + 30
100 + 300
200 + 30
1000 + 15
80 + 40
80 + 40
90 + 40
180 + 90
100 + 100
300 + 80
300 + 800
500 + 80
1200 + 40
100 + 40
100 + 40
140 + 40
400 + 90
100 + 100
500 + 80
500 + 800
800 + 80
1200 + 40
(8 + 1)/°C
(8 + 1)/°C
(8 + 1)/°C
(8 + 1)/°C
(8 + 1)/°C
(50 + 5)/°C
(50 + 50)/°C
(50 + 5)/°C
(50 + 5)/°C
Ratio Accuracy = Accuracy of selected Channel Range + Accuracy of Paired Channel Range
CHANNEL (AVERAGE) 10
Average Accuracy = Accuracy of selected Channel Range + Accuracy of Paired Channel Range
TEMPERATURE
DC MEASUREMENT SPEEDS
(Displayed in °C, °F, or K. Exclusive of probe errors.)
Single Channel, 60Hz (50Hz) Operation
THERMOCOUPLES (ACCURACY BASED ON ITS-90.)
90 Day/1 Year (23°C ± 5°C)
Relative to
Using
Simulated
Plug-In
Type
Range
Resolution
Reference Junction
Module
J
–200 to +760 °C 0.001 °C
0.2°C
1.0°C
K
–200 to +1372°C 0.001 °C
0.2°C
1.0°C
N
–200 to +1300°C 0.001 °C
0.2°C
1.0°C
T
–200 to +400°C 0.001 °C
0.2°C
1.0°C
E
–200 to +1000°C 0.001 °C
0.2°C
1.0°C
R
0 to +1768°C 0.1 °C
0.6°C
1.8°C
S
0 to +1768°C 0.1 °C
0.6°C
1.8°C
B +350 to +1820°C 0.1 °C
0.6°C
1.8°C
FUNCTION
DCV, DCI, Ω (<10M),
THERMOCOUPLE,
THERMISTOR
Temperature
Coefficient
0°–18°C & 28°–50°C
0.03°C/°C
0.03°C/°C
0.03°C/°C
0.03°C/°C
0.03°C/°C
0.03°C/°C
0.03°C/°C
0.03°C/°C
4-Wire RTD:
(100Ω platinum [PT100], D100, F100, PT385, PT3916, or user type. Offset Compensation On)
–200° to
630°C 0.01 °C
0.06°C
0.003°C/°C
Thermistor: (2.2kΩ, 5kΩ, and 10kΩ)20
–80° to
150°C 0.01 °C
0.08°C
READINGS/s
5
(4)
35
(28)
45
(36)
150 (120)
300 (240)
500 (400)
2500 (2000)
PLCs
10
1
1
0.1
0.1
0.1
0.01
4WΩ (<10M)
6.5
6.5
5.5
1.4
15
33
(1.1)
(12)
(25)
10
1
0.1
4WΩ OCOMP, RTD
6.5
6.5
5.5
0.9
8
18
(0.7)
(6.4)
(14.4)
10
1
0.1
CHANNEL (RATIO),
CHANNEL (AVG)
6.5
6.5
5.5
2.5
15
25
(2)
(12)
(20)
10
1
0.1
0.002°C/°C
MULTIPLE CHANNELS, INTO AND OUT OF MEMORY TO GPIB
DC SYSTEM SPEEDS
7703 Scanning DCV
7703 Scanning DCV with Limits or Time Stamp On
7703 Scanning DCV alternating 2W
7702 Scanning DCV
7700 and 7708 Scanning Temperature (T/C)
RANGE CHANGES: 50/s (42/s).
FUNCTION CHANGES: 50/s (42/s).
AUTORANGE TIME: <30ms.
ASCII READINGS TO RS-232 (19.2K BAUD): 55/s.
MAX. INTERNAL TRIGGER RATE: 2000/s.
MAX. EXTERNAL TRIGGER RATE: 375/s.
