ETC PL320QMD-E

CYAN MAGENTA YELLOW BLACK
Technical Specifications
Technical Specifications continued
MODEL SELECTOR GUIDE
MAIN OUTPUT(S)
Output Range:
MODEL
MAIN OUTPUT(S)
LOGIC
OUTPUT
INTERFACES
GPIB
PL310
0 - 32V at 0 - 1A
PL320
0 - 32V at 0 - 2A
PL154
0 - 15.5V at 0 - 4A
PL330
0 - 32V at 0 - 3A
PL310QMD
2 x 0 - 32V at 0 - 1A
ARC
Output Voltage
Setting:
Output Current
Setting:
Output Mode:
or 0 - 64V at 0 - 1A
or 0 - ±32V at 0 - 1A
2 x 0 - 32V at 0 - 2A
or 0 - 32V at 0 - 4A
or 0 - 64V at 0 - 2A
or 0 - ±32V at 0 - 2A
PL330QMD
2 x 0 - 32V at 0 - 3A
or 0 - 32V at 0 - 6A
or 0 - 64V at 0 - 3A
or 0 - ±32V at 0 - 3A
PL310QMT
2 x 0 - 32V at 0 - 1A
5V at 1.5A
or 0 - 32V at 0 - 2A
or 0 - 64V at 0 - 1A
or 0 - ±32V at 0 - 1A
PL320QMT
2 x 0 - 32V at 0 - 2A
LOGIC OUTPUT - PL310QMT
4 - 6V at
0.1 - 4A
or 0 - 32V at 0 - 4A
Output Voltage:
Max.Output Current:
Output Terminal:
Output Protection:
or 0 - 64V at 0 - 2A
or 0 - ±32V at 0 - 2A
PL330QMT
2 x 0 - 32V at 0 - 3A
Load Regulation:
Line Regulation:
4 - 6V at
0.1 - 7A
or 0 - 32V at 0 - 6A
By single logarithmic control.
The power supply operates in constant current or constant
voltage modes with automatic cross-over. Decimal points
flash to indicate constant current mode.
Isolated, True parallel, Series or Series Tracking via front
panel switches.
Configuration
Selection:
(QMD and QMT only)
Output Switch:
Isolates the output and permits voltage and current limits to
be set up before connecting the load.
Output Terminals:
4mm terminals on 19mm (0.75") spacing.
Output Impedance: Constant Voltage: Typically <5mΩ at 1kHz
Constant Current: Typically 50kΩ with voltage limit at maximum
Output Protection:
Up to maximum output voltage +20 Volts forward; diode
clamped for reverse voltages and up to 3A reverse current.
Load Regulation:
<0.01% of maximum output for 90% load change
Line Regulation:
<0.01% of maximum output for 10% line voltage change
Remote Sense:
Eliminates up to 0.5V drop per lead.
Ripple and Noise:
Typically <1mV rms
Transient Response: <20usec to within 50mV of setting for 90% load change
Temperature
Coefficient:
Typically <100ppm/oC
Meter Type:
Dual 3.75 digit (4095 count) with 12.5mm (0.5") LEDS,
(scale length increased to 8190 on PL330QMD/QMT).
Reading rate 4 per second.
Meter Resolution:
Voltage: 10mV over the entire range
Current: 1mA over the entire range
Meter Accuracy:
Voltage: ±(0.1% of reading + 1 digit)
Current: ±(0.3% of reading + 1 digit)
Current Meter
Nominally 20ms switchable to 2 sec for averaging of rapidly
Damping:
varying loads
or 0 - 32V at 0 - 2A
PL320QMD
0 - 32 Volts nominal; 0 - 15.5V (PL154).
0 - 1.1A nominal (PL310);
0 - 2.1A nominal (PL320);
0 - 3.1A nominal (PL330);
0 - 4A nominal (PL154).
By coarse and fine controls; resolution better than 5mV
across the range.
Fixed 5 V ± 0⋅1 V.
>1⋅5 Amps.
4mm terminals on 19mm (0⋅75") spacing.
Output will withstand up to 16 V forward voltage. Diode
clamped for reverse voltages and up to 3 Amps reverse current.
< 0⋅3% for 50% load change.
< 0⋅1% for 10% line change.
LOGIC OUTPUT - PL320QMT
or 0 - 64V at 0 - 3A
or 0 - ±32V at 0 - 3A
PL330P
0 - 32V at 0 - 3A
Yes
Yes
PL330DP
2 x 0 - 32V at 0 - 3A
Yes
Yes
PL330TP
2 x 0 - 32V at 0 - 3A
Yes
Yes
4 - 6V at 1 - 7A
Models illustrated within this brochure:Front cover - PL330, PL320QMD, PL310QMT
Inside pages - PL330QMT, PL330DP
Output Voltage
Range:
Output Current:
Output Switch:
Output Terminals:
Over-Voltage
Protection:
Output Protection:
4 to 6 Volts
0.1 to 4 Amps.
Isolates the output and permits output voltage to be set before connecting the load.
4mm terminals on 19mm (0.75") spacing.
Above 7 Volts
Clamped by the over-voltage protection circuit for forward
voltages over 7 Volts and up to 1 Amp forward current. Diode
clamped for reverse voltages and up to 3 Amps reverse current.
Load Regulation:
<0.01% of maximum output for 90% load change
Line Regulation:
<0.01% of maximum output for 10% line voltage change
Remote Sense:
Eliminates up to 0.5V drop per lead.
Ripple and Noise:
Typically <1mV rms
Transient Response: <20usec to within 50mV of setting for 90% load change
Temperature
Coefficient:
Typically <100ppm/oC
Voltage Setting
Accuracy:
± 0.1V
LOGIC OUTPUT - PL330QMT & PL330TP
Output Voltage
Range:
Output Current:
Output Switch:
Output Terminals:
Over-Voltage
Protection:
Output Protection:
4 to 6 Volts
0.1 to 7 Amps.
Isolates the output and permits output voltage to be set before connecting the load.
4mm terminals on 19mm (0.75") spacing.
Above 7 Volts
Clamped by the over-voltage protection circuit for forward
voltages over 7 Volts and up to 1 Amp forward current. Diode clamped for reverse voltages and up to 3 Amps reverse
current.
Load Regulation:
<0.01% of maximum output for 90% load change
Line Regulation:
<0.01% of maximum output for 10% line voltage change
Remote Sense:
Eliminates up to 0.5V drop per lead.
Ripple and Noise:
Typically <1mV rms
Transient Response: <20usec to within 50mV of setting for 90% load change
Temperature
Coefficient:
Typically <100ppm/oC
Meter Type:
3.75 digit (4095 count) with 12.5mm (0.5") LEDs. Reading
rate 4 per second.
Meter Resolution:
Voltage: 10mV
Current: 10mA
Meter Accuracy:
PL-P MODELS, ADDITIONAL SPECIFICATIONS
MAIN OUTPUT(S) - REMOTE OPERATION
12 bit resolution (10mV steps)
12 bit resolution (1mA steps)
Voltage: ±(0.1% + 10mV)
Current: ±(0.2% + 2mA)
Output Switch:
Electronic by interface command (front panel output
switches must be set to ON)
Readback Resolution:Voltage: 10mV over the entire range
Current: 1mA over the entire range
Readback Accuracy: Voltage: ±(0.1% of reading + 1 digit)
Current: ±(0.3% of reading + 1 digit)
Current Meter
Nominally 20ms switchable to 2 sec and back by remote
Damping:
commands
PL & PL-P Series
GENERAL
Environmental:
Weight:
Internally set for 110, 120, 220 or 240VAC 50/60Hz
±10% of voltage setting
Single Dual
Triple
30V/1A
75VA 150VA 300VA
15V/4A & 30V/2A150VA 300VA 450VA
30V/3A
225VA 450VA 600VA
Operating Range: 5oC to 40oC, 20% to 80% RH
Storage Range: -20oC to +60oC
Single Dual
Triple
30V/1A PL
4.0kg 8.0kg 11.5kg
15V/4A PL
5.0kg
30V/2A PL
5.0kg 9.5kg 13.5kg
30V/3A PL
6.0kg 12.0kg 15.5kg
30V/3A PL-P
6.5kg 12.5kg 16.0kg
Size:
PL310, PL154, &
PL320:
155mm(W) x 170mm(H) x 265mm(D)
PL330:
155mm(W) x 170mm(H) x 300mm(D)
PL330P:
207mm(W) x 170mm(H) x 300mm(D)
PL310QMD, PL310QMT, &
PL320QMD:
350mm(W) x 170mm(H) x 265mm(D)
PL330QMD, &
PL330DP:
350mm(W) x 170mm(H) x 300mm(D)
PL320QMT:
425mm(W) x 170mm(H) x 265mm(D)
PL330QMT, &
PL330TP:
425mm(W) x 170mm(H) x 300mm(D)
Rack Mount Option: 19 inch 4U mount for two PL330P or one PL330DP/TP.
Anti-tamper cover for security.
Electrical Safety:
Designed and manufactured to comply with IEC 348 and
IEC1010-1. Full safety sockets available to special order.
EMC:
Designed and manufactured to comply with EN50081–1
and EN50082–1.
Laboratory Power Supplies
LOGIC OUTPUT (PL330TP) - REMOTE OPERATION
Standard and Programmable models, 35 watts to 240 watts
Output Voltage
Range:
Output Current:
Setting Accuracy:
Output Switch:
4 to 6 Volts in 10mV steps
1 to 7 Amps in approximate 1A steps
Voltage: ±(0.2% + 10mV)
Electronic by interface command (front panel output switch
must be set to ON)
Readback Resolution:Current: 10mA
Readback Accuracy: Current: ±(0.5% of reading + 1 digit )
THURLBY THANDAR INSTRUMENTS
Power Requirements:
Input Voltage:
Input Voltage Range:
Power Consumption:
Voltage: ±(0.2% of reading + 1 digit )
Current: ±(0.5% of reading + 1 digit )
Remote programmable versions in the range feature full control, read back and
status reporting via the GPIB and RS232 interfaces. The GPIB interface conforms to the IEEE 488.1 and 488.2 standards and the RS232 interface is fully
compatible with the Thurlby-Thandar Addressable RS232 Chain (ARC) standard.
8 rear panel DIP switches are used to specify baud rate, bus address and active
interface (GPIB or RS232). Remote/Local operation is selected by a front panel
switch.
LOCAL OPERATION
For a programmable instrument operated in local state, all capabilities and specifications remain unchanged from those of a standard instrument .
REMOTE OPERATION
With the instrument switched to the remote state, all voltage and current adjustment controls become inoperative and commands received over the active interface will be parsed and executed.
Output Voltage
Setting:
Output Current
Setting:
Setting Accuracy:
REMOTE CONTROL INTERFACES - PL-P MODELS
Both interfaces feature full control, readback and status reporting.
RS232:
Variable Baud rate (9600 maximum), 9 pin D-connector (female). Fully compatible with ARC (Addressable RS232
Chain) system.
GPIB:
Conforming with IEEE-488.1 and IEEE-488.2
Address Selection: By rear panel DIP switch.
Remote/Local
Operation:
Selected by front panel switch.
Remote Command Response Time:
Interface:
<15 ms (single command, input buffer empty).
Output Voltage - Up:Time constant typically 2ms, e.g. 10ms to settle within 1%
of a step change, 15ms to settle within 0.1%.
Output Voltage - Down:
Time constant determined by the discharge of the power
supply output capacitor (47uF). Typically <10ms to settle
within 1% for a 10V step change at 50mA load current; typically <200ms to settle within 1% at zero load.
Output Current:
Typically 50ms to settle within 10mA for a 1A change.
Thurlby Thandar Instruments Ltd. operates a policy of continuous development and
reserves the right to alter specifications without prior notice.
Designed and built in the U.K. by:
Thurlby Thandar Instruments Ltd.
Glebe Road, Huntingdon. Cambs. PE18 7DX England
Tel: 01480 412451 Fax: 01480 450409 e-mail: [email protected]
82100-0046
CYAN MAGENTA YELLOW BLACK
Thurlby Thandar PL-P series
programmable power at surprisingly low cost
Thurlby Thandar PL series
the Premier range of laboratory power supplies
Precision with convenience
The Thurlby Thandar PL series of laboratory bench power supplies has established itself in many countries as the
“premier” range.
High resolution controls enable precise
setting of voltage and current levels
whilst high accuracy digital meters provide clear, unambiguous readings.
All of the many features of these PSUs
have been carefully designed to give the
user not just greater precision, but
greater clarity, more control and unrivalled ease of use.
The range has now been improved and
extended to offer even better performance and choice.
Greater resolution and
control
The PL series sets the standard for simple and comprehensive control. Voltages
are set with coarse and fine controls for
speed with precision. Currents are set with
a semi-logarithmic control for increased
resolution at low current levels.
The DC output switch enables voltage
and current levels to be set before the
load is connected.
With the output switch “off” the the current limit set point is displayed. With the
output switch “on” the actual output current flowing is displayed.
This invaluable feature allows delicate
circuits to be protected by accurately setting the current limit level (down to a few
milliamps if necessary) before connecting
the circuit under test.
Safety and protection
PL series PSUs are designed and built to
meet the stringent requirements of
IEC348 and IEC1010.
All outputs are fully protected against
short circuit, reverse voltage and reverse
currents.
PL series units incorporate digital meters
with a 3.75 digit scale length (4095
counts) to provide greater accuracy and
resolution than other PSUs.
Large and bright LEDs give a clear and
unambiguous reading. An update rate of
4 per second provides near instantaneous response.
Separate meters are used for voltage
and current, eliminating the need for meter function switches with their attendant
problems of misinterpretation.
A damping switch for the current meter
simplifies measurements on rapidly varying loads.
PL series units incorporate integrated
band-gap reference diodes as the basis
for stabilisation of both voltage and current.
Remote sense terminals enable the precision to be maintained at high currents
by eliminating the effects of connection
lead resistance.
Without remote sense lead resistance of
just a few tens of milliohms can seriously
degrade regulation and produce misleading results. (Two cables of 0.05Ω
each will drop a total of 0.3V at 3 Amps.)