1.888.KEITHLEY
DIGITS
6.5
6.5
6.5
5.5
5.5
5.5
4.5
Channels/s
200/s
180/s
58/s
65/s
50/s
(U.S. only)
www.keithley.com
A
G R E A T E R
M E A S U R E
O F
C O N F I D E N C E
SWITCH/MEASURE SYSTEMS
FUNCTION
TEST
CURRENT ±5%
OR BURDEN
VOLTAGE
Model 2700 and 2750 Condensed Specifications
CONDITIONS: MED (1 PLC) or 10 PLC or MED (1 PLC) with Digital Filter of 10
02 Switch/Measure Systems
11/16/01
1:35 PM
Page 56
Multimeter/Data Acquisition/
Switch Systems
2700
2750
AC SPECIFICATIONS
Model 2700 and 2750 Condensed Specifications
FUNCTION
VOLTAGE
CURRENT
RANGE
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
750.000 V
RESOLUTION
0.1 µV
1.0 µV
10 µV
100 µV
1.0 µV
1.000000 A
3.00000 A
3 Hz–
10 Hz
0.35 + 0.03
1 Year
0.35 + 0.03
0.06 + 0.03
0.12 + 0.05
0.6 + 0.08
4.0 + 0.5
(TEMP. COEFF.)/°C
0.035 + .003
0.005 + .003
0.006 + .005
0.01 + .006
0.03 + .01
3 Hz–10 Hz
0.30 + 0.04
0.35 + 0.06
0.035 + 0.006
10 Hz–5 kHz
0.10 + 0.04
0.15 + 0.06
0.015 + 0.006
1.0 µA
10 µA
90 Day/1 Yr.
(TEMP. COEFF.)/°C
FREQUENCY
AND PERIOD
100 mV
to
750 V
0.333 ppm
3.33 ppm
33.3 ppm
ACCURACY: ±(% of reading + % of range), 23°C ± 5°C
10 Hz–
20 kHz–
50 kHz–
20 kHz
50 kHz
100 kHz
0.05 + 0.03
0.11 + 0.05
0.6 + 0.08
CALIBRATION
CYCLE
90 Days
90 Day/ 1 Yr.
(3 Hz–500 kHz) (333 ms–2 µs)
100 ppm + 0.333 ppm (SLOW, 1 s gate)
100 ppm + 3.33 ppm (MED, 100 ms gate)
100 ppm + 33.3 ppm (FAST, 10 ms gate)
AC MEASUREMENT SPEEDS
GENERAL
EXPANSION SLOTS: 5 (Model 2750); 2 (Model 2700).
MEMORY SIZE: 110,000 readings (Model 2750); 55,000 readings (Model 2700).
POWER SUPPLY: 100V / 120V / 220V / 240V ±10%.
LINE FREQUENCY: 45Hz to 66Hz and 360Hz to 440Hz, automatically sensed at
power-up.
POWER CONSUMPTION: 80VA (Model 2750); 28VA (Model 2700).
OPERATING ENVIRONMENT: Specified for 0°C to 50°C. Specified to 80% RH at
35°C.
BATTERY: Lithium battery-backed memory, 3 years @ 23°C.
WARRANTY: 3 years.
EMC: Conforms to European Union Directive 89/336/EEC EN61326-1.
SAFETY: Conforms to European Union Directive 73/23/EEC EN61010-1, CAT I.
DIMENSIONS (Model 2700): 89mm high × 213mm wide × 370mm deep (3.5 in
× 8.39 in × 14.563 in).
DIMENSIONS (Model 2750):
RACK MOUNTING: 89mm high × 485mm wide × 370mm deep (3.5 in × 19 in
× 14.563 in).
BENCH CONFIGURATION (WITH HANDLE AND FEET): 104mm high ×
485mm wide × 370mm deep (4.125 in × 19 in × 14.563 in).
SHIPPING WEIGHT: 13kg (28 lbs.) (Model 2750); 6.5 kg (14 lbs) (Model 2700).
SINGLE CHANNEL, 60Hz (50Hz) Operation
Function
ACV, ACI
Digits
6.5
6.5
6.5
6.5
FREQUENCY,
PERIOD
6.5
5.5
4.5
4.5
Readings/s
2s/Reading
1.4 (1.1)
4.8 (4)
40 (32)
Rate
SLOW
MED
MED
FAST
Bandwidth
3 Hz–300 kHz
30 Hz–300 kHz
30 Hz–300 kHz
300 Hz–300 kHz
1 (1)
9 (9)
35 (35)
65 (65)
SLOW
MED
FAST
FAST
3 Hz–300 kHz
30 Hz–300 kHz
300 Hz–300 kHz
300 Hz–300 kHz
MULTIPLE CHANNEL
7703 SCANNING ACV: 180/s.
AC SYSTEM SPEEDS
SWITCH/MEASURE SYSTEMS
RANGE CHANGES: 4/s (3/s).
FUNCTION CHANGES: 4/s (3/s).
AUTORANGE TIME: < 3s.
ASCII READINGS TO RS-232 (19.2K baud): 50/s.
MAX. INTERNAL TRIGGER RATE: 300/s.
MAX. EXTERNAL TRIGGER RATE: 250/s.