PL-P series
The 32V-1A, 32V-2A and 32V-3A supplies are each available as a dual unit incorporating push button selection of four
different modes of operation.
The Thurlby Thandar PL-P series offers
a high performance fully programmable
power supply system at low cost.
Based around the 32V-3A versions of the
standard PL series, the PL-P models include single, dual and triple output units
suitable for bench or rack mounting.
When not connected to the bus, these
PSUs can be operated exactly as a standard PL series PSU.
Isolated
True Parallel
Completely independent operation of
each supply.
Converts the Master unit into a 2 Amp,
4 Amp or 6 Amp supply respectively.
GPIB and RS-232 (ARC)
interfaces
Digital accuracy and
convenience
Remote sense for
precision at high currents
Quad-Mode Dual versions
Series
Series Tracking
Internal linking of the two supplies providing up to 64 Volts.
The Master unit voltage control sets up
equal voltages on both supplies.
Quad Mode Triple versions
●
●
●
●
●
●
●
●
●
●
●
Simultaneous digital metering of voltage and current.
True constant voltage or constant current operation.
Twin 3.75 digit meters with large LED displays.
0.1% accuracy; 0.01 Volts and 0.001 Amps resolution.
Excellent stability, resolution and setting accuracy.
DC output switches, automatic mode indication.
Precise control and monitoring of current limit settings.
Remote sense facility for high-current precision.
Current meter damping switch for fluctuating currents.
Parallel and tracking modes on QMD & QMT models.
High current “logic supply” output on QMT models.
A wide range of models
The PL series includes single, dual and
triple output models from 35 Watts up to
240 Watts.
See the model selector guide for a summary of voltage and current combinations.
The PL series is part of a wider range of
bench PSUs from Thurlby Thandar
which includes models with current capabilities up to 20 Amps.
Each of the quad-mode dual models is alternatively available as a triple supply incorporating one further independent
output.
This is a higher current 5 Volt output intended for powering logic circuits.
The current rating and sophistication of
the logic output varies according to the
model as follows:
PL310QMT
Fixed 5V supply at 1.5A maximum.
Full short-circuit protection.
PL320QMT
Variable output voltage (4V to 6V) and
variable current limit (0.1A to 4A).
Calibrated voltage control.
Remote sense terminals, DC output
switch, over-voltage trip.
PL330QMT
Variable output voltage (4V to 6V) and
variable current limit (0.1A to 7A).
Digital meter for current measurement
and voltage setting.
Remote sense terminals, DC output
switch, over-voltage trip.
Each PL-P series supply is fitted with
both a GPIB (IEEE-488) interface and an
ARC (addressable RS232) interface as
standard.
Both interfaces provide full bus control of
voltage and current settings along with
full readback of actual current and voltage levels.
The GPIB interface conforms fully with
IEEE-488.2 as well as IEEE-488.1. The
ARC interface can be used as a conventional RS-232 interface or as part of a
multi-instrument ARC system.
On dual and triple output models a single
bus address controls all outputs.
Simple and consistent
control
PL-P series supplies use simple and
consistent command structures which
make programming particularly easy regardless of which interface is used.
A National Instruments LabWindows*
device driver is available as an option.
ARC, an exclusive Thurlby
Thandar innovation
ARC stands for “Addressable RS-232
Chain” and is a low-cost system for linking instruments together so that they can
be controlled and monitored by a personal computer.
The ARC interface is an extension of the
industry standard RS-232 interface and
is exclusive to Thurlby-Thandar instruments.
It differs from conventional RS-232 in
that it allows multiple instruments (up to
32) to be controlled using the normal
RS-232 or RS-422/423 port of a PC.
ARC provides a low-cost alternative to
GPIB which utilises lower cost instruments, inexpensive cables, and can be
controlled by any personal computer
without the need for a special interface
card or special software.
* LabWindows is a trademark of National
Instruments Corporation.
Fully isolated outputs for
maximum flexibility
Each output is fully floating and is optoisolated from the bus interfaces.
Outputs can be linked in series or parallel
to produce higher voltages or higher currents as required.
High resolution control
and readback
Voltage and current levels can be set via
the bus to a resolution of 10mV and 1mA
for each main output.
The 7 Amp logic output of the PL330TP
can also be set to a resolution of 10mV
but the current control resolution is limited to 1 Amp steps.
Each main output can be read back via
the bus to a resolution of 10mV and
1mA.
●
●
●
●
●
●
●
Full bus control and readback of voltage and current.
GPIB interface conforms to IEEE-488.2.
ARC (Addressable RS-232) interface for low-cost
PC based control.
Can be operated as a conventional bench PSU.
Single, dual and triple output models available.
Triple output model incorporates fully controllable
high current logic output.
Rack mounting kit (4U) available for all models.
CYAN MAGENTA YELLOW BLACK
Thurlby Thandar PL-P series
programmable power at surprisingly low cost
Thurlby Thandar PL series
the Premier range of laboratory power supplies
Precision with convenience
The Thurlby Thandar PL series of laboratory bench power supplies has established itself in many countries as the
“premier” range.
High resolution controls enable precise
setting of voltage and current levels
whilst high accuracy digital meters provide clear, unambiguous readings.
All of the many features of these PSUs
have been carefully designed to give the
user not just greater precision, but
greater clarity, more control and unrivalled ease of use.
The range has now been improved and
extended to offer even better performance and choice.
Greater resolution and
control
The PL series sets the standard for simple and comprehensive control. Voltages
are set with coarse and fine controls for
speed with precision. Currents are set with
a semi-logarithmic control for increased
resolution at low current levels.
The DC output switch enables voltage
and current levels to be set before the
load is connected.
With the output switch “off” the the current limit set point is displayed. With the
output switch “on” the actual output current flowing is displayed.
This invaluable feature allows delicate
circuits to be protected by accurately setting the current limit level (down to a few
milliamps if necessary) before connecting
the circuit under test.
Safety and protection
PL series PSUs are designed and built to
meet the stringent requirements of
IEC348 and IEC1010.
All outputs are fully protected against
short circuit, reverse voltage and reverse
currents.
PL series units incorporate digital meters
with a 3.75 digit scale length (4095
counts) to provide greater accuracy and
resolution than other PSUs.
Large and bright LEDs give a clear and
unambiguous reading. An update rate of
4 per second provides near instantaneous response.
Separate meters are used for voltage
and current, eliminating the need for meter function switches with their attendant
problems of misinterpretation.
A damping switch for the current meter
simplifies measurements on rapidly varying loads.
PL series units incorporate integrated
band-gap reference diodes as the basis
for stabilisation of both voltage and current.
Remote sense terminals enable the precision to be maintained at high currents
by eliminating the effects of connection
lead resistance.
Without remote sense lead resistance of
just a few tens of milliohms can seriously
degrade regulation and produce misleading results. (Two cables of 0.05Ω
each will drop a total of 0.3V at 3 Amps.)
PL-P series
The 32V-1A, 32V-2A and 32V-3A supplies are each available as a dual unit incorporating push button selection of four
different modes of operation.
The Thurlby Thandar PL-P series offers
a high performance fully programmable
power supply system at low cost.
Based around the 32V-3A versions of the
standard PL series, the PL-P models include single, dual and triple output units
suitable for bench or rack mounting.
When not connected to the bus, these
PSUs can be operated exactly as a standard PL series PSU.
Isolated
True Parallel
Completely independent operation of
each supply.
Converts the Master unit into a 2 Amp,
4 Amp or 6 Amp supply respectively.
GPIB and RS-232 (ARC)
interfaces
Digital accuracy and
convenience
Remote sense for
precision at high currents
Quad-Mode Dual versions
Series
Series Tracking
Internal linking of the two supplies providing up to 64 Volts.
The Master unit voltage control sets up
equal voltages on both supplies.
Quad Mode Triple versions
●
●
●
●
●
●
●
●
●
●
●
Simultaneous digital metering of voltage and current.
True constant voltage or constant current operation.
Twin 3.75 digit meters with large LED displays.
0.1% accuracy; 0.01 Volts and 0.001 Amps resolution.
Excellent stability, resolution and setting accuracy.
DC output switches, automatic mode indication.
Precise control and monitoring of current limit settings.
Remote sense facility for high-current precision.
Current meter damping switch for fluctuating currents.
Parallel and tracking modes on QMD & QMT models.
High current “logic supply” output on QMT models.
A wide range of models
The PL series includes single, dual and
triple output models from 35 Watts up to
240 Watts.
See the model selector guide for a summary of voltage and current combinations.
The PL series is part of a wider range of
bench PSUs from Thurlby Thandar
which includes models with current capabilities up to 20 Amps.
Each of the quad-mode dual models is alternatively available as a triple supply incorporating one further independent
output.
This is a higher current 5 Volt output intended for powering logic circuits.
The current rating and sophistication of
the logic output varies according to the
model as follows:
PL310QMT
Fixed 5V supply at 1.5A maximum.
Full short-circuit protection.
PL320QMT
Variable output voltage (4V to 6V) and
variable current limit (0.1A to 4A).
Calibrated voltage control.
Remote sense terminals, DC output
switch, over-voltage trip.
PL330QMT
Variable output voltage (4V to 6V) and
variable current limit (0.1A to 7A).
Digital meter for current measurement
and voltage setting.
Remote sense terminals, DC output
switch, over-voltage trip.
Each PL-P series supply is fitted with
both a GPIB (IEEE-488) interface and an
ARC (addressable RS232) interface as
standard.
Both interfaces provide full bus control of
voltage and current settings along with
full readback of actual current and voltage levels.
The GPIB interface conforms fully with
IEEE-488.2 as well as IEEE-488.1. The
ARC interface can be used as a conventional RS-232 interface or as part of a
multi-instrument ARC system.
On dual and triple output models a single
bus address controls all outputs.
Simple and consistent
control
PL-P series supplies use simple and
consistent command structures which
make programming particularly easy regardless of which interface is used.
A National Instruments LabWindows*
device driver is available as an option.
ARC, an exclusive Thurlby
Thandar innovation
ARC stands for “Addressable RS-232
Chain” and is a low-cost system for linking instruments together so that they can
be controlled and monitored by a personal computer.
The ARC interface is an extension of the
industry standard RS-232 interface and
is exclusive to Thurlby-Thandar instruments.
It differs from conventional RS-232 in
that it allows multiple instruments (up to
32) to be controlled using the normal
RS-232 or RS-422/423 port of a PC.
ARC provides a low-cost alternative to
GPIB which utilises lower cost instruments, inexpensive cables, and can be
controlled by any personal computer
without the need for a special interface
card or special software.
* LabWindows is a trademark of National
Instruments Corporation.
Fully isolated outputs for
maximum flexibility
Each output is fully floating and is optoisolated from the bus interfaces.
Outputs can be linked in series or parallel
to produce higher voltages or higher currents as required.
High resolution control
and readback
Voltage and current levels can be set via
the bus to a resolution of 10mV and 1mA
for each main output.
The 7 Amp logic output of the PL330TP
can also be set to a resolution of 10mV
but the current control resolution is limited to 1 Amp steps.
Each main output can be read back via
the bus to a resolution of 10mV and
1mA.
●
●
●
●
●
●
●
Full bus control and readback of voltage and current.
GPIB interface conforms to IEEE-488.2.
ARC (Addressable RS-232) interface for low-cost
PC based control.
Can be operated as a conventional bench PSU.
Single, dual and triple output models available.
Triple output model incorporates fully controllable
high current logic output.
Rack mounting kit (4U) available for all models.
CYAN MAGENTA YELLOW BLACK
Thurlby Thandar PL-P series
programmable power at surprisingly low cost
Thurlby Thandar PL series
the Premier range of laboratory power supplies
Precision with convenience
The Thurlby Thandar PL series of laboratory bench power supplies has established itself in many countries as the
“premier” range.
High resolution controls enable precise
setting of voltage and current levels
whilst high accuracy digital meters provide clear, unambiguous readings.
All of the many features of these PSUs
have been carefully designed to give the
user not just greater precision, but
greater clarity, more control and unrivalled ease of use.
The range has now been improved and
extended to offer even better performance and choice.
Greater resolution and
control
The PL series sets the standard for simple and comprehensive control. Voltages
are set with coarse and fine controls for
speed with precision. Currents are set with
a semi-logarithmic control for increased
resolution at low current levels.
The DC output switch enables voltage
and current levels to be set before the
load is connected.
With the output switch “off” the the current limit set point is displayed. With the
output switch “on” the actual output current flowing is displayed.
This invaluable feature allows delicate
circuits to be protected by accurately setting the current limit level (down to a few
milliamps if necessary) before connecting
the circuit under test.
Safety and protection
PL series PSUs are designed and built to
meet the stringent requirements of
IEC348 and IEC1010.
All outputs are fully protected against
short circuit, reverse voltage and reverse
currents.
PL series units incorporate digital meters
with a 3.75 digit scale length (4095
counts) to provide greater accuracy and
resolution than other PSUs.
Large and bright LEDs give a clear and
unambiguous reading. An update rate of
4 per second provides near instantaneous response.
Separate meters are used for voltage
and current, eliminating the need for meter function switches with their attendant
problems of misinterpretation.
A damping switch for the current meter
simplifies measurements on rapidly varying loads.
PL series units incorporate integrated
band-gap reference diodes as the basis
for stabilisation of both voltage and current.
Remote sense terminals enable the precision to be maintained at high currents
by eliminating the effects of connection
lead resistance.
Without remote sense lead resistance of
just a few tens of milliohms can seriously
degrade regulation and produce misleading results. (Two cables of 0.05Ω
each will drop a total of 0.3V at 3 Amps.)
PL-P series
The 32V-1A, 32V-2A and 32V-3A supplies are each available as a dual unit incorporating push button selection of four
different modes of operation.
The Thurlby Thandar PL-P series offers
a high performance fully programmable
power supply system at low cost.
Based around the 32V-3A versions of the
standard PL series, the PL-P models include single, dual and triple output units
suitable for bench or rack mounting.
When not connected to the bus, these
PSUs can be operated exactly as a standard PL series PSU.
Isolated
True Parallel
Completely independent operation of
each supply.
Converts the Master unit into a 2 Amp,
4 Amp or 6 Amp supply respectively.
GPIB and RS-232 (ARC)
interfaces
Digital accuracy and
convenience
Remote sense for
precision at high currents
Quad-Mode Dual versions
Series
Series Tracking
Internal linking of the two supplies providing up to 64 Volts.
The Master unit voltage control sets up
equal voltages on both supplies.
Quad Mode Triple versions
●
●
●
●
●
●
●
●
●
●
●
Simultaneous digital metering of voltage and current.
True constant voltage or constant current operation.
Twin 3.75 digit meters with large LED displays.
0.1% accuracy; 0.01 Volts and 0.001 Amps resolution.
Excellent stability, resolution and setting accuracy.
DC output switches, automatic mode indication.
Precise control and monitoring of current limit settings.
Remote sense facility for high-current precision.
Current meter damping switch for fluctuating currents.
Parallel and tracking modes on QMD & QMT models.
High current “logic supply” output on QMT models.
A wide range of models
The PL series includes single, dual and
triple output models from 35 Watts up to
240 Watts.
See the model selector guide for a summary of voltage and current combinations.
The PL series is part of a wider range of
bench PSUs from Thurlby Thandar
which includes models with current capabilities up to 20 Amps.
Each of the quad-mode dual models is alternatively available as a triple supply incorporating one further independent
output.
This is a higher current 5 Volt output intended for powering logic circuits.
The current rating and sophistication of
the logic output varies according to the
model as follows:
PL310QMT
Fixed 5V supply at 1.5A maximum.
Full short-circuit protection.
PL320QMT
Variable output voltage (4V to 6V) and
variable current limit (0.1A to 4A).
Calibrated voltage control.
Remote sense terminals, DC output
switch, over-voltage trip.
PL330QMT
Variable output voltage (4V to 6V) and
variable current limit (0.1A to 7A).
Digital meter for current measurement
and voltage setting.
Remote sense terminals, DC output
switch, over-voltage trip.
Each PL-P series supply is fitted with
both a GPIB (IEEE-488) interface and an
ARC (addressable RS232) interface as
standard.
Both interfaces provide full bus control of
voltage and current settings along with
full readback of actual current and voltage levels.
The GPIB interface conforms fully with
IEEE-488.2 as well as IEEE-488.1. The
ARC interface can be used as a conventional RS-232 interface or as part of a
multi-instrument ARC system.
On dual and triple output models a single
bus address controls all outputs.
Simple and consistent
control
PL-P series supplies use simple and
consistent command structures which
make programming particularly easy regardless of which interface is used.
A National Instruments LabWindows*
device driver is available as an option.
ARC, an exclusive Thurlby
Thandar innovation
ARC stands for “Addressable RS-232
Chain” and is a low-cost system for linking instruments together so that they can
be controlled and monitored by a personal computer.
The ARC interface is an extension of the
industry standard RS-232 interface and
is exclusive to Thurlby-Thandar instruments.
It differs from conventional RS-232 in
that it allows multiple instruments (up to
32) to be controlled using the normal
RS-232 or RS-422/423 port of a PC.
ARC provides a low-cost alternative to
GPIB which utilises lower cost instruments, inexpensive cables, and can be
controlled by any personal computer
without the need for a special interface
card or special software.
* LabWindows is a trademark of National
Instruments Corporation.
Fully isolated outputs for
maximum flexibility
Each output is fully floating and is optoisolated from the bus interfaces.
Outputs can be linked in series or parallel
to produce higher voltages or higher currents as required.
High resolution control
and readback
Voltage and current levels can be set via
the bus to a resolution of 10mV and 1mA
for each main output.
The 7 Amp logic output of the PL330TP
can also be set to a resolution of 10mV
but the current control resolution is limited to 1 Amp steps.
Each main output can be read back via
the bus to a resolution of 10mV and
1mA.
●
●
●
●
●
●
●
Full bus control and readback of voltage and current.
GPIB interface conforms to IEEE-488.2.
ARC (Addressable RS-232) interface for low-cost
PC based control.
Can be operated as a conventional bench PSU.
Single, dual and triple output models available.
Triple output model incorporates fully controllable
high current logic output.
Rack mounting kit (4U) available for all models.
CYAN MAGENTA YELLOW BLACK
Technical Specifications
Technical Specifications continued
MODEL SELECTOR GUIDE
MAIN OUTPUT(S)
Output Range:
MODEL
MAIN OUTPUT(S)
LOGIC
OUTPUT
INTERFACES
GPIB
PL310
0 - 32V at 0 - 1A
PL320
0 - 32V at 0 - 2A
PL154
0 - 15.5V at 0 - 4A
PL330
0 - 32V at 0 - 3A
PL310QMD
2 x 0 - 32V at 0 - 1A
ARC
Output Voltage
Setting:
Output Current
Setting:
Output Mode:
or 0 - 64V at 0 - 1A
or 0 - ±32V at 0 - 1A
2 x 0 - 32V at 0 - 2A
or 0 - 32V at 0 - 4A
or 0 - 64V at 0 - 2A
or 0 - ±32V at 0 - 2A
PL330QMD
2 x 0 - 32V at 0 - 3A
or 0 - 32V at 0 - 6A
or 0 - 64V at 0 - 3A
or 0 - ±32V at 0 - 3A
PL310QMT
2 x 0 - 32V at 0 - 1A
5V at 1.5A
or 0 - 32V at 0 - 2A
or 0 - 64V at 0 - 1A
or 0 - ±32V at 0 - 1A
PL320QMT
2 x 0 - 32V at 0 - 2A
LOGIC OUTPUT - PL310QMT
4 - 6V at
0.1 - 4A
or 0 - 32V at 0 - 4A
Output Voltage:
Max.Output Current:
Output Terminal:
Output Protection:
or 0 - 64V at 0 - 2A
or 0 - ±32V at 0 - 2A
PL330QMT
2 x 0 - 32V at 0 - 3A
Load Regulation:
Line Regulation:
4 - 6V at
0.1 - 7A
or 0 - 32V at 0 - 6A
By single logarithmic control.
The power supply operates in constant current or constant
voltage modes with automatic cross-over. Decimal points
flash to indicate constant current mode.
Isolated, True parallel, Series or Series Tracking via front
panel switches.
Configuration
Selection:
(QMD and QMT only)
Output Switch:
Isolates the output and permits voltage and current limits to
be set up before connecting the load.
Output Terminals:
4mm terminals on 19mm (0.75") spacing.
Output Impedance: Constant Voltage: Typically <5mΩ at 1kHz
Constant Current: Typically 50kΩ with voltage limit at maximum
Output Protection:
Up to maximum output voltage +20 Volts forward; diode
clamped for reverse voltages and up to 3A reverse current.
Load Regulation:
<0.01% of maximum output for 90% load change
Line Regulation:
<0.01% of maximum output for 10% line voltage change
Remote Sense:
Eliminates up to 0.5V drop per lead.
Ripple and Noise:
Typically <1mV rms
Transient Response: <20usec to within 50mV of setting for 90% load change
Temperature
Coefficient:
Typically <100ppm/oC
Meter Type:
Dual 3.75 digit (4095 count) with 12.5mm (0.5") LEDS,
(scale length increased to 8190 on PL330QMD/QMT).
Reading rate 4 per second.
Meter Resolution:
Voltage: 10mV over the entire range
Current: 1mA over the entire range
Meter Accuracy:
Voltage: ±(0.1% of reading + 1 digit)
Current: ±(0.3% of reading + 1 digit)
Current Meter
Nominally 20ms switchable to 2 sec for averaging of rapidly
Damping:
varying loads
or 0 - 32V at 0 - 2A
PL320QMD
0 - 32 Volts nominal; 0 - 15.5V (PL154).
0 - 1.1A nominal (PL310);
0 - 2.1A nominal (PL320);
0 - 3.1A nominal (PL330);
0 - 4A nominal (PL154).
By coarse and fine controls; resolution better than 5mV
across the range.
Fixed 5 V ± 0⋅1 V.
>1⋅5 Amps.
4mm terminals on 19mm (0⋅75") spacing.
Output will withstand up to 16 V forward voltage. Diode
clamped for reverse voltages and up to 3 Amps reverse current.
< 0⋅3% for 50% load change.
< 0⋅1% for 10% line change.
LOGIC OUTPUT - PL320QMT
or 0 - 64V at 0 - 3A
or 0 - ±32V at 0 - 3A
PL330P
0 - 32V at 0 - 3A
Yes
Yes
PL330DP
2 x 0 - 32V at 0 - 3A
Yes
Yes
PL330TP
2 x 0 - 32V at 0 - 3A
Yes
Yes
4 - 6V at 1 - 7A
Models illustrated within this brochure:Front cover - PL330, PL320QMD, PL310QMT
Inside pages - PL330QMT, PL330DP
Output Voltage
Range:
Output Current:
Output Switch:
Output Terminals:
Over-Voltage
Protection:
Output Protection:
4 to 6 Volts
0.1 to 4 Amps.
Isolates the output and permits output voltage to be set before connecting the load.
4mm terminals on 19mm (0.75") spacing.
Above 7 Volts
Clamped by the over-voltage protection circuit for forward
voltages over 7 Volts and up to 1 Amp forward current. Diode
clamped for reverse voltages and up to 3 Amps reverse current.
Load Regulation:
<0.01% of maximum output for 90% load change
Line Regulation:
<0.01% of maximum output for 10% line voltage change
Remote Sense:
Eliminates up to 0.5V drop per lead.
Ripple and Noise:
Typically <1mV rms
Transient Response: <20usec to within 50mV of setting for 90% load change
Temperature
Coefficient:
Typically <100ppm/oC
Voltage Setting
Accuracy:
± 0.1V
LOGIC OUTPUT - PL330QMT & PL330TP
Output Voltage
Range:
Output Current:
Output Switch:
Output Terminals:
Over-Voltage
Protection:
Output Protection:
4 to 6 Volts
0.1 to 7 Amps.
Isolates the output and permits output voltage to be set before connecting the load.
4mm terminals on 19mm (0.75") spacing.
Above 7 Volts
Clamped by the over-voltage protection circuit for forward
voltages over 7 Volts and up to 1 Amp forward current. Diode clamped for reverse voltages and up to 3 Amps reverse
current.
Load Regulation:
<0.01% of maximum output for 90% load change
Line Regulation:
<0.01% of maximum output for 10% line voltage change
Remote Sense:
Eliminates up to 0.5V drop per lead.
Ripple and Noise:
Typically <1mV rms
Transient Response: <20usec to within 50mV of setting for 90% load change
Temperature
Coefficient:
Typically <100ppm/oC
Meter Type:
3.75 digit (4095 count) with 12.5mm (0.5") LEDs. Reading
rate 4 per second.
Meter Resolution:
Voltage: 10mV
Current: 10mA
Meter Accuracy:
PL-P MODELS, ADDITIONAL SPECIFICATIONS
MAIN OUTPUT(S) - REMOTE OPERATION
12 bit resolution (10mV steps)
12 bit resolution (1mA steps)
Voltage: ±(0.1% + 10mV)
Current: ±(0.2% + 2mA)
Output Switch:
Electronic by interface command (front panel output
switches must be set to ON)
Readback Resolution:Voltage: 10mV over the entire range
Current: 1mA over the entire range
Readback Accuracy: Voltage: ±(0.1% of reading + 1 digit)
Current: ±(0.3% of reading + 1 digit)
Current Meter
Nominally 20ms switchable to 2 sec and back by remote
Damping:
commands
PL & PL-P Series
GENERAL
Environmental:
Weight:
Internally set for 110, 120, 220 or 240VAC 50/60Hz
±10% of voltage setting
Single Dual
Triple
30V/1A
75VA 150VA 300VA
15V/4A & 30V/2A150VA 300VA 450VA
30V/3A
225VA 450VA 600VA
Operating Range: 5oC to 40oC, 20% to 80% RH
Storage Range: -20oC to +60oC
Single Dual
Triple
30V/1A PL
4.0kg 8.0kg 11.5kg
15V/4A PL
5.0kg
30V/2A PL
5.0kg 9.5kg 13.5kg
30V/3A PL
6.0kg 12.0kg 15.5kg
30V/3A PL-P
6.5kg 12.5kg 16.0kg
Size:
PL310, PL154, &
PL320:
155mm(W) x 170mm(H) x 265mm(D)
PL330:
155mm(W) x 170mm(H) x 300mm(D)
PL330P:
207mm(W) x 170mm(H) x 300mm(D)
PL310QMD, PL310QMT, &
PL320QMD:
350mm(W) x 170mm(H) x 265mm(D)
PL330QMD, &
PL330DP:
350mm(W) x 170mm(H) x 300mm(D)
PL320QMT:
425mm(W) x 170mm(H) x 265mm(D)
PL330QMT, &
PL330TP:
425mm(W) x 170mm(H) x 300mm(D)
Rack Mount Option: 19 inch 4U mount for two PL330P or one PL330DP/TP.
Anti-tamper cover for security.
Electrical Safety:
Designed and manufactured to comply with IEC 348 and
IEC1010-1. Full safety sockets available to special order.
EMC:
Designed and manufactured to comply with EN50081–1
and EN50082–1.
Laboratory Power Supplies
LOGIC OUTPUT (PL330TP) - REMOTE OPERATION
Standard and Programmable models, 35 watts to 240 watts
Output Voltage
Range:
Output Current:
Setting Accuracy:
Output Switch:
4 to 6 Volts in 10mV steps
1 to 7 Amps in approximate 1A steps
Voltage: ±(0.2% + 10mV)
Electronic by interface command (front panel output switch
must be set to ON)
Readback Resolution:Current: 10mA
Readback Accuracy: Current: ±(0.5% of reading + 1 digit )
THURLBY THANDAR INSTRUMENTS
Power Requirements:
Input Voltage:
Input Voltage Range:
Power Consumption:
Voltage: ±(0.2% of reading + 1 digit )
Current: ±(0.5% of reading + 1 digit )
Remote programmable versions in the range feature full control, read back and
status reporting via the GPIB and RS232 interfaces. The GPIB interface conforms to the IEEE 488.1 and 488.2 standards and the RS232 interface is fully
compatible with the Thurlby-Thandar Addressable RS232 Chain (ARC) standard.
8 rear panel DIP switches are used to specify baud rate, bus address and active
interface (GPIB or RS232). Remote/Local operation is selected by a front panel
switch.
LOCAL OPERATION
For a programmable instrument operated in local state, all capabilities and specifications remain unchanged from those of a standard instrument .
REMOTE OPERATION
With the instrument switched to the remote state, all voltage and current adjustment controls become inoperative and commands received over the active interface will be parsed and executed.
Output Voltage
Setting:
Output Current
Setting:
Setting Accuracy:
REMOTE CONTROL INTERFACES - PL-P MODELS
Both interfaces feature full control, readback and status reporting.
RS232:
Variable Baud rate (9600 maximum), 9 pin D-connector (female). Fully compatible with ARC (Addressable RS232
Chain) system.
GPIB:
Conforming with IEEE-488.1 and IEEE-488.2
Address Selection: By rear panel DIP switch.
Remote/Local
Operation:
Selected by front panel switch.
Remote Command Response Time:
Interface:
<15 ms (single command, input buffer empty).
Output Voltage - Up:Time constant typically 2ms, e.g. 10ms to settle within 1%
of a step change, 15ms to settle within 0.1%.
Output Voltage - Down:
Time constant determined by the discharge of the power
supply output capacitor (47uF). Typically <10ms to settle
within 1% for a 10V step change at 50mA load current; typically <200ms to settle within 1% at zero load.
Output Current:
Typically 50ms to settle within 10mA for a 1A change.
Thurlby Thandar Instruments Ltd. operates a policy of continuous development and
reserves the right to alter specifications without prior notice.
Designed and built in the U.K. by:
Thurlby Thandar Instruments Ltd.
Glebe Road, Huntingdon. Cambs. PE18 7DX England
Tel: 01480 412451 Fax: 01480 450409 e-mail: [email protected]
82100-0046
CYAN MAGENTA YELLOW BLACK
Technical Specifications
Technical Specifications continued
MODEL SELECTOR GUIDE
MAIN OUTPUT(S)
Output Range:
MODEL
MAIN OUTPUT(S)
LOGIC
OUTPUT
INTERFACES
GPIB
PL310
0 - 32V at 0 - 1A
PL320
0 - 32V at 0 - 2A
PL154
0 - 15.5V at 0 - 4A
PL330
0 - 32V at 0 - 3A
PL310QMD
2 x 0 - 32V at 0 - 1A
ARC
Output Voltage
Setting:
Output Current
Setting:
Output Mode:
or 0 - 64V at 0 - 1A
or 0 - ±32V at 0 - 1A
2 x 0 - 32V at 0 - 2A
or 0 - 32V at 0 - 4A
or 0 - 64V at 0 - 2A
or 0 - ±32V at 0 - 2A
PL330QMD
2 x 0 - 32V at 0 - 3A
or 0 - 32V at 0 - 6A
or 0 - 64V at 0 - 3A
or 0 - ±32V at 0 - 3A
PL310QMT
2 x 0 - 32V at 0 - 1A
5V at 1.5A
or 0 - 32V at 0 - 2A
or 0 - 64V at 0 - 1A
or 0 - ±32V at 0 - 1A
PL320QMT
2 x 0 - 32V at 0 - 2A
LOGIC OUTPUT - PL310QMT
4 - 6V at
0.1 - 4A
or 0 - 32V at 0 - 4A
Output Voltage:
Max.Output Current:
Output Terminal:
Output Protection:
or 0 - 64V at 0 - 2A
or 0 - ±32V at 0 - 2A
PL330QMT
2 x 0 - 32V at 0 - 3A
Load Regulation:
Line Regulation:
4 - 6V at
0.1 - 7A
or 0 - 32V at 0 - 6A
By single logarithmic control.
The power supply operates in constant current or constant
voltage modes with automatic cross-over. Decimal points
flash to indicate constant current mode.
Isolated, True parallel, Series or Series Tracking via front
panel switches.
Configuration
Selection:
(QMD and QMT only)
Output Switch:
Isolates the output and permits voltage and current limits to
be set up before connecting the load.
Output Terminals:
4mm terminals on 19mm (0.75") spacing.
Output Impedance: Constant Voltage: Typically <5mΩ at 1kHz
Constant Current: Typically 50kΩ with voltage limit at maximum
Output Protection:
Up to maximum output voltage +20 Volts forward; diode
clamped for reverse voltages and up to 3A reverse current.
Load Regulation:
<0.01% of maximum output for 90% load change
Line Regulation:
<0.01% of maximum output for 10% line voltage change
Remote Sense:
Eliminates up to 0.5V drop per lead.
Ripple and Noise:
Typically <1mV rms
Transient Response: <20usec to within 50mV of setting for 90% load change
Temperature
Coefficient:
Typically <100ppm/oC
Meter Type:
Dual 3.75 digit (4095 count) with 12.5mm (0.5") LEDS,
(scale length increased to 8190 on PL330QMD/QMT).
Reading rate 4 per second.
Meter Resolution:
Voltage: 10mV over the entire range
Current: 1mA over the entire range
Meter Accuracy:
Voltage: ±(0.1% of reading + 1 digit)
Current: ±(0.3% of reading + 1 digit)
Current Meter
Nominally 20ms switchable to 2 sec for averaging of rapidly
Damping:
varying loads
or 0 - 32V at 0 - 2A
PL320QMD
0 - 32 Volts nominal; 0 - 15.5V (PL154).
0 - 1.1A nominal (PL310);
0 - 2.1A nominal (PL320);
0 - 3.1A nominal (PL330);
0 - 4A nominal (PL154).
By coarse and fine controls; resolution better than 5mV
across the range.
Fixed 5 V ± 0⋅1 V.
>1⋅5 Amps.
4mm terminals on 19mm (0⋅75") spacing.
Output will withstand up to 16 V forward voltage. Diode
clamped for reverse voltages and up to 3 Amps reverse current.
< 0⋅3% for 50% load change.
< 0⋅1% for 10% line change.
LOGIC OUTPUT - PL320QMT
or 0 - 64V at 0 - 3A
or 0 - ±32V at 0 - 3A
PL330P
0 - 32V at 0 - 3A
Yes
Yes
PL330DP
2 x 0 - 32V at 0 - 3A
Yes
Yes
PL330TP
2 x 0 - 32V at 0 - 3A
Yes
Yes
4 - 6V at 1 - 7A
Models illustrated within this brochure:Front cover - PL330, PL320QMD, PL310QMT
Inside pages - PL330QMT, PL330DP
Output Voltage
Range:
Output Current:
Output Switch:
Output Terminals:
Over-Voltage
Protection:
Output Protection:
4 to 6 Volts
0.1 to 4 Amps.
Isolates the output and permits output voltage to be set before connecting the load.
4mm terminals on 19mm (0.75") spacing.
Above 7 Volts
Clamped by the over-voltage protection circuit for forward
voltages over 7 Volts and up to 1 Amp forward current. Diode
clamped for reverse voltages and up to 3 Amps reverse current.
Load Regulation:
<0.01% of maximum output for 90% load change
Line Regulation:
<0.01% of maximum output for 10% line voltage change
Remote Sense:
Eliminates up to 0.5V drop per lead.
Ripple and Noise:
Typically <1mV rms
Transient Response: <20usec to within 50mV of setting for 90% load change
Temperature
Coefficient:
Typically <100ppm/oC
Voltage Setting
Accuracy:
± 0.1V
LOGIC OUTPUT - PL330QMT & PL330TP
Output Voltage
Range:
Output Current:
Output Switch:
Output Terminals:
Over-Voltage
Protection:
Output Protection:
4 to 6 Volts
0.1 to 7 Amps.
Isolates the output and permits output voltage to be set before connecting the load.
4mm terminals on 19mm (0.75") spacing.
Above 7 Volts
Clamped by the over-voltage protection circuit for forward
voltages over 7 Volts and up to 1 Amp forward current. Diode clamped for reverse voltages and up to 3 Amps reverse
current.
Load Regulation:
<0.01% of maximum output for 90% load change
Line Regulation:
<0.01% of maximum output for 10% line voltage change
Remote Sense:
Eliminates up to 0.5V drop per lead.
Ripple and Noise:
Typically <1mV rms
Transient Response: <20usec to within 50mV of setting for 90% load change
Temperature
Coefficient:
Typically <100ppm/oC
Meter Type:
3.75 digit (4095 count) with 12.5mm (0.5") LEDs. Reading
rate 4 per second.
Meter Resolution:
Voltage: 10mV
Current: 10mA
Meter Accuracy:
PL-P MODELS, ADDITIONAL SPECIFICATIONS
MAIN OUTPUT(S) - REMOTE OPERATION
12 bit resolution (10mV steps)
12 bit resolution (1mA steps)
Voltage: ±(0.1% + 10mV)
Current: ±(0.2% + 2mA)
Output Switch:
Electronic by interface command (front panel output
switches must be set to ON)
Readback Resolution:Voltage: 10mV over the entire range
Current: 1mA over the entire range
Readback Accuracy: Voltage: ±(0.1% of reading + 1 digit)
Current: ±(0.3% of reading + 1 digit)
Current Meter
Nominally 20ms switchable to 2 sec and back by remote
Damping:
commands
PL & PL-P Series
GENERAL
Environmental:
Weight:
Internally set for 110, 120, 220 or 240VAC 50/60Hz
±10% of voltage setting
Single Dual
Triple
30V/1A
75VA 150VA 300VA
15V/4A & 30V/2A150VA 300VA 450VA
30V/3A
225VA 450VA 600VA
Operating Range: 5oC to 40oC, 20% to 80% RH
Storage Range: -20oC to +60oC
Single Dual
Triple
30V/1A PL
4.0kg 8.0kg 11.5kg
15V/4A PL
5.0kg
30V/2A PL
5.0kg 9.5kg 13.5kg
30V/3A PL
6.0kg 12.0kg 15.5kg
30V/3A PL-P
6.5kg 12.5kg 16.0kg
Size:
PL310, PL154, &
PL320:
155mm(W) x 170mm(H) x 265mm(D)
PL330:
155mm(W) x 170mm(H) x 300mm(D)
PL330P:
207mm(W) x 170mm(H) x 300mm(D)
PL310QMD, PL310QMT, &
PL320QMD:
350mm(W) x 170mm(H) x 265mm(D)
PL330QMD, &
PL330DP:
350mm(W) x 170mm(H) x 300mm(D)
PL320QMT:
425mm(W) x 170mm(H) x 265mm(D)
PL330QMT, &
PL330TP:
425mm(W) x 170mm(H) x 300mm(D)
Rack Mount Option: 19 inch 4U mount for two PL330P or one PL330DP/TP.
Anti-tamper cover for security.
Electrical Safety:
Designed and manufactured to comply with IEC 348 and
IEC1010-1. Full safety sockets available to special order.
EMC:
Designed and manufactured to comply with EN50081–1
and EN50082–1.
Laboratory Power Supplies
LOGIC OUTPUT (PL330TP) - REMOTE OPERATION
Standard and Programmable models, 35 watts to 240 watts
Output Voltage
Range:
Output Current:
Setting Accuracy:
Output Switch:
4 to 6 Volts in 10mV steps
1 to 7 Amps in approximate 1A steps
Voltage: ±(0.2% + 10mV)
Electronic by interface command (front panel output switch
must be set to ON)
Readback Resolution:Current: 10mA
Readback Accuracy: Current: ±(0.5% of reading + 1 digit )
THURLBY THANDAR INSTRUMENTS
Power Requirements:
Input Voltage:
Input Voltage Range:
Power Consumption:
Voltage: ±(0.2% of reading + 1 digit )
Current: ±(0.5% of reading + 1 digit )
Remote programmable versions in the range feature full control, read back and
status reporting via the GPIB and RS232 interfaces. The GPIB interface conforms to the IEEE 488.1 and 488.2 standards and the RS232 interface is fully
compatible with the Thurlby-Thandar Addressable RS232 Chain (ARC) standard.
8 rear panel DIP switches are used to specify baud rate, bus address and active
interface (GPIB or RS232). Remote/Local operation is selected by a front panel
switch.
LOCAL OPERATION
For a programmable instrument operated in local state, all capabilities and specifications remain unchanged from those of a standard instrument .
REMOTE OPERATION
With the instrument switched to the remote state, all voltage and current adjustment controls become inoperative and commands received over the active interface will be parsed and executed.
Output Voltage
Setting:
Output Current
Setting:
Setting Accuracy:
REMOTE CONTROL INTERFACES - PL-P MODELS
Both interfaces feature full control, readback and status reporting.
RS232:
Variable Baud rate (9600 maximum), 9 pin D-connector (female). Fully compatible with ARC (Addressable RS232
Chain) system.
GPIB:
Conforming with IEEE-488.1 and IEEE-488.2
Address Selection: By rear panel DIP switch.
Remote/Local
Operation:
Selected by front panel switch.
Remote Command Response Time:
Interface:
<15 ms (single command, input buffer empty).
Output Voltage - Up:Time constant typically 2ms, e.g. 10ms to settle within 1%
of a step change, 15ms to settle within 0.1%.
Output Voltage - Down:
Time constant determined by the discharge of the power
supply output capacitor (47uF). Typically <10ms to settle
within 1% for a 10V step change at 50mA load current; typically <200ms to settle within 1% at zero load.
Output Current:
Typically 50ms to settle within 10mA for a 1A change.
Thurlby Thandar Instruments Ltd. operates a policy of continuous development and
reserves the right to alter specifications without prior notice.
Designed and built in the U.K. by:
Thurlby Thandar Instruments Ltd.
Glebe Road, Huntingdon. Cambs. PE18 7DX England
Tel: 01480 412451 Fax: 01480 450409 e-mail: [email protected]
82100-0046